User Manual
1005541
<Number>1
Ta b l e o f C o n t e n ts
1
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
3
4
5
7
8
9
11
11
M easurement . . . . . . . . . . . . . . . . . .
Q uali ty Contro l . . . . . . . . . . . . . . . . .
I dentification of an unknown Substance
Q u a n t i t a t i o n o f a S ample Component .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
13
24
31
35
Transmission Measuring . . . . . . . . . . . . . . . . . . 45
5.1
5.2
5.3
5.4
.
.
.
.
.
.
.
.
Sample Preparation . . . . . . . . . . . . . . . . . . . . . . 41
4.1
4.2
4.3
.
.
.
.
.
.
.
.
G eneral Information . . . .
AL P H A -T . . . . . . . . . . .
Positioning sample in the
ment . . . . . . . . . . . . . .
M easuring Sample . . . . .
. . . . . . .
. . . . . . .
ALPHA-T
. . . . . . .
. . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
sample compart. . . . . . . . . . . . .
. . . . . . . . . . . . .
. 45
. 46
. 47
. 50
ATR Measuring . . . . . . . . . . . . . . . . . . . . . . . . . 51
6.1
6.2
6.3
6.4
6.5
6.6
6.7
G eneral Information . . . . . . . . .
AT R Measurement Module . . . .
Pe r f o r m i n g C l e a n n e s s T e s t . . .
Placing Sample on AT R Crystal
M easuring Sample . . . . . . . . . .
Cleaning ATR Module . . . . . . .
AT R S pectra . . . . . . . . . . . . . .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
51
52
57
57
58
59
60
iii
Table of Contents
Ma i n t e n a n c e & R e p a i r . . . . . . . . . . . . . . . . . . . . 6 1
7.1
7.2
7.3
7.4
7.5
Replacing IR Source . . . . . . . .
Replacing Desiccant Bags . . . .
Re p l aci ng Wi n d o ws . . . . . . . . .
Cl e a n i n g W i n d o w s . . . . . . . . . .
Cl e a n i ng S pe ctrom eter Housing
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
62
65
67
70
70
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . 71
8.1
8.2
8.3
G eneral Information . . . . . . . . . . . . . . . . . . . . . . . . . 7 1
Available Diagnostic M eans . . . . . . . . . . . . . . . . . . . 7 2
Problem - Possible Causes - Sol u t i o n s . . . . . . . . . . . 7 9
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . 93
Consumables . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Dimensional Drawings . . . . . . . . . . . . . . . . . . . . 97
Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
iv
Chapt. 1
Chapt. 1
Chapt. 2
Type of measurement
ALPHA-T
Transmission
(Universal Sampling Module)
ALPHA-P
Attenuated total reflection
(Platinum ATR Module; avail- (ATR)
able as high pressure module
as well)
Solids, liquids
ALPHA-E
(Economic ATR Module)
Solids, liquids
Basic module
Exchangeable
measurement
module
Chapt. 2
2.2
B asic Module
IR source
Interferometer
Beam splitter
(material)
KBr (standard)
ZnSe (in case of high humidity option)
Window
KBr (standard)
ZnSe (in case of high humidity option)
Detector
CDE
H I
Type
Definition
USB port
Not available.
ETH/LAN port
Chapt. 2
Type
Definition
TKD LED
The green TKD LED indicates that the interferometer mirror is within the data acquisition range. When data acquisition actually takes place the light intensity changes to
bright green. The TKD abbreviation means Take Data.
ACC LED
DIR/ERR LED
Ext. Accessory port This port can be used to connect external optional accessories, e.g. detector, remote trigger etc.
TAP port
CFG switch
SBY/RES button
POWER port
Kensington lock
2.3
Chapt. 2
In case of measuring liquid and gaseous samples optional liquid and gas
cells are available. These gas cells can be mounted into the sample
compartment of the transmission measurement module. For further
details on this subject see also chapter 5.
B
E
C
Figure 3: ALPHA-T
Definition
A
Lock/release
Status indicator
2.3.1
A L P H A - T s a m p l e c o m pa r t m e n t
To o p e n t h e s a m p l e c o m pa r t m e n t . . .
...reach under the bottom edge of the sample compartment lid
and lift it up.
Chapt. 2
B
C
Definition
A
Sample holder
2.4
Chapt. 2
Figure 5: ALPHA-P
Definition
A
Anvil
ATR crystal
Basic module
Status indicator
2.5
Chapt. 2
A
E
Figure 6: ALPHA-E
Definition
A
Anvil
Basic module
Status indicator
2.6
Exchanging Measurement
Module
Chapt. 2
10
Operating Spectrometer
Chapt. 2
W h a t h a p p e n s a ft e r e x c h a n g i n g t h e m e a s u r ement module?
OPUS automatically recognizes the newly inserted measurement
module and starts a performance test (accessory test). For more
detailed information refer to the OPUS/MENTOR manual.
2.7
Operating Sp ectrometer
2.8
A c t i v a t i n g Sta n d - b y M o d e
H o w t o s e t A L P H A i n t o s ta n d - b y m o d e ?
1
11
Chapt. 2
12
Chapt. 3
3.1
M easurement
3.1.1
General Information
13
Chapt. 3
W h e n m e a s u r i n g a s a m p l e , a l w a y s pa y a t t e n t i o n
to the following issues:
If the sample needs to be prepared, due to its composition,
prepare it carefully and avoid preparation errors. For information about sample preparation refer to chapter 4.
Avoid measurement errors (e.g. the ATR crystal is not covered
completely with the sample).
3.1.2
14
Measurement
Chapt. 3
4
5
The background measurement takes a few seconds. The measurement progress is shown in the lower status bar.
Put the sample in the spectrometer sample compartment or on
the ATR crystal.
6
7
15
Chapt. 3
3.1.3
M a n i p u l a t i n g a S a m p l e Spe c t r u m
Spectrum before a
baseline correction
Spectrum after a
baseline correction
16
Measurement
Chapt. 3
Note: To undo a baseline correction, select the Undo Changes command in the OPUS File menu.
Spectrum before
an
atmospheric
compensation
Spectrum after
an atmospheric
compensation
17
Chapt. 3
Note: The Atmospheric Compensation command requires a spectrum file that includes, amongst others, the data blocks SSC
(Sample Single Channel) and RSC (Reference Single Channel). (See figure 15.) Otherwise, an atmospheric compensation cannot be performed.
OPUS Browser with a loaded spectrum file.
