Anda di halaman 1dari 128

SERVICE MANUAL

Electrocardiograph
FX-2111
4R2043

FUKUDA DEIMSHI CO., LTD.

Copyright 1997 by Fukuda Denshi Co., Ltd.


No part of this document may be copied
or transmitted in any form without

the prior written permission of Fukuda


Denshi Co., Ltd.

Manufactured in Japan

Foreword

We list up here the warning marks used in Fukuda operation and


service manuals.

When you sen/ice the FX-2111, read this service manual thoroughly and pay attention especially to instructions bearing the following marks:

Warning Marks
Warning marks used in operation and service manuals and labelled on the instrument have the following meanings: Read them carefully to understand the meanings and make sure of the significance of each particular.

A A
A

This mark is used to indicate the direct hazards

Danger which may lead to the death or serious injury of the


person, may wholly damage the instrument, or may
cause fire hazard, unless the instructions written
there are followed.

This mark is used to indicate the indirect potential

Warning hazards which may lead to the death or serious injury


of the person, may wholly damage the instrument or may cause fire hazard, unless the instructions
written there are followed.

This mark is used to indicate the possible hazards

Caution which may lead to a mild or medium injury of the


person, may partially damage the instrument or may erase data from the computer.

NOTE

"NOTE" is not warning instructions but offers information to prevent the person from doing erroneous sen/icing.

Other Marks

Notice to indicate general unspecific prohibited


matters.

Notice to indicate general unspecific caution, warning or hazard.

This service manual describes technical information on FX-2111 to

aid the service engineer in troubleshooting.


The manual is intended to be used by service engineers of Fukuda representatives and authorized technical staff concerned with medical electronic equipment. Description includes repairing and assembling methods of each component unit of FX-2111. For parts lists and diagrams, refer to
the Part II of the service manual.

The service manual consists of the following nine chapters:


1
.

General Description The outline of FX-2111, specifications, controls and indicators are
described.

Circuit Description Circuit configuration and functions are explained.


Troubleshooting
Troubles vs. causes and countermeasures are described.

4 Maintenance
.

Procedures to replace the power fuse, ROM and battery and to perform
self-test are described. 5
.

Periodical Inspection Inspection procedures to prevent troubles and ensure safe and complete operation of the instrument are described.

6
7

Circuit Diagrams Assembly Diagrams

8 Electrical Parts Lists


.

Structural Diagrams

A Caution

. Never remodel Fukuda medical electronic equipment


. The service manual

is intended for the service

engineers of Fukuda representatives and the technical staff concerned with medical electronic equipment. Servicing, reassembling, and adjustment shall be performed by authorized service engineers. . Prepare proper facilities and tools when servicing . Be sure to follow the instructions of operation manual when operating the instrument. For operating precautions, refer to the operation manual.
.

ii

A Servicing Precautions
A Cautions listed below are the instructions of prohibit danger, warn,

ing, and caution described in this service manual. When taking the procedure bearing the following mark, read the description thoroughly,
then start the task.

A Caution

(page 3-2)

When checking the power supply and related circuits for troubleshooting, take sufficient care to avoid a short circuit.

A Caution

(page 3-3)

When checking power fuses, be sure to turn the power off and disconnect the power plug from the wall outlet beforehand.

A Caution

(page 3-8)

The inserting part of key connector is made of carbon. Avoid frequent repeated disconnection and connection.

A Caution

(page 4-1)

When replacing power fuses, be sure to turn the power off and disconnect the power plug from the wall outlet beforehand.

A Caution

(page 4-1)

When replacing the ROM, be sure to turn the power off. Also, take care to install the ROM in correct position.

A Caution

(page 4-2)

When replacing the battery, be sure to turn the power off and disconnect the power plug from the wall outlet beforehand.

iii

A Caution
Disassembling/Reassembling Precautions

(page 4-6)

Be sure to disconnect the power cord and make sure the instrument is turned off before disassembling or reassembling. . Do not remove the battery before disconnect any PC board
.

Take care that repeated disconnection of the key panel and sensor board may result in poor contact.
Use proper tools to loosen screws.

. When reassembling make sure that all screws are securely tightened and all connectors are completely inserted.
,

PC Board Handling Precautions


. PC boards are equipped with highly sensitive components to static electricity. . PC boards are highly sensitive electronic devices Put removed PC boards in a proper protective bag or take appropriate measures to protect them.
.

Handle PC boards carefully. A shock to them may damage the components. Never insert a connector to the powered PC board nor remove the powered PC board.

A Caution

(page 5-2)

If you find a value which exceeds the allowable level, be sure to let the user avoid using the FX-2111. If the user operates the FX-2111 as it is, he/she may receive an hazardous accident.

Equipment Classification
The FCP-2155 is classified into the following equipment:
1
.

Protection against electrical shock


Class I

Type against electrical shock Applied part: Type CF

Degree of protection against harmful water invasion Other equipment Degree of safety in using under air-inflammable anesthetic gases or oxygen/nitrous oxide-anesthetic gases Equipment used under an environment containing no inflammable anesthetic gases or no inflammable cleaning agent.
Running mode Continuous running mode
iv

CHAPTER

1
1-1

General Information
1 Outline of the FX-2111
.

Specifications 2 1 Electrocardiograph Section


.

1-1 1-1
1-2
1-2 1-3

2 2 General
.

23
.

Environmental Conditions

3 Controls and Indicators


.

1 Top Panel 3 2 Side Panels, Left and Right


.

1-4 1-4
1-4

33
.

Bottom Panel

1 Outline of the FX-2111


.

The FX-2111 is the easy-to-use single-channel electrocardiograph featuring a simple operation panel. The compact yet high-performance design makes it suitable not only for use in the consultation room but also for-carrying in a visit to the hospital ward or patient s home and examination at an emergency site.
'

It has a high-density thermal dot printer incorporated to provide clear ECG recording. The FX-2111 can operate on either AC line or rechargeable Ni-MH battery.

2
2

Specifications
Input circuit:
Leads:

1 Electrocardiograph Section
Floating from the ground
Standard and Cabrera 12 leads

Input impedance: Input circuit current: Calibration voltage:

20MQ min. (referred to 10Hz) 5x10"8 max.


1mV within 5% method

Common mode rejection: 10mm or less per IEC 62DC06 test


Polarization voltage:
Time constant: Sensitivities:
300mV min.

3 2 sec min.
.

1/2, 1 and 2 cm/mV

Frequency response:
AC filter: Muscle filter:

0 05
.

to 150Hz (within -3dB)

50/60Hz, -20dB or lower

DC input: Display: Recording system: Paper speeds: Chart papersRoll paper: Z-fold paper:
AID conversion:

35 to 45Hz, -3dB (-6dB/oct) 10mm/0.5V, unbalanced, 100kfi min. Liquid-crystal display, 20 chars, x 2 lines (character: 5x7 dots) Thermal dot printer, 8 dots/mm
25 and 50 mm/sec within 3%

63mm or 50mm wide x 30m long 63mm or 50mm wide x 20m long,
75mm/fold
12 bits 1ms

Sampling rate:

1-1

2 2 General
.

Safety:

I EC 601-1 Class I

.Type CF . Internally powered equipment Type CF (IEC 601-1) Power requirementsAC operation: DC operation:
115V AC, 50/60H2; 19VA max. 230V AC, 50/60Hz; 19VA max. 9 6V, 7W max. (rechargeable Ni-MH battery) Continuous operation: approx. 120 min at 20oC (according to IEC 62D Testing Method) Charging time: within 3 hours 26 (W) x 18.2(D) x 6.3(H) cm Approx. 1.7kg (excluding battery)
.

Dimensions:

Weight:
.

2 3 Environmental Conditions

OperatingTemperature: 10 to 40oC Humidity: 30 to 80%RH (no dew condensation) Atmospheric pressure: 70 to 106kPa Transportation & StorageTemperature: -10 to 40oC Humidity: 10 to 95%RH (no dew condensation) Atmospheric pressure: 70 to 106kPa

1-2

3 Controls and Indicators


.

Top Panel

X Liquid-crystal Display

31
@ Paper Magazine

(D Operation Panel

Left Panel

Right Panel

Potential Equalization
Terminal
0

o
0 0

Power Connector
D

F3

Main Power Switch

@ DC Input Terminal

Lead Connector

Bottom Panel

o
{
*

I
Fuse Holders
0

i<

"

iii
)
~ ~

Rechargeable feattery Room

1-3

1 Top Panel
Indicates current status such as recording mode, lead, sensitivity, heart rate, or program setting.
Turns the FX-2111 on in DC operation. Turns the FX-2111 off in DC operation. Pressing this key during AC operation lets the FX-2111 enter the charge mode. Select a mode in the following order when pressed during cessation of recording: |-> Automatic recording - Manual recording -> Programming -| Holding this key during automatic recording lets the FX-2111 record the current lead continuously until detached. Holding this key during manual recording lets the FX-2111 record
an event mark until detached.

Display:
Operation Panel:
ON OFF

Mode

Sensitivity Lead Select <


Reset

1mV Start/Stop

Selects a recording sensitivity. Select a lead for recording. In programming mode, these keys allow you to select a setting value. Resets the measuring circuit while held during recording. In programming mode, this key selects a setting parameter. Applies a 1mV calibration waveform when pressed during recording. In programming mode, this key selects a setting parameter. Starts the FX-2111 recording the ECG waveform. Another press stops it from recording.
Accommodates a chart paper.

Paper Magazine:
3
.

2 Side Panels, Left and Right


Inputs external DC signals with a sensitivity of 10mm/0.5V.
Connects to the lead cable.

DC Input:
Lead Connector Power Connector Main Power Switch:

Connects to the power cable. Turns AC power on/off. Makes the FX-2111 equipotential to another instrument used in combination with it. To that effect, connect both instruments to a

@ Potential Equalization
Terminal:

common grounding conductor using an optional grounding wire.


3 3 Bottom Panel
.

Battery Room:
Fuse Holders:

Houses the rechargeable Ni-MH battery.


Have power fuses inserted.

1-4

CHAPTER

2
2-1

Circuit Description
1 2
.

Introduction

Isolated Input Circuit 2 1 Buffer Amplifier and RF Driver


.

2-2 2-2
2-3

2 Lead Network and Lead Selector

3 Preamplifier and 1mV Generator

2-3
2-4 2-5 2-5
2-6

2
2

4 R-wave Detector and Overinput Detector 5 Amplifier Control 2 6 Signal Isolator


. . .

27
.

Power Isolator

Middle Amplifier and AID Converter 1 Middle Amplifier and DC Input


.

2-6 2-6
2-7 2-8 2-8

3 2 AID Converter
.

4 Motor Control 5 Sensor Circuit


. .

1 Detection of Magazine Open Condition 5 2 Detection of Paper End and Paper Marks
5
.
.

2-9 2-9
2-9 2-9 2-9

6 CPU Circuit
.

61
.

Reset Circuit

6 2 CPU, ROM and RAM


.

63
.

Gate Array

2-9
2-11 2-11

LCD Control Circuit 8 Thermal Print Head Control Circuit


7
. .

Memory Backup Circuit 10. Power Supply and Charging Circuit


9
.

2-12 2-13
2-13

10.1 Introduction

10.2 Rectifier/Smoothing Unit and ON/OFF Control


10.3 + 10V and +5Vd Power Generators 10.4 5Va Power Generator 10.5 +24V Power Generator

2-13
2-14 2-15 2-15

10.6 Charging Circuit

2-16

1 Introduction
.

The FX-2111 is composed of the following circuits:


(1) Main Board PCB-6409 -Isolated input circuit -Middle amplifier and AID converter circuit
-Motor control circuit
-Sensor circuit -Reset circuit

-CPU circuit (CPU ROM, RAM, gate array)


,

-LCD control circuit

-Thermal print head control circuit -Memory backup circuit -Power supply and charging circuit (2) Sensor Board PCB-6239 (3) AC Inlet Board PCB-6410

ECG signals input through the lead connector are amplified by the isolated input circuit and the middle amplifier circuit, then converted from analog to digital signals by the AID converter circuit. The digitized signals are digitally filtered, if the filter is set to ON, then sent through the gate array to the thermal print head control so that they are recorded as ECG waveform by the thermal print head.
During these processes, the heart rate detected by a hardware technique and operation status such as filter ON/OFF setting are displayed
on the LCD.

The power supply provides circuits with necessary powers. The charging circuit, if activated, charges the Ni-MH battery in AC operation or transfers the power from the Ni-MH battery to the power supply. The state of battery is indicated on the LCD.
PCB-6409
-

:pcb-64io
AC Inlet

i5V
I0V

"

re

Power Supply
AC Power

Isolated Input Circuit

If

-24V

VDO +2 5V
.

ZEE
'

/Charging;-;
NiMH

'

DC

Input

AyiCircuil ;-*

Battery

DC Motor

Motor

Control

5 M mt
J Gate Array

:'
,
CPU Circuit

Thermal ]
Print
*
.

Thermal

Key Panel

Print it
Head [ Control

Head

LCD

LCD Control
Sensor Board

1
PCB.

Backup
Circuit

6239

Overall Block Diagram


2-1

Isolated Input Circuit


The isolated input circuit is composed of the following:
(1) (2) (3) (4) (5) (6) (7) Buffer amplifier and RF driver Lead network and lead selector Preamplifier and 1mV generator R-wave detector and overinput detector Amplifier control Signal isolator Power isolator

L
F

I
-

ECG

2
"

CI
I

D
5

c
D
.

5 5 t a5
-

Preamplifier

"

> Signal
Isolator

1
5

I 2
5

I
"

C6

1mV

RESET

T
RF Drive <-

Amplifier
Control

Signal
Isolator

RF

Overinput/
R-Wave Detector Circuit

(Photo

Coupler)

I
i i

.e-

Power

Isolator <-

Block Diagram of Isolated Input Circuit

1 Buffer Amplifier and RF Driver

oac- Hit icr

I
9 ISO*
0

5
Since the buffer amplifier should input signals with high impedance and output them with low impedance, it is configured as an impedance conversion circuit using an operational amplifier. The circuit shown above is individually applied to all leads except for RF (right foot). Also, a limiter using a dual transistor is mounted for protection against overinput.
The RF driver feeds back the composite signal of limb leads to the

right foot for improved common mode rejection ratio.


2-2

2 2 Lead Network and Lead Selector


.

OK

211 OK

*a?

z 10

Ox

> 304

OK

20 1

0*

Lead Select Signals


Leads It:?
? ior

BCD

ft:

STD
II

0
0 0

0
0

0
0 0

10
Ok

4;

10 0

III
'

110 10 0
0
0

200

0-<

J
m

Z0 OK

I
Tan

Vr

10

aVL

10

El

aVF
VI V2

110
0 0
0

10*

1110 1
1 1 10

V3
l OX t

V4

10

UNS2 L]

V5 V6

110 0 0

11

The lead network is formed with resistors and lead selector, with multi-

plexors IC36, IC38, IC41 and IC4. Input signals to multiplexors are selected
from RA, LA, LF, and CI to C6 according to four signals of A, B, C, and D then synthesized to produce each ECG lead (see table above).
,

3 Preamplifier and 1mV Generator


C-1
o.aou

fcTzOOHK

r-II
"V X
.

ci
10m JOK

cb>

rc

in*
.200 CCC

MOK <

cm. J
o

*.

4= " [ F
? i

.on

-*-

J frt5
i

|Slr4ii

. mi

(C J7I

OIL

1SS
1*3 117 1

1 . IK
t

3C on*

r mioi

5:

A
41J

icia

LIt004C$l-i
.

rc

1 Preamplifier Signals output from the lead selector (multiplexors) are amplified 8 times by a differential amplifier, then 25 times by a noninverting amplifier, thereby letting the isolated input circuit amplify the input signals by 200 times in total. Signals output at the first stage are sent to the overinput
23
. .

detector.
2-3

The 3.5Mfi resistor and 1|jF capacitor set the time constant at 3.5
seconds.

2 3 2 1mV Generator
. .

The 1mV generator divides the high-precision voltage reference output of 1.235V by a high-precision resistor, thereby applying a 8mV voltage to the preamplifier when the 1mV application signal is at a low level.
2
.

4 R-wave Detector and Overinput Detector


era Q.Qit Li

C2

[C4 1C

J
*

*
.

in

>
10QN

4*2

U*f 111

I 4M
0 23
.

lOOH

@
c 111

nrvr-!>

JOCK

1J ..

CP
ft I
r-

a*]

UMf I LI

C 42 0 001 J
.

C12 J

"I

20OK
re

2 41
.

R-wave Detector

RA and LF signals of lead network are synthesized to II lead. The Rwave detector detects R wave by picking up R-wave component from the II lead signal through a band-pass filter, then sending it to a comparator.
24
. .

2 Overinput Detector The signal output at the first stage of preamplifier is sent to a com-

parator and if the signal exceeds 360mV, the output of the comparator
becomes inverted.

Signals output from the R-wave detector and overinput detector are adjusted in the pulse width by the multivibrator IC40 (TC4538), then input into the photo coupler.
+2 9V
.

+0 31V
.

1T63

ov

OV

0 31V
.
-

2 9V
.

TP19

* 170S

TP20

ITOins

R-wave Detection Wavelorm

Overinput Detection Waveform


2-4

5 Amplifier Control
*

"

C S . IK

<5id*

of3

2 n

5
IT

Si a

. 2 17 IK HJU
.

1
XL

lO.Ot
-

v 0

CO

This circuit picks up a reset, 1mV,or lead select signal from the serially
transferred amplifier control signals.
!<-% 1ms 1 KH 2
AMP

;
I

C L K

AM P

DO
i i

DATA

D7

D6

D5

D4

D3

D2

Dl

DO

D7

D6

D5

D4

D3

D2

DL

DO

D7

DO

OLD DATA <-

i i

> NEW DATA

8ms

Data wfitlen (rom CPU are latched.

Response K time: 32ms + a max.


16ms + a mm.
_

0 a 0 8ms

6 Signal Isolator
CT*

IC21 rjJiSJLp-Iilt|
<fl rCLK H

o.c 0+?
.

am

U U
C 73 lM IK

-cTTirClK 1

ICJW y5)*3L. l JLj-l1|


mil Ik

O
frwrr

--W

c
-;

>111 * r i

12
-

aiir

LTT-

CE3d>

T en 1

p y j- v -- -tret? >
0 01
.

