Data Management
with Unity Desktop
Contents
Getting Started
11
21
Choosing a QC Procedure
41
Program Basics
53
61
69
Labs ........................................................................... 69
Adding and Updating Labs ........................................ 70
Duplicating Labs.................................................... 71
Deleting Labs ....................................................... 72
Opening and Closing Labs......................................... 72
4
Bio-Rad Laboratories
Contents
Lots............................................................................73
Adding Bio-Rad Lots ................................................73
Adding Non-Bio-Rad Lots ..........................................74
Duplicating Lots.....................................................75
Deleting Lots ........................................................77
Editing Lots ..........................................................77
Closing and Opening Lots..........................................78
Lot Expiration Notifications.......................................79
Tests
81
Overview of Tests ...........................................................81
Adding Tests .................................................................82
To Add Tests Manually .............................................83
Adding Tests With Instrument Setup ............................84
Adding Tests With an Other Code................................86
Test Rules and Settings ....................................................87
Rules Tab of the Test Settings Dialog Box ......................87
Settings Tab of the Test Settings Dialog Box ...................87
Duplicating Tests ............................................................88
Updating Tests...............................................................88
To Update a Test ...................................................88
Closing and Opening Tests .................................................89
To Close Tests.......................................................89
To Open Tests.......................................................90
Deleting Tests ...............................................................90
To Delete Tests .....................................................91
Sorting Tests .................................................................91
To Sort Tests in Lab Mode.........................................91
To Sort Tests in Panel Mode ......................................92
VITROS Slide Generations ..................................................92
Changing VITROS Slide Generations .............................93
Update VITROS Slide Generations................................94
95
Panels .........................................................................95
Panel Tests Dialog Box .....................................................95
To Add a Panel ......................................................97
To Update a Panel..................................................97
To Delete a Panel ..................................................97
To Sort Tests in a Panel ...........................................98
Data Groups ..................................................................98
To Change the Default Current Data Selection ................99
To Define or Edit a Group .........................................99
Bio-Rad Laboratories
Applying QC Rules
101
Entering Data
109
129
Bio-Rad Laboratories
Contents
135
143
155
165
173
177
Bio-Rad Laboratories
Contents
197
211
217
Bio-Rad Laboratories
Windows Hints
229
Supplemental Information
237
References
249
QC References .............................................................249
Glossary of Terms
251
Index
263
10
Bio-Rad Laboratories
H A P T E R
Getting Started
In This Chapter
Welcome to Unity Desktop ............................................ 11
Where Do I Begin? ...................................................... 14
Essential Startup Tasks for New Users............................... 15
Qualitative Data Entry (on page 125) that allows you to enter
qualitative and semi-quantitative data and send it to the Unity
Interlaboratory Program.
One of the most powerful utilities of Unity Desktop is its ability to connect
your laboratory to the worldwide clinical laboratory community. Bio-Rad's
Unity Interlaboratory Program collects data from thousands of labs
worldwide and combines it to create consensus groups to which your data is
compared.
How to Contact Us
We have made every attempt to ensure that this guide is accurate and
complete. However, if you need to contact Bio-Rad directly for assistance,
you can:
Inside the U.S., phone a Bio-Rad software support representative at 1800-854-6737, option 6. Software support representatives are available
Monday through Friday from 5 am to 5 pm (Pacific Standard Time).
If you phone Bio-Rad outside of normal working hours, please leave a
message and a software support representative will contact you,
typically within 24 hours.
Outside the U.S., please contact your local Bio-Rad software support
representative.
12
Bio-Rad Laboratories
Getting Started
Typographical Conventions
This guide uses consistent styles to aid readability. Throughout this
document:
Items you see on the screen, such as check box names and dialog box
titles, are enclosed in double quotes. For example, the Setup dialog
box.
Menu items are shown in a bold font and appear in the order you select
them with each item separated by a vertical bar (|). For example, Tools
| Security | Administrator.
Buttons that you click appear in a bold font. For example, click Close to
exit the dialog box and return to the main screen.
Names of keys on the keyboard appear in bold and in all capital letters.
For example, CTRL, SHIFT, DELETE.
Keys that you press together (i.e. shortcut keys) appear in bold and are
separated by a plus sign (+). For example, "Press SHIFT+F11 to exit the
program."
Notes are separated from the surrounding text by horizontal lines above
and below the note. For example:
This feature is only available when the test is being evaluated by an
analytical goal.
Service Packs
Occasionally, Bio-Rad releases updates for its existing software products.
When an update is available, Unity Desktop notifies you at login using one
of two screens based on whether or not you have selected the "Automatic
Bio-Rad Laboratories
13
product updates" check box on the "Configure Unity Desktop" tab of the
"Setup" dialog box (see "The Setup Dialog Box" on page 222).
Click Close to close the dialog box without taking any action. Unity
Desktop will notify you again of the update the next time you log
in.
Where Do I Begin?
Once you have completed Installation and Setup, what you do next depends
on what Bio-Rad QC data management software, if any, you were previously
using:
If you are a Unity-PC or Unity Plus/Pro user and want to move your
existing data into Unity Desktop, follow the instructions for moving the
data into Unity Desktop.
If you are a Unity Post user, contact your Bio-Rad software support
representative to request an import disk of data you have submitted to
the Unity Interlaboratory Program.
14
Bio-Rad Laboratories
Getting Started
When the test has a statistically significant number of data points, you
might want to change the default rules as described in Choosing a QC
Procedure (on page 41). If your lab uses the optional Westgard Advisor, it
will suggest rules for you. (Twenty is generally considered the minimum
number of data points for statistical significance.)
Bio-Rad Laboratories
15
If you laboratory wants to have rules suggested for you, contact your BioRad Representative to receive information regarding Unity Real Time
and the optional Westgard Advisor.
The "Password" field should be active (i.e. the cursor appears in it.) If
the password field is inactive, click in the field to select it.
Type sa in the password field, and click OK.
The default administrator login allows new users initial access to the
software. After you create other users, you should delete this user or
change the password.
Add Users
Unity Desktop uses passwords and user profiles to control which program
functions users can access. User initials appear on the data screen, with
actions and comments, and other areas within the software to document
which user performed the activity. User's initials, combined with audit
trails, document who did what and provide useful information when
investigating problems; hence, we strongly recommend that you create a
unique user profile and password for each person who uses Unity Desktop.
3
Click New in the User profile area. All fields on the dialog box are
cleared.
Type the user ID, user name, password, and initials you want to assign
to the new user.
Select the "Set password expiration period to" check box if you want the
user's password to expire after the number of months you choose from
the drop-down list.
16
Bio-Rad Laboratories
Getting Started
Password expiration periods are entirely optional and their use depends
on the level of security required in your laboratory.
8
Click the "Administrator" check box to give the new user all permissions
or select the individual permissions you want the user to have.
Click Apply to create the new user and leave the dialog box open so
that you can add another user or click OK to create the user and close
the dialog box.
Type a new lab number and edit the fields as necessary. Then click
Add to add an the lab number.
Click Lot to display the "Lot" dialog box to add a lot number(s) to
the lab number.
Select the lab number you want to add lots to in the navigation tree.
Bio-Rad Laboratories
17
The navigation trees appear as tabs on the left of the main screen. The
"Lab" tree presents a hierarchical arrangement by lab numbers, lot
numbers, and tests. The "Panel" tree organizes tests by panel name. See
Tree Views (on page 229) for more information.
3
Click the Lot icon on the toolbar. The "Lot" dialog box appears. Notice
that the selected lab number and lab description appear under the
dialog box title.
Select a lot number from the drop-down list under "Lot number." The
list will contain all unexpired lot numbers.
Click Add. The lot number appears in the open lots list.
When all lots are added, click Close to exit the dialog box or click Test
to add tests to the currently selected lot number.
Add Tests
1
Make a selection for each item in the left "Test information" pane.
Alternatively you can add groups of tests for an instrument using Instrument
Setup. (This feature is not available when the RiLiBK module is installed.)
Enter Data
You can enter data for a test as individual data points (point data entry) or
as a monthly mean, SD, and number of points (summary data entry). Rule
evaluation occurs only for point data.
18
Bio-Rad Laboratories
Getting Started
Type values for each level of control you have results for and press Tab
to move to the next data row.
Double-click the test in either the Lab or Panel tree. The data entry
screen appears with either the "Point Data" or "Summary Data" tab
selected.
Type your mean, SD, and number of points for each level, pressing Tab
to move between the cells.
When all values for a row are entered, press Tab to move to the next
blank data entry row.
Bio-Rad Laboratories
19
H A P T E R
Introduction to Quality
Control Statistics
In This Chapter
What is Quality Control? ............................................... 21
Basic QC Statistics ...................................................... 24
SPC Rules................................................................. 31
22
Bio-Rad Laboratories
The following example QC log shows normal and abnormal control results
and patient results for a seven-day period.
Test:
Potassium
Instrument:
Instrument No.1
Level 2 (Abnormal
Control)
3.7-4.3 mmol/L
6.7-7.3 mmol/L
11/1
4.0
7.0
11/2
4.1
7.0
11/3
4.0
6.9
11/4
4.2
7.1
11/5
4.1
7.0
11/6
4.1
7.0
11/7
4.2
8.0
Range:
Date:
Patient Results
The acceptable range for the Level I (normal control) is 3.7 4.3 mmol/L.
The range for Level II (abnormal control) is 6.7 7.3 mmol/L. When the
daily QC result obtained for the normal control is compared to the range
calculated for the normal control, it becomes apparent that each result lies
somewhere within the expected range. This indicates that the analytical
process is "in control" at the normal level on that day of testing.
When the daily QC result for the Level II (abnormal control) is compared to
the defined range for the abnormal control, the analytical process is shown
to be in control for each day of testing except for the last day (11/7). On
November 1 through November 6, both controls were "in control," and
patient values could be reliably reported.
However, the laboratory was "out of control" for abnormal high potassiums
on November 7 because the value obtained for the QC material (8.0
mmol/L) was outside the acceptable range (6.7 7.3 mmol/L). This result
indicates that some error occurred which may have produced abnormally
high patient results that are unreliable. The laboratory should not report
any patient samples with an abnormally high potassium until the error is
resolved and the abnormally high sample(s) are re-tested.
Bio-Rad Laboratories
23
Hence, it is now apparent that the range defined for each level of control is
fundamental to the quality control system. Basic QC Statistics (on page 24)
describes the calculations required to develop an acceptable control range.
CLIA Requirements
In the United States, the Clinical Laboratory Improvement Amendments of
1988 (CLIA), as modified by the final CLIA Rule, require two levels of control
(one normal and one abnormal) be assayed on each day the test is
performed for all non-waived tests, unless the Centers for Medicare &
Medicaid Services (CMS) approves the procedure specified in Appendix C of
the State Operations Manual that provides equivalent quality testing.
In other words, if you test patient samples for potassium on Wednesday, you
must assay at least one normal and one abnormal control product for
potassium on Wednesday unless an equivalent QC Procedure has been
approved by CMS. Blood gas testing is slightly different.
For instruments that verify calibration internally, the U.S. laboratory must
run one control every eight hours using a combination of materials that
include both low and high values on each day of testing.
If the instrument does not verify calibration internally, the laboratory must
assay these same controls with each patient sample.
As with any government regulation, these requirements can undergo change
as a result of the regulatory or political process.
Basic QC Statistics
A total QC system must control both trueness and precision. Your goal is for
a test to have both low bias and low imprecision. Since problems with
imprecision are less likely to cause analytical failure in the modern
laboratory because of computerization and mechanization of the analytical
process, bias is brought to the forefront.
The following figure illustrates low bias and low imprecision as a target,
where the center of the bulls eye represents the target value.
24
Bio-Rad Laboratories
The first figure illustrates the situation where your standard deviation
for a test is small (good precision), but is shifted away from the target
value (high bias).
The second figure represents poor precision and, believe it or not, low
bias because the average of the results is close to the bull's eye. Of
course, individually none of the points are close, and their individual z
scores would reflect this.
Bias and imprecision are most important at the clinical decision levels. For
example, -hCG clinical decision levels are at low concentrations
(corresponding to early pregnancy in the female and early testicular cancer
in the male) or at moderate concentrations (to diagnose the progression of
pregnancy).
Useful Statistics
Evaluation of a test's bias and imprecision utilizes several calculations.
Although Bio-Rad's QC evaluation software performs these calculations for
you, it's useful to understand the math behind the calculations, which
includes:
Mean
The mean of a group of data points is simply their arithmetic average. Your
calculated mean provides your laboratorys best estimate of the analytes
"true" value for a specific level of control. The mean a predetermined
number of standard deviations represents the error you expect to see in a
test when the analytical system is stable.
Bio-Rad Laboratories
25
Standard Deviation
The standard deviation measures a test's precision or how close individual
measurements are to each other. (The standard deviation does not measure
bias, which requires that you compare your results to a target value, such as
your peer group.) The standard deviation provides an estimate of how
repeatable a test is at specific concentrations. Test repeatability can be
consistent (low standard deviation, low imprecision) or inconsistent (high
standard deviation, high imprecision).
Obviously, we want repeated measurements of the same specimen to give
results that are as close to each other as possible. Good precision is
especially needed for tests which are repeated regularly on the same
patient to track treatment or disease progress. For example, a diabetic
patient in a critical care situation may have glucose levels run every 2 to 4
hours. In this case, it is important for the glucose test to be precise because
lack of precision can cause loss of test reliability. If there is a lot of
variability in the test performance (high imprecision, high standard
deviation), the glucose result at different times may not be true.
Standard deviation is often abbreviated as SD or s.
The following formula is used to calculate the SD:
SD =
(x
x)
n 1
Bio-Rad Laboratories
Calculate the mean and standard deviation from the data points collected.
Calculate the statistical control limits from the mean 2s and the
mean 3s.
SD group
Bio-Rad Laboratories
27
SDI value
Interpretation
0.0
1.25
Acceptable
1.25 1.49
1.5 1.99
Marginal performance
Investigation of the test system is recommended
2.0
Unacceptable performance
Remedial action usually required
Bias
Bias (%) measures how far your observed value is from a target value, which
can be determined by some reference value or estimated from outside
sources such as proficiency testing results or interlaboratory groups such as
Bio-Rads Unity Interlaboratory Program.
28
Bio-Rad Laboratories
M lab M group
Lab Bias % =
M group
100
Where Mlab is your lab's mean and Mgroup is the reference value (i.e.
proficiency or interlaboratory value).
Coefficient of Variation
The Coefficient of Variation (CV) is the ratio of the standard deviation to
the mean and is expressed as a percentage as follows:
standard deviation
CV % =
100
mean
Bio-Rad Laboratories
29
Determining an Acceptable CV
How do you determine an acceptable CV? Several sources can provide
expected levels of precision, including:
Proficiency surveys
These sources are useful for evaluating your CV for a test or when
comparing two test systems.
TE
TEbudget (%) = lab 100
TEa
The optional Westgard Advisor allows you to choose a TEa and then suggests
SPC rules based on your test data and Unity Interlaboratory information.
If you laboratory wants to have rules suggested for you, contact your BioRad Representative to receive information regarding Unity Real Time
and the optional Westgard Advisor.
30
Bio-Rad Laboratories
CVR =
CVlab
CVgroup
z-score
The z-score is the number of standard deviations a control result is from the
expected mean and is calculated as follows:
SDexpected
A z-score of 2.3 indicates that the observed value is 2.3 SD away from the
expected mean. A data point with this z-score would violate the 1-2s rule,
but not the 1-3s rule. The z-score appears on the point data screen.
SPC Rules
In 1981 Dr. James Westgard of the University of Wisconsin published an
article on laboratory quality control that set the basis for evaluating
analytical run quality for medical laboratories. The Westgard system is
based on the principles of statistical process control used in manufacturing
nationwide since the 1950s. There are six basic rules in the Westgard
Bio-Rad Laboratories
31
scheme: 1-3s, 2-2s, R-4s, 1-2s, 4-1s, and 10x. These rules are used
individually or in combination (multi-rule) to evaluate the quality of
analytical runs.
The rationale for applying these rules is:
Reduce false rejections made when applying just the 1-2s rule for run
rejection
Increase error detection over that provided when applying just the 1-3s
rule for run rejection
Include rules to detect and distinguish random and systematic error (13s and R-4s to detect random error and 2-2s, 4-1s, and 10x to detect
systematic error)
1-2s
The 1-2s rule is usually a warning rule that is violated when a single control
observation is outside the 2s limits.
Some laboratories consider any quality control value outside its 2s limits to
be out of control, and therefore incorrectly decide that the patient
specimens and QC values are invalid.
An analytical run usually should not be rejected if a single quality control
value is outside the 2s QC limits but within the 3s QC limits.
Approximately 4.5% of all valid QC values will fall somewhere between 2
and 3 standard deviation limits. Laboratories that universally reject values
outside the 2s limit end up rejecting good runs too frequently.
The 1-2s rule was originally designed as a warning rule for manual
application of the Westgard Rules. If one control measurement within a run
exceeds the mean 2 standard deviations in a manual application of
Westgard rules, you should evaluate other controls in the run (within the
run) and in previous runs (across runs) before accepting the run and
reporting the results. With computer-based applications of Westgard rules,
the 1-2s rule is usually not necessary.
32
Bio-Rad Laboratories
Using this rule alone in performing quality control tests causes frequent
rejection of valid runs. According to Dr. Westgard, failure to allow for valid
points between 2s and 3s may result in falsely rejecting:
1-2.5s
Violation of the 1-2.5s rule indicates random error and may also point to
systematic error. This rule is applied within the run only. The following
figure shows an L-J chart with a data point outside the 2.5s limits.
Bio-Rad Laboratories
33
1-3s
The 1-3s rule identifies unacceptable random error or possibly the beginning
of a large systematic error. Any QC result outside 3s violates this rule.
Since only 0.3% or 3 out of 1000 points will fall outside the 3s limits, any
value outside of 3s is usually considered to be associated with a significant
error condition.
While a value outside 3s may be statistically significant, it may not be
biologically or medically relevant due to the fact that modern laboratory
instruments are often more precise than what is needed medically.
The following figure shows an L-J chart with a data point outside the 3s
limits.
1-3.5s
Violation of the 1-3.5s rule indicates random error and may also point to
systematic error. The run is considered out of control when one control
value exceeds the mean 3.5s. This rule is applied within the run only. The
following figure shows an L-J chart with a data point outside the 3.5s
limits.
34
Bio-Rad Laboratories
1-4s
Violation of the 1-4s rule indicates random error and may also point to
systematic error. The run is considered out of control when one control
value exceeds the mean 4s. This rule is applied within the run only. The
following figure shows an L-J chart with a data point outside the 4s limits.
2-2s
The 22s rule detects systematic error only.
Greater than 2s
Within run violationTwo control values in the same run are >2s on the
same side of the mean. Violation of the within run application indicates
that systematic error is present that potentially affects the entire
analytical curve.
