Anda di halaman 1dari 38

Doble Engineering Company

85 Walnut Street
Watertown, MA 02172
(617) 926-4900
(617) 926-0528 (FAX)

M4000
Leakage Reactance Test Procedure
Rev #

Revision History
Rev. #
1.0
1.01
1.02

Date:
5/22/97
7/1/97
8/12/98

Description of Change
Preliminary release
Added section-2.3.5 Test Excitation Level
Updated File locations on network

Author: Corbett J. Weinberg


Date Printed: 11/2/00

Leakage Reactance Test Procedure


K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC

Page 1 of 38

1. SCOPE
2. LEAKAGE REACTANCE
2.1. Running Leakage Reactance
2.1.1. Toolbar
2.1.2. Menu
2.1.3. Exit
2.2. Leakage Reactance Test Screen
2.2.1. Test Screen Inputs ( Below is a list of the inputs unique to Leakage Reactance only.)
2.2.1.1Nameplate Data
5
12.2.1.2 Nameplate Data Window
6
2.2.1.3 Single Phase Transformer Test
7
2.2.1.4 3 Phase Equivalent Test
8
2.2.1.5 3 Phase per Phase - Delta and 3 Phase per Phase - Wye
9
2.2.1.6 Administrative Information
10
2.2.1.7 Test Conditions
11
2.2.1.8 Serial Numbers
12
2.2.1.9 Configuration Menu
12
2.3. Leakage Reactance Testing
2.3.1. Single Phase Transformer Test
2.3.2. Three Phase Equivalent Test
2.3.3. Three Phase per Phase - Delta
2.3.4. Three Phase per - Wye
2.3.5. Test Excitation Levels
2.3.6. Test Screen functions
2.3.7. Calibration Settings test
2.4. Results
2.4.1. % Impedance and % Reactance
2.5. Save
2.5.1. File- Save
2.5.2. Save Icon
2.6. Print Functions
2.6.1. File- Print
2.6.2. F7 function key
2.6.3. Printer Icon
2.7. Print Verification
2.8. Notes
Leakage Reactance Test Procedure
K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC

Page 2 of 38

2.8.1. Note Icon


2.8.2. Enter and Save
2.8.3. Save to Leakreac\Data\02_03_01\

1.

Scope

The scope of this test plan is the testing of the Leakage Reactance portion of the M4000 software
for the Doble Engineering Company. Successful execution of this plan verifies the Doble
Engineering Companys internal requirements for the Leakage Reactance portion of the M4000
software and validates user needs. Development Managers and Engineers should review this
document to ensure all test requirements are included in the plan. Quality Assurance Managers and
Engineers should review this document to ensure all test requirements are being met.
Tests for all general requirements not directly related to Leakage Reactance can be found at
\\Engineering\Archive\QA\M4000\M4000win\HREF="file://///Engineering/archive/QA/M4000/M4
000WIN/TESTPLAN/REL_2_03/Tp0203_1.doc

2.

Leakage Reactance

Equipment Needed
M4110 Leakage Reactance Interface
M4100 Insulation Analyzer (The Instrument)
M4200 Controller running Win 95
750 Ohm Production Test Specimen
0.1 Ohm Specimen for Excitation testing
0.2 Volt meter
Safety switch
Strobe light
Red and Blue Low Voltage Leads
External Ground Cable
Test Leads
Setup
1. Connect the Red and Blue low voltage leads from the M4100 to the low voltage lead
connection on the M4110 Leakage Reactance Interface M4110
2. Connect a couple of leads from M4110 Leakage Reactance Interface to the
3. Connect a Volt meter across the Specimen
4. Ground the M4110 Leakage Reactance Interface
5. Connect the Strobe light to M4110 Leakage Reactance Interface
6. Connect the safety switch to M4110 Leakage Reactance Interface
7. Connect M4200 Controller running Win 95 to M4100 Insulation Analyzer (Instrument)
8. Connect M4200 Controller running Win 95 to AC Power
9. Connect M4110 Leakage Reactance Interface to AC Power
10. Connect M4100 Insulation Analyzer (The Instrument) to AC Power

2.1.

Running Leakage Reactance

2.1.1. Toolbar
Purpose: To Enter into the Leakage Reactance Mode using the Leakage Reactance icon on the
toolbar.
Procedure:
Leakage Reactance Test Procedure
K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC

Page 3 of 38

1. Run the M4000 software.


At the toolbar click on the
button.
Expected Result: The Leakage Reactance Clipboard screen appears.
2.

2.1.2. Menu
Purpose: To Enter into the Leakage Reactance Mode using the Leakage Reactance Test menu
item.
Procedure:
1. Run the M4000 software.
2. Choose Leakage Reactance Test from the Mode menu item on the M4000 main window.
Expected Result: The Leakage Reactance Clipboard screen appears.
2.1.3. Exit
File-Exit
Purpose: Verify that you can exit out of M4000 by selecting File-Exit from the main menu.
Procedure:
1. Choose File-Exit from the main menu of the M4000 Entry Screen.
2. Checkpoint: A pop-up window appears saying Do you wish to exit M4000 for windows ?
3. Select Yes.
Expected Result: You are exited out of the M4000 for Windows program.
X
Purpose: Verify that you can exit out of M4000 by clicking on the X on the top right hand
corner of the screen.
Procedure:
1. Clicking on the X on the top right hand corner of the screen.
2. Checkpoint: A pop-up window appears saying Do you wish to exit M4000 for windows ?
3. Select Yes.
Expected Result: You are exited out of the M4000 for Windows program.
Alt-F4
Purpose: Verify that you can exit out of M4000 by hitting Alt_F4.
Procedure:
1. Hit Alt_F4.
2. Checkpoint: A pop-up window appears saying Do you wish to exit M4000 for windows ?
3. Select Yes.
Expected Result: You are exited out of the M4000 for Windows program.
Corner Icon
Purpose: Verify that you can exit out of M4000 by double-clicking on the M4000 Icon at the top
left corner of the screen..
Procedure:
1. Double-click on the M4000 Icon at the top left corner of the screen..
2. Checkpoint: A pop-up window appears saying Do you wish to exit M4000 for windows ?
3. Select Yes.
Expected Result: You are exited out of the M4000 for Windows program.

Leakage Reactance Test Procedure


K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC

Page 4 of 38

2.2.

Leakage Reactance Test Screen

The Nameplate Information, and the Administration, Test Conditions, and Help command buttons
are the same as the existing Clipboard.
Purpose: To test the M4000 Leakage Reactance Nameplate Information.
Procedure:
1. Test all fields shown below.
Expected Results:
The M4000 Leakage Reactance test fields on the test screen contains the user input fields: Serial
Numbers (A), (B), (C ) . It contains a Nameplate Data command button, and a Test
Configurations choice box. (Location, Manufacturer, Date, Time, Administration, Test Conditions
are explained in the M4000 test procedure).

2.2.1. Test Screen Inputs ( Below is a list of the inputs unique to Leakage Reactance
only.)
Nameplate Data
Verify that the command button brings up the Nameplate data window

Leakage Reactance Test Procedure


K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC

Page 5 of 38

Nameplate Data window


2.2.1.1.1.
2.2.1.1.1.1.
2.2.1.1.1.2.

Verify Cancel option works by not saving, when values are given
Verify OK option works and saves the data
Verify Help option provides the needed information

Leakage Reactance Test Procedure


K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC

Page 6 of 38

Single Phase Transformer Test


2.2.1.1.2.
2.2.1.1.3.
2.2.1.1.4.

