Anda di halaman 1dari 11

KONGSBERG

GEOACOUSTICS L T D

Shuttleworth Cise, Gapton Hall [ndustrial Estate,


Great Yarmouth, Norfolk, NR31 ONQ, UK.
Tel
+44 (0)1493 600666
Fax
+44 (0)1493 651100
W eb
www.krn. kongsberg.com / geoacoustics

KONGSBERG

GeoSwath Plus Compact


Harbour Acceptance Test (HAT)
9-GS10-041/A
Project
/Product
Document
Title

GeoSwath Plus Compact

Document
Number
Author

9-GS10-0041/A

Dis tribu ti on

Master: Drawing File

Issue
A

Print Control

(Printed 24-08-2016)

Harbour Acceptance Test (HAT)

Phil Carpenter

Date
14-07-2015

Status
First Issue

Amendment History
Pages Checked
M.Gentilhomme
11

Approved
F .Gutirrez

Client

Amendment Details
AMD

CTD

Details

K O N G S B E R G G E O A C O U S T I C S LTD currenUy

oprate an accreiHtetl Qnality Management

System to US EN ISO 9001.

Harbour Acceptance Test (HAT)

GeoSwath Plus Compact

1. Introduction
This Harbour Acceptance Test (HAT) describes the test methods for the GeoSwath Plus
Compact interferometric sonar system manufactured by Kongsberg GeoAcoustics. It
verifics the system installed is fully functional, and to serve as a record of the successful
completion of the Harbour Acceptance Test.

2. Definitions
GeoSwath Plus is a seafloor imaging and bathymetric system consisting of:
-

A Deck Unit.

Transducers (Port and Starboard).

Software, for data acquisition and processing.

Sensors (sound velocity, motion, navigation and heading).

Abbreviations
GS4
GPS
1PPS
NMEA
RS-232
DB9
SVP
TVG

: Kongsberg GeoAcoustics acquisition/processing software.


: Global Positioning System.
: One Pulse Per Second
: National Maritime Electronics Association.
: Recommended Standard-232 (two hundred and thirty two)
: 9 way D-Type
: Sound Velocity Profiler
: Time Varied Gain

3. Issues relating to tests to be conducted


Tests wi be conducted under the supervisin of a Kongsberg representativo.

9-GSW-004/A

PngelofU

GeoSwath Plus Compact

Harbour Acceptance Test (HAT)

4. List of Items
Please record the details of all main tems installed on the vessel during this test.

Item

Equipment

Serial bumber

GeoSwath Compact Deck Unit

Toughbook/Laptop

U/K

GS4 Dongle

rJ/K

Transducer

Port Transducer Cable

Starboard Transducer Cable

Sensor Interface Cable

Model:

Sound Velocity sensor

Model:U<mc-

MRU

ModeI:ShQ

10

Sound Velocity Profiler

11

GPS

Model:\cwoi{xModelMmsfait

Freq:

kf'z

ZSO

6S05-039?-O

fiM

3 G m Z .

-\0Z
ftvtfSf?
\&\

Otlier:

5. Software
The GS4 software versin should be noted. This information w i l l be checked during the
H A T procedure.

Software Application

Item
1

GS4

GS4Remote

9-GSW-0041/A

Versin

. l . O , 1*1

Page 3 of I

GeoSwati Plus Compact

Harbour Acceptance Test (HAT)

6. HAT Prerequisites
Check that the GeoSwath Plus system is istalled on the vessel and the following
performed:
,

No.
1

Pre-requisite

Performed (S)

Transducer head securely mounted over-side the vessel i n the


correct orientation and cable runs istalled.

Port transducer cable connected to the Port input on the GeoSwath


Plus Compact Deck Unit.

Starboard transducer cable connected to the Starboard input on the


GeoSwath Plus Compact Deck Unit.

Sensor interface cable connected to +24V OUT on the GeoSwath


Compact Deck Unit.

BNC link lead connected between the GPS receiver and 'PPS I N ' on
the GeoSwath Plus Compact Deck Unit.

GPS receiver RS232 output connected to C O M 1 on the GeoSwath


Plus Compact Deck Unit.

M R U RS232 output connected to C O M 2 on the GeoSwath Plus


Compact Deck Unit.

Navigation + heading connected to the aiding inputs on the M R U .

SVS sensor output connected to C O M 5 on the GeoSwath Plus


Compact Deck Unit.

10

+24V AC-DC power supply connected between the vessels A.C.


mains outlet and the Power Input on the GeoSwath Plus Compact.

