Prerequisites:
This procedure requires the use of part number 06775 (3/32 tube with fittings). This part is
standard in toolkits shipped after April 2006. The tube attaches to the syringe in the toolkit, and is
very useful for this test. If you have an older toolkit that does not contain this part, please order
the part before using this procedure.
Test hose with luer fitting on one end and HP fitting on the other (06775)
Syringe w/ Part
#06775 tube with
fittings
Note-Remove
plunger before
starting test!
1/3
Procedure:
Test Procedure:
1. SET UP: From ZPrint, choose Unpark from the service menu (this will move the carriage
away from the service station, and prevent it from auto parking when you close the top
cover.
IF YOU DONT HAVE ACCESS TO ZPRINT OR COMMUNICATION WITH THE
PRINTER:
The safest option is to power down the printer and manually move the carriage to the
front, right side of the deck.
2.
3.
4.
5.
End of tube
fully extended
(do not force)
2/3
Results:
MINIMUM FLOW RATES (ml/30sec.)
Bottle full
7-8
Bottle empty
4-5
The flow rate test is designed to test whether the system may have a kinked or
blocked fluid line, or a clogged binder filter. If you find a restriction in the flow
and feel that it may be caused by bacteria build-up in the fluid line or filter, it will
be necessary to follow the clean out procedure 08799 to make sure the bacteria is
bleached before returning the machine to operation.