Walking is the most easy and convenient way to travel short distances. The free joint
movement and suitable muscle force will results in efficient walk. As the body moves
forward, one limb typically provides support while the other limb is advanced in preparation
for its role as the support limb [3]. The gait cycle is comprised of 2 phases, stance phase and
swing phase. However, the stance phase is consisting of 5 parts which are initial contact,
loading response, midstance, terminal stance and preswing. Heel strike will represent the
initial contact of heel to the ground. During the heel strike, the knee is extended, hip will
flexed and ankle is neutral or slightly plantar flexed. While, during loading the knee flexes 15
degrees, hip will reduced in flexed and ankle is plantar flexed in about 15 degrees which is the
energy-conserving mechanism [3]. At this stage, the body is slowly transferred to the heel and
will result in the peak of the vertical force (Ground Reaction Force, GRF) [2]. Furthermore,
during the first phase of stance, hamstring and ankle dorsiflexors remain active, while the
quadriceps and gluteal muscles act during loading and throughout early midstance to maintain
hip and knee stability [3]. During midstance, the knee is extended and ankle is in neutral
position again. In this stage, triceps surae acts to control tibial advancement in order to
prevent the tendency for the ankle to dorsiflex due to body weight and inertia. Double support
is occurring at this stage where both feet are in contact with the ground. After that, terminal
stance is taking over and starts the preswing where the knee is flexed and ankle plantar flexed.
In these last phases of stance, the toes in which have been neutral, dorsiflex at the
metatarsophalaneal joint. All the vertical force that acted during the stance phase can be seen
in graph below.
After the 60% of gait cycle complete which is the stance phase, the swing phase is
occur for other 40% of gait cycle. During the leg is swing, the ankle dorsiflexes by the
concentric contraction of anterior tibialis muscle and all other muscles are silent. The
momentum is generated by the gastrocsoleus and hip flexors at terminal stance move the leg
forward. During the initial swing phase, knee is passively flexes and then passively extends at
the end of the swing phase. The sub-talar joint is nearly in neutral position and toes are
dorsiflexed slightly as foot prepares for the next of gait cycle (stance phase) once the toe off
of the ground.
Ideally, each gait cycle will generate identical ground reaction force (GRF) when
walking on the flat surface [2]. However the GRF can be altered by the deviated walking gait.
There are many reasons lead to deviation of walking gait. As in our project, the walking gait is
altered by the transfemoral prosthesis of transfemoral amputee who amputated due to the
osteosarcoma. In other way, ground reaction force can be used to assess prosthesis. A good
prosthesis is the one that can imitate function and movement of normal limb as close as
possible that will enable its user to walk with least energy expenditure and look natural when
walking.
A prosthesis is a device that designed to replace a missing part of the body. It is
prescribed to people that experienced limb loss or amputated. Amputation is inevitable when
it is the only solution in cases like osteosarcoma. Osteosarcoma is a type of bone cancer
where the malignant tissue starts in bone. Most osteosarcoma arises from random and
unpredictable errors in the DNA of growing bone cells during times of intense bone growth.
Chemotherapy will be done to reduce the size of tumor followed by surgery to remove the
tumor. A radical surgery such as amputation will be needed for more extreme cases. Once
amputated patient will need a prosthesis to perform the activities that they able to do prior of
amputation, such as walking.
The subject who participated in this study is a cancer patient, affected with
osteosarcoma almost one year ago. She had undergone the amputation to remove the tumor
that affected her femur. Shes now walking again by wearing prosthesis.
OBJECTIVE
The study is conducted to see the differences caused by wearing prosthesis to the
sound leg. We compare the ground reaction force produced by our subject that is wearing
prosthesis to the other leg that is not wearing prosthesis to see whether the prosthesis is
causing the normal limb to exert more or less force at different stage of stance phase. Other
than that ground reaction force can be used to find out how the prosthesis is imitating normal
sound limb in terms of force exerted to the floor when walking.
Other than that, we would like to find out how wearing prosthesis would affect the
sound leg in terms of knee and hip flexion, extension angles. We compare the knee and hip
flexion and extension angle of the leg that is wearing prosthesis to the other leg that is not
wearing prosthesis. We also compared the results to a published result to see whether the gait
pattern of a prosthesis user deviates from a normal, able bodied person.
METHOD
A cancer patient who had been wearing prosthesis for few months acted as the subject
for this study. She is wearing transfemoral prosthesis with mechanical knee joint. During the
experiment, she walked using sport shoes. 16 markers were placed over her lower limb. The
markers for absent body parts such as right knee, right leg and right ankle were placed
approximately using her sound leg as guide. The markers placement and detailed procedure is
included at Appendix A.
