Designation: D 1742 94

AMERICAN SOCIETY FOR TESTING AND MATERIALS 100 Barr Harbor Dr., West Conshohocken, PA 19428 Reprinted from the Annual Book of ASTM Standards. Copyright ASTM

An American National Standard

Standard Test Method for

Oil Separation from Lubricating Grease During Storage1


This standard is issued under the xed designation D 1742; the number immediately following the designation indicates the year of original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A superscript epsilon (e) indicates an editorial change since the last revision or reapproval. This standard has been approved for use by agencies of the Department of Defense.

1. Scope 1.1 This test method covers the determination of the tendency of a lubricating grease to separate oil during storage in both normally lled and partially lled containers. 1.2 This test method is not suitable for greases softer than NLGI No. 1 grade. 1.3 Values stated in acceptable SI units are to be regarded as the standard except for the dimensions in Fig. 2 where inch-pound units are standard. 1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. For a specic hazard statement, see 7.1. 2. Referenced Documents 2.1 ASTM Standards: D 235 Specication for Mineral Spirits (Petroleum Spirits) (Hydrocarbon Drycleaning Solvent)2 E 11 Specication for Wire-Cloth Sieves for Testing Purposes3 2.2 Other Standard: NLGI Grease Consistency Classication4 3. Terminology 3.1 Denitions: 3.1.1 lubricating grease, na semi-uid to solid product of a thickener in a liquid lubricant. 3.1.1.1 DiscussionThe dispersion of the thickener forms a two-phase system and immobilizes the liquid lubricant by surface tension and other physical forces. Other ingredients are commonly included to impart special properties. 3.1.2 thickener, nin lubricating grease, a substance composed of nely-divided particles dispersed in a liquid lubricant to form the products structure.
1 This test method is under the jurisdiction of the ASTM Committee D-2 on Petroleum Products and Lubricants and is the direct responsibility of Subcommittee D02.G on Lubricating Grease. Current edition approved Jan. 15, 1994. Published March 1994. Originally published as D 1742 60. Last previous edition D 1742 88. 2 Annual Book of ASTM Standards, Vol 06.04. 3 Annual Book of ASTM Standards, Vol 14.02. 4 Available from NLGI, 4635 Wyandotte St., Kansas City, MO 64112.

FIG. 1 Pressure Bleeding Test Cell A

3.1.2.1 DiscussionThe thickeners can be bers (such as various metallic soaps) or plates or spheres (such as certain non-soap thickeners), which are insoluble or, at most, only very slightly soluble in the liquid lubricant. The general requirements are that the solid particles be extremely small, uniformly dispersed, and capable of forming a relatively stable, gel-like structure with the liquid lubricant. 3.1.3 oil separation, nthe appearance of a liquid fraction from an otherwise homogeneous lubricating composition. 4. Summary of Test Method 4.1 The sample of grease, supported on a 75-m (No. 200) sieve, is subjected to 1.72 kPa (0.25 psi) air pressure for 24 h

D 1742

in. 0.0021 0.0029 0.010 116 18 316 14 516 38

mm 0.053 0.074 0.25 1.59 3.17 4.76 6.35 7.94 9.52

in. 158 134 2 212 31116 418 4.360 4.365 438

mm 41.3 44.4 50.8 63.5 93.7 104.8 110.74 110.87 111.1

in. 0.500 0.510 0.520 58 34 1 114 138 3

mm 12.70 12.95 13.20 15.87 19.05 25.4 31.8 34.9 76.2

in. 4716 412 41116 434 42632 514 6 612

mm 112.7 114.3 119.1 120.6 123.03 133.3 152.4 165.1

FIG. 2 Detailed Drawing of Pressure Bleeding Test Cell A

at 25C (77F). Any oil seepage that occurs drains into a beaker and is weighed. 5. Signicance and Use 5.1 When a lubricating grease separates oil, the remaining composition increases in consistency. This can affect the ability of the product to function as designed.

5.2 It has been found that the results of this test correlate directly with the oil separation that occurs in 35-lb pails of grease during storage. 5.3 This test method is not intended to predict oil separation tendencies of the grease under dynamic conditions.

D 1742
7. Reagents 7.1 Mineral Spirits, conforming to Specication D 235. (WarningSee Note 1.)
NOTE 1Warning: Combustible. Vapor harmful.