___

41
ISI3)

O.Ol

o.ot

BIO-

rLji_.
a) i

Cr-

sr:

The signal isolator modulates input analog ECG signals with FETs
Q31 and Q32 and pulses TRF1 + and TRF1 -, then transfers the modulated

signals through the T1 isolation transformer (1:1). FETs Q22 and Q23
demodulate the signals. A low-pass filter formed with capacitors and resistors in two stages eliminates noise in the demodulated signals.

Digital signals (AMPCLK, AMPDATA and HR) are transferred using a photo
coupler.
2-5

2 7 Power Isolator
.

-020
<ianz>

11

ISV

*fsl\v

iSiiQl

'2

<3BJ-pW-i
s

f
'

JC 10 *
i -r-r-

30
TVrr-J ..7,11V

1
ilO 1
11

JH

5p

r
*a
.

<2irD--f-1 J-

ISSIOJ

lisjoj 0 . 1

T
200xhi

C 10

The power is transferred using 10V pulse at 100kHz. At the isolated side, a 3-terminal regulator supplies 5V.

3
3

Middle Amplifier and AID Converter


C4 T

1 Middle Amplifier and DC Input


i
_

:v.\]

at
i

3;
*

1 10

.0.2
1

imrraT>

LA

il-Lt
irrw >
"

21

10. ?

...

ECG signals demodulated at the isolated signal transmission section are output to the middle amplifier just after passing through a low-pass filter formed with capacitors and resistors in two stages. The middle amplifier amplifies the input signals 1.25 times to a total gain of 250 times and sends them to the AID converter. Signals output from the AID converter are digitally filtered with a software technique. The isolated input circuit and the middle amplifier are designed so as to provide a general hardware frequency response of 150Hz (-3dB).
The DC input amplifies input signals 0.5 times, then sends the signals to the AID converter through a buffer amplifier.

2-6

3 2 AID Converter
.

...

icu.

'

it:?
DC J .
*

Li

I -3

100 T * 0.022
"
I

[l 01 l
t<S302

'

Tin

100
1 C*1

Tl
0 O22
.

4 JQ
t 1

1C4l4
I|l 010* ittlQI

HP
L
vvv-r

100

1300f

,1

0110

El00"
w-

-<}
a it

The A/D converter is a single slope type. It is composed of a triangular wave generator, sample & hold circuit, and comparator. The comparator compares input signals with 3.2\/ triangle wave generated at every 1ms. The signals are then converted from the voltage to a pulse width and transferred to the gate array. In the gate array, the pulse width is counted, then converted into a digital value.
<

520ps

5k

480,1s

Si

A/D CLK

--<?

96ps

>

;<

96)13

S/H

(TP35)

Bps
+3 2V
.

Triangular (TP7)

OV

3 2V
.

Start ol counting
A/D Conversion Timing

Besides ECG signals, the AID converter digitizes DC input signals, battery voltage, thermistor signals of thermal print head, and the reference voltage of 2.5V for calibration of A/D conversion.

2-7

4 Motor Control
.

</00

. LJJ s nn? > @


11 -

LI

lOQ-

T
Li

(IT

C*

VCOIm

- -

C3

Li
11%
_

Li
11

Li
10*
_

* t J*

itsio:

. T jiv ;

\ Ten
14 1

I C 1

The motor to drive the recorder is the DC motor which has a photo sensor built in for detection of motor speed. The PLL control IC25 compares a signal detected by the photo sensor with the reference frequency in phase, thereby providing a motor control signal at pin 13 of IC25. The motor control signal goes to the integrator circuit formed with R140, R142, and C64 and is made by a proper motor drive voltage of the
PWM control IC24.
25mm/s
IC25 (pin 14)
SIN
-

(256Hz)

SOmra/s (5L2Hz)
r i
~

r~i

-+5V
ov

+5V

IC25 (pin 3)
CIN

.OV

.+5V

IC25 (pin 13)


P2

. +3 IV
.

\
Delayed
.

K
Early

+2 2V
.

\
Delayed

/
Delayed Delayed
Early

OV

5 Sensor Circuit
The sensor circuit is provided to check for a magazine opening condition and paper end as well as detecting paper marks.
1*4 0

fi2E2>

T
t la
7

1
lit

:
>

il

3 i &
<1 llflOl

if*
-

4 n*

re it

t*ii

-H> -

2-8

51
.

Detection of Magazine Open Condition

When the paper magazine is open, the microswitch SPVC2-1 is turned off to send an interrupt signal to the gate array. Magazine opening is detected with the fall signal and magazie closed is detected with the rise signal.
52
.

Detection of Paper End and Paper Marks

A reflection type photo interrupter acknowledges a black paper mark or paper end status (no paper remaining) if receiving no reflection from the location of paper. It discriminates between paper end status and paper mark based on a duration of the low level starting from a fall signal.

6 CPU Circuit
.

The CPU circuit is composed of a reset circuit, CPU, ROM, RAM, gate array, and backup power circuit.
61
.

Reset Circuit
VOO

1 14

10K

ION

vsa
ViB

vACSEt
*

J
-

OUIC
Cf
C%3
-

QhO

12 !2M

VSC

h
0

k cj
L'ltV

Q-i

M O T
.

Using NJM2103, the reset circuit generates a reset pulse as well as monitoring the battery voltage. It outputs the reset pulse when Vdd falls to lower than 4.27V or when it receives a signal from WDT.
As for the battery voltage, the reset circuit monitors the power voltage divided by resistor and when the voltage becomes lower than 7.5V referred to the battery terminal, it outputs a pulse to the LOWS terminal to turn the power off. To protect the output pulse against power fluctuation noise, a filter formed with a capacitor and resistor in one stage and a Schmidt trigger inverter are provided.
6 2 CPU, ROM and RAM
.

The CPU is HD63B09E. It controls the overall circuit of the instrument

through the gate array. The ROM is 64K-byte AM27C512-150DC and the RAM is 8K-byte SRM2264LM12. By changing the jumper connection, the ROM can be replaced with a 128K-byte ROM and the RAM, with a 32Kbyte RAM (with a data holding current of lower than 2\iA).
6
.

3 Gate Array

The gate array is FD88007-AC with a clock frequency of 16MHz. Its control signals are as follows:
2-9

631
.

2MHz E and Q Clocks for CPU


'

<

-5 00 n s-->

500ns

V,

$->, 125ns

63
.

2 Interrupt Signals

There are three types of interrupt signals-NMI, FIRQ and IRQ. All these signals, active at low level, are output to the CPU. -NMI Output at every 1ms. -FIRQ Output at every 10ms. -IRQ Output when any one of MAGAZIN, MARK, HR, LOWB, CHG IRQ and AC/DC signals is input.
,

!< NM I

S>

Ims

j* 500jis

1 0 0 Hz
1 Oras

FIRQ
. CLEAR

MARK

RUN

PAPER
END

a = 390ms (9.75mm) with a paper speed of 25mm/sec 190ms (9.75mm) with a paper speed of 50mm/sec NOTE: ACDC, CHG IRQ, and MAGAZIN DOWN signals are detected at
their rise and others are at their fall. 63
.

3 Watchdog Timer (WDT)


. '

10ms

<

1 0 0Hz

WDT CLEAR
70ras .
,

a i

WDT

a = Response time of IC
2-10

634
. .

Chip Select Signals

Chip select signals are all active at low level.


WE: Write enable signal RD: Read enable signal
E

ROM ] RAM [ KEY J


WE I

1 1
'

| 1
1 ' '

RD J

NOTE: LCD signal differs from the above since the memory is ready. (See
LCD control circuit.)

7 LCD Control Circuit


.

Since the LCD (NDM202A00) has a controller built in, it is interfaced directly with the CPU. However, E and Q clocks make the bus timing as shown below.
E

LCD

Data Bus

Address Bus

CM ZX

$3040 ~ $30<tF

> X

8 Thermal Print Head Control Circuit


.

A waveform is traced at every 1ms and alphanumerics are at every 6ms with a paper speed of 25mm/sec or every 3ms with a paper speed of 50mm/sec. The RAM for the thermal print head control uses 4K-byte for alphanumeric data. This circuit controls dot heating temperatures based on previously printed data to provide a proper printing condition.

When the paper is not driven or when the magazine is open or there remains no paper in the magazine, the circuit generates TOFF signal to shut off power supply to the thermal print head at the FET Q24, thereby protecting the thermal print head.
There are four control signals of CLOCK, DATAIN, LATCH, and STROBE. The STROBE signal is varied in pulse width according to thermal print
head temperatures.
2-11

1ms

LATCH

e-

us

8,3

STROBE

!<

960vs

(Max)
895,is

CLOCK

9
<

5us
S

(SOOKHz)
DATA

2v s CLOCK
.

Sks

DATA

Memory Backup Circuit


The battery-backed RAM SRM2264LM12 is installed to keep program

contents when the FX-2111 is turned off. The RAM should feature a data holding current of lower than 2nA.

J
O . I

2 k otci 1 . o>

T ."
9

5 "

When the FX-2111 is turned on, Vdd is supplied to the circuit and
turns Q6 and Q7 on to send +5\/ Vdd to the RAM.

When the FX-2111 is turned off, Q6 and Q7 are turned off but the back-

up lithium battery supplies +3V to the RAM to let it keep the stored data.

2-12

10. Power Supply and Charging Circuit


10.1 Introduction
The power supply transforms the AC power by the transformer or receives DC power from the battery, then regulates the power to stabilized voltages required by respective circuits. The FX-2111 is equipped with a charging circuit to permit the user to charge the Ni-MH battery.

AC Power

Transformer

Rectifier

+ 1 0 VN

Regulated
Power

Supplies + I 0 V

DC Power

ON/OFF Control

+ 5 VD

Battery

5 VA

Charging
Circuit

+ 2 4 V

Block Diagram

10.2 Rectifier/Smoothing Unit and ON/OFF Control


10.2.1 AC Operation
The AC power voltage is transformed by the transformer, then rectified by D1 to D4 and smoothed by C136. D1 to D4 are Schottky diodes with low Vpand form a bridge circuit. This rectifier circuit suppresses voltage decrease and features less heat generation.
10.2.2 ON/OFF Control

In battery operation, a press of the ON key switches FET 04 on and electrifies D7, thereby making the base of 05 driven. If the operator detaches the finger from the ON key, 04 is kept on and the power is supplied to the FX-2111.

The battery voltage is detected via D6. If the voltage between battery terminals is lower than 7.5V or so, the gate array generates POWER OFF signal to turn 03 on and cancel the bias voltage of 05, thereby switching 04 off. As the result, the FX-2111 receives no power supply and is turned off.
Similarly, a press of the OFF key cancels the bias voltage of 05, thereby turning the FX-2111 off. In AC operation, 01 is on; therefore the base of 05 cannot be driven and 04 cannot be kept on, thereby making battery operation impossible.
2-13

VDO D 1 -

AC DETECT

rpBHogy

r i

-(sczeo

'

H ' - >fouR

OFT

if"

C t 36

0?

{>-'
OS

tJNS2 1 2

D3FSA
or 3-2P - 20SA

in

gnr Tr

fr T-T r -rrTinr
"

T
T

Shis
2(>

It
-

E32EHZ>

25
tOOK

-OHI7
20OK

"

)
<

azcz-S.iv

12

F
<\
UNS2 1 Z

UMS212

K2LSS301

0-

08 UNS2 12

T P48

{>
2fc 20K

EKZO

ear t

*
-

-> 0 o T I

10.3

+10V and +5Vd Power Generators

10.3.1 10V Power Generator

The dropper type 4-terminal regulator IC22 is used to regulate unstabilized IOVnU) stabilized lOV. R118 and R119 are to set the output voltage.

The 10V power is supplied to the motor and the 5Vd power generator. If the battery voltage becomes lower than 10V in battery operation, IC25 drops the voltage to several ten millivolts.
10.3.2 5Vd Power Generator

C7 is a drop type low-loss 3-terminal regulator. It produces regulated

5V power from 10V. IC7 can operate if an input-output voltage difference


becomes up to 0.5V and thus it can output the stable 5V power if the input voltage fluctuates due to lowered battery voltage.
The 5V power thus produced is supplied to the CPU and other digital
circuits.

IC22 POIQIV11

IPS*

IC7 TP27
-

iovn

->

I
3

U11?

msz

C20
O.Ol

CIS

*
TP38

I
3 119

JO

gno

>

2-14

10.4

5Va Power Generator

Using the unregulated 10Vn power from the AC power supply or the battery, the 3-terminal regulator ICQ produces +5V. IC10 is a CMOS inverting type switching regulator, in which the output voltage is fixed and which
inverts the input +5V and outputs -5V. A capacitor at the output is an organic semiconductor capacitor which features superior characteristics at low temperatures and least leakage current.

5V powers thus produced are supplied to analog circuits such as operational amplifier.
ICS
-

GEC

C I J D Ol
.

.Not hj n ZZ.

VIN

-VOUI

oT*
ir 44*
,

ss

C2

ft

>

10.5

+24V Power Generator

Using the unregulated 10Vn power from the AC power supply or the battery, the +24V power generator, a boosting type DC-DC converter, produces 24V power for driving the thermal print head and charging the
battery.

IC6 is a PWM-controlled switching regulator. R22 and C18 determine the oscillation frequency. The PWM control is performed at approximately 170kHz to drive Q9 at a high speed. While Q9 is on, the current flows to LI which in turn stores the energy. When Q9 is turned off, the stored energy becomes a counter-electromotive force and is stored in C137 via D13. By repeating a series of these operations, the +24V power generator can produce an output voltage larger than the input voltage. R41 and R42 set the output voltage at 24V. R24 checks for an overcurrent and a characteristic is provided to gradually lower the output voltage when the load
current increases to over 0.4A.

The 24V power thus boosted is used to drive the thermal print head. When the thermal print head is printing, a signal from pin 21 of the gate array drives the base of Q15 to switch FET Q24 on so that the -I-24V is supplied to the thermal print head. When the thermal print head is not printing, Q15 and Q24 are off and the thermal print head is not supplied with power.

Pin 16 of IC6 outputs 2.5V, which is used by the AID converter and the thermistor temperature detector of thermal print head.
I
-

UJ

11

;;

i-0-<nm

i 111 * 51 i
= LIL
-

j1UY
_

;v

..

2-15

10.6 Charging Circuit


The charging circuit is equipped with the quick charging IC18 (bq2003), which performs constant current control as a frequency modulation controller for switching regulation of the charging current and monitors the battery temperature and voltage and the charging time to ensure proper charging. It also checks whether or not the Ni-MH battery is connected and enables charging only when the battery is connected. In AC operation, the circuit is supplementarily charging the battery even if the FX-2111 is not placed in charging mode.

10.6.1 Starting the Charging Circuit


When the OFF key is pressed in AC operation, Q11 is turned on to supply +5V for operation of the IC. Since the base of Q12 is driven by turning Q11 on, Q11 is kept on. When +5V is supplied to IC18, the charging circuit automatically starts operating.

10.6.2 Charging Temperature and Voltage Monitoring/Control


Charging is limited by battery temperature and voltage so that it is made in a preset range.
To monitor the battery temperature, a thermistor (with negative temperature coefficient) is installed in the battery pack. The thermistor outputs a voltage signal to pin 6 (TS) of IC18 and charging is enabled if the voltage is within the preset limits in voltage converted from battery temperature. The sensitivity of a battery temperature rising ratio (AT/At) is adjusted by resistances of the thermistor and R59, R61, and R62 on the charging circuit.

The battery voltage is divided to a voltage per cell by R58 and R60 which are connected between battery terminals, then sent to pin 7 (BAT) of IC18. Charging is enabled if the voltage is within the preset limits in voltage per cell. Divided +5V powers of R84 and R85 are used to determine the maximum cell voltage (MCV) of the battery and the maximum voltage (TCO) for the battery temperature to stop charging, respectively. They are output to pin 10 (TCO) and pin 11 (MCU) of IC18. Preset battery temperature limits and cell voltage limits are as follows: . Battery temperature limits Minimum charge enable temperature limit LTF: Approx. -50C (voltage level 2.0V) Maximum charge enable temperature limit HTF: Approx. 60oC (voltage level 1.13V) High temperature to stop charging TCO: Approx. 650C (voltage level 1.01V) . Battery cell voltage limits to enable charging Minimum: Approx. 1.0V
Maximum: Approx. 1.78V

That is, when the charging circuit is turned on, it starts charging the battery if the battery temperatre is within a range of 50C to 60oC and the cell voltage is within a range of 1.0V to 1.78V. If either one of the above

2-16

conditions is not satisfied, it does not start charging the battery but is placed in a standby status until both conditions are satisfied. Once the charging circuit starts charging the battery, the charge complete system described later will be effective. However, if the maximum cell voltage of 1.78V is exceeded before the start of charging, the charging circuit judges the battery abnormal and does not start charging.

10.6.3 Charge Current Control


The power required for charging is supplied from the 24V power generator. The charging current to the battery is detected by the resistor R18 then sent to pin 9 (SNS) of IC18. The output of pin 14 (MOD) of IC18 is switched based on the voltage value of R18 to switch Q17 on/off, thereby making the current constant. The charging current is set at approximately 0 45C (480mA).
.

Q16, D22, and Q10 form a MOS FET driver and level shifter circuit to

heighten the switching frequency.

10.6.4 Charge Complete System


When the battery temperature and cell voltage are within the respective ranges, the charging circuit continues to charge the battery until one of the following five conditions is satisfied:
. - AV detected

. . . .

- AllAt (temperature rising ratio) detected Maximum temperature of 650C to stop charging TCO reached Maximum charging time of 3 hours reached Maximum cell voltage MCV of 1 78V reached
.

Usually, the charging circuit stops charging by detecting -AV. However, if charging is started at high ambient temperatures, the charging circuit may detect the maximum temperature (TCO) to stop charging. Also, as a safety measure against no - AV detection, the maximum charging time of three hours is set by timer and TM1 and TM2 signals at pins 4 and 5 of IC18, respectively, stop the charging circuit from charging if the
time is reached.

......

jLL1Ji -

I
-

s
-

Hit

'

(tl

"

1
,

r s
net

. <1

ix

U l.
-

5
1

I
T..