Verify Cancel option works by not saving, when values are given
Verify OK option works and saves the data
Verify Help option provides the needed information

Leakage Reactance Test Procedure


K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC

Page 7 of 38

3 Phase Equivalent test


2.2.1.1.5.
2.2.1.1.6.
2.2.1.1.7.

Verify Cancel option works by not saving, when values are given
Verify OK option works and saves the data
Verify Help option provides the needed information

Leakage Reactance Test Procedure


K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC

Page 8 of 38

3 Phase per Phase - Delta and 3 Phase per Phase Wye


2.2.1.1.8.
2.2.1.1.9.
2.2.1.1.10.

Verify Cancel option works by not saving, when values are given
Verify OK option works and saves the data
Verify Help option provides the needed information

Leakage Reactance Test Procedure


K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC

Page 9 of 38

Adminstrative Information

2.2.1.1.11.
2.2.1.1.12.
2.2.1.1.13.

Verify Cancel option works by not saving, when values are given
Verify OK option works and saves the data
Verify Help option provides the needed information

Test Conditions

Leakage Reactance Test Procedure


K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 10 of

2.2.1.1.14.
Verify Cancel option works by not saving, when values are given
2.2.1.1.15.
Verify OK option works and saves the data
2.2.1.1.16.
Verify Help option provides the needed information
2.2.1.1.17.
Verify that Convert function converts the Air Temp from
Centigrade to Fahrenite
2.2.1.1.18.
Verify that Convert function converts the Oil Temp from
Centigrade to Fahrenite

Leakage Reactance Test Procedure


K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 11 of

Serial Numbers
2.2.1.1.19.
Single Phase
Description
Transformer Serial Numbers
Field Length
30
Data Type:
Alphanumeric
Source:
Data Entry
Comments:
The S/N field has a command button next to it so you can run 3 different tests (A,
B, or C) on 3 different transformers (3 different S/Ns). Clicking on the button
brings up the test sheet for each transformer (A, B, or C).
2.2.1.1.20.
Three Phase (Equivalent, Delta, and Wye)
Description
Transformer Serial Number
Field Length
30
Data Type:
Alphanumeric
Source:
Data Entry
Configuration Menu
2.2.1.1.21.
Test Configuration
Description
Type of test configuration
Data Type:
Alphanumeric
Source:
Dropdown Menu
Comments:
The Dropdown menu consists of 1-3 choices depending on the Winding
Configuration selected. These choices are Single Phase, 3-Phase Equivalent, Per
Phase - Delta, and Per Phase - Wye.
2.2.1.1.22.
Winding Configuration
Description
Type of Winding configuration
Data Type:
Alphanumeric
Source:
Dropdown Menu
Comments:
The Dropdown menu consists of 7 choices: Single Phase, Delta-Wye, Delta-Delta,
Wye-Delta, Wye-Wye, Delta-Zig\zag, Wye-Zig\zag, and Other.
Note: When you make one of these selections the Test Configuration menu
automatically defaults to a preset list. For Example: If you select Delta-Wye
the Test Config. Menu will default to 3 - Phase Equivalent and the dropdown menu
will have Per - Phase Delta and Per - Phase Wye under it.

Leakage Reactance Test Procedure


K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 12 of

2.3.

Leakage Reactance Testing

2.3.1. Single Phase Transformer Test


Run a Test
2.3.1.1.1.
Phase A
Purpose: Verify that you can run a Single Phase Transformer Test on Phase A.
Procedure:
1. Select Single Phase from the Winding Configuration menu.
2. Push the Nameplate Data command Button.
3. Checkpoint: At this point the Nameplate Data window pops up.
4. Enter
Base VoltAmps (A,B,C)
= 10 MVA
BaseVolts
(A,B,C)
= 75 kV
Nameplate % Impedance (A,B,C)
=2
Benchmark % Impedance (A,B,C)
=2
Benchmark % Reactance (A,B,C)
=2
5.
6.
7.
8.
9.

Press OK
Click on PHASE A to select the column.
Checkpoint: The border of the PHASE A cell is bolded.
Press F2 to run a test.
Checkpoint: The Clipboard Power Factor Test window is displayed. It contains the message
Make sure test parameters are OK before you continue. For the Leakage Reactance Test
Mode, Safety Switches and the Strobe light are not used. To begin testing, select continue
then, adjust the external voltage to the desired test voltage. It has 2 command buttons,
Continue and Abort.
10. Press Continue.
11. Checkpoint: The test screen appears with a message in approximately the middle of the
screen, it reads Preparing to test, then immediately following this another message appears
Adjust Variac until CURRENT measures between 400 m Amps and 100 Micro Amps., and
finally the last message is Press F3 to collect data.
12. Expected Result: The window closes and the test runs
2.3.1.1.2.
Voltage Measurements
Purpose: Verify that the measured voltage value given by the M4000 is correct.
Procedure: Record the value of the volt meter and compare that with the voltage value which
appears on the screen
13. Expected Results: Both values should be equal
2.3.1.1.3.
(F3)Record Results, (F5)Accept Results
Purpose: Verify that pressing F3 records the correct results and pressing F5 Accepts the results
and brings you out to the clipboard screen with the results correctly displayed.
Procedure:
1. Press (F3)-collect data.
2. Checkpoint: A message appears in the middle of the screen Dont change voltage-Collecting
Data, and across the Volts bar it says Ranging.
3. When the Test Done message appears note the values below.
4. Press (F5)-Accept results.
5. Checkpoint: The clipboard screen appears.
6. Mark the %Impedance and % Reactance from the test results below. These will be used for
calculations done later.
Leakage Reactance Test Procedure
K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 13 of

Expected Results: The values you noted from the test are the same as the values that are on the
clipboard, and the Inductance, Reactance, and Impedance values are within 1% of the benchmark
data listed. Resistance should be less than .250
Phase A
Benchmark
Current
Loss
Volts

Resistance
Inductance
Reactance
Impedance
% Impedance
% Reactance

N/A
-64 mH
750
742.7
N/A
4.26

Phase B
Repeat steps from Phase A test Highlighting the Phase B column cell when appropriate.
Phase B
Benchmark
Current
Loss
Volts

Resistance
Inductance
Reactance
Impedance
% Impedance
% Reactance

1632.8
-1.0 H
N/A
1675.9
N/A
67.15

Phase C
Repeat steps from Phase A test Highlighting the Phase C column cell when appropriate.
Phase C
Benchmark
Current
Loss
Volts

Resistance
Inductance
Reactance
Impedance
% Impedance
% Reactance

742.0
-66 mH
N/A
742.4
N/A
4.44

Print Results
Purpose: Obtain a hard copy of the test results.
Procedure:
1. Hit the Print icon.
2. Select all options and press ok.
3. Press ok for generic printer.
Expected Result: A complete hard copy is printed.

Leakage Reactance Test Procedure


K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 14 of

Save Results
Purpose: Save Results to have an electronic copy.
Procedure:
1. Select File - Save As.
2. Enter A:\SP(Rev #).clp
Expected Result: File is saved to disc.
2.3.2. Three Phase Equivalent Test
Run a test
Purpose: Verify that you can run a Three Phase Equivalent Test.
Procedure:
1. Select Delta-Wye from the Winding Configuration dropdown menu.
2. Select 3-Phase Equivalent from the Test Configuration dropdown menu.
3. Push the Nameplate Data command Button.
4. Checkpoint: At this point the Nameplate Data window pops up.
5. Enter
Base VoltAmps
= 25 MVA
BaseVolts
= 138 kV
Nameplate % Impedance
=2
Benchmark % Impedance
=2
Benchmark % Reactance
=2
5.
6.
7.
8.
9.

