Department of Preventive, Restorative and Pediatric Dentistry, School of Dental Medicine, University of Bern, Switzerland
Institute of Mathematical Statistics and Actuarial Science, University of Bern, Switzerland
a r t i c l e
i n f o
a b s t r a c t
Article history:
Objectives. To evaluate if depth of cure DISO determined by the ISO 4049 method is accurately
reected with bulk ll materials when compared to depth of cure Dnew determined by Vickers
microhardness proles.
7 February 2012
Methods. DISO was determined according to ISO 4049; Depth of cure and resin composite
specimens (n = 6 per group) were prepared of two control materials (Filtek Supreme Plus,
Filtek Silorane) and four bulk ll materials (Surel SDR, Venus Bulk Fill, Quixl, Tetric EvoCeram Bulk Fill) and light-cured for either 10 s or 20 s. For Dnew , a mold was lled with one
Keywords:
of the six resin composites and light-cured for either 10 s or 20 s (n = 22 per group). The mold
Increment
was placed under a microhardness indentation device and hardness measurements (Vickers
Light-curing time
hardness, VHN) were made at dened distances, beginning at the resin composite that had
Restorative
been closest to the light-curing unit (i.e. at the top) and proceeding toward the uncured
resin composite (i.e. toward the bottom). On the basis of the VHN measurements, Vickers
hardness proles were generated for each group.
Results. DISO varied between 1.76 and 6.49 mm with the bulk ll materials showing the highest
DISO . Dnew varied between 0.2 and 4.0 mm. Dnew was smaller than DISO for all resin composites
except Filtek Silorane.
Conclusions. For bulk ll materials the ISO 4049 method overestimated depth of cure compared to depth of cure determined by Vickers hardness proles.
2012 Academy of Dental Materials. Published by Elsevier Ltd. All rights reserved.
1.
Introduction
Corresponding author at: Freiburgstrasse 7, CH-3010 Bern, Switzerland. Tel.: +41 316322581; fax: +41 316329875.
E-mail address: simon.ury@zmk.unibe.ch (S. Flury).
0109-5641/$ see front matter 2012 Academy of Dental Materials. Published by Elsevier Ltd. All rights reserved.
doi:10.1016/j.dental.2012.02.002
522
d e n t a l m a t e r i a l s 2 8 ( 2 0 1 2 ) 521528
2.
2.1.
Depth of cure by ISO 4049 was performed with re-usable stainless steel molds according to ISO 4049:2000 [16]. Pretests had
found the absolute length of cylindrical specimens of the
cured resin composite to vary between 3.5 and 13 mm depending on the resin composite. The ISO 4049 method states that
the stainless steel molds shall be at least 2 mm longer than the
absolute length of the cylindrical specimens. Thus, stainless
steel molds of 6 mm, 9 mm, or 15 mm in length and an internal
diameter of 4 mm were custom-made.
Depending on the resin composite, the mold of either
6 mm, 9 mm, or 15 mm in length was placed on a glass slide
covered by a Mylar strip (Hawe Stopstrip Straight, KerrHawe,
Bioggio, Switzerland). The mold was then lled in bulk with
one of the six resin composites. The top side of the mold was
covered with a second Mylar strip and the resin material made
ush with the mold by use of a second glass slide. The mold
was placed on white lter paper (Filter Paper Circles 589/1,
Schleicher & Schuell MicroScience GmbH, Dassel, Germany).
The second glass slide was removed and the resin composite was light-cured for either 10 s or 20 s keeping the light tip
centered and in contact with the second Mylar strip. After
light-curing, the cylindrical specimens were pushed out of the
mold and the uncured resin composite material was removed
with a plastic spatula. The absolute length of the cylindrical
specimens of cured resin composite was then measured with a
digital caliper of 0.01 mm accuracy (Mitutoyo IP 65, Kawasaki,
Japan). The absolute length (AL) was divided by two and the
latter value recorded as DISO . Six specimens were made in each
of the 12 groups (i.e. six materials light-cured for either 10 s or
20 s).
2.2.
