CALIBRATION PROCEDURE (OPTIMA SCALE PLASTIC INDICATOR)

1. Turn on the scale by holding the ON/OFF key. (If the scale does not turn on, plug
in the ac adapter because the battery is low.)
2. Press the SET key and the ON/OFF key simultaneously to access the setup menu.
If done correctly, the display should now show C01.
3. Press the SET key to access the C1 channel. The display should show [C1 X].
4. Press the ZERO key to change the value of C1 to 2. (The value of 2 sets the scale
up in pounds.)
5. Press the SET key when the selection has been made. C02 should now be
displayed.
6. Press the SET key to access the C2 channel. The display should show [C2 X].
7. Press the ZERO key to change the setting to the desired decimal place. (The C2
channel is used to adjust the decimal point on the scale. A value of 1 means
there is one digit behind the decimal point.) Press the SET key when finished.
C03 should now be displayed.
8. Press the SET key to access the C3 channel. The display should show [C3 X].
9. Press the ZERO key to cycle through the values until the desired graduation
appears. (The C3 channel adjusts the divisions on the scale. For example, a
value of 1 selected and C2 set to 1, the scale will read in 0.1 lb increments.)
Press SET to select the value. C04 will now be displayed.
10. Press the SET key to access the C4 channel. The display will show [XXXXXX].
11. Enter in the maximum capacity you want to use for this scale by using the UNIT
and HOLD keys to move the cursor left and right, and the TARE and ZERO keys
move the values down and up. (The C4 channel is used to enter in the max
capacity of the scale; Make sure this doesnt exceed the max capacity of the
scale; Max capacity divided by the increment set in C02 and C03 above cannot
exceed 5000). Press the SET key to set the value. C05 will now be displayed.
12. Press the SET key to access the C5 channel. The display should show [C5 0].
13. The C5 channel calibrates zero on the scale. Make sure the scale is empty.
14. Press the ZERO key to change the value to 1.
15. Press the SET key. The display will count down from 10 while the scale is
calibrating zero. 0 will be on the display when the zero calibration is complete.
16. Press the SET key. C06 should be displayed.
17. Press the SET key to access the C06 channel. The display will show [C6 0].
18. The C6 channel is used to calibrate the scale with a known weight. Press the
ZERO key to set the value of C6 to 1. Press the SET key. The display will flash
SPAN, and then show [XXXXXX].
19. Enter the calibration weight value you will use (at least 10% of max capacity you
set in C04 by using the UNIT and HOLD keys to move the cursor left and right,
and the TARE and ZERO keys move the values down and up.
20. Place the calibration weight you have on the empty scale.
21. Once the load has stabilized, press the SET key.
22. The scale will count down to 0. Once 0 has been reached, the display will show
CALEND.
23. Press the SET key. C07 will now be displayed.
24. Press the ON/OFF key once (do not hold) to save and exit the setup menu.
25. The scale has now been calibrated. The display will show the value of the
calibration weight you put on.
26. If the scale does not show the value of the calibration weight check that the feet
on the platform are not screwed in too tightly, verify that the platform is level.
27. Unload the scale; the display should read 000000.
28. If the scale does not display 00000, check that the feet on the platform are not
screwed in too tightly; Verify that the platform is level.