For the experiment, she required to walk over the force plate for 7 times to produce 7
trials. Breaks were given on interval of each trial. She experienced tiredness and shortness of
breath after 2 trials. The results from all trials were then analyzed. The graphs from joint
angles are plotted against the whole gait cycle while graph for GRF was plotted against
percentage of stance phase.
SUBJECT
DETAILS:
Gender
: Female
Age
: 27 years old
Mass
: 49 kg
Prosthesis weight: 5 kg
Height
Leg length
Knee width
Ankle width
: 153.5 cm
Right leg (cm) (amputate side)
73
9.6
7.1
Push Of
Ground Reaction Force (GRF)
Loading
600
500
400
Force (N)
300
PROSTHETIC LEG
SOUND LEG
BW
Mid
200
100
Toe
0
0
10
20
30
40
50
60
70
80
90
Heel
FIGURE 1 PROSTHETIC LEG AND SOUND LEG GROUND REACTION FORCE FROM
EXPERIMENT
As we can see in figure 1, at the vertical force during loading response of prosthetic leg is
delayed approximately 7% of gait cycle than sound leg. This is because during double support
phase which is starting with heel strike, the subject slowly doing loading response by
transferring weight to heel. It takes longer time from heel strike to loading response than
sound leg.
100
FIGURE 2
At mid stance phase, the vertical force should be below the body weight as shown in figure 1
during the midstance phase of prosthetic leg. Figure 2 show the mid stance of sound leg has
three curve of vertical force. At curve A the vertical force act like normal gait in which below
the body weight. However, a transient force exerted higher than body weight as shown at
curve B and the vertical force reduced below the line of body weigh at curved C. The sudden
transient force created is due to the sudden force produced by hip to swing the prosthetic leg
and lock it in extension. As we can see in figure 1, the vertical force will peaked again during
push off of early terminal stance. The force is exerted due to the persons weight is entirely on
one foot [2]. After that, the vertical force decreased to 0N back as the toe off the ground (force
plate). However, once again we can see the vertical force of prosthetic leg reduces slowly to
0N. This show that the slow movement of the leg to toe off which indicate that the slow
progression of terminal stance.
Based on the Ground Reaction Force (GRF) graph in figure 1 mid stance phase of the
prosthetic leg is shorter than the sound limb. This is because of the subject feel less stable to
stand on the prosthetic leg. As we know, the patient is not regular walking by wearing the
transfemoral prosthesis and still not comfortable with the prosthesis.
DELAYED
SWING
PHASE OF
SOUND
30
20
10
0
0
-10
10
20
30
40
50
60
70
80
90
100
SOUND LEG
MAXIMUM
FLEXION
stability. During 57% of gait cycle, the knee of prosthetic leg starting to flex in order to do
push off and toe off to start swing the ipsilateral leg in swing phase. The knee flexes to 37 at
70% of gait cycle and reduces flexion in order to start heel strike for next gait cycle. The knee
starts extend at 81% of gait cycle. If we compare to the normal gait graph in figure 4, the
prosthetic foot flex at maximum degree and starts extension during swing phase is faster than
the normal gait. This is due to the fast movement of the auto lock system. The prosthesis is
imitating the normal knee joint by locking the knee joint in extension during stance phase and
only allows flexion for swing phase.
For sound leg, knee joint does not present any extension for the whole gait cycle (stance phase
and swing phase). This is because of the subject wants to gain stability counter the leg length
discrepancy (prosthetic leg shorter than sound leg) in which acquired her to flex the knee of
the sound leg. This is also due to the locking of the knee joint during stance phase of the
prosthetic leg. Generally, if we lock one of our leg in knee joint, the other leg will
automatically walk with knee flexion position to gain the stability during stance and swing
phase. As we compare to the normal gait graph, there is no present of knee flexion in stance
phase except at the end of the terminal stance which knee will slowly flex to do a toe off and
starts swing phase of ipsilateral leg.
Besides that, compare the swing phase of sound leg happen more late or delayed when we
with the normal gait in figure 4 and also swing phase of prosthetic leg in figure 3. This is due
to the slow progression in the loading response and terminal stance of the prosthetic leg which
is occurring synchronically with the swing phase of the sound limb.
KNEE
FLEXION
/EXTENS
ION
40
30
20
Angle ()
10
0
0
-10
PROSTHETIC LEG
SOUND LEG
B
10
20
30
40
50
60
70
80
90
100
-20
Percentage of Gait Cycle (%)
GREEN - RIGHT
Minimu
m hip
flexion
(push
starts to flex again as when the ipsilateral foot is starts to enter the swing phase. The
ipsilateral hip is continuously increase in flexion as it undergo swing phase and prepare to do
a next heel strike for next gait cycle.
