Bedside Monitor
PVM-2701/2703
This technical data may be revised or replaced by Nihon Kohden at any time without notice.
TD.PVM2700_D
Specifications
Measuring Parameters
PVM-2701: ECG, respiration in impedance, SpO2, NIBP, temperature
PVM-2703: ECG, respiration in impedance, SpO2, NIBP, IBP, temperature, CO2
Influence on Measuring Accuracy by Electrosurgery/Defibrillation/Electrostatic Discharge
The bedside monitor returns to the previous operating mode within 10 seconds without loss of any stored data. When
performing defibrillation, the filter setting on the bedside monitor must be set to MONITOR on the ECG window to return to
the previous operating mode within 10 seconds without loss of any stored data. Measurement accuracy may be temporarily
decreased while performing electro-surgery or defibrillation. This does not affect patient or equipment safety.
Display
Display size:
Resolution:
Viewing area:
Waveform display:
PVM-2701:
ECG (maximum 2 traces), respiration, SpO2
PVM-2703:
ECG (maximum 2 traces), respiration, SpO2, IBP, CO2 pulse wave
Waveform display mode:
Non-fade fixed
Maximum number of waveform trace: 5 traces
Sweep speed:
6.25, 12.5, 25 or 50 mm/s
Respiration sweep speed:
1.56, 6.25, 12.5 or 25 mm/s
Aspect ratio (ECG display sensitivity ratio to sweep speed):
Standard:
0.4 s/mV
Setting range:
0.05 to 6.4 s/mV
Sweep time (at 25 mm/s sweep speed): 4.8 s
Display delay time:
DIAG and MONITOR mode:
250 ms
MAXIMUM mode:
1s
Waveform display color:
12 colors
Numeric data display:
PVM-2701:
Heart rate, VPC rate (per minute), ST level, respiration rate, NIBP (systolic,
diastolic, MAP), SpO2, pulse rate, temperature
PVM-2703:
Heart rate, VPC rate (per minute), ST level, respiration rate, NIBP (systolic,
diastolic, MAP), IBP (systolic, diastolic, mean), SpO2, pulse rate, temperature,
ETCO2, FiCO2
Synchronization mark:
Heart rate sync mark, pulse rate sync mark, respiratory sync mark
Numeric display color:
12 colors
Recovery time after defibrillation:
Less than 10 s (at MONITOR mode)
Alarm
Alarm classification:
Alarm items:
2/18
Alarm indication*:
Alarm indicator indication:
Crisis:
Warning:
Advisory:
Alarm sound:
Crisis:
Warning:
Advisory:
Alarm silence:
Alarm suspend:
All alarms off:
Alarm volume
3/18
I, II, III
I, II, III, aVR, aVL, aVF, 2 from V1 to V6
ECG input protected against 400 Ws/DC 5 kV
IEC 60601-2-27:2005 17h.101 complied
500 mV
5 mV
30 Vp-p (Referred to input)
maximum 40 dB
1.23 Vrms
95 dB
0.05 to 60 Hz (3 dB)
0.3 to 40 Hz (3 dB)
1 to 18 Hz (3 dB)
NOTE: When performing defibrillation, set <FILTERS> to MONITOR or MAXIMUM. The waveform recovery may
become slow due to electrode polarization when DIAG is set.
Input impedance:
5 M (at 10 Hz)
2.5 M (at 0.67 to 40 Hz)
ESU protection:
Provided
Recovers within 10 seconds after ESU and acquired data is not lost.
IEC 60601-2-27: 2005 complied
Leads-off sensing:
Each leads has own sensing
Active electrode:
< 100 nA (DC)
Reference electrode:
< 500 nA (DC)
Waveform display:
Display sensitivity:
10 mm/mV 5% (at DIAG mode, 10 Hz, at 1 sensitivity)
Number of channels:
2 (maximum, with 6 electrodes on home screen)
Sensitivity control:
1/4, 1/2, 1, 2, 4, or AUTO
Pacing mark display:
Available
Recording sensitivity (when using WS-201P recorder module):
10 mm/mV 5% (same as the display sensitivity)
Heart rate count:
Calculation method:
Moving average/Instantaneous beat to beat
QRS detection (at 1 sensitivity): Adult:
Width: 70 to 120 ms
Range of detection amplitude: 0.5 to 5 mV, rate: 30 to 200 beats/min
Child and neonate: Width: 40 to 120 ms
Range of detection amplitude: 0.5 to 5 mV, rate: 30 to 250 beats/min
Complies with the QRS detection range specified in IEC 60601-2-27: 2005
50.102.15
Counting range:
0, 15 to 300 beats/min
Counting accuracy*:
2 beats/min (0, 15 to 300 beats/min)
* Essential performance in EMC standard
4/18
6/18
A pulse oximeter tester that generates simulated signals can be used to check the difference from the design
specification, but it cannot be used as a replacement for human signals for testing accuracy.
