Anda di halaman 1dari 125

No.

2J313-002E"P

TOSHIBA

PERFORMANCE Ct-tECKMANUAL
FOR
DIAGNOSTIC X-RAY GENERATOR
MODELS KXO-3OR, KXO-30F AND KXO-50F
(23313-OOZE*P>

TQSHIRA

CORPORATION

%? TOSHIBA
CORPORATION
1986
ALLRIGHTS
RESERVED

No.

2J313-002E*P

REVISION RECORD
REV.

DATE!

REASON

PAGE
/AUTHOR

CHANGED

No.

DOC
PRODUCI

--__-_-

INI.

'86
Apr. 18

NC

Mr. Wakai

:A

:B

:C

:D

FE

:F

'G

*H

.I

'J

:K

FL

FM

:N

:O

'87
Sep. 11

Correction

'88
Feb. 20

Correction

'88
May 11

Correction

'88
Dec. 23

Correction

'89
May 26

Correction

'89
Nov. 27

Correction

'90
Mar. 22

Correction

'90
May 21

Correction

'90
Jul. 20

Correction

'90
Sep. 20

Correction

'90
Nov. 30

Correction

'91
Oct. 15

Correction

'92
Jan.

Correction

03/'92

Correction

Mr. Wakai

Mr. Ishiyama

Mr. Kimura

Mr. Ishiyama

Mr. Kimura

Mr. Kimura

Mr. Kimura

Mr. Mochizuki

Mr. Ishiyama

Mr. Ishiyama

Mr. Kimura

Mr. Kimura

Mr. Kimura
NC
Mr. Ishiyama

.P

111'93

P.25, 52

Correction
Mr. Ishiyama

R-l

No,

25313-002E*J

CONTENTS

Page
1.

INTRODUCTION

2.

PREPARATION

3.

______________-_-__---~-------~~~-~--~--~~-------~

______________________________I_________-~~-~---~--

2.1

Test Tools

2.2

Precautions

2.3

Radiation

BEFORE

TURNING

3.1

Visual

3.2

Power

4.

SYSTEM

5.

PERFORMANCE
5.1

5.2

Instruments

---_-_-----__----_----

Required

____-________-___-__~-~----~~----~--~_~___~___

_____________-______-------------___-----

POWER

ON -------_--_------------------a--_--a--_

-----_--_-_-----_----------------~~-~~-~

Source

_--_---_--_------__----~-~~~------~-_-~--_~-_

_---__--___--_--_-_----~~---~~~--~-~--~-~-~-----~

CHECK
Test

-_-_-_---_____-_--__-_~-~-~_--~-~--~____-__-_

___-__-_____-_--____-~-~~~-~-~~~-----~~~--~~~~~~

5.1.1

Power

5.1.2

Checking

voltage

5.1.3

Checking

operations

5.1.4

Checking

the switching

5.1.5

Checking

the starter

circuit

5.1.6

Checking

the battery

_---_-_-_____--____-----------

Radiography

Safety

Inspection

SETTING

Unit

and

ON -----_---__-_-----_------------------

Adjustment

-----

-------__------------~~----~-----and display

circuits

----------

8
10

--------------------

11

----------------------

11

circuit

-_------_-_---___-----__-_--_------

Radiographic

tube current

adjustment

--------------

15

5.2.3

Radiographic

tube voltage

adjustment

--------------

24B

5.2.4

Checking

Automatic

5.4

Fluoroscopy

Exposure

Adjustment

5.4.1

Measuring

5.4.2

Fluoroscopic

-------------------

29

----------------------

30

______-__------------

30

-------------------

30

timer

Control

(KXO-30F/SOF)

instrument
tube

-a-

connection

current

-_-_-

13

5.2.2

the radiographic

_---_______---

13

Measuring

Checking

connection

12

5.2.1

5.3

instrument

adjustment

--------------

30

No.

2J313-002EB

CONTENTS

Page

5.4.3
5.5

ic

Fluoroscop

HHS Tests

tube

voltage

adjl .is tment

31

--_--_-_---___

32

____________________-______-_-_--_~--_--_~-~_~~~

5.5.1

Fluoroscopic

5.5.2

Reproducibility

5.5.3

Automatic

exposure

control

limit

5.5.4

Automatic

exposure

minimum

exposure

entrance
and

exposure
linearity

rate

32

---------------

32

--___________________

32

__________________
time

32

----------

6.

SYSTEM ACCURACY ___--_---__--__-____--_-~-_______-~~--~------~-

33

7.

ATTACHED DATA __-__-------_-_-___-___-~_--~_-----~-~-~~-~~~----

33

7.1

PWB Setting

7.2

Error

Code

and
Table

Adjustment

Tables

_________-__---_-__-----

--_______________________________-__--_

8.

MAINTENANCE SCHEDULE __-___-________-_-_----_----_-e-----------w

9.

APPENDICES

-__-____-___-___-_____________-___~__--__----~-~

-b-

33
---

70

71
---_

71

No. 2J313-002E*A

1.

INTRODUCTION
This manual applies to the diagnostic X-Ray Generators,
models KXO-30R,
KXO-30F, and KXO-50F, and describes the system setting, the electrical
and test to ensure the proper X-ray
performance
check, and the adjustment
exposure precision.
Before adjusting
the generator as described
in this manual, check that the
equipment has been installed according to the installation manual
(2C313-020E)
and understand operating method given in the operation
manual.
Refer to the wiring diagram
the adjustment or test.

supplied

with

These tests are to be performed after


requirements
of the Radiation Control
and all regulations
required in Title
Regulation.

the equipment

when

performing

installation
to meet the
for Health and Safety Act of 1968
21 of the Code of Federal

PREPARATION

2.
2.1

Test Tools and Instruments


Required
Prepare the measuring
instruments
given

Measuring
No.

below.

instrument
Remarks

Hodel

Item
1

; AC voltmeter

I
Digital

, Oscilloscope

Full range:

300/6OOV

voltmeter
Tektronix

434

Or equivalent
Or equivalent

X-ray tube voltmeter


(bleeder resistance
included)

ALCO kV201(D)

X-ray secondary tube


ammeter

ALCO MA-12OlDC

Or equivalent

i<Others>
I
/ Acrylic phantom

180mm thickness

Extension PWB (28 AB


type)
-l-

No.

2.2
2.2.1

2J313-002E*M

Precautions
Introduction
in X-ray
work is protection
the most important
considerations
X-ray
equipment
properly
used does
the hazards
of X-radiation.
careless
or improper
usage can be
not pose any danger;
however,
it is essential
that
all
service
or operating
dangerous.
Therefore,
personnel
of this
equipment
must fully
understand
the problem
themselves.
They should
also
be thoroughly
familiar
and acquainted
with the instructions
in this manual pertaining
to proper
use of the
equipment.

One of
against

2.2.2

Maximum permissible

dose

Various
studies
on the effects
of X-ray
radiation
have provided
the
foundation
for establishing
maximum permissible
dose of X-radiation.
The results
of these
studies
used by the National
Council
on Radiation
Protection
and the International
Comittee
on Radiation
Protection
to
develop
recommendat ions
for the maximum permissible
does
(MPD).
The I.C.R.P.
recommendations
can be interpreted
to say that the maximum
permissible
dose (MPD) to the most critical
organs,
accumulated
at any
by the number of
age,
shall
not exceed
5 x 10 -2Sv (5 rems} multiplied
years
beyond
age 18 and the dose in any 13 consecutive
weeks shall
not
for those
involved
in radiological
Thus,
exceed
3 x 10s2Sv (3 rems}.
works,
accumulation
MPD equals
5(N-18)
x 10W2Sv {5(N-18)
rems),
where N
For
the
hands,
forearms,
and
is
greater
than
18.
is the age in years
the MPD is set at 75 x 10T2Sv
and ankles
of radiologists,
feet,
(75
2.2.3

rems}

a year.

Manufacturers
Although

this

responsibility
equipment

incorporates

protection

against

X-radiation

design
cannot
provide
complete
other
than the useful
beam, practical
does not compel
the operator
protection.
The design
of the equipment
or his assistants
to take adequate
precautions,
or prevent
the
possibility
of authorized
or unauthorized
persons
carelessly,
unwisely
to direct
or secondary
or unknowingly
exposing
themselves
or others
Therefore,
only
authorized
and properly
trained
persons
radiation.
should
operate
this
equipment.
All
persons
authorized
to use the equipment
must be cognizant
of the
The equipment
is sold
danger
of excessive
exposure
to X-radiation.
Medical
Equipment
with the understanding
of the TOSHIBA CORPORATION.
neither
assume nor accept
any
Division,
its
agents
and representatives,
which may result
from exposure
responsibility
for any injury
or danger
TOSHIBA CORPORATION does not accept
any
to X-radiation.
Furthermore,
responsibility
for over-exposure
of patients
or personnel
to the
X-radiation
generated
by this
equipment
due to poor operating
no responsibility
will
be assumed
for
techniques
or procedures.
Also,
any machine
that has not been properly
serviced
and maintained
in
accordance
with the manufacturers
technical
manual or which has been
modified
or tampered
with
in any way.

2 -

No.

2.3

Radiation
(1)

2J313-002E*M

Safety

Radiation

protection

Since
effects
harmful
to the health
may be experienced
by anyone
amount of X-radiation,
the greatest
exposed
to an excessive
consideration
should
be given
to protection
against
the direct
Some
of
these
effects
are
primary
beam from the X-ray
tube.
The best
a
period
of
months
or
years.
cumulative
and may extend
over
is
to
stay
out
of
the
primary
beam
for X-ray
protection
safety
rules
at all
times.
beam produces
a secondary
or
Any object
in the path of the primary
The
intensity
of
this
scattered
radiation
scattered
radiation.
depends
upon the type and amount of the primary
beam and the atomic
This
number of the objects
or tissues
struck
by the primary
beam.
radiation
may actually
be of greater
intensity
than the image-forming
radiation
reaching
the film.
Since
radia

this
tion,

radiation
protective

dangerous
i s as potenti ally
to
meas ures mus t be taken

beam
a s the primary
it.
sa fegua rd against

out with lead


shielding
These protective
measures
may be carried
Items such as movable
lead screens
and lead-impregnated
devices.
can be used to minimize
possible
dangerous
rubber
gloves
and aprons
devices
have
a
lead
thickness
It
is
imperative
that
these
exposures
.
they
should
be
used
by
those
of a least
0.25 mm or the equivalent;
persons
exposed
to radiation
fields
of 12.9 x 10-7C/(kg=h)
{5mR/h)
or
During
servicing
or while
operating
this
equipment
a persons
more.
or other
parts
of the body should
unprotected
hands,
wrists,
arms,
never
be exposed
to the primary
X-ray
beam.
(2)

Monitoring
Monitoring
adequacy
improper
radiation

of

personnel

a valuable
crosscheck
of personnel
provides
It
can
aid
in
revealing
of radiation
safety.
radiation
protection
practices,
and potentially
exposure
situations.

of the
inadequate

or

serious

This
should
be confirmed
by measurement
in rads.
Adequate
protection
measurement
should
be taken
to determine
that the X-ray
radiation
at
A less
any location
does not exceed
the accepted
tolerance
dose.
means of confirming
this
protection
is
but frequently
used,
accurate,
strategic
locations
for a given
to place
unexposed
film
at several
The use of fluorescent
screens,
period
prior
to film
development.
to the darkenss,
may also
aid
after
the eyes have become accommodated
in locating
inadequate
protect
ion areas.
Another
common method of radiation
monitoring
is the use of film
badges.
These
devices
consist
of an X-ray
sensitive
film
inserted
into
a badge which incorporates
metal
filters
of varying
degrees
of
Even though
this
device
can only measure
the
X-ray
transparency
.
radiation
reaching
the particular
area of the body it is worn on, it
does furnish
a clue
to the amount of radiation
that
is being
received.
All
these
regarding

measures
are
X-ray
protect

unless
meaningless
ion are known and

3 -

the rules
are rigidly

of safety
followed.

No.

Refer to the following


requirements.

recommendations

for specific

2J313-002E;kB

safety

(3) Recommendations
It is vitally important that everyone associated with X-ray work
become fully acquainted with the recommendations
of the National
Bureau of Standards, the National Council on Radiation Protection,
The control
and the International
Committee on Radiation Protection.
of diagnostic medical X-ray equipment varies in detail from state to
of the National
state; however, in general, the recommendations
Council on Radiation Protection
based on the other authorities,
are
recommendations
in this manual.
However, prior to operation, all
personnel qualified and authorized
to operate this apparatus should
become familiar with and, where pertinent, conform to the established
recommendations
contained in the documents provided by the
The current sources of information are
aforementioned
authorities.
listed for reference.
a. National Council on Radiation Protection No. 33 (Medical
Gamma-Ray protection
for Energies up to 10 MEV-Equipment
and Use).

X-Ray and
Design

b. National Bureau of Standards Handbook No. 76 (Medical X-Ray


Refer to N.C.R.P. Report
protection
up to Three Million Volts).
No. 33.
C.

Current Recommendations
Radiation Protection.

of the International

-4-

Committee

on

No.

3.

BEFORE

TURNING

POWER

ON

Minimum check points before


Do not turn the main switch
breaker ON before the check
3.1

Visual

visual

inspection

(1) All setscrews


loose.
Various

relays

(3) All connectors


(4) All PWBs
3.2

Power

Note:

SYSTEM
Refer

unit

items

of the units

and distribution

should

be securely

of cables

are securely

terminal

should

inserted

assembly

inserted

not

are as follows:
screws

into the

should

not be

sockets.

be loose.

in place.

Source

. Since
prior
power

4.

turning power on are given below.


of the distribution
panel and the power
is completed.

Inspection

Internal

(2)

2J313-002E*B

the cables were connected for the three-phase


415 V power source
to shipment, reconnect the cables according to the installed
supply (in accordance with section 6.4 of manual No. 2C313-020E).
In the case of a single-phase
power source, change the charging
(Change
the TDR operating time within the
time of the capacitor.
power unit from 30 to 60 seconds.)

SETTING
to Table

7.1 "Setting

and Adjustment

-5-

Table".

No. 2J313-002EfcD

5.

PERFORMANCE CHECK
The performance check flowchart is given below:
Note: Check to see that the shielding terminal of the cable (for serial
transmission) is connected to the cabinet for noise prevention.
If
it is not connected, error 700 (a communication error> may occur.

PWB unit with control box

SPIO
or SIO-1

PWB

Chassis

BSXlZ-0276(5 pairs, 7.2$, 10m)

[I.

-6-

Do not connect the


panel or stand-side
shielding terminal.

No.

Start

2J313-002EkB

1
Cable
power

Note:

connections
supply.

must

comply

with

the

1
Connect

cables

in

accordance

with

the

system

connection

diagram.
I

connected
with
Only the power supplies
system
cables
can be used externally.

Note:

Disconnect

the

connector

from

the

trigger

PWB within

the

INV unit.

t
System setting
when installed

(for
all
columns

items
indicated
of tables
in

Note : The charging


power supply.

Energizing

as necessary
section
7.1. >
time

in

must

the

comply

with

the

I
t

) Unit

test

I sequence
Filament
setting

(sections

5.1.1

5.1.6)

check

heating

check

rConnect

to

the

connector

data

and heating

(cable

length

and

tube)

t
Power

to

off

the

trigger

PWB within

the

INV unit.

I
Connect
the connector
citor
voltage
become

Radiographic

& fluoi!oscopic

adjustment

(sections

5.2

after
the capaless
than 5V.

and

5.4)

ABC/AEC adjustment

ATR adjustment

I
(

End

Note : Enter
)

the

adjustment

data

in

the

data

sheet.

No. 2J313-092E

5.1
5.1.1

Unit

Test

Power

ON

Turn ON the main switch of the distribution


panel and the power unit
Check for abnormality
by turning the
breaker within the control box.
panel or stand power switch ON/OFF.
Within approximately
60 seconds after
the capacitor unit inside the control
approx. 30 seconds).
Note:

5.1.2

After the power is turned off,


approximately
four minutes.

Checking

the power switch is turned on,


box is charged (three-phase:

discharging

is completed

in

voltage

(1) KXO-30R
(a)

Voltage

within

the panel

Panel Cont 1 PWB


Between CH2 (+5 V> and CHl

(GND):

If the above allowab!e voltage


the VR of the stabilized power
Control

box:

Pl witnin

Make sure that


+24 v+3
- v.
(b) Voltages
Mother

the voltage

within

value
source

Unit

is not indicated,
given
below.

adjust

between

the control

PWB of PWB Unit

. Between
. Between
. Between

Power

+5 V,O.l

CH3

(+24 V> and CHl

(GND)

box

Jl (IAB) and Jl (28AB) : +5 V+O.l V


J16 (2AB) and Jl (28AB): +15 ~~0.1 V
J16 (3AB) and Jl (28AB): -15 VzO.1 V

If the above
allowable
voltage
values
are not indicated,
the VRs of the stabilized power source given below.
p2 (+5 v), ~3
(2) KXO-30F
(a>

is

(+I5 V> and P4 (-15 V> within

Power

Unit

adjust

or 50 F

Voltages
. Desk

within
display

Between

CH2

the stand
PWB
(+5 V> and CHl

(GND):

+5 V+O.l V

If the above
allowable
voltage
value
is not indicated,
adjust
Make sure that
the VR of the stabilized
power source (PSZ).
V.
the voltage between CH3 (+24 V> and CHl (GND) is +24 V+3
-

-8-

No.

. Desk

control

Between

PWB

P2 (+5

V> and

pl

If the above allowable


the VR of the stabilized
(b)

Voltages
Mother

within

(3)

ths

PWB 2 of

. Between
. Between
. Between

PJl
PJl
PJl

2J313-002E*B

(GND):

voltage
power

control

PWB Unit

(4AB)
(2AB)
(3AB)

+5 V+O.l
value
source

is not
(PSl).

indicated,

box
2

and PJl
and PJl
and PJl

(28AB):
(28AB):
(28AB):

+5 V+O.l V
+15 V~0.1 V
-15 V+_O.l V

If the above allowable


voltage
values
the VRs of the stabilized
power source

are not indicated,


given
below.

Pl

within

(+5

LV detection

V),

P2 (+15
circuit

V),

adjust

and P3 (-15

V)

Power

adjust

Unit

adjustment

in following
Adjust
VR2, VR3 and VR4 on INV3 PWB as instructed
Tl on the power unit
table
only when the input
tap o f transformer
This adjustment
, however,
is
in the control
box indicates
200 V.
been set.
not required
when other
taps have already

VR to be
1 adjusted

Measurement
point

1
VR2

TP2

VR3
VR4

Tl

220 V,

200 v

415

V,

+4.75

+5.25

TP3

+4.87

TP4,

+5.35

Setting

-9-

380 v,

LED3 through
After
adjustment,
the LV levels
with respect
to
value)
(see
the table
below).

tap

+4.50

+4.75

input

prior

LED7 of the INV-3 PW board


the nominal
value
(Tl input

to

440

V /

shipment
indicates
tap

No. 2J313-OOZEW

Nominal

Other
values

200V

nominal
= Vs

LV>zzoVQc

LV>V$$-10%

220v~Lv>214v

INV-3

Radiography
flu
scopy disabled
area

LV LEVEL

s
m
%
i LED7

Operating area

vstlo%~Lv>vst5%

u
R64

+i

9
LED6

vs t5% 2 LVL vs -5%

214LV2195

Optimum

operat

04

R62

,Q u
LED5

Operating

Vs-5%>LVLVs-10%

195>LV 2 190

area

R61

c
z

5.1.3

Checking
(1)

operations

and

Radiographic

(2)

the

Checking
Repeating
indications
.
.
.
.
.
.

circuits

settings

condition
individual

reading
ID No.

settings
and indications
operation
manuals.

when

the radiographic
according
to the

turning

turning
equal

power
those

power

conditions
operation

of

the

in and from
manual,
check

units

an
for

the

ON/OFF

that
ON, OFF, and ON, check
before
turning
power OFF.

the

following

R kV value

R mAs value
R mA value

R TIME value
TEC. SW
SF/LF

Make sure that


the contents
turning
power ON and OFF.
(4)

u
R60

LED3

ID function
Writing
and
appropriate
ID function.