This file contains, amongst others, a SSC data block and a RSC data block.
2
3
18
Measurement
Chapt. 3
3.1.4
Peak Picking
2
3
Figure 17: a)Peak pick result in case of absorption b) Peak pick result in case of transmission
19
Chapt. 3
Line
20
Using the line in the spectrum window, you can select interactively the peaks of which the frequency value is of interest to you.
symbol and move the line
To do this, place the cursor on the
with pressed left mouse button up- or downwards. The frequency
values of all peaks above this line are displayed (see figure 18).
Click on the Save button to save the currently displayed peak
position values. Thereupon, a PEAK data block is added to the
spectrum file in the OPUS browser.
Alternatively, you can view the peak picking result in form of a
report. The report is displayed below the spectrum window. See
figure 19. For information about the permanent display of the
information view, refer to the OPUS/MENTOR manual.
Measurement
Chapt. 3
Peak picking
result
displayed in
the
spectrum window
Peak picking
report
Figure 19: Peak picking report
21
Chapt. 3
2
3
3.1.5
22
Measurement
Chapt. 3
Note: If you want to preserve the original spectrum select the Save
File As command in the OPUS File menu and define a different
file name and/or path for saving the processed spectrum.
H o w t o s ta r t t h e n e x t s a m p l e m e a s u r e m e n t :
1
Note: If you want to measure the next sample after having processed
a spectrum it is advisable to save this spectrum. To do this,
also activate the Save Changes checkbox. (See section How to
save a processed spectrum.)
Click on the Go button. As a result, the Measurement dialog window (figure 10) opens again allowing you to start the next sample
measurement by clicking on the corresponding button (see
section 3.1.2.)
23
Chapt. 3
3.2
Quality Control
3.2.1
General Information
3.2.2
Quality Control
Chapt. 3
H o w t o s e t u p a n e w Q u i c k C o m pa r e m e t h o d :
1
Click on the Evaluate tab and then on the Quick Compare button
in the OPUS wizard. Thereupon, the following OPUS message
appears:
Note: This message appears only when you click on the Quick Compare button for the very first time because at this stage a quick
compare method has not yet be set up. Otherwise, the quick
compare method set up or loaded at last is used automatically.
25
Chapt. 3
On this page, you select the reference spectrum/reference spectra for Quick Compare.
You can select either:
- only one reference spectrum or
- all reference spectra that are
stored in a certain directory (In
this case, the reference spectra
can originate from the different
sample materials.) or
- the average spectrum of all reference spectra stored in a certain directory. (In this case, the
reference spectra must originate from the same sample
material!)
Click on the Parameters tab and enter the desired parameter values (see figure 26).
The entry of a comment is
optional.
When defining the frequency
region for the quick compare,
you can either:
- use the file limits or,
- enter the values for one or
more regions manually or
- select one or more frequency
regions interactively. (See
fig. 27.)
a) With the threshold value, you determine the degree of similarity between the reference spectrum and the sample spectrum. In other words, you determine how similar reference spectrum and sample spectrum need to be in order that the quick compare result is OK. As a
general guideline: The higher the threshold value is the higher the similarity between the reference and the sample spectrum needs to be in order to get an OK result.
Quality Control
Chapt. 3
H o w t o p e r f o r m Q u i c k C o m pa r e :
1
2
3
Measure the sample of which you want to test the quality and
manipulate the sample spectrum, if required. See section 3.1. (Do
not unload the spectrum yet!)
Click on the Evaluate level in the OPUS wizard.
Click on the Quick Compare button. As a result, OPUS performs
automatically a Quick Compare using the method you have set up
before or the method which has been loaded last.
Immediately after OPUS has executed a Quick Compare evaluation, the result is displayed automatically. (See figure 29, for
example.)
27
Chapt. 3
H o w t o i n t e r p r e t a Q u i c k C o m pa r e r e s u l t :
The Quick Compare result is OK if the correlation value is higher than
the threshold value you have determined. Analogous to this, the Quick
Compare result is not OK if the correlation value is lower than the
threshold value. In case of a Quick Compare evaluation, the correlation
coefficient1 r is a measure for the similarity between the two items
"reference spectrum" and "sample spectrum".
Depending on what you have selected as reference for the Quick
Compare (a single reference spectrum, an average spectrum or all
spectra of a certain directory), keep in mind the following when
interpreting the result:
If you have selected a single file or an average as reference,
you will get either an OK or Not OK as result (see figure 29, 30
and 31). In both cases, the displayed result allows you to see
at first sight whether the analyzed sample meets the specified
quality criteria or not (see figure 29 and 30).
If you have selected all spectrum files of a certain directory as
reference, you will not get a simply OK or Not OK as result. In
this case, only those reference spectra which have a sufficient
similarity with the sample spectrum, due to the threshold value
you have entered and the correlation value OPUS has calculated, are displayed black. All other reference spectra which do
not show a sufficient similarity with your sample spectrum are
displayed gray (see figure 32)1. Normally, the correlation coefficient values can range from -1,0 (i.e. no similarity between the
reference spectrum and sample spectrum at all) to +1,0 (i.e. absolutely identical spectra). For
the quick compare result, this value is converted into percentage with the range 1 r 0
being mapped to 0% and the range 0 r 1 being mapped linearly to 0% ... 100%.
28
Quality Control
Chapt. 3
Figure 29: Result of a Quick Compare using one Reference - Sample Spectrum is OK
Figure 30: Result of a Quick Compare using one Reference - Sample Spectrum is not OK
29
Chapt. 3
Figure 31: Result of a Quick Compare using an average as reference - Sample spectrum is OK
Figure 32: Result of a Quick Compare using several references - 1 hit has been found
30
3.3
Chapt. 3
Identification of an unknown
Substa nce
3.3.1
General Information
Note: Besides the two libraries delivered with the OPUS software,
Bruker also offers a number of additional spectra libraries
which you can purchase.
31
Chapt. 3
In addition, OPUS offers you also the possibility to create your own
spectra library which is adapted to your special needs regarding the
substances you intend to identify. See section 3.3.3 Creating, building
up and maintaining your own spectra library.
3.3.2
T h e g e n e r a l c o u r s e o f a s u b s ta n c e i d e n tification
substance
and
32
Chapt. 3
Substance
information
Structural
formula
Hit quality
value
Hit list
The hit quality value in the hit list gives you a general clue to
what extent the spectrum of the unknown substance (also
called query spectrum) resembles the library spectrum.
