}C
"
alcohols, C
}C
`
aldehydes and C
}C
`
hydrocarbons (para$ns, ole"ns,
aromatics).
This method cannot be used for the analysis of
formic acid which has a very low response in FID
(Dietz, 1967), but is convenient for the analysis of heavier
acids.
3.2. Organic acid sources
In the combustion reactor used as the organic acids
source, the acids can be found only very close to the
burner: within the "rst 5 mm around it. In the "rst 3 mm,
concentrations are high. Reaching the fourth and the "fth
millimetre, they decrease sharply down. No acids are
found from the sixth millimetre and above. Propane
combustion generates only formic, acetic and propionic
acids (Zervas, 1996). It is not clear whether the acids are
formed in the #ame or are products from other com-
pounds through some cooling reactions in the probe.
Figs. 7}9 show the organic acid concentrations obtained
depending on the distance of the burner and on the
equivalence ratio. These concentrations range from
0.09 to 10.6 ppm for formic acid, from 0.11 to 8.3 ppm
for acetic acid and from 0.05 to 0.069 ppm for propionic
acid.
3.3. Collection ezciency
The collection e$ciency of the di!erent traps was
characterised as the percentage collected in the "rst im-
pinger (or cartridge) of the total one.
The e$ciency of water and Na
`
CO
`
solutions for the
"rst two acids is presented in Fig. 10. In the test condi-
tions, it is above 90% for both liquids. It is always 1 or 2
units higher with formic than with acetic acid. Na
`
CO
`
solution traps the acids better than water; the corre-
sponding values are 4 or 5 units higher. Literature re-
ports values reach 99% with water (Smith, 1985) and
92% with Na
`
CO
`
solutions (Bodek, 1980), but e$ciency
values are in#uenced by the gas-phase concentration and
by the #ow rate of the gas passing through the impingers
(Zervas, 1996). The higher these parameters values, the
lower the e$ciency values.
Fig. 11 shows the in#uence of gas volumes, ranging
from 0.038 to 0.095 m`, on e$ciency when concentra-
tions of formic, acetic and propionic acid are, respectively,
9, 8.5 and 0.45 ppm. The gas volume passing through the
impingers does not in#uence the e$ciency for such gas-
phase concentrations. The decrease is 1% for acetic and
propionic acid and 0.4% for formic acid.
The e$ciency of the four cartridges used is presented in
Fig. 12. For formic acid, the four types have an e$ciency
of 100%. For acetic and propionic acids, the two
aluminas have an e$ciency of 100%, but silica and
#uorisil e$ciencies are lower (45}90%), silica being bet-
ter (about ten units) than #uorisil.
4956 E. Zervas et al. / Atmospheric Environment 33 (1999) 4953}4962
Fig. 4. Analysis of nine aliphatic acids and of benzoic acid by gas chromatography/FID detection. Aqueous solution of 6}8 ppm.
Column used: HP Megabore Innowax 30 m;0.53 mm, carrier gas #ow: 4;10` m` s, injector and detector temperatures: 523 K,
initial column temperature: 393 K, initial time: 120 s, "nal temperature: 473 K, heating rate: 0.066 K s.
Fig. 5. FID response for acetic and propionic acid analysis as
a function of the solution concentration.
3.4. Elution ezciency
The e$ciency of the cartridge elution is presented in
Fig. 13. This item is de"ned as the percentage eluted the
"rst time over the total quantity eluted ("rst and second
Fig. 6. Relative standard deviation of acetic and propionic acid
analysis by gas chromatography as a function of the solution
concentration.
time). The elution of silica and of #uorisil is about the
same: 100% for formic acid, 80}85% for acetic acid and
90% for propionic acid. Alumina B leads to lower values,
from 50 to 85%. Last, alumina N has no e$ciency in
E. Zervas et al. / Atmospheric Environment 33 (1999) 4953}4962 4957
Fig. 7. Combustion of propane. Formic acid concentration in
the gas-phase as a function of the equivalence ratio and the
distance probe-burner. Total (propane#air) #ow"3.33;
10" m` s, inspiration #ow"3.33;10` m` s.
Fig. 8. Combustion of propane. Acetic acid concentration in the
gas-phase as a function of the equivalence ratio and the distance
probe-burner. Total (propane#air) #ow"3.33;10" m` s,
inspiration #ow"3.33;10` m` s.
elution. Acids are adsorbed so strongly that hot water
cannot elute them.
The yield of the collection, de"ned as the product ez-
ciency of the collection x ezciency of the elution, reaches
100% in the cases of formic acid collected on silica or on
Fig. 9. Combustion of propane. Propionic acid concentration in
the gas-phase as a function of the equivalence ratio and the
distance probe-burner. Total (propane#air) #ow"3.33;
10" m` s, inspiration #ow"3.33;10` m` s.
Fig. 10. E$ciency of formic and acetic acid collection in
2.0;10` m` of a 1% Na
`
CO
`
aqueous solution or in de-
ionised water. Gas #ow rate"3.33;10` m` s, gas vol-
ume"0.01 m`, gas-phase concentration"formic acid: 1 to
6 ppm, acetic acid: 1}5 ppm.
