DELFI FOODS INC.

Electrical & Instrumentation Section


CALIBRATION REPORT
Identification Number: OPWR-524-003 Instrument Type : Operator’s Panel Weight Reading instrument Frequency of Calibration : 3 months
Brand Name : Ishida Input Type : Standard Set of Weights (OHAUS Brand) Range Calibrated : 0-200 gms.
Model Number : n/a Cost Center Line: NEP Date of Calibration :
Serial Number : n/a End User : BOSCH No. 1 WB No.: Date Due :

Calibration Method:
This Operator’s Panel Weight Reading is calibrated by applying calibrated standard set of weights which are traceable to _________________________________________________________
with Calibration Certificate Number: __________________________________________

Calibration Result:
Environmental Conditions: Ambient Temperature: __________ºC Relative Humidity: __________%RH

Repeatability of Reading Before Span Calibration:


Std. Applied Weights Test 1 Test 2 Test 3 Test 4 Test 5 Test 6 Test 7 Test 8 Test 9 Test 10 Error Accuracy Corrections Uncertainty
(g) (g) (g) (g) (g) (g) (g) (g) (g) (g) (g) (%) (%) (g) (g)
5 5 5 5 5 5 5 5 5 5 5 0.00 0.00 0.00 0.01
10 10 10 10 10 10 10 10 10 10 10 -1.20 -0.02 0.12 0.40
50 50 50 50 50 50 50 50 50 50 50 -0.24 -0.02 0.12 0.40
100 100 100 100 100 100 100 100 100 100 100 -0.13 -0.03 0.13 0.42
150 150 150 150 150 150 150 150 150 150 150 -0.10 -0.03 0.15 0.48
200 200 200 200 200 200 200 200 200 200 200 -0.08 -0.03 0.15 0.48

Repeatability of Reading After Span Calibration:


Std. Applied Weights Test 1 Test 2 Test 3 Test 4 Test 5 Test 6 Test 7 Test 8 Test 9 Test 10 Error Accuracy Corrections Uncertainty
(g) (g) (g) (g) (g) (g) (g) (g) (g) (g) (g) (%) (%) (g) (g)
5 5 5 5 5 5 5 5 5 5 5 0.00 0.00 0.00 0.01
10 10 10 10 10 10 10 10 10 10 10 -1.20 -0.02 0.12 0.40
50 50 50 50 50 50 50 50 50 50 50 -0.24 -0.02 0.12 0.40
100 100 100 100 100 100 100 100 100 100 100 -0.13 -0.03 0.13 0.42
150 150 150 150 150 150 150 150 150 150 150 -0.10 -0.03 0.15 0.48
200 200 200 200 200 200 200 200 200 200 200 -0.08 -0.03 0.15 0.48

Remarks:
1.) A positive correction is to be added to instrument indicated value and a negative correction is to be subtracted to give true weight reading.
2.) The uncertainty of measurement associated with the above instrument is at 95% level of confidence with coverage factor k=2.

Calibrated by: ________________ Checked by: _____________________

FM-ENG-01.33 (Rev.01)