Flow sensor for liquid

media type 236


Flow range
1.8 ... 240 l/min
Nominal diameters
DN 10 / 32
Temperature measurement
-40 ... +125 C

The type 236 is based on the type 210 but incorporates a


brass housing. The Vortex Sensor type 236 has a rugged
construction of brass connection. This flow sensor is available
with a larger variety concerning power supply and outputs.
You can choose between various versions as integrated
temperature measurement. With no moving parts the flow
sensor is not sensitive to debris, has marginal pressure loss and
high accuracy.

Flow measuring with voltage, current


or frequency output
Temperature non-sensitive measuring
principle
Excellent media resistance (measuring
element not in contact with the
media)
CE conformity
Wide application temperature range
Marginal loss of pressure
Measuring element not sensitive to
debris
Direct temperature measurement in
the medium
Drinking water approval
KTW, W270, WRAS

Huba Control type 236 - Technical data subject to change - Edition 10/2014

1/5

Technical Overview
Flow measurement
Measuring principle
Measuring range
Nominal diameters
Accuracy at < 50% fs (water)
Accuracy at > 50% fs (water)
Immediately
Therefore suitable for spigot use.
Response time

Vortex

Frequency output
Analogue output

Temperature measurement
Measuring principle
Resistance
Measuring range
PT1000
Accuracy

@ T = 0 C
@ T 0 C

class B DIN EN 60751


Measuring range
Accuracy

0 ... 10 V

Calculation temperature
Self-heating at temperature sensor
Conduction resistance to connector

Temperature influences

Piezoelectric sensor element


1.8 ... 240 l/min
DN 10 / 25
< 1% fs
< 2% measuring value
Signal delay
< 100 ms
Response time
< 5 ms
Signal delay
< 2s
Response time
< 500 ms

PT1000
-40 ... +125 C
0.3 K
0.3 K 0.005 * T
-25 ... +125 C
0.5 K 0.005 * T
T (C) = +150 C
10 V * UOUT_T - 25 C
1 K/mW
0.8 Ohm

Operating conditions
Suitable for heating circuit water with the usual additives
Drinking water
Media
Ambient
Storage
(for lifetime)
(for lifetime)
(for 600 hours)
(for 2 hours)
(max. test pressure)
The following equation is valid to prevent cavitation:

Medium
temperature

Max. pressure and


medium temperature
Cavitation
Materials in contact with medium (FDA-conform)
Sensor paddle
Case with damming body

Other medium on request


< +125 C
-15 ... +85 C
-30 ... +85 C
12 bar at +40 C
6 bar at +100 C
4 bar at +125 C
4 bar at +140 C
18 bar at +40 C
Pabs outlet / Pdifference > 5.5
ETFE
Brass (CuZn40PbZ), PA6T/6I (40% GF)
EPDM (perox.)
FPM

Sealing material
Electrical overview
Power supply
UIN
Output
Frequency square pulse signal UOUT_Q_frequency
Flow (Q)
Analogue signal
UOUT_Q or IOUT
Output
Resistant signal
ROUT PT1000
temperature (T)
Voltage signal
UOUT_T
Electrical connection and protection class
Load agianst GND or IN
Current consumption load free (IIN)

Frequency output
4.75 ... 33 VDC
< 0.5 ... > UIN - 0.5 V

M12x1 (IP 65)


< 1 mA / < 100 nF
< 2mA

Weight
DN 10 with thread K
DN 10 with thread G
DN 32

Voltage output
11.5 ... 33 VDC
0 ... 10 V
PT1000 class B DIN EN 60751
0 ... 10 V
M12x1 (IP 65)
< 6 mA / < 100 nF 1)
< 5 mA

Current output
8 ... 33 VDC
4 ... 20 mA

M12x1 (IP 65)


< (UIN - 8 V) / 20 mA

~ 170 g
~ 250 g
~ 650 g

Test / Admissions
Electromagnetic compatibility

CE conformity acc. to EN 61326-2-3


WRAS
Plastic parts with KTW and W270 approval

Drinking water approval


Packaging
Single packaging
Multiple packaging

Temperatur in C

Temperature in C

Minimum life span on high flow rate and high temperature


135
130
125
120
115
110
105
100
95
90
85
80
75
70
65
60
55
50

hours
22Std
10Std
std
10
100Std
std
100
0.1Jahr
year
0.1
year
11Jahr
10Jahre
year
10

40

60

80

100

120

140

Flow in
in %
%
Durchfluss

2/5

Huba Control type 236 - Technical data subject to change - Edition 10/2014

Nominal diameters dependent variables


Nominal
diameters

Tube connection

Measuring
range

Quantity per pulse


@ 50% fs

DN 10

1.8 ... 32 l/min

1.416 ml

Flow range

Frequency
range

1.8 ... 32 l/min

DN 10

2.0 ... 40 l/min

1.419 ml

DN 10

2.0 ... 40 l/min

1.386 ml

DN 32

14 ... 240 l/min

27.513 ml

Kf

23 ... 374 Hz
0.265 ... 4.716 m/s

DN 10

Q0

1.386 ml

24 ... 380 HZ

Characteristic line formula


frequency output
QV = Kf * + Q0

Characteristic line formula


voltage output
QV = KU * UOUT

Characteristic line formula


current output
QV = KI * (IOUT - 4 mA)