T
-S

2-17

10.6.5 Other Functions

In AC operation, supplemental charging is made by R66. In AC operation, Q14 is on and the base of Q13 is driven to supply R66 with the 24V power. The current of which the value is determined based on a voltage difference at R66 flows for supplemental charging. However, if the battery is not connected, Q8 is turned on and Q14 is kept on, thereby disabling supplemental charging and at the same time cancelling the bias voltage of Q12. Thus, the 5V power to operate the charging circuit is not held and charging is impossible.
The charging circuit outputs CHG IRQ and CHG STATUS signals. When the circuit is supplied with 5V, the CHG IRQ signal becomes high level to indicate the instrument is placed in charge mode. The CHG STATUS signal is output from pin 13 (CHG) of IC18 to indicate the current charging status. If the battery is not connected, the CHG IRQ signal cannot become higher than 0.7V since D15 is electrified. Thus, the instrument is not placed in charge mode. The CHG signal is output in the following manner: CHG STATUS Output Level High Abnormal power voltage
Continuous
Low

Start of charging
Standby Charging Charged

Continuous (rise)
1 375s 0.225s
.

125ms 20ms

Continuous 125ms 20ms 125ms 20ms

CHG IRQ Signal


CHG swilch ON

Cha/ping Mode

2 0s
.

'

CHG STATUS Signal in ordinary charging

\
CHG STATUS Signal with some pending of charging

Start of Charging Mode

Charolng In Progress

Churging ConipleloiJ

2 0s
.

Rending

Charging in Progress

CHG STATUS Signal with abnormal battery voltage


"

No charging made

CANT CHARGE" message appears aller 5 seconds Irom the start

2-18

CHAPTER

3
3-1

Troubleshooting
1 Introduction
.

1 Internal Power Supplies 1 2 Faulty Power Supplies vs. Troubles Troubleshooting 2 1 All Specified Secondary Power Supplies Are Not Available in AC Operation 2 2 All Specified Secondary Power Supplies Are Not Available in Ni MH Battery Operation 2 3 +24V Is Not Output 2 4 + 10V Is Not Output 2 5 Vdd Is Not Output 2 6 Unisolated 5Va Are Not Output 2 7 Isolated 5V Are Not Output 2 8 Troubles Related to Charging 2 8 1 Charging Mode or "CHARGING" Status Is Not Effected. 2 8 2 Virtually No Charging Is Made Despite "CHARGING COMPLETE" on Display 2 9 LCD Operation Is Abnormal 2 10 No Key Operation Is Possible 2 11 Paper Is Not Driven or Abnormally Driven 2 12 No Mark Detection on Z-fold Paper Is Available 2 13 No Paper End Detection Is Available Or "PAPER END" Alarm Is Constantly Displayed 2 14 Program Contents Change 2 15 Buzzer Does Not Operate 2 16 No Waveform Recording Nor Alphanumeric Printout
1
. .
. .

3-1 3-2 3-3 3-3


3-3 3-4 3-4 3-4 3-5 3-5 3-6

3-6
3-7 3-7 3-8 3-8 3-9
3-9

3-9 3-9
3-10

Is Available

17 18 2 19 2 20
2
.

No DC Input Waveform Recording Is Available Any Waveform of All 12 Leads Is Recorded As Baseline Waveform of Some Specific Lead Is Recorded As Baseline... Checking AID Converter

3-10 3-10 3-11 3-11

1 Introduction
.

11
.

Internal Power Supplies


First, understand internal power supplies and their functions.

10Vn

The IOVn power is output from the secondary side of transformer. Since it is not regulated, it fluctuates depending on the load status and AC input voltage. It is transformed to various regulated voltages to be supplied to respective circuits.
+10V

Inputting 10Vn, the low-loss 4-terminal regulator outputs regulated 10V. However, if the input is lower than 10V, the output voltage will be lower by several ten millivolts than 10V. This voltage is supplied to the motor and the regulator which produces -i-5Vd.

+ 5Yp
Inputting + 10V, the dropper type low-loss regulator outputs regulated -(-5Vd. This voltage is used as the power supply for digital circuits,
such as CPU, of the unisolated section. + 24V

Inputting 10Vn, the switching regulator outputs regulated +24V. A limiter is applied at approximately 0.4A referred to 10V, thereby lowering the output voltage. The +24V voltage is supplied to the thermal print head. In the case of FX-2111N, it is also used as a charging power supply.
+5Va (unisolated) Inputting IOVn, the dropper type 3-terminal regulator outputs regulated -i-5Va.
-

5Va (unisolated)

Inputting +5Va, the inverting type switching regulator produces -5Va. 5Va are used as analog power supplies for DC input, ECG
circuit and AID converter.

5Vf (isolated)

Receiving 10Vn through the isolation transformer which isolates the voltage from other circuits, the 3-terminal regulator outputs regulated 5Vf. The 5Vf are used as power supplies for digital circuits of the
isolated section.

5Vb (isolated)

5Vb are produced from 5Vf through diodes. They are used as power supplies for analog circuits (input amplifiers)
of the isolated section.
3-1

2 Faulty Power Supplies vs. Troubles


.

1 21
.

If 10Vn is not output... The FX-2111 does not operate at all since 5V, 24V and all other powers are not produced.
122
. .

If 10V only is not output... The dropper type low-loss regulator cannot produce +5Vd, thereby disabling the CPU, etc. to operate. The FX-2111 does not apparently operate
at all while 24V and others are available.
12
.

3 If +5Vd only is not output... The CPU and other components are not supplied with power. The FX-2111 does not apparently operate at all while powers other than +5Vd
.

are available.
12
. .

4 If +24V only is not output... The LCD and keys operate normally but waveform recording and al5 If unisolated 5Va are not output...

phanumeric printout are not available at all.


12
. .

The AID converter and comparator do not operate normally, thereby making ECG waveform or DC input waveform a baseline with alphanumerics printed out normally.
12
.

6 If isolated 5Vf and 5Vb are not output... All 12 leads are displayed and recorded as baselines with alphanumerics printed out normally.
.

A Caution
When checking the power supply and related circuits for troubleshooting, take sufficient care to avoid a short circuit.

3-2

2
2

Troubleshooting

1 All Specified Secondary Power Supplies Are Not Available in AC Operation

(1) Check the FX-2111 is securely connected to the wall outlet using the specified power cord.
(2) Check the AC inlet has two specified power fuses inserted and the fuses
are not blown out.

(3) Check AC voltage at the secondary side of power transformer. Disconnect the output jack, which comes from the transformer, from P1 connector and measure the voltage between pins 1 and 3 of the output jack. The AC voltage should be as follows:
9 9V to 12.5V for 115V version
.

8 5V to 12.5V for 230V version


.

(4) Check supply voltages on PCB-6236. 1) Check voltages between the anode side of D2 and D3 and the cathode side of D1 and D4. They should be in a range of 10 to 16V DC. If DC power output is not available on the board while the transformer output is found normal, the contact of PI connector may be faulty or diodes D1 to D4 may be damaged. 2) Also verify that the voltage between the cathode side (TP22) of D5 and GND (TP21) is in a range of 10 to 16V DC. If the voltage between the cathode side of D5 and GND is not available, the fuse F1 on the board may be blown out or D5 may be damaged.

A Caution
When checking power fuses, be sure to turn the power off and disconnect the power plug from the wall outlet beforehand.
2

2 All Specified Secondary Power Voltages Are Not Available in Ni-MH Battery Operation

(1) Open the rear cover of instrument and check the presence of battery. Then check that the battery is securely connected to the FX-2111 via
the connector.

(2) Check supply voltages on PCB-6236. 1) Verify that the battery voltage is available between pins 1 and 4 of P10 and between the plus side (TP25) and GND (TP21). If the battery voltage is not available there, the contact of P10 may be faulty or the fuse F2 may be blown out. 2) Press the ON key and verify that the battery voltage is available between the drain of Q4 (anode side of D6) and GND. If the battery voltage is not available there, the holding circuit around Q4, D7 and Q5 may be faulty or the connection of key panel may
be inferior.

3) Verify that the battery voltage is available between the cathode side (TP22) of D6 and GND. If the battery voltage is not available there, D6 may be faulty.
3-3

3 +24V Is Not Output

(1) Check the presence of IOVn (8V to 16V). Then verify that the voltage is aupplied to pins 10 and 13 of IC6. (2) Check operations of components around IC6. 1) Using the oscilloscope, verify that pin 4 of IC6 outputs a triangular wave at approximately 170kHz. If not, IC6 may be faulty. 2) Using the oscilloscope, verify that the gate of Q9 outputs a rectangular wave at approximately 170kHz. If not, IC6 may be faulty. 3) Using the oscilloscope, verify that the waveform at the drain of Q9 is switched at approximately 170kHz. If not, Q9 or D13 may be faulty. 4) If +24V is supplied to the thermal print head only, conduct the following: Press the START/STOP key to place the instrument in recording condition. Then verify that -I-24V is available at the drain of Q24 and pins 13, 14 and 15 of P5 connector. If the + 24V is not available there, consider Q24 or Q15 may be faulty or check whether or not signals are sent from the gate array (pin 21). Signal level of Q15's base input: High-24V ON
Low- 24V OFF

If +24V is available there, check the cable of thermal print head.

4 + 10V Is Not Output

(1) Check the presence of IOVn at pin 1 of IC22. To make 10V output available, the input 10Vn should be higher than 10V. If the input voltage is present but the output is not available, setting resistors R119 and R118 or IC22 may be faulty.

5 Vdd Is Not Output

(1) Verify that 10V input is available at pin 1 of IC7. If the IOVn output from the secondary side of power transformer is lower than 10V, however, the 10V input at pin 1 of IC7 is lower by several ten millivolts than the specified -I-10V. If the input voltage is present but the output is not available, IC7 may be faulty.

3-4

6 Unisolated 5Va Are Not Output

(1) Verify that the 10Vn input is available at pin 3 of ICQ. If not, both 5Va should not be output. Conduct the steps described
in 2.1 and 2.2.

(2) Verify that +5Va is output at pin 1 of ICQ and pins 3 and 5 of IC10. If not, ICQ may be faulty. (3) Verify that -5Va is output at pin 4 of IC10 (TP2Q, TP53). If not, IC10, L2 or D14 may be faulty.

7 Isolated 5V Are Not Output

(1) Verify that +10Vn (8V to 16V) is supplied to pins 6 and 7 of 12. If not, check whether or not 10Vn is output from the secondary side of power transformer.
(2) Check the circuit around 12. 1) Verify that pin 11 (TP49) of IC20 inputs a 200kHz rectangular wave as clock signal. If not, IC16 or IC1Q on the oscillator circuit may be faulty. 2) Verify that pins 8 and 9 of IC20 output rectangular waves at 100kHz in reversed phases. If not, IC20 may be faulty. 3) Verify that pin 3 (TP60) of D29 and pin pin 3 (TP61) of D30, each at the isolated side, output rectangular waves at 100kHz. If not, check Q20, Q21 and 12, which may be faulty. 4) Verify that rectified and smoothed voltages in plus and minus directions are input into pin 3 of IC33 and pin 2 of IC34, respectively. If not, D29 or D30 may be faulty.
If all the abovementioned inputs and outputs are found normal but isolated 5V are not output, IC33 or IC34 may be faulty.

3-5

8 Troubles Related to Charging


.

28
.

1 Charging Mode or "CHARGING" Status Is Not Effected

If charging mode is not effected, +5V (CHG IRQ) may not be output at pin 9 of IC18 or +5V may not be supplied to the charging circuit. Conduct steps (1) and (2)-1) below. If charging mode is effected but "CHARGING" status is not initiated, + 5V may be available on the charging circuit and at pin 9 (CHG IRQ) of IC18 but either the battery voltage or temperature may not satisfy the charging
condition. Also, if +5V is not correctly output, the voltage may vary con-

siderably and may be beyond the specified range. So conduct steps (1), (2)-1) and (2)-2) below.

(1) Open the rear cover of instrument and check the presence of battery.
Then check the battery is securely connected to the instrument via the connector and the battery cable is not disconnected. (2) Check the circuit around IC18. 1) Press the OFF key in AC operation and verify that pin 16 (TP44) of IC18 outputs + 5V. Also verify that pin 9 (CHG IRQ) of gate array IC8 outputs +5\/. If not, check the operation of + 5V power supply. . If +5V is output only when the OFF key is pressed Q12 may be faulty. . If +5V is not output at all even if the OFF key is pressed Q11, Q12, or D8 may be faulty. . If only +5V for CHG IRQ is not output check whether or not the collector (TP48) of Q8 is at low level. If it is low, Q8 may be faulty. 2) Verify that voltages between pin 6 (TS) and pin 9 (SNS) and between pin 7 (BAT) and pin 9 (SNS) of IC18 are in the specified ranges as
, ,
,

follows:

Pin 6 (TP42) 1.01V to 2.0V Pin 7 (TP2) 1.0V to 1.78V . If the voltage at pin 7 (BAT) is beyond the specified range check the terminal voltage of battery. If the battery voltage is not higher than 8.0V or so, the FX-2111 is not placed in "CHARGING" status. Owing to this, supplemental charging is made in AC operation. Check whether or not 24V output is available at the collector of Q13 in AC operation. . If the voltage at pin 6 (TS) is beyond the specified range remove the battery and check the thermistor in the battery. Measure the resistance between pin 2 (blue) and pin 4 (black) of the collector. The resistance should be in a range of 3kft to 30kl2 (lOkft at 250C). Ambient temperatures should be proper.
, ,

3-6

28
.

2 Virtually No Charging Is Made Despite "CHARGING COMPLETE" on Display

(1) Check the circuit around IC18. 1) Verify that the source side of Q17 is supplied with +24V. If not, conduct the steps described in "8.3 +24V Is Not Output:' 2) Verify that pin 14 (MOD) of IC18 outputs a rectangular wave at approximately 140kHz; the drain (TP52) of Q17 outputs a rectangular wave at approximately 140kHz; and pin 9 (SNS) of IC18 outputs a triangular wave at approximately 140kHz. If not, check the following: . R18 and R82 including their mounting condition
,

. 017 Q10, 016, and D22


,

. If above components are found normal IC18 may be faulty. NOTE: Charging current is detected as a voltage value at pin 9 (SNS), thereby switching the output of pin 14 (MOD).
,

(2) Check the number of charge-discharge times and the number of operating times after charging. Usually, the number of charge-discarge times is 200 max, though it may
decrease depending on the operating environment. The larger the number of operating times after charging, the shorter the service life of battery becomes. Thus, if the charging circuit is found normal but virtually no charging is made despite "CHARGING COMPLETE" on display, the battery may
exhaust the service life.

9 LCD Operation Is Abnormal

(1) Check sure connection between the P2 connector on the main board and the connector of LCD. Check the connector terminals for any possible inferior soldering.

(2) Verify the following voltages on the P2 connector: GND level at pin 1 (Vss) +5V at pin 2 (Vdd) + 0.238V at pin 3 (Vee)
(3) Check the continuity of the following signal lines referring to "4.2 LCD
Control Circuit!'
Rs

signal

R/W

E signal Signals of DB0 to DB7


NOTE: The LCD module used for FX-2111 has a built-in control driver and

is directly accessed by the CPU, with MEMORY READY put.

3-7

2 10
.

No Key Operation Is Possible

(1) Verify sure connection between the P7 connector on the main board
and the flexible cable of key panel. Check the P7 connector for any possible inferior soldering. Check the flexible cable of key panel for any possible disconnection and wear of the carbon part.

(2) Verify that key signal lines except for ON and OFF keys are normally pulled up by checking TP120 to 125 and TP128.
(3) Check the voltage at pin 20 (power supply) of IC15. Check signals at pin 1 (KEY) and pin 19 (AO) of IC15. Signal at pin 1
should be at low level at every one millisecond.

Verify that pins 2 to 9 and 11 to 18 of IC15 are at low level when a key
is pressed.

NOTE: If only some specific key is not effective, the key may have a poor
contact.

A Caution
The inserting part of key connector is made of carbon. Avoid frequent repeated disconnection and connection.
2

11 Paper Is Not Driven or Abnormally Driven

(1>/erify sure connection between the P4 connector on the main board


and the motor cable.

Check the P4 connector for any possible inferior soldering. Check the motor cable for any possible disconnection.

(2Verify that pins 3 and 5 (TP103) of P4 connector are supplied with +5V.

(3yerify that the following signal is available at pin 1 (TP101) of P4 connector.


!< 5|19. 5;s
+ 10V

;<

5*L9. 5,s

JUULJUU
Paper speed 25mm/sec Paper speed 50mm/sec

Check the signal at pin 14 (TP59) of IC25. (Refer to "2.4 Motor Control:") It should be 256Hz with a paper speed of 25mm/sec and 512Hz with
a paper speed of 50mm/sec.

Verify that the signal at pin 2 (JP5Q) of IC12 is at high level when the paper is driven.

3-8

12 No Mark Detection on Z-fold Paper Is Available

(1) Verify that "RECORDING MODE" is set to "MARK" in the program.


(2) Check the mark sensor for any possible dirt deposit and inferior mounting condition.
(3) Verify sure connection between the P6 connector on the main board
and the flexible sensor cable.

Check the P6 connector for any possible inferior soldering. Check the flexible sensor cable for any possible disconnection. NOTE: The mark sensor is the one and only part in the FX-2111, which is adjusted by a variable resistor. Before conducting the steps described above, therefore, check the variable resistor for any possible peeling-off of the paint lock and slippage.
2 13
.

No Paper End Detection Is Available Or "PAPER END" Alarm Is Constantly Displayed

(1) Check the mark sensor for any possible dirt deposit and inferior mounting condition. Verify that the magazine sensor switch is positioned in place and pressed when the magazine is closed. (2) Verify sure connection between the P6 connector on the main board
and the flexible sensor cable.

Check the P6 connector for any possible inferior soldering. Check the flexible sensor cable for any possible disconnection.
NOTE: The mark sensor is the one and only part in the FX-2111, which is adjusted by a variable resistor. Before conducting the steps described above, therefore, check the variable resistor for any possible peeling-off of the paint lock and slippage.
2 14
.

Program Contents Change

(1) Verify that the lithium battery voltage is +2.5V. Check the lithium battery for any possible inferior soldering. (2) Verify that when the FX-2111 is turned off, pin 28 of IC2 is supplied with +3V from the lithium battery to back up the RAM.