(3)

display

condition

Check radiographic
as described
in

zu%
ii:

Radiography
fluoroscopydisabledarea

vs-lO%>LV

19O>LV

LED4

Overload
Check
X-ray
X-ray

written

in

the

ID are

not

destroyed

by

protection

function
under conditions
of the
the overload
protection
tube maximum rating,
filament
characteristics,
kV limit,
generator
rating
(40 kW, 60 kW).

10 -

set
and

No. 2J313-002E"cA

5.1.4

Checking

the

(1)

potential

High

switching

circuit

changer

(KXO-30F

or

50F)

(HRU-A) of the
Operating
the changer
selecting
TECHNIQUE via the panel
or
changer
changes
over normally.
(2)

Focus

change

Checking
There

the

are

(a)

Normal

(b)

High

(c)

Anode

panel

starter

three

of

stand

types

of
for

make

for

for

HSS is controlled
by
the starter
frequency
wiring,
X-ray

indication

focus

anode

radiography

rotation

rotation

tube

fluoroscopy

the
is

rotation

as

(Named

NSS)

radiography

NSS and F-ROT are controlled


The starter
X-ray
control.
(50 or 60 Hz).

Check the
installed

stand,

generator
sure

by

that

the

varies

with

the

set

circuit

rotation
speed

voltage

by R-MA

Check that
the
tube current.

5.1.5

high

(Named

(Named

shown

HSS)

F-ROT)

by the normal
speed
frequency
equals
the

high speed
180 Hz.

starter

phase advance
capacitor
tubes
as described
below.

below.

starter
within
the
power source
frequency

unit

(RS-118A:

value,

and

set

option),

time

of

and

the

[NSS and F-ROT modes]


( 1)

Check
tubes
(a)
(b)

the type
depending

XS-U (standard
stator)
XS-AA (1 ow impedance

For detailed
tube technical
(2)

of the X-ray
tube combined.
on the X-ray
tube stator.

Phase

advance

The relation
X-ray
tube

X-ray

stator
data.

tube

refer

to

are

two

types

phase
below.

Capacity

advance

the

capacitor

Capacitor
the power

TOSHIBA U coil

24uF

Cl

TOSHIBA AA coil

44uF

Cl

individual

value

No. within
unit

+ c2

cl: 24uF, C2: 20uF

11 -

of

tube
X-ray

value

between
the
type is given

type

TOSHIBA X-ray
tube
stator)
TOSHIBA X-ray

characteristics,

capacitor

There

and

the

2J313-002EcA

No.

(3)

Time

setting

The relation
given
below.

between

the

X-ray

tube

I
/
/

type

The oscilloscope
measured.

(a)

Check

(b)

Set

that
the

is

power

on

to

X-ray

tube

type

is

time

be

isolated

observe

from

the

betweenj

power unit within


storage
mode)

and

the

circuit

to

be

off.

oscilloscope

Block
TB2 of the
div,
0.2 sec/div,

and

+o. 1
set
-0

should

power

time

Boost

I
1 0.8
I

TOSHIBA U coil

Note:

boost

select

Tube

the

and / X2
control

box.

of

Terminal

(100

VI

1.

(c)

Turn

(d)

Set a low kV value,


a low mA value,
Normal Speed Rotation
(NSS) mode.

(e)

Put into
the
oscilloscope

and a short-time

Ready state.
Observe
the
and check
tne boost
time.

waveform

on

to

keep

in

the

[HSS mode 1
(a)

5.1.6

Set the stator


coil,
start
voltage,
and control
time
control
voltage,
installation
manual.
speed
starter

(b)

Select

(c)

Set a kV value,
Rotation
(HSS)

(d)

Put into
rotation

Checking

the

the

TECHNIQUE of

the

and

a mA value,
mode.

the Ready
according

state,
to the

tube

as

to

R start
time,
specified
in

F start
time,
the high

be checked.

a time

to

keep

in

and check
the high speed
said
installation
manual.

High

Speed

anode

battery

Each unit PWB is provided


with a chargeable
which stores
the memory contents
when power
both terminals
and check
the voltage
between
given
be low.

12 -

battery
or
turns
off.
according

dry battery,
Turn power
to the table

off,

No. 2J313-OOZE*1

Equipment

KXO-30F/50F

name

Cont

1 PWB

Voltage
3.6

V to 2.4

Heat 2 PWB of PWB Unit


1 within the control box

2.5

Stand

3.6

V to 2.4

Panel

KXO-30R

Unit

desk

control

PWB

Heat 2 PWB of PWB Unit


2 within the control box

5.2

Radiography

V (Chargeable)

V (Chargeable)

2.5 V

Adjustment

The adjustment method for tube voltage and current is given below.
The R.mA and R.kV adjustment procedures
are given separately below.
and R.kV should be adjusted simultaneously
at each mA step.
5.2.1

Measuring

instrument

The measuring

R.mA

connection

instruments

should

be connected

as shown

below.

(1) KXO-30R
<l> Set up the required units as illustrated
the tube voltage and current.
Control

source

generator

\
-9

and adjust

--

Secondary tube current


meter (to be mounted
to the X-ray tube)
\

Anode
Bleeder
resistance

DG-303

Power ground
terminal

both

box

G--J
Power

below

Storage
oscilloscope

13 -

\
f-A-X-ray
tube

Cathode1

No.

<2>

Set up the required


units
tube current
and readjust
It is not necessary
voltmeter-ammeter

Control

is

as
it

to perform
used for

illustrated
if necessary.

below,

this
procedure
measurement.

box

25313-002E*l

check

if

the

the

X-ray

tube

Secondary
tube current
meter
(to be mounted
to the X-ray
tube)

Anode
X-ray
tube

High

Power
source

voltage
-_:l
1

--

--Jr

Cathode

1
Power ground
terminal

-I

. Do not generate
X-rays
Prior
to checking
the
(within
the TNV unit)
l

with Tl and T2 opened.


disconnect
X-ray
sequence,
connector
or the capacitor.

Note : Before
disconnecting
the capacitor
voltage

the
is

- 13A -

connector
or
5 V or less.

the

capacitor,

TRG PW board

check

that

No.

(2)

KXO-30F

and

25313-002E*J

50F
Secondary
tube current
meter
(to be mounted
to the X-ray
tube)
DG - 50F

Power

source

Control
box

n
\
&-7

Tl

Tl

_
T2

T2

---

High
voltage
gener-

lnode

ator

Cathode

Bleeder
resistance

X-ray
tube
Cathod

--A

Tube
voltmeter

-f-

Power ground
terminal

Storage
oscilloscope

. Do not generate
X-rays
to checking
the
. Prior
(within
the INV unit)
Note:

with Tl and T2 opened.


disconnect
X-ray
sequence,
connector
or the capacitor.

Before
disconnecting
the capacitor
voltage

the
is

14 -

connector
or
5 V or less.

the

capacitor,

TRG PW board

check

that

No.

5.2.2

Radiographic
tube current
adjustment
(Applicable
to products
manufactured
from April,
1989 to
Refer
to Radiographic
tube current
adjustment
described
for products
manufactured
in and after
July,
1990.)
(1)

Operation

switch

and

adjustment

The HEAT 2 PW board,


which
with the operation
switches
below:

cl>

Mode indic

x2>

Mode

tor

has an adjustment
function,
and indicators
give
in the

SW

MODE

<3>

Adjustment

<4>

S/R

switches

switch

S/R

June,
1990.
on page 24B

mode

TEST

selec

15 -

25313-002E*J

is provided
drawing

25313-002E*J

No.

<l>

Mode indicator
from 0 to
This LED changes
mode selector
switch.
Each mode is to be used as

1 to

2 and

follows

finally

to

according

to

7 with
the

the

adjustment

type*
Mode

Adjustment

Indicator

Type

moo

Standby

a.0

tube current
Radiographic
(Point
adjustment
>

adjustmnet

000

Radiographic
tube
(Line
adjustment)

current

adjustment

current

adjustment

state

moo

00.

Fluoroscopic

0.0

PRH and

000

Cable

000

Tube setting
error>

tube

PRF level

length

adjustment

setting
(recovery

from

a data

0 ON (Lighted)
Pressing

the

S/R

switch

causes

any mode

to

be

reset

to

Mode

0.

<2> Mode selector


switch
The mode selector
switch
is used to select
the mode and to
check
the main heating
level.
Each time the switch
lever
is set in the downward position,
the
The mode
mode is switched
from 0 to 1 to 2 and finally
to 7.
is then reset
from 7 to 0.
When the switch
lever
is set in the upward position,
only
the
filament
heating
circuit
becomes
ready
for main heating
so that
the heating
adjustment
level
can be checked
operations
(in Modes 0, 1 and 2).
<3> Adjustment
switches
These
switches
are
level
and to input
In Modes 1,
code.

If

that

the

A,
sl

caused

caused

or decrease
cable
length

the heating
and the tube

switch

is

pressed,

voltage

at

Eo increases.

switch

is

pressed,

voltage

at

Eo decreases.

switch

is

pressed,

the

voltage

increases

ten

times

the

voltage

decreases

ten

times

by pressing
switch

that

used to increase
the high-voltage
2 and 4,

is

the
pressed,

by pressing

(Note)

the

- 16 -

NO. 25313-002E*J

Note:

Eo is a check
pin
check
pin AGND is
Eo corresponds
to
transformer.

on the HEAT 1 PWB.


used as the standard
the primary
voltage

Voltage
for
of

at

the

that at Eo.
the filament

Whenever
each of these
switches
is pressed,
the buzzer
will
will
increase
or
sound and,
in Mode 1 or 2, the voltage
decrease
to approximately
0.05 V (to approximately
0.5 V when
If the switch
continues
to be
the tenfold
switch
is pressed).
the buzzer
will
automatically
start
sounding
so that
pressed,
the result
can be achieved
as if the switch
were pressed
by the
number of times
the buzzer
sounded.
for using
these
switches
to input
the
The procedures
in the
high-voltage
cable
length
or the tube code are described
section
<4>

(2)

concerning

adjustment

methods

in

each

mode.

S/R switch
This switch
is used to reset
Mode 0 from any mode or to input
the high-voltage
cable
length
and the tube code
simultaneously
using
the adjustment
switch.
The same result
can be achieved
by setting
the S/R switch
lever
in either
the upward or downward position.

Adjustment
Mode

Mode

procedure

Tube

code

setting
/

Mode
Mode

2
1

Radiographic
adjustment

Mode

tube

current

tube

current

__,__
I
I
I
I
I
I
I
I
I

Mode 4

Fluoroscopic
adjustment

(Not

required

First
perform
setting
for
(Only
the cable
three
tubes.
length
setting
will
not cause
the heating
data to change. >
Note:
When the X-ray
generator
is used with
an X-ray
tube with a
very
small
focus,
such
as the X-ray
tube for
magnification
radiodo
not
connect
graphy >
tube H.T. cables
until
the tube code
is set.
Failure
to do so may
damage the tube focus.

for

KXO-30R)
-

17 -

PRH/PRF

level

adjustment

No.

(3) Setting

the

high

voltage

cable

length:

Mode

The standard
cable
length
is 15 m (for
three
tubes).
change
the cable
length
depending
on the system.
<l>

--+<2>

<3>

Set

Mode

6 with

Set the tube


X-ray
control
If

the

No. to
unit.

swith
1)
cl 0

equipment
<4>

the

goes

is

into

mode

selector

the

data

If

necessary,

switch.

be changed

via

the

operation

the

indicator

input

state.

pressed,

25313-002E*J

panel

blinks,

of

and

the

the

Enter
the new cable
length
using
the adjustment
switches
and
the S/R switch.
If the two-digit
(00 through
99) input
is completed,
the mode
resulting
in the condition
in <2>.
indicator
stops
blinking,
[Example:

6 ml

(Do not
the

/
cl

First
time

press
switches)

aI

[-ZJ

[4

S/R

Press
4

Press
L

[z.][-zj
m 1-q

Second
time

0
If

<5> Perform
voltage

[ZJ
0

Note:

Press

the

data

is

the operations
cable
length

<6> When the


S/R

switcll

the

tube.

correctly
(2)
for

set,

BCD code

0
1

0000
0001

2
3

0010
0011

4
5

0100
0101

6
7

0110
0111

8
9

1000
1001
-

aI

S/R

a buzzer

goes

through
<4> when changing
the 2nd or 3rd tube.

adjustment
is completed,
or set Mode 7 with
the

Value

Press
---f

18 -

reset
mode

on.
the

Mode 0 by pressing
selector
switch
to

high-

the
set

No.

(4)

Tube

code

setting:

Mode

7
Select
set as initial
setting.
from Table
5.2.2
and perform

The DRX-2903HD (for


3 tubes)
is
each tube number to be combined
setting
by the following
steps.
<l>
--+<2)

Set

Mode

Set
the

the tube number to


X-ray
control
unit.

<3> When the


the

7 with

r
I/),

equipment

<4> Enter
Table

the

mode

switch

the binary
5.2.2.

is

code

selector

be

into

goes

25313-002E*J

set

switch.

through

the

operation

panel

pressed,

the

mode

indicator

blinks,

the

input

code

corresponding

of

and

state.

to

the

tube

to

be

used

from

[ Examp le I DRX-1203*
Do not

Press

press

--_)

aI
0

Binary

code

First

time

S/R

Press
L

Press
*

al
0

Second

time

S/R

Note:

-<5>

Upon completion
of correct
Shortly
after
that,
out.
indicator
lights
up once

Perform

<6> After
Mode

this

setting

completion
0. >

for

of

setting,

tube setting
Note : If
data
adjustment
into
memory.
The
small

above

procedure

focuss.

is

the

is
is

used

When the

setting,
the buzzer
again.

2nd

and

press

3rd

the

the indicator
sounds
and

goes
the

tubes.

S/R

switch.

(Restores

completed
after
other
adjustments,
deleted
and preset
data are stored

to

set

the

switch

is

data

with

pressed

both
in

step

large
<3>,

the

and
data

with either
a large
or small
focus
can be set.
This setting
can be used when the tube to be combined
is not listed
in Table
5.2.2
and large
and small-focus
data are to be substituted
or when either
large
or small-focus
data
with different
tube data,
is to be reset.
for example)
is input
for a blank
tube name
If any code
(00 0000,
the buzzer
will
sound for about
two seconds,
column
in Table
5.2.2,
thus indicating
that
there
is no tube for
the input
code.

19 -

No.

23313-002E*J

is noted
during
the inputting
operation
or it is
If an input
error
desired
to input
data once again,
reset
the Mode 7 initial
state
(with
the indicator
lit)
by pressing
the S/R switch
several
times.
(5)

Radiographic

tube

current

adjustment

For the radiographic


tube current
adjustment,
Mode 1 is used to
ad just
each adjustment
point,
and Mode 2 is used to adjust
the tube
current
(MA) for each line.
The radiographic
tube current
by first
adjusting
60 kV in Mode 2
adjustment
can be facilitated
and then adjusting
40 kV, 90 kV and 120 kV in Mode 1.
Adjustment

for

each

Adjustment
for
120 kV in Mode
When the
adjustment
The
are

point

40,
1.

in Mode

60,

Adjustment

90,

for

each

Adjustment

for

MA line

in Mode

60 kV in Mode

2.

tube code
is set for mammography in Mode 7, four
points
at 20, 25, 32 and 45 kV are required.

adjustment
points
for
20, 25, 32 and 45 kV.

mammography

The line
adjustment
should
be a; 25 kV.

for

mammography

25kV
20kV 1 I]

800mA 1
Adjustment
for
20,
45 kV in Mode 1.

25,

Adjustment

32 and

I
for

25 kV in Mode 2.

<Mode l>
tubes)
is set before
shipment.
The standard
data
(for
three
Correct
the data depending
on combinations
of tubes.
Since
the preheating
and preflash
are not checked,
adjust
the
(Generate
X-rays
current
after
the filament
heating
is stabilized.
by pressing
the READY switch
and the X-ray
switch
at intervals
of
them
simultaneously.
Take
note
(Do
not
press
more than 2 seconds.
of the heating
data as shown in table
10. (1) in APPENDLX 01 after
adjustment.)

- 20 -

No.

<1> Set Mode

1 with

the mode

selector

25333-OOZE*5

switch.

<2> Set the tube number (1, 2, or 3) to be adjusted


operation panel of the X-ray control unit.

via

the

<3> Set R.kV and R.mA to be adjusted in the same way.


Some
combinations
of R.kV-R.mA
are limited to Maximum Power (30
kW/SO kW).
R.kV: 40, 60, 90, 120 kV or 20, 25, 32, 45 kV
R.mA: 10, 13, 16, 20, 25, 32, 40, 50, 64, 80, 100, 125, 160,
200, 250, 320, 400, 500, 640 mA
:

<4> Correct

the data with

the

/c
i/(
cl ;

increase)

or

A, '1 (decrease)
n

switches (Note 1) by generating


X-rays under the individually
set conditions
in order to adjust the radiographic
tube
Adjust all conditions
current.
to be used in this way.
<5> Press the Ready SW and the X-ray SW simultaneously
the set conditions
again to see if it is necessary
preheating
or calibration
of preflash level.

and confirm
to perform

<6> After checking the above, if the tube voltage is stabilized


within a specific time, Item (6) "Preflash level adjustment"
level adjustment"
are not required.
and (7) "Preheating
<7> When completing
Mode 0.
Note

1: When

adjustments,

press

switch

the

increased
the B]

the S/R switch

or decreased
or [q

switch

is used,

ten times

to return

the value

as compared

to

can be

to when

is used.

<Mode 2>
in this mode causes 40, 60, 90 and 120 kV to
The adjustment
simultaneously
increase or decrease.
is the same as Mode 1, except that this
Mode 2 adjustment
adjustment should be made with the R.kV tube voltage set to 60 kV.
When the X-ray tube for mammography
is used, the tube current
adjustment in Mode 2 should be done at 25kV.
(6) Preflash

level

adjustment

(PRF): Mode

The preflash level should be adjusted only when the tube current
does not rise within the specified time in (5) <6>.
Preflash is intended to increase the X-ray tube filament rise rate.
After the READY switch is pressed, the preflash is done prior to
main heating.
~Preflash
Heating

level

Main heating

Preheating

I
READY w ON ))
- 21 -

Time

No.

25313-002E*J

time is fixed and the amount of


For this unit, the preflash
preflash can be obtained by multiplying
the amount of main heating
by the preflash factor.
the preflash
level is adjusted by changing the preflash
Therefore,
factor.
The preflash factor curve is generated by preflash factors at three
The three points are the maximum, minimum and middle
points.
points of the filament current on the X-ray tube current
characteristics
curve.
The three preflash factors, PRFl, PRF2 and PRF3 are to be selected
with

the adjustment
Tube

Tube

and

[,,[a,
Current

[Tg

switches*

Characteristics

factor

current
(l.mQ

Middle

Minimum
Filament
PRFl, PRF2, and PRF3
selected as follows:

Maximum

Current

(A)

for the three

adjustment

points

can be

Plot the X-ray conditions


under which the preflash level is to be
If the
curve.
adjusted, on the tube current characteristics
If the
filament current is near the maximum value, adjust PRFl.
filament current is near the middle or minimum value, adjust PRF2
or PRF3, respectively.
If the filament current is between the minimum and middle values,
If the filament current is between the
adjust PRF3 and PRF2.
middle and maximum values, adjust PRF2 and PRFl.
<l>
--+<2>

Set Mode

the mdoe

5 with

selector

switch.