However, figure 5 shows the hip of the prosthetic leg is flexing approximately 1 during heel
strike and reduced flexing as it starts loading. The hip is starting to extend at approximately
10% of gait cycle and increase in extension for the rest of the stance phase. Even during the
toe off of the prosthetic leg, the hip is still in the extension which is act as the minimum
flexion in the normal person gait in which it is the maximum degree of hip extension. As
swing phase of the prosthetic leg stars, the hip extension is reduced and starts to present hip
flexion as it swing and starts to form heel strike position of next gait cycle.
Besides that we can see that the maximum hip extension (refer point of minimum hip flexion
of normal person gait) is late occur if compare to the normal person gait in figure 6. This is
shows that the stance phase of the prosthetic leg is longer compare to the stance phase of
normal person. Same things happen to the sound legs swing phase. It starts late is due to the
longer of stance phase of the prosthetic leg. This is due to the slow progression of loading and
terminal stance of prosthetic leg which has been explained in figure 1 (GRF graph).
For the sound limb, the hip is flexion as the heel strike happen and reduced in flexion as the
ipsilateral is loading but there is slightly increase in flexion (circle A in figure 5). The sudden
increase in the circle A is coinciding happen in circle B. This is due to the synchronizing of
stance phase of sound leg and swing phase of prosthetic leg. At this time, the sound leg is
undergoes a progressing of loading and the prosthetic leg is undergoes auto lock of the knee
joint which produce extra force (inertia) due to the sudden pulling of spring lock to extend the
knee joint. When the knee joint is pulled to extend with the extra force producing by the
spring the hip will automatically pulled to flex due to the extra force. While the hip of the
sound limb which in loading response also flexed in order to counter the sudden pulling force
for the stability and prevention from falling. The flexed of the knee at that moment results the
flexing of the hip. This can be seeing in figure 7.
DELAYED
SWING
PHASE
OF
SOUND
LIMB AND
HAPPEN
Other than that the GRFR also shows that the midstance of prosthetic limb is happening in
short time than the sound limb. This is proves by the flexion of hip and knee flexion is
increased and reduced in short of time.
CONCLUSION
We can conclude that, the prosthesis does affect the ground reaction force (GRF) of
walking gait in an amputee. This have been proves by the stance phase of prosthetic leg is
more longer compare to the normal persons GRF graph. Also, the swing phase of the sound
leg is shorter in time happen than the normal persons GRF graph. Other than that, the
prosthesis affected the knee and hip of the sound limb in term of flexion and extension. We
can see from the graph, the knee flexion of sound limb during stance phase is more than
normal person and the hip extension during push off and preswing of sound limb is happen.
This effect is to gain the stability during walking and prevent from falling.
REFERENCES
1. Pitkin, R Mark. (2006). Biomechanics of Lower Limb Prosthetics. USA:
Springer.
2. Jenkins, J. & Ellis, C. (nd) Using ground reaction forces from gait
analysis: Body mass as a weak biometric, 5-7. Retrieved 19
December 2011 from http://www.jamjenkins.org/
3. Bogey, R. (2009). Gait Analysis. Retrieved 19 December 2011from
http://emedicine.medscape.com/article/320160-overview
APPENDIX A
Skeletal view of markers placement for Vicon Nexus.
To study the variation of parameters involved in gait analysis for different lower limb
prosthetics and orthotics users.
PERSONNEL
Certified Prosthetist and Orthotist (CPO) or Medical Officer (MO) to ensure the
patient safety and give advice on the proper way to handle the patient.
EQUIPMENTS
Force plate
Reflective markers
T-frame
5. Under Resource Pane, select the System tab and expand the MX camera node.
Select all MX cameras.
6. Change the view pane mode from 3D perspective to Camera.
7. Remove any unwanted reflection registered in any of the camera view or mask them
by clicking Start the Stop under the Create MX camera Masks.
8. Then, get someone to wave the 5 Marker Dynamic Calibration frame (T-frame)
throughout the capture volume as soon as you hit start under Calibrate MX
Cameras. (Note: swing the T-frame from head-to-toe and right-to-left alternately, and
walk around the space within the camera view range)
9. Wait until the automatic stop feature enabled and the image error value displayed.
(Note: ensure that the error value for all the cameras is at most 0.2%. If exceed, the
please repeat step 8)
10. After that, place the calibration wand at the capture volume origin. (Note: place the
wand on the force plate)
11. Ensure that no one is standing in the capture volume, and then click on Start under
Set Volume Origin.
12. Reverting back to the 3D Perspective View and zooming out, so that the cameras are
in view. Ensure that Nexus has registered the correct position of all the cameras in the
capture volume.
b) Subject Preparation
Measure the patients body mass, height, leg length, knee width, and ankle
width. (Note: Jot down the value and do the measurement for both sides)
1. In the Subjects resources pane toolbar, click the Create a new Subject from a Template
button.