Publication No. II7049-52


2011-11-21

Installation Instructions
for the

CARESTREAM DRX-1 System Console


on the
QUANTUM QUEST Console
Service Code: 7049
Important
Qualified service personnel must do these procedures.
See the Carestream Health Service Portal for possible updates before doing these procedures.
When doing the procedures in this document, you must use safe work practices and wear the correct personal
protective equipment (for example, safety eyewear) according to your companys standard operating
procedures.

Confidential
Carestream Health, Inc. 2011

PLEASE NOTE

The information contained herein is based on the experience and knowledge relating to the
subject matter gained by Carestream Health, Inc. prior to publication.
No patent license is granted by this information.
Carestream Health, Inc. reserves the right to change this information without notice, and makes
no warranty, express or implied, with respect to this information. Carestream Health shall not be
liable for any loss or damage, including consequential or special damages, resulting from any
use of this information, even if loss or damage is caused by Carestream Healths negligence or
other fault.
This equipment includes parts and assemblies sensitive to damage from electrostatic
discharge. Use caution to prevent damage during all service procedures.

Table of Contents
Description

Page

Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Installing the Hardware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Testing the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Prep Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Power Off Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CR Mode or Film Screen Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
DRX-1 Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3
4
7
7
7
7
7

2011-11-21 II7049-52

Safety

Section 1: Safety
This information defines the safety and information icons used in this publication.

Caution
This icon is used for conditions that could cause injury to a person, or damage to the equipment or software data.

ESD
Possible damage from electrostatic discharge.
This icon is used for conditions that could cause damage to the equipment.
Electrostatic discharge (ESD) is a primary source of:
Equipment failure
Equipment repairs
Electrical components in this equipment can be damaged by a low level electrical discharge imperceptible to a
human.
Preventive Measures
Check for an ESD warning label before doing any procedure with ESD-sensitive components. All sensitive
components have graphic labels that frequently include instructions. Use all label instructions.
Wear a grounding strap when you touch ESD-sensitive components. Check that the clip remains fastened to a
ground that has a clean surface with no paint.
Repair components in an ESD-protection area or use a portable grounding mat.
When moving ESD-sensitive components from area to area, insert and transport the components in the special
material made for the transport of these components.

Important
This icon is used for important information.

Note
This icon is used for additional information.

Maximum Permissible Dose (MPD) and ALARA


Various studies on the effects of X-radiation have provided a foundation for establishing the maximum permissible
dose of radiation to which an individual might be exposed. The results of these studies have been used by the
National Council on Radiation Protection and Measurements (NCRP) and International Commission on Radiological
Protection (ICRP) to develop recommendations for the MPD. Federal and state agencies have incorporated these
recommendations into their regulations for MPD.
See the Safety and Regulatory Information for additional safety information.

II7049-52 2011-11-21

INSTALLATION INSTRUCTIONS

Section 2: Installing the Hardware


Caution
Dangerous Voltage
For the CARESTREAM DRX-1 System Console installation, you must see the Installation Instructions for the
CARESTREAM DRX-1 System Console and System Detector, II7049-1.
[1] De-energize:
DRX-1 System
OM Console

Operator
Control
Panel (OCP)

[2] Remove:
4 screws
Operator console panel (OCP)
Front panel screws
[3] Rotate the OCP.
[4] Lift to open the front panel of the OCP.

Hand switch
Connector

DRX-1 Interface
Cable

[5] Locate the hand switch connector.


[6] Route and push the DRX-1 interface cable through the dedicated hole in the back panel of the OM console.
[7] Connect the interface cable of the handswitch connector to the TB1 terminal.
[8] Locate the OCP control board connector A16A1.
[9] Remove jumpers JP6 and JP7 on OCP Control Board A16A1 to disable the operation of the PREP and
EXPOSE buttons on the OCP.
4

2011-11-21 II7049-52

Installing the Hardware

[10] Install the terminals of the DRX-1 interface cable, 2G7223, to the OM slot 4 of the OM connector block:
4 screws
Operator console panel (OCP)
Front panel screws
[11] Connect:
DRX-1 Console Exposure Switch Connector

DRX-1 interface cable to the DRX-1 Console


interface connector
DRX-1 Console exposure cable to the DRX-1
Console interface connector

DRX-1 Console Interface Connector

II7049-52 2011-11-21

H224_0149ACA
H224_0149AC

INSTALLATION INSTRUCTIONS

Section 3: Testing the System


Important
After completing the installation of the CARESTREAM DRX-1 System, you must do the following tests to check
that the DRX-1 System is correctly connected to the OM console.
The first 2 tests check that the X-ray beam fires as it did before this installation to expose CR or film cassettes.

Prep Test
[1] Push the DRX-1 System hand switch to the Prep state.
[2] Check that the Exposure Ready light on the OM console illuminates.

Power Off Test


Caution
Dangerous Voltage
[1] De-energize the DRX-1 System.
[2] Place a CR cassette on the table with a phantom object on the cassette.
[3] Press the DRX-1 System hand switch to expose the CR cassette.
[4] Process the cassette.
[5] Inspect the phantom image.

CR Mode or Film Screen Mode


[1] Energize the DRX-1 System.
[2] Select the CR/Film icon on the Acquisition screen.
[3] Place a CR cassette on the table with a phantom object on the cassette.
[4] Press the DRX-1 hand switch to expose the CR cassette.
[5] Process the cassette.
[6] Inspect the phantom image.

DRX-1 Mode
[1] Place the DRX-1 System Detector on the table with a phantom object on the detector.
[2] Select the DR icon on the Acquisition screen.
[3] Press the DRX-1 System hand switch to expose the DRX-1 System Detector.
[4] Acquire a phantom image.
[5] Inspect the image for the correct phantom image.

2011-11-21 II7049-52

Testing the System

Publication History
Publication
Date

Publication
No.

ECO No.

Changed Pages

File Name

Notes

2011-11-21

II7049-52

------

------

ii7049_52.fm

New Publication

II7049-52 2011-11-21

CARESTREAM is a trademarks of Carestream Health, Inc.

Printed in U.S.A. ii7049_52.fm

Carestream Health, Inc.