Across run violationIn one run, a single level of control is >2s from the
mean and, in the next run, the same level of control is >2s on the same
side of the mean. Violation of the across run application indicates that
systematic error is present, but affects only a single portion of the
analytical curve.
The following figures show an L-J chart with a data point violating the 2-2s
rule, within and across a run.
Bio-Rad Laboratories
35
R-4s
The R-4s rule identifies random error only and is applied only within the
current run. This rule is violated when there is at least a 4s difference
between control values within a single run.
For example, assume both Level I and Level II have been assayed within the
current run. Level I is +2.8s above the mean and Level II is -1.3s below the
mean. The total difference between the two control levels is greater than
4s (i.e. +2.8s (-1.3s) = 4.1s).
36
Bio-Rad Laboratories
The following figure shows an L-J chart with two points that violate the R-4s
rule.
Exceed 1s
The 3-1s and 4-1s rules can be violated both within a control material (e.g.
all Level I control results) and across control materials (e.g. Level I, II, and
III control results in combination).
The following figures show an L-J chart with a data point violating the 3-1s
rule within a run and a 4-1s rule across a run.
3-1s Within Run
Bio-Rad Laboratories
37
7-T
The 7-T rule is violated when a group of seven consecutive data points for a
single level of control show either a "strict" increasing or decreasing
pattern. A "strict" increasing pattern is defined as a series of points that
increase incrementally from the previous point (each point greater than the
last) without a break in the pattern. A "strict" decreasing pattern is the
same pattern in a negative direction.
The following figure shows an L-J chart with points that violate the 7-T rule.
N-x Rules
The N-x rules are violated when there are 7, 8, 9, 10, or 12 control results
on the same side of the mean.
Each of these rules has two applications: within a control material (e.g. all
Level 1 control results) or across control materials (e.g. Level 1, 2, and 3
control results in combination).
38
Bio-Rad Laboratories
Bio-Rad Laboratories
39
H A P T E R
Choosing a QC Procedure
In This Chapter
Recommended Steps ................................................... 41
Troubleshooting QC Results ........................................... 46
Recommended Steps
Ensuring that your SPC Rules (on page 31) are chosen to maximize error
detection while minimizing false rejections is crucial when setting up your
QC system. CLSI (C24-A2, section 5) recommends that you follow these steps
when setting up your QC system:
1
Rules to apply
Location in the run where the control samples are tested (e.g. at
the beginning, in the middle, at the end, or distributed throughout
the run)
Select a QC procedure
Obviously, this is not a simple process, which may explain why many
laboratories fall back on the 1-2s rule. Although the 1-2s rule produces too
many false alarms, it feels safe.
The optional Westgard Advisor makes this arduous task easier.
Clinicians' opinions
42
Bio-Rad Laboratories
Choosing a QC Procedure
Six Sigma
Sigma metrics are useful for quantifying test performance. During the
1980s, Motorola set out to improve their manufacturing process so that
virtually no defective product would be produced. They defined this as
having six sigmas (standard deviations) of process variation fit within the
product tolerances.
Assuming a normal (Gaussian) distribution, the following table shows the
effect of product specifications (expressed as standard deviations) on the
defect rate and defects per million.
Bio-Rad Laboratories
43
SD range
2 SD
4.5
45,400
3 SD
< 0.27
4 SD
~2,700
0.0063
63
5 SD
0.0057
0.57
6 SD
0.000002
0.002
The advantage of controlling a process to six sigma is that the process can
tolerate small shifts without significantly increasing the defect rate. In an
ideal world, all our processes would be six sigma, and we could monitor
them with very simple QC.
Unfortunately, not all processes are six sigma, and, as process capability
decreases, our choice of QC procedures becomes increasingly important in
detecting significant errors. In fact, some processes may have such low
process capacity (i.e. high total error) that they cannot be controlled to a
defined level of quality. This condition would trigger a maximum QC
condition within the optional Westgard Advisor.
For clinical lab tests, the total error (TE) for a test provides an indication of
the test's process capability, because it combines bias and imprecision. This
allows us to calculate sigma using the following formula:
Sigma =
TEa Bias
CV
The optional Westgard Advisor calculates and displays sigma based on your
data for a test and your selected TEa and consensus group.
We can correlate sigma with the TEa as shown in the following table. (The
table assumes that bias is zero.)
Process classification
4-sigma process
3-sigma process
2-sigma process
When bias is not zero, the sigma classification of a process decreases as its
bias increases.
44
Bio-Rad Laboratories
Choosing a QC Procedure
The information in this chapter was abstracted from the Six Sigma Quality
Management and Desirable Laboratory Precision lesson available on
Westgard.com. Dr. Westgard provides an eye-opening correlation between
CLIA performance requirements and those required for a 5 or 6-sigma
process. As Dr. Westgard concludes, "Six Sigma Quality Management sets
demanding standards of performance for laboratory testing processes."
45
OPSpecs Charts and sigma values, which are available if you are using the
optional Westgard Advisor, provide information about a test's maximum
allowable bias (inaccuracy) and imprecision.
Troubleshooting QC Results
Inevitably, your QC system will eventually indicate an out-of-control
situation. What do you do then? In a lesson titled "QC - The Out-of-Control
Problem" available on Dr. Westgard's web site (Westgard.com), Elsa F. Quam
BS, MT (ASCP) discusses two prevalent bad habits and then provides five
good habits to develop when troubleshooting QC problems.
The two bad habits listed by Ms. Quam will be familiar to anyone working in
a clinical laboratory. They represent the incorrect, yet easy-way-out
approach to solving QC problems.
Bio-Rad Laboratories
Choosing a QC Procedure
Recalibrate
A third risky habit not mentioned by Ms. Quam is frequent recalibration.
Laboratories should be concerned about the number of times a test is
recalibrated because each calibration or recalibration potentially introduces
new or additional systematic errors. Frequent recalibrations may indicate a
defective SPC protocol (rules applied, mean, and range in use) set by the
laboratory, instrument malfunction, sub-optimal reagent quality, or failure
to follow the manufacturers instructions and schedule for maintenance.
So what are the good habits?
Good Habits
If we eliminate the bad habits, what do we replace them with? In her article
"QC - The Out-of-Control Problem ", Ms. Quam lists five good habits:
1
Bio-Rad Laboratories
47
The 1-3s and the R-4s rules usually indicate increased random error
because they test the tails or width of a distribution
The 2-2s, 4-1s, and 10-x rules usually indicate systematic error,
because they examine consecutive QC results that exceed the same
limit
48
Bio-Rad Laboratories
Choosing a QC Procedure
Do all the tests with the problem use the same lamp and tests
without the problem use a different lamp?
Do all the tests use the same mode of detection (endpoint vs. rate,
MEIA vs. FPIA)?
Systematic Error
If you observe a sudden shift, inspect the reagent, calibration, and
maintenance records, and note recent actions. For example, if the shift
occurred immediately following a reagent replacement, verify that the lot
number is correct and has been checked out or calibrated, and that the
reagent has been prepared properly, that the reagent is indeed the correct
reagent.
If you observe a systematic trend, review QC records, including
documentation of function checks, prior to taking actions to resolve the
cause. Trends can be caused by a slowly deteriorating reagent, a calibration
Bio-Rad Laboratories
49
Random Error
The causes of increased random error are generally much more difficult to
determine because of their random nature. Random errors are more likely
due to bubbles in the reagent, reagent lines, sampling or reagent syringes,
or improperly mixed/dissolved reagent, pipette tips not fitting properly, a
clog in the pipettor, imprecise pipettor, the power supply, and even power
fluctuations.
Many of these problems can be detected by inspecting the machine during
operation. If a careful inspection provides no clues, you can consult
troubleshooting guides and manufacturer recommendations.
If you repeat the run, and the controls are acceptable, but you are still
concerned that a problem may exist, you may want to perform a precision
run using 10 back-to-back determinations on the same patient sample. This
precision run may identify further imprecision problems. Duplicate analysis
of patient specimens is also recommended when monitoring random error
problems.
Bio-Rad Laboratories
Choosing a QC Procedure
Are the statistical process control (SPC) rules in effect for the test too
restrictive when the capability of the methodology or technology and
total allowable error are jointly considered?
Are the performance goals for imprecision and bias for the test, which
also affect total allowable error, appropriately set?
How frequently do SPC error flags occur for the test during this review
period? Across review periods? Are these frequent error flags due to
inappropriate selection of SPC rules, larger than expected imprecision,
or the presence of bias, do the mean and range need adjustment?
There is probably not enough time to ask all of these questions during each
review cycle. However, each of these questions represents an opportunity
to measure and appraise the effectiveness of the process control in effect
for a specific test.
Bio-Rad Laboratories
51
H A P T E R
Program Basics
In This Chapter
Starting and Exiting the Program .................................... 53
Current Lab, Lot, Test, and Panel ................................... 53
Toolbars .................................................................. 54
Unity Desktop Menu Bar ............................................... 55
Bio-Rad Laboratories From the Windows Start Menu ............. 57
Shortcut Keys............................................................ 57
Functions and Where to Find Them.................................. 59
Selecting a lot number also determines the current lab, but there is no
current test.
The current lab, lot, and test selections determine the active lab, lot, and
test. For example, if you select a test in the tree and then click the L-J icon
on the toolbar, Unity Desktop displays the Levey-Jennings chart for the
selected test.
Because selecting a test also determines the current lab and lot, these
items will be active if you access the "Lab" or "Lot" dialog boxes. In addition,
menu and toolbar items become available or unavailable depending on
which item (lab, lot, or test) is selected in the tree. For example, charts
and reports are available only when a test is selected in the tree. Similarly,
selecting a lot selects its associated lab number.
Toolbars
The Unity Desktop toolbar contains icons (buttons) that you click to
activate specific program functions. The icons themselves are selfexplanatory, especially when using large icons, but you can hover your
mouse over an icon to display a short description. The default toolbar
contains commonly used icons; however, you can customize your toolbar to
contain the functions you use most frequently.
When you choose View | Toolbar, the "Toolbar" dialog box appears.
54
Bio-Rad Laboratories
Program Basics
Your current toolbar icons appear below the dialog box's title bar. The
"Large toolbar" and "Small toolbar" option buttons control the appearance of
the toolbar icons. Small icons take up less room on the screen; however,
large icons contain labels, which may be useful until you become familiar
with the program. In the preceding figure, the "Large toolbar" option is
selected.
The lower part of the dialog box contains an "Available list" and a "Selected
list" of icons with several buttons:
AddSelect an icon in the "Available list", and click Add to move the
icon to the "Selected list."
RefreshClick to change the icon display under the title bar to your
current selections.
Bio-Rad Laboratories
55
Menu Item
Submenus
File
Open
Save ChartViewing, Printing and Saving Charts
Print Chart
Exit
Select
Lab
Lot
Test
Panel
View
Status Bar
Refresh
Toolbar
Advisors
Reports
General
Supervisory
Charts
Listings
Configure
Tools
Security
Unity Interlab
Utilities
Actions and Comments
Setup
Help
Help Topics
Bio-Rad on the Web
QCNet on the Web
About Unity Desktop
User Guide
56
Bio-Rad Laboratories
Program Basics
Wait while the Reader setup file is recomposed. This may take several
minutes.
When the "Adobe Reader 6.0.1 - Setup" dialog box displays, follow the
Wizard to install Adobe Acrobat Reader.
Shortcut Keys
The following table summarizes the shortcut keys available in Unity
Desktop. In the table, the Key column indicates the function key, while
the remaining columns indicate any additional keys that need to be pressed
along with the function key. For example, to open Bench Review, press the
CTRL key and the F2 key simultaneously. In this guide, CTRL+F2 indicates
these keys (i.e. the two keys to be pressed are shown with a plus sign [+]
between them).
Bio-Rad Laboratories
57
Key Plain
F1
CTRL
ALT
Help
Log off
F2
F3
Shift
Open "Test"
dialog box
Open "Lot"
dialog box
Open "Lab"
dialog box
Open "Panel"
dialog box for
selected test
Exit Unity
Desktop
F4
F5
Next test
Previous test
F6
Net lot
Previous lot
F7
Next lab
Previous lab
F8
Next panel
Previous panel
F9
Point Data
Report
F10
Supervisor
Report
F11
Set Initial
Date in data
entry
F12
58
Youden chart
for selected
test
Statistical Report
"Administrator" "Archive/Restore"
dialog box
dialog box
Import utility
Bio-Rad Laboratories
Program Basics
Where is it?
See
Acrobat Reader Setup (see
"Activation Keys and Licensing"
on page 218)
Archive/Restore
Tools | Utilities |
Archive/Restore
Automatic
Tools | Setup
transmission, monthly
To Activate Automatic
Monthly Transmission (see "To
Activate Automatic Monthly
Transmission for Monthly Unity"
on page 174)
Tools | Setup
F12
F12
Database updates,
automatic
Tools | Setup
Tools | Setup
Email transmission
file to Bio-Rad
Export
Exporting Data
Import
F12
Import Overview
Operator Setup
Bio-Rad Laboratories
59
Password expirations
PDF reports
Tools | Setup
Rejection log
Tools | Utilities |
Archive/Restore
Send/Receive data
Transmissions Files
Toolbar, customize
View | Toolbar
Transmission,
automatic monthly
Tools | Setup
To Activate Automatic
Monthly Transmission (see "To
Activate Automatic Monthly
Transmission for Monthly Unity"
on page 174)
Transmission, manual
Update database
Updating Database
Information (on page 211)
Write transmission
file
60
Bio-Rad Laboratories
H A P T E R
Define, modify, and delete user profiles for individual users, including
their passwords and permission profiles.
The user profile section of the Administrator dialog box consists of the
users ID, name, password, and initials.
The User Name can contain any combination of letters and numbers up
to 60 characters.
To Add a User
1
Click New.
Under "User profile," type the User ID, User name, Password, and
Initials of the new user.
If desired, select the "Set password expiration period to" check box and
select a time period from the drop-down list. See Passwords (on page
63) for more information.
Select a data entry option. Only one data entry option can be chosen at
any given time.
Bio-Rad Laboratories
To Modify a User
1
Select the user to be modified from the drop-down list under "User ID."
Change the users profile, permission, and data entry option as desired.
To Delete a User
Deleting a user removes the user's associated User ID and Password from the
"Login" and "Administrator" dialog boxes. However, the user's initials will
still appear in the OP column of data entry screens for any data points that
the user previously entered.
1
Select the user to be deleted from the drop-down list under "User ID."
Click Delete.
The screen displays a message asking you to confirm that you want to
delete the user.
Passwords
Creating Passwords
Please be aware of the following password facts:
Bio-Rad Laboratories
63
User ID - admin
Password - sa
Initials - sa
The default administrator login allows new users initial access to the
software. After you create other users, you should delete this user or
change the password.
At least one user must have administrator privileges. If only one user on the
system has administrator privileges, the program will not allow you to
delete this user or to clear (uncheck) the Administrator check box until
you define another user with administrator privileges.
Additional users can be given administrator privileges as required; however,
giving all, or even most, users these privileges eliminates the security
precautions gained by using passwords. (Unity Desktop also creates a default
database password as described in Default Database Passwords.)
Password Expiration
Unity Desktop allows users with permission to manage users to specify that
passwords expire after a set length of time. Use of this feature is entirely
optional; however, you may want to use it for added security or to satisfy
regulatory requirements.
You will probably want to set the password expiration period as you add
new users; however, you can update users to add or change the expiration
period. When you apply a password expiration period, you can choose to
apply it to only the current user or to all users.
64
Bio-Rad Laboratories
If every user will have the same expiration period, it is obviously easier to
specify it once and apply it to all users. In addition, specifying "Never" as
the period and applying it to all users is a convenient way to cancel the use
of the password expiration feature.
Choosing Permissions
The user permissions section of the "Administrator" dialog box (ALT+F11)
contains check boxes for a number of program functions. To give a user
permission to perform a particular function, click inside the box next to the
function to place a check mark in it. To clear the permission, click inside
the box again. Assigning permissions allows you to customize what functions
each user has access to.
A check in the Administrator check box automatically selects all user
permissions and sets the Data option to "Edit all data." Removing the check
from the Administrator check box clears all user permissions and sets the
data entry option to "View data only."
Unity Desktop divides permissions into groups which are described in the
following sections.
Administration/Setup Permissions
The Administration/Setup group of permissions contains the following
individual permissions:
Manage users
Users with permission to manage users can:
65
Delete users
Make changes on the Setup dialog box. See The Setup Dialog Box
(on page 222) for more information.
Operator setup
Users with this permission can use operator setup to define the
operator initials to appear on the data entry screen for imported data
when the import file does not contain operator initials.
Data Permissions
The Data section of the "Administrator" dialog box contains option buttons
to allow the Administrator to assign a data entry and editing category.
Enter new data onlyThe user can enter new data but cannot edit or
delete any data.
Edit last lineThe user can enter new data and can edit the last line of
data entered.
Edit all dataThe user can enter new data and can edit any line of data.
View data onlyThe user can view data on the screen but cannot enter,
edit, or delete data.
Database Permissions
The Database group of permissions contains the following individual
permissions:
66
Bio-Rad Laboratories
Condense data
The condense data feature allows you to convert point data to
summarized data within your database. The condense feature is
unavailable when the RiLiBK module is installed.
Reconcile data
The reconcile data feature should only be used when you are instructed
to do so by a Bio-Rad software support representative. This feature will
re-transmit your entire data history to the Unity Interlaboratory
Program instead of just recent changes. This makes the file size much
larger.
Access the Westgard Advisor (optional module for Unity Real Time
only).
Edit Total Error-BV bias and imprecision goals when manually specifying
a target value.
Manage labs/lots/tests
Users with permission to manage labs/lots/tests can:
Add and delete valid Bio-Rad lab numbers. (You cannot delete your
primary lab number.)
67
Manage panels
Users with permission to manage panels can add, edit, and delete
panels.
Import data
Users with this permission can manually import data. Autoimport (e.g.
directly from a LIS or third-party peripheral) runs as a background
process and is not controlled by permissions.
Export data
Users with this permission can manually export data.
68
Bio-Rad Laboratories
H A P T E R
Labs
The Unity Interlaboratory Program uses a hierarchical structure of lab
number, lot number, and test codes to uniquely identify each test in your
laboratory. You must have at least one lab number in order to use Unity
Desktop. Lab numbers are a unique set of 6 digits assigned to your lab by
Bio-Rad. To get started, use the lab number provided with your startup
package. If you need additional lab numbers, contact your Bio-Rad software
support representative.
All lab numbers must be provided by Bio-Rad Laboratories.
Type the six-digit laboratory number provided by Bio-Rad. (If you need
a lab number or cannot remember your lab number, contact your BioRad software support representative.)
Click Add. (If you click Add without completing a required field, a
message will prompt you to supply the missing information.) The lab
number appears in the "Open labs" list.
Click Close to return to the main screen or Lot to define the lot
numbers to be used under this lab number.
