ACM600 Series Anesthesia Machines Service Manual

ACM 600 Series Anesthesia Machines


Service Manual

ACM600 Series Anesthesia Machines Service Manual

Chapter 1 ACM600 Series Anesthesia Machines Gas Passage Tank

1.1 Malfunction and its solution:


Malfunction 1: The gas passage tank leaks
Reason: the inner film of the gas-driving switch is aging
Solution: replacing its inner film

Malfunction 2: When opening the gas supply, the gas passage tank always
has gas whistling sound
Reason: there is something wrong with the shift switch
Solution: replacing the shift switch or disconnect it to clean and reinstall

Malfunction 3: After closing the flowmeter, there is still gas inflating the
foldable bellow
Reason: the O2 flush switch cannot be reset
Solution: replace O2 flush switch

Malfunction 4: The floater in the glass tube of flowmeter calibrates


Reason analysis: unidirectional valve is damaged
Solution: replace the unidirectional direction valve

ACM600 Series Anesthesia Machines Service Manual

Chapter 2 ACM600 Series Anesthesia Machines outer gas circuit

2.1 Bellow assembly


Bellow housing
Foldable bellow
Bayonet table

2.2 Circle Absorber

2.3 common malfunction of the gas passage and the solutions


3

ACM600 Series Anesthesia Machines Service Manual

2.3.1 Simple checking methods of the gas passage


First connect the pipelines, set the manual/mechanical shift switch into
the mechanical mode, press the O2 flush switch to inflate the foldable
bellow to its full, then quickly pull out the connecting pipeline from the
circle absorber and block it with hands. If the foldable bellow doesnt go
down, the bellow and the connecting pipelines are ok and the problem
perhaps lies in the circle absorber. Then join the pipelines and connect
the inspiratory connector and the expiratory connector with one
pipeline. Connect the breathing bag, set the manual/mechanical shift
switch into the mechanical mode, press the O2 flush switch to inflate
the foldable bellow, and pinch the breathing bag with hand to make
the pressure gauge on the circle absorber point to 3Kpa and see the
pointers movements. If the pointer goes down quickly, the circle
absorber has some problem. If the pointer doesnt move, then the
circle absorber is all right.

2.3.2 Outer gas passage malfunction and solutions:


Malfunction 1: foldable bellow goes down continually
Reasons: Probably as follows:
1) Foldable bellow is broken or aging;
2) The corrugated hose that connects bellow and circle absorber is
4

ACM600 Series Anesthesia Machines Service Manual

broken or aging;
3) Patients breathing circuit is broken or
aging;
Foldable Bellow
4) Respiratory valve in the circle absorber is broken or aging or
not well-installed;
5) Soda lime absorber of the circle absorber is broken or aging;
Solution: replacing the spare parts that are
broken or aging

Soda Lime Absorber

Malfunction 2: the foldable bellow goes down to certain place and stands
still
Reason: spill over valves pressure has
been adjusted lower
Solution: readjust it
Spill Valve

Note: the spill over valve can be moved with wrench: clockwise
rotation means pressure decrease and counterclockwise rotation
means pressure increase. Its standard setting is 0.2~0.3Kpa. After
adjustment, connect the gas passage, adjust the flowmeter to 5L, and
5

ACM600 Series Anesthesia Machines Service Manual

at this time if the pressure gauge on the circle absorber is 0.2~0.3Kpa,


then it is ok now.

Malfunction 3: While using the machine, the foldable bellow still inflates
or swells, though the flowmeter isnt opened.
Reason: there is something wrong with the four-tube flowmeter or the O2
flush switch
Solutions:

replace

the

four-tube

flowmeter or the O2 flush switch

Malfunction 4: the bellow cannot be


compressed

or

cannot

be

enough

compressed at inspiration;
Reason: 1. The connecting corrugated
hose falls off
2. Anesthesia ventilators problem
3. The connecting line of the
manual/mechanical shift switch is
broken
4. The below housing is broken or aging
Solution: 1. Reconnect the corrugated hose
2. Overhaul the anesthesia ventilator
6

ACM600 Series Anesthesia Machines Service Manual

3. Replace manual/mechanical shift switchs connecting line


4. Replace bellow housing

Malfunction 5: the airway pressure is


too high at manual ventilation mode
Reason: APL valves pressure hasnt
been adjusted correctly
Solution: readjust APL valve
Note: when necessary to adjust, remove the round lid on the top of
APL valve, and screw off the six-angle screw above with six-angle
wrench, and adjust the six-angle screw below with six-angle wrench.
The upper limit is 6Kpa.

