Background and Objective: A low cost laser imager was designed and fabricated for measurement of wound geometry.
Methods: The accuracy of the imager was validated using reference depressions of known dimensions. Perimeter, area, and
volume were compared to planimetric and packing techniques
on simulated wound models. Results: Wound tracing and alginate measurement methods required approximately 20 times
longer for the reference standards, and 11 times longer for the
simulated wounds than with the laser scanning method (LSM).
LSM consistently overestimated the reference perimeter by 0.73
0.20 cm and the area by 0.98 0.62 cm2. Volume estimates were
not statistically different. The tracing method underestimated
the perimeter by 0.34 0.27 cm and the area by 1.07 1.09 cm2.
Volume measurements by the alginate method were not statistically different. The perimeters of the simulated wounds averaged 1.29 0.27 cm greater using the LSM than obtained by the
tracing method, and areas greater by 2.02 1.30 cm2. Volume
scans averaged 1.04 0.61 cm3 greater than by the alginate
method. Lasers Surg. Med. 23:8793, 1998. 1998 Wiley-Liss, Inc.
Key words: area; perimeter; wound care; volume
INTRODUCTION
88
Smith et al.
1). A cable extension encoder measures the position of the scan head on the track. The scan head
is moved manually along the track in order to
scan the wound. The scan head and linear track
are mounted by a ball joint to a mobile stand. The
ball joint allows the imaging system to be positioned at any arbitrary angle. A laptop computer
used to operate the system is also attached to the
stand. The laptop has an expansion bay for the
ISA card.
A key component of the system is the custom
ISA video capture board. The video capture board
acquires video frames from each of the two video
cameras, monitors the encoder position, then generates a hardware level interrupt. The video card
also processes the image in real time. The position
of the laser line is calculated for each video scan
line. A threshholding process is used to determine
the location of the laser line. For each camera, 10
bit values for the leading and trailing edge positions of the laser line on each scan line are placed
in a buffer. When a video frame is complete a
hardware interrupt is generated signaling the
host computer to transfer the buffer. Video frames
are captured at multiple locations along the linear
track with each frame producing a profile of data
for each camera.
The system is calibrated by presenting a set
of known values to the imager. A calibration
lookup table is used to convert scan line number
and scan line position to x and y values. The z
value is determined by the encoder position.
The scan sequence is initiated by the host
software. The operator holds down a button to
turn on the laser while moving the scan head
along its track over the wound. The scan ends
when the scan head is returned to the starting
position. The data profiles are then displayed on
the screen.
Software Principles
89
90
Smith et al.
Fig. 2. Simulation of an ulceration on the calf carved into a plaster model of the lower leg.
91
Fig. 3. Simulated shin ulceration with wire mesh interpolation of the skin surface.
92
Smith et al.
1
2
3
4
5
6
Reference
Tracing
LSM
15.58
15.71
11.78
11.78
5.84
5.84
15.52 0.36
15.32 0.11
11.70 0.20
11.56 0.25
5.24 0.21
5.14 0.31
16.35 0.02
16.82 0.16
12.48 0.05
12.45 0.11
6.38 0.07
6.43 0.03
Reference
Tracing
LSM
19.32
19.63
11.05
11.05
2.72
2.72
17.00 0.76
17.12 0.54
10.36 0.16
10.24 0.38
2.68 0.17
2.68 0.10
20.93 0.12
21.38 0.26
12.03 0.10
11.95 0.18
3.00 0.07
3.07 0.02
Reference
Alginate
LSM
29.41
34.63
13.05
14.25
1.44
1.44
29.49 1.25
35.11 0.70
13.50 0.28
14.28 0.10
1.59 0.06
1.62 0.72
29.31 0.09
34.62 0.12
13.10 0.09
13.74 0.10
1.41 0.02
1.42 0.01
*Conventional and laser scanning method (LSM) measurements are statistically different from reference and from each
other.
foot
shin
calf
Tracing
LSM
11.83 0.52
15.00 0.84
12.42 0.20
12.90 0.16
16.20 0.26
14.00 0.13
Tracing
LSM
11.83 0.90
14.41 1.12
9.03 0.50
12.44 0.55
17.58 0.72
11.32 0.44
Alginate
LSM
1.04 0.19
1.49 0.19
1.17 0.05
1.42 0.10
2.64 0.09
2.76 0.05
*Conventional and laser scanning method (LSM) measurements are statistically different.
may be covered by debris [16]. Methods for measuring wound volume include:
1. Saline technique [17]. The wound is covered with a transparent adhesive film,
and injected with saline until it appears
filled.