Set conditions
under which the preflash level is short in (5)
<5>, via the operation panel of the X-ray control unit.

<3> Select

the preflash

m,

m,

PRF3

PRFZ

sequence

factor

or [z]

to be

switch.

adjusted

The mode

PRFl

proceeds

to the next

- 22 -

step.

by

pressing

indicator

the

blinks,

and the

No.

<4> While
checking
the tube current
rise
the READY and X-ray
switches,
change
using

the

[z]

or

Increase
L<5>

[a

switch

1).

Decrease

<6> When the adjustment


S/R switch.
1:

by simultaneously
pressing
the preflach
factor
by

(Note

When the S/R switch


is pressed,
and the state
in <2> is reset.
another
focus
or another
factor,
adjusted.

Note

25313-OOZE*5

When the

is

the indicator
In this
state,
tube preflash

completed,

[z]

or

[A

reset

switch

Mode

is

stops
blinking
another
preflash
factor
can be

0 by

pressed,

pressing

the

the

buzzer

Increase
Decrease
will
sound and the preflash
factor
will
increase
or
If either
switch
continues
to be
decrease
by 1%.
the buzzer
will
automatically
start
sounding
pressed,
so that the result
can be achieved
as if the switch
were pressed
by the number of times
the buzzer
sounded.

( 7) Preheating

level

adjustment

(PRH) : Mode

If the adjustment
This adjustment
is not necessary
in principle.
in (6) cannot
be carried
out satisfactorily
depending
on the
system,
this
adjustment
should
be carried
out as an auxiliary
adjustment
to (6).
<l>

Set Mode

5 with

<2> Set the conditions

the

switch

<4> Press

sequence

is

proceeds

11,

selector

switch.

the

next

switch

preflash
panel

the

pressed,

to

the
Increase

(Note

mode

under which
the
via the operation

not satisfactory,
unit.
<3> When the

the

to

mode

of

blinks

is

and

stop.
adjust

the

preheating

level

Decrease

and check

<6> When the adjustment


S/R switch.
1:

adjustment
X-ray
control

indicator

th e

tube

current

rise

by generating

<5> When the S/R switch


is pressed,
the indicator
In this
state,
and the state
returns
to <2>.
another
tube preflash
factor
can be adjusted.

Note

level
the

When the m

is

completed,

or [x]switch

reset

is

Mode

pressed,

X-rays.

stops
blinking
another
focus

0 by pressing

the

buzzer

or

the

will

If either
sound and the preheating
level
will
change.
switch
continues
to be pressed,
the buzzer
will
sounding
so that the result
can be
automatically
start
achieved
as if the switch
were pressed
by the number of
times the buzzer
sounded.
-

23 -

No.

(8)

Changing

the

The following
adjusted
to

Focal

Size

for

Fluorography:

Mode

25313-002E*J

procedures
permit
to change
the focal
size
the small
focus
(SF) at the time of shipment.

<I>

Set

Mode

6 with

the

<2>

Use the operation


the number of the

mode

selector

already

switch.

panel
of the X-ray
control
tube whose fluorographic

unit
to specify
focal
size
needs

to

be changed.
switch
enabling

mode,

the

<4> To adjust
the
switch
only.
For

the

to

place

following

focal

size

focus

(SF),

small

the

mode

to

the

large

tilt

the

focus

S/R

switch.

A buzzer
procedure

sounds
and the indicator
stops
has been performed
correctly

<5> Reset
to mode 0 by means of the
size
adjustment
has been completed.

the

focal

size

in

blinking

steps.

the

To change

indicator

to

S/R

(LF),

switch,

tilt

while

blinking
so far.
switch

the

if

when

pressing

the

the

the

focal

size

to

focal

LF
Do not

To change

S/R

press

any

switch

Tilt

SF

I+
(9)

Recovery
If

from

a data

any

error

data

has

occurred,

recover

When the S/R switch


is pressed
alarm stops
and Mode 7 is set.

(2)

When the

A,
sl

recovery

routine
[z]

down

with

the

this

switch

Tilt

error

<l>

<3> When the

Hold

switch

is

pressed,

while

the

the

following

alarm

indicator

procedure:

sounds,

blinks

the

and

starts.
and

switches

[q

are

pressed

simultaneously,

the indicator
goes
out and the recovery
routine
(for
Upon
regeneration
of an error
check
code)
is executed.
the indicator
lights
up and the
completion
of execution,
sounds.
<4>

the

buzzer

Reset Mode 0 by pressing


the S/R switch
and check
the data.
the power supply
is turned
off
in the adjustment
mode during
is judged
to have be destroyed
tube current
adjustment
, data
-

24 -

If

No.

2J313-O02ECJ

when the power supply


is turned
on again,
and the alarm sounds.
(This
is because
it is not possible
to distinguish
data change
recover
by using
the
from data breakdown.
> In such cases,
After
recovery,
data adjusted
prior
to
above
procedure.
turning
off
the power supply
are all
effective.
Note:

is simply
to regenerate
the error
The recovery
routine
but not to repair
the destroyed
data.
check
code,
it is necessary
to check
the data after
Therefore,
recovery.
If the check
shows that
the data are changed,
correct
If data have been completely
in Mode 1, 2 or 4.
In this
it is necessary
to reset
the tube.
destroyed,
the setting
must start
with the setting
of the
case,
high-voltage
cable
length.

it

> Alarm
<l>

Stored
data check
The stored
data is checked
by heat data when the power supply
is turned
on.
Data is checked
in each tube and each focus
(for
data blocksTube 1 LF, Tube 1 SF . . . Tube 3 SF).
If an
in which an error
has
attempt
is made to use the data block
an alarm
(intermittent
beeping
sound)
and an
been detected,
To recover
from this
state,
error
message
will
be generated.
execute
recovery
procedure
(9) from a data error.

<2> Set condition


check
than 150 kV or the tube
If the set tube voltage
becomes
greater
an alarm
(intermittent
beeping
current
is not set correctly,
If the tube voltage
or the tube
sound)
will
be generated.
the alarm will
be removed.
current
is set correctly,
<3> Maximum filament
current
check
smaller
than the
The heating
output
data is set to a value
maximum filament
current
to prevent
the X-ray
tube filament
If the heating
data exceeds
the maximum
from breaking
down.
filament
current
value
when the READY switch
is pressed,
the
to
the
preheating
level
and
an
alarm
heating
level
will
be set
(intermittent
beeping
sound)
and an error
message
will
be
generated.

24A -

No.

5.2.3

Radiographic
(Applicable
(1) Operation
The HEAT

tube current adjustment


to products manufactured
switch

and adjustment

in and after

July,

25313-002E"J

1990.)

mode

function, is provided
2 PW board, which has an adjustment
switches and indicators given in the drawing

with the operation


below:

xl>

Mode

indicator

C
TEST

<2>

Mode

selector

switch
MODE

<3>

x4>

Adjustment

switches

S/R

S/R switch

- 24B -

No.

<l>

Mode indicator
This LED changes
from 0 to
mode selector
switch.
Each mode is to be used as

1 to

2 and

follows

finally

to

according

to

25313-002E*J

7 with
the

the

adjustment

type*
Mode

Adjustment

Indicator

**

Standby

Radiographic
tube
(Point
adjustment)

current

adjustmnet

00.

Radiographic
tube
(Line
adjustment)

current

adjustment

Fluoroscopic

current

adjustment

moo

00.

T-- 0.0

6
7

tube

PRH and PRF level

adjustment

000

Compensation

for

000

Tube
error

(recovery

0
Pressing

state

0.0

Type

the

S/R

switch

setting
>

tube

current
from

drop
a data

ON (Lighted)

causes

any

mode

to

be

reset

to

Mode

0.

<2> Mode selector


switch
The mode selector
switch
is used to select
the mode and to
check
the main heating
level.
Each time the switch
lever
is set in the downward position,
the
The mode
mode is switched
from 0 to 1 to 2 and finally
to 7.
is then reset
from 7 to 0.
When the switch
lever
is set in the upward position,
only
the
filament
heating
circuit
becomes
ready
for main heating
operations
so that
the heating
adjustment
level
can be checked
(in Modes 0, 1 and 2).
<3> Adjustment
switches
These
switches
are
level
and to input
code.
In Modes 1,

If
that
If
that

the

7
0

caused
the

/)
LI

caused

used to increase
the high-voltage
2 and 4,

or decrease
cable
length

switch

is

pressed,

voltage

at Vset

switch

is

pressed,

voltage

at

switch

is

pressed,

the

by

pressing

switch

is

by pressing
-

the
pressed,
the
24C -

y4
cl

voltage

the heating
and the tube

increases.

Vset

decreases.
(Note >
increases
ten times

switch.
the

voltage

decreases

ten

times

No.

Note:

Vset
is a check
pin on
check
pin AGND is used

25313-OOZE5

the HEAT 1 PWB. Voltage


as the standard
for that

at
at

the
Vset.

Whenever
each of these
switches
is pressed,
the buzzer
will
sound and,
in Mode 1 or 2, the voltage
will
increase
or
If the switch
is held
down, the buzzer
will
autodecrease.
matically
sound at fixed
intervals;
the number of times
the
buzzer
sounds
is the same as pressing
the switch
the same
number of times.
The procedure
for using
these
switches
to input
the tube code
is described
in the section
concerning
adjustment
methods
in
that mode.
<4> S/R switch
This switch
is used to reset
Mode 0 from any mode or to input
the high-voltage
cable
length
and the tube code simultaneously
using
the adjustment
switch.
The same result
can be achieved
by setting
the S/R switch
lever
in either
the upward or downward position.
(2)

Adjustment
Mode

Tube

Mode
Mode

procedure

code

setting

2
1

Mode

Radiographic
adjustment

tube

current

Mode

Note

1:

When the X-ray


generator
is used with an X-ray
tube with a
such as the X-ray
tube for magnification
very
small
focus,
radiography,
do not connect
tube H.T. cables
until
the tube
code
is set.
Failure
to do this
may damage the tube focus.

Note

2:

Diagnostic
X-ray
generators
manufactured
after
April
1990
is turned
ON without
connecting
cause
an error
if the power
X-ray
tubes.
Be sure
to remove
the HEATlA PWB and to
ground
12B on the HEAT2 PWB before
turning
the power ON.

24D -

No.

(3)

Tube

code

setting:

Mode

7
Select
set as initial
setting.
from Table
5.2.2
and perform

The DRX-2903HD (for


3 tubes)
is
each tube number to be combined
setting
by the following
steps.
<l>

Set

Mode

<2>

Set
the

the tube number to


X-ray
control
unit.

<3> When the


the

the

code

is

code

selector

be

into

goes

the binary
5.2.2.

[Example]

mode

switch

/)
CJ

equipment

<4> Enter
Table

Binary

7 with

25313-002E*J

set

switch.

through

the

operation

panel

pressed,

the

mode

indicator

blinks,

the

input

code

and

state.

to

corresponding

of

the

tube

to

be

used

from

DRX-1203B
Do not

press

[d

00 0010

Tilt
&

First

time

S/R

?
Eold

down

Note:

this

setting

<6> After
completion
Mode 0. >

The
small

Upon completion
of correct
Shortly
after
that,
out.
indicator
lights
up once

<5> Perform

for

of

If tube setting
adjustment
data
into
memory.

above

procedure

focus.

is

the

is
is

used

Tilt

S/R

setting,
the buzzer

the indicator
sounds
and

goes
the

again.
and

tilt

3rd

the

tubes.

S/R

switch.

(Restores

completed
after
other
adjustments,
deleted
and preset
data are stored

to

set

switch

When the

2nd

setting,

Note:

this 1 switch

the
is

data
pressed

with
in

both

large

step

<3>,

the

and
data

with either
a large
or small
focus
can be set.
is not listed
This setting
can be used when the tube to be combined
in Table
5.2.2
and large
and small-focus
data are to be substituted
or when either
large
or small-focus
data
with different
tube data,
is to be reset.
for example)
is input
for a blank
tube name
If any code
(00 0000,
the buzzer
will
sound for about
two seconds,
column
in Table
5.2.2,
thus indicating
that
there
is no tube for the input
code.

24E -

No.

25313-002E*J

If an input
error
is noted
during
the inputting
operation
or it is
desired
to input
data once again,
reset
the Mode 7 initial
state
(with
the indicator
lit)
by pressing
the S/R switch
several
times.
(4)

Radiographic

tube

current

adjustment

For the radiographic


tube current
adjustment,
Mode
and Mode 2 is used to
adjust
each adjustment
point,
current
(MA) for each line.
The radiographic
tube
adjustment
can be facilitated
by first
adjusting
60
and then adjusting
50 kV, 90 kV and 120 kV in Mode
Adjustment

for

each

Adjustment
120 kV in

Mode

for

point

50,
1.

60,

in Mode

for

each

Adjustment

for

Adjustment

90,

1 is used to
adjust
the tube
current
kV in Mode 2
1.

MA line

in

Mode

60 kV in Mode

2.

When the tube code


is set for mammography
in Mode 7, four
adjustment
points
at 20, 25, 32 and 45 kV are required.
The
are

adjustment
points
for
20, 25, 32 and 45 kV.

mammography

The line
adjustment
should
be at 25 kV.

for

mammography

1 20kV 1 25kV ( 32kV 1 45kV


20mA 1
I

50mA
lOOmA

lOOmA

.
.
.
800mA 1

800mA

Adjustment
for 20,
45 kV in Mode 1.

25,

32 and

Adjustment

for

25 kV in Mode

2.

<Mode l>
tubes)
is set before
shipment.
The standard
data
(for
three
Correct
the data depending
on combinations
of tubes.
Since
the preheating
and preflash
are not checked,
adjust
the
(Generate
X-rays
current
after
the filament
heating
is stabilized.
by pressing
the READY switch
and the X-ray
switch
at intervals
of
Take note
(Do not press
them simultaneously.
more than 2 seconds.
of the heating
data as shown in table
10. (1)
in APPENDIX 01 after
adjustment.)

24F -

No.

<l>

Set Mode 1 with

<2>

Set the
operation

:I
<3>

Set

the

mode

selector

switch.

tube number (1, 2, or 3) to


panel
of the X-ray
control

R.kV

and

R.mA to

be

23313-OOZE*5

adjusted

in

be adjusted
unit.
the

via

the

same way.

Some

combinations
of R.kV-R.mA
are limited
to Maximum Power (30
kW/50 kW).
R.kV:
50, 60, 90, 120 kV or 20, 25, 32, 45 kV
R.mA: 10, 13, 16, 20, 25, 32, 40, 50, 64, 80,
100, 125,
160,
200,
250,
320, 400,
500, 640 mA

L- <4>

Correct

the

data

with

the

y/
cl ;

(increase)

switches
(Note
1) by generating
X-rays
set conditions
in order
to adjust
the
current.
Adjust
all
conditions
to be

or

/(/I
n

(decrease)

under the individually


radiographic
tube
used in this
way.

<5> Press
the Ready SW and the X-ray
SW simultaneously
the set conditions
again
to see if it is necessary
preheating
or calibration
of preflash
level.

and confirm
to perform

if the tube voltage


is stabilized
<6> After
checking
the above,
Item
(5)
Preflash
level
adjustment
within
a specific
time,
and (6)
Preheating
level
adjustment
are not required.
<7> When completing
Mode 0.
Note

1:

press

adjustments,

switch

When the
increased
the

[q

the

or

decreased

or [q

switch

is

ten

times

is

used.

S/R

switch

to

used,

the

as

compared

return

value
to

can

to

be

when

<Mode 2>
The adjustment
in this
mode causes
50, 60, 90 and 120 kV to
simultaneously
increase
or decrease.
Mode 2 adjustment
is the same as Mode 1, except
that
this
adjustment
should
be made with the R.kV tube voltage
set to 60 kV.
When the X-ray
tube for mammography
is used,
the tube current
adjustment
in Mode 2 should
be done at 25kV.
(5)

Preflash

level

adjustment

(PRF):

Mode

The preflash
level
should
be adjusted
only when the tube current
does not rise
within
the specified
time in (4) <6>.
Preflash
is intended
to increase
the X-ray
tube filament
rise
rate.
the preflash
is done prior
to
After
the READY switch
is pressed,
main heating.
/
Preflash
.
Main heating
I

Heating level

Preheating

READY mON ))

- 24G -

Time

No. 25313-002E*J

and the amount of


the preflash
time is fixed
For this
unit,
preflash
can be obtained
by multiplying
the amount of main heating
by the preflash
factor.
the preflash
level
is adjusted
by changing
the preflash
Therefore,
factor.
is generated
by preflash
factors
at three
The preflash
factor
curve
The three
points
are the maximum, minimum and middle
points.
points
of the filament
current
on the X-ray
tube current
characteristics
curve.
PRFl,
PRF2 and PRF3 are to be selected
The three
preflash
factors,
with

the

[d,

adjustment
Tube

Tube

and [z]

14,

Current

Characterist

switches.
its

current

Pre flash factor

WI

Filament
PRFl, PRF2,
selected
as

Maximum

Middle

Mi nimum

Current

and PRF3 for


follows:

the

(A)

three

adjustment

points

can

be

Plot
the X-ray
conditions
under which
the preflash
level
is to be
If the
on the tube current
characteristics
curve.
adjusted,
If the
filament
current
is near the maximum value,
adjust
PRFl.
filament
current
is near the middle
or minimum value,
adjust
PRF2
or PRF3, respectively.
If the filament
current
is between
the minimum and middle
values,
If the filament
current
is between
the
adjust
PRF3 and PRF2.
middle
and maximum values,
adjust
PRF2 and PRFl.
<l>

Set

<2> Set

Mode

5 with

conditions
the

<5>, via
<3> Select

the

m,

[a,

PRF3

PRFZ

sequence

the

mdoe

under
operation

preflash
or

proceeds

selector

which
the preflash
level
is
panel
of the X-ray
control
factor

[a

switch.

to

switch.

be

adjusted

The mode

PRFl
to

the

next

<

24H -

step.

by

indicator

short
unit.

pressing
blinks,

in

(5)

the
and

the

No.

<4> Change the


(Note
1).
L<5>

preflach

factor

by using

S/R
Note

or

switch

Increase Decrease
the indicator
In this
state,

When the S/R switch


is pressed,
and the state
in <2> is reset.
factor,
another
focus
or another
adjusted.

<6> When the

[z]

the

25313-002E*J

adjustment

is

tube

preflash

reset

completed,

stops
blinking
another
pref lash

Mode

factor

can

be

0 by pressing

the

switch.
1:

When the

[a

or

[q

Increase

switch

is

pressed,

the

buzzer

Decrease

will
sound and the preflash
factor
will
increase
or
If either
switch
continues
to be
decrease
by 1%.
pressed,
the buzzer
will
automatically
start
sounding
so that
the result
can be achieved
as if the switch
were pressed
by the number of times
the buzzer
sounded.
(6)

Preheating

level

adjustment

(PRH):

Mode

If the adjustment
This adjustment
is not necessary
in principle.
in (5) cannot
be carried
out satisfactorily
depending
on the
this
adjustment
should
be carried
out as an auxiliary
system,
adjustment
to (5).
<l>
---+<2>

Set

Mode

5 with

the

Set the conditions


not satisfactory,
unit.

<3> When the


the
<4> Press

mode

switch.

under which the


via
the operation

switch

sequence

selector

is

proceeds

pressed,

to

the

switch

the

preflash
panel

the

next
to

mode

of

level
the

adjustment
is
X-ray
control

indicator

blinks

and

stop.
adjust

the

preheating

level

Increase Decrease
(Note

l),

th e

and check

<5> When the S/R


and the state
another
tube

1:

current

rise

by generating

the indicator
switch
is pressed,
In this
state,
returns
to <2>.
preflash
factor
can be adjusted.

