GAS
CHROMATOGRAPHY
OLEH:
ISMAIL IBRAHIM PPN, PJK
Kenapa sampel yang bercampur
boleh diasingkan?
Arah aliran
Dasar sungai
Apa perbezaan nya?
Ringan
Berat
Interaction is
difference : gravity
Bagaimana pengasingan boleh
dilakukan?
• Pengasingan dapat dilakukan oleh turus/ column.
Turus/
column
packing material, 3-5um
BAGAIMANA PEMISAHAN
TERJADI DALAM KROMATOGRAFI
• split
• splitless.
Split/Splitless Flow System
Split Flow Diagram: Pre-Injection
SEPTUM
(SEPTUM)
2 mL/min PURGE VENT
TOTAL FLOW
50 mL/min
SPLIT VENT
48 mL/min (INLET)
PURGE VALVE
47.4 mL/min
= CARRIER GAS
LINER
0.6 mL/min
SEAL
COLUMN
Split Flow Diagram: Sample/Carrier
Mixing
(SEPTUM)
PURGE VENT
TOTAL FLOW
= SAMPLE MOLECULES
= SOLVENT MOLECULES (INLET) PURGE VALVE
TOTAL FLOW
= SAMPLE MOLECULES
= SOLVENT MOLECULES (INLET)
PURGE
VALVE
COLUMN
LINERS
Column materials
• Tubing
1. Metal - Stainless steel, nickel, copper,
aluminum
2. Glass - Pyrex, fused silica
3. Polymer - Teflon
Turus (Columns)
• Pack Column
• Capillary Column.
Model of the Chromatographic
Process
Flow A
D
How To Improve Column
Efficiency
1. Use smaller diameter column
2. Use a lower % or thinner film of stationary phase
3. Use smaller sample size
4. Use longer column.
5. Use temperature programming for sharper later
eluting peaks.
• 0.53 Megabore
• 0.45 High Speed Megabore
• 0.32 Wide
• 0.20-0.25 Narrow
• 0.18 Minibore
• 0.10 Microbore
• 0.05 “Nanobore”
INLET HEAD PREASSURES
• I.D(mm)Pressure(psig)
• 0.10 225-250
• 0.20 25-35
• 0.25 15-25
• 0.32 10.20
• 0.53 2-4
8
7
Detectors
Capillary Column
End-Position
(1-2 mm from Top of Jet) Jet
H2 Inlet
+
Make-Up
CH
Column
H 4 H
2 2
CH Jet
H 4 H
2 2
Compounds with Little or No
FID Response
Rare gases NH 3 CS 2
Nitrogen oxides H2 COS
Silicon halides CO O2
H 2O CO 2 N2
Perhalogenated cpds HCOH HCOOH
Operating the FID
Recommended Flow Rates
Gas Type Flow Range Suggested Flow
Carrier Gas
(hydrogen, helium,
nitrogen)
Packed Columns 10 - 60 ml/min
Capillary Columns 1 - 5 ml/min
Detector Gases
Hydrogen 24 to 60 ml/min 40 ml/min.
Air 200 to 600 ml/min 450 ml/min
Column plus capillary 10 to 60 ml/min 50 ml/min
makeup
Recommended Detector Temperature
If < 150º C, flame will not light
Detector temp should be 20º C > higher than oven temp.
Detector Type Support Selectivity Detectability Dynamic
gases range
Flame Mass flow Hydrogen Most organic 100 pg 107
ionization and air cpds.
(FID)
• Theory of operation
ECD berdasarkan kepada keelektronegatifan yang bertindak
balas dengan elektron terma yang membentuk ion bercas
negatif. Kehilangan electron bergantung kepada kuantiti analit
di dalam sampel.
Dalam menghasilkan elektron bertenaga rendah, gas
pembawa diionkan oleh zarah beta melalui sumber radioaktif
di dalam sel. Elektron yang mengalir menghasilkan arus yang
mana dikumpul dan diukur. Apabila molekul sampel berada di
dalam sel, elektron ditarik pada elektrod melalui sampel,
menyebabkan arus berkurangan. Perubahan ini dicatat dan
diukur oleh kromatogram.