70
Bio-Rad Laboratories
In the "Open labs" list, select the lab number to be updated. (You
cannot update information for a closed lab. You must reopen it first.)
Click Update.
Click Close to return to the main screen or Lot to go to the Lot dialog
box.
Duplicating Labs
Unity Desktop allows you to copy information defined for one lab number
into a second lab number. After duplication, you can update the Lab profile
for the second lab number as required. Only open lab numbers can be
duplicated.
Duplicating a lab number creates a copy that contains all open lots and tests
in the original lab number. The SPC rule and analytical goal profiles are also
copied for open tests. In addition, you can choose whether or not to
duplicate fixed means and/or fixed SDs.
Duplicating a lab number does not copy QC data.
Select the lab number to be duplicated from the "Open labs" list.
The selected lab's profile appears. (You cannot duplicate a closed lab
number. You must open it first.)
Click Duplicate. The screen displays the "Duplicate Lab dialog box.
Type the target lab number. (Until you type a valid lab number, the OK
button is unavailable and appears dimmed.)
If you would like to copy fixed means and/or SDs, you can select either
or both of the "Duplicate Fixed Mean" and "Duplicate Fixed SD" check
boxes. These check boxes operate independently so that you can copy
just fixed means or fixed SDs.
Click OK. The target lab number appears in the "Open labs" list.
71
Deleting Labs
Deleting a lab number deletes all data for all tests within the deleted lab
number. Information within the deleted lab number cannot be retrieved at
a later date. If you want to make a lab number inactive while retaining its
information, use the Close Lab feature of the program. You can also export
its data as a text, delimited, or import file before you delete it.
Select the lab number to be deleted from the "Open labs" list. (You
cannot delete a closed lab. You must reopen it first.)
Click Delete. A message appears asking you to confirm that you want to
delete the lab number.
Are ignored by Unity Desktop (i.e. data located within the closed lab
number cannot be viewed, printed, or changed.)
Can be reopened if you decide to use the lab number at a future date.
Select the lab number to be closed from the "Open labs" list.
Click Close Lab. The lab number appears in the "Closed labs" list.
Select the lab number to be opened from the "Closed labs" list.
Click Open Lab. The lab number appears in the "Open labs" list.
Lots
You can use both Bio-Rad and non-Bio-Rad lots with Unity Desktop.
However, Unity Interlaboratory reports are only available for Bio-Rad
products.
73
Click the Lot button on the toolbar to display the Lot dialog box.
Select the appropriate Bio-Rad control from the "Control name" dropdown list.
Select the lot number of the control from the "Lot number" drop-down
list.
Click Add. The lot number appears in the "Open lots" list.
Click Close to return to the main screen or, to begin defining tests,
select the lot in the "Open lots" list, and then click Test.
Click the Lot button on the toolbar to display the Lot dialog box.
74
When setting an expiration date, the expiration date will default to one
year, specifically, the last day of the current month plus one year. For
example, if you create a lot on March 13, 2007, its default expiration
date will be March 31, 2008 at midnight (11:59:59).
6
Click Add. The lot number appears in the "Open lots" list of the Lot
dialog box.
Click Close to return to the main screen or select the lot in the "Open
lots" list, and click Test to begin defining tests.
Duplicating Lots
Duplicating lot numbers allows you to create a new lot number with the
tests and test parameters in an existing lot, greatly simplifying lot changes.
(If you import data, Unity Desktop can create new lots for you based on the
lot number in the import file. To activate this feature, you must select the
"Create new lots if necessary" check box on the "Import" dialog box. (Press
F12 to access this dialog box.))
When you duplicate a lot, the new lot number will be configured exactly the
same as the old lot number but will contain no data.
You can duplicate both Bio-Rad and Non-Bio-Rad lot numbers.
If you want to duplicate a closed lot, you must re-open it first. The original
closed lot can then be re-closed so that it is ignored by Unity Desktop and is
omitted from your Unity Interlaboratory reports.
Open tests within the lot (Closed tests are not duplicated.)
75
Fixed mean
Fixed SD
Click the Lot button on the toolbar to display the Lot dialog box.
In the "Open lots" list, select the lot to be duplicated. The product
name and lot number appear under "Lot information."
Select the lot number you want to duplicate to from the drop-down list.
Only unexpired lot numbers in the same product group will appear in
the list.
Select the check boxes for each test parameter that you want to
duplicate.
Click OK. The new lot number appears in the "Open lots" list. The lot
that you duplicated from is unchanged.
Click the Lot button on the toolbar to display the Lot dialog box.
In the "Open lots" list, select the lot to be duplicated. The product
name and lot number appear under "Lot information."
Specify a name and expiration date for the lot. The new lot appears in
the "Open lots" list.
76
Bio-Rad Laboratories
Deleting Lots
You can delete lot numbers at any time. You do not have to wait until after
the lots expiration date. Be aware, however, that deleting a lot number
erases all data in all tests in the deleted lot.
As an alternative to deleting the lot number, you can close it so that it no
longer appears in the navigation tree or on your Unity Interlaboratory
reports. If you choose to delete a lot, you might want to export its data
before deleting it.
Deleting a lot number permanently erases all data in all tests in the
deleted lot.
Click the Lot button on the toolbar to access the Lot dialog box.
Select the lot you want to delete in the "Open lot" list.
Click OK.
Editing Lots
You can edit both Bio-Rad and non-Bio-Rad lots, even if the lot contains
data.
Select the Bio-Rad lot you want to update in the navigation tree.
Click the Lot button on the toolbar. The Lot dialog box appears with
the lot selected.
Select the correct lot number from the drop-down list. Only lots in the
same product group appear in the list.
Bio-Rad Laboratories
77
Click OK.
Select the non-Bio-Rad lot you want to update in the navigation tree.
Click the Lot button on the toolbar. The Lot dialog box appears with
the lot selected.
Enter a new master lot number and expiration date, then click OK.
To Close a Lot
1
Select the lot you want to close from the navigation tree.
Click the Lot button on the toolbar. The Lot dialog box appears
within the lot selected in the "Open lots" list.
Click Close Lot. The lot appears in the "Closed lots" list.
To Open a Lot
1
78
Select the lab number containing the closed lot in the navigation tree.
Click the Lot button on the toolbar to display the Lot dialog box.
Select the lot number to be opened from the "Closed lots" list.
Click Open Lot. The lot number appears in the "Open lots" list.
Bio-Rad Laboratories
79
H A P T E R
Tests
In This Chapter
Overview of Tests....................................................... 81
Adding Tests ............................................................. 82
Test Rules and Settings ................................................ 87
Duplicating Tests ....................................................... 88
Updating Tests .......................................................... 88
Closing and Opening Tests ............................................ 89
Deleting Tests ........................................................... 90
Sorting Tests............................................................. 91
VITROS Slide Generations ............................................. 92
Overview of Tests
Unity Interlaboratory Program defines tests using six parameters. Used
with the lab and lots numbers, these codes uniquely identify your tests in
the Unity Interlaboratory Program.
The six parameters are:
Analyte
The analyte is the constituent being measured (e.g. albumin, calcium,
glucose, etc.).
Instrument/Kit
When using RIA kits, code for the kit rather than the gamma counter.
Reagent type
Factored
Factored is applicable when your quality control data is
mathematically altered to simulate results obtained on another
instrument or at another temperature.
More reagents
More reagents is applicable if you are not using a reagent supplied
by the instrument manufacturer. Click "More reagent selections",
and then select the appropriate reagent under "More reagents." If
the reagent you are using does not appear, choose "Other", and
type a free-text reagent name.
Method
Only methods that Bio-Rad considers "valid" for the selected analyte
appear in the list. If your method does not appear, please contact your
Bio-Rad software support representative.
Unit of measure
Only units that Bio-Rad considers "valid" for the selected analyte appear
in the list. If your unit does not appear, please contact your Bio-Rad
software support representative.
Temperature
For enzymes, you can choose an available temperature. For all other
analytes, the temperature defaults to "no temperature" and cannot be
changed.
Adding Tests
You can add tests to Unity Desktop both directly and indirectly. Direct
methods include manual addition and instrument setup; indirect methods
82
Bio-Rad Laboratories
Tests
include duplication of labs or lots, the create new tests import features,
and changing VITROS slide generation.
Select the lot to which you want to add tests in the navigation tree.
Click the Test icon on the toolbar to display the "Test" dialog box.
Analyte
Bio-Rad Laboratories
83
Instrument/Kit
Alternate formulation/standardization
You will usually only select this option when asked to do so by your
Bio-Rad software support representative or when you receive a
notice asking you to change the reagent type for a test.
Factored
Select factored when your quality control data is mathematically
altered to simulate results obtained on another instrument or at
another temperature.
More reagents
If you choose this option, a comprehensive list of reagents appears.
Simply select a reagent from the list.
When your selections are correct, click Add. The new test appears in
the Open tests list.
If a code you need to use is missing, you may need to update your code list
as described in Updating Database Information (on page 211).
Bio-Rad Laboratories
Tests
You cannot define tests with a mixture of SI and Conventional units at the
same time when using Instrument Setup. You can, however, use the
function twice to add tests using both unit types.
Select the lot to which you want to add tests in the navigation tree.
Click the Test icon on the toolbar. The Test dialog box appears.
Click Instrument Setup. The "Instrument Setup" dialog box appears with
a list of available instruments.
Bio-Rad Laboratories
85
Click OK. The Instrument Setup dialog box appears with a list of all
available tests from the Method Guide. A selected check box preceding
a test indicates that the test will be created. (The Clear All button will
deselect all the check boxes while the Select All button will select all
the check boxes.)
Select and clear the check boxes so that only the tests you want to add
are selected.
10 Click OK to return to the Test dialog box. The newly added tests
appear in the "Open tests" list.
If you added tests for a VITROS instrument, be sure to change your
VITROS Slide Generations (on page 92).
Click the Test icon on the toolbar to display the Test dialog box.
Make a selection for each test code from the drop-down lists. Select
"Other" for any code that does not appear. (Type the letter O to move
to the correct area of the list.)
When all selections are made, click Add. The Other dialog box
appears.
86
Bio-Rad Laboratories
Tests
Type a free-text name in each available field, and then click OK.
Because the Unity Interlaboratory Program does not create reports for tests
that use "Other" codes, the use of "Other" codes should be considered a
stop-gap measure, and we request that you also contact your Bio-Rad
software support representative to request that "Other" codes be added to
the code list.
Levels in use
Levels in use defaults to the number of levels available for the control
product. Change this selection if you do not use all available levels.
When you clear a level's check box, the level is omitted from the data
entry screen, which simplifies tabbing during manual data entry.
This item is available from both the "Lot" or "Test" dialog boxes.
Bio-Rad Laboratories
87
Decimal places
You are allowed to specify the number of decimal places (zero to four)
to which you will report results for a level. You can specify different
decimal formats for each level.
This item is available from both the "Lot" or "Test" dialog boxes;
however, it is not usually possible to accurately assign decimal places at
the lot level. If you use a certain number of decimal places for most
analytes, you might want to set this number at the lot level and modify
individual tests as necessary at the test level.
Duplicating Tests
You cannot duplicate individual tests from the Test dialog box. However,
duplicating a lab or lot number duplicates its tests.
Updating Tests
You can change any test code for an existing test, even if the test contains
data. This feature is useful if you make a mistake in coding a test or if your
Bio-Rad software support representative asks you to make a change (e.g. to
alternate formulation/standardization).
You cannot update a qualitative test to quantitative tests or vice versa.
See Update VITROS Slide Generations (on page 94) for information specific
to tests using slide generation.
To Update a Test
1
88
Tests
Click the Test icon on the toolbar to display the "Test" dialog box.
Change the selections from the appropriate test code lists (analyte,
instrument/kit, reagent, method, unit, and temperature).
Click Update when all selections are correct. The "Update Test" dialog
box appears.
All LabsChoose this option to update all identical tests in all lab
numbers. (This option updates all identical tests in the database.)
Identical tests are those that have the same Unity test codes for
analyte, instrument/kit, reagent, method, unit, and temperature.
6
Click OK.
To Close Tests
1
Click the Test icon on the toolbar to display the "Test" dialog box.
Click Close Test. The tests are removed from the "Open tests" list and
appear in the "Closed tests" list. Closed tests are also removed from the
navigation tree.
Bio-Rad Laboratories
89
To Open Tests
1
Click the Lab icon on the toolbar and select the lab number containing
the closed tests from the "Open labs list."
Select the lot number containing the closed tests from the "Open lots"
list.
Click Open Test. The tests are removed from the "Closed tests" list.
The tests now appear in the "Open tests" list and the navigation tree
and become available for data entry.
Deleting Tests
You can delete tests at several levels:
Lab
Deleting a lab also deletes its associated lots and tests. See Deleting
Labs (on page 72).
Lot
Deleting a lot deletes its associated tests. See Deleting Lots (on page
77).
Test
Within the Test dialog box, you can delete tests that you select.
Bio-Rad Laboratories
Tests
Please notice that the "Test" dialog box has two delete buttonsone in
the "Open tests" list and the other in the "Closed tests" list. Which Delete
button you click determines the test(s) that are deleted. Be careful to
click the button in the correct area to avoid inadvertently deleting a
test.
To Delete Tests
1
Access the "Test" dialog box for the lab/lot containing the test(s) you
want to delete.
Select the test(s) you want to delete from the "Open tests" or "Closed
tests" lists. (If you want to delete both open and closed tests, you will
need to select them separately.)
The Test dialog box supports both SHIFT+Click and CTRL+Click so
you can select multiple tests at a time for opening and closing. See
Selecting Items in Lists (on page 231) for more information.
Click Delete.
Sorting Tests
You can sort tests for display in both lab mode and panel mode. For lab
mode, the order the tests appear in the Open tests list of the Test
dialog box controls the order on the lab navigation tree as well as on the
reports and charts. The Open test list order also controls the nextprevious assignments when using Shortcut Keys.
For panel mode, the order tests appear in the Selected tests list of the
Panel Tests dialog box controls the order on the lab navigation tree as
well as on reports and charts. The Selected tests list order also controls
next-previous assignments when using shortcut keys.
Select the lab/lot combination for which you want to sort tests in the
lab navigation tree.
Click the Test icon on the toolbar to access the "Test" dialog box.
Bio-Rad Laboratories
91
Select a test and drag it to the location where you want it to appear.
Then release the mouse button to drop the test. See Drag-and-Drop
Sorting (on page 232) for more information.
Click the Panel icon on the toolbar to display the "Panel" dialog box.
Select a test and drag it to the location where you want it to appear.
Then release the mouse button to drop the test. See Drag-and-Drop
Sorting (on page 232) for more information.
Click OK.
Bio-Rad Laboratories
Tests
RetainChoose this option if you want to leave the existing test open
and available for data entry. This is the default selection.
CloseChoose this option if you want to stop data entry for the existing
test and eliminate it from the navigation tree and your Unity
Interlaboratory reports.
Selected testChoose this option to apply the new slide generation only
to the currently selected test. This is the default selection.
Current labChoose this option to apply the new slide generation to all
identical tests within the current lab number.
All labsChoose this option to apply the new slide generation to all
identical tests within the current database.
Click the Test icon on the toolbar to open the "Test" dialog box.
In the "Change Slide Generation" dialog box, make a selection for each
of the following:
93
Select the VITROS test you want to modify in the navigation tree.
Click the Test icon on the toolbar to open the "Test" dialog box.
Information for the test you selected in the navigation tree appears
under "Test information."
Select an option for the range of tests to apply the new slide generation
to.
7
94
All labsChoose this option to apply the new slide generation to all
identical tests within the current database.
H A P T E R
Panels
Panels and data groups allow you to customize tests for data entry and
display. A panel is a user-defined group of tests organized for simplified
data entry across lab and lot numbers. Panels provide a useful way to
customize the organization of tests in a convenient manner such as grouping
a number of different tests performed on a single instrument or the same
test performed on multiple instruments.
Unity Desktop allows you to create any number of panels. Tests added to a
panel can come from any lab number and lot number in the system. A single
test may be added to any number of panels, but can only appear once in
any one panel.
When you open a data entry screen for a test from the Panel tab of the
navigation tree, the program is in panel mode, and the shortcut keys for
navigating in data entry apply to the tests in the panel. The panel name
appears with the current status information.
You can delete a panel at any time. Because panels are simply a way of
organizing tests for data entry, tests within a panel are not deleted when
the panel is deleted.
Available tests tree-with the familiar lab, lot, and test hierarchy
Select a lab number in the tree, and click Add to add all the tests in
the lab number.
Select a lot number in the tree, and click Add to add all the tests in the
lot number.
Select an individual test, and click Add to add the test. Alternatively,
you can double-click the test to add it.
Double-click a collapsed lab or lot in the tree to expand it. See Tree Views
(on page 229) for more information.
AddClick to add the selected item (lab, lot, or test) to the panel.
96
Bio-Rad Laboratories
To Add a Panel
1
Click the Panel icon on the toolbar to display the "Panel" dialog box.
Under "Available Tests," select the first test you want to add to the
panel. (Click the + signs to the left of the lab and lot numbers as
necessary to expand the tree to the test level.)
The test appears under "Selected tests." To remove a test from the
panel, select it, and click Remove.
Click OK to create the new panel and close the "Panel Tests" dialog box.
The "Panel" dialog box displays with the new panel in the "Panel" list.
To Update a Panel
1
Click the Panel icon on the toolbar to display the "Panel" dialog box.
Use the features described in The "Panel Tests" Dialog Box (see "Panel
Tests Dialog Box" on page 95) to modify the panel.
To Delete a Panel
1
Click the Panel icon on the toolbar to display the "Panel" dialog box.
Click OK to delete the panel. The "Panel" dialog box appears with the
panel deleted.
97
Click the Panel icon on the toolbar to display the "Panel" dialog box.
Drag and drop tests in the "Selected tests" list until all tests are in the
order you want them to appear. See Drag-and-Drop Sorting (on page
232) for more information.
Data Groups
Categorizing tests by group name is a useful way to denote tests having
"something in common. For example, tests having the same lab and lot
number can be grouped according to reagent type. In this case, all tests
using a specific reagent (say Reagent 1) may be grouped into Group 1 while
all tests using another reagent (say Reagent 2) may be grouped into Group
2. Similarly, tests may also be grouped according to calibrators, the
standardization matrix of a kit, etc.
Overview of groups
The Month and Group options are available for both point data entry
and summary data entry.
From the point/summary data entry screen, select or clear the Group
check box to toggle between Month and Group displays.
When you change the display to Group only, the Summary Statistics
area is changed. Unity Desktop does not arrange data entry rows into
groups.
Until you define a data group, Unity Desktop assigns all data points to
group 0 (zero).
Your Default data entry mode selection from the Setup dialog box
is global (it applies to all tests).
When no Group has been defined for a test, the Summary Data based on
Group will be the same as that based on Cumulative.