<6> When the adjustment


S/R switch.
Note

tube

is

When the[z]or

completed,

14

switch

reset

is

Mode

pressed,

X-rays.

stops
blinking
another
focus

0 by pressing

the

buzzer

or

the

will

If either
sound and the preheating
level
will
change.
switch
continues
to be pressed,
the buzzer
will
can be
automatically
start
sounding
so that the result
achieved
as if the switch
were pressed
by the number of
times
the buzzer
sounded.

241

25313-002E*J

No.

(7)

Changing

the

Set

Size

Mode 6 with

the

<2> Use the operation


the number of the
be changed.
<3>

Press

the

mode,

enabling

<4> To adjust
switch

for

Fluorography:

Mode 6

to change
the focal
size
procedures
permit
the small
focus
(SF) at the time of shipment.

The following
adjusted
to
<l>

Focal

the

the

For

the

the

Ed
.

to

place

following

focal

size

pressing

small

selector

switch.

panel
of the X-ray
control
tube whose fluorographic

switch

yh
cl

while

mode

the

mode

unit
to specify
focal
size
needs

indicator

in

to

blinking

steps.

to

the

large

focus

(LF),

tilt

the

S/R

switch.

the

focus

already

(SF),

tilt

the

S/R

switch,

while

pressing

switch.

A buzzer
procedure

sounds
and the indicator
stops
has been performed
correctly

<5> Reset
to mode 0 by means of the
size
adjustment
has been completed.

To change

the

focal

size

to

S/R

LF [Z]

switch

14
Hold

To change

the

focal

size

to

blinking
so far.

for

tube

current

The dropout
tube current
can
However,
into
the generator.
level
is too low or too high,
procedures
below.

drop:

Mode

Select

<3> Enter

the

<4> Press

the

tube

current

I/

switch.

I/a

mode
which

selector
is

The mode

245

focal

fl

3
switch

Ti;Tt

down

this

switch

Tilt

be compensated
by entering
a tube code
if the preset
compensation
current
it can be adjusted
by following
the

<2>

the

this

the

On the operating
panel
of the generator,
for which the preset
compensation
current
adjusted.
6 with

[a

the

SF

cl>

Mode

when

down

Hold
Compensation

if

to

specify
level

the tube No.


needs
to be

switch.
be corrected.

indicator

starts

(See

Note

blinking.

1. >

No.

<5> Change

the

(increase

preset

compensation

current

ecrease)

> and

level

switches.

<6> When the current


level
is correct,
the indicator
stops
blinking.

using

(See

set

the

S/R

25313-002E*J

the

Notes
switch

yd
cl

2 to

4.)

down;

that
the adjusted
current
level
for
<7> Generate
X-rays
to confirm
(See Note 5.)
the dropout
tube current
is correct.
Repeat
Procedure
<4> if the compensation
current
level
requires
or repeat
from Procedure
<3> if it is
further
adjustment,
the compensation
current
level
for other
necessary
to correct
tube currents.
<8>

(9)

Set

the

Note

1:

Note

2:

Note

3:

Note

4:

Note

5:

Recovery
If
<l>

S/R

down

to

set

When either

the

y
[Y]

Mode

0.

switch

Or [Zl

is

pressed

once,

the buzzer
sounds
once and the compensation
current
If the
level
is increased,
or decreased,
by one step.
will
automatically
switch
is held
down, the buzzer
sound at fixed
intervals;
the number of times
the
buzzer
sounds
is the same as pressing
the switch
the
same number of times.
When the compensation
current
level
value
reaches
the
the buzzer
emits
alternating
upper or lower
limit,
short
and long tones.
time should
be longer
than 200ms.
The X-ray
exposure
any

error

data

has

error
recover

occurred,

When the S/R switch


is tilted
alarm stops
and Mode 7 is set.
switch
recovery

to

Tube current
values
of 20, 50, 100, 200,
500 and 64OmA
can be entered
in Procedure
<3>.
The tube voltage
should
be set to 90kV.
Increase
the compensation
current
level
to correct
for
a fall
in tube current
during
X-ray
exposure
and
decrease
it to correct
for a rise
in tube current.

from

a data

switch

routine

is

pressed,

with

while

the

the

the

following

alarm

indicator

procedure:

sounds,

the

blinks

and

the

starts.
switches

are

pressed

simultaneously,

the indicator
goes
out and the recovery
routine
(for
Upon
regeneration
of an error
check
code)
is executed.
the indicator
lights
up and the
completion
of execution,
sounds.

24K -

buzzer

No.

25313-002E*J

<4> Reset Mode 0 by tilting


the S/R switch
and check
the data.
If
the power supply
is turned
off
in the adjustment
mode during
data is judged
to have be destroyed
tube current
adjustment,
when the power supply
is turned
on again,
and the alarm sounds.
(This
is because
it is not possible
to distinguish
data change
from data breakdown.)
In such cases,
recover
by using
the
data
adjusted
prior
to
After
recovery,
above procedure.
turning
Note:

(10)

off

the

power

supply

are

all

effective.

The recovery
routine
is simply
to regenerate
the error
but not to repair
the destroyed
data.
check
code,
it is necessary
to check
the data after
Therefore,
recovery.
If the check
shows that
the data are changed,
correct
it
If data have been completely
in Mode 1, 2 or 4.
In this
it is necessary
to reset
the tube.
destroyed,
the
setting
must
start
with
the
setting
of
the tube
case,
code.

Alarm
<l>

Stored
data check
The stored
data is checked
by heat data when the power supply
Data is checked
in each tube and each focus
(for
is turned
on.
If an
Tube 1 LF, Tube 1 SF . . . Tube 3 SF).
data blocksattempt
is made to use the data block
in which
an error
has
an alarm (intermittent
beeping
sound)
and an
been detected,
To recover
from this
state,
error
message
will
be generated.
execute
recovery
procedure
(9)
from a data error.

<2>

Set condition
check
If the set tube voltage
becomes
greater
than 150kV or the tube
an alarm
(intermittent
beeping
current
is not set correctly,
If the tube voltage
or the tube
sound)
will
be generated.
the alarm will
be removed.
current
is set correctly,

current
check
<3> Maximum filament
The heating
output
data is set to a value
smaller
than the
maximum filament
current
to prevent
the X-ray
tube filament
If the heating
data exceeds
the maximum
from breaking
down.
filament
current
value
when the READY switch
is pressed,
the
heating
level
will
be set to the preheating
level
and an alarm
(intermittent
beeping
sound)
and an error
message
will
be
generated.

24L -

No. 2J313-002E*P

If the alarm sounds, turn off the READY switch decrease the
heating level.
Table 5.2.2

Tube setting

.
No.

Tube name

Binary code

No.

Binary code

Tube name

00 0000

32

10 0000

DRX-3535HC

00 0001

33

10 0001

DRX-3545HC

00 0010

DRX-1203B

34

10 0010

00 0011

DRX-1403B

35

10 0011

DRX-3603HD

00 0100

DRX-1415B

36

10 0100

DRX-3624HC

00 0101

37

10 0101

DRX-3634HC

00 0110

DRX-1603B

38

10 0110

00 0111

1 DRX-1615B

39

10 0111

DlU+3724HD/HC

00 1000

DRX-1725B

40

10 1000

DRX-3745HD/H

00 1001

DEU-1924HD

41

10 1001

10

00 1010

DRX-1724E

42

10 1010

11

00 1011

43

10 1011

12

00 1100

I DRX-2603HD

44

10 1100

( DRX-4234HC-S

1 DRX-2625C

45

10 1101

DRX-4634HC

13

00

1101

14

00

1110

15
16
17

/
1

18

46

00 1111

1 DRX-2703HD

47

01 0000

1 DRX-2724HD

48

10 1110
10 1111

ISAPPHIRE 150TH

11 0000

IDXB-0335CS(A197)

11 0001

/DXB-0324CS(A192)

01 0001

49

01 0010

50

11 0010

DXB-0434ES(A256)

DFU-2825C

51

11 0011

DRX-B1124B-MO

52

11 0100

DRX-B1146B-MO

11 0101

DRX-B1356EC-MO

19

01 0011

20

01 0100

21

01 0101

DRX-2903HD

53

22

01 0110

DEU-2924HD

54

23

01 0111

24

01 1000

56

11 1000

25

I 01 1001

57

'

11 1001

DRX-8316HD

26

01 1010

58

11 1010

DRX-8336HD

55

11 0110
11 0111

27

01 1011

DRX-3103D

59

11 1011

28

01 1100

DRX-3224D

60

11 1100

29

01 1101

DRX-3224HD-H

61

11 1101

30

01 1110

62

11 1110

31

01 1111

63

11 1111

DRX-3335HC

- 25 -

No.

5.2.3

Radiographic
(1) Prior
Prior
table

tube voltage

2J313-OOZE*1

adjustemnt

adjustment
to X-ray
5.2.3.

exposure,

Table

Power

set SW1 on the INV-2

PWB

according

to

5.2.3

supply

Setting
of SW1

position

Single-phase
200 V
(190 to 215 V>

Single-phase

220 V

(216 to 240 V)
Three-phase

380 V

(380 to 400 V)
Three-phase

415 V

(401 to 425 V>


Three-phase
(426 to

440 V

Set SW5 on the INV2 PWB according to the table


is required for PWBs without SW5.j

Length of the longest


among those connected
the three X-ray tubes

cable
to

Setting
of SW5

to 15m

r-

position

3, 2, 1 or 0

16 to 20m

2, 1 or 0

21 to 25m

1 or 0

26 to

- 26 -

below.

(No setting

No.

(2)

Tube

volt

age

25313-002E*I

adjustment

Perform
tube voltage
adjustments
for each tube current
setting.
Although
tube voltatges,
30 kV, 40 kV, 60 kV, 90 kV, 120 kV and 150
kV, are to be adjusted
in principle,
any voltage
between
tube
voltage
sections
(see
Note 2) can be adjusted.
If the tube current
value
differs
greatly
from the value
preset
by
means of the operation
panel,
the tube voltage
cannot
be adjusted.
In such cases,
therefore,
the tube current
adjustment
should
also
be made along
with
the tube voltage
adjustment.
The delay
time of the tube voltmeter
should
be 20 msec.
[Adjustment
<li

procedure]

Determine
how many kV should
measured
tube voltage
value
within
the tolerance.

be increased
in order
to

or decreased
from
put the tube voltage

the

<2> Look at the display


of 7--segment
LED on the INV-2 PWB. This is
tube voltage
shift
data and stored
into
the RAM.
If an
increase
or decrease
is perfomed
around
this
numeric
value,
the
tube voltage
will
also
be increased
or decreased.
One step
Thus,
corresponds
to about
2 kV.
determine
a value
to be
written
anew.
3 (rotary
switch)
on the INV-2 PWB to
<3> Ad just
SW&
determined
in <2>, then press
SW3.
Consequently,
Check
that
the
display
of
7-segment
rewritten.
to this
value.
<4> Generate
X-rays
tube voltage
is
above
from Item
0

l-i

L!

I
i--

the

value
the RAM is
LED has changed

again
and measure
not found within
(1).

the tube voltage.


If the
the tolerance,
repeat
the

7
-i

_I
I

8
I-i

I I2

26A -

I1
__I

10

11

17

I_

l-i

1-1

12

13

I-l_

14
III

/----

15
I-I-

No.

[Example]

Set 60 kV
(1) Determine
(2) Display

Measure

100 mA.
a decrease

value:

2J313-002E*I

66kV

of 6kV.

of LED

If 7 (H) which is 3 less than this


value is displayed,
a decrease of
6 kV will be accomplished.

(A (H) is an example

value.)

Set SW2 to 7 (H).


Press

The display

changes

to

7.

7(H)

Supplement:

When any value from 40


to 63kV 1OOrn.Ais set,
7(H) is displayed.
A decrease of 6kV is
equally accomplished
within this section
(Refer to note 2)

When SW3 is set in the upward


position, a point lights up
at the lower right corner of
LED as shown on the left.
This point indicates "display
value +1/2 (H)".
When this
point lights up, the tube
voltage is 1 kV greater than
when it does not.
is made

7 + l/2
(H)

Note

1: A tube voltage

Note

2: A tube voltage adjustment


is made for each tube
a tube voltage shift
Therefore,
voltage section.
is performed equally within the same section.

- 27 -

adjustment

SW3.

for each

tube.

No.

Tube voltage
section

20kV to
31kV

Adjustment
value

30kV

32kV to
47kV

48kV to
63kV

64kV to
95kV

96kV to
127kV

128kV to
150kV

90kV

120kV

140kV

60kV

40kV

2J313-002EycB

Note

and an
3: When the tube voltage is not controlled,
OVER kV message is displayed,
turn off the power
switch, and turn it on again after about 30
seconds.
In addition, prior to next exposure,
and other necessary
perform tube current adjustment
countermeasures.

Note

4: If LED2 (kV+7% OVER") on the INV-1 PWB lights up,


If the
adjust the :ube current, tube voltage, etc.
next exposure is correctly made, LED2 goes out.

are generated on the tube


If the following waveforms
voltmeter
under the light load conditions
of 5OmA or less
check TP2 ("ACT kV" 20kV for tube voltage adjustment,
IV"> within the INV-1 PWB using a storage oscilloscope
for
waveform comparison.

Tube voltage
waveform

TP2

T Fine-adjust
it CCW

VR3 by turning
(towards "1").

i
Gradual

increase

T Fine-adjust

(2)
-we--_

I-

Gradual

decrease

it CW

-I

_L

(3)

F=
Gradual

decrease

VR3 by turning
(towards "3").

Gradual

decrease

28 -

If a similar waveform is
observed, do not turn VR3.
I
In a single-phase
2OOV
i
system, the waveform may
show a gradual decrease
I
under a large load.
I

No. 25313-002Ek4L

5.2.4

Checking

the

Check that
radiographic

radiographic

time
correctly

the radiographic
timer
operates
times
as described
below.

(1)

terminal
Connect
the monitor
output
input
terminal
of the oscilloscope

(2)

Turn

power

(3) Select

the

tube

to

which

kV, mA, and


X-ray.

measuring

time

(6) The maximum radiation

Definition

power

and

the

defined
time

as

as

specified

shown

tolerance

is

if

is

in

the

Fig.
given

Item

connected.

7.2.2 (2), and

oscilloscope,

and

5.2.4.
in

6.3.

off.

of

The radiation
increases
to
decreases
to

is

values

instrument

Read each radiation


time from the waveform
on
check
that
it is within
the allowable
range.
radiation

time

the

(5)

Turn

set

on.

Set the
generate

(7)

several

of the tube voltmeter


with a coaxial
cable.

(4)

The

for

radiation
time

is

75% of its
75% of its

time
the interval
peak value,
peak value.

from Point
A, where the tube voltage
to Point
B, where the tube volt age

/ 100%

75%

75%
\

High potential
waveform

Tolerance

Tolerance
Fig.

5.2.4

29 -

No.

5.3

Checking

the

Automatic

The followings
is installed.

5.4.1

check

should

(1)

Press

the

(2)

Check
Refer

that
the settings
to the AEC-lOA/lOB

(3)

Perform
manua 1,

Set

Set the
If this
timer.

Start

as

select

is

adjustment

control

is

(AEC)

performed.

the

AEC-1OA/lOB

installation

below.

kV and mA values.
than an appropriate
AEC radiation
time.
functions
as an AEC backup
or desk timer

generation.

(KXO-30F/50F)

instrument

connect

the

exposure

AEC.

in

given

Adjustment

Measur ing

X-ray

specified

time value
longer
the panel
occurs,

X-ray

Fluoroscopic

to

automatic

of the AEC unit


are correct.
installation
manual,
2C376-519E.

procedure

appropriate

when

after

TECHNIQUE button

the AEC test


26376-519E.

Fluoroscopy

Control

only

be performed

radiation

Connect
5.4.2

necessary

The

The

5.4

are

Exposure

2J313-002E;kB

measuring
tube

ion

instrument

current

as

adjustment:

specified
Mode

in

5.2.1.

can be adjusted
individually
point
In Mode 4, each adjustment
setting
the f luoroscopic
tube (F.kV)
or all
data concerning
fluoroscopic
tube current
can be increased
or decreased
at

When F. kV is set to 60kV in Mode


4, all
data can be increased
or
decreased
at one time.
Any F.mA
setting
can be made.

When F.kV is set to 50, 90, or


120kV in Mode 4, each adjustment
point
can be adjusted
separately.

5OkV /

90kV

2mA

I
I

I
!

3rnA

50kV

/ 120kV

lmll

by
the
one time.

30 -

90kV

/ 120k

253 13-002E+B

No.

the

Therefore,
adjusting

points

with

and

120kV

For

the

The

heating
data
after
in APPENDIX 01.

11.
(1)

tube

fluoroscopic

all

adjustment

F.kV

to

set

be

can

60kV

and

facilitated

then

by

adjusting

first

50,

90,

points.
operation

Select

switches

(T u b e

SPOT

and

adjustment
should

adjustment

1)

via

the

modes,
refer
to 5.2.2
(1).
be recorded
as shown
in table

operation

panel

of

the

X-ray

control

unit.
(2)
(3)

Set

Mode

Set

the

4 with
F.kV

and

part
of
the
Combinations:

(4)

F.kV:

50,

90,

05,

1,
the

switches

Note

3,

values

yA
cl

by

order

generating

If

the

X-ray

correct

tube

Operation
is
(1)

the

is

completed,

or

or

tube

voltage

tube

current

below

VR6

[z]

tube

(2)

is

(VR6
VR6

within

7% of

If

tube

the

(1)
above,
radiographic
possible
the

off,
In addition
shown

,&
Cl

under

operation
of

on

ten

over,

this

Mode

is

F-unit.

pressing

the

as

compared

not

tube

adjustment

PWB)

the

INV-1

PWB so

all

value

be

can

to

possible

current

must

procedure

mA and

that

under
in the

S/R

be

when

the

and

to

obtain

be

adjusted.

voltage

operation

the

measured

the conditions
fluoroscopic

2,

(2)

kV values,

check

that

value
of
mode.

of

90kV

they

the

and

are

within
7% after
adjustment
PWB as in the case
with
the
This
adjustment,
however,
is

adjustment.
When this
mode
fluoroscopic
mode.
tube
voltage
adjustment
is possible.
is performed
for
each
voltage
adjustment
1,

the

value.

voltage
accuracy
is not
yet
use SW2 and SW3 on the INV-2

0.5,

the

procedure.

INV-1

kV accurate

at

by

set
current.

respectively.

may
the

on

the

this

used,

times

used,

it

case

main

setting

tube

individual
tube

only
in the
radiographic

below

spot
part

(decrease)

fluoroscopic

switch

voltmeter
becomes
90kV+lkV
LED4 should
indicate
sT8

1mA.
By manually

and

reset

is

adjustment

adjustment

Fine-adjust

the

operation

adjustment

the

is

supplementary

the

decreased

In

voltage.

(1)

with

the

X-rays

ad just

switch

Fluoroscopic

adjusted
or

(increase)

to

[z]

the

increased

5.4.3

be

table

4mA

adjustment

When

switch.

to

R/F

with
1)

in

1:

selector

120kV
2,

(Note

When the
switch.

F.mA

data

conditions
(5)

mode

conventional

F.mA:
Correct

the

3 and

4mA respectively.

is

turned
section

No.

Tube

voltage

As

for

section

50kV

operating

procedure

not to
Note : Take care
at the boundaries
5.5

to

63kV

of

64kV

SW2 and

to

SW3,

obtain
the very
at each section.

