Analytical Method
Validation
A. Es-haghi
Ph.D.
Dept. of Physico chemistry
Razi vaccine and serum research institute
a.eshaghi@rvsri.ir
http://www.rvsri.ir/
Introduction
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
What is Validation?
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Razi
Vaccine & Serum
Research Institute
SPECIFICITY
LINEARITY
RANGE
DETECTION LIMIT
QUANTITATION LIMIT
ACCURACY
PRECISION
RUGEDNESS
ROBUSTNESS
SYSYEM SUITABILITY
http://www.rvsri.ir/
Specificity / Selectivity
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Specificity
Razi
Vaccine & Serum
Research Institute
Identification
to ensure the identity of the analyte
Purity Test
accurate statement of the content of impurities of an analyte
(related substances, heavy metals, residual solvents, etc.)
Assay
an exact result which allows an accurate statement on the content
of potency of the analyte in a sample
http://www.rvsri.ir/
Specificity
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Specificity
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Linearity
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Linearity
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Linearity
Razi
Vaccine & Serum
Research Institute
USP <1225>
Evaluated across the RANGE
Visual examination of a plot of signals as a function of
concentration
Statistical evaluation of correlation coefficient, y-intercept, slope,
and residual sum of squares of regression line or after
mathematical transformation
ICH Guidance
Performed directly on the drug substance (dilution of standard
stock solution)
Separate weighing of synthetic mixtures of the drug components
Minimum of five concentration is recommended
http://www.rvsri.ir/
Linearity
Razi
Vaccine & Serum
Research Institute
Required for
Assays (Category I)
Impurity tests (Quantitative Category II)
May be for Performance Tests (Category III-Dissolution, etc.)
http://www.rvsri.ir/
Range
Razi
Vaccine & Serum
Research Institute
UPS <1225>
The interval between the upper and lower quantitation levels
of analyte (including these levels) demonstrated by suitable
precision, accuracy, and linearity
ICH Guidance
The following minimum specified ranges should be
considered
Assay of drug substance or finished drug product
from 80% to 120% of the test concentration
Content Uniformity
minimum of 70% to 130% of the test concentration,
unless otherwise justified
http://www.rvsri.ir/
Razi
Vaccine & Serum
Research Institute
QL (LOQ)
the lowest amount of analyte that can be determined with
acceptable precision and accuracy
http://www.rvsri.ir/
Razi
Vaccine & Serum
Research Institute
Non-instrumental Methods
analysis of samples with known concentrations of analyte to
establish the minimum concentration at which the analyte can be
reliably (the ability to reproducibly obtain a signal for the analyte)
detected
Examples
TLC, titrations, color compositions
Instrumental
Same as for non-instrumental methods, plus estimation of the DL
can be made at a value at which the signal to noise ratio (S/N) is
2:1 or 3:1
http://www.rvsri.ir/
Estimation of DL (ICH)
Razi
Vaccine & Serum
Research Institute
DL = 3.3 /S
Where: = Standard deviation (noise) of response
S = slope of the calibration curve for linearity
http://www.rvsri.ir/
Quantitation Limit
Definition
Razi
Vaccine & Serum
Research Institute
USP <1225>
the lowest amount of analyte in a sample that can be determined
with acceptable precision and accuracy under the stated
experimental conditions
http://www.rvsri.ir/
Determination of QL
Razi
Vaccine & Serum
Research Institute
Non-Instrumental
from analysis of samples with appropriate minimum level of
analyte quantified with acceptable accuracy and precision
Instrumental
same as non-instrumental, plus estimation of the QL can be made
at a value at which the signals to noise is 10:1
http://www.rvsri.ir/
Estimation of QL (ICH)
Razi
Vaccine & Serum
Research Institute
QL = 10 /S
where = standard deviation of response
S = slope of the calibration curve for linearity
can be based standard deviation of the blank or of the lowest
level linearity point
http://www.rvsri.ir/
Case Study
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Accuracy
Razi
Vaccine & Serum
Research Institute
Definition
USP
The closeness of the result obtained by the method to the true
value.
ICH
The closeness of the result obtained by the method to a value
that is accepted as conventionally true value or as a reference
value.
