TRADE
Apparatus
Instruments
Glassware
Procedures
WEIGHING
ELECTRONIC BALANCE
VOCABULARY
Pan
Tare
Display
Specifications
Calibration
WEIGHING
Balance Specifications
Maximum Capacity
110 g
Readability
0.1 mg
Repeatability
0.1 mg
Linearity
0.2 mg
Sensitivity
2.5 ppm/C
WEIGHING
The weighing operation consists of:
1. Place a clean receiving vessel on the
balance pan. The mass of the empty
vessel is called the tare.
2. Press the button that resets the tare
reading to zero
3. Add the substance to be weighed to the
vessel and read its new mass
WEIGHING
An Alternate procedure (weighing by
difference) have to be used when dealing
with hygroscopic reagents.
1. Weight a capped bottle containing the
dry reagent
2. Pour some reagent from the weighing
bottle into a receiver
3. Cap the weighing bottle and weight it
again
WEIGHING
WEIGHING
THE DESSICATOR
WEIGHING
Silica-gel:
Hygrometer
indicator
WEIGHING
CALIBRATION
Balances have to be
calibrated against
standard weights in order
to ensure accurate
measurements
Calibration in electronic
balances is now very
easy
WEIGHING
The balance electronics are adjusted in such a way that
the indication of the balance is equal to the mass of the
built-in weight with the built-in weight introduced to the
balance (Rs = ms)
a
Rs kgms 1-
s
k
1
a
g 1-
s
WEIGHING
a
1s
m R
a
1-
MEASURING VOLUMES
BURETTES
A precisely bored glass tube with
graduations that enables us to measure
the volume of liquid delivered.
This is done by reading the level before
and after draining liquid from the burette.
BURETTES
Typical glass burettes have a teflon
stopcock
Sometimes a cap is used to keep dust out
and vapours in
When reading the height of liquid the eyes
have to be at the same level as the top of
liquid to avoid the parallax error
VOLUMETRIC FLASKS
Are used to prepare a solution of a known volume
Are calibrated to contain a particular volume of water at
20C when the bottom of the meniscus is adjusted to the
centre of the mark
After the final dilution, remember to mix your solution
thoroughly, by inverting the flask and shaking
OTHER GLASSWARE
Erlenmeyer flasks and beakers are used
for mixing, transporting, and reacting, but
not for accurate measurements (RSD
about 5%)
OTHER GLASSWARE
Graduated cylinders are useful for measuring liquid
volumes to within about 1%
They are for general purpose use, but not for quantitative
analysis
If greater accuracy is needed, use a pipette or volumetric
flask.
FILTRATION
Precipitates for gravimetric analysis are
collected by filtration, washed and then
dried
If the precipitate does not need to be
ignited, it is collected in a fritted-glass filter
crucible
The filter crucible is used with suction by
using a Kitasato
FILTRATION
FILTRATION
Sometimes the precipitate must be isolated, then igniting
it at high temperature to convert it to a well defined
product of known composition
When the precipitate have to be ignited, it is collected in
ashless filter paper that leaves little residue when
oxidised
The paper should fit snugly and be seated with a little
liquid water
The correct procedure for filtration is shown in the
following figures
FILTRATION
FILTRATION
Particles adhering to the beaker or rod can be dislodged
with a rubber policeman and transferred with a jet of
liquid
The precipitate to be ignited is first collected in the
ashless filter
The filter should be allowed to drain thoroughly
(overnight) with the content protected with a watchglass
Lift the paper out of the funnel, fold it suitably (see figure)
and transfer it to a porcelain crucible
IGNITION
IGNITION
With the filter paper and precipitate inside the crucible,
dry the content cautiously with a small flame by
positioning the crucible above the burner with the help of
a clay triangle as indicated in the figure
After it is dry, char the filter by increasing the flame
temperature