e1 NOTEWarnings were moved from notes to section text editorially January 2003.
1. Scope
1.1 This test method covers the determination of the boiling
range distribution of water-free crude petroleum through
538C (1000F). Material boiling above 538C is reported as
residue. This test method is applicable to whole crude samples,
that can be solubilized in a solvent to permit sampling by
means of a microsyringe.
1.2 The values stated in SI units are to be regarded as the
standard. The values stated in inch-pound units are for information only.
1.3 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the
responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Specific precautionary statements are given in 7.2, 7.5, 7.6, 7.7, and 7.9.
2. Referenced Documents
2.1 ASTM Standards:
D 2892 Test Method for Distillation of Crude Petroleum
(15-Theoretical Plate Column)2
D 4057 Practice for Manual Sampling of Petroleum and
Petroleum Products3
3. Terminology
3.1 Definitions of Terms Specific to This Standard:
3.1.1 area slice, nthe area, resulting from the integration
of the chromatographic detector signal, within a specified
retention time interval.
3.1.1.1 DiscussionIn area slice mode (see 6.2.2), peak
detection parameters are bypassed and the detector signal
integral is recorded as area slices of consecutive, fixed duration
time intervals.
1
This test method is under the jurisdiction of ASTM Committee D02 on
Petroleum Products and Lubricants and is the direct responsibility of Subcommittee
D02.04.0H on Chromatographic Methods.
Current edition approved June 10, 1997. Published October 1997. Originally
published as D 5307 92. Last previous edition D 5307 92.
2
Annual Book of ASTM Standards, Vol 05.01.
3
Annual Book of ASTM Standards, Vol 05.02.
Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
D 5307 97 (2002)e1
provided for programming the entire column, including the
point of sample introduction, up to the maximum column
temperature employed.
6.1.5 Flow ControllerThe chromatograph must be
equipped with a flow controller capable of maintaining carrier
gas flow constant to 6 1 % over the full operating temperature
range of the column. The inlet pressure of the carrier gas,
supplied to the chromatograph, must be sufficiently high to
compensate for the increase of backpressure in the column as
the temperature is programmed upward. An inlet pressure of
550 kPa gage (80 psig) has been found satisfactory with the
columns described in Table 1.
6.2 Data Retrieval System:
6.2.1 RecorderA 01 mV range recording potentiometer
or equivalent, with a full-scale response time of 2 s or less may
be used for graphic presentation of the FID signal.
6.2.2 IntegratorElectronic integrator or computer-based
chromatography data system must be used for detector signal
integration and accumulation. The integrator/computer system
must have normal chromatographic software for measuring
retention time and areas of eluting peaks (peak detection
mode). In addition, the system must be capable of converting
the continuously integrated detector signal into area slices
representing contiguous fixed duration time intervals (area
slice mode). The recommended time interval is 1 s. No time
interval shall be greater than 12 s. The system must be capable
of subtracting the area slice of a blank run from the corresponding area slice of a sample run. Alternatively, the baseline
chromatogram can be subtracted from the sample chromatogram and the net resulting chromatogram can be processed in
the slice mode. A computer program that performs the slice
calculation as a post-run calculation is also used.
6.3 ColumnAny gas chromatographic column that provides separation in order of boiling points and meets the
performance requirements of Section 10 can be used. Columns
and conditions, which have been used successfully, are shown
in Table 1.
6.4 MicrosyringeA 5 or 10 L syringe is used for sample
introduction. The use of an automated liquid sampling device
is highly recommended.
457 (18)
3.17 (18)
10 % UCW-982
610 (24)
3.17 (18)
3 % OV-1
Support material
Chromosorb
PA-AW
Chromosorb Chromosorb
WA-HP
PA-AW
30
30
40
380
350
360
10
N2
25
400
380
10
He
20
380
375
10
N2
28
400
400
457 (18)
3.17 (18)
10 % SE-30
4
Reagent Chemicals, American Chemical Society Specifications, American
Chemical Society, Washington, DC. For suggestions on the testing of reagents not
listed by the American Chemical Society, see Analar Standards for Laboratory
Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeia
and National Formulary, U.S. Pharmacopeial Convention, Inc. (USPC), Rockville,
MD.
