Farooq Hassan
Roll no: 2012-10-0101
LUMS School of Science and Engineering
Abstract
Michelson Interferometer was rst used in the classic experiment in 1887 to detect the presence of aether which was thought to permeate vacuum and act as the
medium in which all EM waves travel. The experiment hoped to detect aether
wind caused by motion of Earth through the medium. Light that traveled parallel to direction of aether would take longer to reach the screen than light moving
perpendicular to it and this dierence would be measurable by the resulting fringe
shift. The experiment did not detect any fringe shift which led to the abandoning
of concept of aether.
In this experiment, we will use Michelson Interferometer to calculate the wavelength of HeNe laser. We will also use Michelson Interferometer to measure the
refractive index of glass.
Introduction
Michelson Interferometer makes use of superposition phenomenon in waves to produce circular fringes. These fringes exit or enter the screen when the optical path
of one of the beams is increased or decreased. By counting the no. of fringes
that pass a reference point, distances of the order of nanometers can be measured
accurately.
Target Screen
Beam Splitter
He Ne Laser
M2
M1
Theoretical background
Light from a monochromatic laser is split into two halves using a beam splitter.
The two components are reected by mirrors M1 and M2 and then superimposed.
In this way, the beams that reach the detector are in phase(both undergo phase
dierence of pi upon reection from beam splitter) and coherent. As a result,
interference pattern is observed on the screen. To an observer on the screen the
set up would look like:
S1
2dco
'O
Detector
M1
S
M2
d
S2
Si
2d
S1
So
2dcos() = N
N = 0,1,2...
2
(1)
4.1
HeNe laser, Beam Splitter, post holders, steel posts, laser and lens mounts, kinematic mounts, convex lens, rotation platform and DC servo motor controller.
4.2
Procedure
The rst task in the experiment was optical alignment. The HeNe laser was
mounted using its clamp and M2 mirror was positioned in front of it such that
the laser beam fell in the centre of M2 and reected back on to the laser. Then M1
was placed at approx. same height as M2. Finally the beam splitter was inserted
and rotated such that the reected beam fell on M1. Once this was done, three
dots were visible on the beam splitter, one from each mirror and one from laser.
The three dots were aligned using the screws on the mirrors. This completed the
alignment.
In order to observe the fringes, we need to diverge the beam (enlargement) so two
plano-convex lenses were place, one right in front of laser and another before the
screen. This required realignment of the beams to keep the centre of fringes in focus. Once realignment was done, circular fringes were observed. These concentric
fringes are characteristic of the plano convex lens used.
Inorder to measure wavelength, the M2 mirror was moved using computer software. The settings were as follows: max vel = 0.003 mm/s, acc = 0.03 mm/s2 ,
step distance = 0.001 mm. The rotor was connected to M2 which was placed
between parallel aluminium sheets to ensure movement only takes place parallel
to the optical path of the beam reected by M2. The software was then used to
move the mirror a distance of 10m.
4.3
Observation
As the mirror M2 moved towards beam splitter, fringes were seen to disappear
from the centre. The no of fringes that disappeared were counted using the centre
as reference point (counting no of fringes was found to be least error prone at the
centre). This procedure was repeated 5 times. The results are given below.
Trial No.
1
2
3
4
5
No of fringes (N )
31
32
32
31
32
4.4
Analysis
Uncertainty
There was an uncertainty of 1 in N at start and end of each trial giving total
uncertainty of 2 in N . Uncertainty in d is same as minimum resolution of DC
motor (29 nm).
Uncertainity in :
2 = (2/N )2 u(d)2 + (2d/N 2 )2 u(N )2
= (2/31.6)2 (29 109 )2 + (2 105 /31.62 )2 (2)2
= 3.37 1018 + 1.605 1015
= 4.01 108
= (632 40) 109 m.
5
5.1
5.1.1
The optical path traveled in presence of glass sheet is nx where n is the refractive
index of glass and x is the length of path travelled inside glass.
From gure 2,
xcos(r) = t where r is the angle of refraction.
x = t/cos(r)
nx = nt/cos(r).
i
o-i
r
5.1.2
i
a
r
f
d b
e c
P
ac = t/ cos(i).
bc = ac t = t/ cos(i) t.
optical path = nt + t/ cos(i) t.
