ECN No.

2009-E-0206

ENGINEERING CHANGE NOTICE (FIELD CHANGE NOTICE)


SUBJECT: Releasing the Revised Reinstallation Procedure for CR-IR 391CL and CR-IR
391V CL
Design Change Reason
Specification/Function Change
Pro-active Improvement
Serviceability
Part Replacement
Corrective Action
Preventive Action
Regulatory Compliance
Other (
)

Action
Mandatory
Recommended
Not required ( Information only )
Other (
)

Time frame
Immediately
(
)
Next maintenance
Upon failure
At your convenience
Preventive maintenance
Not Applicable
Other ( On reinstallation
of
OS
and
application )

Dept.: Medical Systems Business Div.


DATE: Jul. 13, 2009
EDITED BY: K.Ikeda APPROVED BY: Y.Koyanagi
ORIGINATOR: H.Ichikawa
AFFECTED PRODUCTS: CR-IR 391CL
ATTACHMENT:
YES
NO
TOTAL PAGE (S): 6
RELATED INFORMATION:
None
*********************************************************************************************************************************
DESCRIPTION OF CHANGE:
When you reinstall the OS and the application for CR-IR 391CL and CR-IR 391V CL, the reinstallation fails if
you perform the procedure according to the service manual. This ECN is to announce the correct
reinstallation procedure.
Affected document:
CR-IR 391CL Service Manual / 020-221-01E

- APPLICABLE TO (Serial Nos. etc.) :


- ESTIMATED WORK TIME:
- SPECIAL TOOLS / INSTRUMENTS:
- SUPPLY OF THE ARTICLES:
IF YES, WHEN:
NAME / CODE:
QUANTITY:

CR-IR 391CL, CR-IR 391V CL


N/A
YES
NO
YES
NO

- RETURN OF THE REPLACED PARTS TO FUJI:


DOCUMENTS AFFECTED:

YES

YES

NO

NO

AFFECTED
CURRENT
DOCUMENTS
REV
ISSUED DATE
CR-IR
391CL
Service 01E
Apr 29, 2009
Manual 020-221

REV

NEW
ISSUED DATE

DETAILS:
See the attachment for details.

The contents described or shown in our ECN documents and/or any information provided from FUJIFILM in verbal or written communication is intended for the
addressee and/or its organization. Transfer and/or copy distribution of the information without written consent of FUJIFILM is not permitted.
PAGE 1 of 2

K12N0028-6-J

ECN No. 2009-E-0206

NOTE
1) Important information about SUPPLY OF THE ARTICLES:
Charge:
FREE OF CHARGE
UNTIL (mmddyyyy):
WITH YOUR CHARGE
NOT APPLICABLE

The contents described or shown in our ECN documents and/or any information provided from FUJIFILM in verbal or written communication is intended for the
addressee and/or its organization. Transfer and/or copy distribution of the information without written consent of FUJIFILM is not permitted.
PAGE 2 of 2

K12N0028-6-J

ECN 2009-E-0206 1/4


Corrigenda of CR-IR 391CL Service Manual (020-221-01E)

P448 (MC-36)
1.4.1 Installing .NET Framework
The installation fails if you perform "install-ASPNET". Do not perform the procedure 14 and 15.

Incorrect
13. Click [OK].
14. Open folders in the following sequence: D drive,
dotNet, and DNET1.1.
15. Duble-click Install-ASPNET.
The following window opens.

correct
13. Click [OK].
The installation fails if you perform "install-ASPNET".
Do not perform the procedure 14 and 15.

Upon completion of installation, the system returns to the


desktop screen.
16. Restart the PC.
17. When a login window opens after the restart of the PC,

16. Restart the PC.


17. When a login window opens after the restart of the

enter system as the password. Click .


The desktop window open.

PC, enter system as the password. Click .


The desktop window open.

P451 (MC-39)
1.4.3 Adding the Windows Components
Do not install "Tablet PC Optional Components" because it is unnecessary.

Incorrect
4. Put checkmarks for the following items.
ActiveX Installer Service
FTP Publishing Service
Web Management Tools
.NET Extensibility
ASP .NET
ISAPI Extensions
ISAPI Filters
Common Http Features
HTTP Logging
Security
XPS Viewer
Remote Differential Compression
Tablet PC Optional Components
Telnet Client
Windows DFS Replication Service
Windows Process Activation Service

correct
4. Put checkmarks for the following items.
ActiveX Installer Service
FTP Publishing Service
Web Management Tools
.NET Extensibility
ASP .NET
ISAPI Extensions
ISAPI Filters
Common Http Features
HTTP Logging
Security
XPS Viewer
Remote Differential Compression
Telnet Client
Windows DFS Replication Service
Windows Process Activation Service

ECN 2009-E-0206 2/4

ECN 2009-E-0206 3/4


P557 (MU-17)
5.5 Setting the Image Reader
Unfailingly remove the checkmark from "Check the box to use Patient information display". If it is
checkmarked, CR-IR 391RU cannot be connected because it cannot display patient information.
Incorrect

correct

1. Enter an Image Reader name (default: Image Reader) in


the Equipment name field.
2. Enter an Image Reader host name (default: RU0) in the
Host name field.
3. Enter an Image Reader IP address (default: 172.16.1.10)
in the IP Address field.
4. Checkmark Use in the Check the box to use Patient
).
information display field (
NOTE
To have the patient information be displayed, the relevant
setting must be turned ON also on the RU side.
For details of the procedure to use, see Appendix 4: Setting
the RU-Config in MC: Checks, Replacement and
Adjustment of Parts.

1. Enter an Image Reader name (default: Image Reader)


in the Equipment name field.
2. Enter an Image Reader host name (default: RU0) in
the Host name field.
3. Enter an Image Reader IP address (default:
172.16.1.10) in the IP Address field.
4. Remove the checkmark from Check the box to use
Patient information display.
NOTE
If "Check the box to use Patient information display" is
checkmarked, CR-IR 391RU cannot be connected.
Unfailingly remove the checkmark from it.

5. Click [Next].
The Simplified settings completed window opens.

5. Click [Next].
The Simplified settings completed window opens.

ECN 2009-E-0206 4/4

P739 (IN-27)
4.4.6 Setting the Image Reader
Unfailingly remove the checkmark from "Check the box to use Patient information display". If it is
checkmarked, CR-IR 391RU cannot be connected because it cannot display patient information.
Incorrect

1. Enter an Image Reader name (default: Image Reader) in


the Equipment name field.
2. Enter an Image Reader host name (default: RU0) in the
Host name field.
3. Enter an Image Reader IP address (default: 172.16.1.10)
in the IP Address field.

4. Click [Next].
The Simplified settings completed window opens.

correct

1. Enter an Image Reader name (default: Image Reader)


in the Equipment name field.
2. Enter an Image Reader host name (default: RU0) in
the Host name field.
3. Enter an Image Reader IP address (default:
172.16.1.10) in the IP Address field.
4. Remove the checkmark from Check the box to use
.
Patient information display
NOTE
If "Check the box to use Patient information display" is
checkmarked, CR-IR 391RU cannot be connected.
Unfailingly remove the checkmark from it.
5. Click [Next].
The Simplified settings completed window opens.