98
Bio-Rad Laboratories
On the "Configure Unity Desktop" tab, select the "Group" option under
"Current data."
Select (check) the Group check box. The data screen refreshes and a
"Group" column appears.
If the test has no data points, enter a row of data. Then Tab to the
"Group" column and type a group name, which can be
alphanumeric. As you continue to enter data, Unity Desktop will
assign the group name to each new data entry row, until you assign
another group name.
If the test has data points, you can add group names or edit existing
group names.
Unity Desktop assigns successive data entry rows to the same group
until you type another group name in the Group column.
4
Continue entering data or click Save to close the data entry screen.
Bio-Rad Laboratories
99
H A P T E R
Applying QC Rules
In This Chapter
Overview of Rule Evaluation .......................................... 101
Setting SPC Rules ....................................................... 104
Using Fixed Means and SDs with SPC Rules ......................... 105
SPC Rules: Tabular Summary ......................................... 107
If any point within a run violates a rejection rule, Unity Desktop rejects
the entire row.
Rule Statuses
Only point data is evaluated against active SPC rules; summary data is not
evaluated. In addition, inserted or edited data points that are not on the
last line of data are not evaluated.
Each SPC rule can have one of three statuses:
Reject
Unity Desktop rejects data points that violate the rules and excludes
them from summary statistic calculations. The rule(s) violated appear in
the "Rules" column on the data entry screen.
102
Bio-Rad Laboratories
Applying QC Rules
Warn
Unity Desktop accepts data points that violate the rule and the violated
rule appears in the "Rules" column of the data entry screen.
Rules set to reject or warn are the active rules for the test.
Off
Group 1
Group 2
Group 3
Group 4
Group 5
Group 6
Group 7
1-5s
2-2s
(2 of 3)2s
12-x
7T
R-4s
4-1s
1-4s
1-2s
1-2s
10-x
1-3.5s
9-x
1-3s
8-x
1-2.5s
7-x
3-1s
1-2s
Bio-Rad Laboratories
103
(Group numbers in the table are for clarity only and do not imply any
assignment of rules to artificial groups.)
Close any open data entry screens. (You can view a test's settings while
in data entry, but you cannot make changes.)
Click the SPC Rules icon on the toolbar to open the "Test Settings"
dialog box.
On the "Rules" tab, click an option (reject, warn, or off) for each rule.
Optionally, select the "Apply to all lab numbers" check box to apply the
selected profile to all identical tests defined in the database. (Identical
tests in this context are those that share the same Unity codes for
analyte, instrument/kit, reagent, method, unit, and temperature.)
104
Bio-Rad Laboratories
Applying QC Rules
Close any open data entry screens. (You can view a test's settings while
in data entry, but you cannot make changes.)
Click the Lot icon on the toolbar. The "Lot" dialog box displays with the
lot selected. (You can change your lot selection at this time, if
desired.)
On the "Rules" tab, click an option (reject, warn, or off) for each rule.
Optionally, select the "Apply to all lab numbers" check box to apply the
selected profile to the selected lot number in all lab numbers in which
it is defined.
Bio-Rad Laboratories
105
If you set both a fixed mean and a fixed SD, Unity Desktop begins SPC
rule evaluation immediately, even if the test does not have the
specified number of points before rule evaluation. If you do not specify
both a fixed mean and a fixed SD, rule evaluation begins after you have
collected the specified number of points before rule evaluation.
When you specify a fixed mean and/or SD, Unity Desktop uses the
values for new data points entered for the day you set the fixed
statistic(s).
Unity Desktop continues using these values until you either specify new
values or clear a value. When a fixed mean or fixed SD field is blank,
Unity Desktop uses the "floating" mean and/or "floating" SD/percentage.
On the data entry screen, click the + in any data row to view the
evaluation statistics used to evaluate the row.
The fixed mean and fixed standard deviation/percentage statistics are
for your internal use only. They are not transmitted to the Unity
Interlaboratory Program and do not appear on Unity Interlaboratory
reports.
If you duplicate an existing lot number, you can choose to duplicate the
fixed means and/or SDs from the existing lot to the new lot as
described in Duplicating Lots (on page 75).
Close any open data entry screens. (You can view a test's rules and
settings while in data entry, but you cannot make changes.)
Click the SPC Rules icon on the toolbar to open the "Test Settings"
dialog box.
Type a fixed mean and/or SD for each level of control material for
which you want to use fixed statistics.
The fixed mean and fixed SD fields are independent of each other. You
can use them for some levels and not others just as you can set a fixed
mean without a fixed SD and vice versa.
106
Bio-Rad Laboratories
Applying QC Rules
Click OK to save your changes and close the dialog box or Apply to save
your changes and leave the dialog box open.
Rule
Usual status
Error type
When violated
Notes
1-2s
Warning
Random or
systematic
A single control
observation is
outside the 2s
limits
When used as a
rejection rule, 1-2s
will yield a high
proportion of false
rejections.
1-3s
Rejection
Random or
the
beginning
of large
systematic
A single control
observation is
outside the 3s
limits
2-2s
Rejection
Systematic
Two consecutive
QC results are
outside the 2s
limits on the
same side of the
mean
R-4s
Rejection
Random
There is at least a
4s difference
between control
values within a
single run
3-1s
Warning
Systematic
Three consecutive
results exceed 1s
on the same side
of the mean
Bio-Rad Laboratories
107
Rule
Usual status
Error type
When violated
Notes
4-1s
Rejection
Systematic
Four consecutive
results exceed 1s
on the same side
of the mean
Low probability of
false rejections as
long as the betweenrun SD is low. False
rejections increase as
the between-run SD
becomes larger.
(2 of
3) 2s
Warning
Systematic
Two of three
levels of control
within the same
run exceed 2s on
the same side of
the mean
This rule is a
variation of the 2-2s
rule and is applicable
when testing 3 or
more levels of control
material.
7, 8,
and
9-x
Warning
Systematic
x number of
consecutive
results on the
same side of the
mean
Because of the
extreme sensitivity of
these rules, they
should be used
sparingly, if at all.
10-x
Rejection
Systematic
10 consecutive
results on the
same side of the
mean
12-x
Rejection
Systematic
12 consecutive
results on the
same side of the
mean
108
Bio-Rad Laboratories
H A P T E R
10
Entering Data
In This Chapter
Data Entry Overview ................................................... 109
Useful Features of Data Entry ........................................ 110
Unity Desktop Data Screens ........................................... 114
Point Data Entry ........................................................ 117
Summary Data Entry ................................................... 123
Qualitative Data Entry ................................................. 125
Point data
Use this tab to enter individual data points. Point data can be
evaluated by both statistical process control rules (SPC rules) and
analytical goals.
Summary data
Use this tab to enter a monthly mean, standard deviation, and
number of data points. Summary data is not evaluated by SPC rules
or analytical goals.
110
Bio-Rad Laboratories
Entering Data
Within data entry, you can navigate among the tests using the shortcut keys
shown in the following chart:
Shortcut
Action
Within
F5
Next test
SHIFT+F5
Previous test
F6
Next lot
SHIFT+F6
Previous lot
F7
Next lab
SHIFT+F7
Previous lab
F8
Next panel
SHIFT+F8
Previous panel
Bio-Rad Laboratories
111
112
Bio-Rad Laboratories
Entering Data
On the Configure Unity Desktop tab, select a Default data entry mode
option (Point or Summary).
Your selection applies to all tests. You cannot specify a default data entry
mode at the lab, lot, or test levels.
Set Date
Set Date, which can be useful if you manually enter data for several days at
a time, allows you to choose to increment or maintain the date.
Choose Increment if you run each control level only once per day.
Increment increases the day by one each time you move to a new row.
Choose Maintain if you run each control level more than once per day.
Maintain applies the same date/time to each row as you add data rows.
When you are ready to move to the next day, click Set Date, and select
Increment.
You can toggle between the two selections within the same data entry
session.
Once you have chosen a set date, Unity Desktop maintains it until you do
one of the following:
Exit Unity Desktop. If you close a data screen for a test and open
another data screen, the original set date remains in effect, unless the
test has data points entered after the set date. For tests that have data
points entered after the set date, Unity Desktop applies the current
calendar date/time to the active data entry row.
113
Select the part of the date you want to change and use the up and
down arrows to change the date to that of the earliest data point you
want to enter.
Enter your data points. Unity Desktop increments the date by one each
time you move to a new data entry row.
Click the down-arrow in the date field and select the correct date from
the calendar.
114
Bio-Rad Laboratories
Entering Data
Statistics
Displays the mean, SD, CV and number of data points. Select and
clear the "Group" check box to toggle between displaying
monthly/cumulative statistics and group/cumulative statistics for
point data.
Chart
Displays a Levey-Jennings chart based on the available point data.
If you have entered only summarized data, the chart will be blank.
Fixed Mean and SD (on Data Screens) (on page 116)Displays the
fixed mean and/or SD defined on the "Settings" tab of the "Test
Settings" dialog box. Used for evaluation of point data only.
Unity codes used to define the test (i.e. analyte, method, instrument,
reagent, unit, and temperature)
Statistics/Chart Area
The Statistics/Chart area of data entry screens has two tabs"Statistics" and
"Chart."
Bio-Rad Laboratories
115
Statistics
Displays the mean, SD, CV, and number of data points. Select and clear the
"Group" check box to toggle between displaying monthly/cumulative
statistics and group/cumulative statistics for point data.
Chart
The chart option displays a Levey-Jennings chart based on the available
point data. If you have entered only summarized data, the chart will be
blank. Because the display shows the statistics for the currently selected
data row, it is dynamic and changes as you tab through rows of data.
116
Bio-Rad Laboratories
Entering Data
ValueThe cell where you type your QC value when entering data
manually. Imported values also appear in this cell.
Y/NIndicates whether the data point was accepted (Y) or rejected (N).
Unity Desktop automatically rejects any data row that contains a point
that violates a rejection rule. If you have permission to edit data, you
can manually change the accepted/rejected status of a data point as
described in Changing a Data Point's Accepted/Rejected Status (on
page 131).
RulesLists the active rules, if any, that the data point violated. For
simplicity, Unity Desktop divides rules into groups and only displays the
most serious violation in each group. See Rule Precedence When
Displaying Rule Violations (on page 103) for more information.
OPFor manual data entry, these are the initials of the logged in user
when data entry occurs. For imported data, Unity Desktop uses the
initials defined in the import file, if present. Otherwise Unity Desktop
checks Operator Setup (on page 149), and if no initials are defined
there, it places two asterisks in the column.
Bio-Rad Laboratories
117
Additional test informationClick the plus sign (+) to display the additional
information. Click the minus sign () to hide it. See Additional Test
Information on the Point Data Grid for more details.
ValueThe cell where you type your QC value when entering data
manually. Imported values also appear in this cell.
Y/NIndicates whether the data point was accepted (Y) or rejected (N).
Unity Desktop automatically rejects any data row that contains a point
that violates a rejection rule. If you have permission to edit data, you
can manually change the accepted/rejected status of a data point as
described in Changing a Data Point's Accepted/Rejected Status (on
page 131).
RulesLists the active rules, if any, that the data point violated. For
simplicity, Unity Desktop divides rules into groups and only displays the
most serious violation in each group. See Rule Precedence When
Displaying Rule Violations (on page 103) for more information.
zThe data point's z score, which indicates the number of SDs the point
data is from the evaluation mean.
OPFor manual data entry, these are the initials of the logged in user
when data entry occurs. For imported data, the software uses the
initials defined in the import file, if present. Otherwise the software
checks Operator Setup (on page 149), and if no initials are defined
there, it places two asterisks in the column.
For more information on actions and comments, see Annotating Data with
Actions and Comments.
118
Bio-Rad Laboratories
Entering Data
Floating Mean/SD
Fixed Mean/SD
These columns will contain the mean and SD used to evaluate the data
point, which is potentially confusing because you can specify a fixed
mean for a control level and not an SD and vice versa. Evaluation
Means (see "Evaluation Means and SDs" on page 120) explains which
values appear more completely.
Active rulesThe rules against which the data point was evaluated.
These rules are not necessarily the rules that appear in the header,
which are the rules active for the current date/time. This apparent
discrepancy occurs because you can change the rules applied to a test
at any time.
You can expand any number of data rows to compare their information; you
do not need to collapse a row to expand another.
Bio-Rad Laboratories
119
Use of fixed statistics depends on the level of control material. This means
that each level within a control material can have different combinations of
fixed and floating statistics. For simplicity, we will provide examples only
for situations where the fixed statistic choices are the same for each level
within the control material.
Important Facts
The evaluation mean and SD will both display for each control level.
The column the mean and SD appear in is determined by whether the
statistic is fixed or floating.
When both statistics are fixed, rule evaluation begins with the first data
point.
Floating Mean/Floating SD
The preceding figure illustrates extended information for a data point with
no fixed mean or fixed SD specified. In this case:
120
Bio-Rad Laboratories
Entering Data
Until these data points are collected, the Floating mean/SD column will
be blank. (However, the Active Rules column is always populated.)
The Fixed Mean/SD column is blank because you have not specified
either of these values.
Floating Mean/Fixed SD
The preceding figure illustrates extended information for a data point with
a fixed mean but no fixed SD specified. In this case:
z score calculation begins with the third data point because Unity
Desktop uses the first two data points to calculate a mean.
Fixed Mean/Fixed SD
The preceding figure illustrates extended information for a data point with
both a fixed mean and fixed SD specified. In this case:
The Floating Mean/SD column is blank because you are not using
floating values.
Bio-Rad Laboratories
121
Fixed Mean/Floating SD
The preceding figure illustrates extended information for a data point with
a fixed mean and a floating SD. In this case:
z-score calculation begins with the third data point. (Unity Desktop uses
the first two data points to calculate an SD.)
If necessary, click the "Point Data" tab to display it. The first data entry
cell on the last line of data will be selected.
See Specifying a Default Data Entry Mode (on page 112) for
instructions to make point data entry your default data entry mode.
Type a value.
Press Tab or Enter. The active selection moves to the next data entry
cell.
Type a value for this level of control or press Tab or Enter again to
leave the level blank.
When you press Tab or Enter in the last level of the row, Unity Desktop:
7
122
Evaluates the data points against active rules and indicates their
accepted/rejected status.
Continue entering data until all data points you want to enter for the
test appear.
Bio-Rad Laboratories
Entering Data
Click Save to close the data entry screen and save your data.
Click a navigation shortcut key to open data entry for another test
as described in Navigating within Data Entry (on page 110).
Each of these actions saves data for the test and closes its data entry
screen.
Only Bar Charts (on page 161) are available for summary data.
If you view an L-J, Youden, or Yundt chart for a test that has only
summary data, the chart will display but contain no data points. This
includes the L-J chart available within data entry.
Data Groups (on page 98) are not applicable to summary data. If you
select the "Group" check box on the "Summary Data" tab,
<Program_Name> switches to the "Point Data" tab.
Set Date (on page 113) can be used; however, it will increment the
date by one. The set date does not increment by month for summary
data entry.
Bio-Rad Laboratories
123
Click the "Summary Data" tab if not already displayed. You can make
summary data your default data entry mode as described in Specifying
a Default Data Entry Mode (on page 112).
Move the active cell selection to the first blank data entry line.
Click in the Mean column. (A solid border will surround the active cell.)
Press Tab or Enter to move the cell selection to the Pts column.
10 Press Tab or Enter to move the cell selection to the next Mean column.
11 Follow the previous steps to enter the Mean, SD, and number of points
for the levels you are using. If you are entering data for a single month,
press Save. If you are entering data for several months, press Tab or
Enter to start a new data row.
12 Press Tab or Enter. The cell selection moves to the first Mean column on
the next data entry line.
13 When you complete data entry for the test, you have several options:
Each of these actions saves data for the test and closes its data entry
screen.
124
Bio-Rad Laboratories
Entering Data
Autoimmune
Urinalysis
Bio-Rad Laboratories
125
So that the software recognizes your test as qualitative or semiquantitative, it is important to define your test properly and define your
unit selection as Qualitative. Continue this process for each of the tests
you are using.
When a test has qualitative as its unit, the software automatically opens the
qualitative data screen when you double-click the test from either the lab
or panel navigation tree.
Qualitative data is not evaluated by SPC rules.
Rejection Logs
All data points that Unity Desktop cannot import are sent to a rejection log,
which is available by choosing Tools | Utilities | Import | View Rejection Log.
The rejection log will have the following information:
The rejection log lists each error on its own line within the file. Therefore,
a single data row may have multiple entries in the rejection logone for
each error occurring in the data row. This allows you to correct errors
individually.
For example, you might filter the log by the "Lot closed" message; open the
closed lot(s); and then filter the log by the "Date out of range" message and
correct any problems. When you've made all required corrections, you can
click Save Import File and Unity Desktop will create an import file to be
reimported.
You will probably need to use the rejection log only for manually imported
data or when you specify your own interface. If you are using SoftConnect,
you will have resolved these errors prior to importing the file.
126
Bio-Rad Laboratories
Entering Data
If you are manually entering several days' data, you might want to use the
Set Date (on page 113) feature. The selected Set Date will apply to all
qualitative tests that you enter data for. Note that the date/time of each
row of data must be in chronological order.
Bio-Rad Laboratories
127
Click in the cell where you want to enter data. Then click the downarrow to display the list of available result choices.
128
Bio-Rad Laboratories
H A P T E R
11
Edit a value.
Insert a value.
Delete data.
Delete a range of data from the Lab, Lot, or Test dialog boxes.
Users with permission to "Edit all data" can change any value.
Users with permission to "Edit last line of data" can change only the last
line of values entered.
Users with permission to "Enter new data only" or "View data only"
cannot make any changes to existing values.
When you make changes to data, Unity Desktop transmits these changes to
the Unity Interlaboratory Program in the next transmission file you send.
Double-click the test in the navigation tree to open its data screen.
You have two basic options for changing the date-time of a data row:
To change the date and/or time, click the portion of the date-time
you want to change and type over it, or use the small up and down
arrows to change it.
To change only the date, click in the "Date and Time" column, and
click the large down arrow to display a calendar from which you can
select the date as described in Using Calendars (on page 233).
To Edit a Value
1
Double-click the test in the navigation tree to open its data screen.
130
Bio-Rad Laboratories
Double-click the test in the navigation tree to open its data screen.
Click the down arrow to display a list containing two choicesY for
accepted and N for rejected. (The current selection will be
highlighted.)
Bio-Rad Laboratories
131
You can only insert a single data row at a time. If you want to add
multiple data entry rows, you must insert them one row at a time.
Unity Desktop inserts the blank row above the row that you select and
assigns it the same date and time. You can edit the date-time, but data
rows must be in sequential date-time order. If you attempt to assign the
row an out-of-sequence date or time, the software displays a message
and does not save your change.
Within data entry, move the mouse pointer to the left-most column of
the data entry screen in the row where you want to insert the data
point. The mouse pointer changes to a horizontal arrow.
