DEWE-CAL
Technical Reference Manual
Page 1
Page 2
Page 3
Page 4
Page 5
Page 6
WARNING Calls attention to a procedure, practice, or condition that could cause bodily injury
or death.
CAUTION
Page 7
Ground the equipment: For Safety Class 1 equipment (equipment having a protective earth terminal), a
non interruptible safety earth ground must be provided from the mains power source to the product input
wiring terminals or supplied power cable.
DO NOT operate the product in an explosive atmosphere or in the presence of flammable gases or fumes.
DO NOT operate damaged equipment: Whenever it is possible that the safety protection features built into
this product have been impaired, either through physical damage, excessive moisture, or any other reason,
REMOVE POWER and do not use the product until safe operation can be verified by service-trained
personnel. If necessary, return the product to a DEWETRON sales and service office for service and repair
to ensure that safety features are maintained.
Keep away from live circuits: Operating personnel must not remove equipment covers or shields.
Procedures involving the removal of covers or shields are for use by service-trained personnel only. Under
certain conditions, dangerous voltages may exist even with the equipment switched off. To avoid dangerous
electrical shock, DO NOT perform procedures involving cover or shield removal unless you are qualified to
do so.
No modifications are allowed at the instrument. The fuse in the power module has to be replaced by the
same type. For continued protection against fire, replace the line fuse(s) only with fuse(s) of the same
voltage and current rating and type. DO NOT use repaired fuses or short-circuited fuse holder labels and
print on the power module may not be removed.
DO NOT service or adjust alone. Do not attempt internal service or adjustment unless another person,
capable of rendering first aid and resuscitation, is present.
DO NOT substitute parts or modify equipment: Because of the danger of introducing additional hazards, do
not install substitute parts or perform any unauthorized modification to the product. Return the product to a
DEWETRON sales and service office for service and repair to ensure that safety features are maintained.
Before opening the instrument (experts only) or exchanging the fuse in the power module disconnect power!
For voltages above 60 V you are not allowed to use the metal housing of DSUB connectors!
You may not connect higher voltages than 250 V to the DSUB connectors of the DEWETRON modules.
The power-cable and -connector serve as Power-Breaker. The cable must not exceed 10 feet, disconnect
function must be possible without tools.
Safety of the operator and the unit depend on following these rules.
Page 8
Basics
o
All DAQx modules are within specifications when leaving the factory. The time interval for recalibration
depends on environmental conditions. Typically, the calibration should be checked once a year and has to be
done by qualified persons only.
It is recommended that the devices, you use for calibration, are better than 0.01% accuracy in the ranges you
want to calibrate.
Recommended:
o Multi Product Calibrator:
Fluke 5500A
Fluke 5520A
o Multimeter:
Keithley 2000
Keithley 2001
Agilent 34401A
Agilent 3458A
(used by DEWETRON)
(used by DEWETRON)
(used by DEWETRON)
(used by DEWETRON)
Use The DEWETRON Module-Manual for additional information.(How to setup Filters, Gain, Ranges,..)
Page 9
Part Number
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
Contents
Part
Safety Banana Plug Cords red
Safety Banana Plug Cords black
Supply cable
Bridge input cable
Thermo couple cable type T (blue connector)
Thermo couple cable type K (yellow connector)
Thermo couple cable type J (black connector)
BNC to Banana cable
DAQx-V input cable
RTD input cable
PAD input cable
T-adapter
Calibration capacitor 1.0 nF
15 Hz High pass filter
Working part (SUBD9-> SUBD9)
Direct in module
pieces
2
2
1
1
1
1
1
2
1
1
1
2
1
1
2
1
Page 10
DAQxDMM Calibration
4 DAQx-DMM Module
4.1
Specifications
4.1.1
DC-Accuracy
DAQx-DMM
Module
Range
10V
40V
100V
200V
400V
1000V
4.1.2
4.1.2.1
Accuracy
Max absolute
scaled error
% of Value % of Range
0.1% 0.1%
0.1% 0.1%
0.1% 0.1%
0.1% 0.1%
0.1% 0.1%
0.1% 0.1%
V
0.02
0.08
0.2
0.4
0.8
2
Max. Error
mV
10
10
10
10
10
10
dB Definition
Output
Output _ Range
Input
Input _ Range
Table
Max. Offset
mV
5
5
5
5
5
5
Gain
Output
= 20 log10
= 20 log10
dB
Input DC _ Gain
4.1.2.2
5V Module Output
Limits at fo/2
Lower
Upper
-0.8dB
0.5dB
Limits at f0
Lower
-4.5dB
Upper
-1.5dB
Filter Graph
Page 11
DAQxDMM Calibration
4.2
Schematic
DAQx-DMM
GAIN
SW1
ZERO
LEDS
SW2
Supply Cable
GTU
DEWE-DAQP-DMM
t/in
Working Part
Si
S p e c ia l O u
gn
al
Su
Ou
pp
b le
a b le
Ca
DAQ P-D M M
tC
ly
DAQ P-DM M
Direct
Page 12
DAQxDMM Calibration
4.3
Calibration Rack
DAQ P-F REQ DAQ P-DM M
DAQP-DM M
B A LA N C E
OUTPUT
Digital MULTIMETER
Page 13
DAQxDMM Calibration
4.4
Parts List
DAQx-DMM Module
Slot: Any Kind of DEWE-Module-Slot is possible.
Multi Product Calibrator
High accuracy Digital Multimeter
Cables
o Safety Banana Plug Cords (red ,black)
o Supply Cable
To DEW E RACK 1 6
o
o
o
o
o
Su b-D9
M ale
Sub-D9
Fema le
9
8
7
4
3
2
1
Signal
O utput
o
o
7
6
3
2
1
Pin assignment
1 Signal Output +/-5V
2 RS-485 A
3.RS-485 B
4 GND
5 +9V Power Supply
6+1 2V Power Supply
7 Module Input/output
8 Reserved
9 -9V Power Supply
Special
OUT/IN
9 Pin SubD Female 9 Pin SubD Female 1:1. (This is a Working Part and should be changed after
100 calibrations, because oft the limited connector lifetime.)
Isolated screwdriver for adjusting the Module
Page 14
DAQxDMM Calibration
4.5
Calibration Instruction
4.5.1
o
o
DC Calibration
Warm up the module (plug it for 30 minutes into a DEWE-Module Slot )
Connect a Calibrator (for example a Fluke 5500A) to the module.
o The Calibrator Source has to be connected to the input connector of the module. (Banana Plugs)
o The Module output has to be connected to a high accuracy digital Multimeter. Use the supply cable or
the BNC output of a Cal-Rack.
o (DC accuracy better than 0,01% @ 5V Range );
Apply the voltages according to the following table, and check if the module output voltage is between the
limits.
Warning : Dangerous High Voltage!
Step
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
Module Setting
Range
Filter
1000V
20 kHz
1000V
20 kHz
1000V
20 kHz
400V
20 kHz
400V
20 kHz
400V
20 kHz
200V
20 kHz
200V
20 kHz
200V
20 kHz
100V
20 kHz
100V
20 kHz
100V
20 kHz
40V
20 kHz
40V
20 kHz
40V
20 kHz
10V
20 kHz
10V
20 kHz
10V
20 kHz
Calibrator Output
VDC
+1000.0V
-1000.0V
0.0000V
+400.0V
-400.0V
0.0000V
+200.0V
-200.0V
0.0000V
+100.0V
-100.0V
0.0000V
+40.000V
-40.000V
0.0000V
+10.000V
-10.000V
0.0000V
Limits
lower
4.9900 V
-5.0100 V
-0.0050V
4.9900 V
-5.0100 V
-0.0050V
4.9900 V
-5.0100 V
-0.0050V
4.9900 V
-5.0100 V
-0.0050V
4.9900 V
-5.0100 V
-0.0050V
4.9900 V
-5.0100 V
-0.0050V
upper
5.0100 V
-4.9900 V
+0.0050V
5.0100 V
-4.9900 V
+0.0050V
5.0100 V
-4.9900 V
+0.0050V
5.0100 V
-4.9900 V
+0.0050V
5.0100 V
-4.9900 V
+0.0050V
5.0100 V
-4.9900 V
+0.0050V
Page 15
DAQxDMM Calibration
4.5.1
Filter Check
o
o
o
o
o
o
Steps
1
2
3
4
5
6
7
8
9
10
4.6
Input Signal
Voltage
Frequency
7 VRMS
5Hz
7 VRMS
10Hz
7 VRMS
50Hz
7 VRMS
100Hz
7 VRMS
500Hz
7 VRMS
1kHz
7 VRMS
1.5kHz
7 VRMS
3kHz
7 VRMS
10kHz
7 VRMS
20kHz
Output Limits
lower
upper
3.19 VRMS
3.70 VRMS
2.08 VRMS
2.94 VRMS
3.19 VRMS
3.70 VRMS
2.08 VRMS
2.94 VRMS
3.19 VRMS
3.70 VRMS
2.08 VRMS
2.94 VRMS
3.19 VRMS
3.70 VRMS
2.08 VRMS
2.94 VRMS
3.19 VRMS
3.70 VRMS
2.08 VRMS
2.94 VRMS
Module Adjustment
This steps are only necessary if one of the Calibration checks fails, or is only marginal.
4.6.1
Offset adjustment
Open the module cover, and connect it to the digital Multimeter and the slot, using the
supply cable.
o Set the input range to 10V and apply 0.0000V to the module. (applying a
short circuit to the input is also possible). If the signal output is between
1.5 mV it is okay. If not, open the Module, and trim the ZERO
potentiometer, until the Output is between 0.2mV and +0.2mV .
o Set the Module input range to 1000V. If the signal output is between
1.5mV continue with point 4.6.2 , if not (open the module) trim the ZERO
potentiometer, until the output is between 1.5mV, and restart with first step.
4.6.2
Gain adjustment
o
o
o
GAIN
+
+ ZERO
- GAIN
- +
+ ZERO
Set the input range to 10V and apply +10.0000V to the module. The
measured value should be between 4.9970V an 5.0030V. Otherwise
open the module and trim the Gain potentiometer, until the output is
between 4.998V and 5.002V .
Apply 10.0000V to the module.The output voltage should between
-4.9975V and -5.0025V. If not open the module and trim the Gain potentiometer, until the
Output is between -4.998V and -5.002V.
Remount the module cover and proceed with the DC Calibration step.
Page 16
DAQxDMM Calibration
4.7
Calibration Report
Module :
Serial Number:
Customer:
Inspection Date:
DAQP-DMM
Used Standards
Model
Temperature:
Humidity:
Tested by:
Place of inspection:
S/N
Calibration Date
Calibrator
Digital Multimeter
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
1
2
3
4
5
6
7
8
9
10
Calibration Procedure
DC Calibration
Module Setting
Calibrator
Range
Filter
1000V
20 kHz
+1000.0V
1000V
20 kHz
-1000.0V
1000V
20 kHz
0.0000V
400V
20 kHz
+400.0V
400V
20 kHz
-400.0V
400V
20 kHz
0.0000V
200V
20 kHz
+200.0V
200V
20 kHz
-200.0V
200V
20 kHz
0.0000V
100V
20 kHz
+100.0V
100V
20 kHz
-100.0V
100V
20 kHz
0.0000V
40V
20 kHz
+40.000V
40V
20 kHz
-40.000V
40V
20 kHz
0.0000V
10V
20 kHz
+10.000V
10V
20 kHz
-10.000V
10V
20 kHz
0.0000V
Filter Calibration
Module Setting
Calibrator
Range
Filter
10Hz
10V
5Hz
10Hz
10V
10Hz
100Hz
10V
50Hz
100Hz
10V
100Hz
1kHz
10V
500Hz
1kHz
10V
1kHz
3kHz
10V
1.5kHz
3kHz
10V
3kHz
20kHz
10V
10kHz
20kHz
10V
20kHz
Limits
lower
upper
4.9900 V
5.0100 V
-5.0100 V
-4.9900 V
-0.0050V
+0.0050V
4.9900 V
5.0100 V
-5.0100 V
-4.9900 V
-0.0050V
+0.0050V
4.9900 V
5.0100 V
-5.0100 V
-4.9900 V
-0.0050V
+0.0050V
4.9900 V
5.0100 V
-5.0100 V
-4.9900 V
-0.0050V
+0.0050V
4.9900 V
5.0100 V
-5.0100 V
-4.9900 V
-0.0050V
+0.0050V
4.9900 V
5.0100 V
-5.0100 V
-4.9900 V
-0.0050V
+0.0050V
Value
[V]
okay
Limits
lower
upper
3.19 VRMS
3.70 VRMS
2.08 VRMS
2.94 VRMS
3.19 VRMS
3.70 VRMS
2.08 VRMS
2.94 VRMS
3.19 VRMS
3.70 VRMS
2.08 VRMS
2.94 VRMS
3.19 VRMS
3.70 VRMS
2.08 VRMS
2.94 VRMS
3.19 VRMS
3.70 VRMS
2.08 VRMS
2.94 VRMS
Value
[VRMS]
okay
Page 17
DAQP-HV Calibration
5 DAQP-HV Module
5.1
Specifications
5.1.1
DC-Accuracy
DAQP-HV
Range
Accuracy
Max absolute
Error
20V
50V
100V
200V
400V
800V
1400V
% of Value Offset
0.05% 40mV
0.05% 40mV
0.05% 50mV
0.05% 100mV
0.05% 200mV
0.05% 400mV
0.05% 700mV
mV
50
65
100
200
400
800
1400
5.1.2
5.1.2.1
Max. Error
mV
15
6.5
5
5
5
5
5
Definition
Output
Output _ Range
Input
Input _ Range
Table
Bessel
Butterworth
5.1.2.3
Max. Offset
mV
12.5
4
2.5
2.5
2.5
2.5
2.5
Gain
Output
= 20 log10
= 20 log10
dB
Input DC _ Gain
5.1.2.2
5V Module output
Limits at fo/2
Lower
Upper
-1.2dB
-0.3dB
-0.8dB
0.2dB
Limits at f0
Lower
Upper
-4.5dB
-1.5dB
-4.5dB
-1.5dB
Limits at 10*f0
Lower
Upper
-38.7 dB -32 dB
-43 dB
-39.8dB
Diagram
Page 18
DAQP-HV Calibration
5.1.3
CAL-Rack Connection
Calibration Rack
Digital MULTIMETER
DEWETRON DEWETRON DEWETRON
OUTPUT
5.2
Parts List
o
o
o
o
o
o
DAQP-HV Module
Multi Product Calibrator
DEWESoft 6 for module setup
High accuracy digital Multimeter.
Cables
o Safety banana plug test leads (Cal-Kit Contents 3.2)
Calibration Rack
Page 19
DAQP-HV Calibration
5.3
Calibration Instruction
The DAQP-HV Module is a high voltage input module. Turing calibration voltages up to 1000V are applied to the
module. For safety reasons it is not allowed to use this module without a suitable DEWETRON system.
Also do not use the supply cable!
o
o
5.3.1
o
o
o
o
DC Accuracy Calibration
Start DEWESoft 6 for module setup
Set the module to:
Range:
Filter:
Filter Type:
20V
1kHz
Bessel
DAQP-HV DC accuracy
Step
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
Module Setting
Range
Filter
20 V
1kHz
20 V
1kHz
20 V
1kHz
50 V
1kHz
50 V
1kHz
50 V
1kHz
100 V
1kHz
100 V
1kHz
100 V
1kHz
200 V
1kHz
200 V
1kHz
200 V
1kHz
400 V
1kHz
400 V
1kHz
400 V
1kHz
800 V
1kHz
800 V
1kHz
800 V
1kHz
1400 V
1kHz
1400 V
1kHz
1400 V
1kHz
5V Module Output
Calibrator
Specification
Limits
V
mV
Lower [V]
Upper [V]
0.000
12.5
-0.0125
0.0125
20.000
15.0
4.9850
5.0150
-20.000
15.0
-5.0150
-4.9850
0.000
4.0
-0.0040
0.0040
50.000
6.5
4.9935
5.0065
-50.000
6.5
-5.0065
-4.9935
0.00
2.5
-0.0025
0.0025
100.00
5.0
4.9950
5.0050
-100.00
5.0
-5.0050
-4.9950
0.00
2.5
-0.0025
0.0025
200.00
5.0
4.9950
5.0050
-200.00
5.0
-5.0050
-4.9950
0.00
2.5
-0.0025
0.0025
400.00
5.0
4.9950
5.0050
-400.00
5.0
-5.0050
-4.9950
0.00
2.5
-0.0025
0.0025
800.00
5.0
4.9950
5.0050
-800.00
5.0
-5.0050
-4.9950
0.00
2.5
-0.0025
0.0025
1000.00
4.3
3.5671
3.5757
-1000.00
4.3
-3.5757
-3.5671
Page 20
DAQP-HV Calibration
5.3.2
o
5.3.3
Steps
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
26*
27*
28*
*
Filter Check
Set the module to:
Range:
Filter:
Filter Type:
20V
10 Hz
Bessel
Voltage:
14VRMS
Frequency:
10Hz
Wave form:
Sine wave
Offset:
0V
Apply the voltages according to the following table and
measure the module output RMS voltage.
DAQP-HV Filter
Input Signal
Voltage
Frequency
VRMS
Hz
14
10
14
100
14
15
14
30
14
300
14
50
14
100
14
1k
14
1k
14
150
14
300
14
3k
14
500
14
1k
14
10k
14
10k
14
1k5
14
3k
14
30k
14
5k
14
10k
14
100k
14
15k
14
30k
14
50k
14
100k
3.2
180k
Output Limits
Lower
Upper
dB
VRMS
dB
VRMS
-4.5
2.085
-1.5
-38.7
0.041
-32
-1.2
3.048
-0.3
-4.5
2.085
-1.5
-38.7
0.041
-32
-1.2
3.048
-0.3
-4.5
2.085
-1.5
-38.7
0.041
-32
-43
0.025 -36.8
-1.2
3.048
-0.3
-4.5
2.085
-1.5
-38.7
0.041
-32
-1.2
3.048
-0.3
-4.5
2.085
-1.5
-38.7
0.041
-32
-43
0.025 -36.8
-1.2
3.048
-0.3
-4.5
2.085
-1.5
-38.7
0.041
-32
-1.2
3.048
-0.3
-4.5
2.085
-1.5
-38.7
0.041
-32
-1.2
3.048
-0.3
-4.5
2.085
-1.5
-1.2
3.048
-0.3
-4.5
2.085
-1.5
-4.5
0.477
-1.5
2.945
0.088
3.381
2.945
0.088
3.381
2.945
0.088
0.051
3.381
2.945
0.088
3.381
2.945
0.088
0.051
3.381
2.945
0.088
3.381
2.945
0.088
3.381
2.945
3.381
2.945
0.673
The Input voltage is only 3.2 VRMS because the 5500A to not support higher voltages over 100kHz.
These are special tests to verify the filter characteristic.
Page 21
DAQP-HV Calibration
5.4
DAQP-HV
Temperature:
Humidity:
Tested by:
Used Standards
Model
S/N
Calibration Date
Calibrator
Digital Multimeter
Calibration Procedure
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
1
2
4
5
7
8
9
10
11
13
14
17
18
20
21
23
24
26
27
28
Test description
DC Calibration
Module Range
Calibrator
20 V
0.000
20 V
20.000
20 V
-20.000
50 V
0.000
50 V
50.000
50 V
-50.000
100 V
0.00
100 V
100.00
100 V
-100.00
200 V
0.00
200 V
200.00
200 V
-200.00
400 V
0.00
400 V
400.00
400 V
-400.00
800 V
0.00
800 V
800.00
800 V
-800.00
1400 V
0.00
1400 V
1000.00
1400 V
-1000.00
Filter
10 Hz Bessel
30Hz Bessel
10Hz
15Hz
30Hz
100Hz Bessel
50Hz
100Hz
100Hz Bessel
1kHz
100Hz Butterworth 1kHz
300Hz Bessel
150Hz
300 HZ
1kHz Bessel
500 Hz
1k
3kHz Bessel
1k5
3k
10kHz Bessel
5k
10 k
30kHz Bessel
15 k
30 k
100kHz Bessel
50 k
100 k
180kHz Bessel
180 k
Limits
lower
V
-0.0125
4.9850
-5.0150
-0.0040
4.9935
-5.0065
-0.0025
4.9950
-5.0050
-0.0025
4.9950
-5.0050
-0.0025
4.9950
-5.0050
-0.0025
4.9950
-5.0050
-0.0025
3.5671
-3.5757
upper
V
0.0125
5.0150
-4.9850
0.0040
5.0065
-4.9935
0.0025
5.0050
-4.9950
0.0025
5.0050
-4.9950
0.0025
5.0050
-4.9950
0.0025
5.0050
-4.9950
0.0025
3.5757
-3.5671
Value
2.085
3.048
2.085
3.048
2.085
0.041
0.025
3.048
2.085
3.048
2.085
3.048
2.085
3.048
2.085
3.048
2.085
3.048
2.085
0.477
2.945
3.381
2.945
3.381
2.945
0.088
0.051
3.381
2.945
3.381
2.945
3.381
2.945
3.381
2.945
3.381
2.945
3.381
2.945
0.673
Pass
V
Page 22
DAQx-V Calibration
6 DAQx-V Module
6.1
Specifications
6.1.1
DC-Accuracy
DAQx-V
Module Range
Accuracy
Max absolute
Error
10mV
100mV
1V
5V
10V
50V
% of reading Offset
0.05% of reading 40V
0.05% of reading 100V
0.05%of reading 500V
0.05%of reading 2.5mV
0.05%of reading 5mV
0.05%of reading 25mV
mV
0.045
0.15
1
5
10
50
6.1.2
6.1.2.1
Gain
Output
= 20 log10
= 20 log10
dB
Input DC _ Gain
6.1.2.2
Output
Output _ Range
Input
Input _ Range
Table
5V Module output
Max. Error
Max. Offset
mV
mV
20
22.5
5
7.5
2.5
5
2.5
5
2.5
5
2.5
5
Limits at fo/2
Lower
Upper
-0.8dB
0.5dB
Limits at f0
Lower
Upper
-4.5dB
-1.5dB
Diagram
Page 23
DAQx-V Calibration
6.2
Schematic
ZERO
GAIN
DAQx-V
SW1
LEDS
INPUT OFFSET
SW2
Supply Cable
DEWE-DAQP-V-DSUB
t/in
Working Part
Si
S p e c ia l O u
gn
al
Su
Ou
pp
b le
a b le
Ca
DAQP-DM M
tC
ly
Digital MULTIMETER
DAQP-DM M
Direct
Page 24
DAQx-V Calibration
6.3
Parts List
o
o
o
o
o
6.4
DAQx-V Module
Slot: Any Kind of DEWE-Module-Slot is possible.
Multi Product Calibrator
High accuracy digital Multimeter.
Cables
o DAQP-V-B: Safety Banana Plug Cords (red, black).
o DAQP-V-BNC: Safety Banana Plugs (red, black), to BNC.
o DAQP-V-DSUB: Safety Banana Plugs (red, black), to SUB-D 9.
o Supply Cable (see 4.4)
o 9 Pin SUB-D 9 Female, to 9 Pin SUBD Female 1:1. (This is a working part and should be changed
after 100 calibrations, because oft the limited connector lifetime.)
o Isolated screwdriver for adjusting the module.
Calibration Instruction
4.5.2
o
o
DC Calibration
Warm up the module (plug it for 30 minutes into a DEWE-Module Slot )
Connect a Calibrator (for example a Fluke 5500A) to the module.
o The Calibrator Source has to be connected to the input connector of the module. (Banana Plugs)
o The Module output has to be connected to a high accuracy digital Multimeter. Use the supply cable or
the BNC output of a Cal-Rack.
o (DC accuracy better than 0,01% @ 5V Range );
Apply the voltages according to the following table, and check if the module output voltage is between the
limits.
Warning: Dangerous High Voltage!
Step
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
Module Setting
Range
Filter
50V
50 kHz
50V
50 kHz
50V
50 kHz
10V
50 kHz
10V
50 kHz
10V
50 kHz
5V
50 kHz
5V
50 kHz
5V
50 kHz
1V
50 kHz
1V
50 kHz
1V
50 kHz
0.1V
50 kHz
0.1V
50 kHz
0.1V
50 kHz
0.01V
50 kHz
0.01V
50 kHz
0.01V
50 kHz
Calibrator Output
VDC
+50.000V
-50.000V
0.0000V
+10.000V
-10.000V
0.0000V
+5.0000V
-5.0000V
0.0000V
+1.0000V
-1.0000V
0.0000V
+0.1000V
-0.1000V
0.0000V
+0.0100V
-0.0100V
0.0000V
Limits
lower
4.9950 V
-5.0050 V
-0.0025V
4.9950 V
-5.0050 V
-0.0025V
4.9950 V
-5.0050 V
-0.0025V
4.9950 V
-5.0050 V
-0.0025V
4.9925 V
-5.0075 V
-0.0050V
4.9775 V
-5.0225 V
-0.0200V
upper
5.0050 V
-4.9950 V
+0.0025V
5.0050 V
-4.9950 V
+0.0025V
5.0050 V
-4.9950 V
+0.0025V
5.0050 V
-4.9950 V
+0.0025V
5.0075 V
-4.9925 V
+0.0050V
5.0225 V
-4.9775 V
+0.0200V
Page 25
DAQx-V Calibration
6.4.1
Filter Check
o
o
o
o
o
o
Steps
1
2
3
4
5
6
7
8
9
10
6.5
Input Signal
Voltage
Frequency
3.5 VRMS
5Hz
3.5 VRMS
10Hz
3.5 VRMS
50Hz
3.5 VRMS
100Hz
3.5 VRMS
500Hz
3.5 VRMS
1kHz
3.5 VRMS
5kHz
3.5 VRMS
10kHz
3.5 VRMS
25kHz
3.5 VRMS
50kHz
Output Limits
lower
upper
3.19 VRMS
3.70 VRMS
2.08 VRMS
2.94 VRMS
3.19 VRMS
3.70 VRMS
2.08 VRMS
2.94 VRMS
3.19 VRMS
3.70 VRMS
2.08 VRMS
2.94 VRMS
3.19 VRMS
3.70 VRMS
2.08 VRMS
2.94 VRMS
3.19 VRMS
3.70 VRMS
2.08 VRMS
2.94 VRMS
Module Adjustment
5.4.1
GAIN
Offset adjustment
Open the module cover, and connect it to the digital Multimeter and the slot,
using the supply cable.
o Set the Input Range to 1V, the Input Filter to 50kHz, and apply 0.0000V to
the module. If the signal output is between 0.8mV it is okay. If not, trim the
ZERO potentiometer, until the output is between 0.2mV and +0.2mV .
o Set the Module Input Range to 0.01V. If the signal output is between
3mV continue with point 6.5.1 , if not (open the module) trim the ZERO
Potentiometer, until the Output is between +/- 1mV.
o Repeat these two steps as long as both are not in specification.