8. Sampling 8.1 The sample presented for analysis should be large enough to make possible the selection of a representative portion for testing. 8.2 Examine for any indication of non-homogeneity such as oil separation, phase changes or gross contamination. If any abnormal conditions are found, obtain a new sample. 9. Preparation of Apparatus 9.1 Thoroughly clean the strainer, cup, and oil-collecting beaker with mineral spirits and follow by air drying. 9.2 Carefully inspect the sieve to make certain that there is no build-up which would affect the passage of oil. The sieve must also be free of any surface irregularities such as creases, dents, or punctures. Replace, if necessary. 9.3 The funnel must be clean and free of any residues which can retard the ow of oil. 9.4 Inspect the cover and cup for any dents or distortions to the mating surfaces which can interfere with a tight seal between the units. 9.5 When the O-ring seal shows wear it must be replaced. 10. Procedure 10.1 Place the strainer, screen side up, in the funnel and tare the assembly to the nearest 0.05 g. By means of a spatula, completely ll the space between the screen and the top of the funnel with grease. The volume occupied by the grease shall have a depth of 12.7 6 0.3 mm (0.506 0.01 in.). Avoid unnecessary working and entrainment of air. Use a straight edge to remove excess grease and provide a level surface. Avoid forcing any of the grease through the screen. Determine the mass of the sample to the nearest 0.05 g. 10.2 Place the tared beaker in the positioning retainer at the center of the bottom cup, and assemble the entire apparatus as indicated in Fig. 1. By means of a suitable length of rubber tubing, connect the source of air to the air inlet tting in the cover. 10.3 Bring the apparatus and the sample to 25 6 1C (77 6 2F). Before starting the test, regulate the air pressure to 1.72 6 0.07 kPa (0.25 6 0.01 psi) and continue the test for 24 h at this pressure and temperature. 10.4 At the end of the test, weigh the beaker containing the collected oil to the nearest 0.05 g and record. 11. Report 11.1 Report the mass of the oil separated as a percentage of the total mass of the grease sample. Record the value to the nearest 0.1 %. 12. Precision and Bias 12.1 The precision of the test method is not known to have been obtained in accordance with currently accepted guidelines in Committee D2 research report RR: D02-1007, Manual on
3

FIG. 3 Pressure Bleeding Test Cell B

6. Apparatus 6.1 Test Apparatus5The assembled apparatus consists of a tightly tting cup and cover which contains a 75-m (No. 200) sieve strainer for supporting the grease, a funnel for collecting separated oil, and a 20-mL beaker for retaining the separated oil. A tting is provided in the cover for inlet air (1.72 kPa (0.25 psi)) and a hole is provided in the side of the cup to prevent back pressure. Fig. 1 is a photograph of a disassembled apparatus. The details of construction are given in Fig. 2. The 75-m (No. 200) stainless steel sieve shall conform to the requirements of Specication E 11. The diameter of the 75 m (No. 200) sieve shall be 104.8 mm (418 in.), completely clear with no solder showing. 6.2 Air Pressure Supply and RegulationAn air pressure supply, controlled by reducing valves or regulators, capable of maintaining air pressure at 1.72 6 0.07 kPa, should be used. A manometer, or other suitable pressure indicating device, and a pressure relief valve to protect against pressure surge should be included in the pressure system.
5 The apparatus shown in Fig. 1 and Fig. 2 is available from Koehler Instrument Co., Inc., 1595 Sycamore Ave., Bohemia, LI., NY. 11716 as Catalog No. K189; and StanhopeSeta Ltd, Park Close, Engleeld Green, Egham, Surrey TW20 OXD England. Other appropriately designed equipment can be used such as that shown in Fig. 3, Fig. 4, and Fig. 5 but which is no longer available.

D 1742

in. 0.0021 0.0029 0.010 116 18 316 14 516 38

mm 0.053 0.074 0.25 1.59 3.17 4.76 6.35 7.94 9.52

in. 158 134 2 212 31116 418 4.360 4.365 438

mm 41.3 44.4 50.8 63.5 93.7 104.8 110.74 110.87 111.1

in. 0.500 0.510 0.520 58 34 1 114 138

mm 12.70 12.95 13.20 15.87 19.05 25.4 31.8 34.9

in. 4716 412 41116 434 42632 514 6 612

mm 112.7 114.3 119.1 120.6 123.03 133.3 152.4 165.1

FIG. 4 Detailed Drawing of Pressure Bleeding Test Cell CStrainer Assembly

Determining Precision Data for ASTM Methods on Petroleum Products and Lubricants.6 12.1.1 Repeatability (Note 2)The difference between two test results, obtained by the same operator with the same apparatus under constant operating conditions on identical test

material, would in the long run, in the normal and correct operation of the test method, exceed 0.1 M in only one case in twenty where:
M 5 mean of two determinations (1)

Annual Book of ASTM Standards, Vol. 05.03.

12.1.2 Reproducibility (Note 2)The difference between two single and independent results obtained by different operators working in different laboratories on identical test
4

D 1742

FIG. 5 Pressure Bleeding Test Cell C

material would, in the long run, in the normal and correct operation of the test method, exceed 0.17 M in only one case in twenty:
M 5 mean of two determinations (2) NOTE 2These values of precision are based on results by 10 laboratories using four greases that varied in consistency from NGLI Grade No. 1 to Grade No. 3.

from lubricating grease has no bias because the value of the oil loss is dened only in terms of a test method. 13. Keywords 13.1 lubricating grease; oil bleed; oil separation

12.2 BiasThe procedure for measuring the oil separation


The American Society for Testing and Materials takes no position respecting the validity of any patent rights asserted in connection with any item mentioned in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk of infringement of such rights, are entirely their own responsibility. This standard is subject to revision at any time by the responsible technical committee and must be reviewed every ve years and if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards and should be addressed to ASTM Headquarters. Your comments will receive careful consideration at a meeting of the responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should make your views known to the ASTM Committee on Standards, 100 Barr Harbor Drive, West Conshohocken, PA 19428.