SECTION 1: INTRODUCTION
A. Cause of hazards
1. due to energy delivered when functioning properly
2. in case of fault condition
3. due to absence of function
D. Illustration of hazards
1. electrical hazards:- 3 majors effects- i) electrolysis
ii) stimulation of tissue
iii) tissue heating
1.1. low frequency < or = 0.1 Hz A.C.
- electrolysis may occur
1.2. low frequency < or = 1 KHz A.C.:
- excite the muscle and neural cells
- microshock, ventricular fibrillation
- electrocution
1.3. radio frequency burn
2. mechanical hazard e.g.electrical bed
3. radiation hazard
4. explosion hazard
5. temperature or fire hazard
- applied part not intended to supply heat should not have
surface temp. exceeding 41oC
- burning insulation, flaming particles and molten metal should
not fall out of equipment
6. chemical hazard:
e.g. copper in dialysate cause haemolysis, nitrous oxide may have high
risk of spontaneous abortions and congenital abnormality in children as
well as disease of the liver, kidney and CNS, drug may interact with
rubbers, plastics such as catheter, syringes
7. human error/negligence hazard
8. equipment fault hazard
9. system incompatibility hazard
A. Terminology
1. applied parts- the parts which come intentionally into contact with the pt.
2. patient circuit- electrical circuit of which the pt. forms a part
3. basic insulation- insulation applied to live part
4. double insulation- insulation comprising both basic and supplementary
insulation
5. mains part- parts intended to have a conductive connection with the
supply mains
6. high voltage- any voltage over 1000V A.C or 1500V D.C or 1500V peak
value
7. medical safety extra low voltage(MSELV)- a voltage which does not
exceed a nominal value of 24V A.C or 50V D.C
8. F-type isolated(floating) applied part- applied part isolated from all other
parts of the equipment to such a degree that the allowable patient leakage
current under single fault condition is not exceeded when 1.1 times the
highest rated mains voltage is applied between the applied part and earth.
(<<500uA)
D. Other classification
1. drip proof equipment, splashproof equipment, watertight equipment
2. sterilizable equipment
3. anaesthetic proof equipment
E. light indication
red- alarm, amber- alert, green- power on
B. Tissue injuring from high current - burning or destroying tissue in the path -by
heating, charring or electrolysis
A. Electric shock
1. macroshock
- relative large amount of current flows across a person's body
- only a small portion of total applied current pass through the heart
- neither point of contact is directly connected to the heart
- 75 to 400 mA, ventricular fibrillation occurs
2. microshock
- at least one contact on or within the heart
- bypass the protective resistance of the skin and current was
concentrated thro' the heart
- 180 uA can cause fibrillation
- generally accepted safety limit is 10 uA
B. Macroshock hazards
1. electrical fault
2. protection by earthing
C. Microshock hazards
1. leakage current in line-operated equipment
2. leakage current (small current that inevitably flow between any adjacent
insulated conductors) that are at different potentials
3. differences in voltage between earthed conductive surfaces
4. potential difference between 2 earth point
5. prefer to have earthing arrangement in equipotential earthing system
A. General precaution
1 interconnection of equipment - to right connector
- correct interconnecting cable
- what is at the other end of the lead they are connecting
- correct polarity
2. correct handling of equipment after severe shock
- equipment subjected to severe shock should not be used
- thorough check
3. overcharge of battery
- exceeds 100 hours
- shorten battery life, cell explosion
4. muting of alarm
- never try to mute the alarm
- find out the cause and solve
5. test equipment before use