The hit quality value can range from 0 (i.e. no resemblance at
all) to 1000 (i.e. absolute conformity of the library spectrum
with the query spectrum). The hit list is sorted according the hit
quality value, i.e. the hit with the highest quality value ranks
first in the hit list.
Moreover, you can compare visually the query spectrum with
the library spectrum to find out whether there is a certain
resemblance between the two spectra.
3.3.3
C r e a t i n g , b u i l d i n g u p a n d m a i n ta i n i n g
your own spectra library
Initialize Library
Store Spectrum in Library
Library Editor
Library Browser
33
Chapt. 3
1. In an IR spectrum, the position and intensity of the absorption bands are very substance-specific. Similar to a human fingerprint, the absorption spectrum of a substance is well-suited for
the identification of it.
34
3.4
Chapt. 3
3.4.1
General Information
3.4.2
T h e g e n e r a l c o u r s e o f a q u a n t i ta t i v e
analysis
2
3
35
Chapt. 3
W h e n c h o o s i n g t h e c a l i b r a t i o n s a m p l e s , pa y
a t t e n t i o n t o t h e f o l l o w i n g a s p e c ts :
Ideally, the calibration sample set should contain at least 5
samples.
Make sure that the concentration value of the component in
question differs from calibration sample to calibration sample.
3
4
Enter
the
component
name
Enter
the
unit of measurement
36
Chapt. 3
Already at this stage (i.e. on the basis of one calibration spectrum and its
corresponding concentration value), OPUS calculates a calibration line1
(see figure 35). In the further course of the QUANT method setup, this
calibration line will be improved by adding the data of the other
calibration samples.
Note: When you add the first calibration spectrum to a new QUANT
method, make sure that the New option button is activated.
Note: Set the integration area in such a way that the spectral band
resulting from the component in question is within integration
area. For this reason, it is recommended to set the integration
area not until at least 3 or 4 calibration spectra have been
added to the QUANT method because only at this stage you
can recognize clearly the spectral band resulting from the
component in question (see figure 37).
Figure 36: Defining the Integration Area and the Integration Method
37
Chapt. 3
11 Open the Quant Builder dialog again by clicking in the OPUS wizard on the Evaluate tab, selecting in the drop-down list the option
Quant Builder and clicking on the Go button.
12 Enter the component concentration value of the calibration sample in question.
Note: For the second and every following calibration sample, you do
not need to enter component name and unit again.
Figure 37: Defining the integration area on the basis of several calibration spectra
At this stage, the window includes all calibration spectra added to the
QUANT method so far. The display of several calibration spectra
enables you to set the integration area because now the band resulting
from the component to be quantified is clearly recognizable due to
differences in the band height because the band height is quantitydependent.
For detailed information about the Quant Builder command refer to the
OPUS/MENTOR manual.
38
Chapt. 3
H o w t o p e r f o r m a q u a n t i ta t i v e a n a l y s i s :
1
3
4
Note: In case this check box has not been activated double-click on
the QUANT data block of the spectrum file in order to display
the quantitative analysis result.
39
Chapt. 3
H o w t o i n t e r p r e t a q u a nt i ta t i v e a n a l y s i s r e s u l t :
the integration result for the spectral band which results from
the component to be quantitated.
1. Sigma is the standard deviation of the calculated concentration value. The sigma value is determined by the quality of the QUANT method you have set up. Ideally, the sigma value should
be as small as possible.
40
Chapt. 4
S a m p l e P r e pa r a t i o n
Proper sample preparation is crucial to obtain good and meaningful
spectra. This chapter describes the most important sample preparation
methods1. Useful information will give you some help in selecting the
most suitable sample preparation method for a given sample.
How must the sam- homogeneous (constant concentration or composition within the
ple be like?
sample area to be analyzed)
Disadvantages of a misleading or erroneous measuring data
non-homogeneous
sample
What has to be
aggregation state of the sample
considered when
absorptivity of the sample
selecting a sample
preparation
method?
sample has to be very thin
What has to be
considered in case sample has to be diluted by a solvent or powder that is not
of a strongly
strongly absorbing
absorbing sample? (otherwise the spectrum is not meaningful)
What kind of sol Carbon tetrachloride
vents can be used? Carbon disulphide
Chloroform
Cyclohexane
Acetonitrile
Tetrachloroethylene
ALPHA allows to measure both solids and liquids. The most common
sample preparation methods include:
No sample preparation (e.g. self-supporting film or measurement using an ATR unit)
Preparing a KBr pellet (for solids)
Preparing thin film of liquid sample solution between two transparent plates (for liquids)
Caution: Most of the described sample preparation methods involve
the use of hygroscopic materials (such as NaCl or KBr). If
these materials come in contact with water or alcoholic solvents, they begin to dissolve or become cloudy and thus,
impair the measurement results. Therefore, avoid all
sources of water and even alcohol (ethanol and methanol).
1. For more detailed information about the different sample preparation methods refer to the relevant specialist literature.
Bruker Optik GmbH
41
Chapt. 4
Sample Preparation
4.1
Type of sample
Advantage
Disadvantage
Tip
4.2
42
N o Sample Preparation
K Br Pellet
Type of sample
solids
Advantages
meaningful IR spectrum
KBr shows no absorption in the wavenumber area 4000cm-1 to
250cm-1
this prevents the sample spectrum from containing any disturbing absorption bands
Disadvantages
What has to be
considered during
sample preparation?
Tip
A common mistake is to use too much sample material. The concentration of the sample in KBr should be between 0.2% and 1%
(i.e. typically a mixing ratio of 300:1).
4.3
Chapt. 4
Type of sample
Advantages
Disadvantages
What has to be
considered during
sample preparation?
Tips
43
Chapt. 4
44
Sample Preparation
Chapt. 5
Tr a n s m i s s i o n M e a s u ring
5.1
General Information
M e a s u r i n g i n t r a n s m i s s i o n i s s u i t e d f o r t h e f o llowing samples:
Solids, in the form of powder as well (except for highly absorbing substances)
Liquids
Gases
Depending on the sample material, sample preparation can be very
complex in case of measuring in transmission mode, see chapter 4.
A d v a n ta g e s :
Almost all kinds of samples can be analyzed.
Measurement represents the entire sample.
D i s a d v a n ta g e s :
Depending on the sample material a complex and sometimes
difficult sample preparation is required. A successful transmission measurement substantially depends on how carefully the
sample has been prepared.
Depending on the sample preparation method required, measuring in transmission may sometimes destroy the sample.
