1260 Infinity Series HPLC Site Preparation Checklist

Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation is the key first step in ensuring that your instruments and software systems operate reliably over an extended lifetime. This document is an information guide AND checklist prepared for you that outlines the supplies, consumables, space and utility requirements for your equipment for your site. For additional information about our solutions, please visit our web site at http://www.chem.agilent.com/en-US/Pages/HomePage.aspx

Customer Responsibilities
Make sure your site meets the following prior to the installation date using the checklist below. For details, see specific sections within this document, including:
The necessary laboratory or bench space is available. The environmental conditions for the lab as well as laboratory gases, tubing. The power requirements related to the product (e.g. number & location of electrical outlets) The required operating supplies necessary for the product and installation please consult Other/Special Requirements section below for other product-specific information If Agilent is delivering installation and familiarization services, users of the instrument should be present throughout these services; otherwise, they will miss important operational, maintenance and safety information.

Important Customer Information


1. If you have questions or problems in providing anything described as a Customer Responsibilities above, please contact your local Agilent or partner support/service organization for assistance prior to delivery. In addition, Agilent and/or its partners reserve the right to reschedule the installation dependent upon the readiness of your laboratory. 2. Should your site not be ready for whatever reasons, please contact Agilent as soon as possible to rearrange any services that have been purchased.
3.

Other optional services such as additional training, operational qualification (OQ) and consultation for user-specific applications may also be provided at the time of installation when ordered with the system, but should be contracted separately.

Issued: 26 July 2010

Revision: 1.0

1 Copyright 2010 Page 1 of 9

Agilent Technologies

1260 Infinity Series HPLC Site Preparation Checklist

Dimensions and Weight

Identify the laboratory bench space before your system arrives based on the table below. Pay special attention to the total height and total weight requirements for all system components you have ordered and avoid bench space with overhanging shelves. Also pay special attention to the total weight of the modules you have ordered to ensure your laboratory bench can support this weight.
Instrument Description G1310B Isocratic Pump G1311B/C Quat. Pump/VL G1312B/C Binary Pump G1376A Capillary Pump G2226A Nano Pump G1322A / G1379B Degassers G1361A Preparative Pump G2258A Dual Loop Autosampler G1329B/G2260A Autosamplers G1367E, G1377A High Performance Autosamplers G1330B ALS Thermostats G1316AThermostatted Column Compartment G1316C Thermostatted Column Compartment G1314B/C Variable Wavelength Detectors G1314E/F Variable Wavelength Detectors G1315D Diode-Array Detectors G1365D Multiple Wavelength Detectors G1321B Fluorescence Detector G4212B Diode-Array Detectors G1362A Refractive Index Detector G4218A Evaporating Light Scattering Detector G1364B/C/D Fraction Collectors Weight Kg lbs 11 kg 25 lbs. 14.5 kg 32 lbs. 15.5 kg 34 lbs. 17 kg 7.5 kg 15 kg 14.0 kg 14.2 kg 15.5 kg 18.5 kg 11.2 kg 11.2 kg Approx 11kg Approx 11kg 11.5 kg 11.5 kg 11.5 kg 11.5 kg 17 kg 20kg 17 kg 39 lbs. 16.5 lbs. 32.9 lbs. 29.8 lbs. 31.3 lbs. 34.2 lbs 40.7 lbs. 22.5 lbs. 22.5 lbs. Approx 25 lbs Approx 25 lbs 26 lbs. 26 lbs. 25.4 lbs 26 lbs. 38 lbs. 45 lbs 38 lbs. Height cm in 18 cm 7 in 18 cm 7 in 18 cm 7 in 18 cm 8 cm 20 cm 20 cm 20 cm 20 cm 14.4 cm 14 cm 14 cm 14 cm 14 cm 14 cm 14 cm 14 cm 14 cm 18 cm 45cm 18 cm 7 in 3 in 8 in 8 in 8 in 8 in 5.5 in 5.5 in 5.5 in 5.5 in 5.5 in 5.5 in 5.5 in 5.5 in 5.5 in 7 in 17.7 in 7 in Depth cm in 43.5 cm 17 in 43.5 cm 17 in 43.5 cm 17 in 43.5 cm 43.5 cm 43.5 cm 43.5 cm 43.5 cm 43.5 cm 43.5 cm 43.5 cm 43.5 cm 43.5 cm 43.5 cm 43.5 cm 43.5 cm 43.5 cm 43.5 cm 43.5 cm 55.cm 43.5 cm 17 in 17 in 17 in 17 in 17 in 17 in 17 in 17 in 17 in 17 in 17 in 17 in 17 in 17 in 17 in 17 in 21.7 in 17 in Width cm in 34.5 cm 13.5 in 34.5 cm 13.5 in 34.5 cm 13.5 in 34.5 cm 34.5 cm 34.5 cm 34.5 cm 34.5 cm 34.5 cm 34.5 cm 41 cm 41 cm 34.5 cm 34.5 cm 34.5 cm 34.5 cm 34.5 cm 34.5 cm 34.5 cm 25cm 34.5 cm 13.5 in 13.5 in 13.5 in 13.5 in 13.5 in 13.5 in 13.5 in 16 in 16 in 13.5 in 13.5 in 13.5 in 13.5 in 13.5 in 13.5 in 13.5 in 9.8 in 13.5 in

