Anda di halaman 1dari 2

Akashganga

Rakshak
India is on No. 1 position amongst the nations in the field of
milk production. Despite massive increase in the production
of milk, illicit production of the synthetic milk has emerged
as the biggest problem to the milk producing industry. As a
result of which, the confidence of the general consumers is
waning from the dairy industry. People are now reluctant to
consume the dairy products. It is of paramount importance
to soon find the solutions to this raging problem, since the
dairy industry occupies an important place in the Indian
rural economy and attached with the industry are crores of
milk producing farmers including the consumers as well

Adulterant
Detergent
Shampoo
Salt
Synthetic Milk
Ammonium Sulphate
Sodium Hydroxide
Sodium Carbonate /
Bicarbonate
Water
Boric Acid
Salicylic Acid /
Benzoic Acid

Limit of Detection
>= 0.5 %
>= 1%
>= 0.5 %
>= 1%
>= 0.5%
>= 0.5%
>= 0.5%
>= 50%
>= 1%
>= 2%

Advantage of Device:

Tried and tested technology from one of the most reliable organization NHHID of Anna University

Technology at an affordable cost.

Anytime anywhere testing of milk without any complex procedures.

Simple and rapid procedure to detect the quality of milk.

Immediate results within seconds.

No additional equipments to be purchased for testing.

No requirement of milk sample for testing

Easy to carry anywhere without any hazard

Hassle free testing

Most of the test can detect an adulteration level at less than 0.5%

Performance Specification of " the Milk Adulteration Device


Milk Sample

Raw Cow, Buf & Mix Milk / Pasteurized Milk can be tested

Adulteration Screening

If Milk is adulterated with any of the Non Milk Components from the ones listed below, the
device will Spontaneously ALERT the user for a Possible Unsafe Milk.

The device is capable to test and screen for 08 Possible Adulterants. If the Percentage of
Adulteration for any of the following listed adulterant EXCEEDS the Limit of Detection, the
device will indicate the same with a glowing RED LED
Adulterant
Detergent
Shampoo
Salt
Synthetic Milk
Ammonium Sulphate
Sodium Hydroxide
Sodium Carbonate / Bicarbonate
Water
Boric Acid
Salicylic Acid / Benzoic Acid

Screening of Adulterants

Limit of Detection
>= 0.5 %
>= 1%
>= 0.5 %
>= 1%
>= 0.5%
>= 0.5%
>= 0.5%
>= 50%
>= 1%
>= 2%

Time for Measurement

Instantaneous Just Dip and Test ( @ 1 Sec )

Warm Up time

Not Required

Sample Preparation

Not Required

Calibration Check

To be done periodically (@ every 15 days) by user using CALIBRATED REFERENCE Solution


provided with the device. Calibration can be checked and validated in less than 05 minutes.

Probe Cleaning

As always with the Milk Testing Instrument, Regular Cleaning of PROBE is an essential
requirement. The user needs to Clean the PROBE with CLEAN TAP WATER after every use
(Milk Collection Cycle) and with CLEANING SOLUTION Provided on Monthly basis.

Warranty

1 Years against manufacturing Defects.

Operating Voltage

230 Volts / 1 Phase

Power Consumption

Less than 5 Watts

Dimensions

150mm x 75mm x 55 mm

Weight

@ 250 gms