Too much a sample quantity can cause total absorption.
Additional accessories (e.g. liquid and gas cells) are required.
45
Chapt. 5
Transmission Measuring
5.2
A LPHA-T
B
C
Definition
46
Sample holder
5.3
Chapt. 5
Depending on the type of sample used (solid, liquid or gas) you can
position the sample into the ALPHA-T sample compartment by means of
different measurement accessories:
sample holder supplied by default; especially suited for KBr
pellets or a thin film between two IR-transparent plates
plastic film holder (optional)
liquid cell (optional)
gas cell (optional)
5.3.1
Sample Holder
Fixture
Sample holder
5.3.2
The plastic film holder is put into the fixture of the sample holder
(figure 41).
5.3.3
Liquid Cell
47
Chapt. 5
Transmission Measuring
5.3.4
Gas Cell
To insert the gas cell into the ALPHA-T sample compartment you first
have to displace the sample holder in the sample compartment, see
figure 43a and b.
48
Chapt. 5
To d i s p l a c e t h e s a m p l e h ol d e r i n t h e A L P H A - T
s a m p l e c o m pa r t m e n t . . .
1
2
3
TORX screw
Detachable casing
Figure 44: Removing the casing at the front of the sample compartment
4
5
6
Loosen the two TORX screws which fix the sample holder, see
figure 44.
Mount the sample holder into the right position, see figure 43b.
Put the detachable blue casing to the front of the sample compartment again.
To i n s e r t t h e g a s c e l l i n t o t h e s a m p l e c o m pa r tment...
1
2
...put the special gas cell holder into the fixture of the sample
holder, see figure 45a.
Position the gas cell on the sample holder as shown in figure 45b.
49
Chapt. 5
Transmission Measuring
5.4
Measuring Sample
50
Chapt. 6
AT R M e a s u r i n g
6.1
General Information
M e a s u r i n g i n AT R i s s u i t e d f o r t h e f o l l o w i n g
sample type:
Liquid
Powder
Paste
Gel and cream
Polymer film
Surface coating
A d v a n ta g e s :
Non-destructive analysis method, i.e. the sample has not to be
destructed for analysis
No sample preparation required
Only a small number of sample quantity required
Suited for samples which are actually too thick or highly
absorbing for transmission measurement
Due to the multiple reflection of the IR light, the sample
absorbs more light than in case of transmission measurement
D i s a d v a n ta g e s :
Falsified measurement result due to possible contaminations
caused by previous sample material, solvents or fingerprints.
In case of multi-layer materials only the first layer will be spectroscopically analyzed.
The quality of the measurement result substantially depends
on the contact pressure between sample and ATR crystal during measurement.
51
Chapt. 6
ATR Measuring
6.2
6.2.1
b) ALPHA-P
These modules differ from each other in the following technical features:
ATR-Module
Crystal material
available
Actual sampling
Moving pressure arm by
surface on ATR crystal means of a
ALPHA-E
Ge, ZnSe
5mm
knob
ALPHA-P
Diamond, Ge
Diamond:2 x 2mm
Ge: 2mm
lever
ALPHA-P
High Pressure
option
Diamond
2 x 2mm
knob
52
6.2.2
Chapt. 6
AT R C r y s ta l Ty p e s
Crystal type
To be used for
Diamond
5,000 - 550
Refraction index
n = 2.43
n = 4.01
n = 2.43
Caution: Zinc selenide is a toxic material. During normal spectrometer operation the material does not pose a health risk. Be
careful, however, if the crystal is broken. Avoid generating
dust. The dust is harmful by inhalation, ingestion or skin
absorption.
6.2.3
E x c h a n g i n g AT R C r y s ta l P l a t e
The ATR crystal plates can include one of the following crystal material:
ZnSe, Ge and diamond. The color of the crystal indicates which crystal
material is used.
Crystal material
Crystal color
Diamond
transparent
Ge (Germanium)
silver-gray
yellow
53
Chapt. 6
ATR Measuring
Put the stick into the bore of the crystal plate, see figure 47.
Rotating counterclockwise
releases the crystal plate
2
3
4
5
6
6.2.4
ALPHA-E
The pressure application device is moved upwards or downwards by
means of a knob (figure 48):
upwards - you have to rotate the knob counterclockwise
downwards - you have to rotate the knob clockwise
The inner part of the pressure arm is equipped with a slip-clutch
mechanism which is activated as soon as the optimal contact pressure
between sample and anvil has been applied. Further rotating the knob,
which is only possible by substantial force, would not have any effect on
the contact pressure already applied.
54
Chapt. 6
However, the contact pressure actually applied (i.e. how much the anvil
presses the sample against the ATR crystal) depends on how far you
rotate the pressure arm down.
Knob
Pressure arm
Anvil
You can swing away the pressure arm and the anvil from the sampling
surface. The pressure arm is rotatable by 360. If you swing the
pressure arm towards the sampling surface again, it snaps into place
exactly above the ATR crystal.
ALPHA-P
The pressure application device is moved upwards or downwards by
means of a lever (figure 49).
lever in back position - moves the anvil up
lever in front position - moves the anvil down
Lever to move the anvil
Height adjusting knob to move the complete pressure application device
Red control spot indicates optimal contact pressure
Anvil
55
Chapt. 6
ATR Measuring
The optimal contact pressure applied on the sample is ensured if the red
control spot (figure 50) is exactly in the middle of the round recess.
To a d j u s t t h e h e i g h t o f t h e c o m p l e t e p r e s s u r e
application device is required if...
1
56
...the lever is right in the front position but the anvil has not any
contact with the sample. In this case use the height adjusting
knob (see figure 51) and move the complete pressure application
device downwards until the red control spot is exactly in the
middle of the round recess.
You cannot move the lever right to the front position as the sample is too thick. In this case use the height adjusting knob (see
figure 51) and move the complete pressure application device
upwards, until the lever can be moved right to the front position
and the red control spot is exactly in the middle of the round
recess.
Chapt. 6
You can swing away the pressure application device from the sampling
surface. The pressure application device is rotatable by 360. If you
swing the pressure application device towards the sampling surface
again, it snaps into place exactly above the ATR crystal.
Note: For ALPHA-P as High-Pressure option refer to ALPHA-E.
6.3
Performing Cleanness Te st
6.4
Make sure that the sampling surface, ATR crystal and anvil are
clean, see chapter 6.6.