Press OK
Click on PHASE A to select the column.
Checkpoint: The border of the PHASE A cell is bolded.
Press F2 to run a test.
Checkpoint: The Clipboard Power Factor Test window is displayed. It contains the message
Make sure test parameters are OK before you continue. For the Leakage Reactance Test
Mode, Safety Switches and the Strobe light are not used. To begin testing, select continue
then, adjust the external voltage to the desired test voltage. It has 2 command buttons,
Continue and Abort.mps
10. Press Continue.
11. Checkpoint: The test screen appears with a message in approximately the middle of the
screen, it reads Preparing to test, then immediately following this another message appears
Adjust Variac until CURRENT measures between 200 mA and 400 m.A., and finally the last
message is Press F3 to collect data.
12. Expected Result: The window closes and the test runs
2.3.2.1.1.
Voltage Measurements
Purpose: Verify that the measured voltage value given by the M4000 is correct.
Procedure: Record the value of the volt meter and compare that with the voltage value which
appears on the screen
Expected Results: Both values should be equal
(F3)Record Results, (F5)Accept Results
Purpose: Verify that pressing F3 records the correct results and pressing F5 Accepts the results
and brings you out to the clipboard screen with the results correctly displayed.
Procedure:
1. Press (F3)-collect data.
Leakage Reactance Test Procedure
K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 15 of

2. Checkpoint: A message appears in the middle of the screen Dont change voltage-Collecting
Data, and across the Volts bar it says Ranging.
3. When the Test Done message appears note the values below.
4. Press (F5)-Accept results.
5. Checkpoint: The clipboard screen appears.
Expected Results: The values you noted from the test are the same as the values that are on the
clipboard, and the Inductance, Reactance, and Impedance values are within 1 % of the benchmark
data listed. Resistance should be less than .250
Phase A
Benchmark
Current
Loss
Volts

Resistance
Inductance
Reactance
Impedance
Avg. % Impedance
Avg. % Reactance

0
79.5 mH
30
30
2.0
2.0

Phase B
Repeat steps from Phase A test Highlighting the Phase B column cell when appropriate.
Phase B
Benchmark
Current
Loss
Volts

Resistance
Inductance
Reactance
Impedance

0
79.5 mH
30
30

Phase C
Repeat steps from Phase A test Highlighting the Phase C column cell when appropriate.
Phase C
Benchmark
Current
Loss
Volts

Resistance
Inductance
Reactance
Impedance

0
79.5 mH
30
30

Print Results
Purpose: Obtain a hard copy of the test results.
Procedure:
1. Hit the Print icon.
2. Select all options and press ok.
3. Press ok for generic printer.
Expected Result: A complete hard copy is printed.
Leakage Reactance Test Procedure
K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 16 of

Save Results
Purpose: Save Results to have an electronic copy.
Procedure:
1. Select File - Save As.
2. Enter A:\3E(Rev #).clp
Expected Result: File is saved to disc.
2.3.3. Three Phase per Phase - Delta
Run a test
Purpose: Verify that you can run a Three Phase Per Phase-Delta Test.
Procedure:
1. Select Delta-Delta from Winding Configuration dropdown menu.
2. Select Per Phase - Delta from the Test Configuration menu.
3. Push the Nameplate Data command Button.
4. Checkpoint: At this point the Nameplate Data window pops up.
5. Enter
Base VoltAmps
= 10 MVA
BaseVolts
= 69 kV
Nameplate % Impedance
=2
Benchmark % Impedance (A-B,B-C,C-A)
=2
Benchmark % Reactance (A-B,B-C,C-A)
=2
5.
6.
7.
8.
9.

Press OK
Click on PHASE A to select the column.
Checkpoint: The border of the PHASE A cell is bolded.
Press F2 to run a test.
Checkpoint: The Clipboard Power Factor Test window is displayed. It contains the message
Make sure test parameters are OK before you continue. For the Leakage Reactance Test
Mode, Safety Switches and the Strobe light are not used. To begin testing, select continue
then, adjust the external voltage to the desired test voltage. It has 2 command buttons,
Continue and Abort.
10. Press Continue.
11. Checkpoint: The test screen appears with a message in approximately the middle of the
screen, it reads Preparing to test, then immediately following this another message appears
Adjust Variac until CURRENT measures between 1 and 10 amps., and finally the last
message is Press F3 to collect data.
Expected Result: The window closes and the test runs.
Voltage Measurements
Purpose: Verify that the measured voltage value given by the M4000 is correct.
Procedure:
(F3)Record Results, (F5)Accept Results
Purpose: Verify that pressing F3 records the correct results and pressing F5 Accepts the results
and brings you out to the clipboard screen with the results correctly displayed.
Procedure:
1. Press (F3)-collect data.
2. Checkpoint: A message appears in the middle of the screen Dont change voltage-Collecting
Data, and across the Volts bar it says Ranging.
3. When the Test Done message appears note the values below.
4. Press (F5)-Accept results.
Leakage Reactance Test Procedure
K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 17 of

5. Checkpoint: The clipboard screen appears.


Expected Results: The values you noted from the test are the same as the values that are on the
clipboard, and the Inductance, Reactance, and Impedance values are within 1% of the benchmark
data listed. The resistance should be less than .250
Phase A
Benchmark
Current
Loss
Volts

Resistance
Inductance
Reactance
Impedance
% Impedance
% Reactance

0
79.6 mH
30
30
2.0
2.0

Phase B
Repeat steps from Phase A test Highlighting the Phase B column cell when appropriate.
Phase B
Benchmark
Current
Loss
Volts

Resistance
Inductance
Reactance
Impedance
% Impedance
% Reactance

0
79.6 mH
30
30
2.0
2.0

Phase C
Repeat steps from Phase A test Highlighting the Phase C column cell when appropriate.
Phase C
Benchmark
Current
Loss
Volts

Resistance
Inductance
Reactance
Impedance
% Impedance
% Reactance

0
79.6 mH
30
30
2.0
2.0

Print Results
Purpose: Obtain a hard copy of the test results.
Procedure:
1. Hit the Print icon.
2. Select all options and press ok.
3. Press ok for generic printer.
Expected Result: A complete hard copy is printed.

Leakage Reactance Test Procedure


K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 18 of

Save Results
Purpose: Save Results to have an electronic copy.
Procedure:
1. Select File - Save As.
2. Enter A:\3D(Rev #).clp
Expected Result: File is saved to disc.
2.3.4. Three Phase per - Wye
Run a test
Purpose: Verify that you can run a Three Phase Per Phase-Wye Test.
Procedure:
1. Select Wye-Wye from the Winding Configuration dropdown menu.
2. Select Per-Phase Wye from the Test Configuration menu.
3. Push the Nameplate Data command Button.
4. Checkpoint: At this point the Nameplate Data window pops up.
5. Enter
Base VoltAmps
= 25 MVA
Base Volts
= 138 kV
Nameplate % Impedance
=2
Benchmark % Impedance (A-B,B-C,C-A)
=2
Benchmark % Reactance (A-B,B-C,C-A)
=2
5.
6.
7.
8.
9.