Alarm:
30 to 300 beats/min
30 to 300 beats/min
3% 1 beat/min
When <SYNC SOURCE> is set to SpO2:
31 to 300 beats/min in 1 beat/min steps, OFF
When <SYNC SOURCE> is set to SpO2:
OFF, 30 to 299 beats/min in 1 beat/min steps
Selectable from SLOW, NORMAL and FAST.
The following graphs show the response time example when SpO2 changes 0.6%/s.
Pulse Rate = 70 bpm
SpO2 Reference
SpO2 Fast
SpO2 Normal
SpO2 Slow
100
95
90
SpO2 (%)
Pulse rate:
Display range:
Declared range:
Counting accuracy (rms):
Alarm:
Upper limit range:
Lower limit range:
Response time:
85
80
75
70
65
60
0
10
20
30
40
50
60
70
80
90
100
110
120
130
140
Time (seconds)
SpO2 Fast
SpO2 Normal
SpO2 Slow
100
95
90
SpO2 (%)
85
80
75
70
65
60
0
10
20
30
40
50
60
70
80
Time (seconds)
7/18
90
100
110
120
130
140
The following graph shows the response time example when pulse rate changes 10
bpm/s.
SpO2 = 97
PR Reference
PR
150
140
PR (bpm)
130
120
110
100
90
80
70
60
0
10
20
30
40
50
60
70
80
90
100
110
120
Time (seconds)
8/18
9/18
Alarm
Upper limit range:
Lower limit range:
10/18
8 dots/mm
40 dots/mm ( 25 mm/s)
20 dots/mm (50 mm/s)
FQW50-2-100
ZS-900PG/PK Transmitter
Frequency capacity deviation:
Transmission power:
Spurious emission strength:
Occupied bandwidth:
Adjacent channel leaking power:
Transmission frequency range:
Modulation method:
11/18
Temperature:
Humidity:
Atmospheric pressure:
Power Requirement
Line voltage:
AC:
DC (SB-201P):
Line frequency:
Power consumption:
Clock Accuracy
At operating temperature 25C:
At storage temperature 20 to +60C:
Environment
Operating environment:
Temperature:
Humidity:
Atmospheric pressure:
Transport and storage environment:
Temperature:
Humidity:
Atmospheric pressure:
Mechanical Strength
Mechanical strength:
Electromagnetic Compatibility
Safety Standard
Safety standard:
IEC 60601-2-34: 2000 - Particular requirements for the safety of direct blood
pressure monitoring equipment*2*4
IEC 60601-2-49: 2001 - Particular requirements for the safety of multifunction
patient monitoring equipment
EN 12470-4: 2000*3
ISO 9919: 2005
ISO 21647: 2004*4
*1 Not including interbed alarm.
*2 This monitor complies with IEC 60601-2-34: 2000 except for clauses 44.6,
45.101 a) and 45.101 b).
*3 This monitor complies with EN 12470-4: 2000 only for clauses 6.2, 6.3 a), 6.5,
6.6, 6.7, 6.8, 6.9, 6.10 and 8.
4
* PVM-2703 only.
Type of protection against electrical shock:
CLASS I EQUIPMENT (AC Powered)
Internally Powered EQUIPMENT (BATTERY Powered)
Degree of protection against electrical shock
Defibrillator-proof type CF applied part:
ECG, Respiration (Impedance), IBP, Temperature, SpO2, CO2, NIBP
Degree of protection against harmful ingress of water:
IPX0 (non-protected)
Degree of safety of application in the presence of FLAMMABLE ANAESTHETIC MIXTURE WITH AIR, OR WITH
OXYGEN OR NITROUS OXIDE:
Equipment not suitable for use in the presence of FLAMMABLE
ANAESTHETIC MIXTURE WITH AIR, OR WITH OXYGEN OR NITROUS
OXIDE
Mode of operation:
CONTINUOUS OPERATION
Dimensions and Weight (approximate)
PVM-2701/2703 bedside monitor
Dimension:
Weight:
WS-201P recorder module
Dimension:
Weight:
QI-201P interface
Dimension:
Weight:
QI-202P interface
Dimension:
Weight:
13/18
Electromagnetic Emissions
The PVM-2701/2703s essential performances in EMC standard satisfy the following criteria.
This Model PVM-2701/2703 is intended for use in the electromagnetic environment specified below.
The customer or the user of the PVM-2701/2703 should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
Compliance
Group 1
RF emissions
CISPR 11
Harmonic current
emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Class B*1
Class A
Complies
14/18
Electromagnetic Immunity
The PVM-2701/2703s essential performances in EMC standard satisfy the following criteria.