95kV
refer

2J313-002E*B

96kV
to

discontinuous

to

127kV

5.2.3.
tube

voltage

HHS Tests
After
all
of the preliminary
and final
tests
to be necessary
for
the
installation
system
have been successfully
completed,
the HHS test
phase
may be performed.
The

5.5.1

HHS test

is

Fluoroscopic
(1)

Refer

necessary
entrance

to

Refer

to

Automatic
Refer

to

and

Automatic
Refer

to

the

ORH compliance.

required

and

rejection

limit

are

shown

required

and

rejection

limit

are

shown

linearity

APPENDIX 04

exposure

equipment

control

limit

APPENDIX 06

The test
method,
APPENDIX 06.
5.5.4

checking

rate

equipment

The test
method,
in APPENDIX 04.
5.5.3

finally

APPENDIX 03

Reproducibility
(1)

when

exposure

The test
method,
in APPENDIX 03.
5.5.2

only

exposure

equipment

minimum

required

and

exposure

time

required

and

rejection

limit

are

shown

in

rejection

limit

are

shown

in

APPENDIX 07

The test
method,
APPENDIX 07.

equipment

32 -

No.

6.

SYSTEM

ACCURACY

Technique

No.

Applicable

factor

/ Maximum allowable
deviation
I

range

(Radiographic

TRadiographic
3

mA

all

I
I

-+lO%

kV

all

27%

-+3ms

120%

24 to 99ms

I
I

-+20%

0.1

to 9.999sec

-+lO%

0.2

to 2.9 mA

-+0.2mA

3.0

to 4.0 mA

-+lO%

/ Radiographic
time
Setting and
indication
in A.E.C.

1 Setting and
indication
in
Ii
I !ABC

7.

7.1

3 to llms for lOOmA

,
I

for

'Fluoroscopic
kV
Setting and
indication
in ABC
) Fluoroscopic
kV
, time setting
/ and total time
/ indication

ATTACHED

lOOmA

or more
or more

12 to 23ms

4 -/ Fluoroscopic

2J313-002E*B

all

-+7%
I
/
I
I

I
I

+o
_%
-3

1
/

DATA

PWB Setting

and Adjustment

Tables

PWBs which are not listed in 7.1 require no setting and adjustment
For the automatic exposure control PWB, refer
they are installed.
AEC-lOA/lOB
installation
manual.

- 33 -

when
to the

No. 2J313-002E":D

(Panel)
[PANEL CONT l] or
TW CONT] l/8

PWB name:

A:

KXO-30R

B:

KXO-3OF/5OF

Equipment

When

Setting and adjustment

installed

B
0

VRl:

Adjusting the backup battery operating point


There is not a VRl for KXO-30R=B.

Unnecessary

SWl:

Setting the serial transmission baud rate

Unnecessary

Switch settings

I
2

ON

Baud rate

OFF

OFF

ON

1200

ON

/ OFF 1 ON

2400

OFF

/ OFF 1 OFF

4800

OFF

9600

OFF

ON

ON

[W CONT SUB] l/7 or


[W CONT
2/8

PWB name

OFF
ON
,

Equipment
B

/
I
i-

Setting the X-ray generation buzzer timing

Jumper Jl:
l

Open:

Short:

The buzzer goes on during radiation.


goes on for approximately
after the radiation is completed.

The buzzer

Necessary

50msec

L
ON
OFF

Necessary

Setting the system

sw2:

EP
1 2

'&El
3
4

PU
6 7

Turn this switch ON only for USA.


NOTE (Set this switch to OFF)
MAMMO
TOM0
GENE
BKY 2
BKY 1

Turn switches used


for the individual
TECHNIQUES ON.
(Set the other
switches to OFF.)

Unused (Set this switch to OFF.)

NOTE
_
This switch can be used for 100% of OLP testing only in the USA.
ON: 100% Rating
OFF: 90% Rating
100% Rating can be used for only testing; this switch shouid be

turned

OFF after

completing

the

- 34 -

100% Rating

test.

No. 2J313-002E9cD

PWB name:

[W CONT SUB]
TW CONTl 4/8

3/7 or

A:

KXO-30R

B:

Equipment

When
installed

Setting
A
0

B
j sw4:

KXO-3OF/5OF

For BKY 2 TECHNICQUiZ

Necessary

I
0

sw5:

For GENE

TECHNIQUE

Necessary

SW6:

For TOM0

TECHNIQUE

Necessary

For MAMMO
Contents
settings

Necessary

TECHNIQUE

of these
of SW3.

switches

are

- 36 -

the same as the

No. 2J313-002E>kD

PWB name:

[W CONT SUB] 4/7 or


[W CONT] S/8

KXO-30R

A:

B:

KXO-30F/SOF

Equipment

When
installed

Setting

Sw8:

Necessary

Setting Tube 1

ONP1

OFF

Cl 51 Q

-2

LL

X-ray tube type (12


types max.>
Refer to next page.
ISetting

the upper limit of


the X-ray tube voltage.
These switches can set tube
voltages 15kV (maximum) lower
than the maximum X-ray tube
voltage.

l-i-4
SW No.

Upper limit

1 OFF

1 OFF

1 Max. -0kV

1OFF

/ ON

1 Max. -5kV

ON

OFF

Max. -1OkV

ON

ON

Max. -15kV

Setting the starter mode

SW No.

--r---l
1

1 ON

Mode

1 ON 1 Automatic change-mode

1 OFF 1 ON 1 High speed only mode

] OFF 1 OFF1 Normal speed only mode

I
- 37 -

No. 2J313-002E"N

Sw8 (for USA and CANADA)


SW No. 5

SW No. 6

SW No. 7

SW No. 8

CODE

TUBE TYPE

ON

OFF

ON

ON

DXB-0434ES(A256>(USA>

ON

ON

OFF

ON

DRX-3724HD/HC(USA)

ON

ON

ON

OFF

DRX-3624HC(USA)

ON

ON

ON

ON

DXB-0324CS(A192)(USA)

SW8 (for other count ies except USA and CANADA)


SW No. 7

SW No. 8

CODE

TUBE TYPE

OFF

OFF

DXB-0324CS(Al92)

ON

OFF

DRX-3224HD-H

OFF

OFF

ON

ON

DRX-3224HD

OFF

ON

OFF

OFF

DRX-2924HD

OFF

ON

OFF

ON

DRX-3724HD

OFF

ON

ON

OFF

DRX-2703HD/HC

ON

OFF

OFF

OFF

DRX-1924HD

ON

OFF

OFF

ON

DRX-1603B

ON

OFF

ON

ON

DXB-0434EScA256)

ON

ON

OFF

OFF

DRX-2724HD

DRX-1724B

ON

ON

PWB name:

ON

ON

A:

[W CONT SUB] 5/7 or


[W CONT] 6/8

Equipment
Setting
A
0

B
sw9:

Setting Tube 2

- 38 -

KXO-30R

B:

KXO-30F/50F

When
installed

No. 2J313-002E"D

[W CONT SUB] 6/7 or


[W CONT] 7/8

PWB name:

A:

KXO-30R

B:

KXO-30F/50F

Equipment
Setting
A

B
SWlO:

Necessary

Setting the display time of actual R-time


when AEC is ON

-This
value is set
before shipment.

/
sw11:

0
1

Setting the UP/DOiJN key speed

I
I

-~
-

39 -

Necessary

No. 2J313-002E*K

PWB name:

[W CONT SUB] 7/7 or


[W CONT] 8/8

A:

KXO-30R

B:

KXO-3OF/5OF

Equipment

When
installed

Setting
A
0

B
sw12:

Limiting the output


tions of the user's

SW setting

depending on the condipower supply

Necessary

Contents

F . . . . . . . No limit to power
up to 40kW

supply

0 . . . . . . . No limit to power
up to 30kW

supply

1 , . . . . . . OLP
at

for 51 msec or more


25 to 30kW

2 .,...*. OLP for 51 msec or more


at 20 to 30kW

sw13: Setting

the exposure-time
offset value
The offset value added to the preset exposure
time is the actual exposure time.
Added time (msec>
= O.Lt(Msec) x (setting

SW value)

The offset value can be set in 0.4 msec


units in the range of 0 to 6.0 msec.

- 40 -

Necessary

No.

2J313-002E'kB

(Stand)
[DESK CONTROL]

PWB name:

KXO-30R

A:

l/7

B:

KXO-3OF/5OF

Equipment

When
installed

Setting
A
j

SW1

thru SW3:

SWl:
sw2:
sw3:

Baud

rate

for serial

For communication
with the control
For output to the printer
For option RS-232C

Switch

box

settings
Baud

rate

OFF

OFF

ON

ON

1200

Unnecessary

Jumper

Connect
A-B,
DSRA

Unnecessary

transmission

sw4:

jumers

as follows:

RTSA -RTSB,
DSRB

Printer

CTSA -CTSB,

DTRA-

DTRB,

Necessary

connection

(OFF before

- 41 -

shipment)

No.

2J313-002E*B

(Stand)
PWB name:

[DESK CONTROL]

2/7

A:

JLXO-30R

B:

KXO-30F/50F

Equipment

When
installed

Setting
A

B
0

sw5:

Setting

Setting

SW6:

user

Unnecessary

Destination

country

Japan

USA

Countries
other
Japan and USA

3
z
F

for

Setting

than

Undefined

the exposure-time

offset

value

The offset value added to the preset


time is the actual exposure time.
Added time (msec>
= O.Lt(msec) x (setting

exposure

SW value)

The offset vcllue can be set in 0.4 msec


units in the range of 0 to 6.0 msec.

- 42 -

Necessary

No. 2J313-002E

PWB name:

[DESK CONTROL]

3/7

A:

KXO-30R

B:

KXO-30F/50F

Equipment

When
installed'

Setting
A

1 B
1 0
I

sW6:

Necessary

Setting the R-time offset

1
1 The switch sets the offset value of the set time.

(vs actual time)

I
I
,

I
I
I
I

SW7 thru SW12:

Setting the minimum time of mAs for


each TECHNIQUE

I
Necessary

I
/These switches set minimum time of mAs for each TECHNIQUE.
, SW7 :
i SW8 :

SPOT TECHNIQUE
BKYl TECHNIQUE
BKY2 TECHNIQUE
GENE TECHNIQUE
TOM0 TECHNIQUE
MAMMO TECHNIQUE

SW9 :
1 SWlO:
SWll:
sw12:

Switch settings
112

OFF / OFF

OFF / OFF

5
OFF

Minimum time
(msec>

81

OFF OFF

OFF 1

OFF i OFF / OFF 1 OFF , OFF, OFF

OFF

ON

OFF / OFF 1 OFF

ON

OFF (

OFF

OFF

OFF

I
1

1
/ON
1
/ ON

ON

1 ON

ON

1 ON

1 ON
I
/ ON
I
1 ON

ON

ON

/ ON
I
/ ON

ON

ON

ON

/
I

OFF

ON

253

/ ON

ON

OFF j

254

ON

ON

ON

255

- 43 -

No. 2J313-002E*K

PWB name:

A:

[DESK CONTROL] 4/7

KXO-30R

B:

KXO-30F/SOF

Equipment

When
installed

Setting
A

B
0

sw13:

sw14:

01

02

Unnecessary

Output kW limit
Set the limits if necessary
customer's
power conditions
KXO-30F
SW No.

Necessary
due

to the

Limitation

ON

ON

ON

ON

No limit

up to 40kW

OFF

OFF

OFF

OFF

No limit

up to 30kW

OFF

OFF

OFF

ON

OLP

for Slmsec

or more

at 25kW

to 30kW

OFF

OFF

ON

OFF

OLP

for 5lmsec

or more

at 20kW

to .30kW

KXO-50F
SW NO.
Limitation

ON

ON

ON

ON

No

limit

up to 60kW

OFF

OFF

OFF

OFF

No

limit

up to 50kW

OFF

OFF

OFF

ON

OLP

for Slmsec

or more

at 45kW

to 50kW

OFF

OFF

ON

OFF

OLP

for Slmsec

or more

at 40kW

to 50kW

OFF

OFF

ON

ON

OLP

for Slmsec

or more

at 35kW

to 50kW

OFF

ON

OFF

OFF

OLP

for 5lmsec

or more

at 30kW to 50kW

sw15:

ON
OFF

Selectable

APR program

Necessary

setting

00000000
12345678
I

Not us!d

Not used
MAMMO
TOMOGRAPHY

BAS1C

- 44 -

Table Bucky, Wall Bucky


{General, Spot-shot
I

NO.

Equipment

2J313-002E*C

When
installed

Setting
B
0

sw16

to 21:

max. mA code for mAs analysis


when 1.1. is magnified

sw16:
sw17:

upper 4 bits
lower 4 bits

max. mA code for


Normal size (X.1.)

SW18:
sw19:

upper
lower

4 bits
4 bits

max. mA code for


Magnification-l
(1.1.)

sw20:
sw21:

upper 4 bits
lower 4 bits

max. mA code for


Magnification-2
(1.1.)

mA code

(Hex.)

mA (mA)
10

OOH
OlH
02H
03H

13
16
20

OFH

!
320

For single size I.T.,


SW16 and SW17.

set the max.

- 44A -

mA code with

Necessary

No.

PWB name:

A:

[DESK CONTROL] 5/7

KXO-30R

B:

23313-002EAG

KXO-30F/50F

Equipment

When
installed

Setting
A

B
0

SW28:

Selecting TECHNIQUE to be operated when the APR


is installed.
(NOTE) Only one of 1 to 7 on this SW can be selected
when SW30 or SW31 is used.
ON
OFF

PO
1 2

0
3

0
4

P
5

RRP
6 7

AL

1
I

//

1
I__

Unused

Necessary

All TECHNIQUES are


operated.
MAMMO TECHNIQUE
TOMOGRAPHY TECHNIQUE
SPOT TECHNIQUE

BKY2 TECHNIQUE
BKYl TECHNIQUE
GEN/CASSETTE TECHNIQUE

SW29:
ON :
OFF:

SW30 through SW31 are valid.


SW30 through SW31 do not function.
(The switch is set to OFF before shipment.)

Jumper Jl:
Short:

sw30:

Necessary

Approving setting

Selecting KATAKANA

indication

Unnecessary

APR data are indicated in KATAKANA.

Copying APR data from ROM to RAM.

If this switch is turned ON, data equivalent to


TECHNIQUE selected by SW28 is copied into the RAM area.
(NOTE) Only one of 1 to 7 on SW28 can be used.

- 45 -

Necessary

No.

PWB name

[DESK CONTROL]

6/7

A:

KXO-30R

B:

KXO-30F/50F

Equipment

When
installed

Setting
A

B
0

Setting required only when


combinations
of DT-GCU/KDU
.

is used with

Setting of positions/organs
corresponding
AU-APR operation
selector switches
Switches

to be used:

6, 7 and 8 combinations of SW13


1 . .. . . . . . ..
2 ..........
3 ..........
4 ..........
5 ..........
6 ..........
7 ..........
8 of SW28
0 ..........
1 ..........
SW32 ...........
SW17 combinations
1 ..........
2 ..........
3 ..........
4 ..........
5 ..........
6 ..........
7 ..........
SW7 combinations
1 to 10, 15 or 20 .
0 to 12 . . . . . . . . . . .

AU-APR

2J313-002E*B

6, 7 and 8 of SW13,
8 of SW28, SW32, SW17

When

to the

and SW7

Contents
for SW1 of AU-APR unit
for SW2 of AU-APR unit
for SW3 of AU-APR unit
for SW4 of AU-APR unit
for SW5 of AU-APR unit
for SW6 of AU-APR unit
the maximum in 1.1. fluoroscopy mode
Contents
AU-APR spot radiography
mode
AU-APR fluoroscopy mode
Data writing
Contents
Selecting the head
Selecting the breast
Selecting the abdomen
Selecting the spinal cord
Selecting the lumbar vertebrae
Selecting the arms
Selecting the legs
Contents
Selecting an organ from those set with SW17
Determining
the maximum mA code for 1.1.
fluoroscopy
Setting
Setting
Setting
Setting
Setting
Setting
Setting

Setting procedure
a
First set the numeric on SW13 corresponding
to the switch
number on the AU-APR unit to be set.
SW17 and SW7 to the
(@ Set SW28 (spot radiography/fluoroscopy),
desired position/organ
numbers.
3 Press the SW32 pushbutton
switch (for write operation).
If the buzzer
the
setting is completed.
4
If
the
buzzer
sounds,
8
a SW setting
error
has occurred,
and go back
does not sound,
to @
.
@
Terminate
the setting
operation
by setting
6, 7 and 8 of SW13
to "0".
@
After the setting operation,
reset SW13, SW28, SW17 and SW7 tc
their original position.

- 46 -

No. 25313-OOZE"0

PWB name:

[DESK coNTRoLl

A:

7/7

KXO-30R

B:

KXO-30F/50F

Equipment

When
installed

Setting
A

B
0

SW32:

SW33:

Unnecessary

Setting the combined conventional R/F table

This switch sets the types of conventional R/F table.

Switch settings
12

Conventional
R/F table

OFF

OFF

OFF

OFF

OFF

OFF

OFF

OFF

No combination

OFF

OFF

OFF

OFF

OFF

OFF

OFF

ON

DT-GCU/KDU/DUA

OFF

OFF

OFF

OFF

OFF

OFF

ON

OFF

DT-GCS

1 OFF

OFF

OFF

OFF

OFF

OFF ION
8

ON

DT-KEL

OFF

OFF

OFF

OFF

OFF

ON

OFF

OFF

DT-KEL-2

OFF 1 OFF

OFF

OFF

OFF

ON

ON

DT-MCS-2

OFF j OFF

OFF

OFF

OFF

ON

OFF
(ON

OFF

DT-MCI-3
AUX.

i
ON

ON

ON

ON

ON

ON

47 -

ON

ON

AUX.

Necessary

No.

PWB name:

[DESK DISPLAY] l/6

A:

KXO-30R

B:

2J313-002EfcB

KXO-30F/50F

Equipment

When
installed

Setting
A

B
0

Jumper Jl:
Short:
Open :

SWl:

P
4

Necessary

Necessary

System setting

ONEl1

OFF

Setting the X-ray generation buzzer


timing
The buzzer goes on for approximately 50msec
after radiography is completed.
The buzzer goes on during radiography.

u u
5

El P
7

ATR
MAMMO

TOM0
GENE
BKY2
BKYl
SPOT

1 Turn
l

switches used
for the individual
TECHNIQUES ON, and set
the other switches to
OFF.

NOTE
This switch can be used for 100% of OLP testing only
in the USA.
ON: 100% Rating
OFF: 90% Rating
100% Rating can be used for only testing; this switch
should be turned OFF after completing the 100% Rating
test.

48 -

No.

PWB name:

[DESK DISPLAY]

A:

2/6

KXO-30R

B:

KXO-30F/50F

Equipment

When
installed

Setting

SWZ: Setting

SPOT RADIOGRAPHY

Necessary

TECHNIQUE

Setting the maximum mA value


small focus) of mAs
1

Setting the maximum mA value


large focus) of mAS

t
Steps

(for the
1
(for the)
/
I

0 to 3
Contents

Set vlaue
0

2J313-002EfcB

. . . . . mAs

decomposition
by the maximum mA
which is determined
by the tube/unit

1 . . . . . mAs decomposition
by mA at a step below
the maximum mA which is determined
by
the tube/unit
2 . . . . . mAs decomposition
below the maximum
by the tube/unit

by m.A at two steps


mA which is determined

3 . . . . . mAs decomposition
below the maximum
by the tube/unit

by mA at three steps
mA which is determined

Setting
ON:
OFF:

the AEC detector


SOLIDET detector
Image intensifier
or thin detector

Turn this switch ON when


AEC is usable
ON:
OFF: AEC is unusable

49 -

photo

using

pickup

AEC.