523
d e n t a l m a t e r i a l s 2 8 ( 2 0 1 2 ) 521528
Maximum increment
thickness (mm) (according
to manufacturer)
Shade
LOT-number
Universal restorative
A3
N116619
2.5
A3
N138530
Universal
100128
Universal
010030
Posterior restorative
Universal
1007001127
IVA
(reddish universal
shade)
IDS
1.5 mm, 2.0 mm, 2.5 mm, 3.0 mm, 3.5 mm, 4.0 mm, 4.5 mm,
5.0 mm, 6.0 mm, 7.0 mm, 8.0 mm, 9.0 mm, 10.0 mm, 11.0 mm,
12.0 mm, and 13.0 mm. Programming of the hardness indentation device for dened distances and reproducible placement
of the mold ensured that the VHN measurements were made
along the same axis on each specimen. VHN measurements
were made at a load of 3 g for 15 s. For each of the 12 groups (i.e.
six materials light-cured for either 10 s or 20 s), 22 specimens
were prepared and thus 22 VHN measurements were made at
each of the dened distances.
2.3.
Statistical analysis
In each of the 12 groups, the maximum VHNmax of the VHN values obtained at the dened distances = {0.1, 0.2, 0.5, 1.0, 1.5,
2.0, 2.5, 3.0, 3.5, 4.0, 4.5, 5.0, 6.0, 7.0, 8.0, 9.0, 10.0, 11.0, 12.0, 13.0}
was identied for each of the 22 specimens. For each group,
max
the median VHN
of the 22 VHNmax values as well as the
ISO
Fig. 1 Specimen preparation for depth of cure determination by Vickers hardness measurements. A = stainless steel mold
with semicircular notch, B = stainless steel shell, C = semicircular opening for light-curing, and D = mold including the resin
composite specimen.
524
[3.04,4.24]
[1.83,2.57]
[3.49,4.86]
[2.73,3.28]
[2.84,4.48]
[8.15,67.26]
3.57
2.09
3.94
3.04
3.71
38.25
[1.09,1.74]
[1.53,2.51]
[2.31,2.71]
[3.80,4.16]
[2.87,4.60]
[0.10,1.50]
1
0
3
1
0
5
1.5
2.0
2.5
4.0
3.5
0.2
[52.16,61.27]
[76.35,91.77]
[40.43,50.30]
[34.58,36.88]
[28.50,34.71]
[36.75,52.73]
60.7
83.3
46.5
36.0
32.5
51.5
[61.83,73.60]
[52.01,59.10]
[15.23,19.94]
[12.61,13.49]
[64.81,75.15]
[35.03,65.88]
68.50
55.95
17.65
13.05
70.10
51.95
[2.60,2.74]
[2.05,2.13]
[4.89,4.97]
[6.02,6.12]
[6.41,6.60]
[3.73,3.89]
Filtek Supreme Plus
Filtek Silorane
Surel SDR
Venus Bulk Fill
Quixl
Tetric EvoCeram Bulk Fill
2.68
2.09
4.93
6.08
6.49
3.83
95% CI
[3.91,6.57]
[1.07,2.06]
[1.03,6.89]
[3.16,3.25]
[4.58,11.44]
[3.32,33.20]
4.72
1.76
4.33
3.21
5.72
4.43
Median
95% CI
[0.54,1.26]
[1.87,2.16]
[1.55,2.54]
[2.98,3.02]
[1.07,2.35]
[0.23,1.77]
0
0
3
2
0
9
1.0
2.0
2.0
3.0
2.0
1.5
[51.57,65.33]
[83.05,91.10]
[27.32,46.01]
[42.21,47.83]
[23.62,30.21]
[29.32,50.15]
62.7
84.7
44.2
46.0
25.6
40.8
[51.74,65.10]
[54.40,61.08]
[14.34,16.72]
[8.84,10.00]
[57.57,67.10]
[39.87,62.19]
59.45
58.40
15.60
9.45
62.20
48.55
[2.32,2.40]
[1.70,1.80]
[4.28,4.37]
[4.76,4.86]
[5.69,5.76]
[3.29,3.35]
2.36
1.76
4.33
4.81
5.72
3.32
Nmiss
Median
95% CI
Median
95% CI
Median
fnew n = 22
Dnew (mm) n = 22
pISO (%) n = 6
VHNmax n = 22
20 s
max
ISO varied between 25.6%
attained at D
age pISO of VHN
and 84.7%, with the bulk ll materials retaining less than
max
50% of their VHN
whereas Filtek Supreme Plus retained
around 60% and Filtek Silorane retained slightly more than
max
80% of their VHN
. The biggest depth Dnew above which
10 s
95% CI
Results
Median
3.