Press the Insert key on the keyboard. The program inserts a blank row
above the row selected in Step 2 and assigns a row number to maintain
sequential row numbers. The inserted row will have the same date-time
as the row you selected in step 2.
Press Tab until the active cell selection moves to the next data entry
row.
Unity Desktop does not evaluate inserted data against QC rules.
Deleting Data
Unity Desktop allows you to delete data from the data screen or from the
"Lab", "Lot", and "Test" dialog boxes.
Bio-Rad Laboratories
Delete Range of Data from the "Lab", "Lot", and "Test" dialog
boxes
We strongly suggest that you create a database backup or archive before
using this function at the lot, lab, or all data levels. Alternatively, you
could export the data to another application or as an import file. See
Database Archives and Backups.
If you need to delete a larger amount of data, you can click Delete Range
of Data from the "Lab", "Lot" or "Test" dialog boxes. The "Delete Range of
Data" dialog box appears and allows you to select a date range of data to be
deleted and an option for the scope of the data to be deleted. Which
options are available depends on the dialog box from which you clicked
Delete Range of Data.
"Lab" dialog boxThe all data and current lab options will be available.
"Lot" dialog boxThe all data, current lab, and current lot options will
be available.
"Test" dialog boxThe all data, current lab, current lot, and current
test options will be available.
Move the mouse pointer to the row number column of a row you want
to delete. The row number column is the left-most column of the data
entry screen. The mouse pointer changes to a horizontal arrow.
Hold down the SHIFT key, and select additional contiguous rows OR
Hold down the CTRL key, and select additional contiguous or noncontiguous rows
See Selecting Items in Lists (on page 231) for more information.
5
When the ,row(s) you want to delete are selected, press the Delete key
on the keyboard. The screen displays a message asking you to confirm
that you want to delete the selected data.
133
From the "Lab", "Lot", or "Test" dialog boxes, click Delete Range of
Data. The "Delete Range of Data" dialog box appears.
All data
Current lab
Current lot
Current test
All data is never selected by default and you should use this feature
with extreme caution and only after creating a database archive or
backup.
4
Specify a date range for the data to be deleted. The current date is the
default.
134
Bio-Rad Laboratories
H A P T E R
12
Supervisor Reports
Actions
Action logs are intended to standardize documentation of events that occur
repeatedly. Unity Desktop comes with a library of commonly used actions.
You can customize the library for use in your laboratory. However, limiting
the users who have permission to edit actions should also limit the
proliferation of "nonstandard" action logs, which can be important when
reviewing Action Levey-Jennings charts.
Action logs provide a way for you to document actions taken when a
particular data point fails a performance challenge. As such, they are
available only for point data entry.
Unity Desktop provides a library of predefined messages that have been
carefully compiled to provide standardized documentation of most events.
These predefined messages appear in the Action Log dialog box, which is
easily viewed by clicking in a cell within the Action column of the
standard data entry screen.
In the Action Log dialog box, a red check mark next to an action indicates
that it is in use and cannot be deleted or modified. Unity Desktop uses the
action's number code when generating Action Levey-Jennings charts.
Click in the A column of the row where you want to add the action. A
message box appears with any previously entered actions.
Select an action from the list and click Apply. The message box
reappears with your newly added action.
Click Close to exit the message box or Add to append another action.
You cannot delete an action once it is attached to a data row.
136
Bio-Rad Laboratories
Choose Tools | Actions and Comments | Configure Action Logs. The Action
Log dialog box appears with buttons to allow you to:
Add: Type the new log in the Action field and click Add. The new
message displays in the lower section of the dialog box.
Update: Select the log you want to change; edit the message in the
Action field; and click Update.
Delete: Select the log you want to delete and click Delete. The
message is removed from the list of actions.
Bio-Rad Laboratories
137
The Action Log dialog box appears if the Automatic action logs
check box is selected.
If the "Require action logs" dialog box is selected, users cannot close
the "Action Log" dialog box without appending an action to the data
row.
The "Automatic action logs" and "Require action logs" check boxes appear on
the The Setup Dialog Box (on page 222).
Type a free-text message to attach to the data point. This message will
not be added to the list of predefined messages unless you click Add
before clicking OK to close the dialog box.
Click Cancel to exit the Action Log dialog box without appending a
message as long as the "Require action logs" check box is not selected.
You can also click in the Action column for any row to display the Action
Log dialog box. Within the dialog box, all users can select a predefined
message from the drop-down list.
Only users with permission to edit action logs can create new messages to
be included in the "Action Log" list.
138
Bio-Rad Laboratories
Comments
Comment logs are free-text messages that you add to a data row on a data
entry screen. When you click in the C column, Unity Desktop displays the
"Comment" dialog box, which will display any existing comments and allow
you to add additional comments.
Click in the C column of the row where you want to append the
comment. The "Comment" dialog box appears with the cursor in the
new comment area.
An upper section with two check boxesAction and Comment. You can
select one or both check boxes.
Bio-Rad Laboratories
139
If there is no data within your selected date-range and scope, the software
inserts a blank data row and appends your actions/comments to it.
Select an action from the list and click Apply. At this point, you can
click Add again to attach another action.
All other functions of the "Action Log" dialog box are available as
described in To Edit the Action Library (on page 137).
Choose a start and end date for the action(s). These dates will probably
be a single day. By default, the action is applied to the current date
only.
Choose the scope of the action by selecting an item from the "Scope"
tree.
7
140
Select a lab number under the instrument to apply the action to all
tests performed in the selected instrument/lab number
combination.
Choose a start and end date for the comment. By default, these dates
will be set to the current date. You could apply a comment like
"Instrument out of service" to multiple dates.
Choose the scope of the action by selecting an item from the Scope
tree.
Choose a start and end date for the action and comment. By default,
these dates will be set to the current date.
Choose the scope of the action and comment by selecting an item from
the Scope tree.
141
142
Click OK to append the action and comment and close the dialog box.
Bio-Rad Laboratories
H A P T E R
13
Importing Data
The Unity Desktop import utility provides both autoimport and manual
import. As a separate utility, import can run in the background and not
interfere with normal operations of the software.
Please be aware that imported data takes precedence over manually
entered dataif you enter data manually concurrently with an import
process, Unity Desktop will save the imported data rather than your
manually entered data.
Within Unity Desktop, choose Tools | Utilities | Import | Import Data or press F12
to display the "Import" dialog box. The dialog box has two panesthe left
pane allows you to customize how your data is imported and the right pane
displays the status of imported files.
In addition, the "Import" dialog box has the following command buttons:
StopStops the import process. You may need to stop the import
process in order to create an import file from the rejection log and
correct it. If you do not stop the import utility before clicking Save in
the rejection log, Unity Desktop will create the import file and
immediately re-import it.
Reset StatusClears the right pane of the "Import" dialog box. This
pane lists the folders, file names, and status of imported files.
Customizing Import
The left pane of the "Import" dialog box allows you to customize how Unity
Desktop imports your data. When you access this dialog box, you will notice
that each item has a default setting. Typically, you will not need to change
these settings. If you are being assisted by Bio-Rad Connectivity personnel,
they will let you know if any changes are necessary.
The dialog box contains the following, listed in the order they appear:
Bio-Rad Laboratories
Run Length
Unity Desktop assigns data points to runs based on their date-time stamp.
Because the software cannot predict how your laboratory will perform QC
testing, it allows you to specify the time during which you want your
imported data points assigned to the same run. (Runs are used for within
and between evaluations of statistical process control rules.) Unity Desktop
defaults the run length to 15 minutes.
The default run length is 15 minutes, but you can define a run length from 1
to 1440 minutes. (1440 minutes is 24 hours.)
Unity Desktop assigns data points that are within the run length-interval to
the same data entry row (run). Timing for the run length interval begins
with the date-time stamp of the first data point assigned to the run,
regardless of its level.
Unity Desktop starts a new run when:
The date-time stamp for the incoming point exceeds the run-length
interval, measured from the date-time stamp of the first data point in
the active run.
The active run contains values for all available levels as defined on the
"Settings" tab of the "Test Settings" dialog box.
Data Retention
This setting determines how long Unity Desktop retains data on the
rejection log before automatically deleting it. The default is 60 days, but
you can choose Tools | Utilities | Import | Delete Rejection Log at any time to
delete the rejection log.
File Retention
This setting determines how long Unity Desktop retains the .SAV files in the
"Import data archive" folder. The default is 30 days, but you can specify
from zero to 90 days.
Bio-Rad Laboratories
145
QC Output File
Select this check box if you want Unity Desktop to create a file
containing the information in the original file plus each data points
accepted/rejected status and QC rules violated, if any. Unity Desktop
places the file in the folder you specified as the "QC output file" folder.
146
Bio-Rad Laboratories
The default Import Data Source path points to a single folder named
"InstrumentImport" that is created by the program. Although the default
path will be C:\Program Files\Bio-Rad Laboratories\Unity
Desktop\InstrumentImport\. You can edit the path to point to a
different folder or to several folders as described in the example below.
You can create multiple folders and specify each folder in the "Import data
source path". Multiple folders are useful when you have several instruments
and want to separate the data files. If you create multiple folders for data
files, you must specify the complete path to each folder in the "Import Data
Source" path and separate individual paths with a comma (,).
For example:
You installed Unity Desktop to the default folder. (C:\Program Files\Bio-Rad
Laboratories\Unity Desktop.)
Within the Unity Desktop folder on drive C, you created three data folders
(named Inst1, Inst2, and Inst3) into which three instruments will place data.
Bio-Rad Laboratories
147
Field separator
Choose a Field separator to specify how your file is formatted (i.e. Tab,
Semicolon, Comma, Space, Vertical bar, or Other).
String delimiter
Choose the String delimiter (Double quote, Single quote, Space, or
Other), if any, used in your file. String delimiters may be required when
the character used as the field separator is also used within the string.
Decimal point
Choose the character (Comma, Dot, or Other) used to represent a
decimal point.
Polling separator
Choose vertical bar or comma.
Source folderThis column contains the complete path to the folder that
contains the import file.
StatusFor each import file, this column contains a message that the
import process has started and a message that the import process has
completed.
Autoimport
The Import Utility allows you to import data files from your instrument or
LIS that have been translated to a standard format that Unity Desktop can
import. Bio-Rad has a CIS (Connectivity Implementation Services) group
whose sole purpose is to assist you with interfacing.
Contact your Bio-Rad software support representative if you are interested
in using Bio-Rads Connectivity Services. Alternatively, you or your IT group
can develop an interface, which requires that your data be formatted as a
Bio-Rad standard import file as described in Manual Interfacing and
Standard Import Files.
148
Bio-Rad Laboratories
Rejection Logs
All data points that Unity Desktop cannot import are sent to a rejection log,
which is available by choosing Tools | Utilities | Import | View Rejection Log.
The rejection log will have the following information:
The rejection log lists each error on its own line within the file. Therefore,
a single data row may have multiple entries in the rejection logone for
each error occurring in the data row. This allows you to correct errors
individually.
For example, you might filter the log by the "Lot closed" message; open the
closed lot(s); and then filter the log by the "Date out of range" message and
correct any problems. When you've made all required corrections, you can
click Save Import File and Unity Desktop will create an import file to be
reimported.
You will probably need to use the rejection log only for manually imported
data or when you specify your own interface. If you are using SoftConnect,
you will have resolved these errors prior to importing the file.
Operator Setup
Unity Desktop allows you to specify operator initials for imported data.
When importing data, Unity Desktop uses the following cascade to
determine the initials to appear in the OP column of the data entry screen:
If the import file contains initials, Unity Desktop places them in the OP
column.
If the import file does not contain initials, Unity Desktop uses the
initials defined in Operator Setup, if any.
If the import file does not contain initials and the operator is not
defined in Operator Setup, Unity Desktop places two asterisks in the OP
column, indicating that the operator is unknown.
Bio-Rad Laboratories
149
Choose Tools | Utilities | Operator Setup to access the "Operator Setup" dialog
box. From the dialog box you can:
Choose how to assign operator initials. Your options are all data, lab
and lot, or instrument.
Specify the number of work shifts (from one to eight) in use in your lab.
Set the number of shifts before making other selections. The numbers in
the "Shift" column list will be limited to this number.
150
Bio-Rad Laboratories
You cannot edit the start time of shift 1; shift 1 must begin at 12:00
AM.
You cannot edit the end time of any shift. Unity Desktop assigns the
end time based on the start time you specify for the following shift.
The last shift of the day must end at 11:59 PM because the first shift
must begin at 12:00 AM.
All data
When you choose all data, the selection grid will have columns for
you to choose a shift number and operator initials. The operator
initials will be applied to all tests imported during the assigned
work shift.
Instruments
When you choose instruments, the selection grid will have columns
for you to choose an instrument and the operator initials for each
shift. (The list of instruments will contain all instruments in the
code list, not just instruments for which you have already created
tests.)
Bio-Rad Laboratories
151
Exporting Data
One of the most useful features of Unity Desktop is its ability to export
data files in formats compatible with popular business software applications
such as Microsoft Access, Microsoft Excel, and Microsoft Word.
An individual lab number: Select an open lab number in the Lab dialog
box and click Export. Complete the requested information in the
Export Current Lab dialog box and click OK.
An individual lot number: Select an open lot number in the Lot dialog
box and click Export. Complete the requested information in the
Export Current Lot dialog box and click OK.
A single test: Select an open test in the Test dialog box and click
Export. Complete the requested information in the Export Current
Test dialog box and click OK.
152
Bio-Rad Laboratories
Regardless of the area from which you requested the import, Unity Desktop
displays a dialog box requesting the information shown in the preceding
figure including:
File
Choose one of the following options for the export file format:
Text fileChoose this option to create a file for use with word
processing programs. Unity Desktop uses American Standard Code
for Information Interchange (ASCII) characters for exporting text
files.
File name
If you type a file name in this field, Unity Desktop will save the file to
the last location to which you saved an export file. You might want to
create a folder to store these types of files. Alternatively, you can click
the ellipsis (...) and browse to a folder.
Date range
Make selections for from and to dates to export the desired data.
Data
Choose an option (point or summary) that the data should be exported
in. Your choice here applies only to data entered as point data.
Obviously, data that was entered as summarized data cannot be
exported as point data.
Bio-Rad Laboratories
153
H A P T E R
14
Level check boxesEach level you are using has a check box. Selecting
and clearing the check boxes displays or hides the chart for the level.
Overlay (for L-J charts only)When the check box is selected, all levels
of control appear on a single chart. When the check box is cleared,
each level has its own chart.
Printing Charts
Display the chart and then select File | Print Chart. Select the range of charts
you want to print from the drop-down list. Your choices in lab mode are:
All dataPrints the chart for all tests in the database for the from-to
date range.
Current labPrints the chart for the selected lab for the from-to date
range.
Current lotPrints the chart for the selected lot for the from-to date
range.
Current testPrints the chart for the selected test for the from-to date
range.
Your choices in panel mode are current test and current panel.
Saving Charts
1
With the chart displayed, choose File | Save Chart. The "Save As" dialog
box appears allowing you to choose a location, file name, and file
format (JPG, GIF, BMP, or PDF) in which to save the file.
Chart Options
All Unity Desktop chart options include "Header" and "General" tabs. The
selections you make from these tabs will apply to all charts. In addition,
each chart type includes a tab specific to it. Finally, Levey-Jennings Charts
include an additional tab (Lines). The following lists summarize the tabs
available for each chart type:
Levey-Jennings Charts
General (see "General Chart Options" on page 157)
Levey-Jennings Tab (on page 158)
Header (see "Chart Headers" on page 157)
Lines (see "Lines Tab for L-J Charts" on page 160)
Bar Charts
General (see "General Chart Options" on page 157)
Bar Chart Options (see "Selecting the Data to Appear on Reports" on
page 166)
Header (see "Chart Headers" on page 157)
Youden Charts
General (see "General Chart Options" on page 157)
156
Bio-Rad Laboratories
Fill background
Legend
Click the Default button to return to the original Unity Desktop settings.
Fill background
If the "Show background" check box is selected, you can select fill colors for
the 1, 2, and 3 SD ranges from their drop-down lists. If you are printing
charts and want to conserve toner, you can clear (uncheck) the check box
from either the "Print Chart" dialog box or from chart options.
Grid lines
Your selections here determine the color and dash style of the grid lines
marking 1, 2, and 3 SD above and below the mean.
Legend
Applies only to L-J charts. Select the "Show legend" check box if you want
the chart to contain a legend with the symbol and color for each level of
control. You can also select an option to display the legend at the top or
bottom of the chart. By default, the legend appears above the chart.
Chart Headers
Unity Desktop chart headers contain quite a lot of information, some of
which you may not need to include. Unity Desktop allows you to configure
your chart headers to contain just the information you want.
Bio-Rad Laboratories
157
From within a chart, click Options. Then click the "Header" tab. The tab
contains check boxes that you can select (check) or clear (uncheck). When
an item is checked, it appears in your chart headers. When an item is
unchecked, it is omitted. Unity Desktop applies your selections to all charts.
Displaying and hiding items on the chart header is for readability only.
Hiding an item does not resize the headerthe space where the item would
have appeared remains.
Levey-Jennings Charts
Levey-Jennings (L-J) charts plot your lab's data points for a specified time
period against your lab's fixed or cumulative (floating) mean and 3s range.
Unity Desktop uses your fixed mean and standard deviation, if defined.
Otherwise it uses the cumulative statistics.
As discussed in QC Assumptions, using statistics to evaluate QC values
assumes that new control measurements will have a similar distribution to
past measurements while the system is stable and that the distributions of
values is Gaussian (normal).
Using these assumptions, 95.5% of values should be within 2s of the mean
and 99.7% of values should be within 3s of the mean. A value outside 3s of
the mean would be expected only 0.3% of the time if the system is stable.
L-J charts allows you to visually identify these data points as well as shifts
(sudden changes) and trends (gradual changes) in your data.
Lines Tab (see "Lines Tab for L-J Charts" on page 160)
Levey-Jennings Tab
The following options appear on the Levey-Jennings tab:
158
Bio-Rad Laboratories
Days on Charts
By default, days on charts is set to 120, but you can select any number
from 1 to 360.
X-axis Type
You have two options for the x-axis:
Date
When you choose the Date option, the x-axis will have a vertical
column for each day. Each data point collected during that day will
appear in this column, so that the individual data points appear to
be stacked. If you choose this option and select (check) the "Include
point representing mean of data points for the day" check box,
Unity Desktop adds a large point representing the daily mean and
shows the individual data points as smaller points.
Sequence
When you choose the Sequence option, each data point appears in
its own column and the "Include point representing mean of data
points for the day" check box is unavailable.
Symbol
Each control level has a drop-down list that allows you to choose a
symbol to represent that level's data.
Bio-Rad Laboratories
159
An arrow represents points that are off the scale regardless of your
level-symbol choice.
Size
Allows you to specify the size (1-10 pixels) of each level's symbol.