Malfunction 6: though the work pressure of the machine is enough, the


foldable bellow cannot be inflated when pressing the O2 flush switch.
Reason: the pipelines inside the gas passage tank fall off or are broken.
Solution: reinstall or replace pipelines

Malfunction 7: LED display monitors show value has big error


Reasons: Flow sensors wheel has dirt

ACM600 Series Anesthesia Machines Service Manual

Solutions: Clean the flow sensor

Malfunction 8: At the time of O2 flush, gas still goes through the


mechanical mode passageway when shifting to manual mode; vice versa.
Reasons: the inner film of the switch is aging or frayed
Solution: replace its inner film

Malfunction 9: From manual to


mechanical

mode,

the

shift

switchs

inner

switch returns slowly


Reasons:

the

spring has lost its elasticity


Solutions: replace the switch
valve core

ACM600 Series Anesthesia Machines Service Manual

Chapter 3 Flow meter


3.1 Four-tube flow meter
3.1.1

ACM603 ACM603B

ACN606ACM607BACM606B
ACM607 ACM608B

all

use

four-tube flowmeter. (See the picture


on the right)
3.1.2 slim tube: 0.1L/min2L/min
thick tube: 2.5L/min10L/min
3.1.3 four tubes are of O2 and N2O
Compensation adjustment

3.2 Five-tube flowmeter


3.2.1 ACM618B anesthesia machine uses five-tube flowmeter (see the
picture below)
3.2.2 N2O: 0L/min10L/min
O2: 0L/min10L/min
Air: 0L/min10L/min
Base

flow:

130L/min

190L/min
9

ACM600 Series Anesthesia Machines Service Manual

3.3 Flowmeters common malfunctions and their solutions


Malfunction 1: O2/N2O flow linkage ratio is invalid
Reasons: Gear loosens
Solutions: Refasten the gear

Malfunction 2: after closing the knob tightly, the floater cannot return to
the bottom position
Reason: The adjustment knob is frayed and cannot be sealed
Solution: Replace the flowmeter

10

ACM600 Series Anesthesia Machines Service Manual

Chapter 4 Vaporizer
4.1 ACM600 series anesthesia machines equipped vaporizers
4.1.1 ACM102 vaporizer (see the picture on the right)
ACM102 vaporizers dial has a
symbol, which is the knob place to
add anesthetic agent.
1) Concentration adjusting range:
0.55.0VOL
2) Possessing the function of flow
and temperature compensation

4.1.2 ACM105/ACM108 Vaporizer


1) Concentration adjusting range: 0.25.0VOL
2) Possessing the function of flow and temperature compensation

ACM 105 Vaporizer

ACM108 Vaporizer

11

ACM600 Series Anesthesia Machines Service Manual

4.1.3 DRAGER Vaporizer


1) Concentration adjusting range:
0.25.0VOL
2) Possessing the function of flow and
temperature compensation
3)

Temperature

compensation:

15C35C
4) Perfusion between the maximum and minimum graduation
5) It neednt be emptied on the process of transportation

4.2 Vaporizers common malfunctions and their solutions


Malfunction 1: vaporizer leaks
Reasons: 1) Bypass valves seal ring is aging
2) The agent discharge outlet hasnt been screwed down
3) Part of the inject bases seal ring is aging
Solutions: replace the aging seal ring or screw down the agent discharge
outlet

Malfunction 2: dials knob cannot be easily twisted


Reasons: It hasnt been installed well or the spare parts are not of good
quality
Solutions: Reinstall or replace the spare parts
12

ACM600 Series Anesthesia Machines Service Manual

Malfunction 3: the output concentration is not accurate (cannot


anesthetize patients)
Reasons: The inner double-sheet metal or other parts have become
distorted or aging
Solutions: Need to re-regulate or repair (send back to the our company to
re-regulate or repair)

Notice: vaporizer is a kind of precision instruments, needing to be


regulated with high-precision instrument. It is usually hard to mend
it on the spot. If with serious problem, it should be sent back to the
company to be dealt with.

13

ACM600 Series Anesthesia Machines Service Manual

Chapter 5 Anesthesia ventilator


5.1 Summary
5.1.1 The anesthesia ventilators that are equipped on the anesthesia
machines of various types
603

603-1

606

607

60

603B

606B

608B

618B

803

812A

812A 813A 805A

8
802A

802A

802

814A

5.1.2 Misoperation and solutions of other simple malfunctions


Phenomenon
Digital
doesnt

Possible reasons

Solutions

pipeline 1 Power cable disconnect


flash, 2 Power off

1 Connect the cable


2 Turn on the power

ventilator doesnt 3 Fuse is blown out

supply

work

3 Replace the fuse

The
suddenly
working

ventilator
stops

The

power

supply

Manual ventilation

during fails

operation,

the

indicator light is off


with sound alarm
1 Patients pipeline is 1 Check the patients
14

ACM600 Series Anesthesia Machines Service Manual

blocked
Upper

airway 2

pressure limit

pipeline and adjust it

Patients

airway

is 2

blocked

Check

patients

condition

3 Upper airway pressure 3 Reset the alarm limit


limit is set too low

4 Recalculate ventilation

4 Ventilation parameters parameters


change
Airway

pressure 1

gauge

Connecting

between

hoses 1 Reconnect the pipe

inspiration

passage and flow sensor


gets loose or is connected 2 Replace gas supply
reversely
2. Gas supply is out
1 Flow sensors plug gets 1. Re-insert
Tidal

volume

display is too low

loose

2. Connect the pipelines

2 Circuit leaks

properly

3 Expiration valve leaks

3.

