Setup and adjustment of weighing

Metrological setup
All options in this section require INSTALLER level password entry.

Class
  /  / 
Possible values: II, III and IIII.

Capacity
  /  /  /  

  

Legal metrology (trade use):


This option is mandatory if the instrument use corresponds to one of trade uses:
Commercial transaction, medical weighing, etc.
The impacts are many as for metrological adjustments and zero devices.

 

Direct sale to the public:


Type of trade use which imposes certain constraints, particularly with regard to
peripherals.

 

Non legal metrology (non trade use).

  / /


  / 
Unit weight of the indicator system. All weight values displayed or entered are expressed in that unit.
Possible values: Kg, g, t, lb.
  / /
$$
  /$
Possible values: 1 to 3
  /  /  /  "... & / + 
  /  /  /  "... & /  
  /  /  /  "... & /   
  / /
  / 
The operating mode can be specified for an instrument with 2 or 3 ranges:
  / /
  /  
 $

Multi-division:
One area divided into 2 or 3 parts, each with a different division.

$

Multi-range:
2 or 3 ranges, from 0 to its maximum capacity, with different steps.

Taking earth's gravity into account

Record the parameters instead of metrological calibration:


  / /
  /  
  / /
  /  

04-55-00-1 MI

39

i 40 Indicator

Metrological calibration with standard mass


All options in this section require INSTALLER level password entry.

Adjustment with depositing a standard weight is the preferred method of adjusting a weighing instrument. This
is the method that leads to the most accurate results.

Zero offset

Put any standard mass placing accessories (for example, the mass suspension device for a hopper) and start
the following command.
  /  ,  /  

 



 

Zero measurement



To confirm the zero offset acquisition and exit the


function, press key.

As before, the measurement must be made with any standard mass placing accessories, but without any other
load.
  /  ,  / - / -




 



To confirm the zero offset acquisition and exit the


function, press key.

-
Declaring the number of segments
  /  ,  / - / $$ 
Most often, it is not necessary to adjust a weighing instrument over several segments. Leave the default
value: NB SEG=1.
If the load receiver provides a not so linear signal and requires a calibration over several segments, enter
the number of segments, for example: NB SEG=3.
Maximum number of segments: 6.

40

04-55-00-1 MI

Linearisation on a segment
  /  ,  / - /  "












Confirm by pressing key .

 





Input this value and validate by pressing key .

"$./

 



The weight measurement is displayed.




"0



$$ 

Place the standard mass and then validate by pressing


key .

Remove the standard mass.

At this stage, it is advisable to check the proper


consideration of the adjustment by running a verification.
See Verification, page 44.

Linearisation of several segments

Follow the procedure above for each segment, until the maximum number of segments defined in the order of
increasing mass (for example, 3):
  /  ,  / - /  %
  /  ,  / - /  &

04-55-00-1 MI

41

i 40 Indicator

Metrological adjustment without standard weight


All options in this section require INSTALLER level password entry.

The adjustment without standard weight can be used alternatively, in various adverse conditions. This is the
case of a weighing instrument that it is not possible to discharge to perform the adjustment (e.g. a silo loaded with
several tonnes of product to be preserved). This method leads to a precision of the order of 1 to 2%.

"Zero point and slope" method


  /  ,  /   /
   - 
Enter the following data:

  

Sensitivity of sensors used (mV/V).

$

Sensors' nominal load (in system weight unit).



Load supported by sensors when the instrument is unloaded (in system weight unit).

"Downweighing" method

Assessment of the slope by measuring the difference in weight


  /  ,  /    /   /  

 




 

 




42

Record the first measuring point.


Confirm by pressing key

.

Confirm the measurement or not.



04-55-00-1 MI

  /  ,  /   /


     /  

 




 

 



.

Confirm by pressing key

.

Enter the(i) weight difference value between the first and


the second measuring point.

%1$%!

 




Confirm by pressing key







Second measuring point (second screen).

Confirm the entry or not.



(i) Value of the mass loaded or unloaded measured using a reference device.

04-55-00-1 MI

43

i 40 Indicator

Entering the value of zero at a point


This operation is an additional to the "Downweighing" method, but it can be used in addition to using the "zero
slope" method when the tare load is unknown.
  /  ,  /   /
   -"

 

Value of the current load on the weighing instrument


(estimated or known by other processes).




2%$'!

 




Confirm by pressing key

.

Confirm the entry or not.



Verification
This option requires INSTALLER level password entry.
  /  !
Without exiting the configuration mode, this option allows you to immediately assess the weighing
function, in particular the quality of metrological adjustment.




 



To exit the function, press key .