Press OK
Click on PHASE A to select the column.
Checkpoint: The border of the PHASE A cell is bolded.
Press F2 to run a test.
Checkpoint: The Clipboard Power Factor Test window is displayed. It contains the message
Make sure test parameters are OK before you continue. For the Leakage Reactance Test
Mode, Safety Switches and the Strobe light are not used. To begin testing, select continue
then, adjust the external voltage to the desired test voltage. It has 2 command buttons,
Continue and Abort.
10. Press Continue.
11. Checkpoint: The test screen appears with a message in approximately the middle of the
screen, it reads Preparing to test, then immediately following this another message appears
Adjust Variac until CURRENT measures between 1 and 10 amps., and finally the last
message is Press F3 to collect data.
Expected Result: The window closes and the test runs.
Voltage Measurements
Purpose: Verify that the measured voltage value given by the M4000 is correct.
(F3)Record Results, (F5)Accept Results
Purpose: Verify that pressing F3 records the correct results and pressing F5 Accepts the results
and brings you out to the clipboard screen with the results correctly displayed.
Procedure:
1. Press (F3)-collect data.
2. Checkpoint: A message appears in the middle of the screen Dont change voltage-Collecting
Data, and across the Volts bar it says Ranging.
3. When the Test Done message appears note the values below.
4. Press (F5)-Accept results.
5. Checkpoint: The clipboard screen appears.
Leakage Reactance Test Procedure
K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 19 of

Expected Results: The values you noted from the test are the same as the values that are on the
clipboard, and the Resistance, Inductance, Reactance, and Impedance values are within 1% of the
benchmark data listed.
Phase A
Benchmark
Current
Loss
Volts

Resistance
Inductance
Reactance
Impedance
% Impedance
% Reactance

21.2
56 mH
21.2
30
2.0
2.0

Phase B
Repeat steps from Phase A test Highlighting the Phase B column cell when appropriate.
Phase B
Benchmark
Current
Loss
Volts

Resistance
Inductance
Reactance
Impedance
% Impedance
% Reactance

21.2
56mH
21.2
30
2.0
2.0

Phase C
Repeat steps from Phase A test Highlighting the Phase C column cell when appropriate.
Phase C
Benchmark
Current
Loss
Volts

Resistance
Inductance
Reactance
Impedance
% Impedance
% Reactance

21.2
56mH
21.2
30
2.0
2.0

Print Results
Purpose: Obtain a hard copy of the test results.
Procedure:
1. Hit the Print icon.
2. Select all options and press ok.
3. Press ok for generic printer.
Expected Result: A complete hard copy is printed.
Save Results
Purpose: Save Results to have an electronic copy.
Leakage Reactance Test Procedure
K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 20 of

Procedure:
1. Select File - Save As.
2. Enter A:\3Y(Rev #).clp
Expected Result: File is saved to disc.
2.3.5. Test Excitation Levels
5 < Itest < 50 ARMS
2.3.5.1.1.
Single Phase
2.3.5.1.1.1.
Test Excitation Level message
Purpose: Verify that the proper Test Excitation Level message is displayed to the user while
running the Single Phase Test.
Procedure:
1. Pick a value between 0.01 and 0.999 for Z(exp) [which is in Ohms]
2. Pick values in random for %Z which is the same as:
Nameplate % Impedence (A,B,C), Benchmark % Impedence(A,B,C), and Benchmark %
Reactance(A,B,C)
3. Pick a value for Basevolts [In KV]
4. Calculate S The Base Volt Amps (A,B,C) [in MVA], using the following calculation:
Z(exp) = 10(%Z)(V.V(KV))/S(KVA)
5. Use 0.1 Ohm Shunt as Specimen
6. Connect DVM across Chunt
7. Raise Variac until DVM across Shunt measures between 1 and 2 volts
8. Press F2 to run the test.
9. Checkpoint: The test screen pops up.
10. Press Continue
Expected Result: The Single Phase test starts and after initial test preparations the message
Adjust the Variac until current measures between 2 and 20 ARMS
11. Press F3 to measure
12. Verify that when the proper test level is attained, the Test Excitation Level message disappears and
you can measure data.

2.3.5.1.2.
3 Phase Equivalent
2.3.5.1.2.1. Test Excitation Level message
Purpose: Verify that the proper Test Excitation Level message is displayed to the user while
running the Single Phase Test.
Procedure:
1 Pick a value between 0.01 and 0.999 for Z(exp) [which is in Ohms]
2. Pick values in random for %Z which is the same as:

Nameplate % Impedence (A,B,C), Benchmark % Impedence(A,B,C), and Benchmark %


Reactance(A,B,C)
3.

Pick a value for Basevolts [In KV]

4.
Calculate S The Base Volt Amps (A,B,C) [in MVA], using the following calculation:
Z(exp) = 20(%Z)(V.V(KV))/S(KVA)

Leakage Reactance Test Procedure


K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 21 of

5.

Use 0.1 Ohm Shunt as Specimen

6.

Connect DVM across Chunt

7.

Raise Variac until DVM across Shunt measures between 1 and 2 volts

8.

Press F2 to run the test.

9.

Checkpoint: The test screen pops up.

10.
Press Continue
Expected Result: The Single Phase test starts and after initial test preparations the message
Adjust the Variac until current measures between 2 and 20 ARMS
11.

Press F3 to measure

12.

Verify that when the proper test level is attained, the Test Excitation Level message
disappears and you can measure data.

1.3.5.1.3 3 Phase per phase Delta


1.3.5.1.3.1 Test Excitation Level message
Purpose: Verify that the proper Test Excitation Level message is displayed to the user while
running the Single Phase Test.
Procedure:
1. Pick a value between 0.01 and 0.999 for Z(exp) [which is in Ohms]
2. Pick values in random for %Z which is the same as:
Nameplate % Impedence (A,B,C), Benchmark % Impedence(A,B,C), and Benchmark %
Reactance(A,B,C)
3. Pick a value for Basevolts [In KV]
4. Calculate S The Base Volt Amps (A,B,C) [in MVA], using the following calculation:
Z(exp) = 30(%Z)(V.V(KV))/S(KVA)
5. Use 0.1 Ohm Shunt as Specimen
6. Connect DVM across Chunt
7. Raise Variac until DVM across Shunt measures between 1 and 2 volts
8. Press F2 to run the test.
9. Checkpoint: The test screen pops up
10. Press Continue
11. Expected Result: The Single Phase test starts and after initial test preparations the message
12. Adjust the Variac until current measures 2 and 20 ARMS
13. Press F3 to measure
14. Verify that when the proper test level is attained, the Test Excitation Level message disappears and
you can measure data.

1.3.5.1.4 3 Phase per phase - WYE


1.3.5.1.4.1 Test Excitation Level message
Purpose: Verify that the proper Test Excitation Level message is displayed to the user while
running the Single Phase Test.
Procedure:

Leakage Reactance Test Procedure


K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 22 of

1. Pick a value between 0.01 and 0.999 for Z(exp) [which is in Ohms]
2. Pick values in random for %Z which is the same as:
Nameplate % Impedence (A,B,C), Benchmark % Impedence(A,B,C), and Benchmark %
Reactance(A,B,C)
3. Pick a value for Basevolts [In KV]
4. Calculate S The Base Volt Amps (A,B,C) [in MVA], using the following calculation:
Z(exp) = 10(%Z)(V.V(KV))/S(KVA)
5. Use 0.1 Ohm Shunt as Specimen
6. Connect DVM across Chunt
7. Raise Variac until DVM across Shunt measures between 1 and 2 volts
8. Press F2 to run the test.
9. Checkpoint: The test screen pops up
10. Press Continue
11. Expected Result: The Single Phase test starts and after initial test preparations the message
Adjust the Variac until current measures between 2 and 20 ARMS
12. Press F3 to measure
13. Verify that when the proper test level is attained, the Test Excitation Level message disappears and
you can measure data.