DISO (mm) n = 6
max
ISO , VHNISO was divided by
pISO of VHN
was attained at D
max
new
. The biggest depth D
, above which at least 80% of
VHN
max
VHN
was attained, and the factor fnew , by which the absolute length should be divided in order to arrive at Dnew , was
calculated for each specimen. For certain specimens VHN was
max
never above 80% of VHN
. For these specimens Dnew and
new
f
could not be calculated and the number of such incalculable cases was called Nmiss . For DISO , VHNmax , pISO , Dnew and
fnew medians and 95% bootstrap condence intervals for the
medians were determined for each of the 12 groups.
All calculations were performed with R version 2.13.0
(The R Foundation for Statistical Computing, Vienna, Austria;
www.R-project.org), using descriptive methods.
Resin composite
attained).
Light-curing
Table 2 Median and 95% bootstrap condence interval for DISO (depth of cure determined by the ISO 4049 method), VHNmax (maximum of the VHN values obtained at
max
max
ISO ), Dnew (the biggest depth above which at least 80% of VHN
the dened distances), pISO (percentage of VHN
attained at D
was attained), and fnew (the factor by
new
).
which the absolute length of specimens, as determined by the ISO 4049 method, should be divided in order to arrive at D
d e n t a l m a t e r i a l s 2 8 ( 2 0 1 2 ) 521528
d e n t a l m a t e r i a l s 2 8 ( 2 0 1 2 ) 521528
525
L of
Fig. 3 Light-curing for 10 s: Vickers hardness proles for each resin composite (n = 22 per resin composite). Medians A
ISO of depth of cure obtained by the ISO 4049 method, and new depth of cure Dnew for each resin
absolute length, D
composite.
max
at least 80% of VHN
was attained varied between 0.2 and
new
4.
Discussion
This study showed great variation between the six resin composites as regards depth of cure DISO determined by the ISO
4049 method with the bulk ll materials, true to their name,
526
d e n t a l m a t e r i a l s 2 8 ( 2 0 1 2 ) 521528
L of
Fig. 4 Light-curing for 20 s: Vickers hardness proles for each resin composite (n = 22 per resin composite). Medians A
ISO of depth of cure obtained by the ISO 4049 method, and new depth of cure Dnew for each resin
absolute length, D
composite.
4049:2000 of a DISO not less than 1.5 mm. ISO 4049:2000 additionally requires that the depth of cure shall be no more than
0.5 mm below the value claimed by the manufacturer when
using the recommended light-curing time. Only one resin
composite, at only one of the two light-curing times, did not
d e n t a l m a t e r i a l s 2 8 ( 2 0 1 2 ) 521528
meet this requirement: The manufacturer indicates that Tetric EvoCeram Bulk Fill may be light-cured to a depth of 4 mm
using a light-curing time of 10 s provided that the light power
density is 1000 mW/cm2 . Yielding a DISO of 3.32 mm Tetric
EvoCeram Bulk Fill did not quite obtain the minimum DISO
required of 3.5 mm (i.e. 4.0 mm claimed by the manufacturer
minus 0.5 mm). As regards the inuence of light-curing time,
doubling the time from 10 s to 20 s led to an average increase
in DISO of 17%. This increase is in agreement with the 1623%
previously found as a result of doubling the light-curing time
[21,22].
The six resin composites also varied markedly regarding
Vickers hardness proles. The hardness proles of the four
bulk ll materials were wider than those of the control materials (as evidenced by the stretched-out hardness proles)
indicating that the point at which VHN could not be measured
anymore due to softness occurred at a much bigger depth for
the bulk ll materials. It is noteworthy that this point at which
VHN could not be measured occurred at smaller depth than
L of the hardened resin composite specthe absolute length A
imens determined by the ISO 4049 method (Figs. 3 and 4) for
all resin composites except Filtek Silorane. This discrepancy
may be the result of a limited resolution of the microhardness indentation device at relatively low surface hardnesses.