6.5.1
Gain adjustment
Set the input range to 1V and apply +1.0000V to the module. The
measured Value should be between 4.9980V and 5.0020V. Otherwise
open the module and trim the Gain Potentiometer, until the Output is
between 4.9985Vand 5.0015V.
Remount the module cover and proceed with the DC Calibration step.
ZERO
ZERO
GAIN
ZERO
This steps are only necessary if one of the Calibration checks fails, or is only marginal.
-GAIN
Page 26
DAQx-V Calibration
6.6
Calibration Report
Module :
Serial Number:
Customer:
Inspection Date:
DAQP-V
Temperature:
Humidity:
Tested by:
Place of inspection:
Test equipment
Model
S/N
Calibration Date
Calibrator
Digital Multimeter
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
1
2
3
4
5
6
7
8
9
10
Calibration Procedure
DC Calibration
Module Setting
Calibrator
Range
Filter
50V
50 kHz
+50.000V
50V
50 kHz
-50.000V
50V
50 kHz
0.0000V
10V
50 kHz
+10.000V
10V
50 kHz
-10.000V
10V
50 kHz
0.0000V
5V
50 kHz
+5.0000V
5V
50 kHz
-5.0000V
5V
50 kHz
0.0000V
1V
50 kHz
+1.0000V
1V
50 kHz
-1.0000V
1V
50 kHz
0.0000V
0.1V
50 kHz
+0.1000V
0.1V
50 kHz
-0.1000V
0.1V
50 kHz
0.0000V
0.01V
50 kHz
+0.0100V
0.01V
50 kHz
-0.0100V
0.01V
50 kHz
0.0000V
Filter Calibration
Module Setting
Calibrator
Range
Filter
10Hz
5V
3.5 VRMS
10Hz
5V
3.5 VRMS
100Hz 5V
3.5 VRMS
100Hz 5V
3.5 VRMS
1kHz
5V
3.5 VRMS
1kHz
5V
3.5 VRMS
10kHz
5V
3.5 VRMS
10kHz
5V
3.5 VRMS
50kHz
5V
3.5 VRMS
50kHz
5V
3.5 VRMS
Limits
lower
4.9950 V
-5.0050 V
-0.0025V
4.9950 V
-5.0050 V
-0.0025V
4.9950 V
-5.0050 V
-0.0025V
4.9950 V
-5.0050 V
-0.0025V
4.9925 V
-5.0075 V
-0.0050V
4.9775 V
-5.0225 V
-0.0200V
upper
5.0050 V
-4.9950 V
+0.0025V
5.0050 V
-4.9950 V
+0.0025V
5.0050 V
-4.9950 V
+0.0025V
5.0050 V
-4.9950 V
+0.0025V
5.0075 V
-4.9925 V
+0.0050V
5.0225 V
-4.9775 V
+0.0200V
Limits
lower
5Hz
10Hz
50Hz
100Hz
500Hz
1kHz
5kHz
10kHz
25kHz
50kHz
upper
3.19 VRMS
2.08 VRMS
3.19 VRMS
2.08 VRMS
3.19 VRMS
2.08 VRMS
3.19 VRMS
2.08 VRMS
3.19 VRMS
2.08 VRMS
Value
[V]
Value
[VRMS]
3.70 VRMS
2.94 VRMS
3.70 VRMS
2.94 VRMS
3.70 VRMS
2.94 VRMS
3.70 VRMS
2.94 VRMS
3.70 VRMS
2.94 VRMS
okay
okay
Page 27
Specifications
7.1.1
DC-Accuracy
DAQP-V-A
Module Range
Bipolar
Unipolar
20mV
40mV
100mV
200mV
500mV
1V
2V
5V
10V
25V
50V
+20mV
+40mV
+100mV
+200mV
+500mV
+1V
+2V
+5V
+10V
+25V
+50V
5V Module output
Max absolute
Error
Max. Offset Max. Error
% of Value Offset
mV
mV
mV
0.01% 40V
0.042
10
10.5
0.01% 40V
0.044
5
5.5
0.01% 40V
0.05
2
2.5
0.01% 80V
0.1
2
2.5
0.01% 200V
0.25
2
2.5
0.01% 400V
0.5
2
2.5
0.01% 800V
1
2
2.5
0.01% 2mV
2.5
2
2.5
0.01% 4mV
5
2
2.5
0.01% 10mV
12.5
2
2.5
0.01% 20mV
25
2
2.5
0.02% 40V
0.044
20
22
0.02% 40V
0.048
10
12
0.02% 60V
0.08
6
8
0.02% 120V
0.16
6
8
0.02% 300V
0.4
6
8
0.02% 600V
0.8
6
8
0.02% 1200V
1.6
6
8
0.02% 3mV
4
6
8
0.02% 6mV
8
6
8
0.02% 15mV
20
6
8
0.02% 30mV
40
6
8
Accuracy
Page 28
Gain
Output
= 20 log10
= 20 log10
dB
Input DC _ Gain
7.1.2.2
Table
Bessel
Butterworth
7.1.2.3
Output
Output _ Range
Input
Input _ Range
Limits at fo/2
Lower
Upper
-1.2dB
-0.3dB
-0.8dB
0.2dB
Limits at f0
Lower
Upper
-4.5dB
-1.5dB
-4.5dB
-1.5dB
Limits at 10*f0
Lower
Upper
-38.7 dB -32 dB
-43 dB
-39.8dB
Diagram
Page 29
CAL-Rack Connection
Calibration Rack
DA QP-V-A
DAQP-V-A
DA QP-V-A
Digital MULTIMETER
OUTPUT
7.2
Parts List
o
o
o
o
o
o
7.3
DAQP-V-A Module
Slot : Any Kind of DEWE-Module-Slot is possible.
Multi Product Calibrator
DEWESoft 6 or DEWEConfig 5.4.2.d for Module Setup
High accuracy digital Multimeter.
Cables
o DAQP-V-DSUB :Safety Banana Plugs (red ,black), to SUB-D 9.
o Supply Cable or Calibration Rack
o 9 Pin SUB-D 9 Female, to 9 Pin SubD Female 1:1. (This is a Working Part and should be changed
after 100 calibrations, because oft the limited connector lifetime.)
Calibration Instruction
o
o
Page 30
DC Accuracy Calibration
check
Page 31
Unipolar Range
Calibrator
VDC
0.000 mV
+10.000 mV
+20.000 mV
0.000 mV
+20.000 mV
+40.000 mV
0.000 mV
+50.000 mV
+100.00 mV
0.000 mV
+100.00 mV
+200.00 mV
0.000 mV
+250.00 mV
+500.00 mV
0.000 V
+0.500 V
+1.000 V
0.000 V
+1.000 V
+2.000 V
0.000 V
+2.500 V
+5.000 V
0.000 V
+5.000 V
+10.000 V
0.000 V
+12.500 V
+25.000 V
0.000 V
+25.000 V
+50.000 V
Specification
20.0 mV
21.0 mV
22.0 mV
10.0 mV
11.0 mV
12.0 mV
6.0 mV
7.0 mV
8.0 mV
6.0 mV
7.0 mV
8.0 mV
6.0 mV
7.0 mV
8.0 mV
6.0 mV
7.0 mV
8.0 mV
6.0 mV
7.0 mV
8.0 mV
6.0 mV
7.0 mV
8.0 mV
6.0 mV
7.0 mV
8.0 mV
6.0 mV
7.0 mV
8.0 mV
6.0 mV
7.0 mV
8.0 mV
Limits
lower
upper
-5.0200 V
-4.9800 V
-21.0 mV
21.0 mV
4.9780 V
5.0220 V
-5.0100 V
-4.9900 V
-11.0 mV
11.0 mV
4.9880 V
5.0120 V
-5.060 V
-4.9940 V
-7.0 mV
7.0 mV
4.9960 V
5.0060 V
-5.060 V
-4.9940 V
-7.0 mV
7.0 mV
4.9960 V
5.0060 V
-5.060 V
-4.9940 V
-7.0 mV
7.0 mV
4.9960 V
5.0060 V
-5.060 V
-4.9940 V
-7.0 mV
7.0 mV
4.9960 V
5.0060 V
-5.060 V
-4.9940 V
-7.0 mV
7.0 mV
4.9960 V
5.0060 V
-5.060 V
-4.9940 V
-7.0 mV
7.0 mV
4.9960 V
5.0060 V
-5.060 V
-4.9940 V
-7.0 mV
7.0 mV
4.9960 V
5.0060 V
-5.060 V
-4.9940 V
-7.0 mV
7.0 mV
4.9960 V
5.0060 V
-5.060 V
-4.9940 V
-7.0 mV
7.0 mV
4.9960 V
5.0060 V
Page 32
7.3.3
Steps
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25*
26*
27*
28*
29*
*
Filter Check
Set the module to:
Range:
Filter:
Filter Type:
Shunt Type:
Coupling:
Input Mode:
5V
10 Hz
Bessel
none
DC
Bipolar
Voltage:
3.5VRMS
Frequency:
10Hz
Wave form:
Sine wave
Offset:
0V
Apply the voltages according to the following table and measure the module output RMS voltage.
Input Signal
Voltage
Frequency
VRMS
Hz
3.5
10
3.5
100
3.5
15
3.5
30
3.5
300
3.5
50
3.5
100
3.5
1k
3.5
1k
3.5
150
3.5
300
3.5
3k
3.5
500
3.5
1k
3.5
10k
3.5
10k
3.5
1k5
3.5
3k
3.5
30k
3.5
5k
3.5
10k
3.5
100k
3.5
15k
3.5
30k
3.0
300k
3.5
50k
3.5
100k
3.0
150
3.0
300k
Output Limits
lower
upper
dB
VRMS
dB
VRMS
-4.5
2.085
-1.5
-38.7
0.041
-32
-1.2
3.048
-0.3
-4.5
2.085
-1.5
-38.7
0.041
-32
-1.2
3.048
-0.3
-4.5
2.085
-1.5
-38.7
0.041
-32
-43
0.025 -36.8
-1.2
3.048
-0.3
-4.5
2.085
-1.5
-38.7
0.041
-32
-1.2
3.048
-0.3
-4.5
2.085
-1.5
-38.7
0.041
-32
-43
0.025 -36.8
-1.2
3.048
-0.3
-4.5
2.085
-1.5
-38.7
0.041
-32
-1.2
3.048
-0.3
-4.5
2.085
-1.5
-38.7
0.041
-32
-1.2
3.048
-0.3
-4.5
2.085
-1.5
-38.7
0.035
-32
-1.2
3.048
-0.3
-4.5
2.085
-1.5
-1.2
2.613
-0.3
-4.5
1.787
-1.5
2.945
0.088
3.381
2.945
0.088
3.381
2.945
0.088
0.051
3.381
2.945
0.088
3.381
2.945
0.088
0.051
3.381
2.945
0.088
3.381
2.945
0.088
3.381
2.945
0.075
3.381
2.945
2.898
2.524
The Input voltage is only 3 VRMS because the 5500A to not support higher voltages over 100kHz.
These are special test points to verify the filter characteristic
Page 33
Module :
Serial Number:
Customer:
Inspection Date:
DAQP-V-A
Test equipment
Model
S/N
Temperature:
Humidity:
Tested by:
Place of inspection:
Calibration Date
Calibrator
Digital Multimeter
Test description
7.3.1.1
Bipolar Range
1
0.02V Range
0.000 mV
2
+20.000 mV
3
-20.000 mV
4
0.04V Range
0.000 mV
5
+40.000 mV
6
-40.000 mV
7
0.1V Range
0.000 mV
8
+100.000 mV
9
-100.00 mV
10
0.2V Range
0.000 mV
11
+200.00 mV
12
-200.00 mV
13
0.5V Range
0.000 mV
14
+500.00 mV
15
-500.00 mV
16
1V Range
0.000 V
17
+1.000 V
18
-1.000 V
19
2V Range
0.000 V
20
+2.000 V
21
-2.000 V
22
5V Range
0.000 V
23
+5.000 V
24
-5.000 V
25 10V Range
0.000 V
26
+10.000 V
27
-10.000 V
28 25V Range
0.000 V
29
+25.000 V
30
-25.000 V
31 50V Range
0.000 V
32
+50.000 V
33
-50.000 V
Calibration Procedure
Limits
lower
upper
-10 mV
+10 mV
4.9895 V
5.0105 V
-5.0105 V
-4.9895 V
-5 mV
+5 mV
4.9945 V
5.0055 V
-5.0055 V
-4.9945 V
-2 mV
+2 mV
4.9975 V
5.0025 V
-5.0025 V
-4.9975 V
-2 mV
+2 mV
4.9975 V
5.0025 V
-5.0025 V
-4.9975 V
-2 mV
+2 mV
4.9975 V
5.0025 V
-5.0025 V
-4.9975 V
-2 mV
+2 mV
4.9975 V
5.0025 V
-5.0025 V
-4.9975 V
-2 mV
+2 mV
4.9975 V
5.0025 V
-5.0025 V
-4.9975 V
-2 mV
+2 mV
4.9975 V
5.0025 V
-5.0025 V
-4.9975 V
-2 mV
+2 mV
4.9975 V
5.0025 V
-5.0025 V
-4.9975 V
-2 mV
+2 mV
4.9975 V
5.0025 V
-5.0025 V
-4.9975 V
-2 mV
+2 mV
4.9975 V
5.0025 V
-5.0025 V
-4.9975 V
Value
O.K
Page 34
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
Test description
Limits
Unipolar Range
lower
+0.02V Range
0.000 mV -5.0200 V
+10.000 mV-21.0 mV
+20.000 mV4.9780 V
+0.04V Range
0.000 mV-5.0100 V
+20.000 mV-11.0 mV
+40.000 mV4.9880 V
+0.1V Range
0.000 mV-5.060 V
+50.000 mV-7.0 mV
+100.00 mV4.9960 V
+0.2V Range
0.000 mV-5.060 V
+100.00 mV-7.0 mV
+200.00 mV4.9960 V
+0.5V Range
0.000 mV-5.060 V
+250.00 mV-7.0 mV
+500.00 mV4.9960 V
+1V Range
0.000 V-5.060 V
+0.500 V-7.0 mV
+1.000 V4.9960 V
+2V Range
0.000 V-5.060 V
+1.000 V-7.0 mV
+1.000 V4.9960 V
+5V Range
0.000 V-5.060 V
+2.500 V-7.0 mV
+5.000 V4.9960 V
+10V Range
0.000 V-5.060 V
+5.000 V-7.0 mV
+10.000 V4.9960 V
+25V Range
0.000 V-5.060 V
+12.500 V-7.0 mV
+25.000 V4.9960 V
+50V Range
0.000 V-5.060 V
+25.000 V-7.0 mV
+50.000 V4.9960 V
7.3.2
2
4
5
7
8
9
10
11
13
14
17
18
20
21
23
24
26
27
28
29
Value
upper
O.K
-4.9800 V
21.0 mV
5.0220 V
-4.9900 V
11.0 mV
5.0120 V
-4.9940 V
7.0 mV
5.0060 V
-4.9940 V
7.0 mV
5.0060 V
-4.9940 V
7.0 mV
5.0060 V
-4.9940 V
7.0 mV
5.0060 V
-4.9940 V
7.0 mV
5.0060 V
-4.9940 V
7.0 mV
5.0060 V
-4.9940 V
7.0 mV
5.0060 V
-4.9940 V
7.0 mV
5.0060 V
-4.9940 V
7.0 mV
5.0060 V
Filter
10 Hz Bessel
10Hz
30Hz Bessel
15Hz
30Hz
100Hz Bessel
50Hz
100Hz
100Hz Bessel
10kHz
100Hz Butterworth 10kHz
300Hz Bessel
150Hz
300 HZ
1kHz Bessel
500 Hz
1k
3kHz Bessel
1k5
3k
10kHz Bessel
5k
10 k
30kHz Bessel
15 k
30 k
100kHz Bessel
50 k
100 k
300kHz Bessel
150 k
300 k
2.085
3.048
2.085
3.048
2.085
0.041
0.025
3.048
2.085
3.048
2.085
3.048
2.085
3.048
2.085
3.048
2.085
3.048
2.085
2.613
1.787
2.945
3.381
2.945
3.381
2.945
0.088
0.051
3.381
2.945
3.381
2.945
3.381
2.945
3.381
2.945
3.381
2.945
3.381
2.945
2.898
2.524
Page 35
Specifications
8.1.1
DC-Accuracy
DAQP-V-B
Module Range
Bipolar
Unipolar
10mV
20mV
50mV
100mV
200mV
500mV
1V
2.5V
5V
10V
25V
50V
+10mV
20mV
+50mV
+100mV
+200mV
+500mV
+1V
+2.5V
+5V
+10V
+25V
+50V
5V Module output
Max absolute
Error
Max. Offset Max. Error
% of Value Offset
mV
mV
mV
0.02% 40V
0.042
20
21
0.02% 40V
0.044
10
11
0.02% 40V
0.05
4
5
0.02% 50V
0.07
2.5
3.5
0.02% 100V
0.14
2.5
3.5
0.02% 250V
0.35
2.5
3.5
0.02% 500V
0.7
2.5
3.5
0.02% 1.25mV
1.75
2.5
3.5
0.02% 2.5mV
3.5
2.5
3.5
0.02% 5mV
7
2.5
3.5
0.02% 12.5mV
17.5
2.5
3.5
0.02% 25mV
35
2.5
3.5
0.04% 40V
0.044
40
44
0.04% 40V
0.048
20
24
0.04% 40V
0.06
8
12
0.04% 50V
0.09
5
9
0.04% 100V
0.18
5
9
0.04% 250V
0.45
5
9
0.04% 500V
0.9
5
9
0.04% 1.25mV
2.25
5
9
0.04% 2.5mV
4.5
5
9
0.04% 5mV
9
5
9
0.04% 12.5mV
22.5
5
9
0.04% 25mV
45
5
9
Accuracy
Page 36
Gain
Output
= 20 log10
= 20 log10
dB
Input DC _ Gain
8.1.2.2
Table
Bessel
Butterworth
8.1.2.3
Output
Output _ Range
Input
Input _ Range
Limits at fo/2
Lower
Upper
-1.2dB
-0.3dB
-0.8dB
0.2dB
Limits at f0
Lower
Upper
-4.5dB
-1.5dB
-4.5dB
-1.5dB
Limits at 10*f0
Lower
Upper
-38.7 dB -32 dB
-43 dB
-39.8dB
Diagram
Page 37
CAL-Rack Connection
Calibration Rack
Digital MULTIMETER
DEWETRON DEWETRON DEWETRON
OUTPUT
8.2
Parts List
o
o
o
o
o
o
8.3
DAQP-V-B Module
Slot: Any Kind of DEWE-Module-Slot is possible.
Multi Product Calibrator
DeweSoft 6 or DEWEConfig 5.4.2.d for Module Setup
High accuracy digital Multimeter.
Cables
o DAQP-V-B-B: Safety banana plug test leads (red black)
o DAQP-V-DSUB: Safety banana plugs (red ,black) to SUB-D 9
o DAQP-V-BNC: Safety banana to BNC cable
Calibration Rack
Calibration Instruction
o
o
Page 38
Step
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
Bipolar Range
Start DEWESoft or DEWEConfig for Module Setup
Set the module to:
Range:
0.01V
Filter:
1kHz
Filter Type:
Bessel
Shunt Type: none
Coupling:
DC
Input Mode:
Bipolar
Specification
20.00 mV
21.00 mV
21.00 mV
10.00 mV
11.00 mV
11.00 mV
4.00 mV
5.00 mV
5.00 mV
2.50 mV
3.50 mV
3.50 mV
2.50 mV
3.50 mV
3.50 mV
2.50 mV
3.50 mV
3.50 mV
2.50 mV
3.50 mV
3.50 mV
2.50 mV
3.50 mV
3.50 mV
2.50 mV
3.50 mV
3.50 mV
2.50 mV
3.50 mV
3.50 mV
2.50 mV
3.50 mV
3.50 mV
2.50 mV
3.50 mV
3.50 mV
and
Limits
Lower
Upper
-20.0 mV
20.0 mV
4979.0 mV
5021.0 mV
-5021.0 mV
-4979.0 mV
-10.0 mV
10.0 mV
4989.0 mV
5011.0 mV
-5011.0 mV
-4989.0 mV
-4.0 mV
4.0 mV
4995.0 mV
5005.0 mV
-5005.0 mV
-4995.0 mV
-2.5 mV
2.5 mV
4996.5 mV
5003.5 mV
-5003.5 mV
-4996.5 mV
-2.5 mV
2.5 mV
4996.5 mV
5003.5 mV
-5003.5 mV
-4996.5 mV
-2.5 mV
2.5 mV
4996.5 mV
5003.5 mV
-5003.5 mV
-4996.5 mV
-2.5 mV
2.5 mV
4996.5 mV
5003.5 mV
-5003.5 mV
-4996.5 mV
-2.5 mV
2.5 mV
4996.5 mV
5003.5 mV
-5003.5 mV
-4996.5 mV
-2.5 mV
2.5 mV
4996.5 mV
5003.5 mV
-5003.5 mV
-4996.5 mV
-2.5 mV
2.5 mV
4996.5 mV
5003.5 mV
-5003.5 mV
-4996.5 mV
-2.5 mV
2.5 mV
4996.5 mV
5003.5 mV
-5003.5 mV
-4996.5 mV
-2.5 mV
2.5 mV
4996.5 mV
5003.5 mV
-5003.5 mV
-4996.5 mV
Page 39
Step
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
Unipolar Range
DAQP-V-B Unipolar Range
Module Setting
Calibrator
Specification
Range
Filter
VDC
0.01 V
1kHz
0.000 mV
0.04 mV
0.01 V
1kHz
5.000 mV
0.04 mV
0.01 V
1kHz
10.000 mV
0.04 mV
0.02 V
1kHz
0.000 mV
0.04 mV
0.02 V
1kHz
10.000 mV
0.04 mV
0.02 V
1kHz
20.000 mV
0.05 mV
0.05 V
1kHz
0.000 mV
0.04 mV
0.05 V
1kHz
25.000 mV
0.05 mV
0.05 V
1kHz
50.000 mV
0.06 mV
0.10 V
1kHz
0.000 mV
0.05 mV
0.10 V
1kHz
50.000 mV
0.07 mV
0.10 V
1kHz
100.000 mV
0.09 mV
0.20 V
1kHz
0.000 mV
0.10 mV
0.20 V
1kHz
100.000 mV
0.14 mV
0.20 V
1kHz
200.000 mV
0.18 mV
0.50 V
1kHz
0.000 mV
0.25 mV
0.50 V
1kHz
250.000 mV
0.35 mV
0.50 V
1kHz
500.000 mV
0.45 mV
1.00 V
1kHz
0.000 V
0.50 mV
1.00 V
1kHz
0.500 V
0.70 mV
1.00 V
1kHz
1.000 V
0.90 mV
2.50 V
1kHz
0.000 V
1.25 mV
2.50 V
1kHz
1.250 V
1.75 mV
2.50 V
1kHz
2.500 V
2.25 mV
5.00 V
1kHz
0.000 V
2.50 mV
5.00 V
1kHz
2.500 V
3.50 mV
5.00 V
1kHz
5.000 V
4.50 mV
10.00 V
1kHz
0.000 V
5.00 mV
10.00 V
1kHz
5.000 V
7.00 mV
10.00 V
1kHz
10.000 V
9.00 mV
25.00 V
1kHz
0.000 V
12.50 mV
25.00 V
1kHz
12.500 V
17.50 mV
25.00 V
1kHz
25.000 V
22.50 mV
50.00 V
1kHz
0.000 V
25.00 mV
50.00 V
1kHz
25.000 V
35.00 mV
50.00 V
1kHz
50.000 V
45.00 mV
Limits
Lower
-5040.0 mV
-42.0 mV
4956.0 mV
-5020.0 mV
-22.0 mV
4976.0 mV
-5008.0 mV
-10.0 mV
4988.0 mV
-5005.0 mV
-7.0 mV
4991.0 mV
-5005.0 mV
-7.0 mV
4991.0 mV
-5005.0 mV
-7.0 mV
4991.0 mV
-5005.0 mV
-7.0 mV
4991.0 mV
-5005.0 mV
-7.0 mV
4991.0 mV
-5005.0 mV
-7.0 mV
4991.0 mV
-5005.0 mV
-7.0 mV
4991.0 mV
-5005.0 mV
-7.0 mV
4991.0 mV
-5005.0 mV
-7.0 mV
4991.0 mV
Upper
-4960.0 mV
42.0 mV
5044.0 mV
-4980.0 mV
22.0 mV
5024.0 mV
-4992.0 mV
10.0 mV
5012.0 mV
-4995.0 mV
7.0 mV
5009.0 mV
-4995.0 mV
7.0 mV
5009.0 mV
-4995.0 mV
7.0 mV
5009.0 mV
-4995.0 mV
7.0 mV
5009.0 mV
-4995.0 mV
7.0 mV
5009.0 mV
-4995.0 mV
7.0 mV
5009.0 mV
-4995.0 mV
7.0 mV
5009.0 mV
-4995.0 mV
7.0 mV
5009.0 mV
-4995.0 mV
7.0 mV
5009.0 mV
Page 40
8.3.4
Steps
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
26*
27*
28*
*
Filter Check
Set the module to:
Range:
Filter:
Filter Type:
Shunt Type:
Coupling:
Input Mode:
5V
10 Hz
Bessel
none
DC
Bipolar
Voltage:
3.5VRMS
Frequency:
10Hz
Wave form:
Sine wave
Offset:
0V
Apply the voltages according to the following table and measure the module output RMS voltage.