Place the sample on the middle of the sampling surface. Use a
small amount of sample, only enough to cover the ATR crystal.
57
Chapt. 6
ATR Measuring
Note: If the sample cannot be placed onto the ATR crystal as the distance between the ATR crystal and anvil is too small, move the
pressure application device upwards (see chapter 6.2.4).
4
5
Press the anvil against the sample (see chapter 6.2.4). Make sure
that the optimum contact pressure is used to press the sample
against the ATR crystal.
Acquire a sample spectrum (see chapter 6.5). For detailed information to perform a measurement refer to chapter 3.
Move the pressure application device up again and remove the
sample.
Note: Samples should not be left in contact with the crystal for an
extended period of time as some samples may degrade the
crystal material, and discoloration of the metal plate may
occur. Once the measurement has been made, remove the
sample from the crystal.
6.5
M easuring Sample
58
6.6
Chapt. 6
C l e a n i n g s o l v e n ts :
Alcohol
Acetone
Note: When using solvents inspect the materials safety data sheets
associated with the solvents you are using and observe any
recommended handling procedure.
The solvent used for cleaning depends on the sample that has been
analyzed. In all cases it is best to try to clean the sampling surface by the
mildest solvent possible.
Before cleaning...
...swing away the pressure application device from the sampling surface.
When cleaning...
59
Chapt. 6
ATR Measuring
6.7
ATR Sp ectra
60
Chapt. 7
M a i n t e n a n c e & R e pa i r
ALPHA is a low-maintenance instrument, i.e. the operator can replace
components with a limited service life (e.g. IR source). The following
maintenance procedures are described in this chapter:
Replacing IR source
Replacing desiccant
Replacing windows
Cleaning windows
Cleaning spectrometer housing
General Instructions
Perform only the maintenance procedures described in this manual.
Strictly observe the relevant safety precautions. Any failure to do so may
cause property damage or personal injury. In this particular case Bruker
does not assume any liability. Maintenance or repair procedures not
described in this manual should only be performed by a Bruker service
engineer. For service assistance, see appendix E.
Depending on the degree of hazard the safety instructions to be
observed in case of spectrometer maintenance or repair are classified
as follows:
Caution
Note
61
Chapt. 7
7.1
Replacing IR Source
) is yellow. A yellow
If the source is defective, the OPUS status light becomes red. The
following message pops up in OPUS: Source is broken. A red status
light does not allow to continue measuring. The source must be
replaced.
1
2
3
62
Replacing IR Source
Chapt. 7
5
6
Loosen the two fixing screws of the IR source (see figure 53). Use
the TORX TX20 screw driver supplied.
Strip off the green source plug (see figure 53). There will be a
beep.
IR source
Fixing screws
of the source
(TORX
screws)
Source plug
7
8
63
Chapt. 7
B
A
9 Fix the source by fastening the two TORX screws (see figure 53).
10 Put the green source plug into the SRC female connector. There
will be a beep.
11 Attach the cover plate and fasten the 4 TORX screws again (see
figure 52).
12 Connect the power plug to the mains socket again.
A ft e r r e p l a c i n g t h e I R s o u r c e . . .
...you have to perform an instrument test.
To p e r f o r m a n i n s t r u m e n t t e s t . . .
1
2
3
64
Chapt. 7
7.2
Rep l aci n g D e s ic c a n t B a g s
65
Chapt. 7
1
2
3
66
Pull out the used desiccant bags. A pair of tweezers may be helpful.
Replacing Windows
Chapt. 7
Desiccant
bags
The replacement desiccant bags are protected by an extra packaging. Open this packaging and take out the desiccant bags.
Note: Alternatively, you can place the spent bags into an
oven at 130C for at least 10 minutes.
Completely insert the desiccant bags properly into the spectrometer as shown in figure 57.
Caution: Do not cut or tear off the bags into single pieces, nor ingest
the desiccant.
8
9
7.3
Attach the cover plate and fasten the 4 TORX screws again (see
figure 56).
Connect the power plug to the mains socket again.
R eplacing Windows
The ALPHA basic module has two ports: one IR output and one IR input
port (see figure 58). The IR beam enters the measurement module by
the IR output port. Afterwards, the beam is directed to the basic module
again via the IR input port. The ports are closed by means of
exchangeable IR transparent windows.
Note: By default, the spectrometer is equipped with KBr windows. In
case of the High Humidity option the windows are made of
ZnSe.
Caution: Zinc selenide is a toxic material. During normal spectrometer operation the material does not pose a health risk. Be
careful, however, if a window is broken. Avoid generating
dust. The dust is harmful by inhalation, ingestion or skin
absorption.
67
Chapt. 7
Basic module
Exchangeable
measurement
module
2
3
4
68
Replacing Windows
7
8
Chapt. 7
Pull out the assembly tool from the holes on the assembly frame.
After replacing the windows push the measurement module right
against the basic module. Press the lock on the basic module
again right down (see chapter 2.6).
Lock/release
Window assembly
(window in assembly frame)
Fitting
tool
to
replace windows
A ft e r r e p l a c i n g t h e w i n d o w s . . .
...you have to perform an instrument test.
To p e r f o r m a n i n s t r u m e n t t e s t . . .
1
2
3
69
Chapt. 7
7.4
C leaning Windows
7.5
70
Chapt. 8
Tr o u ble s h o o t i n g
8.1
G en era l In fo r m a tio n
Click on the OPUS status light. The status light is in lower right
corner of the OPUS interface (see figure 62).
The Instrument Status dialog window opens. Click on the Send
Report button (see figure 63). The report is sent by e-mail to
opusreports@brukeroptics.
71
Chapt. 8
Troubleshooting
8.2
Sp e c t r o m e t e r Sta t u s I n d i c a t o r
The status indicator is on the spectrometer top side.
The spectrometer status indicator can have the following colors: green,
yellow and red. These colors indicate the following spectrometer status:
Green (permanent):
Spectrometer is in proper operating condition.
Green (flashing):
Spectrometer is in the stand-by mode.
Yellow:
Spectrometer is in the initialization phase or it indicates a warning (e.g. IR source has not yet reached its operating temperature, the air humidity content inside the spectrometer is too
high etc.)
Red:
Spectrometer malfunction, defective spectrometer component
72
Chapt. 8
O P U S Sta t u s L i g h t
The status light is in the lower right corner of the OPUS interface.
Status
light
Figure 62: OPUS interface
The OPUS status light can have the following colors: green, yellow, red
and gray. These colors indicate the following spectrometer status:
Green (permanent):
Spectrometer is in proper operating condition.
