RadiationProtectionDosimetry(year),Vol.0,No.0,pp.00
DOI:10.1093/rpd/nc0000
DetailedinhouseQCprotocolshavebeendeveloped
for the acceptance and routine testing of general
radiographicXraysystems.Theseprotocolsarebased
on recommendations and guidelines published by
International[3]andEuropeanbodies[47].
Thefollowingparameterscommontobothmobile
andfixedsystemswereassessed:
INTRODUCTION
Optimisation of patient dose andimagequality is of
primaryconcerninthefieldofdiagnosticimaging.Itis
recognised that comprehensive Quality Assurance
(QA) programs are a vital component of the
optimisationprocess.
AcceptancetestingandroutineQualityControl(QC)
testing of diagnostic imaging equipment is a
requirementofEuropean[1]andIrish[2]legislation.
Althoughfilm/screentechnologyiscurrentlybeing
replacedbycomputedanddigitalradiography,systems
utilising film/screen combinations still represent a
significant portion of general radiographic Xray
systemsinuseinIreland.
The Medical Physics and Bioengineering (MPBE)
Department, St. Jamess Hospital, Ireland, is
responsiblefortheannualtestingofapproximately100
general radiographic Xray systems in thirtyfour
hospitals within Ireland. This includes thirtynine
film/screen fixed general radiographic Xray systems
andsixtyonemobilegeneralradiographicXrayunits,
suppliedbytenmanufacturers.Theaverageageofthis
equipmentisapproximatelyeightyears.
This work presents the findings from acceptance
testing and routine QC performed during 2006 and
early2007.
AcceptanceandRoutineQATesting
TubeandGeneratorPerformance
RadiationOutput
HalfValueLayer
BeamAlignment
FocalSpotSize,
DoseAreaProduct(DAP)MeterAccuracy
MechanicalSafety
EquipmentCondition
AcceptanceTestingOnly
ElectricalSafety
TubeLeakageRadiation
MATERIALANDMETHODS
*Correspondingauthor:lgray@stjames.ie
1
Theauthor(year).PublishedbyOxfordUniversityPress;allrightsreserved
SAMPLEETAL
Figure2.IssuesidentifiedwithDAPMeters
No. of Syste ms
5
0
Automatic
Beam
Exposure Alignment
Control
DoseAre
Product
Meter
Layer
Tube
Light Beam
Voltage
Diaphragm
Te st Faile d
MeanOpticalDensity
ConsistencywithChambers
ConsistencywithTubeVoltage
ConsistencywithWaterPhantomThickness
Xrayfieldtolightbeamalignmentfailuresarea
particular problem for mobile Xray equipment, as
constantmotionofunitsmaydislodgetheXraytube
orcollimatorhousing.
Tube output varied significantly, typically in the
rangeof28.467.4Gy/mAs.InthecaseofmobileX
raysystems, thishasimplicationsforpatientdoseand
image quality, particularly within hospitals that have
manymobileXraysystemsattheirdisposal.Agiven
mAssettingmayproduceanimageofoptimalquality
on one unit, while the same setting will produce an
imageofinferiorqualityonanother.NotallmobileX
rayunitsincorporateanAECsystemandforthosethat
do,itmaynotalwaysbeused.Inthisstudy,noneofthe
mobilestestedincorporatedandAECsystem.Thereis
apotentialtounderoroverexposethepatient.This
mayleadtorepeatexaminations.
Thetubeoutputforaparticularsystemincreasedby
24%from2005to2006,highlightingtheimportanceof
datatrendingfromyeartoyear.
45%ofsystemsincorporated aDAPmeter forthe
purpose of monitoring patient dose. The majority of
DAP meters were found to be working within
tolerance, howeversome typicalissues identifiedare
depictedinFigure2.
CONCLUSION
73%ofthe systemstestedfailedtomeettherequired
standards,wherebyoneormorefaultswereidentified.
The majority of these failures were minor ones,
requiringattentionbythesuppliersatthenextroutine
service.Significantfaultsthatcouldpotentiallyimpact
uponpatient doseand safetywere onlyidentified in
seven systems. The equipment suppliers were
requested to investigate these issues as soon as
possible,andtakethenecessarycorrectiveaction.
Althoughgeneralmobileunitsandfilm/screenfixed
systemsarenotconsideredhighdosemodalities,useof
these systems comprises a significant portion of a
diagnostic imaging workload. A review of the QC
resultshighlightstheneedtoperformcomprehensive
acceptanceandroutinetestingofthesesystems.
SHORTTITLE
ACKNOWLEDGEMENT
Thisworkwasconductedpartiallywithintheframeof
theEuropeanCommission6thFrameworkProgramme
SENTINELContractNoFP6012909.
REFERENCES
1.
2.
3.
4.
5.
6.
7.
TheCounciloftheEuropeanUnionMedicalExposure
DirectiveonHealthProtectionofIndividualsagainstthe
Dangers of Ionising Radiation in Relation to Medical
Exposure,97/43/EURATOM.
S.I. No. 478/2002: European Communities (Medical
IonisingRadiationProtection)Regulations2002.
International Standard IEC60580, Medical Electrical
Equipment Dose Area Product Meters, (IEC60580:
2000(E)).
European Commission (EC), Radiation Protection 91,
Criteria for Acceptability of Radiological (including
radiotherapy) and Nuclear Medicine Installations
(Luxembourg: Office for Official Publications of the
EuropeanCommunities)(1997).
Institute of Physics and Engineering in Medicine
(IPEM),MeasurementofthePerformanceCharacteristics
of Diagnostic Xray Systems used in Medicine, Report
No.32,Second
Edition. Part 1: Xray Tubes and
Generators. (1996) and Part IV: (Xray Intensifying
Screens, Films, Processors and Automatic Exposure
ControlSystems.(1996).
British Institute of Radiology (BIR), Assurance of
QualityintheDiagnosticImagingDepartment(2 ndEdition)
(2001).
IPEM, Report 91, Recommended Standards for the
RoutinePerformanceTestingofDiagnosticXrayImaging
Systems,2005.