See Footnote 5.
D 5307 97 (2002)e1
9. Preparation of Apparatus
8. Sampling
8.1 Obtain samples for analysis by this test method in
accordance with instructions given in Practice D 4057.
8.1.1 Ensure that samples are received in sealed containers
and show no evidence of leakage.
(1)
D 5307 97 (2002)e1
11.1.2 Following a rigorously standardized schedule, cool
the column to the selected starting temperature, and allow it to
equilibrate at this temperature for at least 3 min. At the exact
time set by the schedule, without injecting a sample, start the
column temperature program.
11.1.3 Acquire the data in area slice mode (see 6.2.2),
recording the area slices for each time interval from the start of
the run until the end of the run. It is essential that all
measurements be on the same time basis for the blank and
sample runs.
11.1.4 Perform a blank analysis at least once each day
analyses are performed.
where:
R
= resolution,
t1 = time for the n-C16 peak apex, in seconds,
t2 = time for the n-C18 peak apex, in seconds,
Y1 = peak width, at half height, of n-C16, in seconds, and
Y2 = peak width, at half height, of n-C18, in seconds.
10.2 Retention Time RepeatabilityThe system must be
sufficiently repeatable, when testing with the calibration mixture, to obtain retention time repeatability (maximum difference between duplicate runs) of 6 s or less for each calibration
peak.
10.3 System Performance CheckAnalyze the detector response test mixture (see 7.8) utilizing identical conditions as
used in the analysis of samples. Calculate response factors
relative to n-decane as follows:
Fn 5 ~Cn/An!/~C10/A10!
(2)
where:
An = area of that n-paraffin peak,
A10 = area of n-decane peak,
Cn = concentration of that n-paraffin in the mixture,
C10 = concentration of n-decane in the mixture, and
Fn = response factor relative to n-decane.
10.3.1 The response factor (Fn) of each n-paraffin must not
deviate from unity by more than 10 %.
10.3.2 With some chromatographs, response factors for
higher boiling n-paraffins (n-C20 to n-C44) have been observed
to change after several crude oil samples have been analyzed.
Check the stability of the system by repeating the performance
test after analyzing ten samples. If the system still meets the
performance specified (see 10.3.1), it is not necessary to repeat
this check after subsequent analyses. However, it is good
practice to repeat the performance test if detector components
are changed.
11. Procedure
11.1 Baseline Compensation AnalysisTo compensate for
baseline drift and signal offset, subtract an area slice profile of
a blank run from the sample run to obtain corrected area slices.
This profile is obtained as follows:
11.1.1 After conditions have been set to meet performance
requirements, program the column oven temperature upward to
the maximum temperature to be used and hold for at least ten
minutes.
4
D 5307 97 (2002)e1
Add 2 to 3 g of drying agent, such as anhydrous calcium
chloride, to a 50-mL vial and fill the vial about half full with
sample. Cap the vial tightly and shake the vial vigorously.
Allow the mixture to stand several minutes to allow the drying
agent to settle out. By means of a disposable pipette, remove
the dried oil layer for sample weighing and analysis.
11.3.5 Weigh at least 10 g of dried sample to the nearest 0.1
mg into a 25-mL vial.
11.3.6 Add approximately 1 g of internal standard mixture
into the same vial. Determine the weight to the nearest 0.1 mg.
11.3.7 Dilute the mixture with an approximately equal
volume of carbon disulfide.
11.3.8 Cap the vial tightly and shake the mixture vigorously
for 3 min, or until the mixture is solubilized completely. Use
this solution for the crude oil plus internal standard analysis
(see 11.4.1).
