product catalogue

PSA-10 & PSA-20


POWER SUPPLIES

AUSTCO

Description
The PSA range of power supplies are 13.8V DC battery-backed power supplies
capable of 10 Amps (PSA-10) or 20 Amp (PSA-20) for use to power Austco
nurse/staff call systems.
The PSA-10 provides two, and the PSA-20 provides four, resettable fused
and individually switched 4 amp power outputs. The power output
switches are key switch protected.
Incorporates internal, or optional external, 12 volt battery back-up as well
as battery charging and monitoring circuits with LEDs and alarm outputs
to indicate battery, mains or output failures. Power for charging the
batteries are provided by the PSA power supplies.
The PSA-10 and PSA-20 are supplied in wall mountable or optional rack
mountable metal cabinets and have IEC950 electrical safety ratings. The
PSA-10 is also available with IEC601 medical rating.

Features

Benefits

Slim unobtrusive design and color.

Can be surface mounted in open areas without being obtrusive or


rack mounted in an electronics equipment cabinet.

Two or four individually fused and switched


power outputs.

Individual parts of the nurse/staff call system can be powered ON/OFF


and isolated separately for more reliable operation and testing.

Key switch protected controls.

Power ON/OFF switches are protected against inadvertent or malicious


operation.

Battery back-up.

Continues to provide power during mains failure.

External battery option.

Allows the addition of external batteries for increased battery capacity.

Continuous status monitoring with LED indicators and alarm outputs.

Clear indication of operational status and immediately alerts staff of any


battery, mains or output failure.

Single assembly mounting with external


connectors.

Installation and services times are minimized.

Ordering Information
Code

Description

PSA-10/240/950

Power Supply 10 Amp 240 V Input IEC950 compliant. (Excludes


battery)

PSA-10/110/950

Power Supply 10 Amp 110 V Input IEC950 compliant. (Excludes


battery)

PSA-10/601

Power Supply 10 Amp 110/240 V Input IEC601 compliant. (Excludes


battery)

PSA-20/240/950

Power Supply 20 Amp 240 V Input IEC950 compliant. (Excludes


battery)

PSA-20/110/950

Power Supply 20 Amp 110 V Input IEC950 compliant. (Excludes


battery)

PSA-BAT

Power Supply Battery --- 7 Amp hour

Order Options
Specify:
- Wall Mount
- Rack Mount

MC 12-1

MediCom

PSA-10 & PSA-20


POWER SUPPLIES

AUSTCO

product catalogue

Product Specifications
Wall mountable or optional rack mountable 13.8V AC/DC battery
backed power supply in cabinet manufactured from 1.2mm thick
beige powder coated steel.
Operates on 220-240 VAC or 110-120 VAC input via a fuse protected
mains input socket and cable.
Provides individually ON/OFF switched 13.8V DC power outputs each
capable of 3.5 Amps continuous power as follows:
PSA-10 - two outputs (total output of 7 Amps)
PSA-20 - four outputs (total output of 14 Amps)

Operational status is continuously monitored and indicated via


externally visible LEDs for;
Mains
Battery OK
Outputs OK
Output ON/OFF
External alarm outputs are provided for mains and/or DC output
failures.
Cabling is to convenient externally accessible plug-in screw
connectors.

Each output is protected against overload by a fuse that


automatically resets after the overload is removed.

Includes a safety earth output.

Operation of the DC power output ON/OFF switches is key-switch


protected to prevent inadvertent or malicious change in settings.

The PSA electronics are protected against power surge, near


lightning strike, cable shorts and power reversals.

Incorporates a 12V battery back-up and charging circuit with optional one (for 10 Amps version) or two (for 20 Amps version) internal
lead gel sealed batteries rated at 7 Amp hour each. The batteries
must be ordered (PSA-BAT) and shipped separately.

Designed to comply with:


AS 3811 Hard-wired patient alarm systems for hospitals.
HTM Bedhead services.
BS EN 60601.1, CSA 601-1-M90 &
UL 2601.1 Medical Electrical Equipment. General
Requirements for Safety.
UL 1069 Hospital Signaling and Nurse Call Equpiment.

Also provides a connector for connecting external batteries (max. 35


Amp hour rating).
Power for charging the back-up batteries are provided by the PSA
power supplies. (2 Amps by the PSA-10 and 4 Amps by the PSA-20)

Manufactured in general accordance with the requirements of


international quality assurance standard ISO9002.

Dimensions & Specifications


Height:

430 mm

Depth:

130 mm

Width:

200 mm

Weight:

2.6 Kg (excluding batteries)

Mounting
Method:

Wall mount or Rack mount

Plate:

N/A

Wall Box size:

N/A

Housing:

1.2mm thick beige powder coated steel


cabinet

Cable Sizes:

MC 12-2

2-core 2mm TPS (AWG 14) power cable


for each power feed

Connectors:

Plug-in screw terminals

Temperature:

-2035 C

Humidity
Range:

090%