Module Range
Filter
Range
Hz
V
10
5 V
10
5 V
30
5 V
30
5 V
30
5 V
100
5 V
100
5 V
100
5 V
100 Butterw. 5 V
300
5 V
300
5 V
300
5 V
1k
5 V
1k
5 V
1k
5 V
1 k Butterw. 5 V
3k
5 V
3k
5 V
3k
5 V
10 k
5 V
10 k
5 V
10 k
5 V
30 k
5 V
30 k
5 V
100 k
5 V
100 k
5 V
180 k
5 V
DAQP-V-B Filter
Input Signal
Voltage
Frequency
VRMS
Hz
3.5
10
3.5
100
3.5
15
3.5
30
3.5
300
3.5
50
3.5
100
3.5
1k
3.5
1k
3.5
150
3.5
300
3.5
3k
3.5
500
3.5
1k
3.5
10k
3.5
10k
3.5
1k5
3.5
3k
3.5
30k
3.5
5k
3.5
10k
3.5
100k
3.5
15k
3.5
30k
3.5
50k
3.5
100k
3.0
180k
Output Limits
Lower
Upper
dB
VRMS
dB
VRMS
-4.5
2.085
-1.5
-38.7
0.041
-32
-1.2
3.048
-0.3
-4.5
2.085
-1.5
-38.7
0.041
-32
-1.2
3.048
-0.3
-4.5
2.085
-1.5
-38.7
0.041
-32
-43
0.025 -36.8
-1.2
3.048
-0.3
-4.5
2.085
-1.5
-38.7
0.041
-32
-1.2
3.048
-0.3
-4.5
2.085
-1.5
-38.7
0.041
-32
-43
0.025 -36.8
-1.2
3.048
-0.3
-4.5
2.085
-1.5
-38.7
0.041
-32
-1.2
3.048
-0.3
-4.5
2.085
-1.5
-38.7
0.041
-32
-1.2
3.048
-0.3
-4.5
2.085
-1.5
-1.2
3.048
-0.3
-4.5
2.085
-1.5
-4.5
1.787
-1.5
2.945
0.088
3.381
2.945
0.088
3.381
2.945
0.088
0.051
3.381
2.945
0.088
3.381
2.945
0.088
0.051
3.381
2.945
0.088
3.381
2.945
0.088
3.381
2.945
3.381
2.945
2.524
The Input voltage is only 3 VRMS because the 5500A to not support higher voltages over 100kHz.
These are special tests to verify the filter characteristic.
Page 41
Module :
Serial Number:
Customer:
Inspection Date:
DAQP-V-B
Test equipment
Model
S/N
Temperature:
Humidity:
Tested by:
Place of inspection:
Calibration Date
Calibrator
Digital Multimeter
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
Test description
Bipolar Range
0.01V Range
0.000 mV
+10.000 mV
-10.000 mV
0.02V Range
0.000 mV
+20.000 mV
-20.000 mV
0.05V Range
0.000 mV
+50.000 mV
-50.000 mV
0.1V Range
0.000 mV
+100.000 mV
-100.00 mV
0.2V Range
0.000 mV
+200.00 mV
-200.00 mV
0.5V Range
0.000 mV
+500.00 mV
-500.00 mV
1V Range
0.000 V
+1.000 V
-1.000 V
2V Range
0.000 V
+2.000 V
-2.000 V
5V Range
0.000 V
+5.000 V
-5.000 V
10V Range
0.000 V
+10.000 V
-10.000 V
25V Range
0.000 V
+25.000 V
-25.000 V
50V Range
0.000 V
+50.000 V
-50.000 V
Calibration Procedure
Limits
lower
upper
-20.0 mV
20.0 mV
4979.0 mV 5021.0 mV
-5021.0 mV -4979.0 mV
-10.0 mV
10.0 mV
4989.0 mV 5011.0 mV
-5011.0 mV -4989.0 mV
-4.0 mV
4.0 mV
4995.0 mV 5005.0 mV
-5005.0 mV -4995.0 mV
-2.5 mV
2.5 mV
4996.5 mV 5003.5 mV
-5003.5 mV -4996.5 mV
-2.5 mV
2.5 mV
4996.5 mV 5003.5 mV
-5003.5 mV -4996.5 mV
-2.5 mV
2.5 mV
4996.5 mV 5003.5 mV
-5003.5 mV -4996.5 mV
-2.5 mV
2.5 mV
4996.5 mV 5003.5 mV
-5003.5 mV -4996.5 mV
-2.5 mV
2.5 mV
4996.5 mV 5003.5 mV
-5003.5 mV -4996.5 mV
-2.5 mV
2.5 mV
4996.5 mV 5003.5 mV
-5003.5 mV -4996.5 mV
-2.5 mV
2.5 mV
4996.5 mV 5003.5 mV
-5003.5 mV -4996.5 mV
-2.5 mV
2.5 mV
4996.5 mV 5003.5 mV
-5003.5 mV -4996.5 mV
-2.5 mV
2.5 mV
4996.5 mV 5003.5 mV
-5003.5 mV -4996.5 mV
Value
Pass
Page 42
Test description
Unipolar Range
+0.01V Range
0.000 mV
+5.000 mV
+10.000 mV
+0.02V Range
0.000 mV
+10.000 mV
+20.000 mV
+0.05V Range
0.000 mV
+25.000 mV
+50.000 mV
+0.1V Range
0.00 mV
+50.00 mV
+100.00 mV
+0.2V Range
0.00 mV
+100.00 mV
+200.00 mV
+0.5V Range
0.00 mV
+250.00 mV
+500.00 mV
+1V Range
0.000 V
+0.500 V
+1.000 V
+2V Range
0.000 V
+1.000 V
+2.000 V
+5V Range
0.000 V
+2.500 V
+5.000 V
+10V Range
0.000 V
+5.000 V
+10.000 V
+25V Range
0.000 V
+12.500 V
+25.000 V
+50V Range
0.000 V
+25.000 V
+50.000 V
Filter
10Hz
15Hz
30Hz
100Hz Bessel
50Hz
100Hz
100Hz Bessel
10kHz
100Hz Butterworth 10kHz
300Hz Bessel
150Hz
300 HZ
1kHz Bessel
500 Hz
1k
3kHz Bessel
1k5
3k
10kHz Bessel
5k
10 k
30kHz Bessel
15 k
30 k
100kHz Bessel
50 k
100 k
180kHz Bessel
180 k
10 Hz Bessel
30Hz Bessel
Limits
lower
Value
upper
-5040.0 mV
-42.0 mV
4956.0 mV
-5020.0 mV
-22.0 mV
4976.0 mV
-5008.0 mV
-10.0 mV
4988.0 mV
-5005.0 mV
-7.0 mV
4991.0 mV
-5005.0 mV
-7.0 mV
4991.0 mV
-5005.0 mV
-7.0 mV
4991.0 mV
-5005.0 mV
-7.0 mV
4991.0 mV
-5005.0 mV
-7.0 mV
4991.0 mV
-5005.0 mV
-7.0 mV
4991.0 mV
-5005.0 mV
-7.0 mV
4991.0 mV
-5005.0 mV
-7.0 mV
4991.0 mV
-5005.0 mV
-7.0 mV
4991.0 mV
-4960.0 mV
42.0 mV
5044.0 mV
-4980.0 mV
22.0 mV
5024.0 mV
-4992.0 mV
10.0 mV
5012.0 mV
-4995.0 mV
7.0 mV
5009.0 mV
-4995.0 mV
7.0 mV
5009.0 mV
-4995.0 mV
7.0 mV
5009.0 mV
-4995.0 mV
7.0 mV
5009.0 mV
-4995.0 mV
7.0 mV
5009.0 mV
-4995.0 mV
7.0 mV
5009.0 mV
-4995.0 mV
7.0 mV
5009.0 mV
-4995.0 mV
7.0 mV
5009.0 mV
-4995.0 mV
7.0 mV
5009.0 mV
2.085
3.048
2.085
3.048
2.085
0.041
0.025
3.048
2.085
3.048
2.085
3.048
2.085
3.048
2.085
3.048
2.085
3.048
2.085
1.787
2.945
3.381
2.945
3.381
2.945
0.088
0.051
3.381
2.945
3.381
2.945
3.381
2.945
3.381
2.945
3.381
2.945
3.381
2.945
2.524
O.K
Page 43
Specifications
DC-Accuracy DAQP-FREQ
Range
Accuracy
% of F.S.
Output
Range
100 Hz
1 kHz
5 kHz
10 kHz*
20 kHz**
100 kHz
200 kHz
+/-0.05%
+/-0.05%
+/-0.05%
+/-0.05%
+/-0.05%
+/-0.05%
+/-0.05%
+/-5V
+/-5V
+/-5V
+/-5V
+/-5V
+/-5V
+/-5V
+/-5mV
+/-5mV
+/-5mV
+/-5mV
+/-5mV
+/-5mV
+/-5mV
* DAQP-FREQ
** DAQP-FREQ-A
Page 44
Schematic
DAQx-FREQ
OUTPUT
FILTER
SLOW
SW1
LEDS
FAST
Supply Cable
SW2
AC
INPUT
OFFSET
DC
COUPLING
GAIN
OFFSET
TRIGGER LEVEL
DEWE-DAQP-FREQ
Us ed for
input Offset
adjus tment
t /i n
Working Part
Si
S p e c ia l O u
gn
al
Su
Ou
pp
b le
a b le
Ca
DAQ P-DM M
tC
ly
Digital MULTIMETER
DAQ P-D M M
Direct
Page 45
Parts List
o
o
o
o
o
o
o
9.4
DAQP-FREQ Module
Slot : Any Kind of DEWE-Module-Slot is possible.
Multi Product Calibrator
Digital Multimeter
Cables
o Connector Cable Banana-> DSUB 9 (+..Pin2; -..Pin7; You find more information in the Module
Manual)
o Supply Cable
o 9 Pin SUB-D 9 Female, to 9 Pin SubD Female 1:1. (This is a Working Part and should be changed
after 100 calibrations, because oft the limited connector lifetime.)
o Isolated Screwdriver for adjusting the Module
Computer with Serial Interface (RS232 or RS485).
DEWEConfig 5.4.2 or later, which is included on the DEWE-System CD-Rom. You will find the actual version
under www.dewetron.com
Calibration Instruction
o
o
o
o
Jumper Setting:
Coupling: AC .
Output Filter: Slow .
Output Filter
S low
Fas t
AC
DC
Coupling
DAQP-FREQ: Turn the Trigger Level completely to the left. (You will hear two Click).
o
o
o
Page 46
Module Range
Filter
Range
kHz
kHz
200
200
200
200
100
100
100
100
20
20
20
20
10
10
10
10
5
5
5
5
1
1
1
1
0.1
0.1
0.1
0.1
Input Signal
VPP
2
2
2
2
2
2
2
2
2
2
2
2
2
2
kHz
200
0
100
0
10
0
10
0
5
0
1
0
0.1
0
Output Limits
lower
V
4.9950 V
-5.0050V
4.9950 V
-5.0050V
4.9950 V
-5.0050V
4.9950 V
-5.0050V
4.9950 V
-5.0050V
4.9950 V
-5.0050V
4.9950 V
-5.0050V
upper
V
5.0050 V
-4.9950 V
5.0050 V
-4.9950 V
5.0050 V
-4.9950 V
5.0050 V
-4.9950 V
5.0050 V
-4.9950 V
5.0050 V
-4.9950 V
5.0050 V
-4.9950 V
* DAQP-FREQ-A
Page 47
9.4.1
o
o
o
o
o
9.4.2
o
o
o
o
9.4.3
Offset Adjustment
Set the module to DC.
Connect The Digital Multimeter To the Signal Out Cable, and
the Range to DCV.
Apply a 2VPP 100kHz Rectangle Signal to the DAQP-FREQ Input.
Adjust the Offset Potentiometer until The Module output voltage is
-0.5mV and 1.5mV .
Offset
Change
between
Gain Adjustment
9.4.3.1
o Apply a 2VPP 200kHz Sine wave Signal to the DAQP-FREQ Input.
o Adjust the Gain Potentiometer until The Module output voltage is
4.9985 V and 5.0015 V (best value would be 5.000 V).
between
GAIN
9.4.3.2
o Apply a 2VDC Signal to the input (that means 0 Hz) .
o Adjust the Gain Potentiometer until The Module output voltage is between -5.0015 V and -4.9985 V
(best value would be -5.000 V).
o If necessary balance the gain, that the Error is in both cases the same. For Example its better to adjust
+5.0005 V and 4.9995V instead of +5.0010V and 5.0000V .
o Continue with the calibration check 7.4.1 .
E-M0243E Printing Version 1.6.9 FEB, 2008
Page 48
Calibration Report
Module :
Serial Number:
Customer:
Inspection Date:
DAQP-FREQ
Temperature:
Humidity:
Tested by:
Place of inspection:
Test equipment
Model
S/N
Calibration Date
Calibrator
Digital Multimeter
Calibration Procedure
Limits
lower
upper
Test
7.4.1
1
2
3
4
5*
6*
7
8
9
10
11
12
13
14
Range Check
200kHz Range @ 200kHz
200kHz Range @ 0kHz
100kHz Range @ 100kHz
100kHz Range @ 0kHz
20kHz Range @ 20kHz
20kHz Range @ 0kHz
10kHz Range @ 10kHz
10kHz Range @ 0kHz
5kHz Range @ 5kHz
5kHz Range @ 0kHz
1kHz Range @ 1kHz
1kHz Range @ 0kHz
0.1kHz Range @ 0.1kHz
0.1kHz Range @ 0kHz
4.9950 V
-5.0050V
4.9950 V
-5.0050V
4.9950 V
-5.0050V
4.9950 V
-5.0050V
4.9950 V
-5.0050V
4.9950 V
-5.0050V
4.9950 V
-5.0050V
Value
O.K
5.0050 V
-4.9950 V
5.0050 V
-4.9950 V
5.0050 V
-4.9950 V
5.0050 V
-4.9950 V
5.0050 V
-4.9950 V
5.0050 V
-4.9950 V
5.0050 V
-4.9950 V
Page 49
Accuracy
% of F.S.
Output
Range
+/- 0.1
+/- 0.2
+/- 0.5
+/- 1
+/- 2
+/- 5
+/-0.05%
+/-0.05%
+/-0.05%
+/-0.05%
+/-0.05%
+/-0.05%
+/-5V
+/-5V
+/-5V
+/-5V
+/-5V
+/-5V
+/-5mV
+/-5mV
+/-5mV
+/-5mV
+/-5mV
+/-5mV
Gain
Output
= 20 log10
= 20 log10
dB
Input DC _ Gain
Output
Output _ Range
Input
Input _ Range
10.1.2.2 Table
Limits at fo/2
Lower
Upper
-1.2dB
-0.1 dB
-0.8dB
0.2dB
Limits at f0
Lower
Upper
-4.5dB
-1.5dB
-4.5dB
-1.5dB
10.1.2.3 Diagram
Page 50
G ND
DAQx-BRIDGE
EXCITATION
LEDS
SW2
J5
FILTER
GTU
GAIN
P4
P3
J3
Fine
Coarse
BALANCE P2
Bessel
Butterworth
2.5 V
5V
10 V
15 V
BALANCE
Supply Cable
P1
EXCITATION
VOLTAGE
J1
DEWE-DAQN-BRIDGE
W orking Part
t/in
Excitation& S ense-
gn
al
Su
Ou
pp
b le
a b le
Ca
tC
ly
Excitation+
& S ens e+
Si
S p e c ia l O u
E xcitation Cable
DAQ P-DM M
Direct
Page 51
BRIDGE CALIBRATOR
50%
+
100%
0%
0,5mV/V
1,0mV/V
2,0mV/V
5,0mV/V
10mV/V
able
Bridge Connector C
DAQx-BRIDGE
SW2
J5
FILTER
GTU
GAIN
P4
P3
J3
Fine
Coarse
BALANCE P2
Bessel
Butterworth
EXCITATION
LEDS
BALANCE
Supply Cable
P1
SW1
2.5 V
5V
10 V
15 V
EXCITATION
VOLTAGE
J1
DEWE-DAQN-BRIDGE
t /in
Working Part
Si
p e c ia l O u
al
Su
gn
Ou
pp
Ca
b le
b le
DAQ P-DMM
t Ca
ly
Digital MULTIMETER
DAQP-DMM
Direct
Page 52
DAQx-BRIDGE Module
Slot : Any Kind of DEWE-Module-Slot is possible.
Multi Product Calibrator or a Bridge Calibrator (such as the HBM K3607)
High accuracy digital Multimeter
Cables
o Bridge Input Cable* (with included high accuracy 100:1 voltage divider).
Resistor specifications : Tolerance0.005%, 5ppm/C
Pin assignment:
1 : Excitation +
6: Sense +
2 : Input +
7: Input 3 : Sense 8: Excitation 4: GND
9: Case
5: Remote
100R
9k9
+
Excitation Cable
+
Input Cable
GND
Supply Cable
9 Pin SubD Female 9 Pin SubD Female 1:1. (This is a Working Part and should be changed after
100 calibrations, because oft the limited connector lifetime.)
Isolated Screwdriver for adjusting the Module
Page 53
Output
Setting
Voltage
lower
upper
V
2.4999
5.0
4.9999
2.5
V
2.5001
5.0001
10
9.9995
10.0005
15
14.999
15.001
Voltage
defaul
F 7.4.2
CMR
10.4.2.1
o
o
Apply +2.500V to the Cable.(that means +25.00mV for the module, because of the voltage divider inside
the cable).
Adjust the gain potentiometer until the output is between 4.9980V and 5.0020V, the optimum value is
5.0000V.
Page 54
Module Setting
Filter
Range
kHz
mV/V
1
5
1
-5
1
+2
1
-2
1
1
1
-1
1
0.5
1
-0.5
1
0.2
1
-0.2
1
0.1
1
-0.1
Calibrator
Limits
Source*
lower
upper
VDC
V
V
2.5000
4.9950
5.0050
-2.5000
-5.0050
-4.9950
1.0000
4.9950
5.0050
-1.0000
-5.0050
-4.9950
0.5000
4.9950
5.0050
-0.5000
-5.0050
-4.9950
0.2500
4.9950
5.0050
-0.2500
-5.0050
-4.9950
0.1000
4.9950
5.0050
-0.1000
-5.0050
-4.9950
0.05000
4.9950
5.0050
-0.05000
-5.0050
-4.9950
*Fluke Calibrator: Fix the Calibrator Reference Voltage to 3.3V
Module Range
Filter
Range
10Hz
+/-5mVV
10Hz
+/-5mV/V
100Hz
+/-5mVV
100Hz
+/-5mV/V
1kHz
+/-5mVV
1kHz
+/-5mV/V
5kHz
+/-5mVV
5kHz
+/-5mV/V
Input Signal
Voltage
Frequency
1.75 VRMS
5Hz
1.75 VRMS
10Hz
1.75 VRMS
50Hz
1.75 VRMS
100Hz
1.75 VRMS
500Hz
1.75 VRMS
1kHz
1.75 VRMS
2.5kHz
1.75 VRMS
5kHz
Output Limits
lower
upper
3.05 VRMS
3.46 VRMS
2.08 VRMS
2.94 VRMS
3.05 VRMS
3.46 VRMS
2.08 VRMS
2.94 VRMS
3.05 VRMS
3.46 VRMS
2.08 VRMS
2.94 VRMS
3.05 VRMS
3.46 VRMS
2.08 VRMS
2.94 VRMS
Module Range
Filter
Range
20kHz
+/-5mV/V
Input Signal
Voltage
Frequency
1.75 VRMS
20kHz
Output Limit
lower
3.19 VRMS
Page 55
Excitation Voltage
o
It doesnt makes sense to calibrate the excitation voltage separate, because it is included in the result, if
you are using a Bridge Calibrator.
o
o
10.5.2.1
o
o
o
Module Setting
Filter
Range
kHz
mV/V
1
5
1
-5
1
+2
1
-2
1
1
1
-1
1
0.5
1
-0.5
1
0.2
1
-0.2
1
0.1
1
-0.1
Calibrator
Source
mV/V
+5.000
-5.000
+2.000
-2.000
1.000
-1.000
0.500
-0.500
0.200
-0.200
0.100
-0.100
Limits
lower
V
4.9950
-5.0050
4.9950
-5.0050
4.9950
-5.0050
4.9950
-5.0050
4.9950
-5.0050
4.9950
-5.0050
upper
V
5.0050
-4.9950
5.0050
-4.9950
5.0050
-4.9950
5.0050
-4.9950
5.0050
-4.9950
5.0050
-4.9950
Page 56
DAQx-BRIDGE
Temperature:
Humidity:
Tested by:
Place of inspection:
Test equipment
Model
S/N
Calibration Date
Calibrator
Digital Multimeter
Test
8.4.1
10.4.1.
1
10.4.3.
1
2
3
4
5
6
7
8
9
10
11
12
10.4.4
1
2
3
4
5
6
7
8
10.4.6
10
Excitation Voltage
2.5 V Excitation
5 V Excitation
10 V Excitation
15 V Excitation
CMR
Range Check
Gain ( + 5mV/V Range)
Gain ( - 5mV/V Range)
Gain ( + 2mV/V Range)
Gain ( - 2mV/V Range)
Gain ( + 1mV/V Range)
Gain ( - 1mV/V Range)
Gain ( + 0.5mV/V Range)
Gain ( - 0.5mV/V Range)
Gain ( + 0.2mV/V Range)
Gain ( - 0.2mV/V Range)
Gain ( + 0.1mV/V Range)
Gain ( - 0.1mV/V Range)
Filter Check (Bessel)
10 Hz Filter @ 5Hz
10 Hz Filter @ 10Hz
100Hz Filter @ 50 Hz
100Hz Filter @ 100 Hz
1kHz Filter @ 500 Hz
1kHz Filter @ 1 kHz
5kHz Filter @ 2.5k Hz
5kHz Filter @ 3kHz
Band width
20kHz Filter @ 20kHz
Calibration Procedure
Limits
lower
upper
2.4999V
4.9999V
9.9995V
14.999V
2.5001v
5.0001V
10.0005V
15.001V
40mVRMS
4.9950
-5.0050
4.9950
-5.0050
4.9950
-5.0050
4.9950
-5.0050
4.9950
-5.0050
4.9950
-5.0050
5.0050
-4.9950
5.0050
-4.9950
5.0050
-4.9950
5.0050
-4.9950
5.0050
-4.9950
5.0050
-4.9950
3.05 VRMS
2.08 VRMS
3.05 VRMS
2.08 VRMS
3.05 VRMS
2.08 VRMS
3.05 VRMS
2.08 VRMS
3.46 VRMS
2.94 VRMS
3.46 VRMS
2.94 VRMS
3.46 VRMS
2.94 VRMS
3.46 VRMS
2.94 VRMS
Value
O.K
2.16VRMS
Page 57
Accuracy
% of F.S. *
Output
Range
+/- 1
+/- 2
+/- 5
+/- 10
+/- 20
+/- 50
+/-0.1%
+/-0.1%
+/-0.1%
+/-0.1%
+/-0.1%
+/-0.1%
+/-5V
+/-5V
+/-5V
+/-5V
+/-5V
+/-5V
+/-10mV
+/-10mV
+/-10mV
+/-10mV
+/-10mV
+/-10mV
* at 5 V excitation voltage
Gain
Output
= 20 log10
= 20 log10
dB
Input DC _ Gain
Output
Output _ Range
Input
Input _ Range
11.1.2.2 Table
Bessel
Butterworth
Limits at fo/2
Lower
Upper
-1.2dB
-0.3 dB
-0.8dB
0.2 dB
Limits at f0
Lower
Upper
-4.5dB
-1.5dB
-4.5dB
-1.5dB
Limits at 10*f0
Lower
Upper
-38.7dB
-32.0dB
-43.0dB
-36.8dB
11.1.2.3 Diagram
Page 58
G ND
Excitation+
& S ens e+
DAQP-BRIDGE-A
Unit under Test
Excitation& S ense-
SW1
LEDS
SW2
Supply Cable
t/in
Working Part
Si
S p e c ia l O u
gn
al
Su
Ou
pp
b le
a b le
Ca
D AQ P-D M M
tC
ly
Digital MULTIMETER
D AQ P-D M M
D irect
Page 59
Function Generator
OUTPUT
Calibration Rack
CAL-BRIDGE
Scanner
Uext
Bridge Amplifier
Control
DA QP-BR IDGE-A
OUTPUT
Digital MULTIMETER
Page 60
DAQP-BRIDGE-A Module
Slot : Any Kind of DEWE-Module-Slot is possible.
For the filter test it is possible to use a waveform generator as the Agilent 33120 instead of a calibrator.
Multi Product Calibrator or Bridge Calibrator:
DEWETRON CAL-BRIDGE ( recommended)
HBM K3607
High accuracy digital Multimeter
Cables
Bridge Input Cable* (with included high accuracy 100:1 voltage divider). Resistor
specifications : Tolerance0.005%, 5ppm/C
Pin assignment:
1 : Excitation +
6: Sense +
2 : Input +
7: Input 3 : Sense 8: Excitation 4: GND
9: Shunt
5: Bridge
100R
9k9
+
Excitation Cable
+
Input Cable
GND
remove Pin 9 and 5
Warning the Bridge Calibration Cable included in the Cal-Kit is designed for the older DAQP-BRIDGE
module, to use this cable for both modules it is necessary to cut out Pin 5 and Pin 9 from the DSUB9
Connector.(Use a side-cutting-pliers. This takes no effect on the calibration procedure of older DAQPBridge) *
Supply Cable
o
o
9 Pin SubD Female 9 Pin SubD Female 1:1. (This is a Working Part and should be changed after
100 calibrations, because oft the limited connector lifetime.)
* not necessary if using Bridge Calibrator
** only necessary if using a Bridge Calibrator
Page 61
Generally
The DAQP-BRIDGE-A is a module from the newer generation of Dewetron. The difference to older modules is that
there are no potentiometers inside anymore. The whole calibration can be done via software. This makes it possible
to calibrate and adjust the module without opening.
Connect the Modules Excitation Voltage with the Test Cable to the digital Multimeter (DCV Range).
Use DEWESoft to select the different Excitation Voltages.
The measured voltage should be the selected excitation voltage 2mV.