No. 2J313-002E

PWB name:

[DESK DISPLAY]

3/6

Equipment
,
B
A
/

A:

KXO-30R

B:

KXO-30F/50F

When
installed

Setting

sw3:

For BKY

1 TECHNIQUE

Necessary

sw4:

For BKY

2 TECHNIQUE

Necessary

sw5:

For GENE

TECHNIQUE

Necessary

SW6:

For TOM0

TECHNIQUE

Necessary

sw7:

For MAMMO

Contents
settings

Necessary

TECHNIQUE

of these
of SWZ.

switches

are

the same as the

iI

- 50 -

No. 23313-002E

[DESK DISPLAY] 4/6

PWB name:

A:

KXO-30R

B:

KXO-3OF/5OF

Equipment
A

When
installed

Setting

sw8:

Necessary

Setting Tube 1

ON a

u u a u

OFF

0 E

X-ray tube type (16


types max.>
Refer to next page.
1

Setting the upper limit of


the X-ray tube voltage.
These switches can set tube
voltages 15kV (maximum) lower
than the maximum X-ray tube
voltage.

Upper limit /

/SW.

Setting the starter mode

I SW
1

No.

Mode
,2

I~_ON ; ON
OFF 1 ON

Automatic change-mode

! High speed only mode


I

OFF j OFF / Normal speed only mode

51 -

No. 2J313-002E*P

TUBE 1
TUBE 2
SWlO: TUBE 3 (for USA and CANADA)
Sw8:
sw9:

SW No. 5

SW No. 6

SW No. 7

SW No. 8

CODE

TUBE TYPE

OFF

OFF

OFF

ON

DXB-0434ES(A256>(USA>

ON

OFF

OFF

OFF

DRX-3535HC/HD

ON

ON

OFF

OFF

DRX-4634HC(USA)

ON

ON

OFF

ON

DRX-3724HD/HC(USA)

ON

ON

ON

OFF

DRX-3624HC(USA)

ON

ON

. ON

ON

DXB-0324CS(A192>(USA>

SW8: TUBE 1
sw9: TUBE 2
SWlO: TUBE 3 (for other countries except USA and CANADA)
TUBE TYPE

SW No. 5

SW No. 6

SW No. 7

SW No. 8

CODE

OFF

OFF

OFF

OFF

DRX-1924HD

OFF

OFF

OFF

ON

DXB-0434EScA256)

OFF

OFF

ON

OFF

DRX-2724HD

OFF

OFF

ON

ON

DRX-1615B

OFF

ON

OFF

OFF

DRX-1603B

OFF

ON

OFF

ON

DRX-B1356EC-MO

OFF

ON

ON

OFF

DXB-0324CS(A192)

OFF

ON

ON

ON

DRX-3745HD-H

ON

OFF

OFF

OFF

DRX-3535HD

ON

OFF

OFF

ON

DRX-3224HD-H

ON

OFF

ON

OFF

DRX-3224HD

ON

OFF

ON

ON

DRX-2924HD

ON

ON

OFF

OFF

DRX-3724HD

ON

ON

OFF

ON

DRX-2703HD/HC

ON

ON

ON

OFF

DRX-ll24B-MO

ON

ON

ON

ON

DRX-1724B
i

- 52 -

No. 2J313-002E*F

PWB name:

[DESK DISPLAY I 516

A:

KXO-30R

B:

Equipment

when

Setting
A

KXO-30F/%)F

installed

B
0

sw9:

Setting

Tube

Necessary

SW10: Setting

Tube

Necessary

Contents
settings

of
of

these
SW~.

switches

are

the

same

as

the

*
-

52-1

No. 25313-002E

PWB name:

A:

[DESK DISPLAY] 6/6

KXO-30R

B:

KXO-30F/50F
When

installed
Necessary

Setting the display time of a ctual R-time when


AEC is ON.

SW11 .

/ Setting [ Display time (

1
2
3
4

0
1
2
3
4

1
I

-Th

is va lue is set
fore shipment.

F
-

15sec.

-I-

Setting the UP/DOWN key speed

- 53 -

Necessary

No.

2J313-002E:cA

(Control box)

[CPU1

PWB name:

A:

KXO-30R

B:

Equipment

When
installed

Setting
A

KXO_30F/50F

B,
0

1 Xl

Short

I Unnecessary

X26:

X26-

Al3

X27:

X27-

+5

/ so -

RAM-S6

s2

1
I

Selecting Ml ROM
27128

: Short

Ml address:

0000 'L 3FFF

.- Short

RAM address: 6000 ?J 7FFF

[FH/INV]

PWB name:
Equipment'

Setting
q---J

/ JP-C
I

Open :
Short:

Clock division by three


Clock division by two

(Set before shipment "short")

- 54 -

When
installed
Unnecessary

No. 2J313-002EkD

PWB name:

A: KXO-30R

[ SIO-2

B:

KXO-3OF/5OF

There is no SIO-2 PWB in the KXO-3OR=B, KXO-30F=B or KXO-50F=B.


Equipment

When
installed

Setting

A
0

Jumper:

Specifying the RX and TX buffer addresses

All short

All short

LRZ-D
LRl-C
LTlw---+B
LT2-A

HRZ-F
HRl-E
HTZ-B
HTl-A

Unnecessary

(Set before shipment "short")


PWB name:

[SIO-11 or
[SPIO]

Equipment

When

Setting

installed

Aj
0

There is no Jumper in the KXO-30R=B,


KXO-30F=B or KXO-SOF=B.

Jumper:

Unnecessary

J: Short
T-G:
Short
SWl:

11

PWB name:

Unnecessary

Baud rate generator

Baud rate

Equipment

ON j OFF

OFF

OFF

4800

30R

ON / OFF

ON

OFF

9600

30F/50F

IPIN

Setting

- 55 -

When
installed

No. 25313-OOZE

PWB name:

A:

[IFI

KXO-30R

B:

KXO-30F/50F

Equipment

When
installed

Setting and adjustment


A
0

/ Necessary

Jumpers 52 through 54

IO

, Connect one jumper of 52, 53, 54 corresponding


; the minimum HU depending on combined tubes.

52:
53:
c 54:

:
0

'

;Jl

PWB name:

Unnecessar}

.
.

Short (only for USA.)


Unnecessary

Adjusting VRl through VR4

140 KHU
200 KHU
300 KHU

Jumper Jl

to

[APR IF]

Equipment

When
installed

Setting
A
I

I Jumpers Jl, 52:

These switches set BKYl or BKY2 when using Bucky for


1 TOM0 TECHNIQUE (Jl and 52 should be set
1 simultaneously.)
l

I
I

Necessary

Setting TOMO, BUCKY selection

When using TOM0 Bucky at BKYl TECHNIQUE,


connect C and Bl.

. When using TOM0 Bucky at BKY2 TECHNIQUE,


connect C and B2.
!

- 56 -

No. 2J313-002E"cL

PWB name:

A:

MQI

KXO-30R

B:

KXO-30F/5OF

Equipment

When
installed

Setting and adjustment


A

B
0

Jumper Jl:

SWl:

Necessary

Short for F. Rotation

Necessary

Setting continuous radiography TECHNIQUE

Unused

SPOT
BKYl
BKY2
GENE
TOM0
MAMMO
105mm spot
camera

When selecting the continuous radiography mode for


TECHNIQUES, turn the switches corresponding to the
individual TECHNIQUES ON.
, (The switches of three TECHNIQUES of SPOT, TOMO, and
105mm Spot Camera are set to ON before shipment.)
0

VRl:

Adjusting the set-up time of the normal speed


starter

Unnecessary

This switch adjusts the set-up time of the normal speed


starter. VRl is set to 0.8sec before shipment.

PWB name:

[RY

SEQl

Equipment
Setting
A

Jumper:

Setting the type of stator coil of X-ray


tube
TUBE 1
TUBE 2

o------ ---Q
TUBE 3

If a combined tube is the XS-AA


coil type or the Eimac tube,
short the pins of the tube.

- 57 -

When
installed
Necessary

No. 2J313-OOZE*M

PWB name:

A:

[F DIG] l/3

KXO-30R

B:

KXO-30F/50F

Equipment
Setting and adjustment
A

B
0

SW1 thru SW5:

Setting the dose limit in F kV Auto


mode

These switches should be set for USA and other


countries where the dose limit is imposed.
An example for USA (limited to 2.58 x lo-IC/(kg*min)
{lOR/min) is given below.
These switches set F.kV values which exceed the limit
of 2.58 x 10_3C/(kg*min) {lOR/min) at F.mA values of
I 0.5mA, l.OmA, 2.0mA, 3.OmA, and 4.0mA.

F*kV
SW1 =A
SW2 =B
SW3 = c
SW4 = D
SW5 = E

F mA
When setting the switches. measure F.kV values
exceeding the limti of 2.58 x 10-3C/(kg=min)
(lOR/min} at individual mA values (binary), and set
them in the individual switches.

- 58 -

When
installed
Necessary

No. 2J313-002E*M

PWB name:

A:

[F DIG] 2/3

KXO-30R

B:

KXO-30F/50F

Equipment
Setting and adjustment
A

When
installed

125
110
100
0
85
80

The above figure gives an example in which SW1


through SW5 are set to 125kV, IlOkV, lOOkV, 85kV,
and 80kV respectively.
0

,946, sw7:

Setting the dose limit in Fully Auto mode

/
~ These switches should be set for USA and other
1 countries where the dose limit is imposed.

/( SW7 requires no setting.)

~ In Fully Auto mode , the F.kV and F.mA values are


~ linked, and the F.kV value should be set for dose
( limit control.
1 When setting SW6, measure the F.kV value in Fully
1 Auto mode which exceeds the limit of 2.58 x 10-X
/C/(kg*min) (lOR/min), and set the value (binary)
I in SW6.

59 -

Necessary

No.

PWB name:

[F DIG]

KXO-30R

A:

3/3

B:

KXO-30F/50F

Equipment

When
installed

Setting
B

SW8,

;
i
j

When SW8 is set to "l", a mode is set which


enables two F.ABC parameters
to be set with the
DIP switches (SW11 and SW12 on the ATR.DIG PWB). .

When SW8 is set to any other


mined parameter
is set.

SW9:

Setting the F.ABC


full-auto mode

parameter

value,

1 When SW9 is set to "O", F.kV/FmA


50kV/O.ZmA
to 125kV/4mA.
I
1 When SW9 is set to "l", F.kV/F.mA
; 50kV/0.2mA
to 125kV/2mA.

ZJ313-002E$cB

mode

and the

Necessary

the predeter-

changes

changes

from

from

1
I
/
I

SWlO:

O
)
I
(

F-ON

test

I
1 Unnecessary

switch

The fluoroscopy
can be done by this Test switch
instead of the F-ON signal from the conventional
R/F table.

- 60 -

No.

PWB name:
Equipment

A:

[F ANA]

Setting

KXO-30R

B:

JXXO-30F/50F

When
installed

and adjustment

B
0

VRl:

Setting

manual

F.kV

2J313-002E*L

Unnecessary

control

Connect 5kS2 variable resistors


(same as the F. ABC
control of R/F table) to FASTON terminals Tl, T2,
and T3.
Adjust VRl to set the T2 voltage to +lO.OV
when turning the variable resistor
fully clockwise
(equivalent
to F.kV=125kV
selected).

SWl, VR3,

VR4,

Adjustment

VR5,

VR6,

Unnecessary

VR7

procedure

Set SW1

to "1".

The set mode is used to adjust the D/A converters


(1C and 2C).
In this mode, adjust VR6 and VR7 so that the
voltage at Pll and P12 is set to a value in the
range of 9.961 to lO.OOOV.

Next,

Adjust

VR4

to set the P8 voltage

to - lO.OOV.

Adjust

VR5

to set the P9 voltage

to +9.961V.

Adjust VR3 to set the Pll voltage


in the range of 9.961 to lO.OOOV.

After
"0" .

set SW1 to "2".

to a value

the adjustment
is completed,
reset SW1 to
(If not, the software will not operate.)

Jl: Selecting the jumper for ABC mode


(Video signal type or photomultiplier
. Video signal type
: Jump "V" side
. Photomultiplier
type: Jump "P" side
0

VR8:

Necessary
type)
connector
connector

ABC gain adjustment VR for video signal type


Adjust the ABC gain according
to the system
requirements.

I
- 61 -

ABC

Necessary
for video
signal
type ABC

No. 25313-002E

PWB name:

[ATR DIG] 1/3

A:

KXO-30R

B:

KXO-30F/SOF

Equipment

When
installed

Setting
A

5, 9:

SWl,

Setting AEC backup mAs values

Necessary

For SPOT-SHOT TECHNIQUE


For photofluorography TECHNIQUE
For special TECHNIQUE

SWl:
sw5:
sw9:
+EC

is automatically selected for ATR, and backup mAs


'values can be set by these switches. The switches are
'for the TECHNIQUES given above.

sw1,5,9

ZFE3
12
A

,
,
I

PEI

00
3

:
I

4
4
i
I

25

24

23

27

26

LIP
6
,I

P
7
ih

22

21

8
A
i

20

[Example]

: ds

,;;-;-1

*2
I

2510

30/o

0,o

35

111

!011/0

40 j O'O/
1
0
~45/O'Oll;O
j
50
10
0~1~1~0~0

Ii
1

o/0(1

12i

I
l/O!0

2.

l:f~al

0)1'

19H

1'0

1EH

1: ON

l/

23H

0: OFF,

o/

28H
2DH

I
I
I
!

/
I

11110
1

1
1'01

Max*
)

1'1'1
/

I
lil

- 62 -

/
;

32H

i
iI255

FFH

No. 25313-002E

PWB name:

A:

[ATR DIG] z/3

JLXO-30R

B:

KXO-30F/5OF

E;i.,,;t

When
installed

Setting

SWZ, 6, 10:

SW2 :
SW6 :
SWlO:

Setting the upper limit of R.kV

Necessary

For SPOT-SHOT TECHNIQUE


For photofluorography TECHNIQUE
For special TECHNIQUE
1

These switches set the upper limit of the R.kV value


when converting F.kV to R.kV.
,
I
I

1
I

SW2,6,10

(Jp$
EIPPUUiiP

a,

12
/

Ii

4
1

27

26

25

24

23

22

21

20

34567

[Example]
Ic

Hexadecimal

Binary
kV
.I

I
~

26

125

zL

! 2

70;

0~1~0~0~0~1~1~0~

801

9oi

l/O
1

/ 100 , 0 / l/

12

12 ~

0i

11i~oii~o~

(H)

46H

1101
0

I-9

0101

50H

,
1
/
/
0 1 0 / 11 0 / 01
1

~AH
64H
6EH

~110~0~1/1,0'1/1~1~0'
I
5

:ii

IMax.

o.

oil~lo~
I

150

96H
,

I O
!

jsw3, 7, 11:

I
I
/

;w:

..:

SWll:

Necessary

Setting the lower limit of R.kV

For SPOT-SHOT TECHNIQUE


For photofluorography TECHNIQUE
For special TECHNIQUE

'These switches set the lower limit of the R.kV value


when converting F.kV to R.kV. (The setting procedure
/is the same as the upper limit setting described
iabove.)
I
I

- 63 -

No. 2J313-002EfcB

PWB name:

[ATR DIG] 3/3

-r---j

A:

JLXO-30R

B:

KXO-30F/50F

Equipment
A

i
I

When
installed

Setting

Necessary

SW4, 8: Setting the maximum mA value of mAs


SW4 :
SW8:

For spot film TECHNIQUE


For photofluorography TECHNIQUE

j In ATR mode, the equipment separates the backup mAs


I value which is set in SWl, 5, 9 using the mA value

/which is set by these switches. When performing small


I focus radiography in ATR mode, set this value to the
1 maximum mA value of small focus.

J,- 640mA is not applicable

to KXO-30F.

SW11, SW12: Setting the F.ABC parameters (effective


only when SW8 and SW9 on the F.DIG PWB
are set to "1" and "0" respectively)
Parameter P (for control sensitivity)
/ SWll:
Parameter I (for reference of output
j sw12:
amount)
I

- 64 -

( When
/ necessary

No. 25313-OOZE

PWB name:

A:

[ATR ANA] 1/2

KXO-30R

KXO-30F/50F

B:

Equipment

When
installed

Setting and adjustment


A

Necessary
P5 input voltage ranges from +3.33V (50kV) to +8.33V
(125kV) and P6 input voltage ranges from +0.5V (0.2mA)
to +lOV (4.OmA). The final output voltage (PIO) is
+lOV (150kV) max.
VRl, VR2, and VR3 should not be set because they are
unused.
VR31 is used for radiography at a constant R.kV value
with the ATR function stopped temporarily in SPOT
(Example: Esophagus radiography)
1
Radiography mode.
Adjust the P9 voltage with 26B grounded.

IVR4 through VR7 adjustment (for spot film radiography) /


I

P7 voltage

PlO voltage

IL

Correction
VR7
\
\ i$

with F.mA

When F.d

> 1mA

When
F.,ti < 1Lti

VR4
1

F.kV
,~/@
lVR4
F.kV

Adjust VR5 to set the P8 voltage to OV when F.mA=lmA. !


VR8 through VR13 adjustment (for spot film radiography)i
These VRs correct R.kV values for film division and
image intensifier (1.1.) magnification.
Division and
1.1. magnification
!

Adjustment VR

Division by 2

VR8

Division by 3

VR9

Division by 4

VRlO

Spare

VRll

1
j

Ml

VR12

M2

VR13

iVR14 through VR17 adjustment


(for 1.1. photofluorography)
These VRs have the same functions as VR4 through
VR7 for spot-film radiography.

- 65 -

No. 25313-002EfcB

PWB name:

[ATR ANA]

KXO-30R

A:

2/z

B:

KXO-30F/50F

Equipment
Setting

When
installed

and adjustment

A
VR18 through VRZO adjustment
(for 1.1. photofluorography)
These VRs have the same functions
and 13 for spot film radiography.

PWB name:

as VRll,

12,

VR21 through VR24 adjustment


(for AUX TECHNIQUE)
These VRs have the same functions as VR4 through
VR7 for spot film radiography.
VR25 through VR30 adjustment
(for AUX TECHNIQUE)
These VRs have the same functions as VR8 through
VR13 for spot film radiography.

[HEAT

A:

11

KXO-30R

B:

KXO-30F/50F

When
installed

Setting

j VR2:

Frequency

adjustment

1 VR3:

Frequency

adjustment

/ VR4:

Offset

o 1-0

/Unnecessary
I
I
IUnnecessary

adjustment

A:

[HEAT 21

PWB name:

/
iunnecessary

KXO-30R

B:

KXO-30F/50F

When
installed

Setting

r
I

l/

VR12:

D/A converter

span

Unnecessary

128/256:

128 crossover

span

- 66 -

1 Unnecessary

No.

PWB name:

Adjustment
point

CNNA

Equipment
.

KXO-30R

A:

[INV-11

Contents

B:

2J313-002EkI

KXO-30F/50F
Adjustment
when
installed

of adjustment

Unnecessary

;'(OVER CURRENT detection level setting +8V (TPl)


VR2
KXO-50F=B or KXO-30F=B
(VRl,VR2) * For the KXO-30R=B,
VR3

Tube

voltage

waveform

m
I

Note *: In the case of


single-phase
power supply:
If 160mA or more is set,
the tube waveform shifts
slightly downwards,
this
cannot be adjusted with
If VR3 is used
VR3.
for adjustment,
the tube
voltage waveform may
become abnormal under
a small load.

Tube

VR5

Tube overvoltage
at 165kV

VR6

Tube voltage output adjustment


in fluoroscopic mode
Set the measured
tube voltage to 90kV
with 90kV and 1mA set and LED4 on the
INV-2 PWB indicating "8".