The hardness proles of the two owable bulk ll materials
were atter than those of the other materials indicating that
max
VHN
of the two owable bulk ll materials was markedly
lower. The inferior hardness can be explained by a lower ller
content necessary for obtaining the reduced viscosity. Furthermore, regardless of resin composite and light-curing time,
most hardness proles showed that VHNmax was not reached
at the very rst measuring depth of 0.1 mm but rather in
subsurface areas at a depth of 0.2 mm to 1.0 mm. This phenomenon has been previously described [23].
max
ISO was
The percentage of VHN
attained at the depth D
ISO
ISO
assessed and expressed as p . Whereas p
was above 80%
for one of the control materials (Filtek Silorane), pISO was much
lower particularly for the bulk ll materials, which showed in
the median a pISO of only 40%, reecting the different forms
of the Vickers hardness proles.
Numerous studies have dened depth of cure based on
hardness measurements performed on the top and bottom
surface of a light-cured resin composite specimen. The hardness values obtained were used to calculate a bottom/top
hardness ratio, and a ratio above 80% has often been used
as a minimum acceptable threshold value [14,24]. Analogically, in the present study a threshold of 80% was used in
order to determine a depth of cure based on the Vickers hardness measurements: Thus, Dnew was dened as the depth
max
was obtained. Dnew varied
at which at least 80% of VHN
between 1 and 4 mm. There was, however, one exception in
that Tetric EvoCeram Bulk Fill yielded a Dnew of only 0.2 mm
when light-cured for 20 s. The reason for this very low value
is mainly that the hardness of a majority of Tetric EvoCeram
Bulk Fill specimens dropped drastically after the rst measuring depth (0.1 mm). Except for Filtek Silorane, the Dnew
values were lower than the DISO values indicating that the
ISO 4049 method overestimated the depth of cure, especially
for the bulk ll materials as already evidenced by the pISO
values. This nding is in corroboration with that of Moore
527
5.
Conclusion
The present study found that for bulk ll materials the ISO
4049 method overestimated depth of cure compared to the
determination by Vickers hardness estimations of the degree
of conversion.
Conicts of interest
The authors declare no conicts of interest, real or perceived,
nancial or nonnancial.
528
d e n t a l m a t e r i a l s 2 8 ( 2 0 1 2 ) 521528
references
[15]
[16]
[1] Price RB, Murphy DG, Drand T. Light energy transmission
through cured resin composite and human dentin.
Quintessence Int 2000;31:65967.
[2] Ruyter IE, Oysaed H. Conversion in different depths of
ultraviolet and visible light activated composite materials.
Acta Odontol Scand 1982;40:17992.
[3] Ferracane JL, Mitchem JC, Condon JR, Todd R. Wear and
marginal breakdown of composites with various degrees of
cure. J Dent Res 1997;76:150816.
[4] Poskus LT, Placido E, Cardoso PE. Inuence of placement
techniques on Vickers and Knoop hardness of class II
composite resin restorations. Dent Mater 2004;20:72632.
[5] Sideridou ID, Achilias DS. Elution study of unreacted
Bis-GMA, TEGDMA, UDMA, and Bis-EMA from light-cured
dental resins and resin composites using HPLC. J Biomed
Mater Res B Appl Biomater 2005;74:61726.
[6] Sakaguchi RL, Douglas WH, Peters MC. Curing light
performance and polymerization of composite restorative
materials. J Dent 1992;20:1838.
[7] Pilo R, Oelgiesser D, Cardash HS. A survey of output
intensity and potential for depth of cure among light-curing
units in clinical use. J Dent 1999;27:23541.
[8] Quixl Scientic Compendium. Dentsply DeTrey; 2003.
[9] Surel SDR ow Directions For Use. Dentsply Caulk; 2009.
[10] Venus Bulk Fill Product Prole. Heraeus Kulzer; 2011.
[11] Tetric EvoCeram Bulk Fill Press Release. Ivoclar Vivadent;
2011.
[12] ISO 4049:1988 (2.). Dentistryresin-based lling materials.
International Organization for Standardization; 1988.
[13] DeWald JP, Ferracane JL. A comparison of four modes of
evaluating depth of cure of light-activated composites. J
Dent Res 1987;66:72730.
[14] Bouschlicher MR, Rueggeberg FA, Wilson BM. Correlation of
bottom-to-top surface microhardness and conversion ratios
[17]
[18]
[19]
[20]
[21]
[22]
[23]
[24]
[25]