Issued: 26 July 2010

Revision: 1.0

2 Copyright 2010 Page 2 of 9

Agilent Technologies

1260 Infinity Series HPLC Site Preparation Checklist

Environmental Conditions
Operating your instrument within the recommended temperature ranges insures optimum instrument performance and lifetime. Special Notes: 1. Performance can be affected by sources of heat & cold e.g. direct sunlight, heating/cooling from air conditioning outlets, drafts and/or vibrations. 2. The sites ambient temperature conditions must be stable for optimum performance.
Operating temp range C (F)

Instrument Description

Operating humidity range (%)

Heat Dissipation (BTU) N/A N/A N/A N/A N/A N/A N/A

G1314B/C/ E, G1315C/D, G1316A/C, G1322A, G1365C/D, G1362A G1321B G4212B G1379B G1330B, G1361A, G1364B/C/D, G2258A, G4212B G4218A1 All other modules

0 to 55C (32 to 131F), constant temperature. 0 to 40C (32 to 104F), constant temperature. 4 to 40C (39 to 104F), constant temperature. 0 to 45C (32 to 113F) 4 to 40C (39 to 104F) 5 to 40C (41 to 104F) 4 to 55C (39 to 131F), constant temperature.

< 95%, noncondensing < 95%, noncondensing < 95%, noncondensing < 95%, noncondensing < 95%, noncondensing <80%, noncondensing < 95%, noncondensing

Issued: 26 July 2010

Revision: 1.0

3 Copyright 2010 Page 3 of 9

Agilent Technologies

1260 Infinity Series HPLC Site Preparation Checklist

Power Consumption
Special Notes: 1. If a computer system is supplied with your instrument, be sure to account for those electrical outlets.
Instrument Description Line Voltage & Frequency (V, Hz) Maximum Power Consumption (VA) Maximum Power Consumption (W)

G1310B Isocratic Pump G1311B/C Quaternary Pump/VL G1312B/C Binary Pump/VL G1376A Capillary Pump G2226A Nano Pump G1361A Prep Pump G1379B Micro Degasser G1322A Degasser G1329B, G2260A ALS G1367E, G1377A High Performance Autosamplers G2258A Dual Loop Autosampler G1330B Sample Thermostat G1316A/C Therm. Column Comp G1314E/F VWD G1315C/D DAD G1365C/D MWD
G4212B DAD

100-240V (AC) 100-240V (AC) 100-240V (AC) 100-240V (AC) 100-240V (AC) 100-240V (AC) 100-240V (AC) 100-240V (AC) 100-240V (AC) 100-240V (AC) 100-240V (AC) 100-240V (AC) 100-240V (AC) 100-240V (AC) 100-240V (AC) 100-240V (AC) 100-240V (AC) 100-240V (AC) 100-240V (AC) 230V (AC) 100V/115V (AC)

