Manual
Model FU20
pH/ORP Combination sensor
IM 12B6J3-02E-E
2nd edition
TABLE OF CONTENTS
1. General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1-1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1-2. Unpacking and Checking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1-3. Warranty and Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
2. General Specifications FU20 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
2-1. Measuring elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
2-2. Construction materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
2-3. Functional specifications (at 25C) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
2-4. Dynamic specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
2-5. Operating range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
3. Installation Site FU20 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
3-1. Typical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
3-2. Preparing the Sensor for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
3-3. Mounting the Sensor into the Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
4. Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
4-1. Conventional pH (& ORP) Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
4-2. Wiring for ORP measurement with normal reference . . . . . . . . . . . . . . . . . . . . . . . . .7
4-3. Wiring for ORP measurement with pH reference . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
5. General Calibration & Maintenance Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
5-1. Calibration for pH measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
5-2. Calibration for ORP measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
5-3. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
6. Model Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
7. Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
IM 12B6J3-02E-E
1. GENERAL
1-1. Introduction
This instruction manual provides
information for the installation and use of
the FU20, four-in-one wide body pH
sensors. The FU20 is the choice for the
majority of typical wastewater and process
applications.
The Model FU20 offers a simple and cost
effective solution for a wide variety of
wastewater and process applications. This
all-in-one sensor provides simultaneous
measurement of pH, redox (ORP) and
temperature if desired. The rugged Ryton
body is designed for easy installation into
on-line and immersion applications via the
3/4" NPT threaded connections provided
on both ends of the sensor. Optional
quick-removal adapters in both stainless
steel and titanium are available to make
calibration and maintenance even easier.
The FU20 include an integral cable of
optional lengths designed to minimize the
adverse affects of noise can have on the
measurement. This cable is pre-finished to
make connection to any industrial pH
instrument as simple as possible. The
FU20 can also include a multiple VP
connector and then requires seperate
cable type WU10-V-S that somes in
variable lengths.
1-2. Unpacking and Checking
Upon delivery, unpack the sensor carefully
and inspect it to ensure it was not
damaged during shipment. If damage is
found, retain the original packing materials
(including outer box) and then immediately
notify the carrier and the relevant
Yokogawa sales office. Make sure the
Model Code on the sensor is the same as
on the packing list. Also, check any
option(s) that were ordered are included
and correct. The sensor has an
identification number on the cable with the
full Model Code and a Serial Number
included.
:
:
:
:
pH glass electrode
Silver chloride reference
Solid platinum electrode
Pt1000 temperature sensor.
:
:
:
:
:
:
:
:
:
:
0 to 14
-10 to 105 C (14 to 221 F)
0 to 10 bar ( 0 to 145 PSIG)
> 50 S/cm
-30 to 50 C (-22 to 122 F)
NOTE: Specifications should not be considered in isolation. For example the pH range can
be 2-12 pH, where the measurement is at elevated temperatures. For advice about
specific applications please contact you local sales office.
IM 12B6J3-02E-E
3. INSTALLATION OF FU20
For optimum measurement results, the
FU20 should be installed in a location that
offers an acceptable representation of the
process composition and DOES NOT
exceed the specifications of the sensor.
The FU20 is designed with threads on
either end of the body 3/4" NPT to allow
installation in a wide variety of applications.
3-1. Typical Installation
The FU20 sensor is designed for in-line
installation, via a bypass loop as an
immersion assembly. For best results the
FU20 should be mounted with the process
coming towards the sensor (Fig 7) and
positioned at least 15 above the horizontal
plane to eliminate air bubbles in the glass
bulb.
FU20
TeflonTape
IMPORTANT NOTICE
Before mounting the sensor, it should be
calibrated. The calibration procedure is
normally described in the Instruction
Manual of the pH transmitter, but a general
calibration is given in section 7 of this
manual.
3.3 Mounting the Sensor
The FU20 can be mounted using the
threads on the body of the sensor (Fig. 6).
Apply Teflon tape to the appropriate
threaded end, then install in the process.
FU20
Process Flow
Wrench Flats
Do NOT
overtighten
Teflon Tape
Teflon Tape
FU20 + FD40
FU20 + ADAPTER
/SN3
/SR3
/FN4
/FR4
Fitting
Process
Flow
Adapter
( /FPS )
Cleaning
Solution
(water, acid, etc.)
