i
i
E. Zolotarevova et al. / Wear 265 (2008) 19141917 1915
Table 1
Summary of studied patients and total joint replacements
Case Sex Birth Implant type Implant duration (years)
H1 M 1940 ABG (Howmedica) 8.5
H2 F 1955 ABG (Howmedica) 9
H3 F 1933 Poldi/Ultima (Poldi) 8
K1 M 1927 PFC (J + J) 8
Case: three cases of total hip replacement (H1H3), one case of total knee replace-
ment (K1).
different centrifugations; Table 2) and the total number of inves-
tigated micrographs was 320 (20 micrographs per each sample).
Image analysis of all the micrographs was performed with pro-
gram Lucia (Laboratory Imaging). Three morphological parameters
were determined: Equivalent diameter (D), Circularity (C) and Elon-
gation (E). Equivalent diameter is a size feature derived from the
object area, giving the diameter of a circle with the same area as
the corresponding object: D=(4 Area/ )
1/2
. Circularity is calcu-
lated from the object area and perimeter: C =4 Area/Perimeter
2
;
C =1 for circles and C <1 for all other shapes. Elongation is cal-
culated as the ratio of maximum and minimum diameter of the
particle: E =MaxDiameter/MinDiameter; E =1 for circles and E >1
for all other shapes.
Distributions of all morphology parameters (C, D and E) were
characterized by their mean values and widths. Mean of the distri-
x )
and width of the distribution, , was calculated as estimated
standard
deviation
(
=
(1/(n
1)
i
(
x
i
)
2
),
where
x
i
is
the ith element of a distribution, n is the number of all ele-
ments in a distribution and summation runs through all elements.
The elements in the above summations were either morphol-
ogy parameters of single wear particles (Table 2) or their mean
values calculated for different centrifugation conditions/patients
(Figs. 2 and 3). Condence intervals from the values of were
not calculated, because we characterized widths of the distribu-
tions (Table 2) or variances of the data (Figs. 2 and 3) rather than
probabilities that morphological parameters are within a particular
range.
Fig. 1. SEM micrographs showing UHMWPE wear particles on 0.1- m polycarbonate membranes. Particles were separated from solution only by otation (a) and by
centrifugation at 500 g (b), 16,000 g (c) and 105,000 g (d). Particles were isolated from the rst patient (case H1 in Table 1).
1916 E. Zolotarevova et al. / Wear 265 (2008) 19141917
Table 2
patients with similar particle size distributions (Fig. 2) so that it
UHMWPE wear particle sizes and shapes obtained by image analysis of SEM micro-
graphs for all patients and centrifugation conditions
was readily possible to compare and average all results (Fig. 3).
After centrifugation, both average morphological parameters
Sample Equivalent diameter (D) Circularity (C) Elongation (E)
(Table 2) and histograms (Fig. 3) demonstrated that particle sizes
( m) ( m) () () () () (given by D) and shapes (given by C and E) remained more-or-less
H1/F 0.32 0.24 0.66 0.24 1.78 0.71
H1/C1 0.32 0.25 0.67 0.25 1.75 0.53
constant and did not change with centrifugation in any systematic
way. This suggested that centrifugation had negligible effect on
H1/C2 0.26 0.16 0.68 0.23 1.71 0.51 particle morphology.