Step
1
2
3
4
5
6
DAQP-BRIDGE-A
Excitation
VDC
0.250*
0.500*
1.000
2.500
5.000
10.000
Limits
lower
V
0.248
0.498
0.998
2.498
4.998
9.998
upper
V
0.252
0.502
1.002
2.502
5.002
10.002
o
o
o
o
o
o
o
o
o
Set the module to 1 mV/V and 5V excitation ( 5mV ) and 1kHz Input Range.
Connect the module input plugs to the calibrator output.
Bridge Input +
< --- > Calibrator+
Bridge Input < --- > Calibrator Bridge GND
< --- > CalibratorConnect the digital Multimeter to the module output.
Signal out cable+ < --- > Digital Multimeter +
Signal out cable- < --- > Digital Multimeter Apply 0.000 V to the input.
Press amplifier zero in DEWESoft. ( Serial command: ##{addr}R8{Carriage Return} )
Press Bridge zero in DEWESoft. ( ##{addr}R7{Carriage Return} )
Now the amplifier is offset adjusted for all ranges.
Apply 5mV to the Module ( 500.00mV calibrator voltage )
The output Voltage should be 5.000V 10mV
Apply -5mV to the Module ( -500.00mV calibrator voltage )
The output Voltage should be -5.000V 10mV
Proceed with the other ranges in the same way, except executing the amplifier zero.
Page 62
1
2
3
4
5
6
7
8
9
10
11
12
Module Setting
Input Range
Filter
mV/V
mV
kHz
1
5
1
-1
5
1
+2
10
1
-2
10
1
5
25
1
-5
25
1
10
50
1
-10
50
1
20
100
1
-20
100
1
50
250
1
-50
250
1
Calibrator
Source
VDC
0.5000
-0.5000
1.0000
-1.0000
2.5000
-2.5000
5.000
-5.000
10.000
-10.000
25.00
-25.00
Limits
lower
V
4.9900
-5.0010
4.9900
-5.0010
4.9900
-5.0010
4.9900
-5.0010
4.9900
-5.0010
4.9900
-5.0010
upper
V
5.0010
-4.9900
5.0010
-4.9900
5.0010
-4.9900
5.0010
-4.9900
5.0010
-4.9900
5.0010
-4.9900
Bessel characteristic
o Connect the Calibrator to the bridge input cable.
o Set the module input to 5mV/V and 10 Hz Bessel.
o Set the Calibrator to 1.75 VRMS (Sine wave) @ 10Hz.
o Check the Output Voltage (RMS-Value) with the digital Multimeter .
o It should be between 2.95VRMS (-1.5 dB) and 2.08VRMS (-4,5dB).
o Set the DAQP-BRIDGE-A module to 5mV/V and 100 Hz Bessel
o Set the Calibrator to 50Hz (f0/2) .
o The Output Voltage now should be between 3.38VRMS (-0.3 dB) and 3.04VRMS (-1.2 dB).
o Continue with the other filters as listed in the table 11.4.5 .
Butterworth characteristic
o Set the module input to 5mV/V and 100Hz Butterworth.
o Set the Calibrator to 1.75 VRMS (Sine wave) @ 1kHz.
o The output should be between 24.7 mVRMS (-43 dB) and 50.6 mVRMS (-36.8dB).
o It is also important that it is 10mVRMS lower than in the test 11.4.5 Nr.4 ( 100Hz Bessel @1kHz)
Module Range
Filter
Range
10Hz Bessel
+/-5mV/V
100Hz Bessel
+/-5mVV
100Hz Bessel
+/-5mV/V
100Hz Bessel
+/-5mV/V
1kHz Bessel
+/-5mVV
1kHz Bessel
+/-5mV/V
5kHz Bessel
+/-5mVV
5kHz Bessel
+/-5mV/V
20kHz Bessel
+/-5mVV
20kHz Bessel
+/-5mV/V
100Hz Butterw. +/-5mV/V
Input Signal
Voltage
Frequency
1.75 VRMS
10Hz
1.75 VRMS
50Hz
1.75 VRMS
100Hz
1.75 VRMS
1 kHz
1.75 VRMS
500Hz
1.75 VRMS
1kHz
1.75 VRMS
2.5kHz
1.75 VRMS
5kHz
1.75 VRMS
10kHz
1.75 VRMS
20kHz
1.75 VRMS
1 kHz
Output Limits
lower
upper
2.08 VRMS
2.95 VRMS
3.04 VRMS
3.38 VRMS
2.08 VRMS
2.95 VRMS
40.6mVRMS
88.0 mVRMS
3.04 VRMS
3.38 VRMS
2.08 VRMS
2.95 VRMS
3.04 VRMS
3.38 VRMS
2.08 VRMS
2.95 VRMS
3.04 VRMS
3.38 VRMS
2.08 VRMS
2.95 VRMS
24.7 mVRMS
50.6 mVRMS
Page 63
o
o
o
o
o
o
o
o
o
Module Setting
Range EXC. Filter
mV/V
V
kHz
1
5
1
-1
5
1
+2
5
1
-2
5
1
+5
5
1
-5
5
1
10
5
1
-10
5
1
20
5
1
-20
5
1
50
5
1
-50
5
1
25
10
1
Calibrator
Source
mV/V
+1.000
-1.000
+2.000
-2.000
5.000
-5.000
10.000
-10.000
20.00
-20.00
50.00
-50.00
24.00
Limits
lower
V
4.9900
-5.0100
4.9900
-5.0100
4.9900
-5.0100
4.9900
-5.0100
4.9900
-5.0100
4.9900
-5.0100
4.9900
upper
V
5.0100
-4.9900
5.0100
-4.9900
5.0100
-4.9900
5.0100
-4.9900
5.0100
-4.9900
5.0100
-4.9900
5.0100
Page 64
Bessel characteristic
Connect the module input plugs to the bridge calibrator .
Bridge DSUB9 < ---> CAL-BRIDGE DSUB9
Connect the multi product calibrator to the CAL-BRIDGE modulation input.
Calibrator source < --- > CAL-BRIDGE UEXT
Connect the digital Multimeter to the module output.
Signal out cable+ < --- > Digital Multimeter +
Signal out cable- < --- > Digital Multimeter
CAL-Rack BNC
< --- > Digital Multimeter *
* In Case of using a CAL-Rack
o Set the module input to 5mV/V and 10Hz Bessel.
o Set the CAL-BRIDGE to 10mV/V; UEXT
o Set the multi product calibrator to 3.5 VRMS 10Hz (Sine wave).
o Check the Output Voltage (RMS-Value) with the digital Multimeter .
o It should be between 2.95VRMS (-1.5 dB) and 2.08VRMS (-4,5dB).
o Set the DAQP-BRIDGE-A module to 5mV/V and 100 Hz Bessel
o Set the Calibrator to 50Hz (f0/2) .
o The Output Voltage now should be between 3.38VRMS (-0.3 dB) and 3.04VRMS (-1.2 dB).
o Continue with the other filters as listed in the table 11.4.9.
Butterworth characteristic
o Set the module input to 5mV/V and 100Hz Butterworth.
o Set the Calibrator to 3.5 VRMS @ 1kHz.
o The output should be between 24.7 mVRMS (-43 dB) and 50.6 mVRMS (-36.8dB).
o It is also important that it is 10mVRMS lower than in the test 11.4.9 Nr.4 ( 100Hz Bessel @1kHz)
Module Range
Filter
Range
10Hz Bessel
+/-5mV/V
100Hz Bessel
+/-5mVV
100Hz Bessel
+/-5mV/V
100Hz Bessel
+/-5mV/V
1kHz Bessel
+/-5mVV
1kHz Bessel
+/-5mV/V
5kHz Bessel
+/-5mVV
5kHz Bessel
+/-5mV/V
20kHz Bessel
+/-5mVV
20kHz Bessel
+/-5mV/V
100Hz Butterw. +/-5mV/V
Input Signal
Voltage
Frequency
3.50 VRMS
10Hz
3.50 VRMS
50Hz
3.50 VRMS
100Hz
3.50 VRMS
1 kHz
3.50 VRMS
500Hz
3.50 VRMS
1kHz
3.50 VRMS
2.5kHz
3.50 VRMS
5kHz
3.50 VRMS
10kHz
3.50 VRMS
20kHz
3.50 VRMS
1 kHz
Output Limits
lower
upper
2.08 VRMS
2.95 VRMS
3.04 VRMS
3.38 VRMS
2.08 VRMS
2.95 VRMS
40.6mVRMS
88.0 mVRMS
3.04 VRMS
3.38 VRMS
2.08 VRMS
2.95 VRMS
3.04 VRMS
3.38 VRMS
2.08 VRMS
2.95 VRMS
3.04 VRMS
3.38 VRMS
2.08 VRMS
2.95 VRMS
24.7 mVRMS
50.6 mVRMS
Page 65
BRIDGE-A
Temperature:
Humidity:
Tested by:
Place of inspection:
Test equipment
Model
S/N
Calibration Date
Calibrator
Digital Multimeter
Test
1
2
3
4
5
6
1
2
3
4
5
6
7
8
9
10
11
12
1
2
3
4
5
6
7
8
9
10
11
Excitation Voltage
0.25 V Excitation
0.5 V Excitation
1 V Excitation
2.5 V Excitation
5 V Excitation
10 V Excitation
Range Check
Gain ( + 1mV/V Range)
Gain ( - 1mV/V Range)
Gain ( + 2mV/V Range)
Gain ( - 2mV/V Range)
Gain ( + 5mV/V Range)
Gain ( - 5mV/V Range)
Gain ( + 10mV/V Range)
Gain ( - 10mV/V Range)
Gain ( + 20mV/V Range)
Gain ( -20mV/V Range)
Gain ( + 50mV/V Range)
Gain ( - 50mV/V Range)
Filter Check
10Hz Bessel
100Hz Bessel
100Hz Bessel
100Hz Bessel
1kHz Bessel
1kHz Bessel
5kHz Bessel
5kHz Bessel
20kHz Bessel
20kHz Bessel
100Hz Butterw.
Calibration Procedure
Limits
lower
upper
0.248
0.498
0.998
2.498
4.998
9.998
0.252
0.502
1.002
2.502
5.002
10.002
4.9900
-5.0100
4.9900
-5.0100
4.9900
-5.0100
4.9900
-5.0100
4.9900
-5.0100
4.9900
-5.0100
5.0100
-4.9900
5.0100
-4.9900
5.0100
-4.9900
5.0100
-4.9900
5.0100
-4.9900
5.0100
-4.9900
2.08 VRMS
3.04 VRMS
2.08 VRMS
40.6mVRMS
3.04 VRMS
2.08 VRMS
3.04 VRMS
2.08 VRMS
3.04 VRMS
2.08 VRMS
24.7 mVRMS
Value
O.K
2.95 VRMS
3.38 VRMS
2.95 VRMS
88.0 mVRMS
3.38 VRMS
2.95 VRMS
3.38 VRMS
2.95 VRMS
3.38 VRMS
2.95 VRMS
50.6 mVRMS
Page 66
Output
Range
0.05%
0.05%
0.05%
0.05%
0.05%
0.05%
0.05%
0.05%
0.05%
0.05%
5V
5V
5V
5V
5V
5V
5V
5V
5V
5V
5mV
5mV
5mV
5mV
5mV
5mV
5mV
5mV
5mV
5mV
at 5 V excitation voltage
Calibration is only possible
with special bridge calibrator
1.
2.
Out of Specification
Not recommended
Suitable calibrator
Accuracy
0.25
0.5
1
2.5
5
10
0.05%0.7mV
0.05%0.7mV
0.05%0.7mV
0.05%0.7mV
0.05%0.7mV
0.05%0.7mV
0.825mV
0.95mV
1.2mV
1.95mV
3.2mV
5.7mV
Gain
Output
= 20 log10
= 20 log10
dB
Input DC _ Gain
Output
Output _ Range
Input
Input _ Range
12.1.3.2 Table
Bessel
Butterworth
Limits at fo/2
Lower
Upper
-1.2dB
-0.3 dB
-0.8dB
0.2 dB
Limits at f0
Lower
-4.5dB
-4.5dB
Upper
-1.5dB
-1.5dB
Limits at 10*f0
Lower
Upper
-38.7dB
-32.0dB
-43.0dB
-36.8dB
12.1.3.3 Diagram
The same specification as for DAQP-BRIDGE-A . Look at 11.1.2.3 .
Page 67
G ND
Excitation+
& S ense+
DAQP-BRIDGE-B
Unit under Test
Excitation& S ense-
SW1
LEDS
SW2
Supply Cable
t/in
Working Part
Si
S p e c ia l O u
gn
al
Su
Ou
pp
b le
a b le
Ca
DAQ P-DM M
tC
ly
Digital MULTIMETER
DAQP-DMM
Direct
Page 68
Function Generator
OUTPUT
Calibration Rack
CAL-BRIDGE
Scanner
Uext
Bridge Amplifier
Control
OUTPUT
Page 69
DAQP-BRIDGE-B Module
Slot: Any Kind of DEWE-Module-Slot is possible.
Multi Product Calibrator
For the filter test it is possible to use a waveform generator as the Agilent 33120 instead of a calibrator.
Bridge Calibrator:
DEWETRON CAL-BRIDGE ( recommended)
HBM K3607
o
o
Pin assignment:
1 : Excitation +
6: Sense +
2 : Input +
7: Input 3 : Sense 8: Excitation 4: GND
9: Shunt
5: Bridge
100R
9k9
+
Excitation Cable
+
Input Cable
GND
remove Pin 9 and 5
Warning the Bridge Calibration Cable included in the Cal-Kit is designed for the older DAQP-BRIDGE
module, to use this cable for both modules it is necessary to cut out Pin 5 and Pin9 from the DSUB9
Connector.(Use an side-cutting-pliers. This takes no effect on the calibration procedure of older
DAQP-Bridge)
Supply Cable
9 Pin SubD Female 9 Pin SubD Female 1:1. (This is a Working Part and should be changed after
100 calibrations, because oft the limited connector lifetime.)
* not necessary if using Bridge Calibrator
** only necessary if using a Bridge Calibrator
Page 70
Connect the Modules Excitation Voltage with the Test Cable to the digital Multimeter (DCV Range).
Connect a 350Ohm Resistor (any resistor between 120 and 10k is possible) between red and black excitation
voltage plug.
Use DEWESoft to select the different Excitation Voltages.
The measured voltage should be between the limits listed in the table below.
( Excitation accuracy 0.05%0.7mV)
Step
1
2
3
4
5
6
DAQP-BRIDGE-B
Excitation
VDC
0.2500*
0.5000*
1.0000
2.500
5.000
10.000
Limits
lower
V
0.249175
0.49905
0.9988
2.49805
4.9968
9.9943
upper
V
0.250825
0.50095
1.0012
2.50195
5.0032
10.0057
o
o
o
o
o
o
o
o
o
Set the module to 1 mV/V and 5V excitation ( 5mV ) and 1kHz Input Range.
Connect the module input plugs to the calibrator output.
Bridge Input +
< --- > Calibrator+
Bridge Input < --- > CalibratorConnect the digital Multimeter to the module output.
Signal out cable+ < --- > Digital Multimeter +
Signal out cable- < --- > Digital Multimeter Apply 0.000 V to the input.
Press amplifier zero in DEWESoft. ( Serial command: ##{addr}R8{Carriage Return} )
Press Bridge zero in DEWESoft. ( ##{addr}R7{Carriage Return} )
Now the amplifier is offset adjusted for all ranges.
Apply 5mV to the Module ( 500.00mV calibrator voltage )
The output Voltage should be 5.000B 5mV
Apply -5mV to the Module ( -500.00mV calibrator voltage )
The output Voltage should be -5.000B 5mV
Proceed with the other ranges in the same way.
Page 71
1
2
3
4
5
6
7
8
9
10
11
12
13
14
Module Setting
Calibrator
Limits
Input Range
Filter
Source
lower
upper
mV/V
mV
kHz
VDC
V
V
0.1
0.5
0.2
1
Not possible to calibrate with a multi product calibrator
0.5
2.5
1
5
1
0.5000
4.9950
5.0050
-1
5
1
-0.5000
-5.0050
4.9950
+2
10
1
1.0000
4.9950
5.0050
-2
10
1
-1.0000
-5.0050
4.9950
5
25
1
2.5000
4.9950
5.0050
-5
25
1
-2.5000
-5.0050
4.9950
10
50
1
5.000
4.9950
5.0050
-10
50
1
-5.000
-5.0050
4.9950
20
100
1
10.000
4.9950
5.0050
-20
100
1
-10.000
-5.0050
4.9950
50
250
1
25.00
4.9950
5.0050
-50
250
1
-25.00
-5.0050
4.9950
100
250
1
25.00
4.9950
5.0050
-100
250
1
-25.00
-5.0050
4.9950
Bessel characteristic
o Connect the Calibrator to the bridge input cable.
o Set the module input to 5mV/V and 10 Hz Bessel.
o Set the Calibrator to 1.75 VRMS (Sine wave) @ 10Hz.
o Check the Output Voltage (RMS-Value) with the digital Multimeter .
o It should be between 2.95VRMS (-1.5 dB) and 2.08VRMS (-4,5dB).
o Set the DAQP-BRIDGE-A module to 5mV/V and 100 Hz Bessel
o Set the Calibrator to 50Hz (f0/2) .
o The Output Voltage now should be between 3.38VRMS (-0.3 dB) and 3.04VRMS (-1.2 dB).
o Continue with the other filters as listed in the table 11.4.5 .
Butterworth characteristic
o Set the module input to 5mV/V and 100Hz Butterworth.
o Set the Calibrator to 1.75 VRMS (Sine wave) @ 1kHz.
o The output should be between 24.7 mVRMS (-43 dB) and 50.6 mVRMS (-36.8dB).
o It is also important that it is 10mVRMS lower than in the test 11.4.5 Nr.4 ( 100Hz Bessel @1kHz)
Page 72
Module Range
Filter
Range
mV/V
10Hz Bessel
+/-5
30Hz Bessel
+/-5
30Hz Bessel
+/-5
100Hz Bessel
+/-5
100Hz Bessel
+/-5
100Hz Bessel
+/-5
100Hz Butterw.
+/-5
300Hz Bessel
+/-5
300Hz Bessel
+/-5
1kHz Bessel
+/-5
1kHz Bessel
+/-5
3kHz Bessel
+/-5
3kHz Bessel
+/-5
10kHz Bessel
+/-5
10kHz Bessel
+/-5
30kHz Bessel
+/-5
30kHz Bessel
+/-5
100kHz Bessel
+/-5
100kHz Bessel
+/-5
200kHz Band width
+/-5
Input Signal
Voltage Frequency
VRMS
Hz
1.75
10
1.75
15
1.75
30
1.75
50
1.75
100
1.75
1k
1.75
1k
1.75
150
1.75
300
1.75
500
1.75
1k
1.75
1500
1.75
3k
1.75
5k
1.75
10k
1.75
15k
1.75
30k
1.75
50k
1.75
100k
1.75
200k
Output Limits
lower
VRMS
2.08
3.04
2.08
3.04
2.08
0.0406
0.0247
3.04
2.08
3.04
2.08
3.04
2.08
3.04
2.08
3.04
2.08
3.04
2.08
2.08
dB
-4.5
-1.2
-4.5
-1.2
-4.5
-38.7
-43.0
-1.2
-4.5
-1.2
-4.5
-1.2
-4.5
-1.2
-4.5
-1.2
-4.5
-1.2
-4.5
-4.5
upper
VRMS
dB
2.95
-1.5
3.38
-0.3
2.95
-1.5
3.38
-0.3
2.95
-1.5
0.0880
-32.0
0.0506
-36.8
3.38
-0.3
2.95
-1.5
3.38
-0.3
2.95
-1.5
3.38
-0.3
2.95
-1.5
3.38
-0.3
2.95
-1.5
3.38
-0.3
2.95
-1.5
3.38
-0.3
2.95
-1.5
3.12
-1.0
o
o
o
o
o
o
o
o
o
Set the module to 0.1 mV/V; 5V Excitation and 1kHz Input Range.
Connect the module input plugs to the bridge calibrator .
Bridge DSUB9 < ---> CAL-BRIDGE DSUB9
Connect the digital Multimeter to the module output.
Signal out cable+ < --- > Digital Multimeter +
Signal out cable- < --- > Digital Multimeter
CAL-Rack BNC
< --- > Digital Multimeter *
* In Case of using a CAL-Rack
Apply 0.000 V to the input.
Press amplifier zero in DEWESoft. ( Serial command: ##{addr}R8{Carriage Return} )
Press Bridge zero in DEWESoft. ( ##{addr}R7{Carriage Return} )
Now the amplifier is offset adjusted for all ranges.
Apply 1mV/V to the Module. ( CAL-BRIDGE: 1mV/V; 100%)
The output Voltage should be 5.000V 5mV
Apply -1mV/V to the Module. ( CAL-BRIDGE: 1mV/V; 100% ;negative output)
The output Voltage should be -5.000V 5mV
Proceed with the other ranges in the same way.
Page 73
Module Setting
Range EXC. Filter
V
kHz
mV/V
5
1
0.1
5
1
0.1
5
1
0.2
5
1
0.2
5
1
0.5
5
1
0.5
5
1
1
5
1
-1
5
1
+2
5
1
-2
5
1
+5
5
1
-5
5
1
10
5
1
-10
5
1
20
5
1
-20
5
1
50
5
1
-50
5
1
100
5
1
-100
Calibrator
Source
mV/V
+0.100
-0.100
+0.200
-0.200
+0.500
-0.500
+1.000
-1.000
+2.000
-2.000
5.000
-5.000
10.000
-10.000
20.00
-20.00
50.00
-50.00
50.00
-50.00
Limits
lower
V
4.9950
-5.0050
4.9950
-5.0050
4.9950
-5.0050
4.9950
-5.0050
4.9950
-5.0050
4.9950
-5.0050
4.9950
-5.0050
4.9950
-5.0050
4.9950
-5.0050
4.9950
-5.0050
upper
V
5.0050
4.9950
5.0050
4.9950
5.0050
4.9950
5.0050
4.9950
5.0050
4.9950
5.0050
4.9950
5.0050
4.9950
5.0050
4.9950
5.0050
4.9950
5.0050
4.9950
Page 74
Output Limits
upper
dB
VRMS
dB
-4.5
2.95
-1.5
-1.2
3.38
-0.3
-4.5
2.95
-1.5
-1.2
3.38
-0.3
-4.5
2.95
-1.5
-38.7
0.0880
-32.0
-43.0
0.0506
-36.8
-1.2
3.38
-0.3
-4.5
2.95
-1.5
-1.2
3.38
-0.3
-4.5
2.95
-1.5
-1.2
3.38
-0.3
-4.5
2.95
-1.5
-1.2
3.38
-0.3
-4.5
2.95
-1.5
-1.2
3.38
-0.3
-4.5
2.95
-1.5
-1.2
3.38
-0.3
-4.5
2.95
-1.5
-4.5
3.12
-1.0
lower
VRMS
2.08
3.04
2.08
3.04
2.08
0.0406
0.0247
3.04
2.08
3.04
2.08
3.04
2.08
3.04
2.08
3.04
2.08
3.04
2.08
2.08
Page 75
BRIDGE-B
Temperature:
Humidity:
Tested by:
Place of inspection:
Test equipment
Model
S/N
Calibration Date
Test
1
2
3
4
5
6
1
2
3
4
5
6
7
8
9
10
11
12
13
14
13
14
15
16
17
18
Excitation Voltage
0.25 V
0.5 V
1V
2.5 V
5V
10 V
Input Range
Range
EXC.
mV/V
V
0.1
5
0.1
5
0.2
5
0.2
5
0.5
5
0.5
5
1
5
-1
5
+2
5
-2
5
+5
5
-5
5
10
5
-10
5
20
5
-20
5
50
5
-50
5
100
5
-100
5
Filter
kHz
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
lower
V
4.9950
-5.0050
4.9950
-5.0050
4.9950
-5.0050
4.9950
-5.0050
4.9950
-5.0050
4.9950
-5.0050
4.9950
-5.0050
4.9950
-5.0050
4.9950
-5.0050
4.9950
-5.0050
Value
O.K
V
upper
V
5.0050
4.9950
5.0050
4.9950
5.0050
4.9950
5.0050
4.9950
5.0050
4.9950
5.0050
4.9950
5.0050
4.9950
5.0050
4.9950
5.0050
4.9950
5.0050
4.9950
Page 1 of 2
Page 76
I
Nr.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
Filter Check
10Hz Bessel
30Hz Bessel
30Hz Bessel
100Hz Bessel
100Hz Bessel
100Hz Bessel
100Hz Butterworth
300Hz Bessel
300Hz Bessel
1kHz Bessel
1kHz Bessel
3kHz Bessel
3kHz Bessel
10kHz Bessel
10kHz Bessel
30kHz Bessel
30kHz Bessel
100kHz Bessel
100kHz Bessel
200kHz Band width
Signal
Frequency
Hz
10
15
30
50
100
1k
1k
150
300
500
1k
1500
3k
5k
10k
15k
30k
50k
100k
200k
Limits
lower
VRMS
2.08
3.04
2.08
3.04
2.08
0.0406
0.0247
3.04
2.08
3.04
2.08
3.04
2.08
3.04
2.08
3.04
2.08
3.04
2.08
2.08
O.K.
upper
VRMS
2.95
3.38
2.95
3.38
2.95
0.0880
0.0506
3.38
2.95
3.38
2.95
3.38
2.95
3.38
2.95
3.38
2.95
3.38
2.95
3.12
Page 77
Range
Accuracy*
% of F.S.
Output
Range
1
3
10
30
100
+/- 5 V
+/- 1.667V
+/- 0.5V
+/- .1667V
+/- 0.05V
0.5%
0.5%
0.5%
0.5%
0.5%
+/-5V
+/-5V
+/-5V
+/-5V
+/-5V
35mVRMS
35mVRMS
35mVRMS
35mVRMS
35mVRMS
*@ 10% of f0
Gain
Output
= 20 log10
= 20 log10
dB
Input DC _ Gain
Output
Output _ Range
Input
Input _ Range
13.1.2.2 Table
Limits at fo/2
Lower
Upper
-0.2dB
0.2dB
Limits at f0
Lower
Upper
-8.0dB
-1.0dB
13.1.2.3 Diagram
Page 78
DAQx-ACC
OFFSET
Supply Cable
Check constant
Current
DEWE-DAQP-ACC
Working Part
S ig
Su
yC
lO
a bl
ut
Ca
b le
ppl
na
Digital MULTIMETER
DAQP-DM M
D irect
Page 79
DAQP-ACC Module
Slot: Any Kind of DEWE-Module-Slot is possible.