Yellow:
Spectrometer is in the initialization phase, or a warning is indicated (e.g. IR source has not yet reached its operating temperature, the validity period of the instrument is expired, the air
humidity content inside the spectrometer is too high etc.)
Red:
Spectrometer malfunction, defective spectrometer component, performance test / instrument test failed
Gray:
In OPUS, no spectrometer is selected or the spectrometer is
off or the spectrometer is in stand-by mode.
Note: In case of a yellow status light measuring is still possible. If
the status light is red or gray, measuring is not possible.
The color of both status lights, in OPUS and on the spectrometer, corresponds in most cases except for one case: if the
performance test or the instrument test has failed the OPUS
status light is red whereas the spectrometer status indicator
lights yellow.
73
Chapt. 8
Troubleshooting
O P U S I n s t r u m e n t St a t u s D i a l o g
Click on the OPUS status light (see figure 62). The Instrument Status
dialog opens.
A The status of the hardware components, e.g. source, laser, interferometer etc. is displayed in the upper icon line. The status can be
as follows:
OK (green):
Component is okay.
WARNING (yellow):
The exact meaning of a warning depends on the component in question. In case of the source a warning means:
End of the specified component lifetime is nearly
reached. In this case, measuring is still possible.
The source is still warming-up (duration: about 7 min.)
In this case, measuring is not possible.
ERROR (red):
Component is defective. In this case measuring is not
possible.
74
Chapt. 8
B The second row of icons refer to the possible active test channel
and indicates the result of the last instrument test performed. The
results can be as follows:
INACTIVE (yellow):
The single tests of the particular test category are disabled.
PASSED (green):
Instrument test passed. Test is still valid.
FAILED (red):
Instrument test failed.
To p e r f o r m a f a u l t d i a g n o s i s o f a pa r t i c u l a r
hardware component...
1
2
75
Chapt. 8
Troubleshooting
Status indication
(for the component in question)
Error message (for
the component in
question)
Component
in question
When you click on the Service Info button the diagnostics page of the
component in question opens. Figure 65 shows the diagnostics page of
the source. These pages contain all relevant information about the
current operating state of the respective spectrometer component.
Note: You can send these diagnostics pages as full report to the
Bruker service by e-mail allowing the technician a first remote
fault diagnostics (see section 8.1).
76
Chapt. 8
Error Message
b) Message window
Note: In case of a spectrometer problem indicated by an error message, the status indicator on the spectrometer and the OPUS
status light are yellow or red.
77
Chapt. 8
Troubleshooting
Sp e c t r o m e t e r D i a g n o s t i c L E D s
The diagnostic LEDs are at the spectrometer rear side.
ABC
Figure 67: ALPHA rear side -Diagnostic LEDs
Type
Definition
TKD LED
ACC LED
The yellow ACC LED indicates that the network accesses the
optics. During transition of measuring data the LED will be light yellow. The ACC abbreviation means Access.
DIR/ERR LED
P e r f o r m a n c e Te s t a n d I n s t r u m e n t Te s t
Both tests check whether the spectrometer achieves the specified
performance. But they differ as follows:
Performance Test
Instrument Test
Signal-to-noise test
Deviation from 100% line
Signal-to-noise test
Deviation from 100% line
Wavenumber accuracy test
11
Test duration
about 1 minute
about 5 minutes
78
8.3
Chapt. 8
Performance Test
Instrument Test
8.3.1
A P r o b l e m i n d i c a t e d b y t h e Sp e c t r o m e t e r
Sta t u s I n d i c a t o r
The status indicator is on the spectrometer top side (see figure 61).
Sp e c t r o m e t e r s ta t u s i n d i c a t o r i s o f f
Possible Cause
Solution
No voltage is applied
If the problem has a different cause... ...contact Bruker service (see appendix E).
Sp e c t r o m e t e r s ta t u s i n d i c a t o r i s y e l l o w
Possible Cause
Solution
Spectrometer is still in the initialization Wait until the spectrometer initialization is completed.
(duration: about 1 minute)
phase.
IR source has not yet reached its operating temperature.
Note: In this case the following
message appears: Device not ready.
79
Chapt. 8
Troubleshooting
Possible Cause
Solution
Click on the OPUS status light (figure 62). The Instrument Status dialog (figure 63) opens. The measurement channel of which the instrument test validity
period is expired is labelled Expired.
Start a new instrument test by clicking on the Instrument Test button in the OPUS wizard. Wait until the test
is completed (duration: about 5 minutes).
Air humidity inside the spectrometer is Replace the desiccant bags (see chapter 7 Repair and
too high. Note: In this case the folMaintenance, section 7.2).
lowing message appears: Humidity
out of range.
Laser wavenumber needs to be recali- Click on the Calibrate button (figure 68) and wait until
brated. Note: In this case a correthe calibration is completed.
sponding message appears in the
lower right corner of the OPUS
interface. See fig. 68.
If the problem has a different cause...
80
Chapt. 8
Sp e c t r o m e t e r s ta t u s i n d i c a t o r i s r e d
Possible Cause
Solution
8.3.2
A P r o b l e m i n d i c a t e d b y t h e O P U S Sta t u s
Light
The OPUS status light is in the lower right corner of the OPUS interface
(see figure 62).
O P U S Sta t u s l i g h t i s y e l l o w
Possible Cause
Solution
Click on the OPUS status light (figure 62). The Instrument Status dialog (figure 63) opens. The measurement channel of which the instrument test validity
period is expired is labelled Expired.
Start a new instrument test by clicking on the Instrument Test button in the OPUS wizard. Wait until the test
is completed (duration: about 5 minutes).
81
Chapt. 8
Troubleshooting
Possible Cause
Solution
Air humidity inside the spectrometer is Replace the desiccant bags (see chapter 7 Repair and
too high.
Maintenance, section 7.2).
Note: In this case the following
message appears: Humidity out of
range.
Laser wavenumber needs to be recali- Click on the Calibrate button (figure 68) and wait until
brated.
the calibration is completed.
Note: In this case a corresponding
message appears in the lower right
corner of the OPUS window. See
figure 68.
If the problem has a different cause...
O P U S s ta t u s l i g h t i s r e d
Possible Cause
Solution
O P U S s ta t u s l i g h t i s g r a y
Possible Cause
Solution
Spectrometer is still in the initialization Wait until the spectrometer initialization is completed.
phase.
(duration: about 1 minute)
For other possible causes...
