CARDIAC/NEURO INSTRUMENTATION LAB

CARDIOLIFE DEFIBRILLATOR LAB 5


BMET 221

LAB 5

CARDIOLIFE DEFIBRILLATOR

NAME:________________

DATE:___________

Equipment Needed: Dynatech Nevada Safety Analyzer


Bio-Tek QED-6 Defibrillator Analyzer
CARDIOLIFE DEFIBRILLATOR
Oscilloscope, DVM, Signal Generator
Resistors: 100K ohm 1/2W, 100 ohm 1/2W
Lab Procedure:
1.
___ Locate the equipment needed.
1a. ___ Read the Specification Data for the Cardiolife, section
11, 107-114 Operators manual.
1b. ___ Read the Functional Descriptions of the Cardiolife,
sec.10, 101-106 Operators manual.
2.
___ In the bench area, connect each device to power and perform an
Electrical Safety Test.
2a. ___ GROUND WIRE RESISTANCE (100 milliohms Limit)
________
2b. ___ CHASSIS LEAKAGE (10 uA Limit)
________
2c. ___ PATIENT LEAD (10 uA Limit)
________
3.
___ Read section 6 of the Cardiolife Defibrillator Service Manual
(Disassembly)
4.
___ Read section 2 of the DEFIBRILLATOR ANALYZER MANUAL.
5.
___ Perform required adjustments as listed in the Service Manual.
5a. ___ ECG/CPU Board 5-1-1 thru 5-1-7
5b. ___ High Voltage Board 5-3-1 thru 5-3-4
5c. ___ CRT Display Adjustment 5-4-2 thru 5-4-4
5d. ___ Motor Control Board 5-5-1 thru 5-5-2
5e. ___ Thermal Head Replacement & Alignment ( 5-6-2 0nly )
6. ___ Re-assemble the Cardiolife and run the SELF TEST, Section 4,
Service Manual.
7. ___ Perform another Electrical Safety Check after re-assembling the
device.
7a. ___ GROUND WIRE RESISTANCE (100 milliohms Limit) __________
7b. ___ CHASSIS LEAKAGE (10 uA Limit)
__________
7c. ___ PATIENT LEAD (10 uA Limit)
__________
8. ___ Contact the Biomedical Instructor and demonstrate the correct
nomenclature identification and proper operation of the Cardiolife
Defibrillator.
Inst Init: ________
9. ___ Demonstrate to the Biomedical Instructor the proper method of
using the Bio-Tek QED-6 Defibrillator Analyzer to check the CARDIOLIFE
for proper operation.
Inst Init: ________
10. ___ Develop a Preventive Maintenance Checklist for the CARDIOLIFE
Defibrillator, that includes a parts needed list, PM Timing intervals
and a itemized check list of the PM.
11. ___ Return all equipment to it's proper place on the shelf.

DIRECTED QUESTIONS ON CONTENT:

--Cardiolife Defibrillator.
--Using the Operator Manual, Guide to Correct Operation Manual, and
Service Manual (open book) answer or respond to the following questions.
a.

What is the maximum charge setting on the cardio-life


defibrillator? __________________

b.

Is the TEC-7300 able to use internal paddles? ______________

c.

How many charge settings are there on the Cardiolife TEC-7300?


______________________

d.

What is the energy setting for the built-in test on the TEC-7300?
_______________________

e.

What is the charge time for the TEC-7300 with AC power supply?
_____________________

f. What type of battery and what is the part number of the battery used
to power the TEC-7300?
_______________________ __________________________
g.

On the CRT CPU chip, pin number 10 controls what function?


_______________________

h.

What function does the symbol RCVFLT stand for?


________________________________

i.

What is the input signal to pin 17 on the UPD-7004C 10-bit A/D


converter?
_________________________________________________________________

j.

What is the Cardiolife part number of the 25 volt 6800 uF capacitor


in circuit number C507 on board UP-7851?
___________________________

k.

What is the component, part number 1090329, on board UP-76351 in


circuit ZD251?
_________________________________________________________________

l.

UPC811C and UPC812C are what type of IC chips?


_________________________________

m.

UPD 74HC4040C is what type of device?


_________________________________________

n.

What happens if Pin number 6 on the HD 637B01X0 CPU chip goes low?
_________________________________________________________________

o.

List the three methods of defibrillation?


1. _________________________________________
2. _________________________________________
3. _________________________________________

DEMONSTRATION OF MEDICAL EQUIPMENT:

1. Demonstrate the ability to check the power supply


voltage +12 Vb and how to adjust if out of
specifications.
2. If the internal discharge relay is not working properly, what would be the
symptoms on the tec7300 defibrillator? Use manual to prove.
3. Demonstrate what changed on the crt driver board with the factory upgrade
UP 7740 DWG NO. 105662A to 105662B.
4. Demonstrate the ability to do a delivered energy check.
5. In general terms, what do you think would happen if C301 on the ecg/cpu
board would short to ground?