at Air Liquide
Martine Carr (Laboratory & Analysis
European Platform)
Jrme Perrin (Centre de Recherche
Claude Delorme)
Outline
ALPHAGAZ
Analytical Methods
H2O and O2
H2, CO, CO2, CH4
NOx and S Compounds
ALMAX : IR diode laser hygrometry
APIX : Atm. Pressure Ion Mass Spec
Analytical services : VOC in ambient Air
Round Robin :
ALPHAGAZ
2 standard
specifications of gas
purity
www.airliquide.com
Shop@Air Liquide
ALPHAGAZ Ar qualities
GC detectors
ALPHAGAZ 1
Publications
CD (Sulfur)
DID or HID
PDHID
M. Gastiger
& B. Jurcik
ELCD
Guideline to hep
to choose the
right gas and
gas equipment
in your
laboratory;
Parts 1, 2 ,3 & 4
AED
PID
TCD
FID
ECD
NPD (N)
NPD (P)
International
Laboratory
(1999,2000)
FPD (S)
FPD (P)
FTIR
MS (SIM)
MS (Scan)
Initial
concentration
in solution
Initial molar
concentration in
gaseous phase
1 g/l
vol %
1 g/l
1 ppmv
Alphagaz 1
EQUIPMENT
1 mg/l
Pure
installation & procedures
Alphagaz 2
1 ppbv
Alphagaz 2
Highly pure
Gas Purity Workshop, LNE, 8 Sept. 2003
Ultra pure
H2 O & O2
Analysis
SAMPLE
ELECTROCHEMICAL
CELL
H2O ANALYSIS
QUARTZ CELL
PURIFIER
SAMPLE
REFERENCE
DIFFERENTIAL ANALYSIS
OF WATER
H2O
trapped
Gas Purity Workshop, LNE, 8 Sept. 2003
CYLINDER
REFERENCE
10
H2
CO
CO2
CH4
Oxygen
2ppb
1ppb
5ppb
2ppb
Nitrogen
2ppb
1ppb
1ppb
1ppb
Argon
2ppb
1ppb
1ppb
1ppb
Hydrogen
4ppb
1ppb
2ppb
Helium
2ppb
1ppb
1ppb
1ppb
11
12
Sulfur Compounds
Analysis
development
In partnership
with Varian
and Sievers
PFPD
dL (ppb)
OCDL*
SCD
10X **
dL (ppb)
OCDL*
10X **
H2S
24.3
4.86
2.07
0.9
0.18
3.95
COS
26.7
5.34
2.10
1.3
0.26
3.94
SO2
29.4
5.88
2.10
Methyl Mercaptan
20.0
4.0
2.09
1.1
0.28
4.07
Ethyl Mercaptan
20.6
4.12
2.09
1.4
0.22
3.83
Thiophene
38.8
7.76
2.08
1.5
0.30
3.93
13
APIX
Atmospheric
Pressure
Ionisation Mass
Spectrometry
On-line analysis of major impurities -down to ppt level
in Semiconductor bulk gases N2, Ar, He , H2 and O2
Multi
Sampler
N2
He
Ar
H2
O2
API
MS
H2O
O2
CO
CO2
CH4
N2
others
14
ALMAX
IR
laser
diode
power
Optical
System
10m
GAS
CELL
Detection Limit :
10 ppb for H2O
(50 ppb for NH3)
Photoelectric signal
Detector
15
VOC
Analysis
Analytical
Service
Applications in electronics:
Clean room air monitoring
Analysis of mini-environment of tools
Clean Room
H2 O
SAMPLING
Tenax
Glass beads
Canister
CO2
Mass Spectrometry
GC MS
Cryogenic
Focuser
CONCENTRATION
SEPARATION ANALYSIS
16
1999-2000
1 mixture
CO in N2
2 mixtures
O2 in N2
CO, CO2, C3H8,
O2 in N2
11 European
laboratories
+ Canada
+ Singapore
11 European
laboratories
+ Canada
+ Singapore
2001
12 European laboratories
+ Singapore
2 mixtures
NO in N2
CH4, C2H2, C2H4,
C2H6, C3H8 in N2
12 American laboratories
(US + Canada + Brazil)
1 mixture
02, CO, CO2, C3H8
in He
17
18
Data treatment
Based on norm ISO 5725:
Description of specific conditions for interlaboratory experiments:
Use one standard measurement method
Samples have to be identical when dispatched to the laboratories
Samples have to remain identical during transport and during the time
interval before measurement
one sample per lab and short time between preparation and analysis
19
Analysis of NO in N2
Evaluation of trueness
4
h, statistique de Mandel
Trueness:
Labo 6
3
2
1
0
Labo 1
Labo 2
Labo 7
Labo 4
Labo 8
Labo 5
-1
Labo 9
Labo 3
-2
-3
h1%
h5%
h1%
h5%
Precision:
Evaluation of precision
k, statistique de Mandel
2.500
Labo 9
2.000
Labo 8
1.500
Labo 2
1.000
0.500
Labo 1
Labo 4 Labo 5
0.000
Labo 3
k
Labo 6
k1%
Labo 7
k5%
Gas Purity Workshop, LNE, 8 Sept. 2003
20
TRUENESS OF LABORATORIES
4
3
NO
CH4
C2H2
C2H4
C2H6
C3H8
2
1
0
-1
-2
La
bo
6
La
bo
7
La
bo
8
La
bo
9
La
bo
10
La
bo
11
La
bo
La
bo
La
bo
La
bo
-3
La
bo
h of Mandel
1% risk limit
21
PRECISION OF LABORATORIES
2.5
2
NO
CH4
1.5
C2H2
C2H4
C2H6
C3H8
0.5
La
bo
10
La
bo
11
La
bo
9
La
bo
8
La
bo
7
La
bo
6
5
La
bo
4
La
bo
3
La
bo
2
La
bo
La
bo
k of Mandel
1% risk limit
22
Results Summary
In general the agreement between the gravimetric value
and the global analytical mean value is good within the
gravimetric uncertainty
A slight discrepancy between gravimetric value and global
analytical mean value has helped identifying a problem
with a C2H2 source
Corrective measures are taken for laboratories outside of
acceptable limit of deviation from global mean value
The regular and systematic practice of these internal
"Round Robin" is continued