NOTE: Program contents are stored in the RAM IC2 and backed by the lithium battery when the FX-2111 is turned off. If program contents change despite sufficient lithium battery voltage, refer to 2 9 Memory Backup Circuit:'
"
.

2 15
.

Buzzer Does Not Operate

(1) Verify that pin 15 (or R73) of the gate array outputs a rectangular wave
at 1kHz under the condition where the buzzer should sound.

If the wave is normally output, R73, soldering of the buzzer or the buzzer itself may be faulty. In the case where no QRS-synchronized sound is generated, make sure that "QRS BEEP" is set to "ON" in the program.
3-9

2 16
.

No Waveform Recording Nor Alphanumeric


Printout Is Available

(1) Verify sure connection between the P5 connector on the main board and the thermal print head cable. Verify sure connection between the thermal print head and the cable. Check the P5 connector for any possible inferior soldering.

(2) Verify that +24V is available between pins 13 and 15 (TP112) of the P5
connector under the condition where a waveform should be recorded.

Verify that pin 9 of P5 connector is supplied with +5V. (3) Verify that the TOFF signal at pin 21 of gate array or the base of Q15 is at high level under the condition where recording should be made. Verify that signals of LATCH (pin 2 of P5 or TP107), CLOCK (pin 3 of P5 or TP108), STROBE (pins 5, 6,and 7 of P5 or TP109) and DATA IN (pin 1 of p5 or TP106) are available at the timings as described "2.8
Thermal Print Head Control Circuit;'

(4) To make sure of the correct operation of the RAM for thermal print head: . Verify that pin 28 of IC5 is supplied with +5V . Verify that JP3 is short-circuited by soldering
. .

NOTE: If electrical signals are normal mechanical failure may make waveform recording and alphanumeric printout impossible. So check the spring mounting condition and other mechanical status around the thermal print head.
,

2 17
.

No DC Input Waveform Recording Is Available

(1) Make sure that "DC RECORDING" is set to "ON" in the program.
(2) Check the P3 connector on the main board for any possible inferior soldering. (3) Verify that pin 4 of IC21 is supplied with +5V and pin 11, with -5V. If IC21 is not correctly supplied with power, the FX-2111 should not be
able to record ECG waveform too.

(4) Check the AID converter referring to "2.20 Checking AID Converter" If the AID converter is faulty, the FX-2111 should not be able to record
ECG waveform too.

2 18
.

Any Waveform of All 12 Leads Is Recorded As Baseline

(1) Check isolated 5V and unisolated 5V referring to 2.6 and 2.7. Verify that power supply pins of IC12,13,17, 21, 31 and 35-45 (operational amplifier, multiplexor, etc.) are supplied with power.
(2) To check the reset circuit and the control signal, verify that pin 13 of IC44 is not fixed to high level.

3-10

19 Waveform of Some Specific Lead Is Recorded As


Baseline

(1) Check the lead cable for any possible internal disconnection and inferior hookup.

(2) In relation to the lead which results in a baseline, check the buffer IC, the resistance of resistance network, power supplies to lead selector ICs (IC36,38,41 and 45) and their connection status such as soldering.
2
.

20 Checking AID Converter

Check the AID converter if the trouble occurs despite the normal condition detected by taking steps described in Section 2.16, 2.17 or 2.18.

(1) Verify that pin 7 (TP7) of IC13 outputs a triangular wave with an amplitude of 3.2V and a pulse width of 520ns. If the triangular wave is normally output, proceed to step (4) below.

(2) Verify that power supply pins of IC17,21 and 13 are supplied with power. Verify that the reference voltage of +2.5V is supplied.
If not, check the power supply.

(3) Verify that a rectangular wave ADCLK (L: 520(is, H: 480|js) of 1ms cycle is output at 1ms cycles from the collector terminal of Q1.
(4) Verify that the low level S/H signal with a pulse width of 96ns is output at every 1ms from pin 6 of IC17, referring to 2 3 AID Converter Circuit!'
"
.

(5) Check the output of comparator as follows. * Verify that pin 1 (ECG signal) of IC13, pin 8 (DC input signal) or pin 14 (battery/thermistor temperature) outputs a pulse which changes in the pulse width corresponding to the changing input signal. Verify that pins 4, 6 and 8 of IC12 output the abovementioned pulse signal of which the amplitude is limited to a range of 0 to 5V.

3-11

CHAPTER

4
4-1 4-1 4-2 4-3 4-3 4-4 4-5 4-6 4-7

Maintenance
1
.

2 3
4

Replacing Power Fuses Replacing ROM Replacing Rechargeable Battery Operation Check by Self-testing 4 1 Self-testing Procedures 4 2 Examples of Test Results 4 3 System Errors Disassembling/Reassembling the FX-2111 5 1 Disassembling Procedure
.
. .

Replacing Power Fuses

Power fuses are hardly blown out. But if they are blown out due to some reason, turn the FX-2111 off and disconnect the power cord from the wall outlet. Then pull fuse holders off the bottom panel as shown below. Insert spare fuses and put fuse holders in original positions.

A Caution
When replacing power fuses, be sure to turn the power off and disconnect the power plug from the wall outlet beforehand.

Replacing ROM

(1) Remove the upper casing.

(2) The ROM will be accessible on the PC board. Remove the ROM. (3) Mount the new ROM, taking care of the direction. The ROM's socket has 36 pins. If 1M-byte ROM is used, it will be mounted taking the full space for it. If the ROM is other than "IM-byte ROM, put it aside the rear of the socket (aside IC3).

(4) Check the ROM is not contacted with feet of R18 when mounted. (5) If there is no problem, put the upper cover and fix it with screws.
(6) Turn the FX-2111 on. Verify that the control program version on the standby display is the same as the label on the ROM.

A Caution
When replacing the ROM, be sure to turn the power off. Also, take care to install the ROM in correct position.

tfC6 tft
i)

4-1

Replacing Rechargeable Battery


outlet.

(1) Turn the FX-2111 off and disconnect the power cord from the wall
(2) Put the FX-2111 upside down. Loosen two screws off the battery cover.

(3) Replace the battery with new one. Insert the connector of battery cable securely in place. (4) Put the battery cover in the original position and fix it with two screws. If the battery cover is not put squarely, rearrange the battery cable, etc. to neatly accommodate the battery in the room. (5) Press the ON key to verify that the FX-2111 is powered. (6) The new battery may have been stored for a long period of time and lowered in capacity due to self-discharge. So be sure to charge the battery after replacing.

A Caution
When replacing the battery, be sure to turn the power off and disconnect the power plug from the wall outlet beforehand.

"v

NOTE: It is recommended to charge the battery in the stock every six months. Also to make full use of the battery in the FX-2111, follow instructions in the operation manual.

4-2

Operation Check by Self-testing


The self-testing function permits you to perform the following tests:
1 Total Test
.

6. LCD Test

Sensitivity Test

7. Key Test
8. Status Test 9. Automatic Power-off Test

3 Time Constant Test


.

4 Print Test
.

5 Recorder Test
.

41
.

Self-testing Procedures

(1) Press the MODE key to place the FX-2111 in the program mode. The display will be as follows:
1 AC FILTER

(OFF)

(2) Press the ImV (A) key. The following message will appear:
MAINTENANCE

PUSH (START) KEY

(3) Press the START/STOP key. The display will be as follows:


M-1 AC FILTER

FREQUENCY

(50HZ)

(4) Press the ImV (A) key. The following message will appear:
SELF TEST

PUSH (START) KEY (5) Press the START/STOP key. The display will be as follows:
1 TOTAL TEST

Press the START/STOP key once more. The total test will start to test sensitivity, time constant, print, recorder, and LCD. To proceed to the Key Test, Status Test or Auto Power-off Test, select it using the < or key.
9 AUTO POWER OFF

4-3

1/2 1
Z

TEST
1

42
.

2 Print Test

n t a *
'

'

1 *
;

42
.

3 Recorder Test

....

;..>

Mm"

RECORDER 'EST

42
.

4 LCD Test

0 o 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0

0 0

4-4

3 System Errors

System errors are caused by troubles with hardware or software. Usually, they cannot be treated by users.

If a system error occurs, the following alarm message blinks and


the buzzer sounds:

(Blinks)
S

m R
9

R
0

CODE

If a system error occurs during recording, the operator should stop the FX-2111 from recording (stop the motor). All the operation keys will be ineffective. Also, the normal operation will not be recovered unless the power is turned off once.
Error codes and causes are as follows: Code
01

Type
Overheat

Description Thermal print head is abnormally heated to over 60oC. Checksum error
RAM's read/write error

02
03

ROM error RAM error

04

Unused interrupt 1
RTM error 1 RTM error 2

Unused software interrupt Task No. to RTM macro is illegal. Task status to RTM macro is improper.
Task's queue area overflows.

05
06
07

Unused interrupt 2 Unused software interrupt

08

RTM error 3

RTM: Real-time monitor

4-5

Disassembling/Reassembling the FX-2111

For maintenance, inspection and replacement of ROM and other components, you need to disassemble the instrument. Use the following steps to disassemble: The FX-2111 will be reassembled in the reverse order. Also, refer to the Chapter 9 Structural Diagrams" when disassembling.
"

When disasssembling or reassembling the FX-2111, take the following cautions:

A Caution
Disassembling/Reassembling Precautions . Be sure to disconnect the power cord and make sure the instrument is turned off before disasasembling or reassembling.

. Do not remove the battery before disconnecting any PC board

. Take care that repeated disconnection of the key panel and sensor board may result in poor contact.
. Use proper tools to loosen screws
,
.

. When reassembling make sure that all screws are securely tightened and all connectors are completely inserted.

PC Board Handling Precautions

. PC boards are equipped with highly sensitive components to static electricity.


. PC boards are highly sensitive electronic devices Put removed PC boards in a proper protective bag or take appropriate measures to protect them.
.

Handle PC boards carefully. A shock to them may damage the components.

. Never insert a connector to the powered PC board nor remove the powered PC board.

4-6

51
.

Diassemling Procedure

Take the following steps to disassemble the FX-2111: (The procedure described here is for the model equipped with the battery.)

Doublt Semus Screw (S)


M3x8
~

e7)

g3

jo

Step 1: Remove the cover of paper magazine and the paper shaft.

Step 2: Open the battery cover on the bottom panel and remove the battery.

Irom Ttanslormer
2-Double

Semus Screw (S)

Irom RecodJer Motor

Thermal Array Head Cable

Sensor Cable
Double Semus Sciew (S)
.

fe

M3x8

1/

51
*
.

0*
4-Double Semus Sciew

(S) M3x8
.

Step 3: Loosen four screws on the bottom panel and remove the upper case.

Step 4: Diconnect connectors from the PC board and loosen two screws on the
PC board. Then, the PC board can be
removed.

4-7

Double Semus Screw (S), M3x8

3-Double Semus

Screw (I). M3xB

r
>

8
8

0
0

0
0

Step 5: Loosen three screws which fix the recorder onto the botton panel and remove the recorder.

Step 6: Loosen the screw which fixes the power board PCB-6238 and the transformer onto the bottom panel. Disconnect
board.

the

connector

from

the

transformer and remove the power


2- Double Semus Screw

(S)

M38

m
8

8
el

Step 7: Loosen the screws whch fix the transformer onto the bottom panel.
Then, remove the transformer and

grounding parts. Now, disassembling is complete.


4-8

CHAPTER

5
5-1 5-1 5-1 5-1
5-1

Periodical Inspection
1
.

Periodical Inspection 1 1 Checking lead cable and power cord 1 2 Visual inspection 1 3 Cleaning the FX-2111
.

1 4 Self-test and total function check


.

Battery check Safety Inspection 2 1 Measuring leakage current 2 2 Measuring protective grounding resistance
. . .

15

5-2 5-2 5-2 5-5


5-6

2 3 Remarks
.

Periodical Inspection

In this chapter, we describe the inspection to prevent troubles and let the FX-2111 keep the sufficient safety and complete operating condition.

Perform the following at least once a year: Check the lead cable, power cord, and grounding wire for any possible damage.
. Conduct visual inspection
. Clean the FX-2111
. .

. Perform the self-test and total function check

. Check the battery . Check the leakage current Measure the protective grounding resistance.
. .

As with simple inspection, recommend the user to make it every day or week. But let the user refer a complete function check and troubleshooting to the service engineer authorized by Fukuda Denshi.
11
.

Checking lead cable and power cord

Visually check the lead cable and power cord for any possible damage. Check also the connectors for any possible looseness. If a damage is found, replace them as required. Merely applying a tape may not repair the cable sufficiently.
Also measure the resistance of each electrode to check for internal

cable disconnection and damage.


1
.

2 Visual inspection
.
.

. . . .
1

Check that mechanical parts such as screws are securely fitted Check that connectors inside the FX-2111 are securely connected Check that there is no trace of damage If a loose part is found tighten it securely.
.

3 Cleaning the FX-2111

The recommended number of cleaning times depends on the operating frequency and environment.

For cleaning, wipe the casing with a properly wetted soft cloth. You
can use the cleanser for tableware by weakening it with water. However, take care that the cleaning solution or water may not enter the inside

through openings. Finally, wipe off wetty substance with a dry cloth.
1 4 Self-test and total function check
.

Perform the self-test described in "4. Maintenance" and check the total

function of the FX-2111 to ensure that the equipment operates safely and completely.

5-1

15
.

Battery check

Measure the voltage between plus and minus sides of battery to check the capacity. If it is below 9V, charge the battery. Also, if the battery has not been charged for six months, charge it. If the battery will not be used for a long period of time, remove it from the FX-2111 and store in a cool
place.

If the battery is below 6V, it may be difficult or impossible to charge. In such a case, replacement is required.

Safety Inspection

To ensure the safety of FX-2111, it is recommended to perform safety inspection. The test methods and measuring instruments are stipulated in the standard for safety test. It is considered extremely difficult to fully follow the standard for the safety test at the health care site. In checking for maintenance and inspection, therefore, measure each specified value as a rough rule of thumb. We describe here examples of simplified measurement of leakage current and protective grounding resistance.

A Caution
If you find a value which exceeds the allowable level, be sure to let the user avoid using the FX-2111. If the user operates the FX-2111 as it is, he/she may receive an hazardous accident.
2

1 Measuring leakage current

Prepare an instrument to measure the leakage current (electronic or digital voltmeter), impedance device and power switch box. The impedance device has the following configuration:
R1: 10kn 5% R2: 1kfi 1%
R 1

C1: 0.015tiF 5%
V: Voltmeter
C 1 -r V

R 2

You can obtain a leakage current by measuring voltages at both ends of the impedance device.

5-2

The digital voltmeter shall indicate a true root mean square value to a composite waveform of a frequency band from DC to 1MHz. If such a voltmeter is not available in hand, you may use a commercially available digital voltmeter for the purpose of simple maintenance and inspection. However, make sure of the frequency band of the voltmeter and note that the voltmeter will not indicate a leakage current at a frequency exceeding the capacity.

(1) Example of checking ground leakage current (current flowing in protective grounding conductor)
Power Switch Box

Polarity Switches
POWER
-

oa

ob
O a

Wall Outlet

Medical Electronic Instalment


GND

GND

'

Oft
8 Point

A Point

Power Cord
M D

MD=lmpedance Measurement Device

See the figure above. A gounding terminal and the power switch box which allows you to switch the polarity facilitate measurement of ground leakage current. The procedure to measure ground leakage current using the power
switch box is as follows:

. Measurement of ground leakage current is made by measuring voltages at B point (grounding conductor of power cord) and A point (wall grounding terminal) under normal condition and single fault condition.

Normal condition

. Measurement of ground leakage current under normal condition is made by switching the polarity switches to a" positions, then to b" positions. Thus, measurement is made in two ways.
" "

Single fault condition . Measurement of ground leakage current under single fault condition is made with a power fuse removed by switching the polarity
switches as above. Since the same measurement is made with

another fuse removed, measurement for single fault condition is made in four ways.

5-3

(2) Example of measuring enclosure leakage current (current flowing from enclosure to the ground)
Power Swdch Box
;

Polanty Switches
-

!
Medical

POWER

Oj-

ob

Wall OuUel
GND
GND
T-O

Electronic Insirument

B Point

A Point

Opsn lor ilngla lauli condillocyPower Cord

| M D \-

Metal Foil

Probe

MO Impedance Measuiement Device

Or

Apply a metal foil of 20 x 10 cm to the isolated enclosure. Put the probe


to the metal foil for measurement.

Measurement of enclosure leakage current is made by measuring voltages at A point (wall grounding terminal) and B point (desired part of enclosure). If the enclosure is isolated, stick a metal foil (e.g. aluminium foil) onto the enclosure with a conductive tape and apply the probe to that
metal foil.

Measurement of enclosure leakage current under normal condition is made by switching the polarity switches to "a" positions, then to b" positions. A value measured under normal condition will be a very small value near zero. Measurement of enclosure leakage current under single fault condition is made with the wall grounding terminal and the grounding conductor of power cord removed and by switching the polarity of
"

power source.

In leakage current measurement, take a maximum value as the measurement result.

The following are maximum allowable ground leakage current and enclosure leakage current: In inspection, it is important to compare measurement values with past ones to check for any outstanding difference.
Maximum allowable leakage currents (117/253V AC)
Normal condition

Single fault condition


1 0mA
.

Ground leakage current Enclosure leakage current

0 5mA
.

0 1mA
.

0 5mA
.

5-4

22
.

Measuring protective grounding resistance

According to the standard, protective grounding resistance shall be measured by letting the testing transformer flow an AC current of 10-20A from a power source of which the voltage with no load does not exceed 6V. Practically, however, it is not easy to prepare such a testing device. We show therefore a simplified method as follows:
V
O
o

Power Iranslormer shall be able lo let a curfent of 13A How at a voltage ol less than 6V with no load
.

Medical Eqjipmenl
Power
A Point

VW
30

Or

C Point

EO

Transformer
A C 6V
2 30 V

(SOW) 6
o

Metal Part

VI (Voltage between A and C)

V2 (Voltage between

R: Protective Grounding
Resistance

B and C)
B Point
.