2.3.5.2
2 < Itest < 20 ARMS
2.3.5.2.1
Single Phase
2.3.5.2.1.1 Test Excitation Level
Purpose: Verify that the proper Test Excitation Level message is displayed to the user while
running the Single Phase Test.
Procedure:
1. Pick a value between 1.00 and 9.99 for Z(exp) [which is in Ohms]
2. Pick values in random for %Z which is the same as:
Nameplate % Impedence (A,B,C), Benchmark % Impedence(A,B,C), and Benchmark %
Reactance(A,B,C)
3. Pick a value for Basevolts [In KV]
4. Calculate S The Base Volt Amps (A,B,C) [in MVA], using the following calculation:
Z(exp) = 10(%Z)(V.V(KV))/S(KVA)
5. Use 0.1 Ohm Shunt as Specimen
6. Connect DVM across Chunt
7. Raise Variac until DVM across Shunt measures between 1 and 2 volts
8. Press F2 to run the test.
9. Checkpoint: The test screen pops up
10. Press Continue
11. Expected Result: The Single Phase test starts and after initial test preparations the message
Adjust the Variac until current measures 2 and 20 ARMS
12. Press F3 to measure
13. Verify that when the proper test level is attained, the Test Excitation Level message disappears and
you can measure data.

2.3.5.2.2
3 Phase Equivalent
Purpose: Verify that the proper Test Excitation Level message is displayed to the user while
running the Single Phase Test.
Leakage Reactance Test Procedure
K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 23 of

Procedure:
1. Pick a value between 1.00 and 9.99 for Z(exp) [which is in Ohms]
2. Pick values in random for %Z which is the same as:
Nameplate % Impedence (A,B,C), Benchmark % Impedence(A,B,C), and Benchmark %
Reactance(A,B,C)
3. Pick a value for Basevolts [In KV]
4. Calculate S The Base Volt Amps (A,B,C) [in MVA], using the following calculation:
Z(exp) = 20(%Z)(V.V(KV))/S(KVA)
5. Use 0.1 Ohm Shunt as Specimen
6. Connect DVM across Chunt
7. Raise Variac until DVM across Shunt measures between 1 and 2 volts
8. Press F2 to run the test.
9. Checkpoint: The test screen pops up
10. Press Continue
11. Expected Result: The Single Phase test starts and after initial test preparations the message
Adjust the Variac until current measures 2 and 20 ARMS
12. Press F3 to measure
13. Verify that when the proper test level is attained, the Test Excitation Level message disappears and
you can measure data.

2.3.5.2.3
3 Phase per phase Delta
2.3.5.2.3.1 Test Excitation Level message
Purpose: Verify that the proper Test Excitation Level message is displayed to the user while
running the Single Phase Test.
Procedure:
1. Pick a value between 1.00 and 9.99 for Z(exp) [which is in Ohms]
2. Pick values in random for %Z which is the same as: Nameplate % Impedence (A,B,C),
Benchmark % Impedence(A,B,C), and Benchmark % Reactance(A,B,C)
3. Pick a value for Basevolts [In KV]
4. Calculate S The Base Volt Amps (A,B,C) [in MVA], using the following calculation:
Z(exp) = 30(%Z)(V.V(KV))/S(KVA)
5. Use 0.1 Ohm Shunt as Specimen
6. Connect DVM across Chunt
7. Raise Variac until DVM across Shunt measures between 1 and 2 volts
8. Press F2 to run the test.
9. Checkpoint: The test screen pops up
10. Press Continue
11. Expected Result: The Single Phase test starts and after initial test preparations the message
Adjust the Variac until current measures 2 and 20 ARMS
12. Press F3 to measure
13. Verify that when the proper test level is attained, the Test Excitation Level message disappears and
you can measure data.

2.3.5.2.4

3 Phase per phase - WYE

2.3.5.2.4.1 Test Excitation Level message


Purpose: Verify that the proper Test Excitation Level message is displayed to the user while
running the Single Phase Test.
Leakage Reactance Test Procedure
K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 24 of

Procedure:
1.

Pick a value between 1.00 and 9.99 for Z(exp) [which is in Ohms]

2.

Pick values in random for %Z which is the same as: Nameplate % Impedence (A,B,C),
Benchmark % Impedence(A,B,C), and Benchmark % Reactance(A,B,C)

3.

Pick a value for Basevolts [In KV]

4.
Calculate S The Base Volt Amps (A,B,C) [in MVA], using the following calculation:
Z(exp) = 30(%Z)(V.V(KV))/S(KVA)
5.

Use 0.1 Ohm Shunt as Specimen

6.

Connect DVM across Chunt

7.

Raise Variac until DVM across Shunt measures between 1 and 2 volts

8.

Press F2 to run the test.

9.

Checkpoint: The test screen pops up

14. Press Continue


15. Expected Result: The Single Phase test starts and after initial test preparations the message
Adjust the Variac until current measures 2 and 20 ARMS
16. Press F3 to measure
17. Verify that when the proper test level is attained, the Test Excitation Level message disappears
and you can measure data.

2.3.5.3 1 < Itest < 10 ARMS


2.3.5.3.1
Single Phase
2.3.5.3.1.1 Test Excitation Level message
Purpose: Verify that the proper Test Excitation Level message is displayed to the user while
running the Single Phase Test.
Procedure:
1.

Pick a value between 10.0 and 39.9 for Z(exp) [which is in Ohms]

2.

Pick values in random for %Z which is the same as: Nameplate % Impedence
(A,B,C), Benchmark % Impedence(A,B,C), and Benchmark % Reactance(A,B,C)

3.

Pick a value for Basevolts [In KV]

4.

Calculate S The Base Volt Amps (A,B,C) [in MVA], using the following
calculation:
Z(exp) = 10(%Z)(V.V(KV))/S(KVA)

Leakage Reactance Test Procedure


K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 25 of

5.

Use 0.1 Ohm Shunt as Specimen

6.

Connect DVM across Chunt

7.

Raise Variac until DVM across Shunt measures between 1 and 2 volts

8.

Press F2 to run the test.

9.

Checkpoint: The test screen pops up

18. Press Continue


19. Expected Result: The Single Phase test starts and after initial test preparations the
message
Adjust the Variac until current measures 1 and 10 ARMS
20. Press F3 to measure
21. Verify that when the proper test level is attained, the Test Excitation Level message

disappears and you can measure data.

2.3.5.3.2
3 Phase Equivalent
Change to the 3 Phase Equivalent configuration and repeat test 2.3.5.3.1
2.3.5.3.2.1 Test Excitation Level message
Purpose: Verify that the proper Test Excitation Level message is displayed to the user while
running the Single Phase Test.
Procedure:
1.

Pick a value between 10.0 and 39.9 for Z(exp) [which is in Ohms]

2.

Pick values in random for %Z which is the same as: Nameplate % Impedence (A,B,C),
Benchmark % Impedence(A,B,C), and Benchmark % Reactance(A,B,C)

3.

Pick a value for Basevolts [In KV]

4.
Calculate S The Base Volt Amps (A,B,C) [in MVA], using the following calculation:
Z(exp) = 20(%Z)(V.V(KV))/S(KVA)

Leakage Reactance Test Procedure


K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 26 of

5.

Use 0.1 Ohm Shunt as Specimen

6.

Connect DVM across Chunt

7.

