Anda di halaman 1dari 11

TABLE OF CONTENTS

1. Features ................................................................................ 2
3. Operating Instructions ........................................................ 6
3.1. AC+DC 1.2W Power(W) and Power Factor (PF)
measurement ........................................................... 6
3.2. AC+DC Voltage Measurement ................................. 8
3.2.1. V + Hz Dual Display ........................................ 8
3.2.2. V + A Dual Display .......................................... 9
3.3. AC+DC Current Measurement .................................. 10
3.3.1. A + Hz Dual Display...................................... 10
3.3.2. V + A Dual Display ........................................ 11
3.4. AC+DC 1.2W Apparent/Reactive Power
Measurement (KVA+KVAR) ................................... 11
3.5. Phase Angle Measurement .................................... 12
3.6. 3. AC+DC Balanced Power Measurement ........... 14
3.6.1. W+ PF Dual Display ..................................... 14
3.6.2. KVA + KVAR Dual Display ............................ 15
3.6.3. V + A Dual Display ........................................ 15
3.7. 3.3W AC+DC Unbalanced Power Measurement . 16
3.8. 3."W AC+DC Unbalanced Power Measurement . 21
3.9. 1.3W Power Measurement ................................... 27
3.10. Calculation of 3.4W Power Factor (PF) ................ 31
3.11. Improving Power Factor of a 3.4W Power
System.................................................................... 31
3.12. Improving Power Factor of a 3. Balanced Power
System.................................................................... 31
3.13. Improving Power Faclor of a 1.2W Power
System.................................................................... 32
3.14. Phase Sequence Indication (for PF > 0.1) ............. 33
3.15. Recording Data in Memory .................................... 35
3.16. Reading Data in Memory ....................................... 35
4. Specifications..................................................................... 36
5. Battery Replacement ......................................................... 39
6. Maintenance 8 Cleaning .................................................... 40
1. Features
1. 3.4W, 3.3W, 3. Balanced, 1.2W and 1.3W Power
Measurement.
2. AC+DC true power, and True RMS AC Voltage and
Current
3. AC+DC 2000A. AC 600V, DC 800V
4. AC+DC 1200KW(1.), AC+DC 2000KW (3.)
5. Dual Display V+Hz, A+Hz, W+PF, KVA+KVAR, V+A
6. Phase Angle Measurement
7. 3. RST(L1L2L3) Sequence Indication.
8. Memory of 4 records.
9. DCA/DCW Auto Zero when power clamp powering on.
10. AC/DC Auto Detection
11. Auto range.

3./1.
POWER
CLAMP
MODFL : 4500
USER MANUAL
1 2
3 4
1. Panel Description
1. Transformer Jaw
This is used to pick up current signal. To measure AC+DC
current or AC+DC power, conductor must be endosed by the
jaw.
2. Transformer Trigger
This is used to open the jaw.
3. Data Hold Button
Once this button is pressed, reading will be held in the LCD.
Press again to release it.
4. Function Selection and On/Off Switch
This is used to sealed the function users desired, such as
KW, V, A, Phase, KVA or 3..
5. LCD
This is a 4 digits Liquid Crystal Display with maximum
indication of 9999. Function symbols, units, sign, decimal
points, low battery symbols, and zero symbol are included.
6. Units Symbols
Once a function is selected, corresponding unit (KW, V, A,
Phase, KVA, or 3.) will be displayed in LCD.
7. 3.3W and 3.W Select Button
If the 3. system is not a balanced system, users can press
this button to select between 3.3W, 3.4W or a balanced
system. Once pressed, LCD will show symbol of 3.3W or
3.4W depends on the function selected.
8. Read/Next Button
When the rotary switch is set at 3. function, the button is
used as a NEXT button In the 3. balanced system mode,
pressing the NEXT button will enable users to select W+PF,
KVA+KVAR or V+A to be displayed.
