Cardio-Aid
360-B
Biphasic defibrillator
Direction for use
R-5656-02-V02
User manual
Cardio-Aid
360-B device........................................... 11
3. The Cardio-Aid
360-B device
100%adaptation of ERC 2010 BLS-AED recommendation
It supports the effort of minimizing CPR interruptions and unintended deliveries of shock,
which can cause the injury of the person performing resuscitation. With the defibrillator it
is possible to start charging before placing the paddle electrodes on the chest. In this case
there is no electric shock hazard, CPR can be continued. When placing the electrode on
the chest, charging is finished in quite a short time and the shock can be delivered.
According to the recommendation of ERC 2010 the first shock can be set to a minimum
level of 150 J.
In order to reach high efficiency, we set the energy of CA360-B in AED mode to 200-
200-270 J, this way most of the time the delivery of the first shock already stops the
fibrillation.
12/89 Cardio-Aid
360-B defibrillator
maintanance, assigments
3.1. Cleaning of the device
Warning!
Do not use any aggressive, acidic chemical for cleaning the device, even if they
are diluted. Do not use any surface damaging granular material or flammable
detergent. Do not sterilize the device and its accessories, only in case it figures in
the instructions of the accessory.
Clean the defibrillator and the cables with soft cloth or sponge, not with aggressive detergent.
During cleaning, make sure that no liquid should enter into the connectors of the device. Pay
special attention when cleaning the tube of the blood pressure measurement cuff, because if
any water gets into the tube, entering inside the device during the operation of the blood
pressure measurement device, it can damage the valve and the compressor.
For cleaning you can use alcoholic solution, ammonium solution with water, peroxide
solution or any non-aggressive domestic detergents dropped on a dampened sponge.
3.2. Regular and periodic maintenance
Continuously make sure that the device and its accessories should be clean and undamaged.
Gel and other materials sticked to it, must be cleaned off from the electrodes. The undamaged
status of the electrode cables (cable insulation, breaking) must be regularly checked.
The battery condition must be followed with attention, if the device indicates the running
down of battery, it must be connected to the mains, in order to be replaced.
After some years of operation the battery capacity can reduce to that extent that it cannot
guarantee the operation conditions described in the technical data. In such cases the device
gives an alarm, the battery must be replaced. Battery replacement can only be carried out by a
skilled person.
Periodically (in 1-2 years) the main parameters of defibrillator must be checked and if needed
a calibration must be made. The calibration can only be made by a skilled service person with
adequate qualifications.
The continuous and reliable operation can be ensured by the allowance of a wake-up test. If
we set it, then the defibrillator will automatically turn on once in a day and it will carry out
some test examinations. Operation of the wake-up self-test is detailed in section no. 13.3.6.5.
This test ensures a continuous check-up, in case of errors or operation faults it will
immediately give an alarm.
The expected lifetime of the device is 8 years.
Cardio-Aid