Raise Variac until DVM across Shunt measures between 1 and 2 volts

8.

Press F2 to run the test.

9.

Checkpoint: The test screen pops up

22. Press Continue


23. Expected Result: The Single Phase test starts and after initial test preparations the message
Adjust the Variac until current 1 and 10 ARMS
24. Press F3 to measure
25. Verify that when the proper test level is attained, the Test Excitation Level message disappears
and you can measure data.

2.3.5.3.3
3 Phase per phase - Delta
2.3.5.3.3.1 Test Excitation Level message
Purpose: Verify that the proper Test Excitation Level message is displayed to the user while
running the Single Phase Test.
Procedure:
1.

Pick a value between 10.0 and 39.9 for Z(exp) [which is in Ohms]

2.

Pick values in random for %Z which is the same as: Nameplate % Impedence (A,B,C),
Benchmark % Impedence(A,B,C), and Benchmark % Reactance(A,B,C)

3.

Pick a value for Basevolts [In KV]

4.

Calculate S The Base Volt Amps (A,B,C) [in MVA], using the following calculation:

Z(exp) = 30(%Z)(V.V(KV))/S(KVA)Use 0.1 Ohm Shunt as Specimen

5. Use 0.1 Ohm Shunt as Specimen

Leakage Reactance Test Procedure


K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 27 of

6. Connect DVM across Chunt


7. Raise Variac until DVM across Shunt measures between 1 and 2 volts
8.

Press F2 to run the test.

9. Checkpoint: The test screen pops up


10. Press Continue
11. Expected Result: The Single Phase test starts and after initial test preparations the message
Adjust the Variac until current measures 1 and 10 ARMS
12. Press F3 to measure
13. Verify that when the proper test level is attained, the Test Excitation Level message disappears
and you can measure data.

2.3.5.3.4
3 Phase per phase - WYE
2.3.5.3.4.1 Test Excitation Level message
Purpose: Verify that the proper Test Excitation Level message is displayed to the user while
running the Single Phase Test.
Procedure:
1.

Pick a value between 10.0 and 39.9 for Z(exp) [which is in Ohms]

2.

Pick values in random for %Z which is the same as: Nameplate % Impedence (A,B,C),
Benchmark % Impedence(A,B,C), and Benchmark % Reactance(A,B,C)

3.

Pick a value for Basevolts [In KV]

4.

Calculate S The Base Volt Amps (A,B,C) [in MVA], using the following calculation:

Z(exp) = 10(%Z)(V.V(KV))/S(KVA)
5. Use 0.1 Ohm Shunt as Specimen

6. Connect DVM across Chunt


7. Raise Variac until DVM across Shunt measures between 1 and 2 volts
8.

Press F2 to run the test.

9. Checkpoint: The test screen pops up


10. Press Continue
11. Expected Result: The Single Phase test starts and after initial test preparations the message
Adjust the Variac until current measures 1 and 10 ARMS
12. Press F3 to measure
13. Verify that when the proper test level is attained, the Test Excitation Level message disappears
and you can measure data.

Leakage Reactance Test Procedure


K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 28 of

2.3.5.4
.5 < Itest < 4 ARMS
2.3.5.4.1
Single Phase
2.3.5.4.1.1 Test Excitation Level message
Purpose: Verify that the proper Test Excitation Level message is displayed to the user while
running the Single Phase Test.
Procedure:
1. Pick a value between 40.0 and 99.9 for Z(exp) [which is in Ohms]
2. Pick values in random for %Z which is the same as: Nameplate % Impedence (A,B,C),
Benchmark % Impedence(A,B,C), and Benchmark % Reactance(A,B,C)
3. Pick a value for Basevolts [In KV]
4. Calculate S The Base Volt Amps (A,B,C) [in MVA], using the following calculation:
Z(exp) = 10(%Z)(V.V(KV))/S(KVA)
5. Use 0.1 Ohm Shunt as Specimen
6. Connect DVM across Chunt
7. Raise Variac until DVM across Shunt measures between 1 and 2 volts
8. Press F2 to run the test.
9. Checkpoint: The test screen pops up
10. Press Continue
11. Expected Result: The Single Phase test starts and after initial test preparations the message

Adjust the Variac until current measures .5 and 4 ARMS


12. Press F3 to measure
13. Verify that when the proper test level is attained, the Test Excitation Level message disappears and
you can measure data.

2.3.5.4.2
3 Phase Equivalent
2.3.5.4.2.1 Test Excitation Level message
Procedure:
1.

Pick a value between 40.0 and 99.9 for Z(exp) [which is in Ohms]

2.

Pick values in random for %Z which is the same as: Nameplate % Impedence (A,B,C),
Benchmark % Impedence(A,B,C), and Benchmark % Reactance(A,B,C)

3.

Pick a value for Basevolts [In KV]

4.
Calculate S The Base Volt Amps (A,B,C) [in MVA], using the following calculation:
Z(exp) = 10(%Z)(V.V(KV))/S(KVA)

Leakage Reactance Test Procedure


K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 29 of

5.

Use 0.1 Ohm Shunt as Specimen

6.

Connect DVM across Chunt

7.

Raise Variac until DVM across Shunt measures between 1 and 2 volts

8.

Press F2 to run the test.

26. Checkpoint: The test screen pops up


27. Press Continue
28. Expected Result: The Single Phase test starts and after initial test preparations the message
Adjust the Variac until current measures .5 and 4 ARMS
29. Press F3 to measure
30. Verify that when the proper test level is attained, the Test Excitation Level message disappears
and you can measure data.
2.3.5.4.3

3 Phase per phase - Delta

2.3.5.4.3.1

Test Excitation Level message

Purpose: Verify that the proper Test Excitation Level message is displayed to the user while
running the Single Phase Test.

Procedure:
1.
2.

Pick a value between 40.0 and 99.9 for Z(exp) [which is in Ohms]
Pick values in random for %Z which is the same as: Nameplate % Impedence (A,B,C), Benchmark %
Impedence(A,B,C), and Benchmark % Reactance(A,B,C)
3. Pick a value for Basevolts [In KV]
4. Calculate S The Base Volt Amps (A,B,C) [in MVA], using the following calculation:
Z(exp) = 30(%Z)(V.V(KV))/S(KVA)
5. Use 0.1 Ohm Shunt as Specimen
6. Connect DVM across Chunt
7. Raise Variac until DVM across Shunt measures between 1 and 2 volts
8. Press F2 to run the test.
9. Checkpoint: The test screen pops up
10. Press Continue
11. Expected Result: The Single Phase test starts and after initial test preparations the message

Adjust the Variac until current measures .5 and 4 ARMS


12. Press F3 to measure
13. Verify that when the proper test level is attained, the Test Excitation Level message disappears and
you can measure data.

2.3.5.4.4
3 Phase per phase - WYE
2.3.5.4.4.1 Test Excitation Level message
Purpose: Verify that the proper Test Excitation Level message is displayed to the user while
running the Single Phase Test.

Procedure:

Leakage Reactance Test Procedure


K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 30 of

1.

Pick a value between 40.0 and 99.9 for Z(exp) [which is in Ohms]

2.

Pick values in random for %Z which is the same as: Nameplate % Impedence (A,B,C),
Benchmark % Impedence(A,B,C), and Benchmark % Reactance(A,B,C)

3.

Pick a value for Basevolts [In KV]

4.

Calculate S The Base Volt Amps (A,B,C) [in MVA], using the following
calculation:Z(exp) = 10(%Z)(V.V(KV))/S(KVA)

5.