11

Ethernet cable connected between the GeoSwath Compact and the

i/

i/

i /

^20v

Toughbook.
12

Toughbook power supply connected between the vessels A.C.


mains outlet and the Toughbook.

13

GS4 Dongle plugged into a spare USB port on the Toughbook.

9-GSW-0041/A

fJ/A

Page 4 of 11

GeoSxvath Plus Compact

Harbour Acceptance Test (HAT)


TEST PROTOCOL
No.
1

Test definition and


purpose
GeoSwath Plus Compact
Turn on check.

UltraVNC Remote
Desktop.

Test Steps

Test acceptance/rejection
criteria

Pass

Notes

K)

Switch-on the power to the GeoSwath Plus


Compact.
Press the system ON/OFF button.

The GeoSwath Plus Compact


starts up. The ON/OFF button
LED illuminates when the system
is on.

Turn-on the Toughbook.

The Toughbook Starts-up.

Start UltraVNC Vlewer.


Enter the IP Address of the GeoSwath Compact:
192.168.0.100 then press Connect.
Type n the password: remote

The Desktop of the operating


system running on the GeoSwath
Plus Compact appears.

UA

Click Log On.

GS4 to GS4Remote link


check.

Cise UltraVNC Viewer.

GS4Remote is running. (Note the


versin).
The remote desktop view closes.

On the Toughbook, start GS4.

GS4 starts.

On the Ribbon click

The Workspace Options appear.

Remte .

Check that Remote Acquisition is icked and the


connect status indicator is green.
4

Frequency detect check.


In GS4 open the Log display Log and check the
reported frequency.

9-GSW-0041/A

GS4 is connected to GS4Remote.


The frequency reported in the
GS4 Log display matches that of
the istalled GeoSwath Plus
bathymetry system.

i**

250 Hs
K

PngeSofH

Harbour Acceptance Test (HAT)

GeoSwath. Plus Compact


TEST PROTOCOL

No.

Test definition and


purpose

New Project configuration.

Test acceptance/rejection
' critera

Test Steps
From the File men choose New Project.

The New Project dialogue


appears. Give the project a ame
e.g. GeoSwathHAT and save.

Checks that:
-

A new project can


be created.

Crate a new Equipment file.

A new Equipment File appears in


Survey Manager.

RS232 strings
from all sensors
are being
decoded.

Edit the Equipment file and decode each sensor


string (remember to set the correct Geodetic
parameters when decoding the nav).

Correctly decoded strings from


the sensors are shown in green
scrolling through the GS4
Equipment Editor.

A Vessel file can


be imported,
The MRU is
outputting valid
data.

Cise the equipment editor when finished.

1 Pass

V
1/

In the Survey Manager right-mouse click on


Vessel Settings and choose Import | Import File.

The Vessel File import dialogue


appears.

Choose Generic.vsf, then click Open. Expand


the Vessel Settings tree in Survey Manager.

The Generic Vessel file is shown


in the Survey Manager.

w.

The vessel attitude and heading


are correctly shown in the Vessel
display.

Select the Generic vessel file.

Click
and open the Vessel display mBm.
Check the vessel attitude and heading.
s t a r t

BHE|

Open the Status display

9-GS10-004/A

cae
PBi
status.

The Status display appears.

In the Status display, check the Navigation.

The current vessel position is


given in the Status display.

In the Status display, check the Heading.

The Heading is shown in the


Status display.

In the Status display, check the SVS valu.

The SVS readings are shown in


the Status display.

y
y
y

</

Notes

Harbour Acceptance Test (HAT)

GeoSwath Plus Compact


TEST PROTOCOL

No.
6

Test definition and


purpose
Security Check.
Checks the istalled
software versin and that
the dongle is licensed to
work with t.

Test acceptance/rejection
criteria

Test Steps
Xr-" Security Settings
From the File men click @

Pass

Notes

K)

The Security Settings dialogue


appears.

i/

Check the Dongle Number matches that of the


attached Dongle.

GS4 deteets the dongle and


indcates which features it is
licensed for, e.g. Acquisition and
Processing.

Check who the dongle is licensed to.

GS4 reports the licensed


customer ame.

Record the software versin.

Test Signal Check.


Generates a test signal
which checks the receive
stages of the internal
sonar electrnics are
functioning correctly.

In the GS4 ribbon, click on the Acquisition tab.


Set the side sean Gain to 4, Pulse Length to 1,
Power to 10 and Ping Length to 95m.
O

Acquisition Settings

50
Sidescan Gain
Puise Length

!_
J

Power
Ping Length

j_j
(j

'8^'"

10
55

,1

9-GS10-004/A

GeoSwath Plus Compact

Harbour Acceptance Test (HAT)


TEST PROTOCOL
No.