#uorisil. It is nil in the case of alumina N. Therefore, solid
cartridges cannot be used to capture all the organic acids.
It was decided to collect organic acids in pure water and
to analyse the solution by the combination of the two
methods described above: ionic and gas chromatography.
4958 E. Zervas et al. / Atmospheric Environment 33 (1999) 4953}4962
Fig. 11. In#uence of the gas volume passing through the
deionised water impinger on its e$ciency. Volume of the
solution"2.0;10` m`, gas #ow rate"3.33;10` m` s,
gas-phase concentration: formic acid 9 ppm, acetic acid 8.5 ppm,
propionic acid 0.45 ppm.
Fig. 12. E$ciency of the four types of cartridges on acid collec-
tion. Gas-phase concentrations: formic acid 1}6 ppm, acetic acid
0.5}5 ppm, propionic acid 0.5}5 ppm, gas volume: 0.01 m`, gas
#ow rate: 3.33;10` m` s.
3.5. Stability of the eluted solutions
The stability of organic acid solutions has to be con-
trolled. Six aqueous solutions of organic acids issued
Fig. 13. E$ciency of elution on the four types of cartridges.
Elution by deionised water, volume"1;10" m`, temperature
363 K.
frompropane combustion were analysed, at weekly inter-
vals, during six weeks. In the intermediate period, they
were refrigerated at 243 K and preserved from light.
During the "rst two weeks, the di!erences in concen-
tration remained under 15%. Later, acetic acid concen-
tration increased in all solutions and reached 150% of
the initial value on week six. Formic and propionic acid
concentrations decreased and reached 10% of the initial
values after six weeks. Obviously, other compounds,
issued from propane combustion and collected in the
impingers were oxidised into acetic acid. Formic and
propionic acid were oxidised to simplest compounds.
3.6. Linearity, repeatability and detection limits of the
method
This method includes three steps: evaporation, collec-
tion and analysis of organic acids. The evaporation of an
aqueous solution of these acids is used as source. The
collection is performed in two impingers each containing
20 mL of deionised water and the analysis of the "nal
solution by two methods: ionic chromatography for the
analysis of the formic acid and gas chromatography for
the analysis of the heavier aliphatic acids. Linearity,
repeatability and detection limits of the global method
are presented here.
Figs. 14 and 15 present the area, as given by the
chromatograms, as a function of the collection time, for
two gas-phase concentrations obtained by the evapor-
ation system: 1 and 10 ppm of each of the three acids:
formic, acetic and propionic acid, using the same condi-
tions of collection and analysis as presented before. The
E. Zervas et al. / Atmospheric Environment 33 (1999) 4953}4962 4959
Fig. 14. Linearity of the global method as a function of the
collection time. Gas-phase concentrations: 1 ppm of each acid,
gas #ow rate: 3.33;10` m` s, volume of the collection solu-
tion: 2.0;10` m` of deionised water.
Fig. 15. Linearity of the global method as a function of the
collection time. Gas-phase concentrations: 10 ppm of each acid,
gas #ow rate: 3.33;10` m` s, volume of the collection solu-
tion: 2.0;10` m` of deionised water.
linearity is high, the r` of these lines are always better
than 0.98.
The relative standard deviation of the global method,
as a function of the solution concentration for a gas-
phase concentration of 1 ppm for the three acids, is
Fig. 16. Relative standard deviation of the global method as
a function of the solution concentration. Gas-phase concentra-
tions: 1 ppm, gas #ow rate: 3.33;10` m` s, volume of the
collection solution: 2.0;10` m` of deionised water, analysis of
the formic acid by ionic chromatography and of the acetic and
propionic acid by gas chromatography.
Fig. 17. Analysis by ionic chromatography of the organic acids
in spark ignition engine exhaust gas. Collection in two impingers
each containing 20 mL deionised water each, gas #ow:
3.33;10` m` s, gas volume: 0.01 m`, equivalent ratio: 1.0.
presented in Fig. 16. The r.s.d. values are lower than 10
for the concentrations of more than 1, 8 and 12 ppm in
the solution for the formic, acetic and propionic acid
respectively. These values are between 3 and 25% higher
4960 E. Zervas et al. / Atmospheric Environment 33 (1999) 4953}4962
Fig. 18. Analysis by gas chromatography of the organic acids in spark ignition engine exhaust gas. Collection in two impingers each
containing 20 mL of deionised water each, gas #ow: 3.33;10` m` s, gas volume: 0.01 m`, equivalent ratio: 1.0.
than the rsd analysis alone (the most important di!erence
is observed in diluted solutions).
The detection limits of the method is under 10 ppb for
the formic acid and below 0.5 ppm for the heavier
aliphatic acids.