Connect Line
Select this check box if you want Unity Desktop to draw a line
connecting the data points on the chart. When displaying by date, the
line connects the mean of the points for the day and only appears when
the "Include point representing mean of data points for the day" check
box is selected.
Default
Click Default to return to the original Unity Desktop selections.
Bio-Rad Laboratories
Multi-L-J Charts
The Multi-L-J feature of Unity Desktop allows you to select up to 18 tests
and display their Levey-Jennings charts on a single screen. This feature
provides an easy way to compare tests, such as the same analyte run on two
different instruments or lot numbers of control material. Like other Unity
Desktop charts, you can customize Multi-L-J charts as described in Chart
Options and Levey-Jennings Chart Options.
Selections you make for standard L-J charts are applied to Multi-L-J charts
and vice versa.
Multi-L-J charts always appear with the levels overlayed on a single chart.
You can, however, select and clear the level check boxes to show or hide
levels.
Select the tests you want to display from the lab or panel trees. (You
can select up to 18 tests.)
Click
Click
Bar Charts
By default, the Bar Chart displays 13 monthly means for the selected test as
vertical bars. The chart also includes a bar for the fixed mean, when
defined, or the cumulative mean when no fixed mean is defined. The fixed
or cumulative mean also determines the scaling of the Y-axis (mean 3SD).
Bio-Rad Laboratories
161
Because Bar Charts plot monthly means against a 3SD range, the Bar chart
allows you to visualize long-term shifts and trends.
You can change the months on the chart and fill colors for each level from
the "Bar Chart" tab of the "Options" dialog box.
Click the Bar icon on the toolbar. Unity Desktop displays the chart for
the selected test.
Months on chart
See Chart Options for information on making changes that will apply to all
charts.
Youden Charts
The Youden chart is a graphical representation used to plot paired Level 1
and Level 2 (or Level 1 and Level 3, etc.) data for a given time period on an
X and Y-axis.
Data points are plotted on the graph and fall within one of four fields:
Pause the mouse pointer on an individual data point to display its two
associated values, their date and time stamp, any violated SPC rules, and
any attached actions or comments.
162
Bio-Rad Laboratories
Youden charts can be used for any two levels of a control material.
Point
163
SizeDefaults to 6 points, but you can choose any number from one
to ten.
Fill check boxSelect to apply a fill color to the point; then choose
a color from the drop-down list.
See Chart Options for information on making changes that will apply to all
charts.
164
Bio-Rad Laboratories
H A P T E R
15
Pause the mouse over an icon on the Crystal Reports toolbar to display a
short description of the item.
All data
Current lab
Current lot
Current test
Current panel
If you select the test from the Lab tree, the first four options are available,
and Current panel is unavailable. If you select the test from the Panel tree,
only the "Current panel" option is available.
166
Bio-Rad Laboratories
Lab Profile (This will be the active tab when you open the dialog
box.)
Lot Profile
Test Profile
Each tab has check boxes for items that appear in the report header.
3
Click a tab and clear (remove check) or select (add check) individual
items. Note:
167
Unity Desktop does not support the use of electronic signatures. However,
you can use a third-party software, such as Adobe Acrobat to implement
electronic signatures.
168
Supervisor's Report
Levey-Jennings Charts
Listings for Labs, Lots, Tests, and Panels (on page 171)
Labs
Lots
Tests
Panels
Configure, which opens to a dialog box that allows you to select the
items to appear in report headers. See Configuring Intralaboratory
Report Headers (on page 167).
Bio-Rad Laboratories
Select a test in the Lab or Panel trees. (Your test selection determines
the choices available for the data set.)
Choose Reports | General | Point Data Report. A dialog box appears allowing
you to select a date range and data set for the report.
Select from and to dates for the report. (You can either click the downarrow to display a calendar from which you can select a date or select
the day, month, or year, and type over it.)
All data
Current lab
Current lot
Current test
If the selected test is on the Panel tree, the only choice is current
panel.
6
Select from and to dates for the report. (You can either click the downarrow to display a calendar from which you can select a date or select
the day, month, or year, and type over it.)
Bio-Rad Laboratories
169
All data
Current lab
Current lot
Current test
Supervisor's Report
The Supervisor's Report allows you to filter a data set to include only data
points that:
When you press CTRL+F10 to request the report, the "Supervisor's Report"
dialog box appears.
170
Bio-Rad Laboratories
By default, the "Reject" check box is selected and the report includes points
that violated SPC rules set to reject. Select or clear the check boxes to
include points that violated a warning rule or have an attached action or
comment.
Labs - shows all lab numbers, divided into open and closed lab numbers.
Lots - shows all open lab numbers and their associated lot numbers,
whether open or closed.
Tests - shows all lab numbers (open and closed) and their associated
tests organized by lot number. Closed lots and their associated tests
are not displayed.
Panels - shows all panels with their associated tests organized by lab
and lot number.
Bio-Rad Laboratories
171
H A P T E R
16
Transmission Files
Because Unity Desktop allows you to specify automatic transmission of
your data, participation in the Unity Interlaboratory Program is almost
effortless.
Submitting your data allows you to compare your laboratory's data with that
of other labs worldwide, which is a major benefit of Bio-Rad's Unity
Intralaboratory Program. The Unity Intralaboratory Program provides both
instant and monthly intralaboratory reports.
Because Unity Desktop allows you to specify automatic transmission of
your data, participation in the Unity Interlaboratory Program is almost
effortless. Activating automatic transmission is as easy as choosing Tools |
Setup and selecting (checking) the "Automatic monthly transmission" check
box.
Monthly Transmission
Monthly transmission ensures that all your data is transmitted for inclusion
in monthly Unity Interlaboratory reports. (These reports appear as PDF files
on QCNet.com.)
Although we refer to manual transmission from the "Send/Receive Data"
dialog box as monthly transmission, you can send data at any time during
the month. If you send another file later, it will contain only new and
changed data.
Click OK to transmit.
Select the day of the month on which you want automatic transmission
to occur. (The day selections are limited to 1 through 7, because the
Unity Interlaboratory Program considers data received after the 7th to
be late.)
174
Bio-Rad Laboratories
The data files created with this procedure must be received at Bio-Rad via
email or on diskettes by the QC Data due date indicated on the Bio-Rad QC
Calendar in order for the data to be included in your monthly/quarterly
Unity Interlaboratory reports.
Transmission files will:
Include new data entries and any changes in data history since the
previous file or diskette was created.
Contain all data that has been changed since the last transmission and
mark the data so that it will not be transmitted in subsequent
transmission files.
When you select Write Transmission File, the transmission file is written to
a file or diskette and the sequence number is advanced. A second file or
diskette, created immediately after the first, will not contain the same
information, because the QC data has been marked as "transmitted" and
thus is not written the second time.
You must send all transmission files to the Unity Interlaboratory
Program, which will generate an error message if it detects a gap in
transmission sequence numbers.
If you want to write the transmission file to your hard disk drive or to a
network drive, create a folder in which to store the file.
Bio-Rad Laboratories
175
176
Bio-Rad Laboratories
H A P T E R
17
Unity Interlaboratory
Reports
In This Chapter
Overview of Unity Interlaboratory Reports ......................... 177
Standard Reports ....................................................... 178
Optional Interlaboratory Reports .................................... 185
Affiliated Reports....................................................... 189
Urinalysis Reports....................................................... 193
Click Reports.
Click My Reports.
Make selections for lab number, lot number, report type, and period.
The selected report displays in Adobe Acrobat Reader.
Standard Reports
Monthly Evaluation Report
The Monthly Evaluation Report provides a quick overview of lab
performance by lot number of control and serves three purposes:
178
Bio-Rad Laboratories
Notifies your laboratory when your laboratorys data was not received
in time for the standard worldwide comparison.
Bio-Rad Laboratories
179
The center of the graph (SDI and CVR both equal to zero) represents perfect
agreement between your laboratory's values and your consensus group (peer
or method) statistics. Bias and imprecision increase as your values move
further away from the center of the graph.
The reverse side of this report has a comment section where you can record
any observations or changes in your system in response to the results of this
report.
Example Report
Bio-Rad Laboratories
For each test, the report contains the following monthly and cumulative
statistics:
Lab mean
Lab SD
Lab CV
The report also contains the following monthly and cumulative Peer and
Method group statistics:
Mean
SD
CV
Example report
Bio-Rad Laboratories
181
Laboratory Histogram
The Laboratory Histogram contains information for each analyte the lab has
reported on in the last 12 months. The histogram has a bar for each
calendar month plus a cumulative bar. The Laboratory Histogram plots the
labs monthly means against the current cumulative peer group mean 2s
range. For each bar, the report also lists the labs mean, SD, CV, and
number of points. Each level of control has a separate bar chart.
The report provides a visual performance comparison, which is useful for
identifying both shifts (abrupt changes in values) and trends (gradual
changes in values).
If a month has no data points, then either no values were submitted, late
values were submitted, or all submitted values were outside of the
statistical window and rejected by the Unity Interlaboratory Program data
filters.
Example report
182
Bio-Rad Laboratories
Bio-Rad Laboratories
183
They were reported using an incorrect code (i.e. invalid units, invalid
method, etc.)
184
Bio-Rad Laboratories
Example Report
Bio-Rad Laboratories
185
Example Report
Manufacturer Report
The Manufacturer Report is a type of Worldwide Report that is further
separated by instrument manufacturer. This report can be used as a
reference, for example, when evaluating a new instrument or kit.
Example Report
186
Bio-Rad Laboratories
A table displaying your labs 2SD and 3SD range for this quarter and last
quarter.
A table displaying your labs and your consensus group and summary
statistics for this quarter and last quarter.
The Unity Interlaboratory Program excludes labs that report fewer than 6
data points during the quarter from the percentile distribution.
Bio-Rad Laboratories
187
Summary Statistics
The following figure shows the Statistical Profile Report summary statistics
table for your laboratory and Bio-Rad's three consensus groups (peer,
method, and all labs).
Histograms
The frequency histogram section of the Statistical Profile Report contains
two frequency histograms:
188
Bio-Rad Laboratories
Affiliated Reports
Affiliated reports allow a group of labs to compare their resultsessentially
becoming their own consensus group. It is likely that an affiliated group SDI
and CVR will be different (probably smaller) than Bio-Rad peer groups
because your affiliated group has "more in common" than the peer
consensus group. For example, your affiliated group may be using the same
reagent or calibrator lots.
Contact your Bio-Rad software support representative to request these
reports.
The Unity Interlaboratory Program provides the following affiliated reports:
Bio-Rad Laboratories
189
190
Bio-Rad Laboratories
Lab mean
Lab SD
Lab CV
The following additional information appears for the Peer, Method, and
Affiliated groups:
Mean
SD
CV
Bio-Rad Laboratories
191
Example Report
192
Bio-Rad Laboratories
Example Report
Urinalysis Reports
If you submit data for a Bio-Rad urinalysis product to the Unity
Interlaboratory Program, you will receive a qualitative report that consists
of a:
Bio-Rad Laboratories
193
Cover page
Dual level microscopy report for each data entry set submitted
The Urinalysis Report can display only one set of results per day. If you
submit multiple sets, this report include(s) the one with the most recent
date-time stamp.
Chemistry Section
The chemistry section of the Urinalysis Report provides a simulation of the
laboratory responses versus a simulation of the group responses using the
visual color changes of reagent strips. On this report, arrows identify the
majority group response. The following figure shows a portion of the report.
The complete report includes leukocytes, nitrite, urobilinogen, protein, pH,
blood, specific gravity, ketone, bilirubin, glucose, and hCG.
Microscopy Section
Graphics on the report simulate your average response and the group's
average response. Your daily responses appear above the graphics with an
arrow to indicate the majority group response.
194
Bio-Rad Laboratories
The following figure shows a portion of the report. The complete report
includes RBC, WBC, and crystals.
Bio-Rad Laboratories
195
H A P T E R
18
Regulatory Requirements
and Reports
In This Chapter
CLIA Requirements ..................................................... 197
Accreditation Requirements (Paraphrased) ........................ 201
ISO 15189 ................................................................ 210
CLIA Requirements
Have a record of
test system
performance
Monitor the
accuracy of the
analytical process
Monitor the
precision of the
analytical process
Detect
immediate
errors
InstantQC Analyte
Report
InstantQC LJ WebChart
Unity Affiliated
Laboratory Comparison
Reports
Unity Laboratory
Comparison Report
Unity Laboratory
Histogram
Unity Laboratory
Performance Overview
Unity Monthly
Evaluation
z
z
Extracted and paraphrased from Part III Department of Health and Human
Services, Centers for Medicare & Medicaid Services, Centers for Disease
Control and Prevention, 42 CFR Part 493 Medicare, Medicaid, and CLIA
Programs; Laboratory Requirements Relating to Quality Systems and Certain
Personnel Qualifications, Final Rule, Friday, January 24, 2003.
198
Bio-Rad Laboratories
Establish or
verify the
criteria for
acceptability
of all control
materials
InstantQC Analyte
Report
InstantQC LJ
WebChart
Establish
statistical
parameters for
each batch and lot
number of control
materials
Document
all control
procedures
performed
z
z
Unity Affiliated
Data Exception
Report
Unity Affiliated
Laboratory
Comparison Reports
Unity Bias &
Imprecision
Histogram
Unity Data
Rejection Report
Unity Laboratory
Comparison Report
Unity Laboratory
Histogram
Unity Laboratory
Performance
Overview
Unity Monthly
Evaluation
Unity Statistical
Profile
Bio-Rad Laboratories
199
Document all
corrective
actions taken as
a result of QC
InstantQC Analyte
Report
InstantQC LJ WebChart
Unity Affiliated Data
Exception Report
Unity Affiliated
Laboratory Comparison
Reports
200
Bio-Rad Laboratories
Document all
analytic
systems
assessment
activities
InstantQC Analyte
Report
InstantQC LJ WebChart
Unity Affiliated Data
Exception Report
Unity Affiliated
Laboratory Comparison
Reports
Unity Bias &
Imprecision Histogram
Unity Laboratory
Histogram
Unity Laboratory
Performance Overview
Unity Monthly
Evaluation
Bio-Rad Laboratories
201
Document calibration or
recalibration when controls
fail to meet established
criteria
InstantQC Analyte
Report
InstantQC LJ WebChart
Unity Affiliated
Laboratory Comparison
Reports
Unity Laboratory
Comparison Report
Unity Laboratory
Histogram
Unity Laboratory
Performance Overview
Unity Monthly
Evaluation
z
z
202
Bio-Rad Laboratories
InstantQC Analyte
Report
InstantQC LJ WebChart
Demonstrate that
QC and instrument
maintenance are
performed and
evaluated
z
z
Unity Laboratory
Histogram
Unity Laboratory
Performance Overview
Unity Monthly
Evaluation
z
z
z
Bio-Rad Laboratories
203
Document
verification results
of controls for
acceptability before
reporting patient
test results
InstantQC Analyte
Report
If commercially
assayed controls are
used for hematology
instruments, verify
the target values
(mean and QC ranges)
z
InstantQC LJ WebChart
Unity Laboratory
Histogram
Unity Laboratory
Performance Overview
Unity Monthly
Evaluation
Unity Statistical Profile
z
z
204
z
z
Bio-Rad Laboratories
InstantQC LJ WebChart
Monitor
precision
data for
significant
changes
z
z
Unity Affiliated
Laboratory Comparison
Reports
Unity Laboratory
Comparison Report
Unity Laboratory
Histogram
Unity Laboratory
Performance Overview
Unity Monthly
Evaluation
z
z
Bio-Rad Laboratories
205
Document
review and
assessment of
quality control
data at least
monthly
InstantQC Analyte
Report
InstantQC LJ WebChart
Unity Affiliated
Laboratory Comparison
Reports
Unity Laboratory
Comparison Report
Unity Laboratory
Histogram
Unity Laboratory
Performance Overview
Unity Monthly
Evaluation
Verify
manufacturer's
calibrations
with control
materials
appropriate for
the system
206
Bio-Rad Laboratories
Document ongoing
evaluation of (QC)
records,
instrument
maintenance and
function,
temperature, etc.
InstantQC Analyte
Report
Organize and
present QC data so
that it can be
evaluated daily by
the technical staff
to detect problems,
trends, etc.
z
InstantQC LJ WebChart
Unity Monthly
Evaluation
z
z
Bio-Rad Laboratories
207
Document that
control
materials for
pH, pCO2, and
pO2 represent
both high and
low values on
each day of
patient testing
InstantQC Analyte
Report
InstantQC LJ WebChart
208
Bio-Rad Laboratories
InstantQC Analyte
Report
InstantQC LJ WebChart
Unity Affiliated
Laboratory Comparison
Reports
Unity Laboratory
Comparison Report
Unity Laboratory
Histogram
Unity Laboratory
Performance Overview
Unity Monthly
Evaluation
Bio-Rad Laboratories
209
ISO 15189
Design internal quality control
systems to verify the intended
quality of results
InstantQC Analyte
Report
InstantQC LJ WebChart
Unity Affiliated
Laboratory Comparison
Reports
Unity Laboratory
Comparison Report
Unity Laboratory
Histogram
Unity Laboratory
Performance Overview
Unity Monthly
Evaluation
210
Bio-Rad Laboratories
H A P T E R
19
Database Basics
The Unity Desktop installation creates a single Microsoft Access
database named QCDAO.MDB on your local hard drive. If you would like to
use multiple databases, contact your Bio-Rad software support
representative as you may want to upgrade to Unity Real Time.
You can specify to automatically update code list and instrument setup
information from the "Setup" dialog box as described in Database Updates
(on page 224). Alternatively, you can update this information manually.
To view the dates of your last database update, choose Help | About Unity
Desktop. The "About Unity Desktop" dialog box displays a list of database
items and their revision dates.
Select the check boxes for the items you need to download. (You can
select any combination of the check boxes.)
Unity Desktop will check for any available updates and install them.
This search may take several minutes, depending on your download
speed. To decrease download times, you might want to download the
items separately.
7
Click the "Information source folder" ellipsis (...), and navigate to your
CD-ROM drive or to the folder where you copied the information.
Bio-Rad Laboratories
213
You must have "read/write" access to the database location. If you saved
your database backup to a network folder, you must have "read/write"
permission to the network folder. If you saved your database backup to a
compact disc, the file will be "read-only"; you must copy the database
file to your local hard disc drive or to a network folder.
Condensing Data
Condense data converts individual point data entries for each calendar
month into a single summarized entry. Users who enter point data can elect
to condense data to conserve disk space. Condensed data points are
removed from the Point Data Entry tab of the data entry screen and are
added to the Summary Data Entry tab. You can condense data from the
"Lab", "Lot" or "Test" dialog boxes or choose Tools | Utilities | Condense All Labs
to condense all data in the current database.
Once you have condensed a group of data points, you cannot retrieve the
individual data points. Therefore, you might want to export the data to
another application before condensing it. See Exporting Data (on page
152).