Formula quantity per pulse [litres/pulse]


quantity
Q * Kf
= V
pulse
60 * (QV - Q0)

3.2

2.000

22.50 * Q2
22.50 * Q2

0.0860

22.50 * Q2

-0.2

4.0

26 ... 479 Hz

9 ... 145 Hz

Pressure drop 1), 2)

0.0847

26 ... 467 Hz

0.290 ... 4.974 m/s

KI

0.0860
-0.2

0.295 ... 5.895 m/s

KU

2.500

0.0840

-1.47

Legend
QV
Q0
Kf
KU
KI

UOUT
IOUT
quantity
pulse

22.50 * Q2

1.64710

24

15.000

Volume flow rate


Axis intercept
Coefficient frequency output
Coefficient voltage output
Coefficient current output
Frequency
Voltage
Current

Nominal diameters and


flow range
Output and power supply

Electrical connection
Sealing material
Tube connection

[l/min]
[l/min]
[(l/min) / ]
[(l/min) / V]
[(l/min) / ]
[Hz]
[V]
[mA]
litres
pulse

Quantity per pulse

Order code selection table


Version

0.25 * Q2

236.

X X X X X X X

Flow
Flow and temperature (PT1000)
Flow and temperature (0 ... 10 V)
DN 10
1.8 ... 32 l/min.
DN 10
2.0 ... 40 l/min.
DN 32
14.0 ... 240 l/min.
Frequency output (Square pulse signal)
4.75 ... 33 VDC
Analogue signal 0 ... 10 V
11.5 ... 33 VDC
Analogue signal 4 ... 20 mA
8 ... 33 VDC
Connector M12x1
2- or 3-pole
(condensation protection)
Connector M12x1
4- or 5-pole
(condensation protection)
EPDM
Ethylene propylene rubber (peroxidically cross-linked)
FPM 3)
Fluoro elastomer
Brass with outside thread K (DN 10 - G , DN32 - G 1 )
Brass with outside thread G (DN 10 - G 1)

9
8
6

3
1
1
3

4
5
5

0
1
2

8,9

2
3
4

8,9
9
8,6

4
5
1
2
K
G

Accessories 4)
Straight-wire box for connector M12x1 with cable
Corner-wire box for connector M12x1 with cable
Straight-wire box for connector M12x1 with cable
Corner-wire box for connector M12x1 with cable
Straight-wire box for connector M12x1 screwing terminal

3-pole
3-pole
5-pole
5-pole
5-pole

200 cm
200 cm
200 cm
200 cm

Order number
114605
114604
114564
114563
115024

(with temperature)
(with temperature)

Dimension diagram DN 10, 32


1
1

DN10

43

57.3

G1

86

19

DN32

50

74.9

G1

134

41

2
3

4
1)

incl. 3xDi inlet and outlet side

2)

Pv in Pa; Q in l/min

Huba Control type 236 - Technical data subject to change - Edition 10/2014

3)

No drinking water approval

4)

Accessories supplied loose

3/5

Tube mounting instructions


Consider the following to ensure the correct function of the sensor.
Only diameter changes from large to small are allowed.
Avoid repeated elbows in the same level at entryside

minimum 5xDN
for alternative elbows

minimum 0.5xDN
for recommended 90 elbow
with min. R 1.8xDN

Electrical connection
Connector M12x1 without temperature measurement
IN

Hz

GND

IN

OUTQ
OUT

current output

Frequency output

GND

1
2

GND

IN

brown
white
blue
black
gray

A
Hz

GND

IN

Colour
brown
blue
black

voltage output

Connector M12x1 with temperature measurement

1
2
3
4
5

OUT

IN

Pin
1
3
4

GND

GND

IN

4
5

OUT
OUTQ
T2

4/5

T1

T2

T1

T2

4
5

GND

IN

OUTQ

T1
OUTT

Frequency output
with PT1000

current output
with PT1000

voltage output
with PT1000

voltage output with


temperature output 0 ...10 V

Huba Control type 236 - Technical data subject to change - Edition 10/2014

Huba Control AG
Headquarters
Industriestrasse 17
5436 Wrenlos
Telefon +41 (0) 56 436 82 00
Telefax +41 (0) 56 436 82 82
info.ch@hubacontrol.com

Huba Control AG
Niederlassung Deutschland
Schlattgrabenstrasse 24
72141 Walddorfhslach
Telefon +49 (0) 7127 23 93 00
Telefax +49 (0) 7127 23 93 20
info.de@hubacontrol.com

Huba Control SA
Succursale France
Rue Lavoisier
Technople Forbach-Sud
57602 Forbach Cedex
Tlphone +33 (0) 387 847 300
Tlcopieur +33 (0) 387 847 301
info.fr@hubacontrol.com

Huba Control AG
Vestiging Nederland
Hamseweg 20A
3828 AD Hoogland
Telefoon +31 (0) 33 433 03 66
Telefax +31 (0) 33 433 03 77
info.nl@hubacontrol.com

Huba Control AG
Branch Office United Kingdom
Unit 13 Berkshire House
County Park Business Centre
Shrivenham Road
Swindon - Wiltshire SN1 2NR
Phone +44 (0) 1993 776667
Fax
+44 (0) 1993 776671
info.uk@hubacontrol.com

www.hubacontrol.com
5/5

Huba Control type 236 - Technical data subject to change - Edition 10/2014

Anda mungkin juga menyukai