11.3.9 In a second vial, dissolve approximately the same
amount of dried sample as 11.3.5 with an approximately equal
volume of carbon disulfide. Use this solution for the separate
crude oil without internal standard analysis (see 11.4.4).
11.4 Sample Analysis:
11.4.1 Using the exact conditions that were used in the
blank and calibration runs (see 11.1 and 11.2), and following
the rigorously defined schedule (see 11.1), inject 1 L of the
diluted crude oil plus internal standard mixture into the
chromatograph. Record the area slices of each time interval
through the end of the run.
11.4.2 Continue the run until the retention time equivalent
to a boiling point of 538C (1000F) is reached. Stop recording
area slices under the chromatogram at this point.
11.4.3 To remove as much as possible of the heavy components remaining on the column, continue heating the column
until the FID signal returns to baseline. The column temperature may be increased to speed this process.
11.4.4 Cool the column to the starting temperature. Use
identical conditions as used in 11.4.1. Inject 1 L of the crude
oil sample without internal standard (see 11.3.9). Record the
area slices of each time interval through the end of the run.
11.4.5 The sample plus internal standard analysis (see
11.4.1) and the sample only analysis (see 11.4.4) may be made
in either order.
BP, C
BP, F
1
2
3
4
5
6
7
8
9
10
162
89
42
0
36
69
98
126
151
174
259
128
44
32
97
156
208
259
304
345
11
12
13
14
15
16
17
18
19
20
196
216
235
254
271
287
302
316
330
344
385
421
455
489
520
549
576
601
626
651
21
22
356
369
674
695
Carbon
Number
BP, C
BP, F
23
24
25
26
27
28
29
30
380
391
402
412
422
431
440
449
716
736
756
774
792
808
825
840
31
32
33
34
35
36
37
38
39
40
458
466
474
481
489
496
503
509
516
522
856
871
885
898
912
925
937
948
961
972
41
42
43
44
528
534
540
545
982
993
1004
1013
12. Calculation
12.1 Area Corrections:
12.1.1 Obtain corrected area slices for both runs (see 11.4.1
and 11.4.4) by subtracting the corresponding area slice of the
blank run profile (see 11.1) from each. (See Note 2.)
12.1.2 Sum the corrected area slices for both runs to obtain
the cumulative corrected area at the end of each time interval
during the run.
12.2 Theoretical Total Area (Refer to Fig. 3):
12.2.1 Based on retention times from the calibration chromatogram (see 11.2.1), select a retention time that is 5 % less
than the retention time of n-C14, and another that is 5 % greater
than the retention time of n-C17. These times define a segment
of the chromatogram that includes the internal standard peaks.
Record the total area within this segment from the chromatogram of the crude oil plus internal standard mixture (see
11.4.1) (area AIS from Fig. 3(a)). Also record the total area of
See Footnote 7.
D 5307 97 (2002)e1
BIS
(5)
where:
AIS = total area of the internal standard segment of the
chromatogram of the sample plus internal standard
mixture,
BIS = total area of the internal standard segment of the
crude oil only chromatogram,
r
= the ratio of areas outside the internal standard
segment through 538C (1000F) for both chromatograms (see Eq 4), and
W
= the mass fraction of internal standard in the mixture
of crude sample plus internal standard (see Eq 3).
12.2.6 Calculate the percent residue (RES) above 538C
(1000F) as follows:
RES 5 100 2 ~B/T 3 100!
where:
B = total area through 538C (1000F) of the crude only
chromatogram, and
T = theoretical total area of the crude oil only chromatogram (see Eq 5).
12.3 Calculation of Boiling Point Distribution:
12.3.1 Record the time at which the cumulative area at the
beginning of the crude only chromatogram is equal to 0.5 % of
the theoretical total area (T, From Eq 5). The temperature
equivalent to this time is the IBP of the sample.
12.3.2 Multiply the corrected cumulative area at the end of
each time interval by 100 and divide by the theoretical total
area (T from Eq 5). This gives the percent of sample recovered
at the end of each time interval.