El

t
Power Cord '

Simplified Measurement of Protective Grounding Resistance


The testing circuit above allows you to measure the protective grounding resistance of equipment using an AC current of 2A or so. Connect the power cord of equipment to El and the metal part (such as equipotentialization terminal) to EO. Make the conductor to EO shortest possible to suppress the contact resistance. Flow a current of 2A or so to the equipment and measure voltage V1 between A and C points and V2 between B and C points. Since the flowing current is the same, the subject protective grounding resistance can be obtained through the relations between the voltage ratio
and resistance ratio as follows: R/3 = V2/V1
"
.

R (protective grounding resistance) = 3 V2A/1 (fi)

Measurement current may be 1A or so.


With the standard, the resistance between an accessible conductive

part and the protective grounding terminal shall be 0.1 Q maximum. In measurement at the grounding conductor of power cord with the power cord connected, the grounding resistance including that of the protective grounding conductor of power cord shall be 0.2ft maximum.
Protective grounding resistance

Resistance between protective grounding terminal and potential equalization terminal Protective grounding conductor of power cord and potential equalization terminal
5-5

0 10 maximum
.

0 20 maximum
.

2 3 Remarks
.

Abovementioned measurement of protective grounding resistance requires the power transformer which can flow an AC current of several amperes and a testing fixture. If you cannot prepare such a testing power transformer, you may measure it with the following method to obtain a rough rule of thumb:
Measure the resistance between an accessible conductive part and the protective grounding terminal to verify that it is lower than 0.1ft. If you measure the protective grounding resistance with the grounding conductor of power cord as the protective grounding terminal, the resistance shall be lower than 0.2ft including that of the protective grounding conductor.

You need also to adjust the resistance of the testing lead or probe
used for measurement.

Reslsianca Measuring InsUument

Medical Ecjjipmenl
Testing Lead
Accessible Conductive Part

< 0 .

1 fi

(e.g. potential equalization terminal)

Protective Grounding Terminal

Potential Equalization Terminal


nvj
po-m

Protective Grounding Terminal

Simplified Method of Resistance Measurement


This method considerably differs from that of the standard. Consider the result as a rough rule of thumb. Protective grounding resistance Resistance between protective grounding terminal and potential equalization terminal
0 1ft maximum
.

Protective grounding conductor of power cord and potential equalization terminal

0 2ft maximum
.

5-6

CHECK SHEET FOR PERIODICAL INSPECTION


Inspection Date:
Model Number

Inspected by:
Installation Site Years of Use

Serial Number

Check Items

Judgment/Measurement : Remarks (Repair Required)

VISUAL INSPECTION
1 Cords and Cables
.

Power cord and lead cable for damage Connections of power cord and lead cable
2
.

Pass/Fail Pass/Fail

Casing and Accessories

. Casing for damage such as crack . Casing for dirt . Labels for fading letters . Screws for loooseness and dropout
. Electrodes for dirt and rust

Pass/Fail Pass/Fail

Pass/Fail '
Pass/Fail

Pass/Fail
Pass/Fail

. Expendables and operation manual


3 Connectors
.

. Power and lead connectors for damage and looseness


r
-

Pass/Fail

4 Others
.

. Power fuses for breakdown

Pass/Fail Pass/Fail
Pass/Fail

Protective grounding terminal for damage Switch panel for damage and dirt Thermal print head for damage and dirt Paper magazine for damage and dirt LCD module for damage and dirt
ELECTRICAL INSPECTION

Pass/Fail Pass/Fail Pass/Fail

. AC operation and battery operation


. Time constant test

Pass/Fail Pass/Fail Pass/Fail Pass/Fail Pass/Fail Pass/Fail Pass/Fail

. Recorder test (paper speed & sensitivity)


. Printout test (for missed dots) . Key test
. LCD test

. Auto power-off test . Battery charge


SAFETY

V)

Pass/Fail

Charge OK/NG

. Ground leakage current (normal condition) (single fault condition)

|jA Pass/Fail

pA Pass/Fail |iA Pass/Fail


|jA Pass/Fail
Q Pass/Fail

. Enclosure leakage current (normal condition) (single fault condition)


. Protective grounding resistance
OTHERS

5-7

CHAPTER

Circuit Diagrams
Overall Block Diagram Overall Block Diagram Main Board Circuit Diagram, 1/9 to 9/9 Power Board Circuit Diagram
Sensor Board Circuit Diagram Motor Module Circuit Diagram

433-5125 6-1 433-4893 6-3 523-2020... 6-5 to 6-21 524-2021 6-23


524-1959 524-1960 6-24 6-25

Battery Pack Circuit Diagram Overall Connection Diagram Overall Connection Diagram LCD Connection Diagram Motor Connection Diagram

524-1961 433-5126 433-4894 434-4895 434-4896

6-26 6-27 6-29 6-31 6-32

r .
.

Main Board
, i

Bult.r

DC Input

Amplifier'

Pr.B-6409
-

Power Supply ' :


i

>

5V

I i
i
'

A C 1 1 5 V
I

5
u
c

rs
Butttr Amplifier
Resislanct NelworK lor Lead Slectior

Isolated Section
-

Time

Preamplilici
it

ConsUnt/
'

Middle

Reset

Ampliner
i25

n ModulaiJon
Circuit

Tranjlormer
I

Demodulation Circuit I->

! fi!
Amplilicr
i US

T
.

J r-

. ""

< IA

I . , AC Input
i
i

Circuit I

Triangular |_
r-Wav*
_

I
l
I
Digital
Power

+5V

: i
.

'

I .

AC Inlet Board

'

'

liaV
Amplifier

Generator ~

jI

\\ \\

PCB-410

'

+5V.

Pustv

i . - - -

i:
*

1
i

Control

Power Circuit

Isolation Translormer

Pull

. I
.

Circuit

+VFB
i

21
Oscillator

+ 5V

Supplies
Rectifier Circuit
Trans I or mc/

R-wav*

Overinpul
Detector

Detector
Circuit

Circuit

f jlse t Generator Circuit

T
i

+ 10'/

: i ; i i
,

Photo

Photo

Coupler
Recorder
t

Coupler

0
A

KM
_

CPU
H063B09

; i 1 i
i
'

2HHz
+ 24V

i
.

Sensor Board:
'

PCB-62 39

Sensor Circuit

T
Gale Array

In

0
R

i i

ROM

i
i

FD88O07-AC
B
Motor Module
Motor

E
S

AM27C512

Charging
Circuit

54Kbyte

U s
4 \

S B
U
-

| A V A T
_

NF-26P1/25|*

Control

7t

:i :i

RAM

ji

TEMP
Thermal
Prinl Head
t

SRM2264-12
SKbvte

r
Ni-MH

QS6-F
L - . - - . .1

RAM

CLK

Battery
PacK

SRH2254-12

16HHz

8Kbyte
lor Thermal Prinl Head

LCD NOH202AOO

Key Panel
i

TITLE

DRAWING NO.

Overall Block Diagram


MODEL NO.

433-5125
DATE

ASSEMBLY NO.

FX-2111 (USA)
6-1

96.10.9

Main Board .
-.1

Butter

DC Inpul

. '

pcb-6409
-

Amplitief

Power Supply 1 ;
sv

i
o

AC
! AC Input

230 V

rs
Resistance 5 Network lor Lead Seltctior

Isolated Section
i

1
Demodulation Circuit

o
V

Time

Preamplilicr
xB

Amplilier

ConslanU
-

Middle
3

1
Modulation
Circuit
Isolation Transformer'

2S1

T
.
,

|i
AC Inlet Board

Reset

'

Amplifier
125

Amplifier
xl.25

5
+ 5V

Circuit
I

fc

I I

Triangular
r-SWave
_

1 ,

pcB-e4io

ItnV
Amplilier Control
i

11.

Generator

i5V
iVFB

Power Circuit

Isoiation k Translormer

PushPull

Circuit
AT
in

I
Oscillator

Digital
Power

+ 5V

i
i

Supplies
Rectifier Circuit

Transformer

R-wave

Overinput
Detector Circuit
-

Oelector Circuit

Pulse Generator Circuit


v

+ 10V
I

Photo

Photo

CPU
HD63809
D
A

Coupler
Recorder
Sensor Board:

Coupler

2MHz FET
i D
+2 V
> i

PCB-5239

Sensoi Circuit

T
X
-

Gate Array

ROM

FD88007-AC 8
Motor Module
Motor Control

E
S S B
U

AM27C512

64Kbyte

Charging Circuit

NF-26P1/25

<

> AVJT
RAM

I i

TEMP
Thermal Print Head 056-F
I

U S

SRM2264-12

J
r

8Kbyte
Ni-MH

R AM
i

CLK

SRH2264-12

16MHz

LC 0
NDM202AOO

i
Key Panel
-

Battery
Pack

8Kbyte
lor Thermal Prim Head

TITLE

DRAWING NO.

Overall Block Diagram


MODEL NO.

4334893
DATE

ASSEMBLY NO.

FX-2111 (CE)
6-3

96.10.9

noi!) i miss omtnsi smcifih


.

AU IEC1ST01S tit IN Q \%, I. ||


,

ALL cmcims tit IK U F

VBF-f VRF -

SPICUl SYHtOlS UUCE

IC37
_

uPC4064G2

1039

4 4
4

<D tfl. tin IIS! PAH


. <M iM 1!S1 PtD
.

IC31
IC35

uPC4054G
TC-4051BF

4 14

JUL
7

IC3Q

TC-4051BF
TC-4S3flRF TC-40S1BF

is 16

7,a 7,ft

VBF-

16
16 3
16
Fl F2 7
4

A
7,6

I1 <

Cl IB

C97 Jcxot, )<:

Ci 19

2.
7 ,a

IC44

TC-40g4BF

T
VDD -14

T
FC

IC451TC-4051BF

TCI? .TC-4Sfl4RF

+RV 4
-

-RV -

.SWF

SVF

IC13

uPC4074C

IC15
TC17

TC7W04F
TC-40S3RF

a
7

11 -

JL4.
14
_

C93 jcil? |ci25 |ci28

C94 \ci Ifc |ci2t ]ct.27

ICig

TC-4Ba4BF

IC20
IC21

TC-74PC74F
u PC40B4G

7 8
4 4

L4_
IB
9
FC

T
FC

T
E 1

T
C I

HEF4046BT

IC26
IC27
TC7W02F

VDD

VDB

6
VDB

IC1 TC2

QM27CS12-150DC SRM2264-12

14

25

14
14 IB 40
65
'

26

I Ci

\ CC

|ci3 jcT |cZ2 ] C30 j C31 | C

4>

IC3 ICS.
ICS

HD53B09E
5RM2264-12

FDa6007-PiC

26 3 26 63

JTTTTTTTT

jJTTTTTTTTT
V

| C3I joi Jclfc | C37 | COS jcfcl lct.6 ] C68 J_C7 3

90
TC11 TC74HC24BftF

1
7

2
J 4
_

IC14

IC15 IC23

TC74HC04nF TC74HCB41ftF
TC74HC04RF

1A

22l
4.

TITLE

DRAWING NO.

Main Board Circuit Diagram, 1/9


MODEL NO.

523-2020
DATE

ASSEMBLY NO.

FX-2111

PCB-6409

95.01.09

6-5

aoac-isscicr >

co1
IC3T0

rc:200H>

Aiit
2 OK

0.0013

XL
H3
o o
-

mil

o o
-

<

i 3

. IIS 2 OK . 1 rnuj . 21 1
'C

s
f

ao
i

JL
n 163
i-

rHI-""

12

2,OK
-

JJ
It
X4

5 4
. 11
IC37B

'0"

-"vw

n tai
20K
" -VW-t

ni&i
3K
VW

caj
0
.

CB7

rc

ICS 10
1 3

P'
t 1*1 LI f

%7

>
>

22

. IBM
,

IC3JC
lT
,_

7 OK
1

140K <

[ c04 X X C04
.

19

>-fl.
uPCIC

01S|

5= |
T

I t11170 > 13.5n > J*


rc
LSS302
1 7 I

LU.
100

WPC4044
Jl J

434

1 SS14

tlK

IJ IQH

-L. > | C 10]

. 210
20K

iC3a,.
40

I) HOP

JL

n ia2
ZuK

IC370

fC

> 30QK

. 294

20K

X I

1
>]

tC3

If
<( a? I NH .

uPC40t*

n i hi
t 40K

H 183

2
1
<1K 10M

. 201
20K

20K

5VF

. IK

2r
] .4KB
13K0

032a T 1SS301

T
017 1

C 104

JK rc

HQ*

'

iBsr~ i i

104 S
10* ?
4 . B

rnua

> 2 If

> JOK

u
C I

xo
It 1

522 <.i/jsv

> 113 19
JL
mi'

A ,c"
03 33 UN3

11214 > 5 IK ?
.

Xft2i4 ? 9. IK
Pa 4

uf*C4044

If
4 I

0
f

UIIO04CSB-1.2
rc

9 HOP
131

[CI 13

Jl

t NH
.

<5
4093
-

oo**r
Paga 4

svr

1cjo A2ao LOK UPC4044

c
'

TWPPrtT A l

'~

1051
I tg41
,

1
4 0?S
?- ai
i- 02
X

ClOl

40

I 3io0i.r
Cl 12
0
.

T
Q4Q runs

Cl 14

HOP
R20

I 12
4
XC
-

sm
O

3:

30K
I cue

%- QJ

CLK

[ j 04

oc
....

. 220

LOK
UPC40ft4

IK

3t
Li fl 2 3 Z
*

I NH
4

1035

IK

0223

0 2
Cl IS
0 01
.

I A
e

JJl
-svr

I OK

S IK

[ C 120
V 1S0P
44 1

TOTi

Ton-

4093

CO

IC400 C42C 41
C I

. 342

F _
rC404

lOK

vr

1C430

. 23ar 13K
}

4S30

<1K

,J lOM
V L0H

.301
*IK

JJL
UX5 212

00.
1C43C

044
-

rnM2

svr

rs
"140

Q
0234
7

JL
ic2o

uPH131

Li
LirC404 C7B
o.oca ?wr
*

11 2

SIM

[ C 123
1S0P
. 1 BO

941

. 174
5 . IK

. It 7

0 22
.

. 169 200K

R7S1

R73

. 234
-

svr

. svr

rnu2
rc IC42n

/W

'
-

C79

iC3in

CI3S

'

6ok
C I I 1
A224 49aK 0

A
1

p.

1
M23T
. 230

IK i J 1 OK JL

1>in

"

PC rc40*4

-S\J~V-

17* 10K

p 9
rc

10K

-T-n <.pc404 4 o.jj ; i?i


5bi7S
} 140K
rc

peo44

10K

HI3 .003 3
PC40%4 rc 1C43U

04 3
UN 112

0 22
.

IJ

..

R210

/300K
1

. 20
-

<fl 1 73
rc

Ji

? . IK

uPC33 5

rc
5

24
--*/W-

IC43n

1124 1
C 134

IK

n2:e e . IK
_

fll4ft

R 2SS
C 1 32

24 oloaii HI-

PC33

JL

CO

C122

"

2 2 OOP
:

2S2
l 30 O . 4 t

? 3 f
/tiOK

IC4aA 4S3a

C133

3or

II
. 24S

zook

I C4 J

SiOP
. 244

. 24*

. 247
-

IOK

S IK

1-0
uFC 4064

25V|

CI g)

C131 HO'

TITLE

DRAWING NO.

Main Board Circuit Diagram, 2/9


MODEL NO.

523-2020
DATE

ASSEMBLY NO.

FX-2111

PCB-6409

95.01.08

6-7

R12(.

2K
-

>
Past
N 1 7 B

v/V

n bfi-

>
K 1 79

Ri 28

2K

<1I> IC47
0.022

f c : 25
j
.

zr.

1' SI* 1age

>
03 2 5514793
T 1

C22

p 10 '

nBD

5511391

120KF
IO00P
CfcO

30Kr
2 C21D

Pa??

Rlt6

rtu

>
1M

E j

Li
1O0OP ce7 1M k 1 25

R4 a 1 n

C56 C5B iOOOP


D O1
.

CS7 12 D . 0 J

R1CS
1- r in
w

0.01 0

I
OOP 5
in

UPCb6(jt)
7

2K

J C 13 3

R43
20K
3

IC 128

IT
02

0
* 1

L3
10O

z
uPC07a

63 1 ce5. 0 J
.

Irc5e
0 11B

5S1393

SSI 4393

T C23
_

2
9

E J

V O 022
.

ISS302

5
FC

zo
c 1

R56
in

6
1 1

I NH

Kb 2

2K

1 C 13 C
.10

R57

rc jjc

Pasff

fa

uPC407a
5V D 19B

20K

E2

L 00
_

] C3e
v 0 02 2
.

at
1 C J

rc58
D20e

1SS302
I C2 lO

1SS302

E 1
K 7B

2
P3

R 106

R 109

3
3 R 1 IS

R5 0 19

n v

vv,
IK

SSI 4393
3
.

I"

UPC4064
.

RSO
ZK z

Rill S-C8036-01 B2K

Dl 90

I C 130
R4

1SS3D2

"ess
5 IP
vr-rr (2 . 5V >

1DO

ICIJO

P9*

SV
R 114

II
6

C5 4
R5

330OP
lOOKB

R4B

JL.
'

1-I

PC4074

20K

t C45e D 12B
1SS302

E 1

78.7KB
I C2 1 C

2 C2 IB R53 39KB

J0D T

1 C 13B

022

Z 1 Rl 13

R130 Par 7

. Z e UPC4064

(EaIIiZ>

10

7F.7KB
Rl 12 UPC4064

n
C 1

R5 1

UPC4074
t-3

30K 2 - *Z .2V J Cfe.25 V LSB)


.

C55

76.7KB

22

S60P
E 1

El

Pgr

R 1D3

4.
018 UNS112

Pave

5 . IK

R 1 OB
B2K

< t enp ysm 1


I cizc 5/M
"

1 p,r 6

P9r

C45HC

nToi

VOL >

TITLE

DRAWING NO.

Main Board Circuit Diagram, 3/9


MODEL NO.
ASSEMBLY NO.

523-2020
DATE

FX-2111

PCB-6409

95.01.08

6-9

<-i

6vn I Pa9,' 7
>
PaF 3

-E13Z> p,*. 3
TP60

-(TftiTPS 1

<Tftf 1<rBFi
Page 3
TP69
~

IC33
~

gg0
030
1 55302
.

Njn78L0SUa
3

ySKT.T .