Replace

expiration

valve
Probe
5.2 The common malfunctions and Solutions about the 802A
804A

812A
813A Anesthesia Ventilators
Malfunction 1: There is no tidal volume
15

ACM600 Series Anesthesia Machines Service Manual

display on the LED window


Reason: The flow sensors probe or the main board is damaged.
Solution: Replace flow sensors probe or main board.

Malfunction 2: The board parameters do not appear when opening the


machine.
Reason: The fuse is burned out or power
module is damaged.
Solution: Replace the fuse or power
module.
Advert: Check the power module whether
it is damaged, multimeter can be used to
check whether the output voltage is
correct.

Malfunction 3: When turn on the machine, it is easy to hang (each


parameter can not be adjusted, and the LED digital tube and indicator
light are all bright);
Reason: There is something wrong with the main board.
Solution: Replace the main board.

Malfunction 4: The anesthesia ventilator can work normally, but the


16

ACM600 Series Anesthesia Machines Service Manual

parameters cant be adjusted.


Reason: the connecting lines of the fore board become loose or aging or
damaged.
Solution: re-insert the surface line or replace
the surface film.

Malfunction 5: When the alarm appears,

804 Fore Board

there is only light but no alarm sound.


Reason: Scale the voltage of the buzzer. If there is voltage, then the
buzzer is damaged; if without voltage, then there is something wrong
with the main board.

Malfunction 6: The tidal volume cant be adjusted or cant be closed


tightly.
Reason: The tidal volume knob falls off.
Solution: Retighten the knob.

Malfunction 7: There is loud noise during


the machines working.
Reason: The expiration valve rubber sheet is
aging or distorted.
Solution: Replace the expiration valve
17

ACM600 Series Anesthesia Machines Service Manual

rubber sheet.

Malfunction 8: There is no driven gas output in the anesthesia ventilator.


Reason: We can use multimeter to check the electromagnetic valve, then
we will know whether the output voltage is normal; If the
electromagnetic valve is normal, then there must be something wrong
with the main board.
Solution: Replace the electromagnetic valve or main board.

Malfunction 9: It appears black on LCD


Screen (only ACM813A).
Reason: Inverter is not connected tightly or
damaged.
Solution: Re-insert the line or replace inverter.

Malfunction 10: There is nothing appears on the screen or no signal (only


ACM813A).
Reason: The signal wire of display control
board looses or the display control board is
damaged.
Solution: Retighten the signal wire or
18

ACM600 Series Anesthesia Machines Service Manual

replace display control board.


5.3 The parameters of Anesthesia Ventilators should be adjusted to
first choose, then adjust.
5.3.1 The adjustment of panel parameters.
1) Irradiance line: It displays the change of airway pressure, the red light
on the right and left means low voltage with high-pressure alarm.
Low pressure alarm: Avoid breath circuits leakage.
High-pressure alarm: It can protect patents lung, and avoid the hurt
from high pressure.
2) The adjustments of I/E ratio and respiration rate.
Press the I/E ratio or respiration rate key on the panel, adjust by the
two symbols of.
3) Tidal volume: Adjust with the flow knob, and calculate by the weight
of patients weight (kg)*(8-12).
* 4) Oxygen concentration: It can only appear but cant be adjusted.
5) Silence for 2 minutes: silence time for alarm.
6) There is an air entrance on the back panel of the ventilator. Using the
Venturi theory, it inhales air through oxygen pressure, thus save the
oxygen.

5.3.2 The adjustment of Ventilators inner parameters


When the ventilators malfunctions are caused by circuit board, we can
19

ACM600 Series Anesthesia Machines Service Manual

solve these problems by replacing it. Then we should regulate and adjust
parameters of the new circuit board to ensure the status of Ventilator.
The corresponding parameters of ACM 813A respective rheostat
6201372 control board: *R3 adjusts the increase of airway pressure, i.e.
the adjustment of Ppeak.
*R4 is used to adjust the zero point of airway pressure
*R35 is used to adjust the oxygen supply failure alarm
S1: SESR, the segments 1, 2, 3 adjust the modulus of tidal volume.
Segment 4 adjusts the airway pressure waveform when it is on, and
adjusts its close when it is off.
Notice: Above with#sign shouldnt be adjust randomly, and above
with *sign need special instrument.