180 VA 180 VA 220 VA 220 VA 250 VA 30 VA 30 VA 300 VA 300 VA 260 VA 260 VA 320 VA 120 VA 160 VA 160 VA
160 VA

55 W 110 W 74 W 75 W 210 W 30 W 30 W 200 W 200 W 210 W 210 W 150 W 100W 160 W 130 W
130 W

G1362A RID G1321B FLD G1364D Fraction Collectors G4218A ELSD

160 VA
180 VA

65 W
70 W

200 VA N/A N/A

180 W 400 W 210 W

Issued: 26 July 2010

Revision: 1.0

4 Copyright 2010 Page 4 of 9

Agilent Technologies

1260 Infinity Series HPLC Site Preparation Checklist

Required Operating Supplies by Customer

Special Notes: 1. For information on Agilent consumables, accessories and laboratory operating supplies, please visit http://www.chem.agilent.com/en-US/Products/consumables/Pages/default.aspx
Item Description, (including dimensions etc) G4218A ELSD Gas N2 (Typical) Vendor/Part Number (if applicable) N/A Recommended Quantity N/A

Other/Special Requirements

G4218A ELSD Gas requirements A supply of inert gas (typically nitrogen) is required to operate the detector. The gas supply needs to be free of oil, humidity and particles, as such contaminations will create background noise in the chromatograms and may damage the built-in pressure sensor. In case of such noise for example for newly installed gas lines, flush the gas lines for sufficient time (might take days) and consider additional filters. A manometer (part number G4218-60100) with filter (0.01 m) is available optionally. Pure gas is not required as the gas is only used as a carrier for the solid sample particles. Exhaust venting and drain requirements The exhaust from the detector must be directed into a fume hood or exhaust vent. If a vacuum is used, it should be moderate so as to avoid turbulence in the glass cell siphon. The potentially hazardous exhaust of evaporated solvent and sample must not be allowed to enter the laboratory atmosphere and any appropriate accessory like solvent filters should be disposed according to local environmental requirements. The drain tube must be directed to a suitable solvent waste container. The user is responsible for decontamination or recycling of any residue, regarding to local environmental requirements. Further requirements The detector is controlled via RS232, not via LAN/CAN, so it must be installed close to the control PC unless special data transmission systems are used. The length of the included straight female/female RS 232 cable is 1.7 m. Customer should have available HPLC grade Acetonitrile and water with a dry residue below 1 ppm or MS grade solvents.

Issued: 26 July 2010

Revision: 1.0

5 Copyright 2010 Page 5 of 9

Agilent Technologies

1260 Infinity Series HPLC Site Preparation Checklist

Solvent cabinet Vacuum degasser

Pump

Instant Pilot

Autosampler or Micro Autosampler or Preparative Autosampler

Column compartment

Detector

Figure 1 Recommended 1-Stack Configuration (without Thermostat)

Issued: 26 July 2010

Revision: 1.0

6 Copyright 2010 Page 6 of 9

Agilent Technologies

1260 Infinity Series HPLC Site Preparation Checklist

Figure 2 Recommended 2-Stack Configuration (with Thermostat)

Detector Column Compartment

Solvent cabinet Autosampler Vacuum degasser

Thermostat

Pump

Issued: 26 July 2010

Revision: 1.0

7 Copyright 2010 Page 7 of 9

Agilent Technologies

1260 Infinity Series HPLC Site Preparation Checklist

Figure 3 Recommended 2-Stack Configuration with Fraction Collector (Analytical Scale System)

Solvent cabinet Vacuum degasser Pump UV-VIS Detector Sampling Device Column Compartment Fraction Collector

Issued: 26 July 2010

Revision: 1.0

8 Copyright 2010 Page 8 of 9

Agilent Technologies

1260 Infinity Series HPLC Site Preparation Checklist

Figure 4 Recommended 2-Stack Configuration with Fraction Collector (Preparative Scale System)

Solvent Cabinet Prep Pump A Column Organizer Prep Pump B

Fraction Collector Sampling Device UV-VIS Detector

Issued: 26 July 2010

Revision: 1.0

9 Copyright 2010 Page 9 of 9