.
-

yj
rl
1SS302

42

i"
.

VDD

:T2

<n2!DT P7C

e s

R123
WV-

C 107 .
0 . 1

22Z

CiOl

IC20B
0 P R D 11

H1.31
13K

T P4 9

Ci02

<zsso7P71

2
3 3/2S<y I 34

"
a

f *3
029 1SS302

ci3e 33QP
5

RL20 I 13K <

1
E2

CLK<Q L

I5_

HJI"179l 0SUn
_

2P
.

? R 12 1

U7'>C74
3

430
D23B 3 Rioa
'W
-

921

D33C '

T o

Pfi'se 2 3C37,39.02
Pauer Pin

T JSKA-

inrn ao

155302

42 .2

1SS302

C93
"

T
(?) |

R10S
-

R79

W** IK

034A

TP72

'm;
C10O

T
IC16B
TC7W04F

"vV

2 . 2K
Cqfe

1SS302

IC16
TC7W04F

100C", 1000 IC16C IClt S TC7W04F

IC19C

T
FQ

O. 1

4584

6
2O0KH2
C74

I C29
NC

PS26S2L I

1|
O . 04 7 0152 430

Pag

ICi90

TPS

Page
C72

tC28
NC 6
_
.

fS26S2L |

1|
0 047
.

P jge

2
V-

R149
430

<ftt1PDaTO

ic j.9e
4

7P9

Page

5VF
1

IC30

PS26S2L
.

Past

NC

[hp~

3
R222 3K
R102 15K E2

IC1.9C

TPli

TC4SB4

TITLE

DRAWING NO.

Main Board Circuit Diagram, 4/9


MODEL NO. ASSEMBLY NO.

523-2020
DATE

FX-2111

PCB-6409

95.01.08

6-11

V/OD

6
Rise
P6

20K

08370061000800

CKYOCERO

ELCO
6

R 159 2K

I C 12F

I
E2

13

r 04584
C77

To Senior

PCe-6239
2

R160

% IK

7
C76 0 3O

I 270

I C27B

ft"
a
.

N.

02 0270

R ISO
IK

R iS6

IC19D

i
k
.

.
155 ISS302 S2K IK

Page

R1S7

ozeo

L3k

1S5302

UNS212

4584

IC7U02F
R 15 X 240K

C7 1

IC7W02F C 75
0

Q .022
0 1

7
C2 E2

E2
C2

9
C2

C2

R 153
ISO R154 R 132

6
/N/V

ISO
iC2oa

929 HI 33 JN5212
027

P4

1 K

925

Dr3A-6P-20SO

2SA14e3 026
IK

S >CLK
L

UNS212 C2

Ra9

as

>

T
E2

cJ>-i
250

UM51 12 1SS302

2
3

Y 74C7<I
1

V0C

PA9e

HZ?

baov

3
>
NC
R 145 I OK IC260

CI N

PP

VCOUT SIN

p i

2NC

IN2R 1 41

GEEEE
Pas*

P2

2
R 1 42

I N 1 -

IN2-

LS
L2

R137

T 24K

&

10 OK

R 138 24K

R 135

HCFOUI

8148 CX VCOIN [MH


Z
10NC
C64

lOOK
NCTT
R 1 R2 OCnO ZEN
4 . T/SSV

Rl 40 I
20K

I 4-3K
.

D T C

QUICNI

L2 L2
7
C2
C2

C i
R T 7

cc

I C26B

D26A

a 14 7

2
16tC

C2

i-<>
2
C62 . .
.

R13

5 1./-0. 2S J

5
1SS302 C70 4 7 X2SV
.

CNO
c i

PC35B

CG5

228

UN51 12

R14*.
5 IK

Lj

L2 C69

a i

C 1

uPC 4 9 4 C5

2200

R 1 43
7K

ce3

R13G

R13?

C2

7 C2

0 . 1

1K

3&K

1 C 1 2A

i
1 t <i5b4

K 1 44
1 K

TITLE

DRAWING NO.

Main Board Circuit Diagram, 5/9


MODEL NO.

523-2020
DATE

ASSEMBLY NO.

FX-2111

PCB-6409

95.01.08

6-13

2SC

R 14 22; Jfcv

2i
6C P5t

S
"

A
|
-

as

O NWI

5-a

ib
ei i

55 I! r,
D7

6iC

iJ

* R92-Bfc

"D* 1 t.

11
AC
AC

D93

fid

2!
E2

il...
B9<

P*9#

-a r mo

It- e

SI
Old

1
26

"1
VOC

A?

17

;3

it vis

DC
C 1

Si

H-<

Y2
V7

ST RST ' ST OF
-

C2

OI127C512-1S6DC
tic

< KE Yd
NSt

it

. NOTE J

at
B9-

0D

- 12
2
20K

IDC
-

< CMC

5 T O T u

2QK

F*q. "

MS
-" P
1
flTTu
[CI 1
i

g...ml!5iii MSfEE rifM ass -ea jfrlff


3 ft?

IE si
etT vol
Thro 12
BIO 1 240k f 10 t A 1 1 aii

i xo?

cS:
C5

JF3

<4

CSI we

8gi ;
s-11
? DE7

-S-5

ft 12

oE

TmmDO

rH
h 01

>

01 R

1*

TMno=
LOC

-S-i

V
< AQCLK
Fa** S

I C 1

ISSI il
f P30

19

NOT I OK
OMFD

sHhiigfsJiI
Fag* 3

ill
0 12 5 ES S

NDFCISSKZR

It23F "":C23C
I C23D C2 IC2SO
i
_

IlOTCH

>

[SJ2__CiiN>

R251

W0TC2
R01 in No.

. NO 1 SI
_

<F_orr

C 3:S 12Kbjt ici-ac


< 3: Jlb it

U* U*>

f c Ktei 1 RAW 1t-.n connect JF1, JP3 256Kb.it RAM t *n connrct JF2.JF4

5 r CHC

{> fe
JPfc

JKC__>

R7? 9

1Kr>
Us*

Soc-t
infen

connect

JPfc

co*-.n*c-t

JPfc. JPB

TITLE

DRAWING NO.

Main Board Circuit Diagram, 6/9


MODEL NO.

523-2020
DATE

ASSEMBLY NO.

FX-2111

PCB-6409

95.01.08

6-15

03

01 -*

nC PCTCCT
"

rpBBOO?
"

3
D3
3
.

1
l a

f
IP M T f R T>

XT

0
OS D3f Si
2*

@
0 3
024 CND

O 1
.

2S J

( *2*v J

3
C 12 2

R.*
C16

2900

R3
09

BL. 0 1 BM 1 -Ofe 2

I
T
lit.

i'3

< mo

2DOM
'

SAH
IOO/2?v UM5ti2 086
R2 3 4BK
.

> 20K

M * - >JW

ON

o IS UM5212

272
CI37

o- B
1-

NMRCN/NM

08

3?

Tf>8

>

C9

I OOP

i
[C22

2 Ion .i >
i

I
/3SV

D 1

013 UN5i12

/1*V

BJ IB

>

ik r
9 11

C17 4*

251
2
Di6a

1553D1

@
Q 17

>
No t kl

>SS ZSJ2

IPS?

J OuM

C 19

W8 L

D.DJ vi n -wout

C29
0
.

"t?

20K

UN3212

ISS3D2
oen

D21

3L
z

Dirs *

KB J 100

f 70uM

TP2a

<n

>

R71
a4

I
o- i

7 2

n
L

v
c

j k? uiza 12
IP3

021

12t(

Rt2

OVEN % T Ml

MOO

BID

2SK

3 . 4 ex

S
R63 C?
*

E2V S

lODw

1
C3

VS5

I T

c i

lOD

6q2oo3

0-15S30J D-1SS102
*JK7aLOSU*

NJMT9LOSUA

2SJ24S

OS-CON

T P45

f f
1 I T )

1
2

out OUf
GWD

I i
2 -

COW COHBON
IN 1

?ft
~ CA'C

In

Out

2 3

[>R*IN SOURCE
DRA1n

TITLE

DRAWING NO.

Main Board Circuit Diagram, 7/9


MODEL NO. ASSEMBLY NO.

523-2020
DATE

FX-2111

PCB-6409

95.01.08

6-17

P7

0637015 1000600
nop E
i
-

CKY0CERO

ELC0>

SCNSI
-

2 3
4

2
3
4

NC

SOU
SZL

s
-

s
6
7

NC

i
7

(key?
_

>
P * S r fc

ST CRT /STOP
-

Oe

4XEY3

~>

1 mV
-

G 9 1 0
1 1
-

fKE Y4
(KEYS

>

O-

INST
-

10
11

(OF F /CMC
-

1 >
_> >
_

O-

P9*

OFF/CHG
-

to

O
ON

12
13

12
L 3
11

{X E Yfc

>

P*9 c

&

NC NC

11
-

E2

IS

IS

rxr2A.U..KKT. poncu

KEY

PRNEL

TITLE

DRAWING NO.

Main Board Circuit Diagram, 8/9


MODEL NO.

523-2020
DATE

ASSEMBLY NO.

FX-2111

PCB-6409

95.01.08

6-19

nxi 2 1 1 90 I
.

P2 n xg j i4qqi
VDO

6 LCO-PCB Vtt
VOO
L 1

0F3-1SS-2C

PS 0F3-15P-2DS
1

<vS5

15
l->

1 5
1

DOT ft r N

2
3

2
3
4

2 3

<i b r ch

VCC

13
-

1 3

CLOCK POT POUT

-rCLOCK

Page

<sr
<R/M

12
1 1

1 2
1 1 10

R /w
_

5 6

STROBES

s
6 7 8 9
-

<e:

10
g

T ST Ro6E

"

ooo
oe i 002 DB3

9 8
7

STW08CI

7
'

8
9
lO

8
9

OB L

THERMIST OR
VDD

B 9

tirnp

>

P .gp

DB 2 0B3
0B4 PBS

JO
1 1 1 2 1 3
1 4

6 5
4

CNQ
CNO GNO
connoN COMnON ConnpN
~ -

10
1 1

10
I 1

081
OPS 086 DB7

1 I

T
E2

C467 0 1
.

12 13
1 <

12 L3
14 IS

1 2
L 3
1 4
-

P a<jr

P85
0B7

3
2

3 2
1

<COMnON

Page

1 S

LCD MODULE

Cpgo-oBfl

Page?

S ISB-Ph-k.-S;

ph5

iS

THERMAL
MODULE

HEPD

TITLE

DRAWING NO.

Main Board Circuit Diagram, 9/9


MODEL NO.

523-2020
DATE

ASSEMBLY NO.

FX-2111

PCB-6409

95.01.08

6-21

o
o

n DD

CD

?
O
KJ

cn
i.

I?
i
c

cnn Dm
Z
r
m

ML/1

I; > !;<
"

n
9

CP o
D XI
O

1
9

s2

n B

I (") C")
B

TITLE

DRAWING NO.

Power Board Circuit Diagram


MODEL NO. ASSEMBLY NO.

524-2021
DATE

FX-2111

PCB-6410

96.01.08

6-23

I
L

s r>

3 5
-

IN

DRAWING NO.
TITLE

Sensor Board Circuit Diagram


MODEL NO.

524-1959
DATE

ASSEMBLY NO.

FX-2111

PCB-6239
6-24

94.07.14

NF-26P . /2'i
.

M
WED

r: i

to

n72 n; 3
MT 4

Or3i>-SP-2D50

ELU

s
6

MOTOR MODULE

TITLE

DRAWING NO.

Motor Module Circuit Diagram


MODEL NO. ASSEMBLY NO.

524-1960
DATE

FX-2111

94.07.14

6-25

VHfl-4M (VHfi-ON

H
9 6V
.

HOOmOh

3
4

-I H H H

'

I
r

Ni-MH

BATTERY

PRCK

6-HRflRFD

TITLE

DRAWING NO.

Battery Pack Circuit Diagram


MODEL NO. ASSEMBLY NO.

524-1961
DATE

FX-2111

94.07.14

6-26

F C P-2 2 2 2 A T

INLET BOARD

PI

LCD

P10 Ni-MH
i

PZ

0
MOTOR

P2

BATTERY PACK
P3
DC [ N

0 0
PC8-6410

NF-26P1/25
MOTOR CABLE

P4

MA I N
THE RMAL HEAD

B 0 A RD iP8

THEMAL CABLE

P5 PATIENT

CABLE

/
Q56-F
P6 P7

S W

PANEL

SENSOR

BOARD

F X - 2 1 1 1

PCB-6239

TITLE

DRAWING NO.

Overall Connection Diagram


MODEL NO.

433-5126
DATE

ASSEMBLY NO.

FX-2111 (USA)
6-27

96.10.09

FCP-2222W INLET BOARD

PI

L CD

: pio
i

P2

P2

Ni-MH BATTERY PACK


P3

u PCB-6410
v
.

MOTOR

DC TN

NF-26P1/25
MOTOR CABLE

P4

MA IN THE RMAL
HEAD

BOARD
P8

THEMAL CABLE

P5

PATIENT

CABLE

Q56-F

P6

P7

SW

PANEL

SENSOR

BOARD

F X-2 1 1 1

PCB-6239

TITLE

DRAWING NO.

Overall Connection Diagram


MODEL NO.
ASSEMBLY NO.

4334894
DATE

FX-2111 (CE)
6-29

96.01.08

SOLDER

NDM202AOO
14

>
33

AXL214901

(7)
m

TO

F4

TITLE

DRAWING NO.

LCD Connection Diagram


MODEL NO.
ASSEMBLY NO.

434-4895
DATE

FX-2111

96.01.08

6-31

MOTOR

NF-26P1/2 5

r
10

SUMITUBE F2 3

xO. 25

SDMI TUBE F2 6x 0. 25

o
HIRING
SOLDER

1 0 0 mm

(BRN)

HIRING
SOLDER

(RED)

HIRING

UL1007
BRN
RED ORG YEL BLU PUR

AHG28
6
1
connector

HIRING

LENGTH

1 6 Omml pin 14 0 2 pin


1 7 0

DF3-6S-2C
contact

3 pin 4 pin Spin Spin

DF3-24 28SC

TITLE

DRAWING NO.

Motor Connection Diagram


MODEL NO.
ASSEMBLY NO.

434-4896
DATE

FX-2111
6-32

96.01.08

CHAPTER

laqram
Main Board Assembly Diagram, 1/2 & 2/2 433-4897 Inlet Board Assembly Diagram 434-4899 7-1 to 7-3 7-5

HOW 10 MOUNT IC22 (PQ30RV11)

COMPONENT SIDE
PCS-6X09

C35

SILK CHART

is FLUX AREA .

65

CI

rfi
ZD

.:iiiiiiisiTnfria
fp
su

5
? o

ir

a
z a
u-l
1 n

=s----flm?ji0

1 c

D
IC4

EES)
I
lr_

PCB-6409
..

lO"'" FUKUOA DEN5HI CO


(
.

how TOMOmr ic socket f en


LTD.

I *

i-i

- "li-=nt-rJ rn0 31

5
0.5

(ZDc
_

ICIO

AT

225

D
CI

.2

HOW TO MOUNT 04 .at

2 M n
B1
1 3
.

(2SJ245)
2J

B2<

= x x X 1 I I 1
-

s [v' Pi LED ys' L I .i-l


-

m AREA :
PC BOARD
MAX OPEN

5
icia

(MOUNT

IC1 :512K

IC1 : 1M)

Li

HOW 7t?MQUNT Cft? CECHEW105JZ)

LEAD CUT LIMIT


St

MAX 1 5 mm
.

IC31

TITLE

DRAWING NO.

Main Board Assembly Diagram, 1/2


PC BOARD
MODEL NO.

433-4897
DATE

ASSEMBLY NO.

FX-2111

PCB-6409

95.12.15

OMPOfCNT SIDE
7-1

is FLUX AREA .

tV

jr''

W AREA ARE NOT REQUI

TITLE

DRAWING NO.

Main Board Assembly Diagram, 2/2


MODEL NO.

ASSEMBLY NO.

DATE

FX-2111

PCB-6409

0
m

>
X)

>
CD

6
CO o
33

5
CO
m

>

ft
s

t
CD

CD 4

5
LTI

CD

O
< ro 30
-

O o

o o

6
O
cz

3
mu

O
z
CO
m

-i

be
3 -I

TITLE

DRAWING NO.

Inlet Board Assembly Diagram


MODEL NO.
ASSEMBLY NO.

434-4899
DATE

FX-2111

PCB-6410

96.01.08

7-5

CHAPTER

8
8-1

Electrical Parts Lists


1 FX-2111 Main Blocks
.

674-11338

2 Main Board, PCB-6409 SAS SMD for FX-2111


.

674-11339
674-11340

8-2
8-9

3 Main Board, PCB-6409 THD for FX-2111


.

AC Inlet Board, PCB-6410 THD for FX-2111 (115V)... 674-11664

8-11

5 6 7

AC Inlet Board, PCB-6410 THD for FX-2111 (230V)... 674-11340


LCD Assembly for FX-2111 Parts Supplied to Wiring/Assembling Process
for FX-2111

8-12
8-13
8-14

674-11342
674-11343

1 FX-2111 Main Blocks


.

674-11338

Parts No.

Symbol

Description

Drawing No.

Q'ty

Remarks

9E3041

PC Board AS

Main Board
9F3419

PCB-6409 with ROM

9F3293
9F3424 9G3739 9G3740

ROM Set
PCB-6409 SAS PCB-6409 SMD PCB-6409 THD

674-10486

674-11339

674-11340

AC Inlet Board
9F3420

PCB-6410 Assembly

674-11341

9F3421

LCD Assembly

674-11342

9D0995 9E2793 9E3042 9D0996

Set of Supplied Parts to Assembling Process (CE)


Set of Assembled Electrical Parts (2307)
674-10513 674-11359 674-11343

1
1

Set of Assembled Electrical Parts (CE)

Set of Supplied Parts to Wiring Process (CE)

8-1

2 Main Board, PCB-6409 SAS SMD for FX-2111


.

674-11339

Parts No.

Symbol

Description

Drawing No.