Multi Product Calibrator or Arbitrary Wave Form Generator ( Agilent 33120A or similar)
High accuracy digital Multimeter
Special Devices
o Input Filter
To DAQP-ACC
47+/-10%
Signal Input
High Pass
Filter 1 5Hz
4mA
2k2
0.1%
Cables
o Banana Plugs (red, black), to BNC.
o Supply Cable
o 9 Pin SubD Female 9 Pin SubD Female 1:1. (This is a Working Part and should be changed after
100 calibrations, because oft the limited connector lifetime.)
o Isolated screwdriver for adjusting the Module
Mathematic background:
Set the sensor supply to 8 mA. The measured value should be 17.6V 1.6 V.
Page 80
o
o
o
o
o
o
the
Offset
Gain
1 (Input Range 5.000V)
Filter
300kHz
Sensor Supply
4mA
Input Time Constant
5sec
Apply a 3.500 VRMS Sine Wave Signal with 2 kHz to the DAQP-ACC. You have to use the AC-Coupling Device
in the input line. In case of using a 5500A it is recommended to lock the 3.3V reference.
Set the Digital Multimeter to ACV and connect it to the module output.
The measured value should be 3.500VRMS 35mVRMS.
There is no possibility to adjust the Gain by hardware.
Continue with the other ranges.
o
o
o
o
o
Step
13.4.2.
1
2
3
4
5
Module Setting
Filter
Gain
kHz
300
1
300
3
300
10
300
30
100
300
Calibrator Source
Voltage
Frequency
VRMS
kHz
3.500
2
1.160
2
0.350
2
0.116
2
0.035
2
Limits
Lower
VRMS
3.465
3.465
3.465
3.465
3.465
Upper
VRMS
3.535
3.535
3.535
3.535
3.535
o
o
o
o
Step
13.4.2.
1
2
3
4
Module Setting
Filter
Gain
kHz
1
1
10
1
100
1
300
1
Calibrator Source
Voltage
Frequency
VRMS
kHz
3.500
1
3.500
10
3.500
100
3.500
300
Limits
Lower
VRMS
1.393
1.393
1.393
1.393
Upper
VRMS
3.120
3.120
3.120
3.120
Page 81
Module :
Serial Number:
Customer:
Inspection Date:
DAQP-ACC
Temperature:
Humidity:
Tested by:
Place of inspection:
Test equipment
Model
S/N
Calibration Date
Calibrator
Digital Multimeter
Test
11.4.1
1
2
13.4.1
1
13.4.2
1
2
3
4
5
13.4.3
1
2
3
4
Constant Current
Constant Current 4 mA
Constant Current 8 mA
Offset Calibration
Offset (Gain =1)
Gain Calibration
Gain 1
Gain 3
Gain 10
Gain 30
Gain 100
Filter Check
1 kHz Filter @ 1 kHz
10 kHz Filter @ 10 kHz
100 kHz Filter @ 100 kHz
300 kHz Filter @ 300 kHz
Calibration Procedure
Limits
lower
upper
8.0 V
16.0 V
9.6 V
18.8V
-0.003V
0.003V
3.465 VRMS
3.465 VRMS
3.465 VRMS
3.465 VRMS
3.465 VRMS
3.535 VRMS
3.535 VRMS
3.535 VRMS
3.535 VRMS
3.535 VRMS
1.393 VRMS
1.393 VRMS
1.393 VRMS
1.393 VRMS
3.120 VRMS
3.120 VRMS
3.120 VRMS
3.120 VRMS
Value
O.K
Page 82
DAQPCHARGE Calibration
14 DAQP-Charge
14.1 Specifications
14.1.1 Gain-Accuracy DAQP-Charge
Gain
Range
Accuracy*
% of F.S.
Output
Range
0 dB
20 dB
40 dB
60 dB
0.1mV/pC
1 mV/pC
10 mV/pC
100 mV/pC
1000 mV/pC
5V
0.5V
0.05V
0.005V
50000pC
5000pC
500pC
50pC
5pC
1%
1%
1%
1%
1%
1%
1%
1%
1%
+/-5V
+/-5V
+/-5V
+/-5V
+/-5V
+/-5V
+/-5V
+/-5V
+/-5V
50 mV
50 mV
50 mV
50 mV
50 mV
50 mV
50 mV
50 mV
50 mV
* at 2kHz
Gain
Output
= 20 log10
= 20 log10
dB
Input DC _ Gain
Output
Output _ Range
Input
Input _ Range
14.1.2.2 Table
Limits at fo/2
Lower
Upper
-0.2dB
0.2dB
Limits at f0
Lower
Upper
-6.0dB
-1.0dB
14.1.2.3 Diagram
Page 83
DAQPCHARGE Calibration
14.2 Schematic
14.2.1
DAQx-CHARGE
SW1
FILTER
LEDS
ON
SW2
OFF
SW3
Supply Cable
J1
DEWE-DAQN-CHARGE
Working Part
S ig
Su
yC
lO
a bl
ut
Ca
b le
pp l
na
Any DEWE-SLot
DAQ P-DM M
Digital MULTIMETER
D AQ P-DM M
Direct
Page 84
DAQPCHARGE Calibration
14.2.2 ICP-Mode Calibration
DAQx-CHARGE
SW1
FILTER
LEDS
ON
SW2
OFF
SW3
Constant Current
Test Point
Supply Cable
J1
DEWE-DAQN-CHARGE
Working Part
S ig
Su
yC
lO
abl
ut
Ca
b le
ppl
na
Digital MULTIMETER
DAQP-DMM
Direct
Page 85
DAQPCHARGE Calibration
14.3 Parts List
o
o
o
o
o
DAQP-CHARGE Module
Slot : Any Kind of DEWE-Module-Slot is possible.
Multi Product Calibrator
High accuracy digital Multimeter
Special Devices
o Input Filter
Signal Input
47 10%
High Pass
Filter 1 5Hz
4mA
2k2
0.1%
WARNING!
Multi product Calibrators are designed to calibrate high impedance input Multimeter. So most
of them have voltage divider for the lower voltage ranges. Also the Fluke 5500A has a 50
Ohm output resistance up to the 3.299999V Range . That means if you apply a voltage to a
module via the input filter, it has to be 2.7% higher.
Mathematical background:
VCalibrator
2k 2 10k
+ 50 R
= VModule 2k 2 + 10k
= VModule 1.0277
2k 2 10k
2k 2 + 10k
Reverence Capacitance
To DAQP-CHARGE
(Charge Mode)
Signal Input
1 .00nF
Reverence
1.00nF
0.5%
This Device should be checked once a year!
o
Cables
o Banana Plugs (red ,black), to BNC.
o Supply Cable
o 9 Pin SubD Female 9 Pin SubD Female 1:1. (This is a Working Part and should be changed after
100 calibrations, because oft the limited connector lifetime.)
o Isolated Screwdriver for adjusting the Module
Page 86
DAQPCHARGE Calibration
14.4 Calibration Instruction
14.4.1 ICP-Mode Calibration Check
14.4.1.1 Constant Current Source Check (ICP-Mode)
o
o
The Calibrator Source has to be connected via a high pass Filter (f0=15Hz) to the Input connector of the
module. (see Point 13.3 -Special Device)
Set the module to:
ICP Mode
Range
0 dB (Input Range 5.000V)
Filter
20kHz
Jumper setting :
FILTER
Filter
ON (default)
ON
OFF
Input
module
Mathematical Background:
o
o
o
o
o
o
Gain
Filter
Apply a 3.083 VRMS Sine Wave signal with 5kHz to the DAQx-CHARGE (that means 3.000 V to the module
Input. See Point 13.3 ). You have to use the AC-Coupling Device in the input line.(see Point 14.2.2)
Connect the Digital Multimeter to the T-Adaptor (Current check point) and set it to ACV .
The measured value should be 3.000VRMS +/-5mVRMS. (This is the real Input Voltage). If its not adjust the
Calibrator Voltage until the measured value is in the tolerance of +/-5mV.
Now connect the Multimeter to the Module output.
The Voltage should be 3.000VRMS+/- 30mV.
Continue with the other ranges in the same way.
Step
13.4.2.
1
2
3
4
Module Setting
Filter
Range
KHz
dB
20
0
20
20
20
40
20
60
Calibrator Source
Voltage** Frequency
VRMS
kHz
3.083
2
0.3083
2
0.03083
2
0.003083
2
Limits
Lower
VRMS
2.950
2.950
2.950
2.950
Upper
VRMS
3.050
3.050
3.050
3.050
Page 87
DAQPCHARGE Calibration
14.4.1.3 Filter Check
o SET the Module gain to 1 and the Filter to 20 kHz.
o Set the Calibrator to 3.083VRMS (Sine wave) @ 20 kHz .
o Check the output voltage (RMS-Value) with the digital Multimeter .
o It should be between 1.500VRMS (-6 dB) and 2.674 VRMS (-1 dB).
14.4.1.4 Check the other Filters at f0 .
Step
1
2
3
4
5*
Module Setting
Filter
Gain
kHz
1
1
5
1
10
1
20
1
Bypass
1
Calibrator Source
Voltage** Frequency
VRMS
kHz
3.083
1
3.083
5
3.083
10
3.083
20
3.083
30
Limits
Lower
VRMS
1.500
1.500
1.500
1.500
1.500
Upper
VRMS
2.674
2.674
2.674
2.674
2.674
*Optional Test
(it is necessary to change the jumper
setting inside the Module )
o
o
o
o
Page 88
DAQPCHARGE Calibration
14.4.2.2 Range check
o Connect the Multi Product Calibrator via the Reverence Capacitance to the DAQx-Charge Module.
o Set the module to:
Charge Mode
Range
1 mV/pC
Filter
20kHz
o Apply a 3.000VRMS sine wave signal with 2kHz to the module.
o The output voltage should be 3.000V +/- 50mV (between 2.950V and 3.050V) .
o If the test fail continue with the module adjustment (Step 14.4.3).
o Continue with the other input ranges.
Step
1
2
3
4
Module Setting
Filter
Range
kHz
mV/pC
1
20
10
20
100
20
1000
20
Calibrator Source
Voltage
Frequency
VRMS
kHz
3.000
2
0.3000
2
0.03000
2
0.003000
2
Limits
Lower
VRMS
2.950
2.950
2.950
2.950
Upper
VRMS
3.050
3.050
3.050
3.050
o
o
o
o
the
and
until
Page 89
DAQPCHARGE Calibration
14.5 Calibration Report
Module :
Serial Number:
Customer:
Inspection Date:
DAQP-CHARGE
Temperature:
Humidity:
Tested by:
Place of inspection:
Test equipment
Model
S/N
Calibration Date
Calibrator
Digital Multimeter
Reference Capacitance
Test
14.4.1.1
14.4.1.2
1
2
3
4
14.4.1.3
1
2
3
4
5
14.4.2.1
14.4.2.2
Constant Current
4mA
Gain ICP Mode
0 dB
20dB
40dB
60dB
Filter
1kHz
5kHz
10kHz
20kHz
Bypass
Offset
Offset after ~1min
Gain Charge Mode
1 pC/mV
10 pC/mV
100 pC/mV
1000 pC/mV
Calibration Procedure
Limits
lower
upper
7.8V
9.8V
2.950 VRMS
2.950 VRMS
2.950 VRMS
2.950 VRMS
3.050 VRMS
3.050 VRMS
3.050 VRMS
3.050 VRMS
1.500 VRMS
1.500 VRMS
1.500 VRMS
1.500 VRMS
1.500 VRMS
2.764 VRMS
2.764 VRMS
2.764 VRMS
2.764 VRMS
2.764 VRMS
-0.05V
0.05 V
2.950 VRMS
2.950 VRMS
2.950 VRMS
2.950 VRMS
3.050 VRMS
3.050 VRMS
3.050 VRMS
3.050 VRMS
Value
O.K
Page 90
DAQPCHARGE-A Calibration
15 DAQP-CHARGE-A
15.1 Specifications
15.1.1 Gain-Accuracy DAQP-Charge-A
Gain
Range
Accuracy*
% of F.S.
Output
Range
0 dB
20 dB
40 dB
60 dB
0.1mV/pC
1 mV/pC
10 mV/pC
100 mV/pC
1000 mV/pC
5V
0.5V
0.05V
0.005V
50000pC
5000pC
500pC
50pC
5pC
1%
1%
1%
1%
1%
1%
1%
1%
1%
5V
5V
5V
5V
5V
5V
5V
5V
5V
50 mV
50 mV
50 mV
50 mV
50 mV
50 mV
50 mV
50 mV
50 mV
* at 2kHz
Gain
Output
= 20 log10
= 20 log10
dB
Input DC _ Gain
Output
Output _ Range
Input
Input _ Range
15.1.2.2 Table
Limits at fo/2
Lower
Upper
-0.2dB
0.2dB
Limits at f0
Lower
Upper
-6.0dB
-1.0dB
15.1.2.3 Diagram
Page 91
DAQPCHARGE-A Calibration
15.2 Schematic
15.2.1
BNC-Banana
DAQx-CHARGE
SW1
Gain ICP
Supply Cable
Constant
Current
ON
J1
1th Integrator
2nd Integrator
Output
Gain
LEDS
SW2
LEDS
SW3
Reverence
Capacitor
1.00nF
DEWE-DAQN-CHARGE-A
Working Part
S ig
Su
yC
lO
a bl
ut
Ca
b le
ppl
na
Digital MULTIMETER
D AQ P-DM M
Direct
Page 92
DAQPCHARGE-A Calibration
15.2.2 Parts List
o
o
o
o
o
DAQP-CHARGE-A Module
Slot : Any Kind of DEWE-Module-Slot is possible.
Multi Product Calibrator
High accuracy digital Multimeter
Reverence Capacitance
To DAQP-CHARGE
(Charge Mode)
Signal Input
1.00nF
0.5%
1 .00nF
Reverence
Cables
o Banana Plugs (red ,black), to BNC.
o Supply Cable
o 9 Pin SubD Female 9 Pin SubD Female 1:1. (This is a Working Part and should be changed after
100 calibrations, because oft the limited connector lifetime.)
o Isolated Screwdriver for adjusting the Module
Software: DEWECONFIG 5.4.2 or newer.
Mathematical Background:
Input with a
module
should be
Page 93
DAQPCHARGE-A Calibration
15.3.1.2 Gain Check ( ICP Mode)
o
ICP
0dB (Input Range 5000mV)
50kHz
1Hz
none
o
o
o
o
o
o
Step
1
2
3
4
o
o
Module Setting
Calibrator Source
Gain
Range Filter Voltage
Frequency
dB
mV
KHz
VRMS
kHz
0
5000
50
3.500
2
20
500
50
0.3500
2
40
50
50
0.03500
2
60
5
50
0.00350
2
Calibrator
Source
Limits
Lower
VRMS
3.450
3.450
3.450
3.450
Upper
VRMS
3.550
3.550
3.550
3.550
o
o
o
o
Connect the Multi Product Calibrator via the Reference Capacitor to the DAQP-CHARGE-A Module.
Set the module to:
Input
CHARGE
Range
50000 pC
Filter
50kHz
High pass filter 1Hz
integration
none
Apply a 3.500VRMS sine wave signal with 2kHz to the module.
The output voltage should be 3.500V +/- 50mV (between 2.950V and 3.050V) .
If the test fail continue with the module adjustment (Step 14.4.3).
Continue with the other input ranges.
Step
1
2
3
4
5
Module Setting
Filter
Range
kHz
pC
50000
20
5000
20
500
20
50
20
5
20
Calibrator Source
Voltage
Frequency
VRMS
kHz
35.00
2
3.500
2
0.3500
2
0.03500
2
0.003500
2
Limits
Lower
VRMS
3.450
3.450
3.450
3.450
3.450
Upper
VRMS
3.550
3.550
3.550
3.550
3.550
Page 94
DAQPCHARGE-A Calibration
15.3.3 Filter Check
Set the module to:
Input
CHARGE
Range
5000 pC
Filter
100Hz
High pass filter 1Hz
integration
none
Set the Calibrator to 3.000VRMS (Sine wave) @ 100 Hz .
Check the output voltage (RMS-Value) with the digital Multimeter .
It should be between 1.505VRMS (-6 dB) and 2.674VRMS (-1 dB).
o
o
o
1
2
3
4
5
Module Setting
Gain
Filter
kHz
1
0.1
1
1
1
3
1
10
1
50
Calibrator Source
Voltage
Frequency
VRMS
kHz
3.000
0.1
3.000
1
3.000
5
3.000
10
3.000
20
Limits
Lower
VRMS
1.505
1.505
1.505
1.505
1.505
dB
-6
-6
-6
-6
-6
Upper
VRMS
dB
2.674
-1
2.674
-1
2.674
-1
2.674
-1
2.674
-1
o
o
o
Connect the Multi Product Calibrator via the Reference Capacitor to the DAQP-CHARGE-A Module.
Set the module to:
Input
CHARGE
Range
5000 pC
Filter
100Hz
High pass filter 1Hz
integration
single
Set the Calibrator to 3.500VRMS (Sine wave) @ 15.9 Hz.
Check the output voltage (RMS-Value) with the digital Multimeter.
It should be between 3.192VRMS (-0.8 dB) and 3.838 VRMS (0.8 dB).
o
o
o
o
o
Input
CHARGE
Range
5000 pC
Filter
100Hz
High pass filter 1Hz
integration
double
Set the Calibrator to 3.500VRMS (Sine wave) @ 15.9 Hz.
Check the output voltage (RMS-Value) with the digital Multimeter.
It should be between 3.192VRMS (-0.8 dB) and 3.838 VRMS (0.8 dB).
Page 95
DAQPCHARGE-A Calibration
15.3.5 Module adjustment
Important: This steps are only necessary if one of the Calibration checks fails, or is only marginal.
15.3.5.1 Charge Mode Gain Adjustment
o Connect the Multi Product Calibrator via the Reference Capacitor to the DAQP-CHARGE-A Module.
o Set the module to:
Input
CHARGE
Range
5000 pC
Filter
50kHz
High pass filter 1Hz
integration
none
o
o
Page 96
DAQPCHARGE-A Calibration
15.4 Calibration Report
Module :
Serial Number:
Customer:
Inspection Date:
DAQP-CHARGE
Temperature:
Humidity:
Tested by:
Place of inspection:
Test equipment
Model
S/N
Calibration Date
Calibrator
Digital Multimeter
Reference Capacitance
Test
1
2
3
4
1
2
3
4
5
1
2
3
4
5
Constant Current
4mA
Gain ICP Mode
5000mV (0dB)
500mV (20dB)
50mV (40dB)
5mV (60dB)
Gain Charge Mode
50000 pC
5000 pC
500 pC
50 pC
5 pC
Filter
0.1
1
3
10
50
Calibration Procedure
Limits
lower
upper
8.36 V
12.32 V
3.450 VRMS
3.450 VRMS
3.450 VRMS
3.450 VRMS
3.550 VRMS
3.550 VRMS
3.550 VRMS
3.550 VRMS
3.450 VRMS
3.450 VRMS
3.450 VRMS
3.450 VRMS
3.450 VRMS
VRMS
dB
1.505
-6
1.505
-6
1.505
-6
1.505
-6
1.505
-6
3.550 VRMS
3.550 VRMS
3.550 VRMS
3.550 VRMS
3.550 VRMS
VRMS
dB
2.674 -1
2.674 -1
2.674 -1
2.674 -1
2.674 -1
Value
O.K
Page 97
DAQx-TRQ Calibration
16 DAQx-TRQ
16.1 Specifications
16.1.1 Linearity DAQx-TRQ
Gain
Range
mV/V
Linearity
% of F.S.
Output
Range
High
Low
164
10640
+/- 0.1%
+/- 0.1%
+/- 5V
+/- 5V
+/- 10mV
+/- 10mV
Gain
Output
= 20 log10
= 20 log10
dB
Input DC _ Gain
Output
Output _ Range
Input
Input _ Range
16.1.2.2 Table
Limits at fo/2
Lower
Upper
-0.8dB
0.2dB
Limits at f0
Lower
Upper
-4.2dB
-1.8dB
16.1.2.3 Diagram
Page 98
DAQx-TRQ Calibration
16.2 Schematic
BRIDGE CALIBRATOR
50%
+
100%
0%
0,5mV/V
1,0mV/V
2,0mV/V
5,0mV/V
10mV/V
ble
TRQ Connector Ca
DAQP-TRQ
SW1
0.2 mV/V
0.56 mV/V
2.2 mV/V
7.3 mV/V
24 mV/V
80 mV/V
GAIN
ZERO
Bessel
SW2
GAIN
low
FILTER
Butterworth
high
Supply Cable
LEDS
FREQ PHASE
DEWE-DAQP-TRQ
t/in
Working Part
Si
S p e c ia l O u
gn
al
Su
le
b le
ab
t Ca
C
ly
DAQP-DM M
Ou
pp
DAQP-DM M
Direct
Page 99
DAQx-TRQ Calibration
16.3 Parts List
o
o
o
o
o
DAQP-TRQ Module
Slot: Any Kind of DEWE-Module-Slot is possible.
Bridge Calibrator (such as the HBM K3607)
High accuracy digital Multimeter
Cables
TRQ Connector Cable
Pin assignment:
DAQP-TRQ
D-SUB9
1 : Excitation +
6: Excitation +
2 : Input +
7: Input 3 : Excitation 8: Excitation 4: GND
9: Case
5: Remote
K3607
BU1
BU2
BU3
BU4
BU5
Supply Cable
9 Pin SUBD Female 9 Pin SUBD Female 1:1. (This is a Working Part and should be changed after
100 calibrations, because oft the limited connector lifetime.)
Isolated Screwdriver for adjusting the Module
Page 100
DAQx-TRQ Calibration
16.4 Calibration Instruction
o
Jumper Setting :
Filter Characteristic: Bessel.
Gain: High.
Zero: 2.2mV/V
Open the module (the two screws on the back of the Module).
Connect a Digital Multimeter to the Module Output Connectors of Your
Supply Cable, and set it to DCV.
Apply 5mV/V to the Module Input (that means 10mV/V for the Calibrator)
Adjust the Phase Potentiometer until the Output DC Voltage reaches its
maximum.
Page 101
DAQx-TRQ Calibration
16.4.2 Gain Linearity
o
o
o
o
o
o
This Check is only to provide that nothing is damaged inside the module. It is based on the Adjustment you
did before in point 9.4.4 .
That means that if you have adjusted Range D (4..8mV/V) to 5mV/V with the Gain Potentiometer, the other
Ranges have approximately the Sensitivities listed up in the table below.
Set the Module to Range A and apply 5mV/V to the input (that means 10mV/V for the Calibrator). The Output
should now be 0.5V +/- 0.1V .
Continue with the other Rangers as specified in the Table below.
Step
DAQP-TRQ
Range
Sensitivity
9.4.5.
1
2
3
adjusted
4
5
6
A
B
C
D
E
F
F
High
High
High
High
High
High
Low
mV/V
40
20
10
5
2.5
1.25
12.5
Calibrator K3607
Input from
Calibrator
Calibrator
Setting
mV/V
mV/V
5
10
5
10
5
10
5
10
2
4
1
2
5
10
Digital Multimeter
Output
Voltage
V
0.5 +/- 0.1V
1.0 +/- 0.1V
2.5 +/- 0.1V
5.0 +/- 0.1V
4.0 +/- 0.1V
4.0 +/- 0.1V
2.0 +/- 0.1V
Page 102
DAQx-TRQ Calibration
16.5 Calibration Report
Module :
Serial Number:
Customer:
Inspection Date:
DAQP-TRQ
Temperature:
Humidity:
Tested by:
Place of inspection:
Test equipment
Model
S/N
Calibration Date
Calibrator
Digital Multimeter
Test
16.4.2.
1
2
3
3
2
16.4.3.
1
2
3
4
5
6
Gain Linearity
D
0 mV/V
D
5mV/V
D
2.5mV/V
D
-2.5mV/V
D
-5mV/V
Coarse Range Check
A
5mV/V
B
5mV/V
C
5mV/V
E
2mV/V
F
1mV/V
Gain low F
5mV/V
Calibration Procedure
Limits
lower
upper
-0.001V
4.999 V
2.495 V
-2.505 V
-5.005 V
0.001 V
5.001 V
2.505 V
-2.495 V
-4.995 V
0.50 V
1.15 V
2.4 V
3.85 V
3.85 V
1.90 V
0.70 V
1.35 V
2.6 V
4.15 V
4.15 V
2.10 V
Value
O.K
Page 103
DAQN-RTD-x Calibration
17 DAQN-RTD-x
17.1 Specifications
17.1.1 Gain-Accuracy DAQN-RTD-x
RDT-Type
Range
Accuracy*
% of F.S.
Output
Range
RTD-1
RTD-2
RTD-3
RTD-Spec
PT100 -30..170
PT100 -100..200
PT100 -50..600
Costumer defined
+/-0.1
+/-0.1
+/-0.1
+/-5v
+/-5v
+/-5v
+/-10mV
+/-10mV
+/-10mV
-2.5
Ohm -5
+2.5
+5
165
155
3: Zero
145
1: Zero
135
125
4:Linearization
115
105
2: Gain
95
85
-30 -20 -10
10
20
30
40
50
60
70
80
90
17.1.2 Mathematical
17.1.2.1 Temperature Conversion
F = (1.8 C) + 32
Page 104
DAQN-RTD-x Calibration
17.2 Schematic
DAQN-RTD
ZERO
LINEARITY
Supply Cable
GAIN
DEWE-DAQN-RTD
t /i n
Working Part
Si
p e c ia l O u
gn
al
Su
Ou
pp
Ca
b le
b le
D AQ P-D M M
t Ca
ly
Digital MULTIMETER
DAQ P-D M M
Direct
Page 105
DAQN-RTD-x Calibration
17.3 Parts List
o
o
o
o
o
DAQN-RDT-x Module
Slot : Any Kind of DEWE-Module-Slot is possible.