2. Dental mold technique [18].
3. Alginate cast technique [810].
Positioning of the patient can be problematic for
use of the saline technique in certain circumstances. In addition, removal of the adhesive film
required to perform the technique has been noted
to be painful and unacceptable for some patients.
The technique is not of use in large scale studies.
The dental mold technique has already given
way to the cheaper and faster alginate cast technique. However, objections to its use have included the difficulty in accurately filling the
wound, contact with the wound, and pain due to
packing of the wound. The need to prevent the
cast from drying out requires immediate measurement of the wound volume by suspending it
in a water bath and measuring the amount of displaced water (Archimedes principle) or by calculating the volume based upon the specific gravity
of the material. Such time consuming techniques
are cumbersome in clinical practice and impractical for large scale trials requiring precise, frequently repeated, wound measurements.
The three-dimensional laser scanning device
described herein appears to provide consistent,
The authors thank Wayne Skloss of Instrumentation Services (University of Texas, Health
Science Center at San Antonio, TX) for his clever
engineering implementations and precise machining.
93
REFERENCES
1. Baharestani M. Clinical decision making in wound care
management: The need for a paradigmatic shift. Wounds
1995; 7 Suppl A:84A89A.
2. Harding K. Defining the well-healed wound. Wounds
1995; 7 Suppl A:65A69A.
3. Forrest RD, Gamborg-Nielsen P. Wound assessment in
clinical practice. Acta Med Scand 1984; 687:6974.
4. Dyson M, Suckling J. Stimulation of tissue repair by ultrasound: A survey of the mechanisms involved. Physiotherapy 1978; 64:105108.
5. Yucel VE, Basmajian JV. Decubitus ulcers: Healing effect
of an enzymatic spray. Arch Phys Med Rehabil 1974; 55:
51519.
6. Etris MB, Pribble J, LaBrecque J. Evaluation of two
wound measurement methods in a multi-center, controlled study. OWM 1994; 40:4448.
7. Lancaster JL, Walsh N, Faulkner VW. Basic theory, design, and preliminary evaluation of a laser scanner for
shape sensing below-the-knee amputees. Med Phys 1990;
17:30510.
8. Pories WJ, Schear EW, Jordan DR, Chase J, Parkinson
G, Whittaker R, Strain WH. The measurement of human
wound healing. Surgery 1966; 59:821824.
9. Sutherland F, Cruse P. Wound volume: Measurement
and radiology of casts. Canadian J Surg 1982; 25:364.
10. Resch CS, Kerner E, Robson MC, Heggers JP, Scherer M,
Boertman JA, Schileru R. Pressure sore volume measurement. J Amer Geriat Soc 1988; 36:444446.
11. Forrest RD, Gamborg-Nielsen P. Wound assessment in
clinical practice. Acta Med Scand 1984; 687:6974.
12. Lazarus GS, Cooper DM, Knighton DR, Margolis DJ, Pecorardo RE, Rodeheaver G, Robson MC. Definitions and
guidelines for assessment of wounds and evaluation of
healing. Arch Derm 1994; 130:489493.
13. Sutton J. Accurate wound assessment. Nursing Times
1989; 85:6871.
14. Plassman P, Melhuish JM, Harding KG. Methods of measuring wound size: A comparative study. OWM 1994; 40:
5060.
15. Cuzzell JT. Tell it like it is. A realistic approach to wound
documentation. Amer J Nursing 1986; 86:600601.
16. Margolis DJ, Cohen JH. Management of chronic venous
leg ulcers: A literature-guided approach. Clin In Derm
1994; 12:1926.
17. Berg W, Traneroth C, Gunnarson A, Lossing C. A method
for measuring pressure sores. The Lancet 1990; 335:
14451446.
18. Covington JS, Griffin JW, Mendius RK, Tooms RK, Clifft
JK. Measurement of pressure ulcer volume using dental
impression materials: Suggestion from the field. Physical