Q'ty

Remarks

Resistors

1A7510A

R1, 15, 23, 75-77, 145, 163, 168, 176, 181, 192, 197, 200, 202, 207, 209, 212, 219, 220, 226, 227, 229, 230, 232, 236, 239, 242,
245

RK73H2AF, Square Chip

lOkfi

30

1A7503A

R2~11, 14,

RK73H2AF, Square Chip

20kfi

51

26, 29, 30, 34-40,


43, 46 57, 68-70, 86-92, 140, 158, 182, 183, 189-191,
,

195, 201, 210, 225,

196, 204, 211, 241,

254, 255

1A7685A 1A9662A

R12, 84, 110 R13, 41, 51, 72, 80, 199, 206,
215

RK73H2AF, Square Chip RK73H2AF, Square Chip

12kG 30kn

1A7641A

R16, 117,
148

RK73H2AF, Square Chip

560kn

1A9666A

R17, 32, 58, 116, 169, 173, 231,


252

RK73H2AF, Square Chip

200kn

1A9657A

R19, 59, 119, 164,


222

RK73H2AF, Square Chip

3kil

1A9655A

R20, 28, 64, 65, 67, 82, 93-100, 105, 109, 132, 133, 136, 144, 150, 156,
217, 248

RK73H2AF, Square Chip

1kfi

14

8-2

2 Main Board, PCB-6409 SAS SMD for FX-2111. continued


.

674-11339

Parts No.

Symbol

Description

Drawing No.

Q'ty

Remarks

1A7512A
1A7502A 1A9263A
1A7522A

R85, 147 R22


R24

RK73H2AF, Square Chip


RK73H2AF, Square Chip
SR73K2EJ, Metal Film

16kn
6 8kG
.

2
1

0 1fi/0.5W
.

R25, 63, 81,


141,142

RK73H2AF, Square Chip

lOOkfi

1A7545A

R27, 61, 103, 171, 174, 214, 216, 228,


249, 250

RK73H2AF, Square Chip

5 1k0
.

10

1A7571A
1A9631A 1A7534A

R31 R33 R42, 62

RK73H2AF, Square Chip

300fl
100 3 48kU
.

1
1

RK73K2AF, Square Chip


RK73H2AF, Square Chip

2 11

1A7584A

R44, 45, 56, 78, 124, 125, 165,


166
,

RK73H2AF, Square Chip

1Mfi

218,

244, 246 1A9653A

R47, 48, 55, 74, 115,


161

RK73H2AF, Square Chip

1001]

1A9656A

R49, 50, 52, 71, 73, 126-129, 143, 159, 178, 179,
187

RK73H2AF, Square Chip

2kU

14

1A9396 1A9398
1A9303A 1A9271 1A7681A R7519A

R53
R54 R60 R66 R79

RN73F2ATD-B, Metal Film

RN73F24TD-B, Metal Film

RH73H2AF, Square Chip


RK73H3AF, Metal Film

28.7k(] IkO/IW
2 2kQ
.

1
1

RK73H2AF, Square Chip


RK73H2AF, Square Chip

R83, 146, 160, 193,


198, 205, 213, 223, 240,
247

51 kl)

15

203, 208, 221, 237, 243,

8-3

2 Main Board, PCB6409 SAS SMD for FX-2111., continued


.

674-11339

Parts No.

Symbol

Description

Drawing No.

Q'ty

Remarks

1A7547A 1A7656A

R85, 147

RK73H2AF, Square Chip RK73H2AF, Square Chip

16kn

2 3

R101, 151,
162

240ki]

1A7558A 1A7689A

R104, 122

RK73K2AF, Square Chip RK73H2AF, Square Chip

42.2fl 82kn

R106, 108, 111, 155,


167

1A9665A 1A9397
1A7674A

R107 R112~114

RK73H2AF, Square Chip


RN73F2ATD-B, Metal Film

120kQ 78.7ki] 430n

3 5

R118, 121, 123, 149,


152

RK73H2AF, Square Chip

1A7652A

R120, 131,
157, 238

RK73H2AF, Square Chip

13kn

1A7616A 1A9270A 1A7604A 1A7540A

R130

RK73H2AF, Square Chip


SR73H2EF, Metal Film

56kn 0 51Q/0.25W
.

R134 R135

RK73H2AF, Square Chip RK73H2AF, Square Chip

1 43kl
.

R137, 138,
233, 251

24kQ

1A7554A 1A7578A

R139 R153, 154


R170

RK73H2AF, Square Chip RK73H2AF, Square Chip

36kfi 150fi

1A7202
1A7591A

TRN60LG, Cylindrical Tantalum Alloy


3 5Mn
.

1 4

R175, 180,
234, 253

RK73H2AF, Square Chip

160kfi

1A7526A

R172, 177

RK73H2AF, Square Chip


RN73F2ATD-B, Metal Film
11011 140kfi 3 4kQ
.

1A9393
1A7640A

R184
R185, 186 R188 R194
R256~259 R224

RK73H2AF, Square Chip RK73H2AF, Square Chip


RN73F2ATD-B, Metal Film

2
1

1A7587A
1A9395
1A7668A 1A7528A

13kn 220n 698kH

RK73H2AF, Square Chip

RK73H2AF, Square Chip

8-4

2 Main Board, PCB-6409 SAS SMD for FX-2111, continued


.

674-11339
Parts No.

Symbol

Description

Drawing No.

Q'ty

Remarks

Capacitors
1B5021A

01,2, 4, 6, 7 11, 13, 14, 22, 25, 26, 28-34, 36, 37, 39, 40, 42, 43,
,

GRM40F104Z50, Layered Ceramic Chip

0.1 pF

59

45, 49, 50, 52, 53, 61 -63, 66-68, 73, 76, 77, 85, 86, 89,

93, 94, 96, 97, 99,


100,

105-108, 116-119,
125-128
1B1701 C3

ECSTCY105, Tantalum Electrolytic 1.0(jF EEV-H81CV220, Aluminum Electrolytic 22nF/16V EEV-HBIHV0010, Aluminum Electrolytic 1mF/50V GRM40CH1U2J50, Layered Ceramic lOOOpF

1B5412A 1B5408A 1B5073A

C5, 27 C8

2 1 7

C9, 16, 41, 56, 58, 87,


88

1B0702

C10

25CV100GX, Aluminum Electrolytic Chip 100uF/25V for High Frequency


GRM40B222K, Layered Ceramic 2200pF

1B5055A

C12, 51, 69,


134

1B0209
1B0208

C15
C17, 21

16SM6R8M, OS
10SM22M, OS

6.8mF/16V
22tiF/10V

1 2

1B5019A 1B5022A

C18

GRM40CH101J, Layered Ceramic


GRM40B103K, Layered Ceramic

100pF
O.OIpF

C18, 20, 57, 59, 60, 65,


75, 115

1B5062

C23, 24, 38,


47, 71

MCH215C223K, Layered Ceramic

0.022mF

1B5002A

C35

GRM40CH510J, Layered Ceramic

51pF

1B8501

C44, 90, 91

35CV47GX, Aluminum Electrolytic Chip for High Frequency 47pF/35V


ECHU1H102G, Layered Film

3
1

1B5815

C46

1000pF

8-5

2 Main Board, PCB-6409 SAS SMD for FX-2111


.

674-11339
Parts No.

Symbol

Description

Drawing No.

Q'ty

Remarks

1B5027

C48, 131,
133

MCH215A561J, Layered Ceramic

1B5816 1B5402A

C54

ECHU1C332G, Layered Film

3300pF

C64, 70, 95

EEV-HB1 EV4R, Aluminum Electrolytic 4 7nF/25V


.

3
2

1B5513A 1B5011A

C55, 83 C72, 74
C78

GRM42-6F224Z, Layered Ceramic

0.22nF

GRM42-6B473K, Layered Ceramic


ECHU1C683J, Layered Film ECWU1C224J, Layered Film
GRM40B332K, Layered Ceramic

0.047mF
0.068mF 0.22nF
3300pF

1B5818
1B5261

1 3 6

C79, 80, 92
C81, 100, 110, 112,
124, 135

1B5076A

1B5044A 1B5003A

C84

GRM40B153K, Layered Ceramic

0.015nF

C98, 103, 104, 113, 114, 120, 121, 123,


129

GRM40CH101J, Layered Ceramic

150pF

1B5403A

C101, 102

EEV-HB1EV330, Aluminum Electrolytic 4 7nF/16V


.

2 2
1

1B5808 1B5014C
1B5001A 1B5074A

C111,113

ECHU1H140J, Layered Film

O.lpF

C130
C132 C138

MCH325F274Z, Layered Ceramic GRM40CH300J, Layered Ceramic


GRM40CH331J, Layered Ceramic

0.47nF 30pF
330pF

Diodes 3929 1D4200 1B3885 1D3886 D1~6


D7

D3FS4A (2.6A, 40V)


02CZ-5.1V, Zener 1SS301 1SS302

6
1

D8, 1, 15, 32 D9, 11, 12, 16, 17, 19, 20, 23-31,
33, 34

3 18

1D4113

D10

02CZ-3.0K, Zener
D1FS4A

1D3928

D13, 14, 18,


21

8-6

2 Main Board, PCB-6409 SAS SMD for FX-2111, continued


.

674-11339
Parts No.

Symbol

Description

Drawing No.

Q'ty

Remarks

1D4552

D22

U1ZB12, Zener

Transistors
1D1735

Q1~3, 5, 8, 12 14, 15, 27, 29, 30,


,

UN5212

13

33, 38

1D989
1D0116

Q6
Q7
Q9

2SC-2713GR
2SA-1163G 2SK-739-2, FET
2SK-1062, FET 2SA-1162V

1 1

1D2385
1D2392 1D0117 1D1734

6
1
1

Q10
Q11

Q13, 18, 26, 28, 42, 43


Q16
Q17

UN5112, Digital

1D0968

2SC-1623(L5/L6)
2SJ245S, FET
SST4393, FET

1D2394
1D2382

Q19, 22, 23, 31, 32, 34


Q20, 21
Q25

1D2342 1D0153
1D1713

2SK1920-FA, FET

2
1

2SA-1463(1 K)
FMW2

Q35~37, 39-41,
44-46

Integrated Circuits
1E5037A
1E2769 1E6168 1E6245 IE2621 1E6235 1E7205
1E5390A

IC2, 5
IC4 IC6

HM6264BLFP-10LT, RAM
NJM2103M, Reset IC

2
1

HA16120FP, Switching Regulator


PQ05SZ5, Low-loss Regulator
FD88007-AC, Gate Array

IC7
ICS IC9, 33
IC10 IC11

1
2

NJM78L05UA, 3-terminal Regulator


S-8437AF
,

Inversion-type Switching Regulator

TC-74HC245AF, CMOS Logic

8-7

2 Main Board, PCB-6409 SAS SMD for FX-2111, continued


.

674-11339

Parts No.

Symbol

Description

Drawing No.

Q'ty

Remarks

1E4424

IC12, 19 IC13 IC14, 23


IC15 IC18 IC16

TC-4584BF, CMOS Logic

2 1
2

1E0374 1E5397A
1E5412A 1E6167

HPC4074G2, Low-noise FET-input OP Amp.


TC-74HC04AF, CMOS Logic TC-74HC541AF, CMOS Logic bq2003S> Quick Charging IC TC7W04F, L-MOS Logic

1E5543
1E4169

IC17
IC20

TC-4053BF, CMOS Logic


TC-74AC74F, CMOS Logic

1E5595
1E0198

IC21, 31

pPC4064G2, Low Power CoSsumption J-FET Input Operational Amplifier pPC4064G2(2), Low Power Oonsumption J-FET
Input Operational Amplifier

1E0715

IC24

1E2613A 1E0266 1E5542 1E8961 1E2083 1E7221 1E4177 1E4170

IC25 IC26
IC27

HEF4046BT, PLL

MPC358G2, Comparator
TC7W02F, L-MOS Logic

1 3

IC28~30 IC32 IC34 IC35

PS2652L2-V(K), Photo Coupler

LT1004CS8-1.2, Voltage Reference


NJM79L05UA, 3-terminal Regulator TC-4093BF, CMOS Logic TC-4051BF, CMOS Logic

1 1

IC36, 38, 41,


45

1E0264

IC37, 39, 42

MPC4064G2(2), Low Power Oonsumption J-FET Input Operational Amplifier


TC-4535BF, CMOS Logic pPC339G2, Comparator TC-4094BF, CMOS Logic
1
1

1E4423 1E0365 1E4182

IC40
IC43 IC44

Coil
1T6047 L2

LPC4045TE-471K

470mH

1H6409

Printed Circuit Board, PCG-6409

8-8

3 Main Board, PCB-6409 THD for FX-2111


.

674-11340

Parts No.

Symbol

Description

Drawing No.

Q'ty

Remarks

Resistor 1A7040
R18

RSSX3L20 0.01J/3W, Metal Film

Capacitors
1B4701 1B0866 1B0713 C82 IC136 CI37

ECHE1H105JZ, Metalized

URS1E222MHZ 2200/25V, Aluminum Electrolytic URS1V102MHZ 1000/35V, Aluminum Electrolytic

Transistors 1D2393

Q4, 24

2SJ245L, FET

Crystal Oscillator
1D7620 XI

JX0-5S 16MHz (with stand)

Integrated Circuits
1E1957
1E7021 1E9198 1E9199

IC3
IC22

HD63B09E, CPU

PQ30RV1/11, Low-loss Regulator


DIC?iG11S1 IC Socket
,

DIC640G11S1, IC Socket

Lithium Battery
1U0231 BT1

CR-1/3-FT2-1

684-2428

Connectors
1F0215 1F1068 1F3208 1F0562
1F9376

P1 P2
P3 P4 P5
P6

B3P-VH AXB114001 BB S-G8036 #01


DF3A-6P-2DSA

DF3A-15P-2DSA 008370061000800 008370151000800

1F0548
1F1131

P7

1F1004

P8

RDAG-15SE1(F)

8-9

3 Main Board, PCB-6409 THD, continued


.

674-11340 Parts No.

Symbol

Description

Drawing No.

Q'ty

Remarks

1F0365

P10

B4P-VH

Arrestor

1L4807A

NE1

Y06S-230B

Fuses
1L4057

F1, 2

TR-5 K19374

3 15A
.

Coils & Ferrite Core


1T6242
L1

HK-05S035-2010 RBP HK-04D030-1510 RBP BL01RN1-A62, Ferrite Core

200mH 150nH

684-2109

1T6243
1L4584

L3
L4

684-2108

Isolation Transformers
1T6015A T1 T2 FT-2111 684-2153 684-2154
1

1T6016A

FT-2155

Piezoelectric Buzzer 1S3200


BZ1

PKM13EPP-4002

8-10

AC Inlet Board, PCB-6410 THD for FX-2111 (115V)


674-11664

Parts No.

Symbol

Description

Drawing No.

Q'ty

Remarks

Fuses
1L4476

239001 MF-561A, Fuse Holder

1A

1L4905

Locker Switch

1G3508

SJ-2S4A-07BB

AC Inlet 1D2393 NC176-1.5


1

GND Terminal
1K1151

FX-2111GND (A)

Connector
1F0159 B2P-VH
1

Choke Coil
1T6342

TF1028S-801Y2R0-01

Spacer
1Z4311

LR-1.5

1/2

Printed Circuit Board


1H6410 PCB-6410 323-3961
1

8-11

AC Inlet Board, PCB-6410 THD for FX-2111 (230V)


674-11340

Parts No.

Symbol

Description

Drawing No.

Q'ty

Remarks

Fuses

1L4091 1L4905

No. 19195

500mA

2 2

MF-561A, Fuse Holder

Locker Switch

1G3508

SJ-2S4A-07BB

AC Inlet
1D0288 NC176-1.5
1

GND Terminal 1K1151A

GND (A)

314-3471

Connector

1F0159

B2P-VH

Choke Coil

1T6342

TF10283-801Y2R0-O1

Spacer
1Z4311 LR-1.5
1/2

Printed Circuit Board


1H6410

PCB-6410

8-12

LCD Assembly for FX-2111


674-11342

Parts No.

Symbol

Description

Drawing No.

Q'ty

Remarks

LCD

1 LI 709

NDM202A00

Connector
1F1085 AXL214901
1

Spacers
6B8308 SP-15
2

6B9216

Grounding Piece

Shielding Fingers
6B9217

(A) (B)

1 1

6B9218

8-13

Parts Supplied to Wiring/Assembling Process for FX-2111


674-11343

Parts No.

Symbol

Description

Drawing No.

Q'ty

Remarks

1M0385 1M0466 1M0482

Motor Assembly

434-4725

Inlet Board Grouding Cable 3 with Core

Inlet Board Grounding Cable 3


Ferrite Core, 5977-000601

684-2415

1L4581
6G0234 1K0031

Heat Shrink Tube, F2 25x0.4, 30mm long


Press-fit Terminal, TMEV1.25-3

1 1

8-14

CHAPTER

9
9-1 9-2

Spare Parts Lists


PCB-6238 2 PCB-6236
1
. .

674-10554 674-10555

3
4

Paper Magazine
Recorder

674-10556
674-10557

9-3
9-4

5 6

Upper Case

674-10558
674-10559 674-10560

9-5
9-6 9-7

Lower Case 7 Screws


. .

1 PCB-6238
.

674-10554
Parts No.

Symbol

Description

Drawing No.

Q'ty

Remarks

9E2963
9F3331 5T3111 5T2009

PCB-6238B (2) Assembly for 115V PCB-6238 Assembly for 115V Nuts, Type 3 M6

2
1

Spring Washer, Nominal Dia. 6mm

9E2964 9F3310 5T3111


5T2009

PCB-6238B (3) Assembly for 230V


PCB-6238 Assembly for 230V Nut, Type 3 M6

1 1

2
1

Spring Washer, Nominal Dia. 6mm

1L4476 1L4475

Fuse, 239001 for 115V


Fuse, 239500 for 230V

2 2

9-1

2 PCB-6236
.

674-10555
Parts No.

Symbol

Description

Drawing No.

Q'ty

Remarks

9E2965 9F3307 5R1604

PCB-6236 (1) Assembly

PCB-6236 Assembly for FX-2111(N)


Double Semus Screw, M3x6

9E2966 9F3309
5R1604

PCB-6236 (2) Assembly

PCB-6236 (with ROM) for FX-2111 (N)


Double Semus Screw, M3x6

1E8531 1L4057 1U0231 9F3305


9F3305

Programmed ROM for FX-2111 (36-202)


Fuse, TR-5 K19374, 3.15A

2 684-2428 634-4726
1

Lithium Battery, CR-1/3-FT2-1


LCD Assembly

Spacer, SP-15

9-2

Paper Magazine
674-10556

Parts No.