In the 3.3W system mode, Pressing the NEXT button to store
is pressed, the data number will be shown in LCD. A REC
symbol will be shown in LCD if any data is stored. If the memory
is full, FULL will be show in LCD. To clear memory, users need
to turn the power off and on again.
13. Low Battery Symbol
When this symbol appears, it means the battery voltage
drops below the minimum required voltage. Refer to Section
V for battery replacement.
14. REC and NO. symbols
If users observe REC is showri. that means data is stored in
memory. If users observe both REC and NO. are shown, that
means readings shown in LCD are data stored in memory not
current data measured.
the measured values W
RS(L1L2)
and W
TS(L3L2)
After two values are
measured and stored, the microprocessor inside the power
Clamp will add the two vaIues together, display the result in LCD
and show the symbol of W
RST
to represent W
3.3W.
To start another
W
3.3W
measurement, press the NEXT button again.
In the 3.4W system mode. Pressing the NEXT button to store
the measured values W
R(L1)
W
S(L2)
and W
T(L3)
After three values are
measured and stored, the microprocessor inside the clamp will
add the three values together, display the result in LCD and show
the symbol ot W
RST
to represent W
3.4W
To start another W
3.4W
measurement, press lhe NEXT button again.
If the rotary is not set at the 3. function, the button is used as
a READ button.
If users ever store data in the memory by:pressing REC button,
pressing the READ button will retrieve the data from the memory.
First the data number will be shown in LCD, then the data stored.
Once READ fwction is enabled, symbol of REC and NO. will be
shown in LCD to indicate the power clamp is in READ mode. The
reading shown in LCD is not current reading but data stored in
memory. To exit READ function. turn the rotary switch to change
funclion.
9. V input Terminal
This terminal is used as input for voltage measurements.
10. COM Terminal
This terminai is used as common reference input.
11. DC A/W ZERO button
When users find the reading of A or W is not zero, press this
button once will zero the A or W reading (Users do not need
to press and hold the button). When the power clamp is doing
the zero action, a symbol of ZERO will be shown in LCD.
12. REC button
The clamp meter can store 4 data in memory. Once the button
5 6
7 8
3. Operating instruction
3.1 AC + DC 1.2W Power (W) and Power Factor (PF)
measurement.
WARNING :
Do not clamp on to any conductor when turning on the power of
power clamp. Because the power clamp will aulo zero the
residual magnet remained in the jaws at the moment the power
clamp is tumed on. If users clamp on to conductor with current
flowing through, the reading measured later will be inaccurate.
3.1.1. Turn the power on without jawsdamping on to any wire.
3.1.2. Set the rotary switch at W (refer to figure 2).
3.1.3. If the watt reading is not zero. press the DCA/DCW
ZERO button once to zero the watt reading
3.1.4. Insert the test leads into the inpul jack.
3.1.5. Connect the test prod of COM (black) terminal to the
neutral line.
3.1.6. Connect the test prod of V (red) terminal to the power
line.
3.1.7. Clamp on to the line where V (red) terminal is
connected.
3.1.8. The power clamp with automatically select proper
range.
3.1.9. Read the Watt and PF values displayed in LCD.
NOTE :
The + sign printed on jaw must face the power source for
accurate measurement.
WARNING : Maximum input for DC V is 1000, and for AC V is
750. Do not attempt to take any voltage measurement that
exceeds the limits. Exceeding the limits could cause electrical
shock and damage to the clamp meter.
a. Set the rotary switch at V (refer to figure 3).
b. Insert the test leads into the input jack.
c. Connect the test prods of the test leads in PARALLEL to
the circuit to be measured.
d. The power clamp will automatically select proper range.
e. Read the voltage and Frequency values displayed on the
LCD.
NOTE :
The sensitivity for voltage frequency measurement is 1V. and the
frequency range is 10 - 400Hz. If the frequency is < 10 Hz, LCD
will show 0 Hz. If the frequency is > 400 Hz, LCD will show OL.
3.2.2. V + A Dual Display
Follow Ihe instruction in section 3.6.3 3. AC+DC Balanced
Power Measurement V + A. Then users can monitor V and A
at the same time from LCD.