H1/C3 0.30 0.22 0.67 0.26 1.75 0.62
Selection of the patients for this study was made as follows:
H2/F 0.30 0.36 0.66 0.26 1.92 0.83
H2/C1 0.24 0.21 0.70 0.23 1.70 0.49
H2/C2 0.29 0.26 0.65 0.26 1.79 0.77
H2/C3 0.23 0.17 0.70 0.24 1.76 0.64
Our isolation procedure based on HNO3 digestion [7] was applied
to samples from more than 100 patients. In 25 cases we corre-
lated total amounts of wear particles with osteolysis in various
H3/F 0.35 0.32 0.70 0.28 1.76 0.78 zones around TJR [13], while the remaining results are being pro-
H3/C1 0.36 0.27 0.62 0.25 1.82 0.67
H3/C2 0.31 0.22 0.64 0.25 1.74 0.66
H3/C3 0.30 0.26 0.67 0.25 1.75 0.67
K1/F 0.26 0.23 0.73 0.23 1.69 0.57
K1/C1 0.34 0.30 0.67 0.26 1.76 0.72
K1/C2 0.29 0.23 0.67 0.26 1.72 0.73
K1/C3 0.36 0.32 0.64 0.27 1.86 0.72
Samples are denoted as Case/Centrifugation, where Cases are TJRs described in
Table 1 and Centrifugations are (F or C1 or C2 or C3), where F = otation and
C1/C2/C3 = centrifugation at 500/16,000/105,000 g. Each numerical value in the
table was calculated from a distribution obtained by image analysis of exactly
twenty SEM micrographs with magnication 10,000 and real width 24 m. Mean
of the distribution, , was calculated as the arithmetic mean ( = 1/n i xi)
and width of the distribution, , was calculated as estimated standard deviation
(
=
(1/(n
1)
i
(
xi
)
2
),
where
x
i
is
the
ith
element
of
a
distribution,
n
is the number of all elements in a distribution and summation runs through all
elements.
3. Results
Illustrative SEM micrographs of in vivo UHMWPE wear particles
isolated from the rst patient are shown in Fig. 1. No qualitative dif-
ference was found between the particles separated by spontaneous
otation (Fig. 1a) and centrifugation at 500, 16,000 or 105,000 g
(Fig. 1bd). All micrographs showed both spherical and elongated
particles indicating that centrifugation did not impact on particle
morphology signicantly.
Quantitative information from SEM micrographs was conrmed
by image analysis, whose results are summarized in Table 2. Particle
morphology was described by equivalent diameter D, circularity C
and elongation E. After otation, the morphology of wear particles
was quite similar for all four cases (H1/F, H2/F, H3/F and K1/F in
Table 2). This resulted from the fact that we intentionally selected
Fig. 2. Particle size distributions in all four studied patients. Each column represents Fig. 3. Distributions of particle sizes (a) and shapes (b and c). Each column rep-
the average through all centrifugation conditions (which are denoted as F, C1, C2 and resents the average through all four patients (which are denoted as H1, H2, H3
C3 in Table 2); error bars were calculated as estimated standard deviations of the and K1 in Table 2); error bars were calculated as estimated standard deviations
corresponding average. The columns are given in quadruplets; the columns in each of the corresponding average. The columns are given in quadruplets; the columns in
quadruplet, from left to right, correspond to case H1, H2, H3 and K1 in Table 1, each quadruplet, from left to right, correspond to otation, centrifugation at 500 g,
respectively. 16,000 g and 105,000 g, respectively.
E. Zolotarevova et al. / Wear 265 (2008) 19141917 1917
cessed [14]. In 10 cases we compared morphology of the particles
isolated with or without centrifugation and found no qualitative dif-
ferences in SEM micrographs. Detailed quantitative morphological
study was performed in four cases listed in Table 1; this means that
the presented results are based on image analysis of 320 SEM micro-
graphs (4 patients 4 centrifugation speeds 20 micrographs per
sample).
4. Discussion and conclusion
From the physico-chemical point of view there is little chance
that centrifugation forces would change the morphology of
UHMWPE wear particles. Centrifugation at 105,000 g is generally
used for separation of cell organelles without any signicant dam-
age to them [15]. However, when such a possibility was suggested
for polyethylene wear particles [9], it was desirable to investi-
gate the problem and verify the correctness of both our own [7]
and other isolation techniques [1012], which all employ centrifu-
gation. There are two theoretical possibilities how centrifugation
could have inuenced wear particle morphology: (i) elongated par-
ticles might have been changed to spherical due to mutual collisions
and collisions with vessel edges at very high centrifugation speeds
and (ii) very small particles might have been isolated in higher
amounts at higher centrifugation rates. Nevertheless, neither of the
above-mentioned effects was observed and even ultracentrifuga-
tion at 105,000 g did not change the morphology of UHMWPE
wear debris.
Acknowledgement
This work was supported by grant of the Ministry of Education,
Youth and Sports of the Czech Republic (MSMT 2B06096).