Test definition and


purpose

Test acceptance/rejection
criteria

Test Steps

Test Signal Check.


Continued

Turn the Test Signal on ^2

then press

<2>
. Open the Depth, Waterfall, Sidescan
Analogue and Sidescan Waterfall displays.
s t a r t

Transmit Check.
To verify that:
-

the GeoSwath
Plus system can
display
bathymetry and
sidescan data.
Port and
Starboard data is
received
correctly.
Noise is of an
acceptable level.

9-GSW-0041/A

Turn off the Test Signal

An inverted V is seen in the


depth display and a ramp
waveform in the Sidescan
Analogue window indicating that
the interna! receive elements of
the sonar electronics are
operating correctly.

Pass

Notes

WKij and turn Transmit

On
. Set the Sidescan Analogue display
TVG Gain levis appropriately to render the
greyscale waterfall.
Check the Depth display.
If possible, look for Port and Starboard features
(e.g. a harbour wall).

A depth profile of the seabed is


displayed in the depth display.
Features are shown on the
correct channel (Port/Starboard).

Zoom out on the depth display and check for


noise.

No strong interference lines are


present in the Depth display.

Check the Waterfall display.

A history of depth profiles are


shown in the Waterfall display.

Check the Sidescan Analogue display.

Amplitude (sidescan) data is


shown in the Analogue display.

Check for noise in the Sidescan Analogue display.

No electrical (or other) noise is


present.

s/

Page8ofU

Harbour Acceptance Test (HAT)

GeoSwath Plus Compact


TEST PROTOCOL

No.

Test definition and


purpose
Transmit Check.

Test acceptance/rejection
criteria

Test Steps
Check the Sidescan Waterfall display.

A history of the sidescan returns


are shown in the Sidescan
Waterfall display.

Continued....

Once the test is complete, press


9

Logging Check.
Verifies that:
-

ThelPPS
interface is
working.

The MRU is
being fully aided.

Data can be
logged.

<>

Press
, then press s . Type in a line
ame (if desired), click 'Apply'.
Check that none of the Status indicators for
decoded strings are red. Check the Logging
status indicator.
STar

The LOG status indicator is


green. The file size increments.
GS+ successfully records data
and indicates the logging and
status from the sensors.
The MRU status indicator is green
(fully aided).

Check the PPS status indicator.

The PPS status indicator is blue


indicating that PPS Sync Pulse
mode is working.

"-3

GS4 enables PPS Sync Pulse


mode.

Check the MRU status indicator.

Notes

o .

On the Ribbon set the PPS Clock Synchronisation


modeto
v
.

Stop Logging
complete.

9-GS10-0041/A

5 t

Pass

and Transmitting

stop

K
/

once

Page9ofU

GeoSwath Plus Compact

Harbour Acceptance Test (HAT)


TEST PROTOCOL
No.
10

Test definition and


purpose
Processing/playback
check.
To verify that the
GeoSwath Bathymetry
system successfully
processes/re-plays the
logged data.

11

SVP Interface check.


To verify that files can be
retrieved from the Sound
Velocity Profiler.

Test Steps

Test acceptance/rejection
criteria

Pass

K)

Select the previously logged Raw Data File then


process the line. (Note: If the vessel is stationary it
may be necessary to mark the Navigation 'Good'
beforehand).

GS4 successfully processes


logged raw data files.

Connect the DB9 on the mini-SVP interface cable


to the COM port on the Toughbook.

The GeoSwath deck unit can


communicate with the mini-SVP
and retrieve files from the internal
memory.

Connect the 10way SubConn connector onto the


mini-SVP.

Notes

Use Datalog Express to communicate to the miniSVP.


Set the pressure tare and also retrieve the files
from the sensors memory.

9-GS10-0041/A

Page 10 of 11

GeoSwath Plus Compact

Harbour Acceptance Test (HAT)

7. Test Report
Test Result
Tested in accordance t o HAT:

GeoSwath Plus Compact

Manufacturen

Kongsberg GeoAcoustics

Customer:

The test items . . 1 . to ...11

of this Harbour Acceptance Test have been performed


Accepted, no non-conformity detected

Test No.

Accepted, with following non-conformities


Non-Conformity

Not Accepted; re-test required

Please use capital letters:


Test performed by

Position

Company
Kongsberg Maritime

Test accepted by

Position

Kema Rcbb \j<?^as


Date

Company

INTRA peo EC

Signature

oto \o&ntp
9-GSW-004/A