3.2. AC + DC Voltage Measurement.
3.2.1. V + Hz Dual Display
9 10
11 12
3.3 AC+DC Current Measurement
3.3.1. A + Hz Dual Display
WARNING :
1. Do not clamp on to any conductor when turning on the power
of power clamp.
2. Make sure that all the test leads are disconnected from the
power clamps terminals for current measurement.
a. Set the rotary switch at A (refer to figure 4).
b. Push and hold the DCAIDCWZERO once to zero the
reading.
c. Press the trigger to open the jaw and fully endose the
conductor to be measured. No air gap is allowed
between the two half jaws.
d. The power clamp will automatically select proper range.
e. Read the current and frequency values displayed in
LCD.
NOTE :
The sensitivity for current frequency measurement is 5A, and
the frequency range is 10 - 400Hz. If the trequency is < 10 Hz,
LCD will show 0 Hz. If the frequency is > 400 Hz, LCD will show
OL.
3 3.2. V + A Dual Display
Follow the instruction in section 3.6.3 3g AC+DC Balanced
Power Measurement V + A. Then users can monitor V and A
at the same time from LCD.
3.4. AC+DC 1.2W Apparent/Readive Power Measurement
(KVA+KVAR)
Before any measurement, zero the current (A) reading. Then set
the rotary switch at KVA. The rest of the procedvres are the same
as section 3.1 (AC+DC 1.2W Power(W) and Power Factor (PF)
measurement). Refer to figure 2 for test leads connection and
clamping of jaw.
KVAR is a calculated value and its accuracy greatly depends on
the accuracy of V, A and KW. To obtain a more accurate KVAR
value when PF is greater then 0.91(.<25
0
) user can measure the
phase angle and obtain the KVAR by themselves from the following
equation for pure sine wave. KVAR = KVA * sin .
3.5.1. Make sure the current reading is zero by setting the
rotary switch at A. If the current reading is not zero,
press the DCA ZERO button.
3.5.2. Set the rolary swiich at Phase (refer to figure 5).
3.5.3. Insert the test leads into the input jack.
3.5.4. Connect the test prod of the COM (black) terminal to
Ihe reference line.
3.5.5. Connect the test prod of the V (red) terminal to the
voltage signal to be detected.
3.5.6. Clamp on to the wire where V (red) terminal is
connected.
3.5.7. If current signal is detected from the jaws, the phase
angle will be displayed in degree in LCO together with
the frequency of voltage.
3.5.8. If the current signal is not detected from the jaws, only
the frequency of voltage will be displayed while the
phase angle is left blank.
WARNING :
Before taking measurement, users must zero the current (A)
reading to avoid incorrect reading of phase angle. To check if
current is zero, turn the rotary switch to current (A) function.
NOTE :
INDUCTIVE LOAD : Negative phase angie means currenl signal
lags behind voltage signal. If connection is correct, negative phase
angle also means inductive load.
CAPACITIVE LOAD: Positive phase angle means current signal
leads ahead voltage signal. If connection is correct, positive phase
angle also means capacitive load.
NOTE:
The + sign printed on jaw must face the power source for correct
result.
3.5 Phase Angle Measurement
13
14
15 16
WARNING:
Before taking measurement. users must make sure the current
(A) reading is zero by setting the rotary switch at A position. If the
current reading is not zero, users might get an incorrect KVA and
KVAR values.
19 20
17 18
a. Tum the power on without clamping on to any wire.
b. Set the rotary switch at 3. (refer to figure 6).
c. LCD will show 3.3W, 3.".W, and BAL symbols indicate
balanced power measurement.
d. If the watt reading is not. zero. press the DCA/DCW
ZERO button once to zero the watt reading
e. Insert the test leads into the input jack.
f. Select one phase (eg. R or L1) as COM and Connect
the test prod of COM (black) terminal to that phase
(eg. R or L1).
g. Connect another test prod to the second phase
(eg. S or L2).
h. Clamp on to the third phase (eg. T or L3).
i. The power clamp will automatically select proper range.
j. Read the Watt and PF values displayed in LCD.
k. To see KVA and KVAR. press the NEXT button.