References
[1] J. Charnley, Tissue Reactions to Implanted Plastics, Acrylic Cement in
Orthopaedic Surgery, Williams and Wilkins, Baltimore, 1970.
[2] R.A. Cooper, C.M. McAllister, L.S. Borden, T.W. Bauer, Polyethylene debris-
induced osteolysis and loosening in uncemented total hip arthroplasty, J.
Arthroplasty 7 (1992) 285290.
[3] T.P. Schmalzried, M. Jasty, W.H. Harris, Periprosthetic bone loss in total hip
arthroplasty, J. Bone Joint Surg. 74A (1992) 849863.
[4] E. Ingham, J. Fischer, The role of macrophages in osteolysis of total joint replace-
ment, Biomaterials 26 (2005) 12711286.
[5] H.W. Fang, Y.C. Ho, C.B. Yang, H.L. Liu, F.Y. Ho, Y.C. Lu, H.M. Ma, C.H. Juany, Prepa-
ration of UHMWPE particles and establishment of inverted macrophage cell
model to investigate wear particles induced bioactivites, J. Biochem. Biophys.
Methods 68 (2006) 175187.
[6] A.P.D. Elck, S.M. Green, S. Krikler, A. Unsworth, The nature and dissemination
of UHMWPE wear debris retrieved from periprosthetic tissue of THR, J. Biomed.
Mater. Res. 65A (2003) 95108.
[7] M. Slouf, S. Eklova, J. Kumstatova, S. Berger, H. Synkova, A. Sosna, D. Poko-
rny, M. Spundova, G. Entlicher, Isolation, characterization and quantication of
polyethylene wear debris from periprosthetic tissues around total joint replace-
ments, Wear 262 (2007) 11711181.
[8] S. Niedzwiecki, C. Klapperich, J. Short, S. Jani, M. Ries, L. Pruitt, Compari-
son of three joint simulator wear debris isolation techniques: acid digestion,
base digestion, and enzyme cleavage, J. Biomed. Mater. Res. 56 (2001)
245249.
[9] M. Visentin, S. Stea, S. Squarzoni, B. Antonietti, M. Reggiani, A. Toni, A new
method for isolation of polyethylene wear debris from tissue and synovial uid,
Biomaterials 25 (2004) 55315537.
[10] P. Campbell, S. Ma, B. Yeom, H. McKellop, T.P. Schmalzried, H.C. Amstutz,
Isolation of predominantly submicron-sized UHMWPE wear particles from
periprosthetic tissues, J. Biomed. Mater. Res. 29 (1995) 127131.
[11] J.D. Mabrey, A. Afsar-Keshmiri, G.A. Engh, C.J. Sychterz, M.A. Wirth, C.A.
Rockwood, C.M. Agrawal, Standardized analysis of UHMWPE wear parti-
cles from failed total joint arthroplasties, J. Biomed. Mater. Res. 63 (2002)
475483.
[12] J. Bell, J.L. Tipper, E. Ingham, M.H. Stone, J. Fischer, The inuence of phos-
pholipid concentration in protein-containing lubricants on the wear of
ultra-high molecular weight polyethylene in articial hip joints, Proceedings
of the Institution of Mechanical Engineers Part H-J. Eng. Med. 215 (2001)
259263.
[13] M. Slouf, D. Pokorny, G. Entlicher, J. Dybal, H. Synkova, M. Lapcikova, Z. Fej-
farkova, M. Spundova, F. Vesely, A. Sosna, Quantication of UHMWPE wear
debris in periprosthetic tissues of hip arthroplasty: description of a new method
based on IR and comparison with radiographic appearance, Wear. 265 (2008)
674684.
[14] M. Slouf, D. Pokorny, G. Entlicher, Z. Fejfarkova, H. Synkova, M. Lapcikova,
F. Vesely, A. Sosna, UHMWPE wear particles in different zones around total
hip replacements, in: Proceedings of UHMWPE, Madrid, Spain, 14.9.2007 and
15.9.2007, 2007, p. 27.
[15] R.H. Garrett, Biochemistry, Saunders College Publishing, 1995, p. 566.