Anda di halaman 1dari 5

We confirm that, the instrument meets or exceeds the manufacturers published specifications at the

points tested. All measurements are traceable to national and/or international standards or have been
derived by approved ratio techniques. This certificate may not be reproduced other than in full. Calibration
certificates without signature are not valid.
Son Calibration location
Certificate of Calibration
G.J.J. Sprik
5
Model
Description
Manufacturer
Serial number
Inventory number
Tested by
Customer
FLUKE NETWORKS
DTX-SFM2
SINGLEMODE FIBER MODULE
2001611
-
A.R Figarella Gomez
TELFONE LDA
PORTO
Site number
Number of pages:
Date of calibration
Date of recalibration
Certificate No.:
Tel: +31 (0)40 2675308 Fax: +31 (0)40 2675321
Fluke Nederland B.V. / Science Park Eindhoven 5108 / 5692 EC Son
Head of laboratory
This calibration is performed by a DEKRA certified lab for ISO 9001:2008
Issue date:
CTAG:feacd03690ee3441ac9e6da2ed65807a
5870968
1342910
03 Sep 2014
03 Sep 2015
03 Sep 2014
Certificate of Calibration
Certificate No.:
Issue date:
2 of 5 Page:
1342910
03 Sep 2014
- If the unit under test is used under rough conditions we recommend to decrease the calibration interval period,
the calibration interval (due date) is the responsibility of the end user;
- According to the European norm 'Operation of electrical installations' NEN-EN 50110-1 release 2005 and the
Dutch norm NEN 3140 release 2011 paragraph 5.102.12 through 5.102.16, is a safety test not required.
Therefore not performed.
(45 20) %rh
FLUKE :DTX-SFM2 :CAL :RS232 :CNET :FIB 1.40
0
Conform specifications
Completed
(23 3) C
IDENTIFICATION:
CALIBRATION CONDITIONS:
SUMMARY CALIBRATION INFORMATION:
Unit under test
Serial number
Inventory number
Environmental temperature
Humidity relative
Calibration procedure Rev:
Procedure
Outgoing status
Failed test(s)
REMARK:
Remarks
-
2001611
DTX-SFM2
Certificate of Calibration
Certificate No.:
Issue date:
3 of 5 Page:
1342910
03 Sep 2014
Standards and test-equipment used for this calibration:
Model: Serial No: Inventory No: Due to: Certificate No:
IQS-510 228460 SL0538 28 Nov 2014 1213131
IQS-510 2651250 SL0539 28 Nov 2014 1213132
MPS-8033 80331685 SL0542 28 Nov 2014 1213129
IQS-1502 335896 SL0574 13 Nov 2014 284476-14
MPS-8033 80331733 SL0605 28 Nov 2014 1213130
FOS79800FSRV 79800K447 WP1350 27 Nov 2014 1204039
Certificate of Calibration
Certificate No.:
Issue date:
4 of 5 Page:
1342910
03 Sep 2014
Supplied value Lower limit Upper limit Measured value Unit Range
% of
Tol
TUR
<4
Absolute Power calibration, input
wavelength 850 nm
3 dBm -20.26 -19.76 -20.02 -20.011 dBm
wavelength 1310 nm
16 dBm -20.25 -19.75 -20.04 -19.995 dBm
wavelength 1550 nm
10 dBm -20.28 -19.78 -20.00 -20.027 dBm
Linearity calibration at 1310 nm
Local Linearity calibration
Reference set at -19.982 dBm
Measured at -14 dBm
6 dB -0.1000 0.1000 -0.0063 0.00000 dB
Measured at -17 dBm
9 dB -0.1000 0.1000 0.0092 0.00000 dB
Measured at -20 dBm
1 dB -0.1000 0.1000 -0.0008 0.00000 dB
Measured at -23 dBm
0 dB -0.1000 0.1000 0.0002 0.00000 dB
Measured at -26 dBm
2 dB -0.1000 0.1000 0.0022 0.00000 dB
Measured at -29 dBm
4 dB -0.1000 0.1000 -0.0043 0.00000 dB
Measured at -32 dBm
6 dB -0.1000 0.1000 0.0057 0.00000 dB
Measured at -35 dBm
4 dB -0.1000 0.1000 -0.0043 0.00000 dB
Measured at -38 dBm
6 dB -0.1000 0.1000 -0.0063 0.00000 dB
Measured at -41 dBm
1 dB -0.1000 0.1000 0.0007 0.00000 dB
Measured at -44 dBm
1 dB -0.1000 0.1000 -0.0013 0.00000 dB
Measured at -47 dBm
1 dB -0.1000 0.1000 -0.0013 0.00000 dB
Measured at -50 dBm
0 dB -0.1000 0.1000 0.0002 0.00000 dB
Total Linearity calibration
0.0095dB
Absolute Power calibration, output
wavelength 1310 nm
Pass -6.76 dBm > -7 dBm
wavelength 1550 nm
Pass -6.63 dBm > -7 dBm
Visual fault locator calibration
wavelength 650 nm
Pass Visual Fault Locator test
Certificate of Calibration
Certificate No.:
Issue date:
5 of 5 Page:
1342910
03 Sep 2014
Supplied value Lower limit Upper limit Measured value Unit Range
% of
Tol
TUR
<4
Remark:
1310/1550 output calibrations are measured into 9 um singlemode fiber
VFL output calibrations are measured into 9 um singlemode fiber