I. To see V and A, press the NEXT button again.
m. To return to display of W and PF, press the NEXT button
again.
3.6.2. KVA + KVAR Dual Display
Follws steps from a to j in section 3.6.1 Then press the
NEXT once, and wait for 1.5 seconds. The LCD display
will show the values of KVA and KVAR.
3.6.3. V + A Dual Display
Follows, steps from a to j in section 3.6.1. Then press the
NEXT twice, and wait for 1,5 seconds. The LCD display
will show the absolute values of V and A.
NOTE : Specific selection of each phase is not required in 3..
balanced power measurement. As long as each prod and the jaws
are connected to different phases, the reading is always correct.
The + sign printed on jaw must face the power source for accurate
measurement.
NOTE : To find out it load is capacitive or inductive, refer to section
3.5 Phase Angle Measurement for details.
3.6 3. AC+DC Balanced Powre Measurement
3.6.1. W + PF Dual Display
3.7. 3.3W AC+DC Unblanced Power Measurment.
23 24
21 22
3.7.1. Two measurements of W
RS
(or W
L1L2
) and W
ST
(or W
L3L2
) are
required
3.7.2. Firstly, measure W
RS
(or W
L1L2
) (refer to figure 7).
a. Turn the power on without clamping on to any wire.
b. Set the rotary switch at 3..
c. Press the 3.button once. and only the 3.3W symbol
is lett in LCD. At this momeni, the W
RS
symbol blinks to
instruct users to take measurement of W
RS
d. If the watt reading is not zero, press the DCA/DCW
ZERO button once to zero the watt reading
e. Insert the test leads into the input jack.
f. Select one phase (eg. S or L2) as COM and connect
the test prod of the COM (black) terminal to that phase
(eg. S or L2).
g. Connect the test prod of V (red) terminal to the second
phase (eg. R or L1).
h. Clamp on to the same phase as in step g. (eg. R or L1).
i. The power clamp will automalically select proper range.
j. Wait until the reading is stable, press the NEXT button,
and W
RS
(W
L1L2
) symbol will disappear. At this moment,
W
RS
is stored in the memory, and W
ST
(W
L1L2
) symbol
appears and blinks to instruct users to take measurement
of W
ST
(W
L3L2
).
3.7.3. Secondly, measure W
ST
, (or W
L3L2
) (refer to figure 8).
a. Disconnect the test prod from the phase where jaws is
clamp on in previous measurement.
b. Connect the test prod to the third phase (eg. T or L3).
c. Open and close the jaws, so power clamp is clamping
t o
nothing
d. If the watt reading is not zero, press the DCA/DCWZERO
button once to zero the watt reading
e. Clamp on to the third phase where test prod is connected
to (eg. T or L3)
f. The power clamp will automatically select proper range.
g. Wait until the reading is stable, press the NEXT button,
and W
ST
(or W
L3L2
) symbol will disappear. At this moment,
W
ST
(or W
L3L2
) is stored in the memory,
3.7.4. Once the NEXT button is pressed after measurements of
W
RS
(or W
L1L2
) and W
ST
) (or W
L3L2
), the power clamp will add
the two values together and show the result on Ihe LCD.
W
RST
symbol will be shown to indicate the watt of 3.3W
unbalanced Power (refer to figure 9). PF is not shown in
3.3W unbalanced power mode.
W
3.3W
= W
RST
= W
RS(L1L2)
+ W
ST(L3L2)
NOTE:
Once a phase is selected as COM, users can not change this
selection in the subsequent measurement. For example, if S
(or L2) phase is selected, S (or L2) phase is always connected to
the COM of power clamp during measurement of W
RS
(or W
L1L2
)
and W
ST
(or W
L1L2
) in 3. 3W unbalanced power.