The OPUS status light is gray while the spectrometer status light has
one of the three possible colors. This state indicates a communication
problem between ALPHA and the PC. This problem can be caused by:
82
Chapt. 8
Possible Cause
Solution
Connect the spectrometer to the PC using only the supplied cross-over, Cat5-cable. See Installation Instructions, chapter 3.
The RJ45-plug(s) of the data cable has Check the data cable cord connection at both RJ45(have) not been connected properly.
ports, the ETH/LAN port at the spectrometer rear side
and the Network/LAN port of the PC. See Installation
Instructions, section 3.2.
The factory-set standard IP address of Reassign the standard IP address to the spectrometer
the spectrometer (10.10.0.1) has been using the DIP-switch at the spectrometer rear side. See
changed.
Installation Instructions, chapter 6, section Assigning
the standard IP address using the DIP switch.
Spectrometer is in stand-by mode. In
this case, the status light on the spectrometer top side flashes green.
Spectrometer
IP address
83
Chapt. 8
Troubleshooting
84
8.3.3
Chapt. 8
P e r f o r m a n c e t e s t d o e s n o t s ta r t a u t omatically
Solution
No measurement module is placed Snap the measurement module correctly into place
in the basic module, or the measure- (see also chapter 2, section 2.6).
ment module is not snapped properly in place.
No ATR-crystal plate is installed or
not installed properly.
8.3.4
P e r f o r m a n c e Te s t f a i l e d
85
Chapt. 8
Troubleshooting
Possible Cause
Solution
Wait until the source has reached its operating temperature. Then, perform an instrument test by clicking on
the Instrument Test button in the OPUS wizard. Wait
until the test is completed (duration: about 5 minutes).
8.3.5
I n s t r u m e n t Te s t f a i l e d
86
Chapt. 8
A l l t h r e e t e s ts f a i l e d
Possible Cause
Solution
A l l t h r e e t e s ts o r o n l y t h e s i g n a l - t o - n o i s e - t e s t
or only the 100%-line test failed
Possible Cause
Solution
Order a new crystal plate with the corresponding ATRcrystal material. After receipt, replace the crystal plate
(see chapter 6, section 6.2.3).
87
Chapt. 8
Troubleshooting
Possible Cause
Solution
Solution
Laser wavenumber needs to be recali- Select the Setup Instrument Test command in the Valibrated.
dation menu. Click on the Measure LWN button in the
Setup OPV dialog.
If the problem has a different cause...
8.3.6
A Sp e c t r o m e t e r P r o b l e m i n d i c a t e d b y a n
Error Message in OPUS
Possible Cause
Solution
Source is broken.
IR source is defect.
88
Chapt. 8
Possible Cause
The spectrometer is in
stand-by mode.
Deactivate the stand-by mode by pressing the SBY/RES button for about 1 second. This button is at the spectrometer
rear side (see chapter 2, section 2.8).
Solution
An instrument calibration
is required. ...
Wrong accessory
inserted, unable to run
OVP tests.
Can not run OVP tests,
please check OVP Setup.
Wrong accessory
inserted, unable to run
OVP tests.
Can not run OVP tests,
please check OVP Setup.
If an error message
appears which is not
listed above...
89
Chapt. 8
Troubleshooting
8.3.7
The diagnostic LEDs are at the spectrometer rear side (see figure 67).
N o L E D l i g h ts
Possible Cause
Solution
No voltage is applied
Solution
Connect the spectrometer to the PC using only the supplied cross-over, Cat5-cable. See Installation Instructions, chapter 3.
The RJ45-plug(s) of the data cable has Check the data cable cord connection at both RJ45(have) not been connected properly.
ports, the ETH/LAN port at the spectrometer rear side
and the Network/LAN port of the PC. See Installation
Instructions, chapter 3 section 3.2.
The IP address of the ALPHA spectrometer is not selected in OPUS or
the wrong IP address has been
selected.
The factory-set standard IP address of Reassign the standard IP address to the spectrometer
the spectrometer (10.10.0.1) has been using the DIP-switch at the spectrometer rear side. See
changed.
Installation Instructions, chapter 6, section Assigning
the standard IP address using the DIP switch.
90
Chapt. 8
Possible Cause
Solution
T h e r e d E R R L E D l i g h ts p e r m a n e n t l y
A permanent red ERR LED indicates an interferometer and/or laser
problem.
Note: During a trouble-free spectrometer operation, this red LED
flashes in the rhythm of the scanner movement.
Possible Cause
Solution
91
Chapt. 8
92
Troubleshooting
A
General
Sp e c i f i c a t i o n s
Parameter
Specification
Dimension
22 x 30 x 25cm (w x d x h)
Weight
approx. 7kg
Power consumption
Environmental conditions
Optik
Electronics
Resolution
Wavenumber accuracy
Wavenumber reproducibility
Photometric accuracy
0.1 % T
Design
Source
Interferometer
Beamsplitter
Detector
DTGS
HeNe laser
A/D converter
Connectors
Ethernet
93
Appendix A
94
Specifications
10
Consumables
Part Number
Description
1005855
1000917
Desiccant bags
1005847
1006041
95
Appendix B
96
Consumables
11
Dimensional Drawings
A L P H A - T:
97
Appendix C
Dimensional Drawings
98
Appendix C
99
Appendix C
Dimensional Drawings
ALPHA-P
100
Appendix C
101
Appendix C
Dimensional Drawings
102
Appendix C
ALPHA-E
103
Appendix C
Dimensional Drawings
104
Appendix C
105
Appendix C
106
Dimensional Drawings
12
Glossary
ATR
Beamsplitter
A device used to produce two separate beams from one incident beam.
Generally a coated substrate that reflects and transmits approximately
equal portions of the incident radiation.
Calibration
Cross-over cable
Cable used for the direct connection of two workstations Network Interface Cards (NICs). NICs have been dedicated to send and receive
lines. A crossover cable connects from the send line of one NIC to the
receive line of the other NIC.
Detector
Ethernet
Ethernet is the most widely installed Local Area Network (LAN). A LAN
is a network of interconnected workstations sharing the resources of a
single processor or server within a relatively small geographic area
(e.g. an office building). Ethernet is a set of hardware and signaling
standards used for LANs. The most commonly installed systems are
10/100 BaseT.
