Doc Number
ABS.III.q/TC-006
00
Prepared by
Approved by
Siti Margiastuti
(Technical Laboratory)
Teguh Sudrajat
(DeputyTech Manager )
Safety First
Action
Details
Plan
Do
Heat sample (135-140C, 30-60min)
Thorough stirring and pour sample into the mold until it more than level full
Pour sample back and forth from end to end of the mold
Avoid incorporating bubbles
Clean chemical agent with water tap and then place it in the testing bath
(30min)
Control bath temp (Testing temp.0.1C)
Place the plate and mold with sample into testing bath (90 min 5min)
Temperature: +25 degC
Use Thermometer ASTM 63C
Sample immersed in testing bath, depth not less than 10cm
Then remove the side pieces and attach the rings at each end of the clips to
the hooks in testing machine.
Start machine and pull the two clips apart at a uniform speed to an
elongation of 100.25 cm
Stop the elongation and immediately cut the test specimen into two halves at
the midpoint using the scissors
Samples shall not come cross-sectional or come to the surface of the water,
nor touches the bottom of the bath at any time during the test.
Allow the specimen to remain in the testing machine in an undisturbed
condition at the specified temperature for a period of 60 min
Record the total length of the specimen with the severed ends just touching
each other
Check
Calculation and Report
Clean all the equipments (device, sample containers, accessories) after use
Picture