NOTE:
The + sign printed on jaw must face the power source, and
make sure all the connections and clamping are corred for correct
measurement.
NOTE: In the 3.3W unbalanced power measurement, one of
W
RS
or W
ST
could be negative. So users must make sure all the
connections and clamping are correct to obtain correct power.
3.8. 3.4W AC+DC Unblanced Power Measurment.
25 26
28
27
3.8.1. Three measurements of W
R
(or W
L1
), W
S
(or W
L2
) and W
T
(or W
L3
) are required.
3.8.2. Firstly, measure W
R
(or W
L1
) (refer to figure 10).
a. Turn the power on without clamping on to any wire.
b. Set the rotary switch at 3..
c. Press the 3. button twice, and only the 3.4W symbol
is left in LCD. At this moment, the W
R
symbol blinks to
instruct users to take measurement of W
R.
d. If the watt reading is not zero, press the DCA/DCW ZERO
button once to zero the watt reading.
e. Insert the test leads into the input jack.
f. Connect the neutral line to the COM (black) terminal.
g. Connect the test prod of the V (red) terminal to the first
phase (eg. R or L1).
h. Clamp on to the same phase (eg. R or L1).
i. The power clamp meter will automatically select proper
range.
j. Wait until the reading is stable; press the NEXT button,
and W
R
symbol will disappear. At this moment, W
R
is
stored in the memory and W
S
symbol appears and blinks
to instruct users to take measurement of W
S
.
3.8.3. Secondly, measure W
S
(or W
L2
) (refer to figure 11)
a. Disconned the test prod from the phase where jaws is
clamp on in 3.7,2.
b. Connect the test prod of the V (red) terminal to the
second phase (eg. S or L2).
c. Open and Close the jaws, so Clamp meter is clamping
to nothing.
d. If the watt reading is not zero, press the DCA/DCW ZERO
button once to zero the watt reading
e. Camp on to phase where test prod is connected to
(eg. S or L2 phase)
31 32
29
30
f. The power clamp will automatically select proper range.
g. Wait until the reading is stable, press the NEXT button,
and W
S
symbol will disappear. At this moment, W
S
is
stored in the memory.
3.8.4. Thirdly, measure W
T
, (or W
L3
) (refer to figure 12)
a. Disconnect the test prod from the phase where jaws is
clamp on in 3.7.3.
b. Connect the test prod of the V (red) terminal to the third
phase (eg. T or L3 phase).
c. Open and dose the jaws, so power clamp is Clamping
to nothing.
d. If the watt reading is not zero, press the DCA/DCW ZERO
button once to zero the watt reading
e. Camp on to the phase where test prod ia aonneded to
(eg. T or L3).
f. The power damp will automatically select proper range.
g. Wait until the reading is stable, press the NEXT button,
and W
T
symbol will disappear. At this moment,W
T
is
stored in the memory.
3.8.5. Once the NEXT button is pressed after measurements of
W
R
W
S
and W
T
the power clamp will add the three values
W
R
, W
S
and W
T
together and show the result on the LCD.
W
RST
symbol will be shown to indicate the watt of 3.4W
unbalanced power (refer to figure 13). PF is not shown in
3.4W unbalanced power measurement.
W
3.4W
= W
RST
= W
R(L1)
+ W
S(L2)
+ W
T(L3)
NOTE :
The + sign printed on jaw must face the power source, and make
sure all the connetions and clamping are correct for correct
measurement.
Note :
In the 3."W unbalanced power measwement, all three W
R
or W
S
and W
T
must be positive. If users find one negative power, check
the connection of test leads and clamping ot jaw. Make sure all
the connections and clamping are correct to obtain correct power.
3.3. 1.3W Power Measurment.
