Note: Although FASTEST requires two units, you initiate it only from the master unit.
Both units use FASTEST settings from the master unit.
17
FOT-930
Reference patchcord
Bulkhead
adapter
Reference patchcord
Bulkhead
adapter
Note: The FOT-930 has been designed not to require offset nulling under normal operation.
However, you should perform it whenever environmental conditions change significantly.
19
Note: Your FOT-930 has been designed not to require a new setup each time you perform a FASTEST.
It keeps your parameters in memory even when you turn off the unit.
Press.
18
2
20
21
21a
Press to turn
the unit on.
4
Select.
21b
2006 EXFO Electro-Optical Engineering Inc. All rights reserved.
Printed in Canada (2006-08)
P/N: 1047282
3
Configure FASTEST parameters.
Select.
Press.
Note: Your FOT-930 has been designed not to perform a new side-by-side reference each time you
perform a FASTEST. It keeps the value in memory even when you turn off the unit.
On the master
unit, press.
12
14
Select.
Reference patchcord
8
9
13
Reference patchcord
Bulkhead
adapter
Press.
Connect as shown.
FASTEST port
Termination
10
Component
under test
Terminate the fiber as close as possible before the component under test.
Wrap it at least 10 turns around a mandrel or small diameter tool,
adding turns until the reading stabilizes.
Press.
15
Note: After a few seconds, the unit displays new reference values for each wavelength on both units.
16
Press to save,
then remove the termination.
11
Disconnect the two patchcords from the bulkhead only and connect them to the fiber under test
(using the bulkhead adapter or the system patch panels).