Necessary

VR7

Error signal level


-0.6 to 1OV (TP~)

Unnecessary

VR8

Frequency
adjustment
When Pin 7 at location
IC:
2.5kHz at TP12
Variable-width
timing

20kV

detection

to 1V (TP2)

Unnecessary

VR4

VR9,VRll

gain

In the case
of a singlephase power
supply, the
'""=I
tube voltage waveform may
gradually descend under a
heavy load.

signal

voltage

When
necessary
(fine
adjustment

adjustment

If the tube voltage


waveform under a small
load * (about 5OmA) is
overshoot or undershoot,
Set the
adjust VR3.
waveform by a tube voltmeter to be the same as
that at TP2.

Unnecessary

level

Detection

adjustment

adjustment

- 67 -

Unnecessary

Unnecessary
2D is 5V,

of phase

control

Unnecessary

No. 25313-002E*I

A:

[INV-21 l/2

PWB name:

KXO-30R

B:

KXO-30F/50F
Adjustment
when
installed

Equipment
Setting

and adjustment

D/A converter
adjustment
VRl, VR3, VR5: Zero-position
adjustment
Gain adjustment
VR2, VR4:

Unnecessary

SW2,

Necessary
at tube
voltage
adjustment

SW3:

Switches to write the tube voltage


adjustment
data (kV shift data) in
radiographic
or fluoroscopic
mode
into the .RAM on the PWB.
kV to be set:
For each section 20kV to 31kV,
32kV to 47kV, 48kV to 63kV, 64kV to
95kV, 96kv to 127kV, and 128kV to
150kV, kV shift data can be written
into the RAM for each mA and each
tube.

Necessary

SW1
POWER
source

SW4:

Setting

Position

19 2oov

l$ 220v

3cP 380V

39 415v

39 440v

Tnverter

operation

disable

of SW1

switch

When this switch is set to "DIS" (DISABLE),


LED3 lights up and the INV-UNIT operation
is
set this switch to "EN".
disabled.
Normally,
Note:

If SW4 is set t0 "EN" (ENABLE), LED3


will light up in the following cases:
@

If LV

If SW1

is not

If the

INV unit

During

charging

is too high

or too low

set as specified

- 68 -

fuse

is blown

Necessary

No. 25313-OOZE*1

PWB name:

[INV-21

2/2

A:

KXO-30R

B:

KXO-30F/50F

Equipment
Setting
A

SW5:

(YWA1977

Adjustment
when
installed

and adjustment

only)

Necessary

Length of the longest


among those connected
the three X-ray tubes

cable
to

to 15m

Setting
of SW5

position

3, 2, 1 or 0

16 to 20m

2, 1 or 0

21 to 25m

1 or 0

26 to

68A -

No. 2J313-002E*D

PWB name:

A:

[INV-31

KXO-30R

B:

Adjustment
when
installed

Equipment
Setting

and adjustment

VRl

When the power voltage


Line voltage signal:
has the nominal value, signal at TPl is 5Vp.

VR2

TP2:

VR3

VR4

VR5

I
I
1

cases>

TP3: +4.87V
+4.75V

(at 14 2OOV),
(in all other

cases>

TP4:

(at l$ 2OOV),
(in all other

cases>

+5.5OV
+5.5OV

VR6

TP6:

CNNA

Set to "NEGA"

SW1

Turn

(at l$ 2OOV),
(in all other

= 1V for tube current

Unnecessary
cases>
Unnecessary

lOOmA

1 Ucnecessary

(at lb 2OOV),
(in all other

TP5:

Unnecessary

Necessary

+4.75V
+4.5OV

+5.35V
+5.25V

HO-3OF/5OF

on 1 of SW1 for a single-phase


power
Turn
it
off
for
all
other
power
suPPlY*
supplies.

it is unnecessary to set SWl, for


KXO-30R=B, KXG30F=B
or KXO-5OF.

Necessary

the

69 -

No.

7.2

Error

Code

.k

_k

Code

Table
Meaning

Transmission
and the box

system
error
is not carried

(Data communication
out correctly)

between

the

198

RAM error

(Warning.

RAM on

the

panel

or

desk

is

defective)

199

RAM error

(CPU stop.

RAM on

the

panel

or

desk

is

defective)

201

TECHNIQUE setting

202

Tube No. setting


not set correctly)

301

TECHNIQUE setting

700

Box-side

CPU or

913

1 Normal
starter
/ was not ready)
I

914

; Over MAS
I
1 Filament
heater

88
89
t

2J313-002Ey:B

i Filament
1 greater

switch

(on

tghe

(Tube

error

switch

panel

setting

(on

the

(Current

RAM down

desk)

switch

panel

serial-transmission
error

or

or

on

the

desk)

is

CPU down
was

applied

(Heating

heater
current
abnormal
than the specified
value)

data
(The

is

the

set

panel

not

(Either

when

are

not

panel/desk

or

desk

is

set

CPU malfunctions
normal

starter

abnormal)

heating

current

has

become

90

Filament
hardware

heater
output
is abnormal)

95

Over current
(The
discharge,
etc. )

96

Over

kV (Output

kV value

97

Fuse

blown

fuse

98

Line voltage
smaller
than

(The

section

inverter

of

abnormal

thyristor

has
the

become
INV-UNIT

(Heating

70 -

system

has

malfunctioned

due

greater

than

specified

is

the

to

tube

value

blown)

(The power source


voltage
error
the specified
value
and X-rays

output

has become
greater
cannot
be generated.

or

No.

8.

MAINTENANCE

SCHEDULE
described

The schedule of periodic maintenance


procedures
conducted
for both the X-ray tube and generator
listed below.
First Servicing
Second Servicing
Note:

2J313-002E*B

-------------________

system

here

with

shall

be

the frequency

6 month after installation


1 yaer after installation

All subsequent servicing shall be performed every 12 months


When a new X-ray tube is installed, repeat the first
thereafter.
servicing.

APPENDICES
LIST
APPENDIX

01

Recording

Form

APPENDIX

02

Relations Between Focal


Tube Current Range

APPENDIX

03

Test Method

of Fluoroscopic

APPENDIX

04

Test

of Reproducibility

APPENDIX

05

The Setting Technique


Factor for
Reproducibility
and Linearity Test without

Method

for Tests
Spot

Size and

Entrance

Exposure

Rate

and Linearity

A.E.C.

APPENDIX

06

Test

Method

of Automatic

Exposure

Control

Limit

APPENDIX

07

Test Method
Time

of Automatic

Exposure

Minimum

Exposure

- 71 -

No.

APPENDIX

01

Recording

Form

2J313-OOZE*B

for Tests

This

data sheets are to be filled out and copied at the time of the
Record test results in copied sheets and return
installation
and maintenance.
to:
-------------

U.S.A.

TOSHIBA

MEDICAL

SYSTEM

DEVISION

AMERICA INC. 2441 MICHELLE


CALIFORNIA
92680
The

---------

others

TOSHIBA'S

Service

Date:

2.

Installed
Branch

Maintenance

by:

Office:

Address:
3.

Hospital:
Address:

4.

Serial
0

No. of X-ray

Generator

DC-30R:

q DC-30F:

5.

DC-50F:

17

DG-30F:

DG-5OF:

X-ray

Tube

to be Combined

Manufacturer

6.

Tube

Tube

Tube

X-ray

Table

Model

name

to be Combined

Manufacturer:
Model
Serial

name:
No.:

- 72 -

DRIVE,

TUSTIN,

representative

: Installation,

1.

OF TOSHIBA

Serial

No.

Inspection

No.

X-ray

beam

limiting

device

combined

2J313-002E~kB

with

Manufacturer:
Model

name:

Serial:
8

Power

source
0
0

(1) Power

phase

(2) Power

frequency

q 5OHz
0

(3) Power
G
0
9.

Single phase
Three phase

60Hz

voltage
0 22ov
i? 415V

2OOV
38OV

Preliminary

test

440v

48OV

results
Result

Item
Section
Section
Section
Section
Section
Section
Section

Section

Tube

3.2
5.1.2
5.1.3
5.1.4
5.1.5
5.1.6
7.1

5.1.5

No.

Power Source
Checking voltage
and display circuits
Checking operations
Checking the switching circuit
Checking the starter circuit
Checking the battery
PWB Setting and Adjustment
Tables

Check

on starter

Tube model
name

circuit

Normal speed
boost time

- 73 -

starter
(set>

OK,
OK,
OK,
OK,
OK,
OK,
OK,

NG
NG
NG
NG
NG
NG
NG

No.

10.

Radiographic

KV, mA

and time

2J313-002E*K

test

TUBE NO.
Tube model
High

(1) R.kV

potential

cable

R.Time:

and R.mA

lx

71A
L

kV

mA

length

kV

mA

0.05

kV

set

mA

kV

mA

13
16

- 74 -

No.

TUBE

2J313-OOZE*K

NO.

Tube model
High

R.kV

potential

cable

length

R.Time:

and R.mA

0.05

set

140

120

kV

75 -

mA

kV

mA

No.

2J313-002E*K

TUBE NO.

R.kV

10

and

Tube

model

High

potential

cable

R.Time:

R.mA

kV

length

mA

kV

mA

kV

13
16
20
25
32
40
50
64
80
100
125
160
200
250
320
400
500
640

76

mA

0.05

set

kV

mA

kV

mA

2J313-002E"B

No.

(2)

R.time

R.time

R.kV

0.005sec

60kV

Result

R.mA

(set>

320mA

111
+

lOOmA

90kV

0.2sec

11.

Fluoroscopic

test

(KXO-30F

or 50F)
Tube model
Table

model

F. KV and F. mA

/
I

I
I
I
j

12.

F. ABC

mL4

mA

mA

mA

mA

mA

mA

mA

InA I

mA
3

In U.S.A,
limit.

/
mA

4
Note

m&4 /
I

test

should

be performed

within

the entrance

test

. Thickness

of the acrylic

cm

phantom:

. Peack value of the video signal waveforms:


(including
the signal pedestals)
. Voltage

at P6 of the PTW unit:

requirements:
. Fluoroscopic
(during stable operation
of the ABC circuit)

V
kV,

- 77 -

mA

exposure
,

No.

13.

Automatic
exposure control test
(in combination
with the automatic
AEC-lOA/B)
Record
manual

14.

exposure

2J313-002E*B

.
unit, model

control

test results, using data sheets attached to the installation


(No. 2C376-519E).
for A.E.C. UNIT, model AEC-lOA/B,

ATR adjustment
(1) X-ray
X-ray

(KXO-30F

or 50F)
(FFD

table:

>

tube:

Grid:
Intensifying

screen:

Film:
(2) Set value
R.kV
BACK

kV
kV

Clamp

DOWN:
UP:
UP mAs:

(3) Correction

ITAS

value

Correction

Correction
mA correction

Division
correction

0.5mA
1mA
2mA
3mA
4mA

value
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV

l/2
l/3
l/4
l/6
Ml
M2

- 78 -

No.

(4) ATR

data

in FKV-AUTO

--

--.-

---

i L :;
-.
.
.
:-

;_

I!

;_

__

.:

.I

F_

_. __~

._

____L_~-.-

gheet:~;~___i_I~

ATR.d.ata

.-... . .
_
_.
_-.

-_
..
_.._-.

.
-_--

-----

:I

::f

_~~~:-

s-___._.-----_
-1.
-f -_

--!

._. -.
_.
:-

..
.-

.--.

I
I

_
.

__

__. ..__._.
.
_._
. .

_.

.
_

--i

t-

.
.

. ,
I-

__

-.-

&_...- cm22--.
-_--_
. . -_.
_
____
. _.________I___-_-_
. _
--.--_ _- ---.__. _ ._____-----Acrylic phantom thickness :L.LL--L- :.__________
._._.---.
__ ]
~
_-

li --__l&___---_-16

-_-

(mA: Manual)

._.
.
_.
- ._
. _._-..__
_ _.

._______.&_______

mode

2J313-002E*B

_~____~&-_-__2O__

-L_

T
Acrylic
phantom
12 cm

Fluoroscopic

conditions

mA

kV

14
16
18
20
22
24

79 -

Radiographic

kV

conditions

mA

No.

15.

23313-002E*M

HHS Test
The following tests are applied to the X-ray generator.
The HSS test is necessary
only for ORH regulations.

15.1

Entrance

exposure

(1) Installed
(2) Measured
a> Manual

rate

system:
entrance

limit
1. Under
exposure

tube

2. Over

tube

3. C-arm

rate

mode

b) ABC mode
Brightness
Entrance

control:
exposure

limit

F.mA

limit

F.kV

maximum
rate:

position
(R/min)
(1 R/min = 2.58 x
lo-*C/(kg*min))
mA
kV

- 80 -

No.

15.2

Reproducibility

kV

and linearity

60 kV

without

automatic

exposure

TUBE

NO.

TUBE

TYPE

2J313-002E*M

control

60 kV

60 kV

60 kV

60 kV

mA

InA

mA

mA

mA

mA

sec.

Focus
Line

regulation

Exposure
No.

-+%

-+%

I
I

-+%

-+%

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

10

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

x
S

-+%

mR/mAs

mR/mAs

mR/mAs

mR/mAs

mR/mAs

Xl +X2
(1 mR = 2.58 x 10-7C/kg)
Note:

Make a sufficient
results.

number

of copies

- 81 -

of this

sheet

to record

test

No.

TUBE

NO.

TUBE

TYPE

2J313-002E*M

kV

90 kV

90 kV

90 kV

90 kV

90 kV

mA

mA

mA

mA

mA

mA

sec.

Focus
Line

regulation

Exposure
No.

j;-

+
-

+
-

-+

+
-

-+

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

10

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR/mAs

mR/mAs

mR/mAs

mR/mAs

mR/mAs

C
?i
I)il - x21
Xl + x2
(1 mR = 2.58 x IO-7C/kg)

- 82 -

No. 2J313-002EkM

TUBE NO.
TUBE

TYPE

kV

140 kV

140 kV

140 kV

140 kV

140 kV

mA

mA

mA

mA

mA

mA

sec.

Focus
Line

regulation

Exposure
No.

+
-

+
-

-+

+
-

+
-

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

10

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR/mAs

mR/mAs

mR/mAs

mR/mAs

mR/mAs

C
51

Xl +X2

(1 mR = 2.58 x lo-7C/kg)

- 83 -

No.

15.3

Automatic

exposure

Setting technique
Measured mAs
15.4

Automatic

exposure

(1) AEC Type:


Measured
(2) AEC Type:

Field

control

limit.

minumum

mA

50 kV

factors:

exposure

time

phototimer
time
Solidtimer

Measured

time

1
2
3
1+2
1+3
2+3
/1+2+3/

- 84 -

ms

2J313-002E*B

No.

APPENDIX

02

RELATIONSHIP

BETWEEN

FOCAL

SPOT

SIZE AND TUBE

2J313-002E9cB

CURRENT

RANGE

Small
X-ray

Large

-r

focus

tube
Focal

type

Al92

focus

(EIMK)

spot

The current

size

range

Focal
size

spot

0.6 x 0.6mm

20 to 320mA

1.2 x 1.2mm

Tube

current

range

400

to 1OOOmA

Tube
voltage
range

40 to 150
kV
1

85

No,

APPENDIX 03

2J313-002E*M

TEST METHOD OF FLUOROSCOPIC ENTRANCE EXPOSURE RATE

A.

Objective
Fluoroscopic
equipment
which
is provided
with automatic
exposure
rate
of tube potential
and current
control
is not operable
at any combination
rate
in excess
of 2.58 x lo-3C/
(kgemin)
which results
in an exposure
of the useful
beam enters
the
{lOR/min}
at the point
where the center
(See the section
1020.32
(d).)
patient.

B.

Test method
(1) The position
as follows.
a.

Testing

of

test

when

the

equipment

source

is

and

below

flip

diagnostic

table

system

Spot

film

Lead

sheet(300

device
x 300 x Zmm)

Aluminum attenuation
(200 x 200 x 19mm)

3oomm

assembly

table

Imaging

X-ray

block

4Dosimeter

P
Beam-limiting
X-ray

the

Set

2.

Place
the attenuation
block
(200 x 200 x 19mm) at the
approximate
center
of the table
and 150mm (6in)
above
tabletop
and conform
the center
of the X-ray
field
to
observing
the image.
imaging

to

3.

Set

4.

Place
field
image.

5.

Place
the lead sheet
X-ray
field
between
system.

the
and

the

system

horizontal

tube

1.

the

tabletop

device

to

probe
of the
1Omm (0.4in)

position.

300mm (12in)

dosimeter
above
the

at the
tabletop

above

86 -

tabletop.

center
of the
by observing

(300 x 300 x 2mm) at


the attenuation
block

the

the
and

the
it by

X-ray
the

center
of the
the imaging

is

2J313-002E*B

No.

b.

Testing

when

the

source

is

above

the

table

Beam-limiting

device

Aluminum attenuation
(200 x 200 x 19mm)

Spot

Lead

sheet

(300

x 300

film

device

Imaging

to

the

horizontal

sys tern

position.

Set

7
_.

If
of
be

3.

Place
the attenuation
block
(200 x 200 x 19mm) at the
center
of the table
and 150mm (6in)
above
approximate
the center
of the X-ray
field
to
tabletop
and conform
observing
the image.

5.

tabletop

x 2mm)

1.

4.

the

the source-to-image
the beam-limiting
placed
provided

block

distance
(SID)
is variable,
set the
as
close
to
the
tabletop
as it
device
shall
not
be
closer
than
30Omm.
it

probe
of the
30Omm (12in)

dosimeter
above the

at

the

center

Place
field
image.

the
and

Place
X-ray

(300 x 300 x 2mm) at


the lead
sheet
field
on the tabletop.

57 -

tabletop

by

the

of

the

the
it by

X-ray

observing

center

the

of

the

end
can

2J313-002E*M

No.

Testing
table)

c.

in

a C-arm

type

fluoroscope

(in

a Gyroscopic

diagnostic

Imaging
Spot

system

film

device

Lead sheet
(300 x 300 x Zmm)
Aluminum attenuation
(200 x 200 x 19mm)

300mm
850.5mm
(MIN. SID)

Beam-limiting
X-ray

1.

If necessary,
primary
beam.

2.

Place
the probe
of the
surface
of the imaging

3.

Place
the aluminum attenuation
block
(200
lead sheet
(200 x 200 x 2mm) at the center
shown in the above
figure.
open

shutter

the

tabletop

dosimeter
system.

(2)

Fully

(3)

Set the Rotary


DIP SW, SW5 on the
to the position
0.
This
setting
means that
the X-ray
entrance
exposure
rate
limit.

(4)

Push

(5)

Select
the
not select

(6)

Set the F.mA control


film
device.

(7)

Perform

(8)

Monitorig

(9)

When the dosimeter


(kgamin)
{9R/min},
record
it.

(10)

Repeat

the

the

position

blades

of

out

of

the

path

the

the

input

x 200 x 19mm) and the


of the X-ray
field
as

device.

DESK CONTROL PWB in


is

of

from

beam-limiting

generator

device

tube

300mm (12in)

the

block

the

excluded

control

from

desk

the

POWER ON switch.
spot-film
the ABC

radiography
mode on the
for

technique
spot
film

4mA and F.kV

on the
device.

control

for

control

panel,

5OkV on

the

spot

fluoroscopy.
increase

a dosimeter,

Steps

(6)

the

reading
reaches
read the setting

to

(9)

at

the

tube

to

125kV.

the limit
level
23.22
x lo-*C/
F.kV on the control
panel
and

F.mA setting

88 -

voltage

of

3,

2,

1 and

0.5mA.

and

No.

(11)

Set the limit


kV which
is
1 for each F.mA setting.
numbers are shown below.

F.mA

in Step (9)
according to the Table
settings and the settings switch

SW No. on the F.CONTROL


control box

setting

0.5

recorded
The F.mA

2J313-002E*M

DIG.