NITROGEN PHOSPHORUS DETECTOR
• Theory of operation
The NPD (also called a thermionic detector) uses a jet
and collector similar in appearance to a Flame Ionization
Detector. In an NPD, however, ions of alkali metal are
introduced into a flame where hydrogen and air flows are
less than those for an FID, minimizing the normal
hydrocarbon ionizations, and increasing the ionization of
nitrogen or phosphorous compounds. This causes the
NPD to be both sensitive and selective for organic
compounds containing nitrogen and/or phosphorous.
This thermionic source efficiently ionizes nitrogen and
phosphorous containing organic molecules. Ions are
collected and the resulting current measured for the
chromatogram.
NITROGEN PHOSPHORUS
DETECTOR
• NPD juga dikenali sebagai pengesan thermoionic
menggunakan jet dan Collector adalah lebih kurang
sama dengan FID. Walau bagaimanapun, didalam NPD
ion logam alkali diajukan(introduced) kepada nyalaan api
hidrogen dan udara yang kurang berbanding FID,
pengionan hidrokarbon yang kurang dan meningkatkan
kadar pengionan sebatian nitrogen atau fosforus.
• Ini menyebabkan NPD lebih sensitif dan berupaya untuk
memilih sebatian organik yang mengandungi sebatian
nitrogen dan/atau fosforus.
• Ion yang dikumpul dan arus yang terhasil diukur untuk
kromatogram.
Flame Photometric Detector (FPD)
Theory of operation
In the Flame Photometric Detector (FPD), the sample burns in a hydrogen rich
flame, where some species are reduced and excited. The gas flow moves
the excited species to a cooler emission zone above the flame where they
decay and emit light. A narrow bandpass filter selects light unique to one
species, while a shield prevents intense carbon emission from reaching the
photomultiplier tube (PMT).
The light strikes a photosensitive surface in the PMT where a light photon
knocks loose an electron. The electron is amplified inside the PMT for an
overall gain of up to a million.
The current from the PMT is amplified and digitized by the FPD electronics
board. The signal is available either as a digital signal on the communications
output or as a voltage signal on the analog output.
FLAME PHOTOMETRIC DETECTOR
How a photoionization detector works
•
A photoionization detector uses an ultraviolet (UV) light
source to ionize components of the incoming sample.
When a compound enters the detector, it passes by the
UV lamp and is bombarded by high energy photons.
Components that have a low enough ionization potential
are ionized, producing free electrons that are directed to
a polarizing electrode within the detector. The
photoionization detector senses the electron stream as a
current that is proportional to the amount of ionizable
gases in the sample. The signal from the photoionization
detector is then amplified and output to a display or
external device.
THERMAL CONDUCTIVITY DETECTOR
• Theory of operation
The TCD responds to any compounds whose thermal conductivity is
different from the thermal conductivity of the carrier gas alone. The
TCD cell is a dual channel device, with an empty flow path and a
path containing a detector filament. A switching valve alternates
between sending the column effluent (containing analytes) through
the empty and the active flow paths. When the column effluent flows
through the empty channel, a pure stream of reference gas
maintains an equilibrium through the filament path. The reference
gas is used to compare thermal conductivity changes caused by the
column effluent.
QUALITATIVE AND
QUANTITATIVE
Data Acquisition
1. Chart recorder
2. Integrator
3. Chemstation
- Computer & chromatographic
software
Qualitative
Analysis
1. Direct Comparison of Retention Time
The simplest procedure in the identification of
unknown. A known standard is first analysed under
the specific GC conditions, followed by the
unknown also under the similar conditions.
The difference in the retention times should not be
more than +/-0.1 min for proper identification
Qualitative
Analysis Retention time
ISTD Known component added to both sample Must add a component to the sample
and standard More complex sample and standard
Injection size not critical preparation steps
Calibrate peaks of interest
Correct for detector response
Results reported in units of choice
Quantitative Analysis
Concentration
1. External Standard
Method
- Concentration of an unknown A
CA = C1/A1 x A A
C =concentration , A = peak area Peak Area
2. Internal Standard
Method concentration ratio
- Concentration of an unknown X
Cx = (CA /CS)/(AA /AS) x AX /AIS x WiS
IS, S=internal standards
Area ratio
( As/AIS)
Concentration
INTERNAL STANDARD
(Corrected Response) y
Amount of y = X ISTD Amount X MF
(Corrected Response) ISTD
EXTERNAL STANDARD
Target Target
Disadvantage of external
standard calibration method
• Injection error will directly influence the
quantitative result.
IS T T
IS
T
IS
Disadvantage of internal
standard calibration method