35
33 34
36
d. If the watt reading is not zero, press the DCA/DCW
ZERO button once to zero the watt reading
e. Camp on to the L2 line where test prod is connected
to.
f. The power clamp will automatically select proper
range.
g. Wait until the reading is stabIe, press the NEXT
button. and W
ST
(or W
L2G
) symbol will disappear. At
this moment, W
ST
(or W
L2G
) is stored in the memory,
3.9.4. Once the NEXT button is pressed after measurements
of W
RS
(or W
L2G
) and W
ST
(or W
L2G
) the power clamp will
add the two values together and show the result on the
LCD. WR
RST
symbol will be shown to indicate the watt
of 1.3W unbalanced Power (refer to figure 16). PF is
not shown in 1.W unbalanced power mode.
W
1.3W
= W
RST
= W
RS(LlC)
+ W
ST(L2C)
3.10. Calculation of 3F4W Power Factor (PF)
3.10.1.As measuring 3F4W unbalanced power, use the REC
button to record the individual power factor (KW, PF,
KW, PF, KW and PF) of each phase in mernory.
3.10.2. With those data, 3F4W power factor can be calculated
by the fallowing equation.
KVA
3.4W
= KVA
R(LI)
+ KVA
S(L2)
+ KVA
T(L3)
PF
3.4W
=
KW
3.4W
KVA
3.4W
KVA
R(LI)
=
KW
R(LI)
; KVA
S(L2)
=
KW
S(L2)
; KVA
T(L2)
=
KW
S(L2)
PF
R(LI)
PF
S(L2)
PF
T(L3)
3.11. Improving Power Factor of a 3.4W Power System
3.11.1. Measure KVAR
R
(or KVAR
L1
), KVAR
S
(or KVAR
L2
),
KVAR
T
(or KVAR
L3
) values of each phase.
3.11.2. Based upon the measured values, users can purchase
required 3. or 1. capacitor at rated voltage and
frequency to improve power factor.
3.11.3.If value of capacitance is needed, users can obtained
the value by the following equation.
where
f : frequency in Hz
V : phase voltage
3.11.4. It is recommended that the KVAR value of the capacitor
should be a little less than the value measured.
3.12. Improving Power Factor of a 3. Balanced Power System
3.12.1. Measure KVAR value ef.a balanced system.
3.12.2. Based upon the measured value, users can purchase
required 3. capacitor at rated voltage and frequency
to improve power factor.
3.12.3.If value of capacitance is needed, users can obtained
the value by the
following equation.
where
f : frequency in Hz
V : phase voltage
3.12.4. It is recommended that the KVAR value of the
capacitor should be a little less than the value
measured.
Capacitance (Farad) =
KVAR *1000
2FfV
2
Capacitance (Farad) =
KVAR *1000
2FfV
2
1.3W power measurement is similar to 3.3W unbalanced power
measurement except the nomendature is different.
3.9.1. Two measurements of W
RS
(or W
L1G
) and W
ST
(or W
L2G
)
are required
3.9.2. Firstly, measure W
RS
, (or W
L1G
) (refer to figure 14).
a. Tum the power on without damping on to any wire.
b. Set the rotary switch at 3.
c. Press the 3. button once, and only the 3.3W
symbol is left in LCD. At this moment. the W
RS
symboI
blinks to instruct users to take measurement of
W
RS(L1G)
.
d. If the watt reading is not zero, press the DCA/DCW
ZERO button once to zero the watt reading
e. Insert the test leads into the input jack.
f. Connect the test prod of the COM (black) terminal
to ground.
g. Connect the test prod of V (red) terminal to the
second phase (eg. L1).
h. Clamp on to the same phase as in step g. (eg. L1).
i. The power clamp will automatically select proper
range.
j. Wait until the reading is stable, press the NEXT
button, and W
RS
(W
L1G
) symbol will disappear. At this
moment, (W
RS(L2G)
) is stored in the memory and W
ST
(W
L2G
) symbol appears and blinks to instruct users
to take measurement of W
ST
(W
L2G
).
3.9.3. Secondly, measure W
ST
(or W
L2G
) (refer to figure 15).
a. Disconnect the test prod from the phase where jaws
is clamping on in the previous measurement.
b. Connect the test prod to the L2 line.
c. Open the jaws, so damp meter is clamping to
nothing.