5.1.3
5.2
Procedure
First, we place the glass slide horizontally in front on M1. Then the rotation platform on which glass slide was mounted was rotated by hand. Care was taken to
ensure no weight was placed on the desk as it caused fringe pattern to be destroyed.
The zero degree mark in front of rotation platform was used as reference point. We
counted the no. of fringes that disappeared once again using centre as reference
point. After every 20 fringes that disappeared, the degree of rotation was noted
and the counting procedure continued upto 160 fringes. This experiment was repeated 5 times using the same mark on the dial as starting point. The results are
given below.
Next, the micrometer screw gauge was used to calculate the thickness of glass,
t, at the point where the laser beam passed the glass. The calculated value was
0.99m.
t = 1.15mm 0.16mm = 0.99mm.
0.16mm is the zero error.
N
020
040
060
080
100
120
140
160
rotated(o )
10
15
18
21
24
26
28
30
rotated(o )
11
16
19.5
22.5
25
27
29
31
rotated(o )
11
16
18.5
22
25
27.5
29.5
31
rotated(o )
11
15.5
19
22.5
24.5
26.5
28.5
30.5
rotated(o )
11
15.5
19
22
24.5
26.5
28.5
30.5
No of Fringes
250
data 1
quadratic
data 2
200
150
100
50
10
15
20
25
30
No of Degrees
35
5.3
Analysis
n (Refractive Index)
1.52
1.52
1.51
1.50
1.50
1.51
1.51
1.51
5.3.1
n2 =
)sin(2tN )(1cos)2tsin 2
()2 ( (2t(1cos)N )(2tN
) +
(2t(1cos)N )2
)(1cos)() 2
(N )2 ( (2t(1cos)N )(1cos)()(2tN
)
(2t(1cos)N )2
n2 =
)N )sin(1cos )2tsin)) 2
) +
(1)2 ( ((2tN )((2t(1cos
(2t(1cos )N )2
2t(1cos )+(2tN ))) 2
(2)2 ( ((1cos )(N
)
(2t(1cos )N )2
)(2tsinN )) 2
)(2t cos )) 2
n2 = ( ((2tN
) + 4( ((1cos
)
(2t(1cos )N )2
(2t(1cos )N )2
n(uncertainty in n)
0.08
0.07
0.03
0.08
0.03
0.02
0.01
0.01
Conclusion
Precautions:
The Michelson Interferometer is an extremely sensitive experiment. Several precautions were taken while performing the experiment:
1) While taking readings, no vibrations or noise in nearby surrounding had to be
ensured. The fan had to be switched o.
2) The fringes could appear or disappear as a result of movement of steel post
holding the glass as well. It had to be ensured that steel post was rmly clamped
to the rotation platform.
3) The starting point of all readings was take to be the same for each trial.
4) We had to make sure that all lens, glasses were clean.
5) Direct eye contact with the laser was avoided for personal safety. Care was taken
while performing optical alignment to ensure that ray did not deect in arbitrary
direction.
Suggested Improvements:
The experiment can be further improved by: 1) Use computer software to generate
rotation as well so as to minimize uncertainty in its value.
2) Increase optical path by causing beam to move multiple times between beam
splitter and mirror before falling on the screen.
3) Use compensation mirror. This would reduce the error that occurs because 1
beam passes beam splitter once while the other beam passes beam splitter thrice.
4) Conduct experiment in dark environment in mercury bath( as the original experiment in 1887).
References
http://physlab.lums.edu.pk/images/c/c5/Michelson nal1.pdf.
http://physlab.lums.edu.pk/images/1/1d/Errorsv3.pdf.
http://physlab.lums.edu.pk/images/5/50/Optics Hecht.pdf.
http://physlab.lums.edu.pk/images/e/ed/Refractive index reference Book.pdf.
http://physlab.lums.edu.pk/images/d/dc/Lab Michelson.pdf.
http://physlab.lums.edu.pk/images/c/c7/Michelsonpaper.pdf.
10