Fourier Transform
107
Appendix D
Glossary
Frequency
The frequency is the number of periods per time unit. Since all electromagnetic waves travel at the speed of light, the frequency corresponding to a given wavelength can be calculated as: = c/
FT-IR
Instrument test
The instrument test checks the spectrometer performance. It automatically runs when ALPHA communicates with the PC for the very first
time. The test has to be repeated if the validity period (7 days) set by
Bruker has been expired.
Interferogram
108
Interferometer
Inside the interferometer the light beam is splitted into two beams by
the beam splitter. The most-commonly used Michelson interferometer
consists of a beam splitter, stationary and movable mirror. Due to the
continuous movement of one mirror the optical path lengths of the two
separated beams changes in time. Interference is effected when the
two light beams are recombined.
IP address
Laser
Micrometer (m)
Appendix D
MIR
Mid InfraRed; covers the spectral range from 4,000 to 400cm-1 wave
numbers. In this range fundamental vibrations are typically excited.
Performance test
The performance test checks the spectrometer performance. It automatically runs after starting OPUS, and exchanging a measurement
module or ATR crystal plate.
Reflectance
Reflection
The process by which incident illumination reacts with the sample and
is converted to radiation energy that subsequently travels back from
the sample surface at a defined wavelength.
Relative humidity
Resolution
Resolution in FT-IR is determined by the maximum optical path difference reached in the measurement of an interferogram. The spectral
resolution corresponds to the minimum possible spectral distance
between 2 adjacent spectral lines which can be resolved by the spectrometer. It is produced by the instrumental spectral line shape and the
line width of the samples absorption.
The instrumental line shape is reciprocally related to the travel distance
of the movable mirror in the interferometer, and also influenced by the
apodization function and divergence in the interferometer. The higher
the spectral resolution (smaller values), the higher the noise in the
spectrum at a given measurement time.
Sample
Special kind of substance which is put into the sample position and
measured.
Single-channel
spectrum
The arithmetical transformation of the interferometer, the Fourier transformation, made by the spectroscopic software results in the so-called
single-channel spectrum. This spectrum shows the intensity of each
single wavenumber. The single-spectrum includes the entire spectral
characteristics of the spectrometer, i.e. the signal detected by the
detector as function of the wavenumber. This signal is influenced by
the source type, beamsplitter transmittance, detector response etc.
Spectrometer
From the Latin word specere meaning image, and the Greek word metron meaning to measure. A spectrometer uses some sort of mechanical or electrical detection device to obtain the infrared spectrum of a
sample. It contains a source of infrared radiation, a sample compartment to allow the radiation to interact with a sample, a detector for radiation and units to determine and display the intensity of radiation.
Spectrum
A spectrum is an image of radiation energy dispersed into its wavelength constituents or a two-dimensional plot of radiation energy or
radiation energy ratio versus wavelength. It includes a set of absorption, reflection or transmission values which have been measured by
an IR spectrometer based on defined wavelengths. The features in an
infrared spectrum correlate with the presence of functional groups of
molecules of a sample.
109
Appendix D
Glossary
Transmission
Transmission measurement
Transmittance
Validation
Wave number
The term wave number is the commonly used dimension for the x-axis
1
in FT-IR spectroscopy. It is defined as = ----. The unit is cm-1.
0
Wavelength
110
13
Service Addresses
Bruker Optik has an international network of branch offices and
representations to ensure a competent customer service throughout the
world. The following list includes the addresses of Bruker headquarters.
For a complete list with the addresses and telephone numbers of the
Bruker branch offices and representations worldwide, refer to the
internet:
http://www.brukeroptics.com/offices.htmll
North America
Bruker Optics Inc
19 Fortune Drive, Manning Park
Billerica, MA 01821 - 3991
USA
Phone: +1 978 439 9899 (ext. 5227)
Fax: +1 978 663 9177
www.brukeroptics.com
info@brukeroptics.com
Europe
Bruker Optik GmbH
Rudolf-Plank-Str. 27
76275 Ettlingen
Germany
Phone: +49 7243 504 619/600
Fax: +49 7243 504 698
www.brukeroptics.com
ir_service@brukeroptics.de
Asia
Bruker Hong Kong Ltd.
Unit 509, 5/F, Tower II
Enterprise Square No 9
Sheung Yuet Road
Kowloon Bay
Hong Kong
www.brukeroptics.com
Phone: +852 2796 6100
Fax: +852 2796 6109
hk@brukeroptics.com.hk
111
Appendix E
112
Service Addresses
Index
In OPUS 88
Error Messages
In OPUS 77
ETH/LAN Port 4
Ethernet 107
A
ALPHA-E 8, 9, 54
ALPHA-P 7, 8, 55
ALPHA-T 5, 6, 46
Gas Cell 48
Liquid Cell 47
Sample Compartment 6, 47
Sample Holder 47
Atmospheric Compensation 17, 18
ATR
Cleaning 59
Crystal Types 53
Height Adjustment 56
Measuring 51
Measuring Sample 58
Module 52
Placing Sample 57
Pressure Application Device 54
Sample Measurement 58
Spectra 60
B
Baseline Correction 16
Basic Module 3
I
Instrument Test 78, 86, 108
Interferometer 108
IR Source 62
Replacing 62, 64
K
KBr Pellet 42
Kensington Lock 5
L
Laser 108
LEDs 4
M
Maintenance Instructions 61
Measurement 13, 14
Background 15
Procedure 14
Sample 15
Measurement Module
Exchanging 9
O
C
Calibration Line 37
Calibration Samples 36
CFG Switch 5
Communication Connection 83
Connecting Port 4
Consumables 95
Desiccant 65
Replacing 66
Detector 107
Diagnostic LEDs 78
Troubleshooting 90
Diagnostics LEDs
Troubleshooting 90, 91
Quality Control 24
Procedure 24
Quantitative Analysis 35, 39
Method 36
Result 40
Quick Compare 24, 27
Method 25
Result 28
Quick User Guide 13
E
Error Message
Bruker Optik GmbH
Index
R
Reference Sample 24
Relative Humidity 109
S
Sample Measurement 50
Sample Preparation 41
Methods 41
Service Addresses 111
Sigma 40
Specifications 93
Spectra Library 33, 34
Spectrometer Status Indicator 72, 79,
81
Spectrum Search 31
Result 32
Stand-by Mode 11
Substance Identification 31, 32
T
TAP Port 5
Transmission Measurement 45
Troubleshooting 71, 74
V
Validation 110
W
Wave Number 110
Wavelength 110
Windows 67
Cleaning 70
Replacing 67, 68, 69
Chapt. 14