DILUCION
LEONARDO A. VELA ORIHUELA.

Leonardo Vela Orihuela. HNGAI

Leonardo Vela Orihuela. HNGAI

30/03/2013

Leonardo Vela Orihuela. HNGAI

DILUCION
[VAPORIZADOR] [INSPIRADA]

[2%] [1.2%] Donde est el otro 0.8%?

Leonardo Vela Orihuela. HNGAI

30/03/2013

Leonardo Vela Orihuela. HNGAI

Leonardo Vela Orihuela. HNGAI

30/03/2013

Leonardo Vela Orihuela. HNGAI

DILUCION
VT = FGF + VR VT = FGF + VR VT = FGF + VR

VT = Volumen Tidal. FGF = Flujo de Gas Fresco. VR = Volumen Recirculante.

Leonardo Vela Orihuela. HNGAI

30/03/2013

DILUCION.
FGF

> V.M. Circuito Lineal, no reinhalacin. [FGF] = [Inspirado] FGF < V.M. Circuito circular, hay reinhalacin. [Inspirada] = [FGF] + [Gas reinhalado] [Inspirada] < [FGF]
Leonardo Vela Orihuela. HNGAI

VAPORIZADOR FRACCION INSPIRADA [2%] [2% ] VAPORIZADOR FRACCION INSPIRADA [2%] [1.8% ] VAPORIZADOR FRACCION INSPIRADA [2%] [1.2% ]

Leonardo Vela Orihuela. HNGAI

30/03/2013

MONITORIZACION.
FRACCION INSPIRADA Y ESPIRADA:

Oxgeno. Oxido Nitroso. Agente anestsico.

Leonardo Vela Orihuela. HNGAI

DILUCION SE INCREMENTA:
Ventilador

compensa el Volumen

Tidal. Uso de analizadores de gases. Uso de cal sodada seca.

Leonardo Vela Orihuela. HNGAI

30/03/2013

[Vaporizador]
Cantidad de anestesico entregada al circuito anestesico.

[2%] y 3 litros/min.60 ml. de vapor anestesico. 2. [2%] y 0.5 litros/min.10 ml. de vapor anestesico.
1.
Leonardo Vela Orihuela. HNGAI

[Inspirada] Cantidad de anestesico entragada por el circuito al paciente. [2%] y 600 ml /min.12 ml. de vapor anestesico.
Leonardo Vela Orihuela. HNGAI