FU20
O-ring
6
3/4" TAPERED THREAD
26.0 (1.02")
-NPT
19.7(0.78")
19.7 (0.78")
22.5 (0.89")
1 NPT or ISO 7/1-R1
51.0 (2.3")
150.0 (5.9")
22.0 (0.86)
-FSM
10.0 (0.39")
25.0 (1")
99.0 (3.9")
1 NPT or
ISO 7/1-R1
39.0 (1.54")
5.0 (0.2")
35.0 (1.38")
30.0 (1.18")
30.0 (1.18")
36.0 (1.42")
Figure 11. Dimensions FU20 adapter for FF40, FS40 and FD40 fittings
IM 12B6J3-02E-E
4. WIRING
(Temp)
F
Red (11)
(Temp)
E
Blue (12)
C
6.5
FU20-VP-
(pH)
A
Core coax (15)
(Ref.)
B
Shield coax (13)
VP-connector
A
(pH)
B
(Ref.)
C
(Temp)
F
FU20-VP-
(pH)
A
(Ref.)
B
FU 20-VP-
(ORP/LE)
D
Black
(Temp)
(ORP/LE) E
D
F (Temp)
E (Temp)
D
(ORP liquid Earth)
WU10 cable
VP connector
color
A
Clear
B
Brown
C
D
Black
E
Red
F
Blue
Yellow
8
4.2.1. Settings for PH402G and PH202G(S)
Function
pH
pH&rH
Setting
(default) pH&ORP
code 01
0(pH)
0(pH)
0(pH)
code 02
0(off)
1(ORP)
2(rH)
code 03
1.1.1
1.1.1
1.1.1
code 04
0.0.1
0.0.1
0.0.1
code 31
0
2
2
jumper input 1
no
no
no
jumper input 2
yes
yes
yes
4.2.2. Settings for PH150 and PH450G
Function
pH
Setting
(default) pH&ORP
sensor setup
pH
pH+ORP
measurement
pH+ORP
setup
Impedance
setting
input1
High
High
input 2
Low
Low
jumper input 1
no
no
jumper input 2
yes
yes
1(ORP)
0(off)
0.0.1
0.0.1
pH compensated
ORP
1(ORP)
0(off)
0.0.1
1.1.1
yes
yes
yes
no
ORP
pH&rH
ORP
pH+ORP
ORP
pH compensated
ORP
ORP
Low
Low
yes
yes
Low
High
yes
no
pH+rH
High
Low
no
yes
IM 12B6J3-02E-E
9
The FU20 has a reference system of
Saturated Silver/Silver Chloride and the
values are given in the following table as
function of the measurement function ( 4.2)
5-3. Maintenance
All pH sensors require routine maintenance
to keep the measuring elements clean and
functioning. Depending on the process,
different cleaning solutions may be required.
EXAMPLES:
1. Deposits of limes, hydroxides or
carbonates can be removed by immersing
the sensor in a solution containing dilute
hydrochloric acid (5% is recommended).
Afterwards rinse with water.
6. MODEL CODES
Model
Code
FU20
Cable length
Temp. element
Model
Options
Suffix
code
Option
Description
/HCNF
/FPS
/NSS
/NTI
/BSP
/BTI
-03
-05
-10
-20
-VP
-T1
-NPT
-FSM
IM 12B6J3-02E-E
7. SPARE PARTS
Spare part
Description
FU20
K1523DD
K1547PK
K1547PL
K1547PM
K1547PN
K1500 FR
K1500FS
K1500FT
YOKOGAWA HEADQUARTERS
9-32, Nakacho 2-chome,
Musashinoshi
Tokyo 180
Japan
Tel. (81)-422-52-5535
Fax (81)-422-55-1202
www.yokogawa.com
Buffer
Buffer
Buffer
Buffer
powder
powder
powder
powder
pH
pH
pH
pH
1.68
4.01
9.18
6.87
;
;
;
;
IEC746-2
IEC746-2
IEC746-2
IEC746-2
Yokogawa has an
extensive sales and
distribution network.
Please refer to the
European website
(www.yokogawa.com/eu)
YOKOGAWA
IM 12B6J3-02E-E
Subject to change without notice
Copyright