12.3.3 Tabulate, in pairs, the cumulative percent recovered
and the retention time at the end of each time interval.
12.3.4 Using linear interpolation where necessary, determine the time associated with each percent between 1 % and
the percent eluted at the time equivalent to 538C (1000F).
12.3.5 For each 1 % and its associated retention time,
determine the corresponding temperature from the table of
boiling point-retention time calibration data (see 11.2.2).
(3)
where:
I = mass of internal standard, g, and
S = mass of sample, g.
12.2.4 Calculate the ratio of areas (r) outside the internal
standard segment and through the retention time equivalent to
a boiling point of 538C (1000F) of the crude oil only
chromatogram to the chromatogram from the mixture of
internal standard plus crude as follows:
r 5 ~B2BIS!/~A2AIS!
(6)
13. Report
13.1 Report the following information:
13.1.1 The temperature to the nearest 0.5C (1F) at the IBP
and at 1 % intervals, and
13.1.2 The total residue above 538C to the nearest 0.1 %.
(4)
where:
A
= total area through 538C (1000F) of the crude plus
internal standard mixture chromatogram,
AIS = total area of the internal standard segment of the
crude plus internal standard mixture chromatogram,
B
= total area through 538C (1000F) of the crude oil
only chromatogram, and
5
This precision was obtained from an interlaboratory cooperative study by eight
laboratories on five samples. The results of this study have been filed at ASTM
headquarters. Request RR:D02-1295.
6
API Project 44, October 31, 1972.
D 5307 97 (2002)e1
14.1.1 RepeatabilityThe difference between two successive test results, obtained by the same operator with the same
apparatus under constant operating conditions on identical test
material would, in the long run, and in the normal and correct
operation of the test method, exceed the values shown in Table
3 only in 1 case in 20.
14.1.2 ReproducibilityThe difference between two single
independent results obtained by different operators working in
different laboratories on identical test material would, in the
long run, and in the normal and correct operation of the test
method, exceed the values shown in Table 3 only 1 case in 20.
NOTE 3Samples included in the study had residues ranging from
about 3 to 30 %. Samples with residues outside this range may have
different precision.
14.2 BiasThe procedure in this test method for determining the boiling range distribution of crude petroleum by gas
chromatography has no bias because the boiling range distribution can only be defined in terms of a test method.
14.2.1 A rigorous, theoretical definition of the boiling range
distribution of crude petroleum is not possible due to the
complexity of the mixture as well as the unquantifiable
interactions among the components (for example, azeotropic
behavior). Any other means used to define the distribution
would require the use of a physical process such as a
conventional distillation or gas chromatographic characterization. This would therefore result in a method-dependent
definition and would not constitute a true value from which
bias can be calculated.
Repeatability, C
Reproducibility, C
IBP
5
10
20
30
40
50
60
70
80
85
90
Residue
3.7
4.7
6.9
6.8
7.6
9.3
10.6
11.8
17.6
24.8
18.8
20.7
2.6 Mass %
10.6
14.8
11.3
15.4
20.4
24.6
30.3
25.9
39.2
38.8
38.8
44.9
8.1 Mass%
15. Keywords
15.1 crude oil; gas chromatography; petroleum; simulated
distillation
APPENDIX
(Nonmandatory Information)
X1. AGREEMENT WITH CONVENTIONAL DISTILLATION
McTaggart, N. G., Glaysher, P., and Harding, A. F., ASTM STP 577, ASTM,
1973, p. 81.
8
Green, L. E., Chromatograph Gives Boiling Point, Hydrocarbon Processing,
May 1976, p. 205.
9
Worman, J. C., and Green, L. E., Anal. Chem., Vol 37, 1965, p. 1620.
10
Ceballo, C. D. et al. Rev. Tc. INTEVEP, 7(1), 1987, pp. 8183.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org).