Multi Product Calibrator
High accuracy digital Multimeter.
Cables
o 4-Wire RDT Connector Cable
Pin assignment:
1 :Sense +
6: Input 2 :Input+
7:Sense 3 : N.C.
8: N.C.
4: GND
9: N.C.
5: N.C.
+
To Calibrator RTD Source
+
To Calibrator Sense
9 Pin SUB-D 9 Female, to 9 Pin SubD Female 1:1. (This is a Working Part and should be changed
after 100 calibrations, because oft the limited connector lifetime.)
Isolated Screwdriver for adjusting the Module
o
Page 106
DAQN-RTD-x Calibration
17.4.1.4 Linearity Check
o Set the Calibrator to the Zero Calibration temperature of the module, and check the output voltage. It should
be within a +/- 3mV tolerance.
RTD 1
RTD 2
RTD 3
116.33C
120.26C
425,91C
2.3170 V +/-3mV
2.3419 V +/-3mV
2.3263 V +/-3mV
For
Zero
-
V +/-
GAIN
adjust
+/-
Linearity
minimum (-5V)
C
-30
-100
-50
Input Range
Zero Calibration
half (0V) maximum (+5V)
Temprature
Output Voltage
C
C
C
Ohm
V
70
170
116.33
144.678
2.3170
50
200
120.26
146.148
2.3419
275
600
425,91
255.922
2.3263
Contact Dewetron
Page 107
DAQN-RTD-x Calibration
17.5 Calibration Report RTD-1
Module :
Serial Number:
Customer:
Inspection Date:
DAQP-RTD-1
Temperature:
Humidity:
Tested by:
Place of inspection:
Test equipment
Model
S/N
Calibration Date
Calibrator
Digital Multimeter
Test
17.4.1.1
17.4.1.1
17.4.1.3
17.4.1.4
Maximum
170C
Minimum
-30C
Half Input Range
70C
Linearity Check
116.33C
Calibration Procedure
Limits
lower
upper
4.9900 VDC
5.0100 VDC
-5.0100 VDC
-4.9900 VDC
-0.0030 VDC
0.0030 VDC
2.3140 VDC
2.3200 VDC
Value
O.K
Page 108
DAQN-RTD-x Calibration
17.6 Calibration Report RTD-2
Module :
Serial Number:
Customer:
Inspection Date:
DAQP-RTD-2
Temperature:
Humidity:
Tested by:
Place of inspection:
Test equipment
Model
S/N
Calibration Date
Calibrator
Digital Multimeter
Test
17.4.1.1
17.4.1.1
17.4.1.3
17.4.1.4
Maximum
200C
Minimum
-100C
Half Input Range
50C
Linearity Check
120.26C
Calibration Procedure
Limits
lower
Upper
4.9900 VDC
5.010 VDC
-5.010 VDC
-4.9900 VDC
-0.0030 VDC
0.0030 VDC
2.3389 VDC
2.3449 VDC
Value
O.K
Page 109
DAQN-RTD-x Calibration
17.7 Calibration Report RTD-3
Module :
Serial Number:
Customer:
Inspection Date:
DAQP-RTD-3
Temperature:
Humidity:
Tested by:
Place of inspection:
Test equipment
Model
S/N
Calibration Date
Calibrator
Digital Multimeter
Test
17.4.1.1
17.4.1.1
17.4.1.3
17.4.1.4
Maximum
600C
Minimum
-50C
Half Input Range
275C
Linearity Check
425.91C
Calibration Procedure
Limits
lower
upper
4.9900 VDC
5.010 VDC
-5.010 VDC
-4.9900 VDC
-0.0030VDC
0.0030 VDC
2.3233VDC
2.3293 VDC
Value
O.K
Page 110
DAQN-TERM-x Calibration
18 DAQN-THERM
18.1 Specifications
18.1.1 Accuracy DAQN-TERM
DAQN- THERM
TC-TYPE
THERM-1
THERM-2
THERM-3
THERM-4
THERM-5
K(NiCr-Ni)
K(NiCr-Ni)
K(NiCr-Ni)
K(NiCr-Ni)
J(Fe-CuNi)
Range
C
-30..170
-30..370
0..1000
-100..1350
-100..760
Accuracy*
% of F.S.
+/-0.5
+/-0.5
+/-0.5
+/-0.5
+/-0.5
Output Range
V
+/-5
+/-5
+/-5
+/-5
+/-5
maximum
Max Output Error*
uncertainty
mVDC
+/-1.0C
+/-50
+/-2.0C
+/-50
+/-5C
+/-50
+/-7.25C
+/-50
+/-4.3
+/-50
*Only approximate
Value, it depends on the
Seebk voltage
DAQN-THERM
transfer funktion
0V
Module
Output
Voltage
-5V
-30C
70
70.99C
170
Input Range
Output Zero
Calibration Point
18.1.2 Mathematical
18.1.2.1 Temperature Conversion
F = (1.8 C) + 32
Page 111
DAQN-TERM-x Calibration
Multi Product Calibrator
TC-OUT
DAQN-THERM
ZERO
GAIN
Supply Cable
Cold
Junction
Compensation
DEWE-DAQN-THERM-x
t/in
Working Part
Si
S p e c ia l O u
gn
al
Su
Ou
pp
b le
a b le
Ca
D AQ P-D M M
tC
ly
Any DEWE-SLot
Digital MULTIMETER
DAQ P-DM M
Direct
Schematic
Page 112
DAQN-TERM-x Calibration
18.2 Parts List
o
o
o
o
o
DAQN-RDT-x Module
Slot : Any Kind of DEWE-Module-Slot is possible.
Multi Product Calibrator
High accuracy digital Multimeter.
Cables
o Thermocouple Compensation Cable Type K or J
o 9 Pin SUB-D 9 Female, to 9 Pin SubD Female 1:1. (This is a Working Part and should be changed
after 100 calibrations, because oft the limited connector lifetime.)
Pin for adjusting the Module. (The potentiometers are on the upper side of the module, so it isnt possible to
adjust it with a screwdriver.)
Page 113
DAQN-TERM-x Calibration
18.3.1.4 Offset Adjustment
o
o
o
o
o
o
Page 114
DAQN-TERM-x Calibration
18.4 Calibration Report DAQN-THERM-1
Module :
Serial Number:
Customer:
Inspection Date:
DAQP-THERM-1
Temperature:
Humidity:
Tested by:
Place of inspection:
Test equipment
Model
S/N
Calibration Date
Calibrator
Digital Multimeter
Test
17.4.1.1
17.4.1.1
18.3.1.3
Minimum
-30C
Maximum
170C
Linearity
70C
Calibration Procedure
Limits
lower
upper
-5.050 VDC
-4.950 VDC
4.950 VDC
5.050 VDC
-0.050VDC
0.050 VDC
Value
O.K
Page 115
DAQN-TERM-x Calibration
18.5 Calibration Report DAQN-THERM-____
Module :
Serial Number:
Customer:
Inspection Date:
DAQP-THERM-1
Temperature:
Humidity:
Tested by:
Place of inspection:
Test equipment
Model
S/N
Calibration Date
Calibrator
Digital Multimeter
Calibration Procedure
Limits
lower
upper
Test
17.4.1.1
Minimum
17.4.1.1
Maximum
18.3.1.3
Linearity
-5.050 VDC
-4.950 VDC
4.950 VDC
5.050 VDC
-0.050VDC
0.050 VDC
Value
O.K
Page 116
DAQN-POT Calibration
19 DAQN-POT
19.1 Specifications
19.1.1 Accuracy DAQN-POT
Range
%
0..100
Accuracy
% of F.S.
+/-0.1
Output Range
V
+/-5
19.2 Schematic
+
RDT-Input
-
RDT-Input Cable
+
RDT-Sens
DAQN-POT
+EXCITATION
Supply Cable
-EXCITATION
ZERO
GAIN
DEWE-DAQN-POT
t /i n
Working Part
ly
a b le
b le
DAQP-DMM
Ca
ut C
pp
lO
Su
na
S ig
p e c ia l O u
Digital MULTIMETER
DAQP-DM M
Direct
Page 117
DAQN-POT Calibration
19.3 Parts List
o
o
o
o
DAQN-pot Module
Slot : Any Kind of DEWE-Module-Slot is possible.
High accuracy digital Multimeter.
Cables
o RTD Input Cable
Pin assignment:
1 :Excitation 6: Excitation+
2 :Input
7:NC
3 : GND
8: N.C.
4: N.C.
9: N.C.
5: N.C.
+
RDT Input Cable
Warning this cable is designed for the DAQN-RDT Module, so the wire names doesnt fit.
9 Pin SUB-D 9 Female, to 9 Pin SubD Female 1:1. (This is a Working Part and should be changed
after 100 calibrations, because oft the limited connector lifetime.)
Page 118
DAQN-POT Calibration
19.4.2 Module Adjustment (only if the calibration check dont succeed)
19.4.2.1 Positive Excitation Voltage Adjustment
o
o
o
Page 119
DAQN-POT Calibration
19.5 Calibration Report DAQN-POT
Module :
Serial Number:
Customer:
Inspection Date:
DAQN-POT
Temperature:
Humidity:
Tested by:
Place of inspection:
Test equipment
Model
S/N
Calibration Date
Digital Multimeter
Test
17.4.1.1
19.4.1.2
19.4.1.3
Minimum
0%
Maximum
100%
Half Input Range
50%
Calibration Procedure
Limits
lower
upper
-5.0100 VDC
-4.9900VDC
4.9900VDC
5.0100 VDC
-0.010VDC
0.010 VDC
Value
O.K
Page 120
PAD-V8-P Calibration
20 PAD-V8-P
20.1 Specifications
20.1.1 DC-Accuracy PAD-V8-P
Input Range
Specified Accuracy
Output Format
+/- V
+/- % of reading
+/- V
(High Precision Mode)
50
0.02
500
7Digits ASCII
10*
0.02
500
7Digits ASCII
1*
0.02
500
7Digits ASCII
0.5*
0.02
500
7Digits ASCII
0.1*
0.02
500
7Digits ASCII
*this ranges are only software limited sub ranges of the 50V hardware range, and implemented to be compatible to
the older PAD-V8 modules.
20.2 Schematic
DAQ P-DM M
PAD-V8-P
1:Channel 0 +
2:Channel 0 3:Channel 1 +
4:Channel 1 5:Channel 2 +
6:Channel 2 7:Channel 3 +
8:Channel 3 9:Channel 4 +
10:Channel 4 11:Channel 5 +
12:Channel 5 13:Channel 6 +
14:Channel 6 15:Channel 7 +
16:Channel 7 17:Digital Input 1
18:Digital Input 2
19:Digital Input 3
20:+ 12 VDC
21:Reset
22:GND
23:Reserved
24:Reserved
25:Reserved
DSUB 25 Connector
PAD-V8-P
E-M0243E Printing Version 1.6.9 FEB, 2008
Page 121
PAD-V8-P Calibration
20.3 Parts List
o
o
o
o
o
o
PAD-V8-P Module
Slot : Any Kind of DEWE-Module-Slot with Serial Module Interface is possible.
Multi Product Calibrator (Fluke5500A recommended)
Cables
o Connector Cable DSUB25 to Banana Plug (see 20.2 Schematic).
Computer with Serial Interface (RS232 or RS485).
DEWEConfig 5.4.2 or later, which is included on the DEWE-System CD-Rom. You will find the actual version
under www.dewetron.com .
o
o
Page 122
PAD-V8-P Calibration
20.4.1 Check the Calibration
o
o
o
Terminal Command
Slot Number
Select Pad
Send ${address in hex }A to the module or press two times get data .
You should receive the following string:
>+00000.00+00000.01+00000.00-00000.01-00000.02+00000.00+00000.01-00000.00 . This are the input
voltages in mV from all 8 channels.
Every Value must be between-00000.50mVDC and +00000.50mVDC .
o
o
o
o
It is not necessary calibrate the other ranges, because the analogue circuit of the module is the same for all
input ranges.
If one channel fails the test or scrap through, perform the adjustment step .
o
o
Apply 0.00000VDC to the 8 Input channels. We recommend that you short circuit the banana plugs, of the PAD
input cable, instead of setting the calibrator to zero.
Wait 5 seconds.
Send the command ${address}1 . (older zero calibration command)
The response should be !{address}
If the response is ?{address}old command send %{address}ZA to the module ( set all channels to zero).
The response should be !{address}
Apply 10.0000VDC to the 8 Input channels, and wait 5 seconds.
Send the command ${address}0 . (older span calibration command)
If the response is ?{address}old command send %{address}SA ( span calibration to all channels).
The response should be !{address}
Check the calibration again.
Page 123
PAD-V8-P Calibration
20.5 Calibration Report
Module :
Serial Number:
Customer:
Inspection Date:
PAD-V8-P
Temperature:
Humidity:
Tested by:
Place of inspection:
Test equipment
Model
S/N
Calibration Date
Calibrator
Calibration Procedure
Limits
lower
upper
Test
Value
O.K
Zero calibration
CH0
CH1
CH2
CH3
CH4
CH5
CH6
CH7
Gain calibration
CH0
CH1
CH2
CH3
CH4
CH5
CH6
CH7
Linearity Check
CH0
CH1
CH2
CH3
CH4
CH5
CH6
CH7
-0.0005
-0.0005
-0.0005
-0.0005
-0.0005
-0.0005
-0.0005
-0.0005
+0.0005
+0.0005
+0.0005
+0.0005
+0.0005
+0.0005
+0.0005
+0.0005
+49.9895
+49.9895
+49.9895
+49.9895
+49.9895
+49.9895
+49.9895
+49.9895
+50.0105
+50.0105
+50.0105
+50.0105
+50.0105
+50.0105
+50.0105
+50.0105
-50.0105
-50.0105
-50.0105
-50.0105
-50.0105
-50.0105
-50.0105
-50.0105
-49.9895
-49.9895
-49.9895
-49.9895
-49.9895
-49.9895
-49.9895
-49.9895
Page 124
PAD-TH8-P Calibration
21 PAD-TH8-P & PAD-CB8-x-P
21.1 Specifications
21.1.1 DC-Accuracy PAD-TH8-P
Input Range
Specified Accuracy
+/- V
+/- % of reading
+/- V
50
0.02
200
10
0.02
200
1
0.02
200
.5
0.02
200
.1
0.02
200
Thermo Couple
Range
+/- C *
Type J: 210..1200C -210C-100C
1
-100C..150C
0.3
150C..400C
0.4
400C..1200C
1
Type K: 270..1372C -270C-25C
1
-25C..120C
0.4
120C..400C
0.6
400C..1327C
1
Type T: 270..400C -210C-150C
1
-150C..400C
0.4
Output Format
(High Precision Mode)
7Digits ASCII
7Digits ASCII
7Digits ASCII
7Digits ASCII
7Digits ASCII
7Digits ASCII
7Digits ASCII
7Digits ASCII
7Digits ASCII
7Digits ASCII
7Digits ASCII
7Digits ASCII
7Digits ASCII
7Digits ASCII
7Digits ASCII
*Only in combination with a
Connector block PAD-CB8-xP
21.2 Schematic
19.2.1 Voltage Calibration
D AQ P -D M M
PA D-V8-P
1:Channel 0 +
2:Channel 0 3:Channel 1 +
4:Channel 1 5:Channel 2 +
6:Channel 2 7:Channel 3 +
8:Channel 3 9:Channel 4 +
10:Channel 4 11:Channel 5 +
12:Channel 5 13:Channel 6 +
14:Channel 6 15:Channel 7 +
16:Channel 7 17:Sensor Identifications 1
18:Sensor Identifications 2
19:Sensor Identifications 3
20:+ 12 VDC
21:Reset
22:GND
23:CJC
24:CJC
25:CJC
DSUB 25 Connector
PAD-TH8-P
E-M0243E Printing Version 1.6.9 FEB, 2008
Page 125
PAD-TH8-P Calibration
21.2.1 Temperature Calibration (with PAD-CB8-x-P)
TC-OUT
D AQ P-D M M
PA D-V8-P
CH0
CH1
CH2
CH3
CH4
CH5
CH6
CH7
PAD-CB8-x-P
21.2.2 Parts List
o
o
o
o
o
o
o
PAD-TH8-P Module
Slot : Any Kind of DEWE-Module-Slot with Serial Module Interface is possible.
Multi Product Calibrator (Fluke5500A)
Cables
o Connector Cable DSUB25 to Banana Plug (see 19.2.1 Schematic).
o Thermo couple Cable according to the thermo couple type you are using.
Computer with Serial Interface (RS232 or RS485).
DEWEConfig 5.4.2 or later, which is included on the DEWE-System CD-Rom. You will find the actual version
under www.dewetron.com
PAD-CB8-X-P
Page 126
PAD-TH8-P Calibration
21.3 Calibration Instruction
19.3.1 Configuring the Module
o
o
o
o
o
o
o
Page 127
PAD-TH8-P Calibration
21.3.1 Calibration Check
o
o
o
Terminal Command
Slot Number
Select Pad
Send ${address in hex }A to the module or press two times get data .
You should receive the following string:
>+0000.000+0000.001+0000.000-0000.001-0000.002+0000.000+0000.001-0000.000 . This are the input
voltages in mV from all 8 channels.
Every value must be between-0000.200mVDC and +0000.200mVDC .
o
o
o
o
It is not necessary calibrate the other ranges, because the analogue circuit of the module is the same for all
input ranges.
If one channel fails the test or scrap through, perform the adjustment step .
o
o
Page 128
PAD-TH8-P Calibration
o
o
o
Example: >+00000.44+01200.00+01200.00+01200.00+01200.00+01200.00+01200.00+01200.00
If there is an offset temperature higher than +/-0.1C continue with the point Adjust CJC Offset 21.3.3.1 , if not
continue with 21.3.3.2 .
Page 129
PAD-TH8-P Calibration
21.3.3.2 Check all thermocouple input channels
o
Apply the following temperatures to all input channels one by one, check the measured temperature. It
should not exceed the specified limits.
Connector Block
Type
J
-210C..1200C
K
-270C..1372C
T
-270C..400C
Calibrator
Source
C
-200
0
1100
-200
0
1200
-200
0
380
Page 130
PAD-TH8-P Calibration
21.4 Calibration Report
Module :
Serial Number:
Customer:
Inspection Date:
Temperatur
e:
Humidity:
Tested by:
Place of inspection:
PAD-TH8-P
Test equipment
Model
S/N
Calibration Date
Calibrator
Calibration Procedure
Limits
lower
upper
Test
21.3.3.2
Value
O.K
Zero calibration
CH0
CH1
CH2
CH3
CH4
CH5
CH6
CH7
-0.200 mV
-0.200 mV
-0.200 mV
-0.200 mV
-0.200 mV
-0.200 mV
-0.200 mV
-0.200 mV
+0.200 mV
+0.200 mV
+0.200 mV
+0.200 mV
+0.200 mV
+0.200 mV
+0.200 mV
+0.200 mV
+999.600mV
+999.600mV
+999.600mV
+999.600mV
+999.600mV
+999.600mV
+999.600mV
+999.600mV
+1000.400mV
+1000.400mV
+1000.400mV
+1000.400mV
+1000.400mV
+1000.400mV
+1000.400mV
+1000.400mV
+1449.510mV
+1449.510mV
+1449.510mV
+1449.510mV
+1449.510mV
+1449.510mV
+1449.510mV
+1449.510mV
+1450.490mV
+1450.490mV
+1450.490mV
+1450.490mV
+1450.490mV
+1450.490mV
+1450.490mV
+1450.490mV
Gain calibration
CH0
CH1
CH2
CH3
CH4
CH5
CH6
CH7
Linearity Check
CH0
CH1
CH2
CH3
CH4
CH5
CH6
CH7
Page 131
PAD-TH8-P Calibration
21.5 Calibration Report Connector Block Type J
Module :
Serial Number:
Customer:
Inspection Date:
PAD-CB8-J-P
Temperature:
Humidity:
Tested by:
Place of inspection:
Test equipment
Model
S/N
Calibration Date
Calibrator
Calibration Procedure
Limits
lower
upper
Test
21.3.3.2
Value
O.K
0C
CH0
CH1
CH2
CH3
CH4
CH5
CH6
CH7
-0.40 C
-0.40 C
-0.40 C
-0.40 C
-0.40 C
-0.40 C
-0.40 C
-0.40 C
+0.40 C
+0.40 C
+0.40 C
+0.40 C
+0.40 C
+0.40 C
+0.40 C
+0.40 C
-201.0C
-201.0C
-201.0C
-201.0C
-201.0C
-201.0C
-201.0C
-201.0C
-199.0C
-199.0C
-199.0C
-199.0C
-199.0C
-199.0C
-199.0C
-199.0C
1099.0C
1099.0C
1099.0C
1099.0C
1099.0C
1099.0C
1099.0C
1099.0C
1101.0C
1101.0C
1101.0C
1101.0C
1101.0C
1101.0C
1101.0C
1101.0C
-200C
CH0
CH1
CH2
CH3
CH4
CH5
CH6
CH7
1100C
CH0
CH1
CH2
CH3
CH4
CH5
CH6
CH7
Page 132
PAD-TH8-P Calibration
21.6 Calibration Report Connector Block Type K
Module :
Serial Number:
Customer:
Inspection Date:
PAD-CB8-K-P
Temperature:
Humidity:
Tested by:
Place of inspection:
Test equipment
Model
S/N
Calibration Date
Calibrator
Calibration Procedure
Limits
lower
upper
Test
21.3.3.2
Value
O.K
0C
CH0
CH1
CH2
CH3
CH4
CH5
CH6
CH7
-0.40 C
-0.40 C
-0.40 C
-0.40 C
-0.40 C
-0.40 C
-0.40 C
-0.40 C
+0.40 C
+0.40 C
+0.40 C
+0.40 C
+0.40 C
+0.40 C
+0.40 C
+0.40 C
-201.0C
-201.0C
-201.0C
-201.0C
-201.0C
-201.0C
-201.0C
-201.0C
-199.0C
-199.0C
-199.0C
-199.0C
-199.0C
-199.0C
-199.0C
-199.0C
1199.0C
1199.0C
1199.0C
1199.0C
1199.0C
1199.0C
1199.0C
1199.0C
1201.0C
1201.0C
1201.0C
1201.0C
1201.0C
1201.0C
1201.0C
1201.0C
-200C
CH0
CH1
CH2
CH3
CH4
CH5
CH6
CH7
1200C
CH0
CH1
CH2
CH3
CH4
CH5
CH6
CH7
Page 133
PAD-TH8-P Calibration
21.7 Calibration Report Connector Block Type T
Module :
Serial Number:
Customer:
Inspection Date:
PAD-CB8-T-P
Temperature:
Humidity:
Tested by:
Place of inspection:
Test equipment
Model
S/N
Calibration Date
Calibrator
Calibration Procedure
Limits
lower
upper
Test
21.3.3.2
Value
O.K
0C
CH0
CH1
CH2
CH3
CH4
CH5
CH6
CH7
-0.40 C
-0.40 C
-0.40 C
-0.40 C
-0.40 C
-0.40 C
-0.40 C
-0.40 C
+0.40 C
+0.40 C
+0.40 C
+0.40 C
+0.40 C
+0.40 C
+0.40 C
+0.40 C
-201.0C
-201.0C
-201.0C
-201.0C
-201.0C
-201.0C
-201.0C
-201.0C
-199.0C
-199.0C
-199.0C
-199.0C
-199.0C
-199.0C
-199.0C
-199.0C
379.6C
379.6C
379.6C
379.6C
379.6C
379.6C
379.6C
379.6C
380.4C
380.4C
380.4C
380.4C
380.4C
380.4C
380.4C
380.4C
-200C
CH0
CH1
CH2
CH3
CH4
CH5
CH6
CH7
380C
CH0
CH1
CH2
CH3
CH4
CH5
CH6
CH7
Page 134
PAD-CB8-RTD Calibration
22 PAD-CB8-RTD
22.1 Specifications
Input Range
Range
Code
70
Resistance
PT100
a=0.00385
71
PT200
a=0.00385
721
PT500
a=0.00385
73
PT100
a=0.003916
74
Ni120
75
PT1000*
a=0.00385
72
PT2000*
a=0.00385
73
Input
0 to 500 Ohm
>500 to 1000 Ohm
-200C to 100C
>100C to 400C
>400C to 800C
-200C to 100C
>100C to 400C
>400C to 800C
-200C to 100C
>100C to 400C
>400C to 800C
-200C to 100C
>100C to 400C
>400C to 800C
-200C to 100C
>100C to 400C
-200C to 100C
>100C to 400C
>400C to 800C
-200C to 100C
>100C to 400C
>400C to 800C
Constant Current
[A]
1250
1250
1250
1250
1250
1250
200
200
Accuracy
0.25 Ohm
0.5 Ohm
0.25 C
0.50 C
0.80 C
0.25 C
0.50
0.80
0.25
0.50
0.80
0.25
0.50
0.80
0.25
0.50
0.25
0.50
0.80
0.25
0.50
0.80
Calibration
factors
linearization
table
8 analog
Channels
digital
Interface
PAD-TH8-P
24 bit
ADC
linearization
RS 485
Converter
Data acquisition
software
Micro controller
Page 135
PAD-CB8-RTD Calibration
22.3 Schematic
RTD-Input Cabl e
CH 2
CH 3
CH 4
CH 5
CH 6
CH 7
T
8R
B
-C
D
A
P
CH 0
Page 136
PAD-CB8-RTD Calibration
22.4.2 Calibration Check
o
o
o
o
Apply the following resistances to the module in 4-wire technique, and check if the measured values are
within the specifications.
The testing of the resistance range guarantees that the analog amplifier is working.
Change the input range to PT100 (terminal command: %{address}IA71[Cr]) and apply 20C to the module.
Check if the measured value is between the limits.