3.13. Improving Power Factor of a 1.2W Power System
3.13.1. Measure KVAR value of a 1.2W power
system.
3.13.2. Based upon the measured value, users can
purchase required capacitor at rated voltage
and frequency to improve power factor.
3.13.3. If value of capacitance is needed, users can
obtained the value by the following equation.
following equation.
where
f : frequency in Hz
V : phase voltage
3.13.4. It is recommended that the KVAR value of the
capacitor should be a little less than the value
measured.
Capacitance (Farad) =
KVAR *1000
2FfV
2
40
37 38
39
NOTE :
The indication of phase sequence is not limited to 3. balanced
power. It can be used in general as long as the PF displayed is
greater than 0.1.
Set the power clamp In 3. balanced power mode, the power clamp
will automatically detect phase.
3.14.1. Set the rotary at 3..
3.14.2. Connect the voltage input (red terminal) to R (or L1)
phase, connect the COM input (black terminal) to S
(or L2) phase, and clamp on the T (or L3) phase. When
clamp on the T (or L3) phase, the power clamp must
face the power source.
3.14.3. If the phase sequence is clockwise, R, S, T sequence
will be shown and flashed. (refer to figure 17 )
3.14.4. If the phase sequence is counterdockwise, T, S,
R sequence will be shown and flashed. (refer to
figure 17)
3.14.5. If the current signal (where jaws damp on) is not
detected, only one of the R, S or T will be left in LCD.
NOTE :
The + sign printed on jaw must face the power source for correct
measurement.
3.15. Recording Data in Memory
The power clamp can store 4 data in its own memory. To store
any data displayed in LCD, press the REC button. If the power
clamp has stored 4 data already, LCD will display FULL. The
data is lost if power is turned off.
The REC symbol will be displayed in LCD if any data is stored.
NOTE:
If users observe REC (only) is shown in LCD, that means some
data is stored in memory.
3.16. Reading Data in Memory
To retrieve data in memory, press the READ button as long as
rotary switch is not set at 3.function.
Once the read button is pressed, the record number will be
first displayed. then data is displayed after. Once in the READ
function, the REC. NO. symbols will be displayed in LCD.
To exit the READ function, turn the rotary switch to change
function.
NOTE:
If users observe both REC and NO. symbols are shown in LCD,
users should know that the readings shown in LCD are not cur-
rent data measured. They are data stored in the memory.
4. Specifications (23
0
C 5
0
C)
AC DC True Power ( PF 0.2 - 1.0, 3.3W, 3.4W, 1.2W and
1.3W) :
Range Resolution Accuracy (of rdg) Range
0 - 0.01KW 2.0%0.05KW AC 600V, DC 800V,
99.99KW ACA/DCA 2000A
100 - 0.1KW 2.0%0.5KW AC 600V, DC 800V,
999.9KW ACA/DCA 2000A
1000 - 1KW 2.0%5KW AC 600V DC 800V,
1200KW ACA/DCA 2000A
KW Autoranging Map (PF 0.2 - 1.0, 3.3W, 3.4W, 1.2W and
1.3W)
0V - 200V 200V - 600V
0A - 200A 0.00 - 40.00KW 0.00 - 99.99KW
100.0 - 999.9KW
200A - 2000A 0.0 - 400.0KW 0.0 - 999.9KW
1000 - 1200KW
AC+DC True Power ( Power Factor 0.2 - 1.0, 3. Balanced
Power) :
Range Resolution Accuracy (of rdg) Range
0 - 0.01KW 2.0%0.5KW AC 600V, DC 800V.