http://www.rvsri.ir/
Accuracy
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Accuracy
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Accuracy
assay of a drug substance
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Precision
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Precision
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Precision
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Razi
Vaccine & Serum
Research Institute
Repeatability
agreement within a short period or the same analyst and
instrumentation
Intermediate precision
agreement in results intra-laboratory but from different days,
analysts and equipment (as appropriate)
Reproducibility
agreement in results between laboratories (as in a collaboration
study)
http://www.rvsri.ir/
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Precision
choice of acceptance criteria
Razi
Vaccine & Serum
Research Institute
Practical Considerations
Ability to meet criteria
may be instrument dependent
Maximum acceptance criteria for precision should be linked to
acceptance criteria for claim or specification to be able to obtain a
reliable result (reduce Out-Of Specifications)
http://www.rvsri.ir/
Precision
Razi
Vaccine & Serum
Research Institute
Max, % RSD
0.37
0.55
0.73
1.9
3.9
5 to 25
http://www.rvsri.ir/
Precision
choice of acceptance criteria
Razi
Vaccine & Serum
Research Institute
Intermediate Precision
FDA recommends a minimum of 2 different days with different
analysts
Acceptance criteria
Perform F-test
results between labs should not be statistically different
Typically about 2x repeatability (2 x RSD)
However, a statistical difference may still not be
significant (practical consideration). Need to evaluate
suitability for methods intended use
http://www.rvsri.ir/
Reproducibility Precision
Razi
Vaccine & Serum
Research Institute
FDA
at least 2 laboratories within the company
FDA
for multi-laboratory follow ASTM 691
Standard practice for conducting an laboratory study to
determine the precision of a test method
AOAC Protocol
8 samples, 8 laboratories
http://www.rvsri.ir/
Ruggedness
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Ruggedness
Razi
Vaccine & Serum
Research Institute
Determination
Analysis of an homogeneous sample in different laboratories, by
different analysts, under prevalent environmental conditions using
the specified parameters
The degree of reproducibility of test results is determined as
function of the variables
http://www.rvsri.ir/
Ruggedness
Razi
Vaccine & Serum
Research Institute
Analyst
Day
Lab
Column
A, B, C
1, 2, 3
L, M N
X, Y, Z
http://www.rvsri.ir/
Ruggedness
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Robustness
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Robustness
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Robustness
Razi
Vaccine & Serum
Research Institute
Two approaches
Uni-variable
Systematically varying each parameter sequentially
Typically, each parameter may be varied by 5 10% above
and below the value set in the method
Multi-variable
Use of the statistical design
Central Composites
Fractional factorials,
Two-factor Plackett-Burman statistical approach
Hendrix
http://www.rvsri.ir/
System Suitability
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
System Suitability
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
System Suitability
Razi
Vaccine & Serum
Research Institute
Resolution specificity
Column efficiency specificity
Relative standard deviation precision
Tailing factor accuracy and precision
QL sensitivity
Capacity factor specificity
Reference Standard Check - analyte
http://www.rvsri.ir/
Razi
Vaccine & Serum
Research Institute
Category I
Quantitative test for major active ingredients (chromatographic
and titration)
Category II
Impurities or degradation compounds; quantitative or limit tests
Category III
performance tests (dissolution, drug release, tapped density,
particle size)
Category IV
Identification tests (TLC, IR, UV, Chemical ID <191>
http://www.rvsri.ir/
USP <1225>
Data Elements for Validation
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
ICH Q2B
Data Elements for Validation
Razi
Vaccine & Serum
Research Institute
May
http://www.rvsri.ir/
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Validation Protocols
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Pre-validation Assessment
Razi
Vaccine & Serum
Research Institute
Equipment
Suitable for expected accuracy?
Reference Materials
Suitability characterized?
Analytical Method
Is procedure finalized?
Validation Protocol
Management / QA approved?
http://www.rvsri.ir/
Pre-validation Assessment
Razi
Vaccine & Serum
Research Institute
Equipment
Demonstrate that equipment used in validation studies is suitable
for use and is comparable to equipment used for routine analysis
Calibrated (as applicable)
Qualification should have been performed
Installation Qualification
Operation Qualification
Performance Qualification
Routine maintenance performed- proper working order
http://www.rvsri.ir/
Pre-validation Assessment
Razi
Vaccine & Serum
Research Institute
Reference Materials
evaluate suitability of use
FDA requires no additional evaluation of USP Reference
Standards
Standardization of reagents vs. Primary standards when available
(i.e., use of reagent grade chemical need qualification)
Principal analyte (purity and homogeneity)
Impurity Standards
chemical purity is very critical if used to determine
relative response
Solution stability of standard preparations
http://www.rvsri.ir/
Pre-validation Assessment
Razi
Vaccine & Serum
Research Institute
Analytical Method
must be final before validation
Individual procedure steps may impact overall method accuracy
or precision
Solution preparations
concentration of analyte in solution for analysis (suitable
range)
System suitability
necessary parameters and acceptance criteria must be
specified
Changes may negatively impact validation studies
previously performed
Later changes in tailing factor or resolution criteria from
robustness evaluation may have a significant impact on
integration in accuracy evaluations
http://www.rvsri.ir/
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Revalidation
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Razi
Vaccine & Serum
Research Institute
Equipment
Personnel
Reference Materials
Analytical Methods
Validation Protocol
Change control
Weak links between different chain to final product be
completely assured
http://www.rvsri.ir/
Quick Quiz
Cumulative errors in serial dilution
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Razi
Vaccine & Serum
Research Institute
http://www.rvsri.ir/
Quick Quiz
Cumulative errors in serial dilution
Razi
Vaccine & Serum
Research Institute
A. 1.0/10; 1.0/10L
B. 5.0/50; 5.0/50
C. 10.0/100; 10.0/100
D. 1.0/100
E. 2.0/200
F. 5.0/500
http://www.rvsri.ir/