MIT200 Series

Digital/Analog Insulation and Continuity


Testers
DESCRIPTION
The MI T200 i sone of the smallest i nsulati on testersavai lable
on the market today. Wi th opti onsof one, two and three
voltage test voltage i nstruments, the MI T200 i nstruments
offer a range of safety and operati on features.
The di splay offersa combi nati on of di gi tal readout and
analog di splay, usi ng Megger spatented DART di splay
technology, whi ch i ncludesthe benefi tsof an LCD di splay,
such asrobust, clear and accurate measurement, wi th an
analog poi nter response for evaluati ng ci rcui t charge and
di scharge characteri sti cs.
The i nstrument housi ng i si n tough ABS, desi gned to wi th-
stand the ri goursof hard use, and i ssmall enough to drop
i nto your pocket when not i n use.
Batteryrequi rementsare 6 AA batteri esof ei ther standard
alkali ne or ni ckel metal hydri de ( Ni MH) rechargeable type. A
low battery warni ng i ndi cator gi vesadvanced warni ng of
exhausted batteri es.
Continuity testing
Automati c conti nui ty testi ng i sperformed at 200 mA to
ensure compli ance wi th i nternati onal requi rements. There i s
no need to pressthe test button.
All i nstrumentswi ll measure up to 100 on conti nui ty, of
whi ch 0-10 i sperformed at greater than 200 mA to meet
i nternati onal electri cal testi ng requi rements.
Lead null i spossi ble up to 9.99 of test lead resi stance,
ensuri ng the abi li ty to null fused test leadsaswell as|
standard leads.
Continuity buzzer
A conti nui ty buzzer provi desa meansof rapi d cable testi ng
and ci rcui t i denti fi cati on, wi th voltage protecti on should you
acci dentally touch a li ve ci rcui t.
The buzzer operatesat a 2 threshold.
Insulation testing
The i nstrumentsoffer four confi gurati onsasdetai led on
page 2, provi di ng an i deal soluti on to most low voltage
i nsulati on testi ng appli cati ons.
I nsulati on testi ng i spossi ble up to 1000 M on all ranges.
Auto di scharge ensuresall ci rcui tsare safely di scharged after
testi ng.
1000 V i nsulati on test rangeshave a hi gh voltage warni ng
pri or to test voltage bei ng appli ed.
I Insulation testing to 1000 M
I Continuity testing at 200 mA down to
0.01
I Live circuit warning (voltage display) and
test inhibit
I Digital/Analog display
I Alkaline or rechargeable batteries
I 10C to +55C operating temperature
I CATIII 600 V
I Conf orms to EN61557-1
MIT200 Series
Digital/Analog Insulation and Continuity Testers
FEATURES AND BENEFITS SUMMARY
I Meetsthe i nternati onal EN61557 requi rementsof the
rated test voltage i nto a 1 mA load.
I Di gi tal di splay of i nsulati on measurement up to
1000 M on a logari thmi c analogue arc and a di gi tal
di splay.
I Conti nui ty range has0,01 resoluti on and a short ci rcui t
current i n excessof 200 mA.
I Automati c conti nui ty testi ng leavesboth handsfree. No
need to pressthe test button.
I Automati c power-off i f left unattended reduced wasted
battery li fe.
I Automati c voltage detecti on avoi dsacci dental contact
wi th li ve ci rcui ts.
I Test lead zero allowscompensati on for test lead
resi stance.
I Buzzer range operatesat < 2 .
Saf ety
Every Megger i nstrument i sdesi gned wi th safety asi ts
pri mary objecti ve. All i nstrumentsmeet or exceed the
requi rementsof safety di recti ve I EC 61010 and EN61557 for
i nsulati on and conti nui ty testi ng..
Def ault f ault meter
A bui lt-i n voltmeter automati cally swi tchesi n when the
i nstrument i sconnected to a ci rcui t wi th an AC or DC
voltage greater than 25 V.
Insulation Test inhibit
Ci rcui tsi n excessof 25 V wi ll generate a voltage warni ng.
Ci rcui tsover 50 V wi ll i nhi bi t testi ng of the i nsulati on,
protecti ng the operator and the i nstrument from i njury or
damage.
600V CATIII
The MI T200 seri eshasbeen desi gned for use on
appli cati onsup to 600 V CATI I I .
APPLICATIONS
The MI T200 seri eswi ll fi nd appli cati onsi n electri cal
contracti ng, both on domesti c and i ndustri al systems, as
well assi te mai ntenance and servi ce departments.
The MI T200 seri esof i nsulati on and conti nui ty testersare
i deal for testi ng transformers, motors, generators
swi tchgear, panel bui ldi ng, domesti c appli ances, power
toolsetc., aswell asfi xed electri cal wi ri ng systems.
Thei r small si ze and li ght wei ght make them i deal for those
engi neersthat need to carry them for extended peri ods.
All i nstrumentsmeet the requi rementsof most
I nternati onal Standardsi ncludi ng VDE0413 Part 1 and
BS7671 ( the 16th Edi ti on of the I EE Wi ri ng Regulati ons) .
