This is cerify, that the Hospital / Institution accepted the following equipment confirmation yhe
completion of installation :
EQUIPMENT :
No
.
1.
2.
3.
4.
5.
6.
7.
DESCRIPTION
Model
Serial No.
VENTILATOR
BUBLE CPOP
PASIEN MONITOR
PASIEN MONITOR
PASIEN MONITOR
PASIEN MONITOR
KOMPRESOR
MV 2000 NA I EO 150
150871022/1408030006
JH051500084 V3 CCE
JH051500540 V3 CCE
JH051500545 V3 CCE
JH051500543 V3 CCE
THC 13131
THC 13129
Phone
Date of Installation
: 15 NOVEMBER 2015
Hospital / Institution,
Authorised Representative
SERTIFIKAT KALIBRASI
IDENTITAS ALAT
Nama Alat
Bed Side
Monitor
Bed Side
Monitor
Bed Side
Monitor
Bed Side
Monitor
Merk Pabrik
Zonecare
Tipe
PM-7000C
No. Sertifikat
KSI 15100391
PM-7000C
Nomor seri
JH051500543V3CC
E
JH051500084V3CC
E
JH051500540V3CC3
Zonecare
PM-7000C
Zonecare
Zonecare
PM-7000C
JH051500545V3CC3
KSI 15100389
KSI 15100390
KSI 15100388
IDENTITAS PEMILIK
Nama
Alamat
I.
II.
III.
IV.
V.
: A-036
: KSI 15100391
: Bed Side Monitor
: Zonecare
: PM-7000C
: JH051500543V3CCE
:: Ruang Lab Kelistrikan KSI
: 01 Oktober 2015
Kondisi Lingkungan
1. Suhu
2. Kelembaban
: ( 24.50.4 ) C
: ( 53.752.6) %RH
: BPFK Jakarta
: Fluke Biomedical
: PT. Mutuagung Lestari
: BPFK Jakarta
: Baik
: Baik
PARAMETER
Main Voltage
Protective Eart Resistance
Insulation Resistance
Earth Leakage Current Normal Polarity
Earth Leakage Current Reverse Polarity
Enclosure Leakage Current Normal Polarity
Enclosure Leakage Current Normal Polarity
No Earth
Enclosure Leakage Current Reservel
Polarity
Enclosure Leakage Current Reservel
AMBANG BATAS
TERUKUR
22010%V
0.2
2 M
500 A
500 A
100 A
500 A
214.2A
0.175
OL M
73.2 A
73.6 A
5.0 A
5.0 A
100 A
5.0 A
500 A
5.0 A
Polarity No Earth
Parameter
SpO2
(%)
Pembacaan
standar
Standart
Reading
%
70
75
80
85
90
95
97
98
99
100
: A 036
: KSI 15100391
Pembacaan alat
Equipment
Reading
Koreksi
Correction
Ketidakpastian
Uncertainty
(U95% K=2)
%
70.3
75.0
80.0
85.0
91.0
95.2
97.5
98.0
99.0
99.0
%
-0.3
0.0
0.0
0.0
-0.1
-0.2
-0.25
-0.0
0.0
1.0
0,5
The result of measurement uncertainty has a confidence level (CL) of 95% with a coverage
factor (K)=2
2. Nilai sebenarnya = Nilai Pembacaan Alat + Nilai Koreksi
True Value = Equipment Reading Value + Correction Value
3. Hasil Kalibrasi tertelusur ke sistemsatuan internasional (SI)
Calibration result are traceable to the international unit system
4. Alat dikalibrasi menurut metode KSI-321, dengan ECRI 410-0595, 454-0595, 425-0595
sebagai standar.
The equipment is calibated according to KSI-312 method, with ECRI 410-0595, 454-0595,
425-0595 as a standard.
Manajer Teknik
Technical Manager
Ahmad Taufik, ST