Anda di halaman 1dari 4

Plant Documentation

15.02.2016 15:52:46
! = out of range
# = not read or communication error
* = parameter changed

Page 1/3

DTM Version: 3.26.00

Flowmeter Verification Certificate Transmitter


Plant

Customer

-------Order code

Tag Name

PROMASS 80 E DN50

1.9177 - 1.9177

Device type

K-Factor

K6142302000

99

Serial number

Zero point

V3.01.02

V1.06.00

Software Version Transmitter

Software Version I/O-Module

26.01.2016

19:25

Verification date

Verification time

Verification result Transmitter: Failed


Test item
Amplifier
Density
Temperature Measuring tube
Temperature Carrier tube
Current Output 1
Current Output 2
Pulse Output 1
Test Sensor

Result
Passed
Passed
Passed
Not tested
Passed
Not tested
Passed
Failed

FieldCheck Details
322763

Simubox Details
8672294

Production number

Production number

1.07.02

V1.00.00

Software Version

Software Version

01/2008

01/2008

Last Calibration Date

Last Calibration Date

Date

Operator's Sign

Applied Limits
Basis: 0.51 %
0.5 %
1.0 C
1.0 C
0.05 mA
0.05 mA
2P

Inspector's Sign

Page 2/3

FieldCheck - Result Tab Transmitter


Customer
Order code
Device type
Serial number
Software Version Transmitter
Verification date

Plant
Tag Name
K-Factor
Zero point
Software Version I/O-Module
Verification time

PROMASS 80 E DN50
K6142302000
V3.01.02
26.01.2016

-------1.9177 - 1.9177
99
V1.06.00
19:25

Verification Flow end value ( 100 % ): 28274.330 kg/h


Flow speed 4.00 m/s
Verification range: Standard range
Type of flow unit: MASS FLOW

Passed / Failed

Test item

Simul. Signal

Limit Value

Deviation

1413.718 kg/h (5%)


8199.557 kg/h (29.0%)
14137.166 kg/h (50.0%)
28274.331 kg/h (100%)

0.75 %
0.54 %
0.52 %
0.51 %

-0.13 %
-0.10 %
-0.10 %
-0.10 %

0.400 kg/l
0.800 kg/l
1.200 kg/l
2.000 kg/l

0.5 %
0.5 %
0.5 %
0.5 %

-0.0133 %
-0.0085 %
-0.0074 %
-0.0055 %

-23.4 C
123.4 C

1.0 C
1.0 C

-0.29 C
-0.39 C

---

---

---

Current Output 1

4.000 mA (0%)
4.800 mA (5%)
8.640 mA (29.0%)
12.000 mA (50.0%)
20.000 mA (100%)

0.05 mA
0.05 mA
0.05 mA
0.05 mA
0.05 mA

0.008 mA
0.008 mA
0.006 mA
0.006 mA
-0.009 mA

Current Output 2

-----

-----

-----

1000 P

2P

0P

---

---

-------

-------

Test Transmitter
Amplifier

Density

Temperature Measuring tube

Temperature Carrier tube

Pulse Output 1
Test Sensor
Erregerstrom- Kreis
Tube not oscillating
Frequency parameter
Temperature measuring points
Pick-up coils
Tube not oscillating

Legend of symbols
Passed

Failed

not tested

not testable

Attention

Page 3/3

FieldCheck: Parameters Transmitter


Customer
Order code
Device type
Serial number
Software Version Transmitter
Verification date

PROMASS 80 E DN50
K6142302000
V3.01.02
26.01.2016

Plant
Tag Name
K-Factor
Zero point
Software Version I/O-Module
Verification time

Curent Output

Assign

Current Range

Value 0_4mA

Terminal 26/27

MASS FLOW

4-20 mA activ

0.0 kg/h

Terminal 24/25

OFF

Pulse Output

Assign

Pulse Value

Terminal 22/23

MASS FLOW

0.100 kg/P

Output signal
Passive/Positiv
e

a)
b)
c)

System parameter
Flow Damping
Density Damping
Actual System Ident.

-------1.9177 - 1.9177
99
V1.06.00
19:25

Value 20
mA
17000.01
kg/h

Pulse width
10.00 ms

0.0 s
0.0 s
133.0