Measurement
journal homepage: www.elsevier.com/locate/measurement
Centro Nacional de Metrologa, km 4.5 carr. a Los Cus, El Marques Quertaro, Mexico
Instituto Nacional de Tecnologa Industrial, Av. General Paz y Albarellos, San Martn Pcia, de Buenos Aires, Argentina
a r t i c l e
i n f o
Article history:
Available online 11 August 2013
Keywords:
Differential pressure
Comparison
a b s t r a c t
A differential pressure comparison was performed between CENAM (Mexico) and INTI
(Argentina) by means of a liquid column manometer. The measuring range was 5
500 Pa. CENAM calibrated the transfer standard at the beginning and at the end of the
comparison. The transfer standard used was a Dwyer liquid column manometer model
Microtector with an accuracy class of 0.013% of the reading. The compared pressure points
were (5, 75, 125, 200, 250, 300, 350, 400, 450, 500) Pa. The uncertainty sources to be evaluated included at least the following: (a) Uncertainty due to the standard used by the laboratory; (b) Uncertainty due to repeatability; (c) Uncertainty due to resolution; (d)
Uncertainty due to hysteresis; (e) Uncertainty due to zero drift. The criteria used to compare the results obtained were the normalized error equation (En). The results obtained by
the laboratories were compatible according to the criteria |En| 6 1.
2013 Published by Elsevier Ltd.
1. Introduction
2. Comparison conditions
Corresponding author. Tel.: +52 442 2110572; fax: +52 442 2110578.
E-mail address: jorge.torres@cenam.mx (J. Torres-Guzman).
0263-2241/$ - see front matter 2013 Published by Elsevier Ltd.
http://dx.doi.org/10.1016/j.measurement.2013.06.023
2.2. Procedure
The procedure used for the comparison consisted of
the TS calibration by measuring 11 differential pressure
points distributed on the instrument range (5, 75, 125,
200, 250, 300, 350, 400, 450, 500) Pa. Six series of measurements were taken, both ascending and descending
order. The procedure was the SIM technical protocol,
supplementary comparison SIM.M.P-S6 in low pressure
50500 Pa send for registration and approval to the
SIM MWG7 (SIM working group 7, mass and related
quantities).
4329
Table 3
INTIs reference standard.
Manufacturer
Junkalor, Dessau, D
Type
Diving Bell
Table 1
Transfer standard general specications.
Instrument
Make
Model
Serial number
Range
Accuracy class
Dwyer
M1430
S48H
150 mm H2O
0.013% of the reading
Table 2
CENAMs reference differential pressure measurement system.
Manufacturer
Model
Serial number
Piston cylinder
(differential)
Piston cylinder
(reference)
DH Instruments
PC-7607-5
231
DH Instruments
PC-7607-5
216 A
3. TS characterization
s calibrations. Results for
Table 4 shows the two CENAM
each measured pressure included error and uncertainty.
Fig. 4 shows a graph of the measurements presented in
Table 4.
Table 4
Transfer standard uncertainty and error of the 2 calibrations made by
CENAM.
Differential
pressure
(Pa)
TS error
1st cal
(Pa)
TS error
2nd cal
(Pa)
TS U (k = 2)
1st cal (Pa)
TS U (k = 2)
2nd cal (Pa)
50
100
125
150
200
250
300
350
400
450
500
0.04
0.05
0.04
0.02
0.18
0.17
0.16
0.09
0.04
0.08
0.03
0.05
0.03
0.02
0.01
0.14
0.11
0.13
0.07
0.04
0.06
0.02
0.19
0.13
0.13
0.13
0.16
0.16
0.14
0.18
0.13
0.12
0.13
0.16
0.13
0.13
0.13
0.14
0.13
0.14
0.18
0.12
0.14
0.16
4330
Table 6
INTIs transfer standard error and uncertainty.
TS pressure
TS error
U (k = 2)
50
100
125
150
200
250
300
350
400
450
500
0.00
0.03
0.06
0.10
0.13
0.09
0.13
0.02
0.02
0.07
0.00
0.30
0.30
0.30
0.30
0.30
0.30
0.30
0.30
0.30
0.30
0.30
5. Comparison results
The calibration results from CENAM and INTI are shown
in Fig. 7. The graph shows the results of both laboratories
for each pressure point and includes the associated uncertainty in the range, 50500 Pa.
5.1. Uncertainty and comparison criteria
The uncertainty sources to be evaluated included at
least the following: (a) Uncertainty due to the standard
used by the laboratory; (b) Uncertainty due to repeatabil-
Table 5
Transfer standard error and U used by CENAM.
TS pressure
TS error
U (k = 2)
50
100
125
150
200
250
300
350
400
450
500
0.04
0.04
0.03
0.01
0.16
0.14
0.14
0.08
0.04
0.07
0.02
0.17
0.13
0.13
0.13
0.15
0.14
0.14
0.18
0.12
0.13
0.14
Fig. 7. Transfer standard error and uncertainty (k = 2) from the calibrations made by CENAM and INTI.
Normalized
error
50
100
125
150
200
250
300
350
400
450
500
0.14
0.03
0.06
0.26
0.10
0.69
0.04
0.17
0.18
0.43
0.07
6. Conclusions
The calibration of the differential pressure transfer
standard obtained by CENAM (Mexico) and INTI (Argentina) in the range from 50 Pa to 500 Pa clearly shows that
there is compatibility between the two laboratories. The
results of the comparison were |En| < 1 in all pressure
points compared, which meets the criteria of compatibility.
Acknowledgements
CENAM and INTI Fsica y Metrologa want to thank
their pressure laboratories people for their involvement
and collaboration in this comparison.
xINTI xCENAM
En q
U 2INTI U 2CENAM
4331
References
[1] A. Calcatelli, F. Arrhen, M. Bergoglio, J. Greenwood, R. Kangi, K.
Jousten, J.-C. Legras, M. Rantanen, J. Verbeek, C. Matilla, D. Szaulich,
Results of the regional key comparison EUROMET.M.P-K1.a in the
pressure range from 0.1 Pa to 1000 Pa, Metrologia 42 (Tech. Suppl.)
(2005) 07004.
[2] J.C. Torres-Guzman, S. Ruiz, P. Olvera, Ma. Nieves Medina, Differential
pressure comparison from 20 Pa to 3 500 Pa between CEM-Spain and
CENAM-Mexico, Memories of XIX IMEKO World Congress
Fundamental an Applied Metrology, Lisbon, Portugal.
[3] J.C. Torres-Guzman, S. Ziga, P. Olvera, Caracterizacin de un
manmetro diferencial digital usado como patrn de transferencia a
3.5 kPa [Characterization of a digital differential manometer used as
transfer standard at 3.5 kPa], in: Simposio de Metrologia 2008,
Queretaro, Mexico, October 2008.