Compensation for earth gravity variations


All options in this section require INSTALLER level password entry.

When the weighing instrument is adjusted to a different place from the point of use, it is necessary to
record the latitude and altitude values of each site. Thus, the system can calculate the compensation of the gravity
difference between the two locations.
So that the compensation for the variation of gravity is effective, the parameters of the instrument's usage location
must be recorded after adjustment.
Record the place of use parameters:
  / /
  /  
  / /
  /  

44

04-55-00-1 MI

Zero Filtering Stability (Installation)

Initial zero
This reset to zero can be made when the indicator is powered on. To do this, the user starts the indicator
by pressing and holding the  and
keys at the same time.
This zero is the reference to other zero devices.
  /   / - 
- 
Limit value below the real zero. For example, to enter a limit at -10%, enter 10.
Possible values: From 0 to -100% of the maximum capacity.
Trade use: From 0 to -20% of the maximum capacity (default value: +/- 5 %)
- 
Limit value below the real zero. For example, to enter a limit at -10%, enter 10.
Possible values: From 0 to -100% of the maximum capacity.
Trade use: From 0 to -20% of the maximum capacity (default value: 15 %)

Automatic zero
If enabled, this reset to zero occurs automatically when the gross weight remains stable:
- in a range from 0 to -2 of the maximum capacity from the initial zero;
- during a period greater than or equal to the  time.
  /   / $-$
 



Function enabled after the expiration of  delay time described below.



Function not in service.

Default value: Yes.


 
Possible values: From 0 to 65 s.
Legal Metrology 5 s.
Default value: 5000ms.

User zero function action limit


This zero occurs if the user presses the
key depending on the ! parameter configuration.
See Keyboard, page 33.
  /   / - 
Possible values: 0 to 99% of the maximum capacity.
Legal Metrology < 2% of the maximum capacity from the initial zero.
(Default value: 2 %

Zero tracking
This device allows maintaining the weight at zero when the scale is unloaded.
The action of this reset to zero is limited at the range 0 2% of the maximum capacity from the initial
zero.
  /   / - 

%'

Device not in service.

 

Active device capable of gross weight only.

 

Active device in all cases.

Default value: ACTIVE B/N.

04-55-00-1 MI

45

i 40 Indicator

Flow correction
This function corrects a slow drift of the indication supplied by the loaded weighing cell.
  /   / 
Possible values: ACTIVE OR INACTIVE.
Default value: INACTIVE.

Filtering
This function allows reducing the signal disturbance provided by the weighing sensor. A strong filtering
results in longer response times after the removal or addition of a load.
  /   / 
  
Possible values: 0 to 4.
Predefined filtering code details.
Filter 1

Filter 2

Filter 3

Code

Frequency
(meas. / s)

Type

Window

Type

Window

Type

Window

50

AVERAGE

10 meas.

AVERAGE

10 meas.

AVERAGE

10 meas.

50

AVERAGE

10 meas.

AVERAGE

10 meas.

SORTING

9 meas.

50

AVERAGE

7 meas.

AVERAGE

6 meas.

AVERAGE

5 meas.

100

AVERAGE

10 meas.

AVERAGE

10 meas.

AVERAGE

8 meas.

If CODE = 0, the previously loaded parameters (code 1 to 4) can be modified.


 
Possible values: From 1 to 100 measurements/s.
-  " to  &

Possible values: SORT, INACTIVE or AVERAGE.
 
Possible values: From 1 to 100 measurements (from weighing signal).

Stability
This function allows adjusting the measurement instability range in which it is declared unstable (symbol
 present). This only has consequence on the auxiliary devices that are only triggered when stability is
reached (recording in data storage memory, for example).
  /   / 
  
Possible values: 0 to 9.
Details of predefined stability codes.
Code

Window

2 meas.

3 meas.

4 meas.

4 meas.

8 meas.

12 meas.

8 meas.

16 meas.

20 meas.

Limit

10 pts

10 pts

10 pts

5 pts

5 pts

7 pts

6 pts

5 pts

5 pts

If CODE = 0, the previously loaded parameters (code 1 to 9) can be modified.


  
Possible values: from 1 to 99 measurements.
 
Possible values: from 1 to 9,999,999 measurement points (internal).

46

04-55-00-1 MI

Default parameter restoration

In case of need, the default parameters can be re-established, according to three independent categories:
Adjustment, setup and installation.
  / $
The first two options require INSTALLER level password entry.

04-55-00-1 MI

 , 

Restoration of factory default metrology adjustment.



Restoration of factory default metrology settings.

 

Restoring the default zero, filtering and stability parameters.

47

i 40 Indicator

48

04-55-00-1 MI