Procedure for Taking Radiation Patterns Using Antenna

Measurement Utility Software and VNA


1-Start the VNA
2- Calibrate the cableCalibration Step12345-

Go
Go
Go
Go
Go

to measurement Select S21


to Stimulus Select Start and Stop Frequency 1GHz to 8GHz
to Sweep Setup Power-10dB
to Calibration- Calibration Kit-85032F
to Calibrate-2-port Cal Reflection:
1-Connect Open male from the kit to port1 cable then click on Open
male at the VNA.
2- Connect Short male from the kit to port1 cable then click on Short
male at the VNA.
3-Connect Load male from the kit to port1 cable then click on Load male
at the VNA.
4-Connect Open male from the kit to port2 cable then click on Open
male at the VNA.
5- Connect Short male from the kit to port2 cable then click on Short
male at the VNA.
6-Connect Load male from the kit to port2 cable then click on Load male
at the VNA.
7-Return
6- Transmission:
1-Connect to port1 cable to port2 cable via feed thru then click on Thru at
the VNA.
2- Press Done at the VNA
7- Go to Save and Recall
8- Save State
9- Choose state
10-Save settings
If we have done the calibration and have the save settings already
then we do not need to re calibrate the VNA we just recall the
calibration settings byGo to save and Recall
Recall state
Select state in which you have stored your settings already (like state 1).
3- Connect the cables to transmitter and receiver
4- Connect the antenna positioner and controller as shown in figure-1;

Figure-1

5-Switch on the controller box


6- Close all other applications in PC like Excel and others.
7- Disable the antivirus during Positioner movement.

8- Click on the AMU_SLTL_S.exe shortcut on your desktop.


9- It will Show like Figure-2, now click on the Start button.
10-It will open anew window shown in figure-3.

Figure-2

Figure-3

11-For Getting serial information port Go to Computer-properties- Device


Manager-Ports(COM& LPT) as shown in figure-4-5.

Figure-4

Figure-5
12-Now enter the serial port -5 as shown in figure 5

Figure-6

13-Now set the Direction clockwise or Anticlockwise by selecting the popupmenu .

14- Step has been fixed by 1 Degree.


15- We can set the Delay from 0.25 second to 20 second by selecting Delay(s).
as shown in figure-7.

Figure-7

16- Now click on the connect button as shown in figure -8.

Figure-8

17- Now click on The SCAN button as shown in figure-9.

Figure-9

18- After pressing the SCAN button Generation of polar plot has been started as
shown in figure-10.

Figure10
19- After Completion of all 361 degree wait for a second and then press save
button it will show a window asking the filename, and then please give any file
name this will save your data in .mat format.

Figure11

Figure12
20- After saving data press Polar plot and browse your saved data file. This will
Generate polar plot as shown by figure-14.

Figure13

Figure14

21- You can also save your data in .excel format by pressing Generate report
button as shown in figure-15.

Figure15
22- Report in Excel shown in figure-16.