The condense procedure DOES NOT change the monthly summarized data
values; therefore, it is not necessary to retransmit condensed data to BioRad.
Determine the scope of the data you want to condense and click the
appropriate icon (lab, lot, or test) on the toolbar.
214
Bio-Rad Laboratories
Choose a start and end date for the range of data you want to
condense.
Click OK. The program displays the message: Individual data points will
be reduced to a single summarized point. This operation is NOT
reversible. Do you want to continue?
Choose Tools | Utilities | Condense All Labs. The "Condense Data" dialog box
displays.
Choose a start and end date for the range of data you want to
condense.
Click OK. The program displays the message: Individual data points will
be reduced to a single summarized point. This operation is NOT
reversible. Do you want to continue?
Bio-Rad Laboratories
215
H A P T E R
20
Installation
1
Insert the Unity Desktop CD into your CD-ROM drive. The Choose
Setup Language dialog box appears.
Wait while setup prepares the InstallShield Wizard and configures the
Microsoft .NET Framework. A welcome message displays.
Click Yes to accept the license agreement. (If you click No, the
installation will stop.)
Select your country from the list of available countries and click Next.
If your country does not appear, please contact your Bio-Rad software
support representative. The Choose Destination Location dialog box
appears.
Activation Keys
Unity Desktop installation automatically provides a 60-day trial period for
all license types. During this trial period, all license types have access to all
program features, except the Westgard Advisor. However, at some time
during the trial period, you must obtain and enter your Activation Key:
218
Bio-Rad Laboratories
If you have not entered your activation key when 15 days remain in
your trial period, the program will begin to warn you that your trial
period is going to expire and that you should contact Bio-Rad
Laboratories for an activation key.
If you do not enter your activation key before the end of your 60-day
trial period, you will not be able to use Unity Desktop until you enter it.
You can update your activation key from the "Update Activation Key" dialog
box (Tools | Security | Update Activation Key).
Choose Tools | Security | Update Activation Key, and make a note of the
serial key.
If you plan to email Bio-Rad, you can copy (Ctrl+C) the serial key and
paste (Ctrl+V) it into your email message.
Have your purchase order number available.
Contact Bio-Rad to obtain your activation key. (If you email us for your
activation key, you can use the shortcut keys Ctrl+C to copy the Serial
Key from the "Update Activation Key" dialog box. Then use Ctrl+V to
paste it into your email.)
Type the activation key, or copy and paste it from your email.
Click OK.
Bio-Rad Laboratories
219
Licensing
You must accept the Unity Desktop license agreement (see "License,
Warranty, and Trademark Information" on page 245) to complete the
installation process. In addition, you agree to the license each time you log
in to Unity Desktop.
To determine your license type and period, choose Tools | Security | Update
Activation Key. The screen displays the "Update Activation Key" dialog box,
which contains your license type and license period.
Each time you start Unity Desktop, your license expiration date appears in
the "Login" dialog box.
Export the data from QC OnCall, Unity Plus/Pro, or Unity-PC, and then
import it into Unity Desktop. This strategy will preserve your point
data.
220
Bio-Rad Laboratories
Create the lab numbers in the import file within Unity Desktop
You must create lab numbers in the import file manually. Unity Desktop
will reject any data from lab numbers that are not defined in its
database. See Adding and Updating Labs (on page 70).
Select the "Create new lots if necessary" check box in the Import Dialog box
From within Unity Desktop, press F12 to display the "Import" dialog
box.
Select (check) the "Create new lots if necessary" check box so that
Unity Desktop will automatically create the unexpired lots in your
import file.
Notice that the "Create new tests if necessary" check box is
selected by default so that Unity Desktop will also create tests for
you. See Import Check Boxes (on page 146) for a more detailed
discussion of available import features.
Bio-Rad Laboratories
221
222
Configure Unity DesktopFrom this tab, you customize how the software
operates. Users with permission to edit the setup options can edit the
following using this tab:
Database updates
Notifications
Reports
Transmission
Bio-Rad Laboratories
All selections you make from the Setup dialog box are globalThey apply
throughout the program.
If you select Month, the summary statistics will be those for the
current calendar month plus your cumulative statistics. Month is
the default selection.
If you select Group, the summary statistics will be those for the
current data group plus your cumulative statistics.
You can also toggle between the Month and Group displays by selecting and
clearing the "Group" check box on the point/summary data screen.
Bio-Rad Laboratories
223
Database Updates
You must have Internet access in order to utilize these features.
The "Database updates" section of the "Setup" dialog box contains check
boxes to control the following:
When you select one of these check boxes, Unity Desktop will automatically
update the selected item when the files are updated in the Unity
Interlaboratory Program. Selecting these check boxes ensures that you
always have the most up-to-date information.
None of these dialog boxes are selected by default. You must select them to
activate this feature.
If you choose not to activate the database update check boxes, you should
periodically update the files you use as described in Updating Database
Information (on page 211). Everyone should regularly update their code list
files; however, if you do not use analytical goals or instrument setup, you
might choose not to update these files.
Notifications
The "Notifications" section of the "Setup" dialog box contains a single check
box that controls whether or not Unity Desktop displays a message each
time you open data entry for a test in a lot that will expire in 30 or fewer
days. Select the "Do not display this message again" check box to stop the
message.
224
Bio-Rad Laboratories
Reports
The "Reports" section of the "Setup" dialog box contains the "Generate PDF
reports" check box. If you select this check box, Unity Desktop will generate
intralaboratory reports as Adobe Acrobat PDF files. These files can be saved
and read using the free Acrobat Reader. If you leave the check box
unselected, Unity Desktop will generate reports that can be printed or
exported to a file using Crystal Report *.rpt format.
Transmission
The "Transmission" section of the "Setup" dialog box has two check boxes:
Laboratory Histogram
Bio-Rad Laboratories
225
Create six JPEG files and name them backgrd1 through backgrd6.
(These are the file names the software looks for when displaying the
main screen graphic.)
Right-click the Windows Start button and select EXPLORER from the popup menu.
Drag the new files you created into the Bitmaps folder.
If the Bitmaps folder contains files with the same names, the software
asks if you want to overwrite the existing files with the new ones. Click
OK to overwrite the files.
226
Bio-Rad Laboratories
Choose the JPEG file you want to use and create a total of 6 copies of
it.
Name the copies backgrd1 through backgrd6. (These are the file
names the software looks for when displaying the main screen graphic.)
Right-click the Windows Start button and select EXPLORER from the popup menu.
Drag the new files you created into the Bitmaps folder.
If the Bitmaps folder contains files with the same names, the software
asks if you want to overwrite the existing files with the new ones. Click
OK to overwrite the files.
Bio-Rad Laboratories
227
H A P T E R
21
Windows Hints
If you are entirely new to Windows-based programs, the hints in this
chapter may be helpful to you. These hints are just to get you started
and not intended to replace documentation provided by Microsoft.
In This Chapter
Tree Views ............................................................... 229
Command Buttons ...................................................... 229
Check Boxes and Option Buttons ..................................... 230
Selecting Items in Lists ................................................ 231
Available/Unavailable Items .......................................... 232
Drag-and-Drop Sorting ................................................. 232
Paging Arrows ........................................................... 233
Using Calendars ......................................................... 233
Creating Shortcuts ...................................................... 234
Setting a Short Date Format .......................................... 234
Removing Unity Desktop ............................................... 235
Tree Views
Items in a tree can be collapsed or expanded. Collapsed items are indicated
by a plus sign (+). Expanded items are indicated by a minus sign (-). Doubleclick a collapsed item to expand it.
Command Buttons
Common Command Buttons
When you click a command button, you are telling the software to perform
an action (e.g. execute a command). The most common command buttons
are:
ApplyClick to save your changes and leave the dialog box open.
230
Bio-Rad Laboratories
Windows Hints
Option buttons allow you to choose only one of the available options. If you
have selected an option and then click another option, the original option is
cleared automatically. Unselected option buttons appear as hollow circles.
Selected option buttons have a solid circle within the hollow one.
To select noncontiguous items, press and hold the CTRL key, and click
individual items in the list. For the following example, the user held down
the CTRL key and then clicked Albumin, Creatinine, and Magnesium. Items
can be clicked in any order.
Bio-Rad Laboratories
231
Available/Unavailable Items
When an item is not available to you, it appears dimmed. Items that may
appear dimmed include menus, toolbar icons, and command buttons. Items
may be dimmed because you do not have permission to perform them or you
need to perform some action before they become available.
In the Reports menu shown in the following figure, RiLiBK Reports is
dimmed and unavailable because the RiLiBK module is not installed.
Drag-and-Drop Sorting
You can sort tests in both the "Lab" and "Panel Tests" dialog boxes using
Windows drag-and-drop sorting.
1
Open the dialog box with the items you want to sort.
Keeping the mouse button pressed, drag the item to the location where
you want it to appear.
Release the mouse button to drop the item in the new location.
232
Bio-Rad Laboratories
Windows Hints
Paging Arrows
When the data screen or a report has multiple pages, you can use the
paging arrows to move quickly through the pages.
| Click to go to the first page.
| Click to go to the last page.
Click to go to the previous page.
Click to go to the next page.
Using Calendars
Several areas within Unity Desktop provide calendars to allow you to
quickly change the date. Within the program, you will use two types of
calendars: one to change the date-time of a data row in data entry; the
other calendar is in several areas (including set date on the data entry
screen, adding an "Other" lot, and selecting report and chart date ranges).
Bio-Rad Laboratories
233
Creating Shortcuts
The following instructions detail how to create a shortcut for Unity Desktop;
however, the instructions can be applied to any other program on the
Windows Start menu.
1
Select a Short date style that uses four digits to specify the year. We
recommend that U.S. customers choose MM/dd/yyyy as the Short date
style.
234
Bio-Rad Laboratories
Windows Hints
Click Apply.
If you installed both components on the same machine (e.g. express install),
the two components will appear one after the other in the "Currently
installed programs" list. In some cases, you may want to remove one or the
other component but not both.
Locate the first component you want to remove, and click on the
component to select it. If you need to remove both the application and
the database utilities, it does not matter which you remove first.
Bio-Rad Laboratories
235
H A P T E R
22
Supplemental Information
In This Chapter
Manual Interfacing and Standard Import Files ..................... 237
Messages ................................................................. 242
License, Warranty, and Trademark Information ................... 245
Place the file in a folder specified in the Import Data Source path
Because Unity Desktop periodically polls folders in the Import Data Source
path and imports any new data files found there, all data files to be
imported must be placed in one of these folders. This procedure is defined
in the Import utility.
238
Bio-Rad Laboratories
Supplemental Information
date-time: Date (and, optionally, time of day) the test was performed,
specified in the form yyyymmdd [hh[mm[ss[.xx]]] |. Characters
enclosed in brackets ([ ]) are optional. This format conforms to the
ASTM E 1238-91 (6.6.19.1) and ASTM E 1394-91 (6.6.2) specifications.
level: Level number of the control being tested (the number 1, 2, or 3).
lab: Lab number assigned by Bio-Rad for use with the test (6 digits).
lot: Lot number for the control (5 digits). The fifth digit must be a zero
(0).
Operator initials: you can place the initials of the operator performing
the test in this field. This field can be blank.
action: you can place a pre-defined action log here. This field can be
blank.
comment: you can place a free-text comment here. This field can be
blank.
value: (Point record only) The value of the test result, specified as a
positive number less than or equal to 99999.0 with up to 3 decimal
places. The characters > (greater than) and < (less than) are not valid.
mean: (Summary record only) The mean of the test results expressed as
a positive number less than or equal to 99999.0 with up to 3 decimal
places.
sd: (Summary record only) the standard deviation of the test results
expressed as a non-negative number less than or equal to 99999.0 with
up to 3 decimal places.
Bio-Rad Laboratories
239
240
Bio-Rad Laboratories
Supplemental Information
Second Instrument
Third Instrument
DiaSTAT1001.txt
DiaSTAT2001.txt
DiaSTAT3001.txt
DiaSTAT1002.txt
DiaSTAT2002.txt
DiaSTAT3002.txt
DiaSTAT1003.txt
DiaSTAT2003.txt
DiaSTAT3003.txt
Naming Convention 2
Another simple file naming convention uses the current date and time in the
file name.
For example:
200201010820.txt
200201010825.txt
200201010855.txt and so forth
Users with permission to edit setup options can change the default paths
from the import utility. Within Unity Desktop, press F12 to access the
Import utility. Outside of Unity Desktop, choose Start | All Programs | Bio-Rad
Laboratories | Import.
Bio-Rad Laboratories
241
Alternatives to Interfacing
If you are unable to connect your LIS directly to Unity Desktop, you may be
able to use another program such as Microsoft Excel to reformat the
generated file to the standard format required by Unity Desktop. Using the
following steps:
1
Messages
Rejection Log Messages
The following messages appear on the rejection log to allow you to identify
and correct import errors:
Date/Time invalid
Invalid Day
Invalid Hour
Invalid Mean
Invalid Minute
242
Bio-Rad Laboratories
Supplemental Information
Invalid Month
Invalid SD
Invalid Second
Invalid Time
Invalid Value
Invalid Year
Lab closed
Lot closed
Lot expired
Result invalid
Action Messages
The following list contains the default action messages and their identifying
number. As you add actions, Unity Desktop numbers them sequentially and
appends them to this list.
1
Calibrator: changed
Bio-Rad Laboratories
243
Inspection: regulatory
10
Instrument: bleached
11
Instrument: calibrate
12
13
14
15
Instrument: service
16
Maintenance: corrective
17
Maintenance: daily
18
Maintenance: monthly
19
Maintenance: semi-annual
20
Maintenance: weekly
21
22
Pipette: calibrate
23
Proficiency testing
24
25
244
Bio-Rad Laboratories
Supplemental Information
26
27
28
Reagent: changed
29
30
Test/assay repeated
31
Test: calibrate
Bio-Rad Laboratories
245
246
Bio-Rad Laboratories
Supplemental Information
Warranty Information
The Purchaser must notify Bio-Rad of its claim of any defect. If the
Software is found to be defective by Bio-Rad, Bio-Rads sole obligation
under this warranty is to remedy such defect in a manner consistent with
Bio-Rads regular business practices. For a defect which materially
adversely affects the performance of the Software, Bio-Rad shall use
commercially reasonable efforts to cure such defect as soon as reasonably
practicable after receipt of Purchasers notice, and for minor defects, BioRad shall use commercially reasonable efforts to correct such minor defects
in the next release of its Software. If, however, Bio-Rad is unable to cure a
major defect, the Purchasers sole remedy shall be the option to cancel this
Agreement, whereupon Bio-Rad shall refund only the Software fees paid,
and Bio-Rad shall have no other liability whatsoever.
The warranties set forth in this Agreement are in lieu of all other
representations and warranties, express or implied, including warranties of
merchantability and fitness for a particular purpose, any warranty of noninfringement, and any other statutory or common-law warranty. Bio-Rad on
its own behalf hereby expressly disclaims and excludes any and all such
other representations and warranties. Liability of Bio-Rad to the Purchaser,
if any, for breach of warranty, or any other claim relating to this Agreement
shall in no event exceed the total amount of Software fees paid by the
Purchaser to Bio-Rad. In no event shall Bio-Rad be liable for incidental or
consequential damages (including but not limited to damages to any
laboratory instruments, computer equipment or data arising from the use of
the Software), loss of business or profits, special or indirect damages of any
nature whatsoever, even if Bio-Rad has knowledge of the potential loss or
damage. Bio-Rad accepts no responsibility for misuse of the Software or its
data modifying capabilities.
Trademark Notices
DirectConnect, InstantQC, QCNet, SoftConnect, UnityWeb, Unity
Interlaboratory Program, Unity Desktop, Unity Real Time and Westgard
Advisor are trademarks or registered trademarks of Bio-Rad Laboratories,
Inc.
All other trademarks and registered trademarks used in this guide are the
property of other companies.
Please consult the individual manufacturers for specific information.
Bio-Rad Laboratories
247
H A P T E R
23
References
In This Chapter
QC References .......................................................... 249
QC References
1
250
Bio-Rad Laboratories
Glossary of Terms
A
Accuracy
The closeness of agreement
between a result and the
accepted reference value. The
term accuracy, when applied to a
set of test results, involves a
combination of random
components and a common
systematic error or bias
component. [ISO 3534-I]
Active cell
The cell in the data-entry grid
where values will appear as you
type. A solid border surrounds
the active cell.
Active rules
SPC rules with a status of reject
or warn.
Affiliated group
A group of labs that the Unity
Interlaboratory Program groups
together to form an ad hoc
consensus group for which it
generates specific Affiliated
Reports (on page 189). You must
contact your Bio-Rad software
support representative to request
inclusion in an affiliated group.
Analyte
The constituent or characteristic
of the sample to be measured.
Note: this includes any element,
ion, compound, substance,
factor, infectious agent, cell,
organelle, activity (enzymatic,
hormonal, or immunological) or
property, the presence or
absence, concentration, activity,
intensity or other characteristics,
of which are to be determined.
Reproduced with permission from
NCCLS publication NRSCL8-P3,
Terminology And Definitions For
Use In NCCLS Documents Third
Edition, Proposed Standard.
Copies of the current edition may
be obtained from: NCCLS, 940
West Valley Road, Suite 1400,
Wayne, Pennsylvania 19087, USA.
Analytical process
A series of steps taken in the
analysis or testing of patient
specimens or samples.
B
Bias
Close
M lab M group
100
LabBias % =
M
group
C
CIS
Connectivity Implementation
Services
CLSI
Clinical and Laboratory Standards
Institute. Formerly NCCLS.
Concentration
A measure of the amount of
dissolved substance per unit of
volume.
Reproduced with permission,
from NCCLS publication NRSCL8P3, Terminology And Definitions
For Use In NCCLS Documents
Third Edition, Proposed Standard.
Copies of the current edition may
be obtained from: NCCLS, 940
West Valley Road, Suite 1400,
Wayne, Pennsylvania 19087, USA.
Connectivity
Implementation Services
CLIA
Abbreviation for the Clinical
Laboratory Improvement
Amendments of 1988, which
regulate laboratory practice in
the U.S.
252
Bio-Rad Laboratories
Glossary of Terms
Consensus group
A group of labs submitting data
to the Unity Interlaboratory
Program. In order of decreasing
specificity, the groups are peer,
method, and all labs.
D
Data conversion
The process that copies your user
profiles, labs, lots, tests, and
test data from earlier Bio-Rad QC
software into Unity Desktop.
Constituent
1) component of a sample; 2)
analyte.
Cumulative mean
Used synonymously with floating
mean. The mean of all accepted
data points entered for a test. If
you do not define a fixed mean,
Unity Desktop uses the
cumulative (floating) mean for
rule evaluation.
CV
A measure of precision expressed
as a percent.
standard deviation
CV % =
100
mean
CVR
Coefficient of variation ratioa
statistic that compares your lab's
precision to that of other labs in
a consensus group.