Symbol

Description

Drawing No.

Q'ty

Remarks

9H0402 6B8324A 6A5350 6B8325 5R3003 9F2648 6B8323 9H0402

Platen Roll Assembly


Platen Roll 224-0890

Bearing, LF-740ZZ
Gear, 38A Countersunk Screw, M2x6
224-0313

2
1

Paper Magazine Assembly Paper Magazine Platen Roll Assembly


222-0889

9-3

4 Recorder
.

674-10557
Parts No.

Symbol

Description

Drawing No.

Q'ty

Remarks

9H0617 6B8320 5R8001


9H0812

Gear 38B Assembly


Gear, 38B Setscrew, M3x3
244-0314

Thermal Array Head Assembly


Thermal Array Head, Q56-F Thermal Array Head Mounting Plate Tension Coil Spring, DE542
Semus Screw, M3x5 684-2092 213-3448

6B8316
6B8316A

6B8394 5R0602 9F3354


9H0812

2
1

Recorder Assembly

Thermal Array Head Assembly


Recorder Chassis SU

9H3215 1M0385 9H0617


9H3216 6B8307 6B8395

Motor Assembly Gear 38B Assembly

Grounding Piece (C) SU

313- 3461 6B8307 314- 3465

Grounding Flexible Conductor Sensor Presser Sponge


PCB-6295SMD Sensor Board

9F3289B
1M0381

674-10490
684-2093

Thermal Array Head Cable


Double Semus Screw, M3x4 Double Semus Screw, M3x8

5R1851 5R1604 9H3522

Paper Shaft SU

224-0891

9-4

Upper Case
674-10558

Parts No.

Symbol

Description

Drawing No.

Q'ty

Remarks

9F3355

Upper Case Assembly

9H3211C
6B8305 6B9171

Upper Case SU
LCD Filter

111-6621 114-6625 114-6811 684-2146 154-4573 154-4575

Lock Spring
Key Panel

1G9340A
5H9128 5H9130

Rated Power Label (B) for 115V


Rated Power Label (C) for 230V

9-5

6 Lower Case
.

674-10559 Parts No.

Symbol

Description

Drawing No.

Q'ty

Remarks

9F3356 9H3212C 6B8306 6B8327 9H3213A 1T1558 1T1559 6B8326A


5H9127

Lower Case Assembly (1)


Lower Case SU Insulation Sheet
111-6622 314-3463 314-3466 313-3459 684-2158 684-2157 112-6624 154-4572 154-4577 154-4574
154-4576 154-3574

Batter Presser Sponge

Grounding Piece (A) SU


Transformer, FCP2222AT for 115V Transformer, FCP-2222UT for 230V

Battery Cover
Instructions Label (B)

5H9138 5H9129
5H9131

Instructions Label (C) for 230V Fuse Label (B) for 115V Fuse Label (C) for 230V Lithium Battery Label
TM-166 No. 12

5H5639 6B6796

9-6

7 Screws
.

674-10560
Parts No.

Symbol

Description

Drawing No.

Q'ty

Remarks

9H3238 5R1604

Set of Screws for Ni-MH Double Semus Screw, M3x8

1 24

5R0602 5R1851 5R3003


5R8001

Semus Screw, M3x5


Double Semus Screw, M3x4 Countersunk Screw, M2x6
Setscrew, M3x3

2
1

9-7

CHAPTER

10
10-1 10-3
10-5

Structural Diagrams
FX-2111 (USA) External Appearance (1) External Appearance (2)
AC Inlet Board

413-0994 413-0995
424-3619

Upper Case Lower Case Assembly (1) Lower Case Assembly (2) Lower Case Assembly (3) Exploded View FX-2111 (CE) External Appearance (1) External Appearance (2)
AC Inlet Board
Main Board*

423-3620 423-3621 422-3622 422-3623 412-0996


413-0849 413-0927
424-3443
423-3444

10-7 10-9 10-11 10-13 10-15


10-17 10-19
10-21
10-23

>

Paper Magazine Assembly* Paper Shaft Assembly* Recorder Assembly* Upper Case Assembly Lower Case Assembly (1) Lower Case Assembly (2) Lower Case Assembly (3) Exploded View

423-3250 424-3251 423-3252 423-3445 423-3446 423-3447 422-3448 412-0928

10-25 10-27 10-29 10-31 10-33 10-35 10-37 10-39

Asterisked diagrams are common to both FX-2111 (USA) and FX-2111 (CE).

3sg

II 8
.S5

ii
A
Of

as

TO

[J
a

TITLE

DRAWING NO.

External Appearance (1)


MODEL NO.

413-0994
DATE

ASSEMBLY NO.

FX-2111 (USA)
10-1

96.09.30

JUL

JUL

0)
ClITIOI-Ti II.t r iin til It r i 11 if iti(tci| t*tIi(i d if f i 111

Q
dtclric ibid, ti ul rii*Tc tirtr. firiooi. 1,

D
DASOBI-liil *f ciflniiB if ttti
Ii

ii
pn

the grtietct tf
dpUr*' ei

<1

flmtblt

ittilkitici.
'

DA NG B K - 1 i ii it

d'tiplntoi.

o -

pruiii* d lBtltkttilstt

ilflilBikltl.

ruse

2jnv

51

10DE

HIJITIIltT

[ Vrjcuwdensh)
C A R D I MA X
Pi- 2 111
HAD

$1

START
Dl DFF

1ISBT

taV

@
V A

0
a

<3

Mil
o

IS

@
STOP

3J
2 6 0

CA8T10R:FBD8RAl Ul SISTIICTS T8IS DtVICt TO SALS BT 01 01 71! 0IDI1 01 A PB1S1CIA1.

Tizr

6 5mm
DRAWING NO.

TITLE

External Appearance (2)


MODEL NO.

413-0995
DATE

ASSEMBLY NO.

FX-2111 (USA)
10-3

96.09.30

\
a*
o

2-M8 Nut & Plain Washer

Key
No.

Parts No.

Drawing Na

Description
Fuse, 239001

Q'ty

Remarks

1L4476 1K1151A 314-3471

2
1

GND Terminal (A)

TITLE

DRAWING NO.

AC inlet Board
MODEL NO.

424-3619
DATE

ASSEMBLY NO.

FX-2111 (USA)

PC 6-6410

96.09.30

10-5

3
6

''

A
-

55

33

ti

Si'S
II

Of

S!|!Li

TO
8

Key
No.

Parts Na

Drawing Na

Description
Upper Case Assembly

Q'ty

Remarks

1 1 1
TITLE

2 3
4

9H3494 1G9340A
6B8305 5H9128 6B9171

111-6862
684-2146

Upper Case Assembly SU for USA


Key Panel
LCD Filter

DRAWING NO.

114-6625 154-4573

Upper Case
MODEL NO.

423-3620
DATE

Rated Power Label (B)

1
1 1 ASSEMBLY NO.

6
7

114-6811
154-4670

Lock Spring
Caution Label

5H9237

FX-2111 (USA)
10-7

96.09.30

4
2-Double Semus Screw M3x8
,

2-Taping

Pan-head Screw M2.6x6


,

6
A

A
C

3
Pasted

2
A

C
f_

TV

Key
Na

Parts Na

Drawing Na

Description
Lower Case Assembly (1)

Q'ty

Remarks

2 3
4

9H3495 6B8327

111-6863 314-3466
112-6624 154-4669

Lower Case Assembly SU for USA Battery Presser Sponge


Battery Cover
Fuse Label (D) Rating label

1
TITLE

DRAWING NO.

1 1
1 1
MODEL NO.

Lower Case Assembly (1)


ASSEMBLY NO.

423-3621
DATE

6B8326 5H9374
9E7329

5 6

FX-2111 (USA)
10-9

96.09.30

Key
No.

Parts No.

Drawing No.

Description

Q'ty

Remarks

Lower Case Assembly (1)


423- 3621 9H3213A 1T1558 9F3420 6B8306 6B9198A

1 1 1

Lower Case Assembly (1) Grounding Piece


Transformer, FCP2222AT

3
4

313- 3459 684-2156 424- 3619


314- 3463 314-3601

1 1 1 1

5 6
7

PCB-6410 Assembly
Insulation Sheet Insulation Cover

.v

3
B

3
o

0
6

TITLE

DRAWING NO.

Lower Case Assembly (2)


MODEL NO.

422-3622
DATE

ASSEMBLY NO.

FX-2111 (USA)
10-11

96.09.30

9-Double Semus Screw M3x8


,

IF
D

D
A G

it
5

E B

F O

5
11

0
-1 p

6
V

Key
Na

Parts Na

Drawing Na
422- 3622

Description

Q'ty

Remarks

1 2

Lower Case Assembly (2)

9F3354
1M0381

423- 3252 684-2093


423-3444

Recorder Assembly

1 1
TITLE
DRAWING NO.

3
4

Thermal Array Head Cable


PCB-6409 Assembly Lead Label (C)

1T1558
5H9615

1
1

Lower Case Assembly (3)


MODEL NO. ASSEMBLY NO.

422-3623
DATE

154-4836

6
7

5H5639
5H9417

154-3574 154-4821

Lithium Battery Label


Patient Label

1 1

FX-2111 (USA)
10-13

96.09.30

5S>
'

II
D
At
r

3*

Li
6

Double Semus Screw


M3x8

ii
s"

2-D-jubl

e Semus Screw M3x8


,

11

Key
Na

Parts No.

Drawing Na

Description

Q'ty

Remarks
DRAWING NO.

TITLE

423-3623 423-3620

Lower Case Assembly (3)

Exploded View
MODEL NO.

412-0996
DATE

2 3
4

Upper Case Assembly Paper Magazine Assembly


Paper Shaft SU

1
1 1

9F2648 9H3522

423-3250
224-0891

ASSEMBLY NO.

FX-2111 (USA)
10-15

96.09.30

2?5

>5-

TITLE

DRAWING NO.

External Appearance (1)


MODEL NO.

413-0849
DATE

ASSEMBLY NO.

FX-2111 (CE)
10-17

94.10.07

o
Cilil01-Ti TtAtEi tki nil f ilttltit tkit t* tit rtiif* ititt.

/Ft,

Itfir

itttitin

11

tiflid

nrvlci

pt r 11 d i * 1.

u
DkHQtl'litl If tlfltiitt if tie* ii tk* frimci sf
f 11 mm t k It (inlbitiei.

D A B C E S -1 i n *'h|Iiiii. It in tMplertf > ptititu d inttl*Miii iifliaiiMn.

n
I 2TI
;

5i
CA R D I MA X
FX-Z11 J
IBID
PIT OTf

oo

1DDI

UISITITITT

5 I

STAKT

0
] a

11 oil:?:
"

'

!=.

ii

si

A
STOP

3J
v
.

2 6 0

nzr

TITLE

DRAWING NO.

External Appearance (2)


MODEL NO. ASSEMBLY NO.

413-0927
DATE

FX-2111 (CE)
10-19

96.01.24

2_M8 Nut & Plain Washer

T
T
1

Key
No.

Parts Na

Drawing Na
Fuse, 239500 314-3471

Description

Q'ty

Remarks

1L4475 1K1151A

2
1

GND Terminal (A)

TITLE

DRAWING NO.

AC Inlet Board
MODEL NO. ASSEMBLY NO.

424-3443
DATE

FX-2111 (CE)

PCB-6410

96.01.24

10-21

5
7
r

2-Binding

Screw M3x8
,

CD

.j
3
o o

3
0

v.

6
y
y

0
y

4
y

y
y

y y

0
y
y y

2-Double Semus Screw M3x8


,

Key
No.

Parts No.

Drawing No.
PCB-6409 SAS

Description

Q'ty

Remarks

9F3424
9F3419 1L4057

PCB-6409 with ROM Fuse, TR-5 K1937, 3.15A 684-2428

3
4

2
1
1 1 1
TITLE DRAWING NO.

1U0231
1E8631
9F3421

Lithium Battery, CR-1/3N-FT-2-1

5
6 7

Programmed ROM, FX-2111 (36-202) LCD Assembly


LCD, NDM202A0O

1L1709 6B8308 6B9216


6B9217

8
9
10
11

Spaeer, SP-15
314-3602 314-3603 314-3604

2
1

Main Board
MODEL NO.

423-3444
DATE

Grounding Piece (D)

Shielding Finger (A)

1
1

ASSEMBLY NO.

6B9218

Shielding Finger (B)

FX-2111 (CE)
10-23

96.01.24

5
Countersunk Screw M2x6
,

J6
1

2 3

Key
Na

Parts No.

Drawing No.
222-0889

Description
Paper Magazine Platen Roll Assembly

Q'ty

Remarks

6B8323 9H0402 6B8324A 6A5350 6B8325

1 1

2
3 4

Platen Roll

LF-740ZZ Bearing
224-0313 Gear, 38A

2
1

TITLE

DRAWING NO.

Paper Magazine Assembly


MODEL NO. ASSEMBLY NO.

423-3250
DATE

FX-2111

94.10.01

10-25

Key
No.

Parts Na

Drawing Na
223-0892

Description
Paper Shaft

Q'ty

Remarks

6B8321 6B8322

224-0893

Back Tension Spring

TITLE

DRAWING NO.

Paper Shaft Assembly


MODEL NO. ASSEMBLY NO.

424-3251
DATE

FX-2111

94.10.07

10-27

2-Double Semus Screw M3x5


,

1
1
4

Setscrew, M3x3

)
a

1
v
.

Key
Na

Parts Na

Drawing Na
211-3447 434-4725

Description
Recorder Chassis SU

Q'ty

Remarks

9H3215

2
3

1M0385
9H0617 6B8320 9H0812

Motor Assembly Gear 38B Assembly

1
1 1 1 1 1

1 1 9
Double Semus Screw M3x4
,

224-0314

Gear 38B

5 6

Thermal Array Head Assembly


213-3448

6B8316A
6B8317
1W3006

Thermal Array Head Mounting Plate Tension Coil Spring, DE511

684-2092
674-10490 313- 3461 314- 3465 314-3464

Thermal Array Head Q56-F


,

1
1 1 1 1

9 10
11

9F3289B

PCB-6239 SMD, Sensor Board

9H3216
6B8395 6B8307

Grounding Piece C SU

Sensor Presser Sponge


Flexible Grounding Conductor

2
D
3-Double Semus Screw, M3x8

12

TITLE

DRAWING NO.

Recorder Assembly
Double Semus Screw,
M3x8

423-3252
DATE

MODEL NO.

ASSEMBLY NO.

FX-2111

94.10.07

10-29

II

II
It

.St

.4>

2
4

Key
Na

Parts Na

Drawing Na

Description Upper Case Assembly

Q'ty

Remarks

9F3355
9H3211C

1
1 1
TITLE

111-6621 684-2146 114-6625

Upper Case Assembly SU


Key Panel
LCD Filter

3
4

1G9340A

6B8305 5H9130
6B9171 5H9428

1 1
1
MODEL NO.

DRAWING NO.

5 6
7

154-4575
114-6811
154-4694

Rated Power Label (C)

Upper Case Assembly


ASSEMBLY NO.

423-3445
DATE

Lock Spring

CE Marking Label

FX-2111 (CE)
10-31

96.01.24

b
2-Double Semus Screw
,

M3x8

rJ5

A
A

4
Pasted
<S>

A
<s>

1/

OS

Key
No.

Parts Na

Drawing Na

Description

Q ty

'

Remarks

9F3356 9H3212D 6B8327


6B8326 111-6622

Lower Case Assembly (1) Lower Case Assembly SU Battery Presser Sponge Battery Cover
Fuse Labell

2
4

1
TITLE

DRAWING NO.

314-3466
112-6624

5 6 7

1
MODEL NO.

Lower Case Assembly (1)


ASSEMBLY NO.

423-3446
DATE

5H9131 9E7293

154-4576

Rating Label

FX-2111 (CE)
10-33

96.01.24

Key
No.

Parts Na

Drawing Na

Description

Q'ty

Remarks

1
B

9F3356 423- 3446 9H3213A 1T1559 9F3420 6B8306

2
A

3
4

313- 3459 684-2157


424- 3443

Lower Case Assembly (1) Lower Case Assembly (1) Grounding Piece (A) SU
Transformer FCP2222UT
,

1 1 1 1 1 1 1 1

5
6 7

PCB-6410 Assembly
Insulation Sheet Insulation Cover TM-166 No 12
.

314- 3463 314-3601

6B91108A
6B6796

IB

ft
o

8
o

TITLE

DRAWING NO.

Lower Case Assembly (2)


MODEL NO.

422-3447
DATE

ASSEMBLY NO.

FX-2111 (CE)
10-35

96.01.24

2
4-Double Semus Screw M3x&
,

E
B

F
O

0
6
41
V

si fi'

ii

/I

i"

Key
Na

Parts Na

Drawing Na
422-3271

Description

Q ty

'

Remarks

Lower Case Assembly (2) Recorder Assembly

9F3354

1
TITLE

DRAWING NO.

3
4

1M0381

684-2093
423-3444

Thermal Array Head Cable PCB-6409 Assembly Lead Label (C) Lithium Battery Label

1
1

Lower Case Assembly (3)


MODEL NO.

422-3448
DATE

5 6

5H9138 5H5639

154-4577 154-3574

1 1

ASSEMBLY NO.

FX-2111 (CE)
10-37

96.01.24

1
it r

35
5S>
"J

ir.>
r

re

Double Semus Screw, M3x8

si

'

4-Doubte

Semus Screw M3x8


,

t'

Key
Na

Parts No.

Drawing Na

Description

Q'ty

Remarks
TITLE

Lower Case Assembly (3) Upper Case Assembly


9F2648

DRAWING NO.

2 3
4

1
1 1
MODEL NO.

Exploded View
ASSEMBLY NO.

412-0928
DATE

Paper Magazine Assembly


224-0891

9H3522

Paper Shaft SU

FX-2111 (CE)
10-39

96.01.24

FUKUDA DENSHI CO., LTD.


39-4, Hongo 3-chonne, Bunkyo-ku, Tokyo 113, Japan
Phone:(03)3815-2121 Fax: (03)5684-3791

Anda mungkin juga menyukai