3.7. Practical use with engine exhaust gases
In the exhaust gases of a spark ignition engine fed with
several commercial fuels and within a large range of
tuning, formic, acetic, propionic, acrylic, butyric and
isovaleric acids can be detected, in concentrations of
0.1}15, 2}40, 0.5}80, 0}7, 0.8}3.5 and 0}5 ppm, respec-
tively. These concentrations depend on the fuel and the
equivalence ratio used. The other acid's concentrations
are below the detection limits. A Diesel engine, fed with
two di!erent commercial fuels, produces detectable con-
centrations of the "rst three acids (formic 2}6, acetic 4}8
and propionic 0.4}3 ppm).
The other compounds of the exhaust gases do not
interfere with the analysis of these acids. Figs. 17 and 18
are two chromatograms, corresponding to the analysis,
by ionic and by gas chromatography respectively, of
the acids produced by the spark ignition engine at
stoichiometry.
4. Conclusion
The evaporation of an aqueous solution containing
organic acids can be successfully used to generate a gas
#ow with a known and stable concentration in these
acids. Propane combustion on a #at burner can also
generate the "rst three acids in the "rst "ve mil-
limetres of the burner. It is not clear whether the organic
acids are formed in the #ame or in the probe, but this
method can be used as a stable source of these com-
pounds.
Organic acid trapping can be performed in deionised
water or in Na
`
CO
`
aqueous solution. The e$ciency of
these liquid traps is over 92% for the "rst two acids,
under the conditions we used. All four solid traps are
e$cient in collecting formic acid but only the two
aluminas can collect quantitatively acetic and propionic
acids. The elution e$ciency using hot water reaches
100% in the case of formic acid adsorbed on silica or
#uorisil; the other acids have lower elution rates
(80}90%) of these solids. Alumina B has a very low and
alumina N no e$ciency in elution; only silica could be
used for formic acid collection.
The analysis of the collected acids can be performed by
two methods, isocratic ionic chromatography for formic
acid and gas chromatography for the other acids. Or-
ganic acids have no interference with other compounds
of the exhaust gases and the detection limits are low,
some ppb in the case of formic acid and under 0.5 ppm in
the case of the other acids.
This method is applied to the analysis of the organic
acids produced by two internal combustion engines, a SI
and a Diesel one, using commercial fuels. Six acids are
detected in the exhaust gases of the "rst engine emissions
and three in the second.
E. Zervas et al. / Atmospheric Environment 33 (1999) 4953}4962 4961
References
Bodek, I., Menzies, T., 1980. Ion chromatographic analysis of
organic acids in diesel exhaust and mine air. Proceedings of
Second Symposium on Process Measurements for Environ-
mental Assessment, February 25}27, Atlanta, Report NTIS
PB82-211574, pp. 155}168.
Dietz, W.A.J., 1967. Gas chromatography 5 (2), 68}71.
Grosjean, D., Tuazon, E.C., Fujita, E., 1990. Ambient formic
acid in Southern California air: a comparison of two
methods, Fourier Transform Infrared Spectroscopy and Al-
kaline Trap-Liquid Chromatography with UV detection.
Environmental Science Technology 24, 144}146.
Hekmat, M., Smith, R.G., 1991. Determination of low molecular
organic acids collected on silica gel air sampling tubes by
using ion-exclusion chromatography. American Industrial
Hygine Association Journal 52, 332}335.
Kawamura, K., Ng, L.L., Kaplan, I.R., 1985. Determination
of organic acids in atmosphere, motor exhausts and
engine oils. Environmental Science and Technology 19,
1082}1086.
Lawrence, J.E., Koutrakis, P., 1992. Evaluation of annular de-
nuders to collect gas-phase organic acids. Air and Waste
Management Association 91-53.2.
Levi, V., Wehr, T., Talmade, K., Zhu, M., 1993. Analysis of
organic acids in wines by capillary electrophoresis and
HPLC. International Chromatography Laboratory 15, 4}7.
Lopez, B., Masclet, P., Person, A., Degobert, P., 1987. Emissions
diesel: analyse des polluants non reH glementeH s. Pollution At-
mosphere 29 (No speH cial), 113}123.
McCalley, D.V., Tomas, C.W., Leveson, L., 1984. Analysis of
carboxylic acids by gas chromatography, derivation using
chloroacetone. Chromatographia 18 (6), 309}312.
Slanina, J., De Wild, P.J., Wyers, G.P., 1992. The application of
denuder systems to the analysis of atmospheric components.
In: Nriagu, J.O. (Ed.), Gaseous Pollutants: Characterisation
and Cycling. Wiley, New York, pp. 129}154.
Smith, L.R., 1985. Characterisation of exhaust emissions alco-
hol-fuelled vehicles. NTIS Report PB85-238582.
Tanaka, K., Fritz, J.S., 1986. Separation of aliphatic carboxylic
acids by ion-exclusion chromatography using a weak-acid
eluent. Journal of Chromotography 361, 151}160.
Yokouchi, Y., Ambe, Y., 1986. Characterisation of polar or-
ganics in airborne particulate matter. Atmospheric Environ-
ment 20 (9), 1727}1734.
Zervas, E., 1996. Mechanisms of speci"c pollutants' formation in
internal combustion engines, Ph.D. Thesis, University of
Mulhouse, France.
4962 E. Zervas et al. / Atmospheric Environment 33 (1999) 4953}4962