CVR =
Data group
A user-defined set of data points
that have something in common,
such as a calibrator lot.
Frequently shorted to group. See
Data Groups (on page 98).
CVlab
CVgroup
Bio-Rad Laboratories
253
Data Sets
Groups of data points that have
some elements in common (e.g.
their lab or lot numbers, panel
assignments, entry dates etc.)
The terms Data Set A and Data
Set B (or DS:A and DS:B) are used
to refer to the two data sets
compared on Data Analysis Grids.
The "Graph Against" tab of the "LJ Options" tab also allows you to
specify a data set.
DBA
E
Evaluation Mean/SD
The mean and SD that Unity
Desktop used to evaluate a data
point. When using cumulative
(floating) statistics, these values
can be different for each data
point. You can view the
evaluation statistics by clicking
the plus sign (+) for the data row.
F
Fixed statistics
Within this guide, fixed statistics
refers to specifying (fixing) the
evaluation mean and/or SD for a
test.
Database Administrator
DirectConnect
One of Bio-Rads connectivity
solutions, whereby QC data from
an instrument, LIS, or
middleware solution can be
imported into Unity Real Time,
with a hardware solution,
immediately for run-time
decisions.
DSN
Data Source Name. The Data
Source Name provides
connectivity to a database
through an ODBC driver. The DSN
setup to a database allows the
Unity Desktop application to call
information from the database.
Floating mean
Used synonymously with
cumulative mean. The mean of
all accepted data points entered
for a test. If you do not define a
fixed mean, Unity Desktop uses
the floating (cumulative) mean
for rule evaluation.
G
Global
Within this document, global
means that a selection is applied
to all lab numbers, lot numbers,
and tests.
Group
Data group. A user-defined set of
data points that have something
in common, such as a calibrator
lot. See Data Groups (on page
98).
254
Bio-Rad Laboratories
Glossary of Terms
I
Identical tests
L
Level
Imprecision
Imprecision (%) expresses how far
apart numbers are from each
other. Unity Desktop calculates
lab imprecision using the
following formula:
standard deviation
Lab I% =
100
mean
In control
Used as a adjective, indicates
that the analysis of patient
samples is reliable.
Levels in use
The levels (e.g. level 1, level 2,
level 3) of a control product that
the customer is using in the lab.
Customers define the levels in
use as part of the test setup.
Unused levels of a control
material are omitted from data
screens and Unity Desktop
reports.
Levey-Jennings chart
Inactive rule
An SPC or analytical goal rule
with a status of off. (See Active
rule.)
LIME
InstantQC
An instant interlaboratory
program option that allows you to
compare point data submitted via
Bench/Supervisor Review to
other lab's data on demand.
InstantQC reports and charts are
available on QCNet.com.
ISP
Internet Service Provider. A
company that provides access to
the Internet.
Bio-Rad Laboratories
255
M
Master lot number
Median
An alphanumeric string of 15 or
fewer characters that uniquely
identifies a lot number.
Traditionally, for Bio-Rad
controls, the master lot number
is the 5-digit lot number, ending
in zero, which includes all levels
of the control product (e.g.
40910). To identify levels of a
control product, the final zero is
changed to the level number
(e.g. 40911 is level 1 of master
lot number 40910).
Matrix
The base to which the
constituents of a control material
are added (e.g. serum, urine,
spinal fluid, whole blood, etc.)
Maximum QC
The condition that exists when
the Westgard Advisor is unable to
suggest rules that meet your
quality specification (e.g. TEa).
This condition indicates that the
test process has such low process
capacity (i.e. high total error)
that it cannot be controlled to a
defined level of quality. The
default maximum QC procedure
is a Westgard multirule applying
every possible rule.
Mean
The arithmetic average of a set
of data points.
256
Method
Analytical methodthe means by
which an analyte is measured, for
example hexokinase. The Unity 3digit method code is one of the 6
codes used to define a Test.
Method group
The Unity Interlaboratory
consensus group of labs reporting
an analyte using the same
method code (e.g. all customers
reporting glucose by hexokinase).
N
N
The total number of control
measurements per analytical run.
An N of two could refer to 2
control measurements on a single
control material (e.g. level) or to
one measurement on each of two
control levels.
Bio-Rad Laboratories
Glossary of Terms
O
Open
P
Panel
Operating point
The point on an OPSpecs chart
that represents the intersection
of a test's bias and imprecision.
The operating point represents
the current performance of a
test.
Path
Out of control
Peer group
OPSpecs chart
Bio-Rad Laboratories
257
Precision
Probability of False
Rejection
Abbreviated as Pfr.
A QC performance characteristic
that describes how often a run is
rejected when these are no
errors.
Pupil
The inner circle on Yundt plots
that represent either the Data
Set A or Data Set B CVR,
depending on which is smaller.
Q
QC
Quality Control. In the clinical
laboratory, QC is a system
designed to increase the
probability that each result
report is valid and can be used
with confidence by a health care
provider when making diagnostic
or therapeutic decisions.
QCNet
QCNet.com. A Web portal for
clinical laboratory information
that provides QC management
tools. For example, from
"UnityWeb Interlab" tab of
QCNet, you can view reports and
enter data.
258
R
Random error
Any random deviation from the
laboratory mean. There is
"expected" or "acceptable"
random error which generally lies
anywhere between 3s of the
mean. It follows that any
deviation greater than 3s is
considered "unacceptable"
random error. Because of its
random nature, this type of error
is unpredictable.
Range
The difference between the
largest and the smallest observed
value of a quantitative
characteristic or statistical limit.
Reject rule
(Also known as a rejection rule) A
QC rule with Reject assigned as
the rule status. When a data
point violates a QC rule with a
Reject status, the point is not
accepted and is excluded from
routine QC calculations.
RiLiBK table
A table of analyte/matrix
combinations that labs in
Germany are required to QC
using the RiLiBK QC protocol.
Bio-Rad Laboratories
Glossary of Terms
RiLiBK test
A test being evaluated by RiLiBK
rules. (Contrast with standard
test.)
Run
1
Run length
The customer-specified time
interval in which QC values are
assigned to the same run during
autoimport as long as there is no
value already existing in the run
for the level. Values do not need
to be received in level-number
order to be assigned to the same
run.
S
Series
Used for RiLiBK tests.
A "sequence of determinations of
the same measurable quantity
performed with the same
measuring instrument and the
same calibration under identical
conditions.
In automated analysis where
human intervention does not
occur during the measurement
process, a series cannot exceed
one work shift." (Deutsches
rzteblatt, Vol. 98, No 42, 19
October 2001, pp. A2747 to A
2759.)
Set Date
A Unity Desktop mechanism that
allows customers to specify a
starting date/time for manual
data entry. When using set date,
customers can specify whether to
increment the date (increase it
by one each time the user tabs to
a new row) or maintain the date
until the user changes it.
Shift
A type of systematic error caused
by a sudden and eventually stable
change in control values and
possibly patient values.
SoftConnect
Whereby QC data from an
instrument, LIS, or middleware
solution can be imported into
Unity Real Time.
SPC rules
Statistical Process Control rules.
This terminology encompasses
the original Westgard rules and
additional rules not included in
the original six Westgard rules.
Bio-Rad Laboratories
259
Standard deviation
Abbreviated as SD or s. The SD
quantifies the degree of
dispersion of data points around
the mean and is defined by the
formula:
SD =
(x
x)
n 1
Systematic error
Where:
SD=standard deviation
Suspect Data
Standard Test
A test being evaluated by SPC
rules and possibly an analytical
goal. A test not defined as a
RiLiBK test.
Statistical limits
1
Used to differentiate an
analytical process that is in
control from one that is not
in control.
T
Test
Unity Desktop and the Unity
Interlaboratory Program define a
test using six codes: analyte,
instrument/kit, reagent, method,
unit, and temperature.
Trend
A type of systematic error
causing a gradual, often subtle,
increase or decrease in control
values and possibly patient
values.
Statistical process
A series of steps that results in
production of one or more
statistics.
260
Bio-Rad Laboratories
Glossary of Terms
Trueness
Westgard Rules
U
UnityWeb Interlab
A tab within QCNet.com where
your Unity Interlaboratory reports
are posted and from which you
can generate monthly
interlaboratory and InstantQC
reports and charts. The
"UnityWeb Interlab" tab also
allows point and summary data
entry.
Z
z-score
The number of standard
deviations a control result is from
the expected mean, calculated
as:
SDexpected
W
WA
Abbreviation for Westgard
Advisor.
Westgard Advisor
An optional module within Unity
Real Time that suggests SPC rules
for tests based on a quality
specification you select and your
labs imprecision. Abbreviated as
WA.
Bio-Rad Laboratories
261
Index
1
1-2.5s 33
1-2s 32
1-3.5s 34
1-3s 34
1-4s 35
2
2-2s 35
3
3-1s and 4-1s 37
Abbreviated Summary
189, 190
7
7-T 38
A
A Note About Screen Shots 13
Accreditation Requirements
(Paraphrased) 202
Accuracy 251
Action Messages 159, 243
Actions 59, 136
Actions and Comments 223
Actions and Comments by
Instrument 59, 139
Activation Keys 218
Activation Keys and Licensing
59, 218
Active cell 251
Active rules 251
Add a Lab Number 15, 17
Add a Lot Number 15, 17
Add Tests 15, 18
Add Users 15, 16
Adding and Editing Action
Messages 136
Adding and Updating Labs 70,
221
Adding Bio-Rad Lots 73
B
Bar Charts 123, 161, 168
Basic QC Statistics 24
Before You Begin 217
Bias 25, 28, 252
Bias and Imprecision Histogram
Report 177, 183
Bio-Rad Laboratories From the
Windows Start Menu 57
C
Calculating a Control Mean and
Range 27
D
Data conversion 253
Data due date 253
Data Entry Configuration 223
Data Entry Grid 114, 115
Data Entry Overview 109
Data group 253
Data Groups 98, 123, 253, 254
Data Handling Permissions 68
Data Permissions 66
Data Rejection Report 177, 184
Data Retention 144, 145
Data Screen Header 115
Data Sets 254
Database Basics 211
Database Permissions 66
Database Updates 59, 211, 224
DBA 254
Define Paths for Interface Data
237, 241
Defining Initials With Operator
Setup 151
Defining, Modifying, and Deleting
Users 61
Deleting Data 132
Deleting Labs 72, 90
Deleting Lots 77, 90
Deleting Tests 90
Determine Quality Requirements
for the Test 30, 42
Determine the Type of Error 47
Determining an Acceptable CV
30
Documenting Report Reviews
165, 167
Drag-and-Drop Sorting 92, 98,
232
DSN 254
Duplicating Labs 71
Duplicating Lots 75, 106
Duplicating Tests 88
Bio-Rad Laboratories
Index
E
Editing a Bio-Rad Lot 77
Editing a Non-Bio-Rad Lot 78
Editing Lots 77
Email 176, 219
Enter Data 15, 18
Entering Data 109
Entering Point Data 122
Entering Qualitative Data 110,
115, 127
Entering Summary Data 115,
124
Essential Startup Tasks for New
Users 15
Evaluate Test Performance 43
Evaluation Mean/SD 254
Evaluation Means and SDs 116,
119, 120
Example Point Data Grid 117,
119
Example Report 190, 192
Exporting Data 152, 214
Extended Row Information on the
Point Data Grid 119
F
File Retention 144, 145
First Time Log In 15, 16
Fixed Mean and SD (on Data
Screens) 115, 116
Fixed Mean/Fixed SD 120, 121
Fixed Mean/Floating SD 120,
122
Fixed means and SDs 105
Fixed statistics 254
Floating mean 254
Floating Mean/Fixed SD 120,
121
Floating Mean/Floating SD 120
Functions and Where to Find
Them 59
G
General Chart Options 156,
157, 158
Getting Started 11
H
Histograms 188
How Are QC Results Used? 22
How Do I Use Six Sigma? 45
How to Contact Us 12
I
Identical tests 255
Import Check Boxes 59, 83,
144, 146, 221
Import Data Archive 144
Import Data Source 144, 146
Import File Format Selections
144, 147
Import File Status 148
Importing and Exporting Data
143
Importing Data 143
Imprecision 255
In control 255
Inactive rule 255
Individual Unity Desktop Reports
168
Inserting Data Rows 131
Installation 217
Installation and Setup 217
InstantQC 255
Internet Service Provider 225
Interpreting the SDI 28
Introduction to Quality Control
Statistics 21
ISO 15189 210
ISP 255
K
Keys to a Productive Review of
the Laboratory Quality System
47, 50
L
Laboratory Comparison Report
177, 180
Bio-Rad Laboratories
265
M
Managing Your Data 129
Manual Database Updates 212
Manual Interfacing and Standard
Import Files 237
Manufacturer Report 186
Master lot number 256
Matrix 256
Maximum QC 256
Mean 25, 256
Median 256
Medically relevant error 256
Messages 242
Method 256
Method group 256
Monthly Evaluation Report 177,
178
Multi-L-J Charts 161, 168
N
N 256
Navigating within Data Entry
110, 123, 124
266
O
Open 257
Opening and Closing Labs 72
Operating point 257
Operator Setup 59, 117, 118,
149
Operator Setup Shifts 150
OPSpecs chart 257
Optional Interlaboratory Reports
178, 185
Organization of this Guide 12
Out of control 257
Overview of Actions and
Comments 135
Overview of Data Management
129
Overview of Regulatory
Requirements and Reports
197
Overview of Rule Evaluation
101
Overview of Tests 81
Overview of Unity Interlaboratory
Reports 177
P
Paging Arrows 166, 233
Panel 257
Panel Tests Dialog Box 95, 97
Panels 95
Panels and Data Groups 95
Passwords 62, 63
Path 257
Peer group 257
Percent AQA (SE) 257
Percentile Distribution Table
189
Point Data Entry 117
Point Data Grid 118
Point Data Report 169
Precision 258
Probability of False Rejection
258
Bio-Rad Laboratories
Index
Product Inserts 27
Program Basics 53
Pupil 258
Q
QC 258
QC Output File Folder 144
QC References 249
QCNet 258
Qualitative Data Entry 11, 125
Qualitative Evaluation of a Test's
Bias and Imprecision 45
Quality Control Product(s) 258
R
R-4s 36
Random error 258
Range 258
Recommended Steps 41, 104
Reconciling Your Database 215
References 249
Regulatory Requirements and
Reports 197
Reject rule 258
Rejection Log Messages 126,
149, 242
Rejection Logs 126, 149
Relate the Error to Possible
Causes 48
Relate the Problem to Recent
Changes 49
Removing Unity Desktop 235
Reports 225
Reviewing and Annotating Data
135
RiLiBK table 258
RiLiBK test 259
Rule Precedence When Displaying
Rule Violations 103, 117, 118
Rule Statuses 102
Rules and Settings Permissions
67
Rules Tab of the Test Settings
Dialog Box 87
Run 259
Run length 259
Run Length 144, 145
Bio-Rad Laboratories
267
T
Test 260
Test Rules and Settings 87
Tests 81
The Configure Unity Desktop Tab
223
The Setup Dialog Box 14, 60,
66, 86, 138, 222
To Activate Automatic Monthly
Transmission for Monthly Unity
11, 59, 60, 174, 178, 225
To Add a Bio-Rad Lot 74
To Add a Lab Number 70
To Add a Non-Bio-Rad Lot 74
To Add a Panel 97
To Add a User 62
To Add an Existing Action to a
Data Entry Row 136
To Add Tests Manually 83
To Add Tests Using Instrument
Setup 85
To Add Tests With an Other Code
86
To Attach a Comment by
Instrument 141
To Attach an Action by
Instrument 140
To Attach Both an Action and a
Comment by Instrument 141
To Call a Point Data Report 169
To Change the
Accepted/Rejected Status
131
To Change the Default Current
Data Selection 99
To Change the Main Screen
Graphics 226
To Change VITROS Slide
Generations 93
To Change Your Password 60,
65
To Close a Lab Number 72
To Close a Lot 78
268
To Close Tests 89
To Condense Data by Lab, Lot, or
Test 214
To Condense Data for All Labs
215
To Configure Intralaboratory
Reports 167
To Define or Edit a Group 99
To Delete a Lab Number 72
To Delete a Lot Number 77
To Delete a Panel 97
To Delete a Range of Data 134
To Delete a Row(s) of Data 133
To Delete a User 63
To Delete Tests 91
To Display a Summary Data
Report 169
To Display Multi-L-J Charts 161
To Duplicate a Bio-Rad Lot
Number 76
To Duplicate a Lab Number 71
To Duplicate a Non-Bio-Rad Lot
Number 76
To Edit a Value 130
To Edit the Action Library 137,
140
To Edit the Date-Time 130
To Email a Transmission File to
Bio-Rad 59, 176
To Enable/Disable Automatic
Action Logs 138
To Enter Qualitative Data 127
To Insert a Data Row 132
To Modify a User 63
To Open a Lab Number 73
To Open a Lot 78
To Open Tests 90
To Set a Fixed Mean and/or
Standard Deviation 106
To Set SPC Rules at the Lot Level
105
To Set SPC Rules at the Test
Level 104
To Sort Tests in a Panel 98
To Sort Tests in Lab Mode 91
To Sort Tests in Panel Mode 92
To Transmit Data Manually for
Monthly Unity 174
To Update a Lab Number 70
Bio-Rad Laboratories
Index
To Update a Panel 97
To Update a Test 88
To Update VITROS Slide
Generations 94
To Use Set Date
U
Unity Desktop Charts 155
Unity Desktop Data Screens
109, 114
Unity Desktop Database Basics
211
Unity Desktop Menu Bar 55
Unity Desktop Reports 165
Unity Interlaboratory Reports
177
Unity Interlaboratory Reports Tab
177, 225
UnityWeb Interlab 261
Update VITROS Slide Generations
88, 92, 94
Updating Database Information
60, 84, 211, 224
Updating Tests 88, 90
Updating Your Activation Key
219
Upgrading from QC OnCall, Unity
Plus/Pro, or Unity-PC 220
Upgrading From Unity Post 221
Upgrading to Unity Desktop 220
Urinalysis Reports 193
Useful Features of Data Entry
110
Abbreviated Summary
Report 191
V
Verify and Document 50
Viewing Charts and Reports
within Data Entry 111
Viewing Interlaboratory Reports
on QCNet 178
Viewing Rules/Settings and AGs
within Data Entry 111
Viewing, Printing, and Saving
Charts 155
VITROS Slide Generations 83,
86, 92
W
WA 261
Welcome to Unity Desktop 11
Westgard Advisor 261
Westgard Rules 261
What is Quality Control? 21
Where Do I Begin? 14
Windows Hints 229
Worldwide Report 185
Y
Youden Chart Options 157, 163
Youden Charts 162, 168
Your Lab's Two and Three SD
Ranges 187
Z
z-score 31, 261
Bio-Rad Laboratories
269