Input Range
Resistance
PT100 a=0.00385
Calibrator output
10 Ohm
100 Ohm
990 Ohm
-100C
20C
790C
Lower Limit
9.75
99.75
989.50
-100.2C
19.75C
789.20C
Upper Limit
10.25
100.25
990.50
-99.25C
20.25C
790.80C
Page 137
PAD-CB8-RTD Calibration
Typical Calibration Certificate
Module :
PAD-CB8-RTD
Temperature:
Serial Number:
Humidity:
Customer:
Inspection Date:
Tested by:
Place of inspection:
Test equipment
Model
S/N
Calibration Date
Calibrator
PAD-TH8-P
Test
Channel 0
Channel 1
Channel 2
Channel 3
Channel 4
Channel 5
Channel 6
Channel 7
10 Ohm
100 Ohm
990 Ohm
PT 100 -100 C
PT 100
20 C
PT 100 790 C
10 Ohm
100 Ohm
990 Ohm
PT 100 -100 C
PT 100
20 C
PT 100 790 C
10 Ohm
100 Ohm
990 Ohm
PT 100 -100 C
PT 100
20 C
PT 100 790 C
10 Ohm
100 Ohm
990 Ohm
PT 100 -100 C
PT 100
20 C
PT 100 790 C
10 Ohm
100 Ohm
990 Ohm
PT 100 -100 C
PT 100
20 C
PT 100 790 C
10 Ohm
100 Ohm
990 Ohm
PT 100 -100 C
PT 100
20 C
PT 100 790 C
10 Ohm
100 Ohm
990 Ohm
PT 100 -100 C
PT 100
20 C
PT 100 790 C
10 Ohm
100 Ohm
990 Ohm
PT 100 -100 C
PT 100
20 C
PT 100 790 C
Calibration Procedure
Limits
lower
9.75
99.75
989.50
-100.2C
19.75C
789.20C
9.75
99.75
989.50
-100.2C
19.75C
789.20C
9.75
99.75
989.50
-100.2C
19.75C
789.20C
9.75
99.75
989.50
-100.2C
19.75C
789.20C
9.75
99.75
989.50
-100.2C
19.75C
789.20C
9.75
99.75
989.50
-100.2C
19.75C
789.20C
9.75
99.75
989.50
-100.2C
19.75C
789.20C
9.75
99.75
989.50
-100.2C
19.75C
789.20C
Value
upper
10.25
100.25
990.50
-99.25C
20.25C
790.80C
10.25
100.25
990.50
-99.25C
20.25C
790.80C
10.25
100.25
990.50
-99.25C
20.25C
790.80C
10.25
100.25
990.50
-99.25C
20.25C
790.80C
10.25
100.25
990.50
-99.25C
20.25C
790.80C
10.25
100.25
990.50
-99.25C
20.25C
790.80C
10.25
100.25
990.50
-99.25C
20.25C
790.80C
10.25
100.25
990.50
-99.25C
20.25C
790.80C
O.K
Page 138
PAD-RTD-3 Calibration
23 PAD-RTD-3
23.1 Specifications
23.1.1 Accuracy PAD-RTD-3
Input Range
Pt 100 (a=0.00385)
-100100C
0600
Code
20
23
Specified Accuracy*
+/- % of reading
+/- C
0.25%
0.5
0.1%
0.6
Calibration Resistor
Zero
Span
50.03
200.05
50.07
350.15
* Only valid for 4 Wire Measurement and
latest revision ( since 2003 )
23.2 Schematic
Normal
Aux
RTD
Sense
D A Q P-D M M
CH0
PA D-RT D -3
CH1
CH2
Page 139
PAD-RTD-3 Calibration
23.2.1 Parts List
o
o
o
o
PAD-RTD-3 Module
Slot : Any Kind of DEWE-Module-Slot with Serial Module Interface is possible.
Multi Product Calibrator (Fluke5500A)
Cables
o RTD-Input cable
2
o PAD-RTD-3 split cable ( not included in the cal-kit)
CH0
1
9 Pol DSUB
female
CH1
9 Pol DSUB
female
CH3
9 Pol DSUB
female
o
o
7
6
2
1
7
6
2
1
7
6
PAD-RTD-3
25 Pol DSUB
male
1
14
2
15
3
16
4
17
5
18
6
19
7
20
8
21
9
22
10
23
11
24
12
25
13
o
o
o
o
address 00h
select Pad
get data
100C Range
Page 140
PAD-RTD-3 Calibration
o
Page 141
PAD-RTD-3 Calibration
23.3.1 Calibration Check
o
o
Connect the Calibrator via the split cable and the RTDto channel 0 . (see 23.2 )
Set the Calibrator to
o 95.00C
o RTD Source
o 4-Wire
o
o
o
o
o
o
o
o
o
o
o
If one Channel fails the test or scrap through, perform the adjustment step .
Input cable
o
o
o
o
o
o
o
o
Page 142
PAD-RTD-3 Calibration
23.4 Calibration Report
Module :
Serial Number:
Customer:
Inspection Date:
PAD-RTD-3
Temperature:
Humidity:
Tested by:
Place of inspection:
Test equipment
Model
S/N
Calibration Date
Calibrator
Test
Channel 0 (100C Range)
-95.00C
00.00C
95.00C
Channel 0 ( 600C Range)
10.00C
300.00C
550.00C
Channel 1 (100C Range)
-95.00C
00.00C
95.00C
Channel 1 ( 600C Range)
10.00C
300.00C
550.00C
Channel 2 (100C Range)
-95.00C
00.00C
95.00C
Channel 2 ( 600C Range)
10.00C
300.00C
550.00C
Calibration Procedure
Limits
lower
upper
-95.50
-0.50
94.50
-94.50
0.50
95.50
9.40
299.40
549.40
10.60
300.60
550.60
-95.50
-0.50
94.50
-94.50
0.50
95.50
9.40
299.40
549.40
10.60
300.60
550.60
-95.50
-0.50
94.50
-94.50
0.50
95.50
9.40
299.40
549.40
10.60
300.60
550.60
Value
O.K
Page 143
MDAQ-V-x Calibration
24 MDAQ-V Module
24.1 Specifications
24.1.1 DC-Accuracy MDAQ-V without Software correction
MDAQ-V-10
MDAQ-V-100
MDAQ-V-250
Range
Accuracy
% of F.S.
MDAQ-V Output
voltage
Max Output
Error (VDC)
0.5V
1V
5V
10V
5V
10V
50V
100V
12.5V
25V
125V
250V
0.25%
0.25%
0.15%
0.15%
0.25%
0.25%
0.15%
0.15%
0.7%
0.7%
0.7%
0.7%
5.000 V
10.000V
5.000V
10.000V
5.000V
10.000V
5.000V
10.000V
5.000V
10.000V
5.000V
10.000V
12.5 mV
25 mV
7.5 mV
15 mV
12.5 mV
25 mV
7.5 mV
15 mV
35mV
70mV
35mV
70mV
1.25 mV
2.5 mV
7.5 mV
15 mV
12.5 mV
25 mV
75 mV
150 mV
87.5mV
175mV
875mV
1.750mV
0.5V
:
:
100V
DEWE-ORION-1624
PCI-6071-E
PCI-6055-E
MDAQ-V
10V
AD-Card
Correction
Values
DEWESoft 6
Software
Correction
88888
RS485
Over All Accuracy:
Page 144
MDAQ-V-x Calibration
24.1.2.2 AD-Card Overview
If the AD-Card type is unknown it can be determined in the DEWESoft Hardware Setup.
AD Card Accuracy
AD-Card
Range
Orion-1624
PCI-6071
PCI-6052
NI-625x
NI-628x
% of reading
% of range
Absolute
Maximum*
5V
0.06
0.01
0.00337 V
10 V
0.06
0.005
0.00624 V
5V
0.03
0.064
0.005193 V
10 V
0.07
0.0638
0.014366 V
5V
0.007
0.00952
0.000875 V
10 V
0.037
0.00947
0.004744 V
5V
0.0070
0.002
0.00101 V
10 V
0.0060
0.002
0.00192 V
5V
0.0045
0.0008
0.00051 V
10 V
0.0040
0.0008
0.00098 V
All A-D card accuracy specifications for 3rd party products have been taken from the manufacturers published data
sheets and are subject to change without notice.
24.1.2.3 AC Accuracy Specification
Frequency
MDAQ-V-10
0..500 Hz
500..1kHz
1..5kHz
5..10kHz
MDAQ-V-100
0..1KHz
1..5kHz
5..10kHz
10..50kHz
MDAQ-V-250
0..1KHz
1..5kHz
5..10kHz
10..50kHz
0.03%
0.05%
0.15%
3%
Page 145
MDAQ-V-x Calibration
Absolute Accuracy specifications for MDAQ-V-xx depending on AD-Board
DEWE
ORION-1624
ORION-xx16 series
PCI
6071E
Absolute
Offset
Absolute
Offset
Absolute
Offset
Absolute
Offset
Absolute
Offset
mV
mV
mV
mV
mV
mV
mV
mV
mV
mV
mV
0.2
0.637
0.215
0.465
0.47
0.8193
0.1976
0.3875
0.16
0.401
0.154
0.351
0.35
1.224
0.415
0.915
0.938
2.0366
0.3947
1.0744
0.32
0.792
0.308
0.698
MDAQ-V
AD-Input
Range
Range
Offset
mV
0.5 V*
5V
1V
10 V
PCI 6052E
NI-625x
NI-628x
Absolute
5 V*
5V
6.37
2.15
4.65
4.7
8.193
1.976
3.8745
1.6
4.01
1.54
3.51
10 V
10 V
3.5
12.24
4.15
9.15
9.38
20.366
3.947
10.744
3.2
7.92
3.08
6.98
50 V*
5V
20
63.7
21.5
46.5
47
81.93
19.76
38.745
16
40.1
15.4
35.1
100 V
10 V
35
122.4
41.5
91.5
93.8
203.66
39.47
107.44
32
79.2
30.8
69.8
12.5 V*
5V
15.925
5.375
11.625
11.75
20.483
4.94
9.6863
10.025
3.85
8.775
25 V
10 V
8.75
30.6
10.375
22.875
23.45
50.915
9.8675
26.86
19.8
7.7
17.45
125 V*
5V
50
159.25
53.75
116.25
117.5
204.83
49.4
96.863
40
100.25
38.5
87.75
250 V
10 V
87.5
306
103.75
228.75
234.5
509.15
98.675
268.6
80
198
77
174.5
24.2 Schematic
A few Calibrators allow the connection of all MDAQ Inputs together to the Calibrator source.
Also the Fluke 5500 for Voltages except in the lowers ranges.
na
nan
-Ba
BNC
MDAQ-V-xxx
CH0 CH2
CH4
CH6
CH8
CH10
CH14
CH1 CH3
CH5
CH7
CH9
CH11
CH15
Page 146
MDAQ-V-x Calibration
24.3 Parts List
o
o
o
MDAQ-V-x Module
Multi Product Calibrator
Cables: Safety Banana Plugs (red ,black), to BNC.
By clicking on the slot 0 icon it is easy to Copy and Paste one channel setting to all the other.
Apply 0.0000V to the input. The measured Value has to be 0.0000V the maximum offset defined in table 24.1.2.4
For example a system with a PCI-6071-E together with an MDAQ-V-100 in the 10V range has to be between the
maximum offset error of 9.38mV
Page 147
MDAQ-V-x Calibration
Check the Points 24.4.1; 24.4.2 for all selectable Ranges and all 16 Channels.
24.4.3 AC Calibration
Set the MDAQ-V-100 to the 10V Input Range. Apply a 7.000VRMS (9.899 VP) , 1kHz sine wave signal via a shielded
cable to the input.
Set the AD Sampling rate to the maximum. Be sure that it is more than the double of the calibration frequency.
Otherwise aliasing effects take influence on your measurement.
AC-RMS
Slow (1s)
Caution: DeweSoft indicates an overload when the signal is more than 98% of
the range. This does not have any influence on the calibration.
24.4.3.1 Recommended Calibration Points and specified Limits
AC-Accuracy
Input Signal
MDAQV -10
MDAQV -100
MDAQV -250
Input
Range
Input
Voltage
Frequency
AC
Accuracy
DEWEPCI-6071- PCI-6052ORION1624
E
E
NI-625x
NI-628x
VRMS
kHz
% of F.R.
10
0.5
0.03
0.0070
0.0080
0.0056
0.0034
0.0033
10
0.05
0.0090
0.0100
0.0076
0.0054
0.0053
10
0.15
0.0190
0.0200
0.0176
0.0154
0.0153
10
10
0.3040
0.3050
0.3026
0.3004
0.3003
10
0.03
0.0070
0.0080
0.0056
0.0034
0.0033
10
0.05
0.0090
0.0100
0.0076
0.0054
0.0053
10
10
0.15
0.0190
0.0200
0.0176
0.0154
0.0153
10
50*
0.3040
0.3050
0.3026
0.3004
0.3003
25
15
0.03
0.0161
0.0182
0.0131
0.0084
0.0081
25
15
0.05
0.0211
0.0232
0.0181
0.0134
0.0131
25
15
10
0.15
0.0461
0.0482
0.0431
0.0384
0.0381
25
15
50*
0.7586
0.7607
0.7556
0.7509
0.7506
*Some AD-Cards are not able to run all 16 Channels at sample rates higher then 50kHz. In that case skip the 50kHz
test point, or activate channel by channel.
Page 148
MDAQ-V-x Calibration
24.5 Calibration Report Example for MDAQ-V-100
Module :
Serial Number:
Customer:
Inspection Date:
MDAQ-V-100
Temperature:
Humidity:
Tested by:
Place of inspection:
Test equipment
Model
S/N
Calibrator
AD-Card
Test description
Channel 0
DC
5V Range
50V Range
10V Range*
100V Range*
AC
Channel 1
DC
10V Range
5V Range
50V Range
10V Range*
100V Range*
AC
Channel 2
DC
10V Range
5V Range
50V Range
10V Range*
100V Range*
AC
10V Range
0.000V
4.900V
-4.900V
0.000V
49.000V
-49.000V
0.000V
9.900V
-9.900V
0.000V
99.000V
-99.000V
7.000V 1kHz
7.000V 5kHz
7.000V 10kHz
7.000V 50kHz
0.000V
4.900V
-4.900V
0.000V
49.000V
-49.000V
0.000V
9.900V
-9.900V
0.000V
99.000V
-99.000V
7.000V 1kHz
7.000V 5kHz
7.000V 10kHz
7.000V 50kHz
0.000V
4.900V
-4.900V
0.000V
49.000V
-49.000V
0.000V
9.900V
-9.900V
0.000V
99.000V
-99.000V
7.000V 1kHz
7.000V 5kHz
7.000V 10kHz
7.000V 50kHz
Calibration Date
Ni-6071-E
Calibration Procedure
Limits
Value
lower
upper
-0.0047 V
4.8918 V
-4.9082 V
-0.0470 V
48.918 V
-49.082 V
-0.0094 V
9.8181 V
-9.9204 V
-0.0938 V
98.796 V
-99.204 V
6.992 VRMS
6.990 VRMS
6.980 VRMS
6.595 VRMS
-0.0047 V
4.8918 V
-4.9082 V
-0.0470 V
48.918 V
-49.082 V
-9.38 mV
9.8181 V
-9.9204 V
-0.0938 V
98.796 V
-99.204 V
6.992 VRMS
6.990 VRMS
6.980 VRMS
6.595 VRMS
-0.0047 V
4.8918 V
-4.9082 V
-0.0470 V
48.918 V
-49.082 V
-9.38 mV
9.8181 V
-9.9204 V
-0.0938 V
98.796 V
-99.204 V
6.992 VRMS
6.990 VRMS
6.980 VRMS
6.595 VRMS
0.0047 V
4.9082 V
-4.8918 V
0.0470 V
49.082 V
-48.918 V
0.0094 V
9.9204 V
-9.8181 V
0.0938 V
99.204 V
-98.796 V
7.008 VRMS
7.010 VRMS
7.020 VRMS
7.305 VRMS
0.0047 V
4.9082 V
-4.8918 V
0.0470 V
49.082 V
-48.918 V
9.38 mV
9.9204 V
-9.8181 V
0.0938 V
99.204 V
-98.796 V
7.008 VRMS
7.010 VRMS
7.020 VRMS
7.305 VRMS
0.0047 V
4.9082 V
-4.8918 V
0.0470 V
49.082 V
-48.918 V
9.38 mV
9.9204 V
-9.8181 V
0.0938 V
99.204 V
-98.796 V
7.008 VRMS
7.010 VRMS
7.020 VRMS
7.305 VRMS
O.K
Warning
High Voltage!
Warning
High Voltage!
Warning
High Voltage!
Warning
High Voltage!
Warning
High Voltage!
Warning
High Voltage!
Page 149
MDAQ-SUB-V-200-BNC Calibration
25 MDAQ-SUB-V-200-BNC
25.1 Signal Path
RS485
AD- Boar d
Ch0 to Ch15
5V
Ch0 to Ch15
dig ital 5 V
in
areG
oftw
S
orectin
C
Ch0 to Ch15
Input
1 25mV to 200 V
Dewe so ft 6
BNC-OUT-5
Ch1 to Ch15
Con ditione d S igna l Out
5V
Ch0 to Ch15
5V
25.2 Specification
25.2.1 DC Accuracy
MDAQ-SUB-V-200-BNC
Software corrected
Analog out
*Input divider on
Range
% of reading
% of range
% of reading
% of range
0.125V
0.03
0.32
0.15
0.32
0.25V
0.03
0.16
0.15
0.16
0.5V
0.03
0.08
0.15
0.08
1V
0.03
0.04
0.15
0.04
1.25V
0.03
0.08
0.15
0.08
2.5V
0.03
0.04
0.15
0.04
5V
0.02
0.03
0.15
0.03
Source:
10V
0.02
0.03
0.15
0.03
2.5V*
0.06
0.32
0.25
0.32
5V*
0.06
0.16
0.25
0.16
10V*
0.06
0.08
0.25
0.08
20V*
0.06
0.04
0.25
0.04
25V*
0.03
0.08
0.25
0.08
50V*
0.03
0.04
0.25
0.04
100V*
0.02
0.03
0.25
0.03
200V*
0.02
0.03
0.25
0.03
Page 150
MDAQ-SUB-V-200-BNC Calibration
25.2.2 AC Accuracy
0Hz
>1
>10kHz
>10kHz
divider off
divider on
MDAQ-SUB-V-200-BNC
to 1kHz DC Accuracy
to 10kHz DC Accuracy + 0.2% of reading
to 50kHz DC Accuracy + 3% of reading
to 50kHz DC Accuracy + 10% of reading
AD-Board DC accuracies
AD-Card
Range
% of reading
% of range
Orion-1624
5V
0.0574
0.01
5V
0.02
0.013
Specification Source
5V
0.0093
0.001094
DE-D0421E DEWE-ORION-1624.
Printing version 1.0.6 Dec. 04, 2006
DE-M060304E DEWE-ORION-1616-10x
Printing version 1.0.2 November 24,2006
B/E/M/S Series Calibration
PCI-628x
5V
0.0072
0.00308
PCI-6071
5V
0.0314
0.0724
370937d-01 Jan07
PCI-6052
5V
0.0080
0.01
DT-9834 (USB)
5V
0.00
0.08
Orion-xx16 series
PCI-625x
AD-Board
AD-Board AC accuracies
additional Gain error
-3dB
% at
% at
Range
bandwidth
10kHz
50kHz
Orion-1624
5V
200kHz
0.4
Orion-xx16-10x
5V
100kHz
0.5
0.4
11
Orion-xx16-50x
5V
300kHz
1.5
Orion-0816-1000
5V
750kHz
0.3
PCI-625x
5V
1.7MHz
PCI-628x
5V
750kHz
0.3
PCI-6071
5V
1.6MHz
PCI-6052
5V
480kHz
0.6
DT-9834
5V
1.1MHz
PCI-622x
5V
700kHz
0.3
INFO:
Orion-1624 AC Accuracy:
Flatness + % of reading @200kS/sec
All other AD-Board AC accuracies are
calculated by using the specified
bandwidth as base for an 1st order low
pass filter. The filter attenuation on the
test frequencies is the AC gain accuracy
at these points.
All AD-Board accuracy specifications for 3rd party products have been taken from the manufacturers published data
sheets and are subject to change without notice.
E-M0243E Printing Version 1.6.9 FEB, 2008
Page 151
MDAQ-SUB-V-200-BNC Calibration
25.2.4 Absolute System DC Accuracies with multiplexed boards
System Accuracy MDAQ-SUB-V-200-x using DeweSoft 6.3 or later
Range
PCI-622x
PCI-625x
PCI-628x
PCI-6071
PCI-6052
DT-9834 (USB2)
max
offset
max
Error
max
offset
max
Error
max
offset
max
Error
max
offset
max Error
max
offset
max
Error
max
offset
max
Error
mV
mV
mV
mV
mV
mV
mV
mV
mV
mV
mV
mV
0.125
0.40
0.44
0.40
0.45
0.40
0.45
0.49
0.57
0.41
0.46
0.50
0.54
0.25
0.40
0.48
0.40
0.50
0.41
0.50
0.58
0.73
0.43
0.52
0.60
0.68
0.5
0.40
0.55
0.41
0.60
0.42
0.60
0.76
1.07
0.45
0.64
0.80
0.95
0.40
0.70
0.41
0.80
0.43
0.80
1.12
1.74
0.50
0.88
1.20
1.50
1.25
1.00
1.37
1.01
1.50
1.04
1.50
1.91
2.67
1.13
1.60
2.00
2.38
2.5
1.00
1.75
1.03
2.01
1.08
2.01
2.81
4.35
1.25
2.20
3.00
3.75
1.50
2.50
1.55
3.02
1.65
3.01
5.12
7.69
2.00
3.40
5.50
6.50
10
3.00
5.00
3.11
6.04
3.31
6.03
10.24
15.38
4.00
6.80
11.00
13.00
2.5*
8.00
9.50
8.03
9.76
8.08
9.76
9.82
12.10
8.25
9.95
10.01
11.51
5*
8.00
11.00
8.05
11.52
8.15
11.51
11.63
16.20
8.50
11.90
12.01
15.01
10*
8.00
14.00
8.11
15.04
8.31
15.03
15.25
24.39
9.00
15.80
16.01
22.01
20*
8.00
20.00
8.22
22.08
8.62
22.05
22.49
40.77
10.00
23.60
24.01
36.01
25*
20.00
27.50
20.27
30.09
20.77
30.07
38.11
53.47
22.50
32.00
40.02
47.52
50*
20.00
35.00
20.55
40.19
21.54
40.13
56.21
86.92
25.00
44.00
60.02
75.02
100*
30.00
50.00
31.09
60.38
33.08
60.26
102.42
153.83
40.00
68.00
110.02
130.02
200*
60.00
100.00
62.19
120.75
66.16
120.52
204.84
307.66
80.01
136.01
220.05
260.05
max offset
Orion-0816-100x
max offset
mV
mV
mV
mV
mV
mV
0.125
0.41
0.52
0.42
0.48
0.43
0.49
0.25
0.43
0.64
0.43
0.56
0.46
0.58
0.5
0.45
0.89
0.47
0.72
0.52
0.77
0.50
1.37
0.53
1.03
0.63
1.13
1.25
1.13
2.22
1.16
1.79
1.29
1.91
2.5
1.25
3.44
1.33
2.58
1.58
2.83
2.00
5.87
2.15
4.15
2.65
4.65
10
4.00
11.74
4.30
8.30
5.30
9.30
2.5*
8.25
11.19
8.33
10.33
8.58
10.58
5*
8.50
14.37
8.65
12.65
9.15
13.15
10*
9.00
20.74
9.30
17.30
10.30
18.30
20*
10.00
33.49
10.60
26.60
12.60
28.60
25*
22.50
44.36
23.25
35.75
25.75
38.26
50*
25.00
68.71
26.50
51.51
31.50
56.51
100*
40.00
117.41
43.00
83.01
53.01
93.01
200*
80.01
234.83
86.01
166.02
106.01
* ranges with activated input divider (250V maximum common mode voltage)
186.02
Page 152
MDAQ-SUB-V-200-BNC Calibration
25.3 Schematic
A few Calibrators allow the connection of all MDAQ Inputs together to the Calibrator source.
Also the Fluke 5500 for Voltages except in the lowers ranges.
B NC
na
na n
-B a
MDAQ-SUB-V200-BNC
CH0 CH2
CH4
CH6
CH8
CH1 CH3
CH5
CH7
CH9
CH11
CH13 CH15
MDAQ-SUB-V200-BNC System
Multi Product Calibrator
Cables: Safety Banana Plugs (red ,black), to BNC.
Page 153
MDAQ-SUB-V-200-BNC Calibration
Set the MDAQ-SUB-V-200-BNC to the first input Range (0.125V ) and connect channel 0 to the calibrator.
By clicking on the slot 0 icon it is easy to copy and paste the channel setting to all the others.
Apply 0.0000V to the input. The measured value has to be 0.0000V the maximum offset defined in table 25.2.4 .
For example a system with a PCI-6071-E together with an MDAQ-SUB-V-200-BNC in the 125mV range has to be
between the maximum offset errors of 0.490mV.
Page 154
MDAQ-SUB-V-200-BNC Calibration
25.5.1 Gain Calibration
Apply a value between 90% and 99% of full scale with the multi product calibrator to the input. The measured value
has to be the applied value the absolute system accuracy defined in table 25.2.4.
For example:
o PCI-6071 + MDAQ-SUB-V-200
o 125mV Range
o Applied Signal :120mV
At this system the measured value has to be 120mV 0.570mV. Apply the same value negative and the output has to
be -120mV 0.570mV.Check the Points 25.5.1 and 25.5.1 for all selectable Ranges and all 16 Channels.
Caution: There are a few ranges twice inside the range selection field in DeweSoft. The difference is that on half of
them the input divider is activated and so the maximum common mode voltage is 250V. These ranges are not that
accurate as the ranges with 12V common mode voltage, and so they are separately calibrated.
Caution: DeweSoft indicates an overload when the signal is more than 98% of
the range. This does not have any influence on the calibration.
25.5.2 AC Calibration
If filters are installed, deactivate them. Or switch them to the highest bandwidth.
Set the MDAQ-SUB-V-200 to the 10V divider off Input Range ( 12V max common mode). Apply a 7.000VRMS (19.799
VPeakPeak) , 1kHz sine wave signal via a shielded cable to the input.
Set the AD Sampling rate to the maximum. Be sure that it is more than the double of the calibration frequency;
otherwise aliasing effects take influence on your measurement.
