Anda di halaman 1dari 3

G4 PHOENIX DH

Analyzer for automatic determination of diffusible Hydrogen in solids


Hot extraction, carrier gas method

Description and Technical Data

G4 PHOENIX DH
Determination of diffusible Hydrogen
The Analyzer
The G4 PHOENIX DH has been developed for
the determination of the diffusible hydrogen in different sample matrices. The analysis system
comprises a rapid heating infra-red clamshell
furnace and/or a wire-heated tube furnace, both
equipped with a quartz tube for heating of the
sample. The diameter of the tube in the infrared
furnace is 30 mm so that even big samples like
weld seams according to EN/ISO 3690 and AWS
A4.3 can be analyzed without problems.
A simple and reliable calibration is guaranteed by
the integrated automatic gas dosing unit with 10
different volumes. The heart of the analyzer is a
highly sensitive and long term stable thermal
conductivity detector for reading even smallest
hydrogen contents.

The Application
The G4 PHOENIX DH is suitable for the automatic determination of the diffusible Hydrogen content in solid material, like steel, aluminum, welding materials, weld seams etc.

The Software
The analysis software of the G4 PHOENIX DH is
clear and simply structured. All tasks to be performed by the software are clearly divided into
four different screens.

Analysis screen

Performance of the analysis, manual input of


sample weight or by direct transfer from the
balance, display of weight buffer, graphical
display of signal data, display of the results of the
last five analysis

Program settings

Setting of analysis parameters, e.g. furnace temperatures, timing, storage of configuration


for different applications and methods with complete set of data.

Statistics screen

The statistical evaluation of the analyses, calculation of mean value, standard deviation, variance, variance coefficient, display of minimum and maximum value, print-out

Calibration

Different methods for the calibration of the instruments by means of the integrated gas calibration
unit are selectable: single-point, two-point or multipoint calibration.

Page 2 of 3

The Benefits

Temperature programmable infrared furnace, temp. up to 900C, rapid heating and cooling,
tube diameter 30 mm, e.g. for welding samples according to EN/ISO 3690 and AWS A4.3
Wire heated tube furnace, temp. adjustable up to 1200C, tube diameter 18 mm
Long term stable thermal conductivity cell with automatic base line compensation/zero adjustment, Automatic Level Control (ALC) and optimum range selection
Integrated gas calibration unit with 10 volumes
Easy to operate via external PC with Windows-Software, automatic evaluation, storage of
sample information, data storage for follow-up assessment of all analyses
Evaluation selectable in ppm or ml/100 g
Data transfer via FTP or local network connection possible
The G4 PHOENIX analysis as well as reporting structure are compliant to norms AWS
A4.3 and ISO 3690

The Technical Data


Measuring range

0.05 1000 ml/100 g

Analysis time

approx. 3 min 2 h
for welding samples typically 15 20 min
depending on sample material and weight

Resolution

0.01 ml/100 g

Reproducibility

0.01 ml/100 g or 1 % rel.


depending on sample material and weight

Carrier gas

Nitrogen 99,999% purity, 2 bar

Cooling water

approx. 1 l/min (only for IR furnace)

Dimensions & weight

460 x 750 x 500 mm (wxdxh); 50 kg


18.11 x 29.52 x 19.68 inches; 110 lbs.

Electrical supply

Analyzer: 230 V, 1 kVA


Peripherals: 230 V, 300 VA

PC requirements

Operating system: Windows XP/Vista


CD-ROM, 2 serial ports RS 232
*) Windows is a registered trademark of MICROSOFT Corp.

*)

Our service to you:


We would be pleased to examine your samples in our applications laboratory. We invite you to
convince yourself of the high performance of our advanced technology analyzers.
Just contact us and make an appointment for a demonstration in our facilities.

Bruker Elemental GmbH


Kalkar, Germany
Phone +49 (28 24) 9 76 50-600
Fax
+49 (28 24) 9 76 50-629
info@bruker-elemental.de
www.bruker-elemental.com

All configuration and specifications are


subject to change without notice.
2008 Bruker Elemental, HES Rev. 1

Page 3 of 3

Anda mungkin juga menyukai