Use 0.1 Ohm Shunt as Specimen

6.

Connect DVM across Chunt

7.

Raise Variac until DVM across Shunt measures between 1 and 2 volts

8.

Press F2 to run the test.

31. Checkpoint: The test screen pops up


32. Press Continue
33. Expected Result: The Single Phase test starts and after initial test preparations the message
Adjust the Variac until current measures .5 and 4 ARMS
34. Press F3 to measure
35. Verify that when the proper test level is attained, the Test Excitation Level message disappears
and you can measure data.

2.3.5.5 .5 < Itest < 400/Zexp ARMS


2.3.5.5.1
Single Phase
2.3.5.5.1.1 Test Excitation Level message
Purpose: Verify that the proper Test Excitation Level message is displayed to the user while
running the Single Phase Test.
Procedure:

Leakage Reactance Test Procedure


K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 31 of

1. Pick a value between 100 and 750 for Z(exp) [which is in Ohms]
2. Pick values in random for %Z which is the same as: Nameplate % Impedence (A,B,C),
Benchmark % Impedence(A,B,C), and Benchmark % Reactance(A,B,C)
3. Pick a value for Basevolts [In KV]
4. Calculate S The Base Volt Amps (A,B,C) [in MVA], using the following calculation:Z(exp) =
10(%Z)(V.V(KV))/S(KVA)
5. Use 0.1 Ohm Shunt as Specimen
6. Connect DVM across Chunt
7. Raise Variac until DVM across Shunt measures between 1 and 2 volts
8. Press F2 to run the test.
9. Checkpoint: The test screen pops up
10. Press Continue
11. Expected Result: The Single Phase test starts and after initial test preparations the message
Adjust the Variac until current measures .5 and 400/Z(exp) ARMS
12. Press F3 to measure
13. Verify that when the proper test level is attained, the Test Excitation Level message disappears
and you can measure data.
2.3.5.5.2
3 Phase Equivalent
2.3.5.5.2.1 Test Excitation Level message
Procedure:
1.

Pick a value between 100 and 750 for Z(exp) [which is in Ohms]

2.

Pick values in random for %Z which is the same as: Nameplate % Impedence (A,B,C),
Benchmark % Impedence(A,B,C), and Benchmark % Reactance(A,B,C)

3.

Pick a value for Basevolts [In KV]

4.
Calculate S The Base Volt Amps (A,B,C) [in MVA], using the following calculation:
Z(exp) = 20(%Z)(V.V(KV))/S(KVA)

Leakage Reactance Test Procedure


K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 32 of

5.

Use 0.1 Ohm Shunt as Specimen

6.

Connect DVM across Chunt

7.

Raise Variac until DVM across Shunt measures between 1 and 2 volts

8.

Press F2 to run the test.

36. Checkpoint: The test screen pops up


37. Press Continue
38. Expected Result: The Single Phase test starts and after initial test preparations the message
Adjust the Variac until current measures .5 and 400/Z(exp) ARMS
39. Press F3 to measure
40. Verify that when the proper test level is attained, the Test Excitation Level message disappears
and you can measure data.
2.3.5.5.3
3 Phase per phase - Delta
2.3.5.5.3.1 Test Excitation Level message
Purpose: Verify that the proper Test Excitation Level message is displayed to the user while
running the Single Phase Test.

Procedure:
1. Pick a value between 100 and 750 for Z(exp) [which is in Ohms]
2. Pick values in random for %Z which is the same as: Nameplate % Impedence (A,B,C), Benchmark %
Impedence(A,B,C), and Benchmark % Reactance(A,B,C)

3. Pick a value for Basevolts [In KV]


4. Calculate S The Base Volt Amps (A,B,C) [in MVA], using the following calculation:
Z(exp) = 30(%Z)(V.V(KV))/S(KVA)
5. Use 0.1 Ohm Shunt as Specimen
6. Connect DVM across Chunt
7. Raise Variac until DVM across Shunt measures between 1 and 2 volts
8. Press F2 to run the test.
9. Checkpoint: The test screen pops up
10. Press Continue
11. Expected Result: The Single Phase test starts and after initial test preparations the message
Adjust the Variac until current .5 and 400/Z(exp) ARMS
12. Press F3 to measure
13. Verify that when the proper test level is attained, the Test Excitation Level message disappears and
you can measure data.

2.3.5.5.4
3 Phase per phase - WYE
2.3.5.5.4.1 Test Excitation Level message
Purpose: Verify that the proper Test Excitation Level message is displayed to the user while
running the Single Phase Test.

Procedure:

Leakage Reactance Test Procedure


K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 33 of

1.

Pick a value between 100 and 750 for Z(exp) [which is in Ohms]

2.

Pick values in random for %Z which is the same as: Nameplate % Impedence (A,B,C),
Benchmark % Impedence(A,B,C), and Benchmark % Reactance(A,B,C)

3.

Pick a value for Basevolts [In KV]

4.

Calculate S The Base Volt Amps (A,B,C) [in MVA], using the following
calculation:Z(exp) = 10(%Z)(V.V(KV))/S(KVA)

5.

Use 0.1 Ohm Shunt as Specimen

6.

Connect DVM across Chunt

7.

Raise Variac until DVM across Shunt measures between 1 and 2 volts

8.

Press F2 to run the test.

41. Checkpoint: The test screen pops up


42. Press Continue
43. Expected Result: The Single Phase test starts and after initial test preparations the message
Adjust the Variac until current measures .5 and 400/Z(exp) ARMS
44. Press F3 to measure
45. Verify that when the proper test level is attained, the Test Excitation Level message disappears
and you can measure data.
2.3.5.6 750 Ohms and up
2.3.5.6.1Test Excitation Level message: N/A
2.3.6 Test Screen functions
1.3.6.1 F1-Screen Help
Purpose: To test the F1-Screen Help function key.
Procedure:
Run any test.
While the test is running press F1.
Expected Result: The M4000 for Windows Help is displayed.
1.3.6.2 F4-Stop Test
Purpose: To test the F4-Stop Test function key.
Procedure:
1. Run any test.
2. At the test screen, while the test is running, press F4.
Expected Result: The test stops, voltage, current, and loss bars go down to 0, and the bottom of
the screen shows F1-Screen Help, F2-Restart Test, F6 Discard Results.
1.3.6.2.1 F2-Restart Test
Purpose: To test the F2-Restart Test function key after press F4.
Procedure:
Follow the procedure from 2.3.5.2.
Leakage Reactance Test Procedure
K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 34 of

Hit F2.
Expected Result: The Clipboard Power Factor Test window is displayed, you press continue and
the test starts.
1.3.6.3 ESC-Stop Test
Purpose: To test the ESC-Stop Test key.
Procedure:
1. Run any test.
2. At the test screen, while the test is running, press F4.
Expected Result: The test stops, voltage, current, and loss bars go down to 0, and the bottom of
the screen shows F1-Screen Help, F2-Restart Test, F6 Discard Results.
1.3.6.3.1
F2-Restart Test
Purpose: Verify that you can restart a test after pressing ESC.
Procedure:
1. Follow the procedure from 2.3.5.3.
2. Hit F2.
Expected Result: The Clipboard Power Factor Test window is displayed, you press continue and
the test starts.
1.3.6.4 F6-Discard Results
Purpose: To test the F6-Discard Results function key.
Procedure:
1. Follow the procedure from 2.3.5.3.
2. Hit F6.
Expected Result: The message This will discard the results of the test just run. Do you wish to
do this?
1.3.6.5 F8 - Save Test Results to File
Expected Results:
Pressing F8 after Tests have completed successfully, Saves the Test Results to a file. See
Save As in the Menu Items and Toolbar section.
1.3.7 Calibration Settings test
Purpose: Verify the calibration settings while the L.R.M. test is running.
Procedure:
1. Connect the external testing cable from a separate controller to the M4000 board connection.
2. Start the controller and go into DOS.
3. Change directory to c:\dta .
4. At the c:\dta prompt type m4ktests/2m.
5. Checkpoint: The m4ktest runs and a list of choices appears on the screen.
6. Enter 6) Request Calibration Status.
Expected Result: The calibration results are:
Enable Guard Board Calibration =1 (Enabled)
Enable M4000 Reference Calibration =1 (Enabled)
Enable M4000 System (Factory) Calibration =1 (Enabled)
Enable M4000 System (Field) Calibration =1 (Disabled)
1.4
Results
All values in the Test Results grid have been verified except % Impedance and % Reactance.
Leakage Reactance Test Procedure
K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 35 of