This Model PVM-2701/2703 is intended for use in the electromagnetic environment specified below.
The customer or the user of the PVM-2701/2703 should assure that it is used in such an environment.
Immunity test
Compliance level
Electrostatic discharge
(ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Surge
IEC 61000-4-5
1 kV differential mode
2 kV common mode
Electromagnetic environment
- guidance
Floors should be wood, concrete or
ceramic tiles. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.*2 *3
*1
*2
*3
15/18
Immunity test
Compliance level
Conducted RF IEC
61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
d = 1.2 P
(1 Vrms*3)
(d = 3.5 P *3 *4)
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
1
* Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the PVM-2701/2703 is used exceeds the applicable
RF compliance level above, the PVM-2701/2703 should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the PVM2701/2703.
*2 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
*3 For impedance respiration.
*4 Impedance respiration measurement is very sensitive and affected by electromagnetic interference.
An electromagnetic environment that exceeds the compliance value may cause an incorrect measurement result. Do
not use any device that generates electromagnetic waves near the system during impedance respiration measurement.
16/18
0.01
0.12 (0.35*)
0.12
0.23
0.1
0.38 (1.1*)
0.38
0.73
1
1.2 (3.5*)
1.2
2.3
10
3.8 (11*)
3.8
7.3
100
12 (35*)
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
* For impedance respiration
System Composition for EMC Test
The PVM-2701/2703 bedside monitor is tested to comply with IEC 60601-1-2: 2001, Amendment 1: 2004 with the following
composition. If any part which is not specified by Nihon Kohden is used, the EMC specifications might not comply.
Units
PVM-2701/2703 bedside monitor
BR-903P electrode lead
JC-906P ECG connection cable
JL-900P SpO2 connection cord
TL-201T finger probe
YN-921P air hose for neonate
YP-821P disposable cuff for neonate
401J thermistor probe for adult
TG-950P CO2 sensor kit
JP-900P IBP connection cord
DX-300 monitoring kit
SB-201P battery pack
QI-202P interface
QW-100Y (HIT-100) hyper isolation transformer
YS-100P7 network connection cable
Network cable
Power cord W (ND01-AC2500-JPNK)
Power cord N (ND01-AC2500-EUNK)
Grounding lead
WS-201P recorder module
YS-100P6 serial cable
Cable length
0.8 m
3.0 m
2.5 m
1.6 m
3.5 m
0.2 m
3.5 m
4.0 m
3.5 m
0.2 m
10 m
2.5 m
2.5 m
5m
2.5 m
Standard Accessories
Name
Operators Manual
PVM-2701/2703 Manuals CD-ROM
Qty
1
1
Code No.
0614-904413D
0644-900076E
17/18
1b Items A and B in
PATIENT
ENVIRONMENT
1c Item A powered
from specified power
supply in item B in
PATIENT
ENVIRONMENT
2 2a Item A in
PATIENT
ENVIRONMENT and
item B in medically
used room
2b Item A in
PATIENT
ENVIRONMENT and
item B in medically
used room
3 3a Item A in
PATIENT
ENVIRONMENT and
item B in non-medically
used room
IEC
60601
IEC
60601
IEC
60601
IEC
XXXXX
Outside the
PATIENT
ENVIRONMENT
Non-medically used
room
For B: Additional
protective earth or
separating transformer
For B: Additional
protective earth or
separating transformer
A
IEC 60601
B
IEC XXXXX
A
IEC
60601
IEC
60601
IEC
60601
IEC
XXXXX
B
IEC 60601
or
IEC XXXXX
IEC
60601
3b Item A in
PATIENT
ENVIRONMENT and
item B in non-medically
used room
B
IEC 60601
or
IEC XXXXX
IEC
60601
For B: Additional
protective earth or
SEPARATION DEVICE
protective earth
KEY TO TABLE
Additional protective earth: If necessary, provide additional protective earthing, which is permanently connected (see also
58.201).
NOTE: Equipment modification may be required.
Separating transformer: If necessary, limit the ENCLOSURE LEAKAGE CURRENT, by using an additional separating
transformer according to annex EEE.
NOTE 1: No equipment modification is required.
NOTE 2: A separating transformer is a transformer with one or more input winding(s) separated from the output winding(s)
by at least basic insulation [IEC 60989].
SEPARATION DEVICE: If necessary, apply SEPARATION DEVICE.
IEC 60601: MEDICAL ELECTRICAL EQUIPMENT in compliance with IEC 60601.
IEC XXXXX: Non-medical equipment in compliance with relevant IEC safety standards.
18/18