99.99KW ACA/DCA 2000A
100 - 0.1KW 2.0%0.5KW AC 600V, DC 800V,
999.9KW ACA/DCA 2000A
1000 - 1KW 2.0%0.5KW AC 600V, DC 800V,
2000KW ACA/DCA 2000A
Power Factor (PF)
3.14. Phase Sequence Indication (for PF > 0.1)
PF =
KVA
KW
AC+DC Voltage ( True RMS, Crest Factor < 4, Autorange,
Overload Protection 800VAC for all range)
Accuracy (of reading) Input
Range Res. DC, 50/60Hz 40-400Hz Impedance
0-200V 0.1V 1.5%5dgts 2.0%5dgts 10M8
200-500V 0.1V 1.5%5dgts 2.0%5dgts 10M8
500-800V 1V 1.5%5dgts 2.0%5dgts 10M8
AC+DC Current (True RMS, Crest Factor < 4) :
Accuracy (of reading) Overload
Range Res. DC, 50/60Hz 40-400Hz Protection
0 - 200A 0.1A 1.5%5dgts 2.0%5dgts AC 3000A
200 - 500A 0.1A 2.0%5dgts 2.5%5dgts AC 3000A
500-2000A 1A 2.55dgts 3.0%5dgts AC 3000A
AC+DC KVA (Apparent Power)
(KVA) =
1000
AC+DC KVAR (Reactive Power)
(KVAR) = (KVA)
2
- (KW)
2
VA
44
41 42
6. Maintenance & Cleaning
Servicing not covered in this manual should only be performed by
qualified personnel. Repairs should only be performed by qualified
personnel. Periodically wipe the case with a damp doth and
detergent; do not use abrasive or solvents.
43

Certificate of Calibration
We hereby cer tify that this product has been
calibrated and found to be in accordance
with the applicable SPECIFICATIONS and
STANDARDS.
Accuracies of the standard equipment used in
this calibration are traceable to the National
Standards.
SR. NO. :
CHECKED BY :
DATE :
MODEL NO. :
MECO METERS PVT. LTD.
Block 9, Plot 270, 2nd Floor, Rup-Udey Niwas,
Sion (E), Mumbai - 400 022 (INDIA)
Correspondance Address :
Plot No. EL-1, MIDC Electronic Zone, TTC Industrial
Area, Mahape, Navi Mumbai - 400710 (INDIA)
Tel : 0091-22-27673311-16, 27673300 (Board)
Fax : 0091-22-27673310, 27673330
E-mail : sales@mecoinst.com
Web : www.mecoinst.com
Frequency ( if < 10 Hz, Hz = 0)
Range Accuracy Sensitivity
50/60 Hz 2dgts V: >1V, A: >5A
10 - 400 Hz 0.5% 2 dgts V: >1V, A: >5A
Indoor Use
Conductor Size : Cable .55mm. (approx.)
Bus Bar 65mm(D) x 24mm(W)
Battery Type : 9V
Display : 2 X 4 Digits Dual Display LCD
Range Selection : Auto
Overload Indication : OL
Power Consumption : 25 mA (approx.)
Low Battery Indication :
Sampling Time : 0.5 sec. (V and A)
1.6 sec. (W)
Operating Temperature : 4
0
C to 50
0
C
Operating Humidity : less than 85% relative
Altitude : up to 2000M
Storage Temperature : -20
0
C to 60
0
C
Storage Humidity : less than 75% relative
Dimension : 271mm (L) x 112mm (W) x 46mm (H)
10.7" (L) x 4.4" (W) x 1.8" (H)
Weight : 647 g/22.8 oz (batteries induded)
Accessories : Carrying bag x 1
Users manual x 1
9V battery x 1 (Installed)
5. Battery Replacement
When the low battery symbol is displayed on the LCD, replace
the old battery with new battery.
5.1. Turn the power off and remove the test leads from the
power damp.
5.2. Remove the screws of the bottom case.
5.3. Lift and remove the bottom case.
5.4. Remove the old battery.
5.5. lnsert new 9V battery.
5.6. Replace the bottom case and secure the screws.
WARNING :
do not touch or adjust any parts inside the power clamp when
the bottom case is open.
Phase Angle (must zero current reading before
measurement)
Range Accuracy Sensitivity Remark
50/60 Hz 2.0 V > 100V, A > 10A Zero crossing

Anda mungkin juga menyukai