Anda di halaman 1dari 10

Shree Renuka Sugars Ltd., Haldia Automatic Digital Saccharimeter - Technical Specification Doc No.

REN/ HAL/ LAB/ INT / SPEC-01 SNO. Instrument Particulars 1 Name 2 Make 3 Measurement Scale

4 5 6 7

Range of Angle of Rotation Readability Measurement Accuracy Source of Light

Auto - Digital Saccharimeter / Polarimeter Rudolph Autopol - 880 - US Angle of Rotation International Sugar Scale User defined mode + 225 Deg Z 0.01 Z Measurement Accuracy + 0.01 Z Tungsten-Halogen Lamp Wave Length : 589 nm D Line & 880 nm Prism Make : Glan Thompson Calcite Quartz

Accessories

Jacketed Flow tube and funnel Observation Tube 200 mm Long and 100 mm Quartz Control Plate ISS 99 Z RS-232 port to connet to computer / printer RS-232 interface cable Purity Option Automatic Digital Polarimeter with Automatic Temperature Correction Automatic calibration by push - button User Interface - Touch Screen Measures samples with transmittance as low as 0.01 % Colour LCD with black light Uses Glan Thompson Calcite Prism Output to Digital Printer DP-63 of DP-AD optional, RS -232 Input / Output ports for connecting to a computer system

Special Features

Page 1 of 10

Shree Renuka Sugars Ltd., Haldia Automatic Digital Refractometer - Technical Specification Doc No. REN/ HAL/ LAB/ INT / SPEC-02 SNO. Instrument Particulars 1 Make / Model 2 Measurement Scale 3 4 Measurement Range & Accuracy Details Rudolph Refractometer - J157 - US Brix with automatic temperature compensation and Refractive Index 0 to 95 Brix Minimum Indication 0.01 % brix ATC Range for brix : 18 to 40 deg C Temperature Accuracy : 0.05 deg 5 Measurement Accuracy + 0.01 % Brix 1, 2 or full multi-point calibration Temperature measurement and control : prism surface and in the presser. Communcaiton : RS232 cable Prism : Aritificial Sapphire 6 Accessories All additional accessories like Purity analyzer comes along with Autopol model

Shree Renuka Sugars Ltd., Haldia Semi Automatic Polarimeter - Technical Specification Doc No. REN/ HAL/ LAB/ INT / SPEC-03 SNO. Instrument Particulars 1 Make / Model 2 Measurement Scale 3 4 5 6 7 Measurement Range Minimum Indication Measurement Accuracy Source Accessories Details Polax - 2L Atago - Semi Automatic Angle of Rotation International Sugar Scale Angle of Rotation: +180 ~ -179.95 Deg International Sugar Scale : -130.00 ~ +130.00 Z 0.05Deg / 0.1 Z + 0.1 Deg Wave Length : 589 nm D Line Observation Tube 100 mm and 200 mm one each Maximum Length of observation tube : 200 mm

Shree Renuka Sugars Ltd., Haldia Refractometer (Pocket type ) - Technical Specification Doc No. REN/ HAL/ LAB/ INT / SPEC-04 SNO. 1 2 3 Instrument Particulars Make / Model Measurement Scale Measurement Range Details Atago - PAL 3 - Japan (Pocket size) Brix 0 to 93 Brix

4 5

Minimum Indication Measurement Accuracy

0.1 % brix + 0.2 % Brix Temperature range - 10 to 75 Deg C AAA alkaline battery X 2 With automatic tempreature compensation and water resistance features Washable under running water - water resistance model

6 7

Accessories Special Features

Shree Renuka Sugars Ltd., Haldia UV-Visible Spectrophotometer - Double Beam - Technical Specification Doc No. REN/ HAL/ LAB/ INT / SPEC-05 SNO. Instrument Particulars 1 Make 2 Wave Length Range and Accuracy 3 Photometric Range and Accuracy Details Thermo electron / Merck-UV1 Wave Length Range: 190nm to 1100 nm Wave Length Accuracy : + 1 Band Width : 2.0 nm Photometric Range : -3.0 to 4.0 Abs Photometric Accuracy : + 0.002 Abs @ 0.5 Abs + 0.005 Abs @ 1.0 Abs Photometric Reproducibility : + 0.002 Abs @ 1 Abs Photometric Mode : Abs, T, Conc 4 Cell Holder type and length Cell Holder : 1-100 mm inbuil - 10 mm auto cell changer for measuring beam 10 mm holder at reference beam Data storage: Floppy and Memory Computer interface port with cable RS 232C Tungsten Halogen Lamp + UV lamp (D2) Silicone Photodiode 1.44 mb floppy disk drive Quartz Cell 1 CM, 2 CM, 4 CM, 5 CM and 10 CM Cell Holder for variable length upto 10 CM for measuring beam Cell holder for variable length upto 5 CM for reference beam

5 6 7 8

Data Processing Source Detector Accessories

Shree Renuka Sugars Ltd., Haldia Filter Assembly - Technical Specification Doc No. REN/ HAL/ LAB/ INT / SPEC-06 SNO. Instrument Particulars 1 Make 2 Specification Bharati Chemicals Softtech Filter funnel - 250 ml borosilicate glass with clamp with neoprene cover and neoprene cord Memberane filter 47 mm - 0.45 um pack 100 Vacuum flask 1 lit Syringe type nylon membrane filter 25 mm / 0.45 um with needle Flexible tubing 1 Mt Wooden box to house all the above (including vacuum pump)

Shree Renuka Sugars Ltd., Haldia Vacuum Pump - Technical Specification Doc No. REN/ HAL/ LAB/ INT / SPEC-07 SNO. Instrument Particulars 1 Make 2 Specification Details Riviera Oil free Vacuum pump 0 - 720 mm of Hg With Vacuum release valve

Shree Renuka Sugars Ltd., Haldia Analytical Balance - Technical Specification Doc No. REN/ HAL/ LAB/ INT / SPEC-08 SNO. Instrument Particulars 1 Make 2 Specification Details Sartorius BT 224 S Capacity - 220 Grams Readability : 0.0001 g Linearity : 0.0002 grams Auto calibration with internal weight 80 mm Interface RS 232 port along with cable

3 4

Pan Size Accessories

Shree Renuka Sugars Ltd., Haldia Technical / Cost Comparitive Statement For pH Meter Doc No. REN/ HAL/ LAB/ INT / SPEC-09 SNO. Instrument Particulars 1 Make 2 Specification Details Eutech Singapore - Cyberscan pH 510 pH Range : 0.00 to 14.00 Resolution & Accuracy : 0.01 pH & + 0.01 Temperature Range : 0.0 to 100 Deg C Temperature Compensation Automatic Parallel Multipoint Calibration Parallel temperature display

Accessories

Epoxy body general purpose combined electrode Temperature Probe TEMP-01P, Buffer Gel for pH 4.01, 7.00 & 10.01 Stand with flexible arm & electrode holder

Shree Renuka Sugars Ltd., Haldia Conductivity Meter - Technical Specification Doc No. REN/ HAL/ LAB/ INT / SPEC-10

SNO. Instrument Particulars 1 Make 2 Specification

Details Cyberscan PC 510 Can Measure pH and conductivity together Conductivity Resolution 0.01 uS/CM to 0.01 mS/CM 5 Point calibration Temperature compensation automatic Calibration upto 5 Points Measurement range 0.5 uS/cm to 199.9mS/cm Temperature Probe pH Electrode Conductivity Cell Integral electrode holder