MIT200 Series
Digital/Analog Insulation and Continuity Testers
MIT200 OPTIONS
Insulation testing
250 V I I
500 V I I I
1000 V I I
1000 M range I I I I
Auto-rangi ng I I I I
Auto di scharge I I I I
Test i nhi bi t I I I I
Li ve ci rcui t voltage di splay I I I I
Continuity testing
Conti nui ty @
> 200 mA I I I I
Conti nui ty to 0.01 I I I I
Test lead null ( 9.99 ) I I I I
Automati c conti nui ty test I I I I
Conti nui ty buzzer wi th 2
threshold I I I I
Default voltswarni ng I I I I
General
Di gi tal di splay + arc I I I I
Batterycondi ti on I I I I
Auto power down I I I I
Tough carry case I I I I
Test leads I I I I
CATI I I 600 V I I I I
Environmental
O perati on temperature -10C to + 55C
Storage temperature -20C to + 65C
I P rati ng I P40
M
I
T
2
0
0
M
I
T
2
1
0
M
I
T
2
2
0
M
I
T
2
3
0
MIT200 Series
Digital/Analog Insulation and Continuity Testers
SPECIFICATIONS
Insulation ranges
Nominal test voltage:
1000 V, 500 V, 250 V ( d.c.)
Measuring range
10 k -1000 M on all ranges
Terminal voltage on open circuit (d.c.):
-0% + 25% of rated voltage
Short circuit current:
205 mA + 10 mA 5 mA
Test current on load:
> 1 mA at mi ni mum passvaluesof i nsulati on asspeci fi ed i n
BS7671, HD384, I EC364 and VDE0413 part 1
Accuracy (at 20 C)
MIT200, 210, 220, 230:
3% of readi ng 2 di gi tsup to 10 M
5% of readi ng 2 di gi tsup to 100 M
30% of readi ng up to 1000 M
Terminal characteristics
Continuity ranges
Measuring range:
0,01 -100,0
( 0 -50 on analogue scale)
Open circuit voltage:
5 V 1 V
Short circuit current:
205 mA, 5 mA ( 0-10 ) > 20 mA ( 0 -100 )
Accuracy (at 20 C)
MIT200, 210, 220, 230:
0,01 to 9,99 3% 2 di gi ts
10,0 to 99,9 5% 2 di gi ts
Zero offset adjust:
MIT200, 210, 220, 230:
0 to 9,99
Continuity buzzer
MIT200, 210, 220, 230:
O peratesat < 2
Def ault voltmeter
MIT200, 210, 220,230:
> 25 V ac or dc. i sappli ed di splay wi ll operate asa voltmeter.
Test inhibit
I f more than 50 voltsi sdetected, testi ng wi ll be i nhi bi ted.
Range:
25 V to 600 V @ 50/60 Hz & dc
Accuracy:
25 V to 450 V ac/dc 1% 1 di gi t
450 V to 600 V ac 2% 1 di gi t
Auto power down
Auto power down operatesafter 10 mi nutesi f left i n standby mode.
Temperature and humidity
Operating range:
-10C to + 55C ( 14F to 130F)
Operating humidity:
93% R.H. at + 40C max. ( 93% R.H. at 104F)
Storage range:
-25C to + 65C ( -13F to 150F)
Environmental protection:
I P40
Fuses
Terminals:
500 mA ( F) 600 V, 32 x 6 mm Cerami c HBC 50 kA mi ni mum.
Di splay showsi f fuse i sruptured.
Saf ety
Meetsthe requi rementsof EN61010-1 Cat I I I 600V phase to earth.
Automatic discharge
After an i nsulati on test the i tem under test wi ll be di scharged
automati cally. Any voltage present wi ll be i ndi cated on the di splay
so that the di scharge can be moni tored.
Power supply
Battery 6 x 1,5 V cellsI EC LR6 type( AA alkali ne) .
Rechargeable Ni MH cellsmay be used.
Battery condi ti on i sconstantly shown on the di splay asa four-
secti on bar-graph.
Battery lif e
3000 consecuti ve tests( 5 secondsper test) on any test usi ng 2Ah
batteri es.
Weight
All units: 550gms approx. ( 1.2lbs)
Dimensions
All units: 195 x 98 x 40mm ( 7.7 x 3.9 x 1.6 i nches)
E.M.C
I n accordance wi th I EC61326-1
UK
Archcliff e Road Dover
CT17 9EN England
T +44 (0) 1304 502101
F +44 (0) 1304 207342
UNITED STATES
4271 Bronze Way
Dallas TX 75237-1088 USA
T 800 723 2861 (USA only)
T +1 214 333 3201
F +1 214 331 7399
OTHER TECHNICAL SALES OFFICES
Norristown USA, Sydney AUSTRALIA,
Toronto CANADA, Trappes FRANCE,
Kingdom of BAHRAIN, Mumbai INDIA,
Johannesburg SOUTHAFRICA and
Conjure THAILAND.
Registered to ISO 9001:2000 Cert. no. Q 09290
Registered to ISO 14001-1996 Cert. no. EMS 61597
MIT200_DS_us_V06
www.megger.com
Megger is a registered trademark
MIT200 Series
Digital/Analog Insulation and Continuity Testers
ORDERING INFORMATION
Item (Qty) Order No.
250 V/500 V Insulation and continuity tester MIT220-EN
250 V/500 V/1000 V Insulation and continuity tester
MIT230-EN
Non UK instruments
500 V Insulation and continuity tester MIT200-EN
1000 V Insulation and continuity tester MIT210-EN
Included accessories
Lead set with prods and clips 6220-779
Carry case 5410-419
Item (Qty) Order No.
Optional accessories
Replacement lead set 6220-779
Fused lead set 6220-789
Removable protective holster 5410-346
Carry pouch 6220-773