The calculations involved in determining these values are complicated, therefore we have utilized
Excel to perform them.
1.4.1
% Impedance and % Reactance
1.4.1.1 Single Phase Transformer Test
Purpose: Verify that the values on the results screen are correct.
Procedure:
1. Open the excel worksheet R:\m4000\leakreac\data\calcs\template\spcalc.xlt.
2. Save as R:\m4000\leakreac\data\calcs\(rev #)\spcalc.xls
3. Enter the Nameplate data.
4. Enter the Measured Inductance from the test results.
5. Enter the Impedance from the test results.
Expected Result: The values for % Impedance and % Reactance as listed on the Single Phase
Transformer line of the worksheet are the same as the values on the Test Results grid.
1.4.1.2 Three Phase Equivalent
Purpose: Verify that the values on the results screen are correct.
Procedure:
1. Open the excel worksheet R:\m4000\leakreac\data\calcs\template\3ecalc.xlt.
2. Save as R:\m4000\leakreac\data\calcs\(rev #)\3ecalc.xls
3. Enter the Nameplate data.
4. Enter the Measured Inductance from the test results.
5. Enter the Impedance from the test results.
Expected Result: The values for % Impedance and % Reactance as listed on the 3 Phase
Equivalent line of the worksheet are the same as the values on the Test Results grid.
1.4.1.3 Three Phase per Phase - Wye
Purpose: Verify that the values on the results screen are correct.
Procedure:
1. Open the excel worksheet R:\m4000\leakreac\data\calcs\template\3ycalc.xlt.
2. Save as R:\m4000\leakreac\data\calcs\(rev #)\3ycalc.xls
3. Enter the Nameplate data.
4. Enter the Measured Inductance from the test results.
5. Enter the Impedance from the test results.
Expected Result: The values for % Impedance and % Reactance as listed on the 3 Phase - Wye
line of the worksheet are the same as the values on the Test Results grid.
1.4.1.4 Three Phase per Phase - Delta
Purpose: Verify that the values on the results screen are correct.
Procedure:
1. Open the excel worksheet R:\m4000\leakreac\data\calcs\template\3dcalc.xlt.
2. Save as R:\m4000\leakreac\data\calcs\(rev #)\3dcalc.xls
3. Enter the Nameplate data.
4. Enter the Measured Inductance from the test results.
5. Enter the Impedance from the test results.
Expected Result: The values for % Impedance and % Reactance as listed on the 3 Phase - Delta
line of the worksheet are the same as the values on the Test Results grid.
1.5 Save
1.5.1 File- Save
Leakage Reactance Test Procedure
K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 36 of

Purpose: Verify that you can save the clipboard data to a file by using File-Save from the menu.
Procedure:
1. Run a Three Phase Equivalent test.
2. Print the test results including the detailed option.
3. When the testing is complete select File-Save.
4. When the save screen pops up save the file to R:\m4000\leakreac\data\(rev #)\filesave.clp
5. When function is complete exit out of clipboard.
6. Re-enter clipboard.
7. Press File-Open.
8. Open R:\m4000\leakreac\data\(rev #)\filesave.clp
Expected Result: The saved file is retrieved and the values are correct.

1.5.2
Save Icon
Purpose: Verify that you can save the clipboard data to a file by pressing the save icon on the
toolbar.
Procedure:
1. Run a UST Reversal test.
2. Print the test results including the detailed option.
3. When the testing is complete press the Save icon.
4. When the save screen pops up save the file to R:\m4000\leakreac\data\(rev #)\iconsave.clp
5. When function is complete exit out of clipboard.
6. Re-enter clipboard.
7. Press File-Open.
8. Open to R:\m4000\leakreac\data\(rev #)\iconsave.clp
Expected Result: The saved file is retrieved and the values are correct.
1.6 Print Functions
1.6.1 File- Print
Purpose: Verify you can print the Clipboard from the File -Print menu options.
Procedure:
1. At the clipboard screen select File -Print from the menu.
2. Checkpoint: At this point the Clipboard Print Options screen pops. ( The function of the check
boxes is described in the M4000 Test Procedure.
3. Select the options you wish to print.
4. Press the Print command button.
Expected Result: The Clipboard screen prints out, with only the options you selected.
1.6.2 F7 function key
Purpose: Verify you can print the Clipboard by pressing the F7 function key.
Procedure:
1. At the clipboard press F7
2. Checkpoint: At this point the Clipboard Print Options screen pops. ( The function of the check
boxes is described in the M4000 Test Procedure.
3. Select the options you wish to print.
4. Press the Print command button.
Expected Result: The Clipboard screen prints out, with only the options you selected.
1.6.3

Printer Icon

Leakage Reactance Test Procedure


K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 37 of

Purpose: Verify you can print the Clipboard by pressing the


icon.
Procedure:
1. At the clipboard printer icon.
2. Checkpoint: At this point the Clipboard Print Options screen pops. ( The function of the check
boxes is described in the M4000 Test Procedure.
3. Select the options you wish to print.
4. Press the Print command button.
Expected Result: The Clipboard screen prints out, with only the options you selected.
1.7 Print Verification
Purpose: Verify the following values on the printout are the same as the values on the clipboard
screen.
Procedure:
1. Open file R:\m4000\leakreac\testplan\printing.clp
2. Print a full report.
3. Check each of the following fields on the printout against the corresponding field on the
clipboard screen.
Expected Result: The values are the same as the values on the clipboard screen.
1.8 Notes
1.8.1 Note Icon
Purpose: Verify that the notepad icon brings up the Leakage Reactance Test Note screen.
Procedure:
Hit the Notepad icon
.
Expected Result: The Leakage Reactance Test Note screen appears.
1.8.2 Enter and Save
Purpose: Verify that you can enter and save notes.
Procedure:
Type in a note in the note screen and hit OK.
Checkpoint: The note screen closes.
Hit the note icon again.
Expected Result: The Leakage Reactance Test Note screen appears with the previously entered
note.
1.8.3 Save to Leakreac\Data\(rev #)
Purpose: If all the .CLP files saved in this test procedure were saved locally, to put them on \\QALab1\1g_arche
Procedure:
Using a file management system put all .CLP files saved in this test procedure to
\\QA-Lab1\1g_arche\m4000\leakreac\data\(rev #).
Expected Result: All files are now on the net under the correct Qa directory.

Leakage Reactance Test Procedure


K:\QA\M4000\LEAKREAC\TESTPLAN\LREACTP.DOC
38

Page 38 of

Anda mungkin juga menyukai