AC-RMS
Slow (1s)
Multiplexed boards
PCI-62 PCI-60 PCI-60
8x
71
52
DT-983
4
mVRMS
33
Range
frequ.
voltage
1624
1616-10x
1616-50x
0816-1000
PCI-625x
10
kHz
10
mVRMS
7
mVRMS
59
mVRMS
61
mVRMS
26
mVRMS
26
mVRMS
23
mVRMS
23
mVRMS
36
mVRMS
24
10
50
255
992
327
243
219
240
232
262
229
10 divider on
50
752
1489
824
740
716
737
729
759
726
*Some AD-Cards are not able to run all 16 Channels at sample rates higher then 50kS/s. In that case skip the 50kHz
test point, or activate channel by channel.
E-M0243E Printing Version 1.6.9 FEB, 2008
Page 155
MDAQ-SUB-V-200-BNC Calibration
25.6 Calibration Report Example for MDAQ-SUB-V-200
Unit under Test :
Serial Number:
Inspection Date:
MDAQ-SUB-V-200
Standards Used
Model
Temperature:
Humidity:
Tested by:
S/N
Calibration Date
Calibrator
AD-Card
ORION-3216-100
Calibration Procedure
Test description
Limits
Range
Calibrator
lower
upper
Channel 0
DC
125mV
250mV
500mV
1.0V
1.25V
2.5V
5.0V
10V
2.5V
Divider On
5V
Divider On
10V
Divider On
20V
Divider On
25V
Divider On
50V
Divider On
100V
Divider On
200V
Divider On
AC
0.0mV
120.0mV
-120.0mV
0.0mV
240.0mV
-240.0mV
0.0mV
480.0mV
-480.0mV
0.0mV
990.0mV
-990.0mV
0.0mV
1200mV
-1200mV
0.000V
2.400V
-2.400V
0.000V
4.900V
-4.900V
0.000V
9.900V
-9.900V
0.000V
2.400V
-2.400V
0.000V
4.900V
-4.900V
0.000V
9.900V
-9.900V
0.000V
19.00V
-19.00V
0.000V
24.00V
-24.00V
0.000V
49.00V
-49.00V
0.000V
99.00V
-99.00V
0.000V
190.00V
-190.00V
7VRMS10kHz
7VRMS50kHz
7VRMS50kHz
-0.416 mV
119.524 mV
-120.476 mV
-0.433 mV
239.448 mV
-240.553 mV
-0.465 mV
479.295 mV
-480.705 mV
-0.00053 V
0.98898 V
-0.99103 V
-0.00116 V
1.19824 V
-1.20176 V
-0.0013 V
2.3975 V
-2.4025 V
-0.0022 V
4.8959 V
-4.9041 V
-0.0043 V
9.8917 V
-9.9083 V
-0.0083 V
2.3898 V
-2.4102 V
-0.0087 V
4.8874 V
-4.9126 V
-0.0093 V
9.8828 V
-9.9172 V
-0.011 V
18.974 V
-19.026 V
-0.023 V
23.965 V
-24.035 V
-0.027 V
48.949 V
-49.051 V
-0.043 V
98.917 V
-99.083 V
-0.086 V
189.838 V
-190.162 V
6.939 Vrms
6.008 Vrms
5.511 Vrms
Value
O.K
0.416 mV
120.476 mV
-119.524 mV
0.433 mV
240.553 mV
-239.448 mV
0.465 mV
480.705 mV
-479.295 mV
0.00053 V
0.99103 V
-0.98898 V
0.00116 V
1.20176 V
-1.19824 V
0.0013 V
2.4025 V
-2.3975 V
0.0022 V
4.9041 V
-4.8959 V
0.0043 V
9.9083 V
-9.8917 V
0.0083 V
2.4102 V
-2.3898 V
0.0087 V
4.9126 V
-4.8874 V
0.0093 V
9.9172 V
-9.8828 V
0.011 V
19.026 V
-18.974 V
0.023 V
24.035 V
-23.965 V
0.027 V
49.051 V
-48.949 V
0.043 V
99.083 V
-98.917 V
0.086 V
190.162 V
-189.838 V
7.061 Vrms
7.992 Vrms
8.489 Vrms
Page 156
MDAQ-SUB-ACC Calibration
26 MDAQ-SUB-ACC(-A)
26.1 Signal Path
RS 485
MDAQ- SUB-ACC(- A)
AD- Boar d
Ch0 to Ch15
5V
Ch0 to Ch15
dig ital 5 V
in
areG
oftw
S
orectin
C
Ch0 to Ch15
Inp ut
125 mV to 1 0V
Dewe so ft 6
BNC-OUT-5
Ch1 to Ch15
Con ditione d Sign al Out
5V
Ch0 to Ch15
5V
26.2 Specification
26.2.1 DC Accuracy
MDAQ-SUB-ACC
Software corrected
Analog out
Range
% of reading
% of range
% of reading
% of range
0.125V
0.03
0.32
0.15
0.32
Source:
0.25V
0.03
0.16
0.15
0.16
0.5V
0.03
0.08
0.15
0.08
1V
0.03
0.04
0.15
0.04
1.25V
0.03
0.08
0.15
0.08
2.5V
0.03
0.04
0.15
0.04
5V
0.02
0.03
0.15
0.03
10V
0.02
0.03
0.15
0.03
limits
Range
Lower
Upper
4mA
3.8 mA
4.2 mA
8mA
7.6 mA
8.4mA
Page 157
MDAQ-SUB-ACC Calibration
26.2.3 AC Accuracy
MDAQ-SUB-V-200-BNC
0..1kHz DC Accuracy
1 to 10 kHz DC Accuracy + 0.2% of reading
10kHz to 50kHz DC Accuracy + 3% of reading
10kHz to 50kHz DC Accuracy + 10% of reading
divider off
divider on
AD Board DC Accuracy
AD-Card
Range
% of reading
% of range
Orion-1624
5V
0.0574
0.01
Orion-xx16 series
5V
0.02
0.013
PCI-622x
5V
0.016
0.016
DE-D0421E DEWE-ORION-1624.
Printing version 1.0.6 Dec. 04, 2006
DE-M060304E DEWE-ORION-1616-10x
Printing version 1.0.2 November 24,2006
B/E/M/S Series Calibration
PCI-625x
5V
0.00928
0.01094
PCI-628x
5V
0.00718
0.00308
370937D-01 Jan07
PCI-6071
5V
0.0314
0.0724
PCI-6052
5V
0.008
0.01
DT-9834
5V
0.00
0.08
AD-Card
Orion-1624
Orion-xx16-10x
Orion-xx16-50x
Orion-0816-1000
PCI-625x
PCI-628x
PCI-6071
PCI-6052
DT-9834
PCI-622x
Specification Source
AD Board AC Accuracy
additional Gain error
-3dB
Range
% at 10kHz
% at 50kHz
bandwidth
5V
0.4
0.4
200kHz
5V
0.5
11
100kHz
5V
0
1.5
300kHz
5V
0
0.3
750kHz
5V
0
0
1.7MHz
5V
0
0.3
750kHz
5V
0
0
1.6MHz
5V
0
0.6
480kHz
5V
0
0
1.1MHz
5V
0
0.3
700kHz
INFO:
Orion-1624 AC Accuracy:
Flatness + % of reading @200kS/sec
All other AD-Board AC accuracies are
calculated by using the specified
bandwidth as base for an 1st order low
pass filter. The filter attenuation on the
test frequencies is the AC gain accuracy
at these points.
All AD-Board accuracy specifications for 3rd party products have been taken from the manufacturers published data
sheets and are subject to change without notice.
E-M0243E Printing Version 1.6.9 FEB, 2008
Page 158
MDAQ-SUB-ACC Calibration
26.2.5 Absolute System DC Accuracies with multiplexed boards
System Accuracy MDAQ-SUB-V-200-x using DeweSoft 6.3 or later
Range
PCI-622x
max
offset
PCI-625x
max
Error
max
offset
PCI-628x
max
Error
max
offset
max
Error
PCI-6071
max
offset
PCI-6052
max Error
max
offset
max
Error
DT-9834 (USB2)
max
offset
max
Error
mV
mV
mV
mV
mV
mV
mV
mV
mV
mV
mV
mV
0.125
0.40
0.44
0.40
0.45
0.40
0.45
0.49
0.57
0.41
0.46
0.50
0.54
0.25
0.40
0.48
0.40
0.50
0.41
0.50
0.58
0.73
0.43
0.52
0.60
0.68
0.5
0.40
0.55
0.41
0.60
0.42
0.60
0.76
1.07
0.45
0.64
0.80
0.95
0.40
0.70
0.41
0.80
0.43
0.80
1.12
1.74
0.50
0.88
1.20
1.50
1.25
1.00
1.37
1.01
1.50
1.04
1.50
1.91
2.67
1.13
1.60
2.00
2.38
2.5
1.00
1.75
1.03
2.01
1.08
2.01
2.81
4.35
1.25
2.20
3.00
3.75
1.50
2.50
1.55
3.02
1.65
3.01
5.12
7.69
2.00
3.40
5.50
6.50
10
3.00
5.00
3.11
6.04
3.31
6.03
10.24
15.38
4.00
6.80
11.00
13.00
max offset
Orion-0816-100x
max offset
mV
mV
mV
mV
mV
mV
0.125
0.41
0.52
0.42
0.48
0.43
0.49
0.25
0.43
0.64
0.43
0.56
0.46
0.58
0.5
0.45
0.89
0.47
0.72
0.52
0.77
0.50
1.37
0.53
1.03
0.63
1.13
1.25
1.13
2.22
1.16
1.79
1.29
1.91
2.5
1.25
3.44
1.33
2.58
1.58
2.83
2.00
5.87
2.15
4.15
2.65
4.65
10
4.00
11.74
4.30
8.30
5.30
9.30
Page 159
MDAQ-SUB-ACC Calibration
26.3 Schematic
A few calibrators allow the connection of all MDAQ inputs together to the calibrator source.
Also the Fluke 5500 for voltages except in the lowers ranges.
B NC
n
-B a
a nn
MDAQ-SUB-ACC
CH4
CH6
CH8
CH1 CH3
CH5
CH7
CH9
CH11
CH0 CH2
CH13 CH15
MDAQ-SUB-ACC System
Multi Product Calibrator
Digital Multimeter ( for constant current calibration)
Cables: Safety Banana Plugs (red ,black), to BNC.
Caution: DeweSoft indicates an overload when the signal is more than 98% of
the range. This does not have any influence on the calibration.
E-M0243E Printing Version 1.6.9 FEB, 2008
Page 160
MDAQ-SUB-ACC Calibration
26.5.1 Offset Calibration
o
Copy the settings to the other channels by clicking on the slot 0 icon.
Connect the calibrator to channel 0.
By clicking on the Slot 0 icon it is easy to copy and paste the channel
setting to all the others.
Apply 0.0000V to the input. The measured Value has to be 0.0000V the
maximum offset defined in table 26.2.4.
For example a system with a PCI-6071-E together with an MDAQ-SUB-ACC
in the 125mV range has to be between the maximum offset error of
0.490mV.
Page 161
MDAQ-SUB-ACC Calibration
26.5.2 AC Calibration
Set the MDAQ-SUB-V-200 to the 10V Input Range. Apply a 7.000VRMS (19.799 VPeakPeak) , 1kHz sine wave signal via a
shielded cable to the input.
Set the AD sampling rate to the maximum. Be sure that it is more than the double of the calibration frequency.
Otherwise aliasing effects take influence on your measurement.
AC-RMS
Slow (1s)
frequ.
voltage
1624
1616-10x
1616-50x
0816-1000
PCI-625x
PCI-62
8x
PCI-60
71
PCI-60
52
DT-983
4
kHz
mVRMS
10
10
mVRMS
59
mVRMS
61
mVRMS
26
mVRMS
26
mVRMS
23
mVRMS
23
mVRMS
36
mVRMS
24
mVRMS
33
10
50
255
992
327
243
219
240
232
262
Range
229
*Some AD-Cards are not able to run all 16 channels at sample rates higher then 50kHz. In that case skip the 50kHz
test point, or activate channel by channel.
26.5.2.2 Check the constant current
Connect a Digital Multimeter in Current measurement mode to the first Channel of the MDAQ-SUB-ACC.
Set the module to :
Range:
Filter:
Input Mode:
Coupling:
Current:
10000mV
highest
ICP
AC, any frequency
4mA
Page 162
MDAQ-SUB-ACC Calibration
26.6 Calibration Report Example for MDAQ-SUB-ACC
MDAQ-SUB-ACC
Temperature:
Humidity:
Tested by:
Standards Used
Model
S/N
Calibration Date
Calibrator
AD-Card
Orion-3216-100
Calibration Procedure
Test description
Limits
Range
Calibrator
lower
upper
Channel 0
DC
125mV
250mV
500mV
1.0V
1.25V
2.5V
5.0V
10V
AC
Constant
Current
10V
10V
4 mA
8 mA
0 mV
120 mV
-120 mV
0 mV
240 mV
-240 mV
0 mV
480 mV
-480 mV
0V
0.99 V
-0.99 V
0V
1.2 V
-1.2 V
0V
2.4 V
-2.4 V
0V
4.9 V
-4.9 V
0V
9.9 V
-9.9 V
7VRMS10kHz
7VRMS50kHz
-0.416 mV
119.524 mV
-120.476 mV
-0.433 mV
239.448 mV
-240.553 mV
-0.465 mV
479.295 mV
-480.705 mV
-0.00053 V
0.98898 V
-0.99103 V
-0.00116 V
1.19824 V
-1.20176 V
-0.0013 V
2.3975 V
-2.4025 V
-0.0022 V
4.8959 V
-4.9041 V
-0.0043 V
9.8917 V
-9.9083 V
6.939 Vrms
6.008 Vrms
3.8 mA
7.6 mA
Value
O.K
0.416 mV
120.476 mV
-119.524 mV
0.433 mV
240.553 mV
-239.448 mV
0.465 mV
480.705 mV
-479.295 mV
0.00053 V
0.99103 V
-0.98898 V
0.00116 V
1.20176 V
-1.19824 V
0.0013 V
2.4025 V
-2.3975 V
0.0022 V
4.9041 V
-4.8959 V
0.0043 V
9.9083 V
-9.8917 V
7.061 Vrms
7.992 Vrms
4.2 mA
8.4mA
Page 163
MDAQ-SUB-BRIDGE-D Calibration
27 MDAQ-SUB-BRIDGE-D and MDAQ-SUB-STG-D
27.1 Signal Path
RS 485
MDAQ-S UB-BRIDGE
AD- Boar d
Ch0 to Ch15
5V
Ch0 to Ch15
dig ital 5 V
in
areG
oftw
S
orectin
C
Ch0 to Ch15
Inp ut
2.5mV to 10 V
Dewesoft 6
BNC-OUT-5
Ch1 to Ch15
Con ditione d Sign al Out
5V
Ch0 to Ch15
5V
Software corrected
Analog out
Range
% of reading
% of range
% of reading
% of range
2.5mV
0.03
0.60
0.25
0.60
5mV
0.03
0.30
0.25
0.30
10mV
0.03
0.15
0.25
0.15
20mV
0.03
0.12
0.25
0.12
25mV
0.03
0.06
0.25
0.06
50mV
0.03
0.05
0.25
0.05
100mV
0.03
0.03
0.25
0.03
200mV
0.03
0.25
0.25V
0.03
0.03
0.16
0.15
0.03
0.16
0.5V
0.03
0.08
0.15
0.08
1V
0.03
0.04
0.15
0.04
1.25V
0.03
0.08
0.15
0.08
2.5V
0.03
0.04
0.15
0.04
5V
0.02
0.03
0.15
0.03
10V
0.02
0.03
0.15
0.03
Source:
DE-M041203E MDAQ-INT-Series Modules
Printing version 1.1.1 November 07, 2007
Page 164
MDAQ-SUB-BRIDGE-D Calibration
27.2.2 Excitation Voltage Accuracy
MDAQ-SUB-BRIDGE & MDAQ-SUB-STG
Excitation
0.05%of selected Excitation 0.7mV
27.2.3 AC Accuracy
MDAQ-SUB-BRIDGE & MDAQ-SUB-STG
0..1kHz DC Accuracy
1 to 10 kHz DC Accuracy +7% of reading
AD-Board DC accuracies
AD-Card
Range
% of reading
% of range
Orion-1624
5V
0.0574
0.01
5V
0.02
0.013
5V
0.0093
0.001094
DE-D0421E DEWE-ORION-1624.
Printing version 1.0.6 Dec. 04, 2006
DE-M060304E DEWE-ORION-1616-10x
Printing version 1.0.2 November 24,2006
B/E/M/S Series Calibration
Orion-xx16 series
PCI-625x
Specification Source
PCI-628x
5V
0.0072
0.00308
PCI-6071
5V
0.0314
0.0724
370937d-01 Jan07
PCI-6052
5V
0.0080
0.01
DT-9834 (USB)
5V
0.00
0.08
Page 165
MDAQ-SUB-BRIDGE-D Calibration
AD-Board
AD-Board AC accuracies
additional Gain error
-3dB
% at
% at
Range
bandwidth
10kHz
50kHz
Orion-1624
5V
200kHz
0.4
Orion-xx16-10x
5V
100kHz
0.5
0.4
11
Orion-xx16-50x
5V
300kHz
1.5
Orion-0816-1000
5V
750kHz
0.3
PCI-625x
5V
1.7MHz
PCI-628x
5V
750kHz
0.3
PCI-6071
5V
1.6MHz
PCI-6052
5V
480kHz
0.6
DT-9834
5V
1.1MHz
PCI-622x
5V
700kHz
0.3
All AD-Board accuracy specifications for 3rd party products have been taken from the manufacturers published data
sheets and are subject to change without notice.
PCI-622x
max
offset
max
Error
PCI-625x
max
offset
max
Error
PCI-628x
max
offset
max
Error
PCI-6071
max
offset
max
Error
PCI-6052
max
offset
max
Error
DT-9834
max
offset
max
Error
mV
mV
mV
mV
mV
mV
mV
mV
mV
mV
mV
mV
2.5mV
0.015
0.016
0.015
0.016
0.015
0.016
0.017
0.018
0.015
0.016
0.017
0.018
5mV
0.015
0.016
0.015
0.017
0.015
0.017
0.019
0.022
0.016
0.017
0.019
0.021
10mV
0.015
0.018
0.015
0.019
0.015
0.019
0.022
0.028
0.016
0.020
0.023
0.026
20mV
0.024
0.030
0.024
0.032
0.025
0.032
0.038
0.051
0.026
0.034
0.040
0.046
25mV
0.015
0.022
0.015
0.025
0.016
0.025
0.033
0.048
0.018
0.027
0.035
0.043
50mV
0.025
0.040
0.026
0.045
0.027
0.045
0.061
0.092
0.030
0.049
0.065
0.080
100mV
0.030
0.060
0.031
0.070
0.033
0.070
0.102
0.164
0.040
0.078
0.110
0.140
0.2
0.060
0.120
0.062
0.141
0.066
0.141
0.205
0.328
0.080
0.156
0.220
0.280
0.25
0.40
0.48
0.40
0.50
0.41
0.50
0.58
0.73
0.43
0.52
0.60
0.68
0.5
0.40
0.55
0.41
0.60
0.42
0.60
0.76
1.07
0.45
0.64
0.80
0.95
0.40
0.70
0.41
0.80
0.43
0.80
1.12
1.74
0.50
0.88
1.20
1.50
1.25
1.00
1.37
1.01
1.50
1.04
1.50
1.91
2.67
1.13
1.60
2.00
2.38
2.5
1.00
1.75
1.03
2.01
1.08
2.01
2.81
4.35
1.25
2.20
3.00
3.75
1.50
2.50
1.55
3.02
1.65
3.01
5.12
7.69
2.00
3.40
5.50
6.50
10
3.00
5.00
3.11
6.04
3.31
6.03
10.24
15.38
4.00
6.80
11.00
13.00
Page 166
MDAQ-SUB-BRIDGE-D Calibration
27.2.6 Absolute System DC Accuracies with synchronised sampling boards
System Accuracy MDAQ-SUB-V-200-x Synchron Sampling Boards using DeweSoft 6.3 or later
Range
Orion-1624
max offset
Orion-xx16-100/500
max abs.
error
max offset
Orion-0816-1000
max abs.
error
max offset
max abs.
error
mV
mV
mV
mV
mV
mV
0.0025
0.015
0.017
0.015
0.017
0.016
0.017
0.005
0.016
0.020
0.016
0.018
0.016
0.019
0.01
0.016
0.025
0.016
0.021
0.017
0.022
0.02
0.026
0.043
0.027
0.037
0.029
0.039
0.025
0.018
0.039
0.018
0.031
0.021
0.033
0.05
0.030
0.074
0.032
0.057
0.037
0.062
0.1
0.040
0.127
0.043
0.093
0.053
0.103
0.2
0.080
0.255
0.086
0.186
0.106
0.206
0.25
0.43
0.64
0.43
0.56
0.46
0.58
0.5
0.45
0.89
0.47
0.72
0.52
0.77
0.50
1.37
0.53
1.03
0.63
1.13
1.25
1.13
2.22
1.16
1.79
1.29
1.91
2.5
1.25
3.44
1.33
2.58
1.58
2.83
2.00
5.87
2.15
4.15
2.65
4.65
27.3 Schematic
A few Calibrators allow the connection of all MDAQ Inputs together to the Calibrator source.
Also the Fluke 5500 for Voltages except in the lowers ranges (below 3.3V).
a na
B an
14
H
C
12
H
C
15
H
C
13
H
C
1
H
C
MDAQ-SUB-BRIDGE
9
H
C
7
H
C
5
H
C
3
H
C
1
H
C
10
H
C
8
H
C
6
H
C
4
H
C
2
H
C
0
H
C
SUB
to D
MDAQ-SUB-BRIDGE
Page 167
MDAQ-SUB-BRIDGE-D Calibration
27.4 Parts List
o
o
o
o
MDAQ-SUB-BRIDGE System
Multi Product Calibrator
Digital Multimeter ( for Excitation voltage calibration)
Cables: Banana to DSUB 9
o Voltage input calibration: DAQx-V cable.( 3.2 cable number 9)
o Excitation voltage calibration: DAQx-Bridge cable + 1kOhm resistor.( 3.2 cable number 4)
o
o
o
o
Page 168
MDAQ-SUB-BRIDGE-D Calibration
o
For example a system with a PCI-6071-E together with an MDAQ-SUB-BRIDGE in the 2.5mV range has to be
between the maximum offset error limits of 17V.
10000mV
highest
Full bridge
1V
DC ( MDAQ-SUB-BRIDGE only)
Channel
0
0
0
1
2
3
4
5
6
7
Page 169
MDAQ-SUB-BRIDGE-D Calibration
27.5.3 AC Calibration
Set the MDAQ-SUB-BRIDGE to the 10V Input Range. Apply a 7.000VRMS (19.799 VPeakPeak), 1kHz sine wave signal via
a shielded cable to the input.
Set the AD Sampling rate to 25kSamples/sec.
AC-RMS
Slow (1s)
PCI-6
071
mV
490
PCI-6 DT-98
052
34
mV
mV
490
490
Page 170
MDAQ-SUB-BRIDGE-D Calibration
27.6 Calibration Report Example for MDAQ-BRIDGE-D
Unit under Test :
Serial Number:
Inspection Date:
MDAQ-SUB-BRIDGE-D
Standards Used
Model
S/N
Temperature:
Humidity:
Tested by:
Calibration Date
Calibrator
Multimeter
AD-Card
ORION-3216-100
Calibration Procedure
Test description
Limits
Range
Calibrator
lower
upper
Channel 0
DC
2.5mV
5mV
10mV
20mV
25mV
50mV
100mV
200mV
250mV
500mV
1.0V
1.25V
2.5V
5.0V
10V
AC
Excitation
Voltage
10V
0 mV
2.4 mV
-2.4 mV
0 mV
4.9 mV
-4.9 mV
0 mV
9.9 mV
-9.9 mV
0 mV
19.9 mV
-19.9 mV
0 mV
24 mV
-24 mV
0 mV
49 mV
-49 mV
0 mV
99 mV
-99 mV
0 mV
199 mV
-199 mV
0 mV
240 mV
-240 mV
0 mV
480 mV
-480 mV
0V
0.99 V
-0.99 V
0V
1.2 V
-1.2 V
0V
2.4 V
-2.4 V
0V
4.9 V
-4.9 V
0V
9.9 V
-9.9 V
7VRMS10kHz
1V
5V
12V
-0.015 mV
2.383 mV
-2.417 mV
-0.016 mV
4.882 mV
-4.918 mV
-0.016 mV
9.879 mV
-9.921 mV
-0.027 mV
19.863 mV
-19.937 mV
-0.018 mV
23.970 mV
-24.030 mV
-0.032 mV
48.944 mV
-49.056 mV
-0.043 mV
98.908 mV
-99.093 mV
-0.086 mV
198.815 mV
-199.186 mV
-0.433 mV
239.448 mV
-240.553 mV
-0.465 mV
479.295 mV
-480.705 mV
-0.00053 V
0.98898 V
-0.99103 V
-0.00116 V
1.19824 V
-1.20176 V
-0.0013 V
2.3975 V
-2.4025 V
-0.0022 V
4.8959 V
-4.9041 V
-0.0043 V
9.8917 V
-9.9083 V
6.510 VRMS
0.9988V
4.9968 V
11.9933 V
Value
O.K
0.015 mV
2.417 mV
-2.383 mV
0.016 mV
4.918 mV
-4.882 mV
0.016 mV
9.921 mV
-9.879 mV
0.027 mV
19.937 mV
-19.863 mV
0.018 mV
24.030 mV
-23.970 mV
0.032 mV
49.056 mV
-48.944 mV
0.043 mV
99.093 mV
-98.908 mV
0.086 mV
199.186 mV
-198.815 mV
0.433 mV
240.553 mV
-239.448 mV
0.465 mV
480.705 mV
-479.295 mV
0.00053 V
0.99103 V
-0.98898 V
0.00116 V
1.20176 V
-1.19824 V
0.0013 V
2.4025 V
-2.3975 V
0.0022 V
4.9041 V
-4.8959 V
0.0043 V
9.9083 V
-9.8917 V
7.490 VRMS
1.0012 V
5.0032 V
12.0067 V
Page 171