HBU31 series

30W Open Frame Medical Grade Power Supply

The HBU31 series of AC/DC switching mode power supplies provide


30 Watts of continuous output power . All supplies are UL94V-1
min compliant. All models meet FCC Part-18 class B and CISPR-11
EN55011 class B emission Limits and are designed to comply with
UL/c-UL(UL 60601-1:2ndEdition), TUV/T-mark(EN 606011:2ndEdition) and new CE requirements. All units are 100% burned
in and tested.

FEATURES:
* Wide Operating Voltage, 80 to 275 VAC, 47 to 63 Hz
* Single Output
* Over Voltage and Over Load protection
* Class II system
* 3 year warranty

RoHS2
2011/65/EU

APPLICATIONS:
* Home Healthcare Equipment
* Patient Monitor
* Blood Pressure system
* Portable medical devices
* ECG machine

GENERAL SPECIFICATION:
APPROVALS:

CB

* Short Circuit Protection: Auto Recovery


* Cooling: Free Air Convection
* Flammability Rating: UL94V-1
* Protection Classes: Double insulated, Class II
* Safety: UL/c-UL(UL 60601-1:2ndEdition), TUV/T-mark(EN 60601-1:2ndEdition)

IEC 60601-1

Electrical Characteristics:
Vins

Safety Approval Input Voltage Range

Safety Approval & Specification in Label

100

Typ. Max. Unit


240 VAC

Vin

Input Operate Voltage Range

Detail to see Fig.1 (Derate linearly from 100% load at 90VAC to 80% load at 80VAC)

80

275

Fi

Input Frequency

Sine wave

47

63

Symbol

Condition

Characteristic

Min.

Po

Output Power Range

See Rating Chart

30

Iil

Low Line Input Current

Full Load, Vin=100VAC

0.8
0.4

Iih

High Line Input Current

Full Load, Vin=240VAC

Irl

Low Line Input Inrush Current

Full Load, 25C, Cool start, Vin=100VAC

25

Irh

High Line Input Inrush Current

Full Load, 25C, Cool start, Vin=240VAC

50

Efficiency

Full Load, Vin=230VAC, Detail to see Rating Chart

Line Regulation

Full Load, Vin=100~120VAC or 200~240VAC

Voi

OVP

Over Voltage Protection

OLP

Over Load Protection

Recovers automatically after fault condition is removed

VAC
Hz
W
A
A
A
A

See Rating Chart


1

112

132

110

150

ms

ttr

Time of Transient Response

Full Load, Vin=110VAC

thu

Hold-Up Time

Full Load, Vin=100VAC

12

ts

Start-up time

Full Load, Vin=100~240VAC

0.3

Tc

Temperature Coefficient

All Condition

HV

Dielectric Withstanding Voltage (P-S)

Primary to Secondary, limit current <10mA

EMI

EMC Emission

Compliance to EN55011 (CISPR11), EN61000-3-2,-3

ms
2

0.04 %/C
4000 VAC

Class

Environmental:
Symbol

Condition

Characteristic

Min.

Typ. Max. Unit

70

To

Operating Temperature

Detail to see Fig.2 (Derate linearly from 100% load at 40C to 50% load at 70C)

Ts

Storage Temperature

10 ~ 95% RH

-40

85

Ho

Operating Humidity

non-condensing

95%

RH

Hs

Storage Humidity

See Rating Chart

95%

RH

Vsg

All Condition
Air Discharge, IEC61000-4-2

2
8

kV

ESDa

Surge Voltage
Electro Static Discharge

ESDc

Electro Static Discharge

Contact Discharge, IEC61000-4-2

kV

MTBF

Mean Time Between Failure

Operating Temperature at 25C, Calculated per MIL-HDBK-217F

ELEV

Operating Altitude (Elevation)

All condition

3000

VBR

Vibration

10 ~ 500Hz, 10min./1cycle, 60min. each along X, Y, Z axes

100k
5

kV
h
G

medical
HBU31 series

30W Open Frame Medical Grade Power Supply


110

SPECIFICATION NOTE :
1. Output can provide up to peak load when the power supply starts up.
Continuous staying in more than rated load is not allowed.
2. At factory, in 60% rated load condition, each output is checked to be
within voltage accuracy.
3. Line regulation is defined by changing 10% of input voltage from
nominal line at rated load.
4. Load regulation is defined by changing 40% of measured output load
from 60% rated load.
5. Ripple & noise is measured by using 20MHz bandwidth limited
oscilloscope and terminated each output with a 0.47uF capacitor at
rated load and nominal line.
6. Hold up time is measured from the end of the last charging pulse to the
time which the main output drops down to low limit of main output at
rated load and nominal line.
7. Efficiency is measured at rated load, and nominal line.
8. The specifics for testing the energy efficiency of this Series are outlined
in a separate document titled "Test Method for Calculating the Energy
Efficiency of Single-Voltage Interchangeable AC-DC and AC-AC Power
Supplies (August 11, 2004)," which is available on the ENERGY STAR
Website.

100

LOAD (%)

90

80

70

60

50
100

80

120

140

160
180
200
220
INPUT VOLTAGE (VAC)
(FIG.1) INPUT VOLTAGE DERATING CURVE

240

260

280

110
100
90
80
70
60
50
40
30
20
10
0
0

MECHANICAL DIMENSIONS: ( UNIT: mm )

10

20

30
40
50
60
TEMPERATURE (C)
(FIG.2) TEMPERATURE DERATING CURVE

70

80

PACKING :
1. Net weight: 85g approx.
2. Input connector mates with Molex housing
09-50-3031and Molex 2478 series crimp terminal.
3. Output connector mates with Molex housing
09-50-3041 and Molex 2478 series crimp terminal.

PIN CHART

L
N
4

PIN

MODEL

HBU31-1XX

Vout

Vout Vout Vout

RTN

RTN

Rating Chart:
Typ. Efficiency

No Load
Consumption

Hold-Up Time
(ms)

77.2

0.5

12

OLP

79.8

0.5

12

OLP

90

81.9

0.5

12

OLP

30

100

86.1

0.5

12

OLP

2.31

30

100

86.1

0.5

12

OLP

1.88

30

100

87.0

0.5

12

OLP

1.12

1.43

30

100

87.7

0.5

12

OLP

33.0

0.91

1.12

30

100

87.7

0.5

12

OLP

40.0

0.76

0.91

30

100

87.7

0.5

12

OLP

max

(VDC)

(A)

(A)

(W)

(mVp-p)

HBU31-102

5.0

6.0

3.33

4.00

20

50

HBU31-103

6.0

8.0

2.87

3.83

23

70

HBU31-104

8.0

11.0

2.45

3.37

27

HBU31-105

11.0

13.0

2.31

2.72

HBU31-106

13.0

16.0

1.88

HBU31-107

16.0

21.0

1.43

HBU31-108

21.0

27.0

HBU31-109

27.0

HBU31-110

33.0

Output Current
(Based on the output volt.)

MODEL NO.

Protection Mode

Total Regulation

(W)

min

Ripple & Noise

(%)

max

Maximum
Output Power

(%)

min
(VDC)

Voltage Range

Anda mungkin juga menyukai