PWB

in the

SW1

mA
1mA

SW2

2mA

SW3

3mA

SW4

4m.A

SW5

(12)

Select the F.ABC and MA-AUTO mode switches.


Set the brightness control to the minimum position.

(13)

Perform

(14)

Monitoring

(15)

When the dosimeter reading reaches the limit level 23.22 x lo-4C/
(kgemin) {9R/min), read the indicated F.kV and F.mA on the control
panel and record them.

(16)

Set the limit kV recorded above according to Table 1. Set the limit
mA recorded above according
to Table 2. The setting switches are
shown below.

fluoroscopy.
a dosimeter,

(17)

Note:

limit

mA

limit

the brightness

SW No. on the F.CONTROL


control box

Limit

kV

increase

DIG. PWB

setting

in the

SW6
SW7

Set the Rotary DIP SW, SW5 on the DESK CONTROL


to the position 1.
When

control.

the DIP

switch,

the power

89 -

should

PWB

in the control

be off.

desk

No.

Table

Limit

kV setting

2J313-002E*B

code

(1; ON, 0; OFF)


F.kV

SW.
b

No.
5

F.kV

1.

I
50

I 51 lolo
52

10

l/1lOll:o

54

0,o

55

010

58

011/o
/

1 1

IO

61

1 ; 1

l/O

62

I 0

1 / 1 ( 1 / 1

Note:

When

l~o;O,o

1 0

0
0

74

75

76

77

l/O

/
O/l

93

Ojl

11

94

95

10

01

96

10

IO

jl

010
1

Oil

1 1

/ 0

III

llil/l/l
1010

011

l/O

1 0 /

/ 1 1 0

1o7

1 1

: 0

l/

108

109

O/l

110

0[1

i0j

011

99
100

1/00/0/1~0

011

011/011

ojo

11

111

112

O/l

j 0

1 1 1 0

1 0

1 1

10

10
1

11

ljOlO(

-?

l/l

1 1

o/oJo

80

1 / O/

81

10

1;O

82

011

83

01/0/L~0~0/1

84

l/O

85

011/o

86

87

01

l,r)iiil

1
l/O

ilk

Oil

11

115

O/l

OOll/l

116

101110/l
1

O/

o/1/+

O/l
/

10

0111

Ijo;1/0
1

90 -

on

11

11

119

01

120

01111

I/

121

Ojl

lql;olo~l

122

O/l

IjO

123

121,

125

oj

lllll

all

I
1 /

11

turn

O/l/O

j 0 / 1/ 1j 0 /
I

118

010/i

125kV,

1/1/o

I
1 I 1

10

011

/o

)0~1,1/0/0,0/0/1

198

1)

79

than

1
11

1 1

/O/l

01

,l
1

o/1
0

o/

73

kV is more

,oo

10

j 0

~]

im

011 !

97

1(1,1

72

1011

90

O/l

O/l

1/110!1,1/

I.0

91

l\l)Oil

71

89

,
IO
I

92
111

60

10

o/101
I

[ 69

I/

100

1
011

011

l/l

IO,

88

100

I59

0
1

ijOll/l/O

57

o/o

II

53

,
l!O/Ojl/lj

O/l

56

1I 0

switches.

0 j 0
1

10

0
.

No.

Table

F.mA

SW. No.
4
5

0.2

0.3

0.4

0.5

0.6

Limit

mA setting

code

SW 7

SW. No.
4
5

F*mA

2.6

2.7

2.8

2.9

0.7

3.1

0.8

0.9

_O

1.0

1.1

3.2

3.3

3.4

1 0

( 0

/ 11

3.5

1.2

LL

o(

1.3

1.4

1.5

1.6

1.7

1.8

1.9

2.0

2.1

2.2

2.3

2.4

2.5

1:
Note:

ON, 0: OFF
When mA is

more

than

4.OmA,

1 0 1 0 1 1 1 0 1 0

4.0

91 -

3.8

turn
-

0
0
I
1
/ 0 / 0 1 1 j

3.6

2J313-002E*K

on all

switches.

1 1 1 1

No.

2J313-002E*M

(18) After adjustment of the entrance exposure rate limit, repeat the Steps
(6) to (8) for the F.mA setting of 4, 3, 2, 1 and 0.5mA and record the
tube voltage and the dosime ter readings at the limit points in manual
mode
(19) Perform
current

fluoroscopy
in the FULL-AUTO mode and record
and dosimeter readings at the limit point.

the tube voltage,

(20) If the dosimeter reading exceeds the rejection


limit,
follows and then repeat the Steps (18) to (19).
(a) Manual mode
If the test results are not within the rejection
the limit kV which has been set in Step (11).

as

limit,

decrease

(c) FULL-AUTO mode


If the test results are not within the rejection limit,
the limit kV and mA which have been set in Step (16).

decrease

(b) kV-AUTO (F.ABC)


Same as above.

C.

proceed

Test

instruments

required

(1) Dosimeter
This measurement
a. Keithley
006A).

mode

requires

model

35020

one of the following

with

96020

b. Victreen model 550 with 550-3,


to TOSHIBA APPENDIX 026A.)
C.

Victreen
APPENDIX

(2) Aluminum

model 555 with


008A.j

attenuation

lmA,

chamber

550-4,

test

(Refer

550-5

instruments:

to TOSHIBA

or 550-6

1OmA or 1DAS chamber

APPENDIX

probe

(Refer

(Refer

to TOSHIBA

block

200 x 200 x 19 (mm>


(3) Lead

sheet

300 x 300 x 2 (mm>


D.

Rejection

limit

Maximum entrance dose limit: 23.22 x 10_4C/(kg=min)


{9R/min).
The maximum limit means a reading value on the dosimeter specified
C.

92 -

in item

No. 2J313-002E*B

APPENDIX
A.

04

TEST

METHOD

OF REPRODUCIBILITY

AND LINEARITY

Objective
are performed
A series of 10 radiation exposure measurements
the coefficient
of variation and the average exposure ratio.
Section 1020.31 (b) and cc>.>

B.
B-l

Test

to calculate
(See the

method

Testing

without

automatic

exposure

control

(1) Set the X-ray apparatus and test


following block diagram.

instrument

according

to the

s Dosimeter

Focus detector
distance:
1OOcm
Collimator

Primary
voltmeter

High-voltage
generator
1

(2) Turn
(3) Set
and
two
and

on the equipment

X-ray
control

r~:7.

Power
source

&

line switch

and calibrate

test

instruments.

the technique factors (radiographic


kV, mA and time selectors)
at intervals of one or
during the 10 consecutive measurements,
factors
are
adjusted
to alternate settings
minutes the technique
reset the test setting after each measurement.

For a detailed description of this step refer to APPENDIX 05


"Combination
of Specified Technique Factors" in this manual.
(4) While confirming
that the respective percent line voltage regulations
are within +l% of the mean value for all the 10 measurements,
record
the value of a dosimeter.
(5) The coefficient
of variation "C", 1020.30(b)(8),
of the 10 exposure
measurements
is calculated using the following equation;

where

Xi = the exposure measurement


x = mean value of the exposure measurements
S = estimated standard deviation

93 -

No.

(6)

Confirm

that

the

above

product

(7)

The average
ratios
of exposure
product
(mR/mAs),
(1020.31(c)),
calculated
using
the following

is

B-2

= average
mR/mAs values
tube current
settings.

Confirm

above

Testing

with

the

automatic

product

exposure

X- ray apparatus
block
diagram.

( 1) Set

the
following

than

0.045.

in

x lo-TC/kg)

where : xl and x2
consecutive
that

greater

to the indicated
milliampere-second
of the 10 exposure
measurements
equation;

(1 mR = 2.58

(8)

not

2J313-002E*M

is

not

obtained

greater

at

than

each

0.09

of

two

times.

control

and

instruments

test

according

to

the

Dosimeter

Beam
limiting

device
__--------_____

___------

-----_
Aluminum attenuation
block
(200 x 200 x 38mm)
SID

(i.e.

Source
Image
1ooomm

Distance)

Primary
voltmeter
Power

H.P.
Igenerator

(2)

Set the cassette


type)
posterior

(3)

Set the tube


film
density

(4)

Adjust

(5)

Adjust
the
the inside

(6)

Set

the

X-ray
control

the

with
film
(Sakura
to the detector
as

voltage
regulator

AEC so
beam
size

timer

of

for
for

that

the

limiting
of the
the

90kV,
0

A type)
and screen
shown in the above

the tube
posit ion

film

density

X-ray

control

94 -

to

current
for
lOOmA and
and make an exposure.
of

device
that
the
attenuation
block.

(Toshiba
figure.

0.85

to

exposure

0.3sec.

or

1.15
field

longer.

is

EM

the

got.

conforms

to

No.

(7)

Take

10

consecutive

percent

measurements

line-voltage
All

measurement.
within

-+l%

of

Measurement

/ X-ray
1

I(

conditions
The
The
au

(Tube

1 the

tube

I
1 The

maximum

voltage
tube

proper

time

current

to

1 The

proper

; p roper

tube

time

This
func

(1)

time
t ion.

can

to

be

Exposure

that

greater

than

by

equipment

to

b.
C.

1 0

Keithley

b.

position

pos it ion

time.

that

by

of

1 0

position

an

ON/OFF

for

equation

the

oscilloscope
SIGNAL

with

terminal

a
of

storage
each

of AEC.)
plural
detecting
fields,
test
For example,
test
should
field.

detecting
fields

get

2.

using

diagram
with

can

1,

the

detector

with

fields.

the
in

10 exposure
step
6 of

measurements,
Paragraph
B-l,

the

following

dosimeters.

is

not

0.045.

model

35020

one
with

of
96020

chamber

(Refer

to

TOSHIBA

APPENDIX

>
550-4,

550 with
550-3,
Victreen
model
to TOSHIBA APPENDIX 026A.j
Victreen

model

APPENDIX

008A.

555

with

lmA,

1OmA or

550-5

or

550-6

1DAS chamber

(Refer

probe

(Refer

to

TOSHIBA

>

measurement

ALCO primary
012A.

required

Electrical
a.

I positions

change

F:q:POSURE

Radiation
measurement
This
measurement
requires

006A.

got,

measured

the product
using
the

Density

time)

time
to 0.3

proper

condition

each

table.

change

current

3 detecting

following

level.

connection
the detector

taken

to

be

the

the

system
the

not

the

is

the
of

the

get

and

be

level.

proper

be

point

done

of

for

got,

The

measurements.

current,

step.)
not
be

can

should

Confirm
calculated

a.

(2)

all

according

proper

the

except

should
be

the

voltage

AEC.
(See
In the case

Test

done

Tube

this
can

that

Measuring

C.

be

current

the
tube

Note:

to

voltage

/ the

(8)

for

value

voltage,

tube

When

mean

should

(See
NOTE in
proper
time

the

The

regulation
values
for

within
a time
period
of
1 hour.
shall
be determined
for
each
line-voltage
regulation
shall
percent

minimum
tube
voltage
of
the system
tube
current
that
can get
the proper
omatic
exposure
control
time
of 0.1

1 sec.>.
When

12

the

ZJ313-002E$cB

voltmeter

model

PV-401

(Refer

to

TOSHIBA

APPENDIX

>

Tektronix
059B. >

storage

oscilloscope

model

95 -

464

(Refer

to

TOSHIBA

APPENDIX

No.

D.

Rejection

limit

(1)

Coefficient

(2)

Average

of
exposure

variation

c =
< 0.045

ratio

1x1 - x2/
Xl

+ x2

96 -

2 0.09

2J313402E*B

No.

APPENDIX 05
THE SETTING TECHNIQUE FACTOR FOR REPRODUCIBILITY
TEST WITHOUT A. E. C.
1.

2J313-002E+B

AND LINEARITY

General
which determines
how the combination
of technique
fundamental
idea,
d, is to select
a peak tube potential
factors
for
the test
are selecte
middle
and high kV) and the product
of tube current
and exposure
(low,
time (approximately
5 mAs and 30 mAs).

The

During
a series
of 10 radiation
controls
for
technique
factors
reset
to the test
setting
after
According

60kV
1OmA
3.0sec

to

the

following

I1

4th
exposure

all

variable
settings

exposures

are

13.0sec

:Alternate
setting(l))

(Test
setting)

Resetting
60kV
1OmA
3.0sec

/I

(Alternate
setting(2))

-/

10 radiation

procedure,

13.0sec

(Test
setting)

Setting
70kV
13mA
2.31sec

exposure
measurements,
are adjusted
to alternate
each measurement.

I-

_I

/
(Alternate
setting(l))

(Test
setting)

_ ____---

----

Test

Procedure

97

1OmA
/ 3.0sec
/
(Test
setting)

-*

and

measured.

No.

2.

Combination

No.
Test
Settings

of Specified

1
60kV
160mA
0.188sec

Technique

2J313-OOZE*K

Factors

60kV
20OmA
0.15sec

60kV

60kV

60kV

60kV

60kV

250mA
0.12sec

32OmA
0.094sec

400mA
0.075sec

500mA
0.06sec

640mA
0.047sec

Alternate
Settings
(1)

50kV
125mA
,0.24sec

50kV
160mA
0.188sec

5OkV
200mA
0.15sec

50kV
250mA
0.12sec

50kV
320mA
0.094sec

50kV
40OmA
0.075sec

50kV
500mA
0.06sec

Alternate
Settings
(2)

70kV
,200mA
0.15sec

70kV
250mA
0.12sec

70kV
320mA
0.094sec

70kV
400mA
0.075sec

70kV
500mA
0.06sec

70kV
640mA
0.047sec

70kV
640mA
0.047sec

No.

11

10

12

13

14

Test
Settings

90kV
1OmA
0.5sec

90kV
13mA
0.385sec

90kV
16mA
0.313sec

90kV
20mA
0.25sec

90kV
25mA
0.2sec

90kV
32mA
0.157sec

90kV
40mA
0.125sec

Alternate
Settings
(1)

80kV
1OmA
0.5sec

80kV
1OmA
0.5sec

80kV
0.385sec

80kV
16mA
0.313sec

80kV
20mA
0.25sec

80kV
25mA
0.2sec

80kV
32mA
0.157sec

Alternate
Settings
(2)

1OOkV
13mA
0.385sec

1OOkV
16mA
0.313sec

1OOkV
20mA
0.25sec

1OOkV
25mA
0.2sec

1OOkV
32mA
0.157sec

1OOkV
40mA
0.125sec

1OOkV
50mA
O.lsec

No.

15

13mA

18

17

16

19

Test
Settings

140kV
50mA
O.lsec

140kV
64mA
0.079sec

140kV
80mA
0.063ec

140kV
lOOmA
0.05sec

140kV
125mA
0.04sec

Alternate

130kV

130kV

130kV
64mA
0.079sec

130kV
80mA
0.063sec

130kV
lOOmA
0.05sec

145kV
lOOmA
0.05sec

145kV
125mA
0.04sec

145kV
160mA
0.032sec

(Note)

No. 6 and No. 7 should

be tested

98 -

only

for KXO-50F.

No.

APPENDIX

TEST

06

METHOD

OF AUTOMATIC

EXPOSURE

CONTROL

2J313-002E*B

LIMIT

A. Objective
To determine
that the product of X-ray tube current and exposure time is
limited to not more than 6OOmAs per exposure when an automatic exposure
control is provided.

B . Test

method

(1) Connect

High resistance voltage MA meter


divider (KV-201)
(MA-1201DC)
/

as follows.

Power
SUPPl

X-ray
tube
--&

\)

High-voltage
cables

KV indicating
unit
(KV-201)

Exposure
time
counter
(TD-3/-3A)
(2) Turn

on the main

(3) Press the


control.
(4) Turn
(5) Set

circuit

technique

on the automatic

switch

which

exposure

can use

control

the POWER

ON pushbutton.

the automatic

exposure

switch.

the exposure condition as follows.


: 50kV
R.kV
: more than GOOmAs
R.mA and time

(6) Remove the detector


X-ray field.
(7) Make
back

and press

breaker

for automatic

an exposure and record


up operates.

exposure

the mA reading

99 -

control

outside

and exposure

the

time when

No.

The operation
of over-mAs
lamp and the buzzer.
(8) Verify

C. Test

that

equipment

the product

: &CO
time

(3) High voltage


meter

D. Rejection

circuit

of indicated

is indicated

value

is not

over

by the BACK UP

540mAs.

required

(1) MA meter
(2) Exposure
counter

back-up

2J313-OOZE$cB

Model

MA-120lDC

: ALCO Model

TD-3/-3A

: ALCO Model

KV-201

(Refer
(Refer

(Refer

to TOSHIBA
to TOSHIBA

to TOSHIBA

APPENDIX
APPENDIX

APPENDIX

004B.)
018A.)

003A.)

limit

The product of tube current and exposure time measured


on the mA meter
the exposure
time counter must not exceeded the following
limits:
Maximum

product

of tube current

and exposure

100

time

limit

540mAs.

and

No.

APPENDIX 07
A.

2J313-002E*B

TEST METHOD OF AUTOMATIC EXPOSURE MINIMUM EXPOSURE TIME

Objective
When the X-ray
tube potential
is equal
to or greater
than 50kV, the minimum
time for
field
emission
equipment
rated
for pulsed
operation
shall
exposure
less
than
a
time
interval
equivalent
to
two
pulses,
and
the
be equal
to or
time
for
all
other
equipment
shall
be
equal
to
or
less
minimum exposure
than l/60
second
or a time interval
required
to deliver
SmAs, whichever
is
greater
(See
102.31(a)(3)(ii)).

B.

Test
(1)

method
Set

the

X- ray

apparatus

and

test

instruments

according

X-ray

X-ray

Beam limiting

device

Acrylic

tube

phantom

to

Fig.

2.

detector

(5cm)

Power
SUPPlY
I

High resistante voltage


divider
(KV-201)

H.P.
transformer

AEC selector

diagram

of

test

(2)

Turn

(3)

Set the
density

the
tube voltage
for 90kV,
regulator
for
0 position.

tube

(4)

Set

acrylic

: 5cm)

(5)

Perform

the

the

Block

phantom
and

an exposure

Confirm

that

the

exposure

Measured
time on the
than l/120
second.

method

switch.

(thickness
observe

the

time

is

within

time

exposure

101

counter

current

actual

In the case
of the detector
with multiple
be taken
to each combination
of detecting
should
be done under
7 conditions
(23 fields.
(6)

L
Fig.

on

Automatic
exposure
control

X-ray
control

in

for

lOOmA and

front

of

exposure

detector.

fields,
test
should
For example,
test
the

rejection

should

film

time.

detecting
fields.
1 = 7) for

the

the

the

detector

with

limits.

be equal

to

or

less

C.

Test

(2)

reqE:lrea

eqci?men:

Es?osure

tine

-+a
couLl,,r

: AK0

?lodel

ID-I/

-3A

(Eiefer

time

counter

to

TOSF-:I3A

_AI??EIiDIX

0!33.?

D. Rejection

limit

The exposure
time
following
limits.
Measured

measured

zirne

on the

5 l/120

exposure

second

102

must

no:

exceed

the

AP?ENDIX
A list

TEST

OS
of

test

TOQLS

AND

equipment

required
Test

Test

INSTRUMENTS

REQUIRED

is shown
for

equipment

below.
installation

equipment
Remarks

Item

Type
I
Full

AC Voltmeter
Oscilloscope
High

Voltage

Analyzer

Tektronix

464

Machlett
Dynalizer

II

Digital

V 0 M

Simpson

Primary

Voltmeter

ALCO
PV-401

Acrylic phantom
Water phantom

or equivalent

or equivalent

thickness

are necessary

300/6OOV

or equivalent

260

or

test equipment
The following
the above equipment.

range:

cnly

for

18Omm

HHS test

in addition

'Test equipment
Remarks
Item

Type
. Keithley model 35020 with
96020 chamber
. Victreen model 550 with
550-3, 550-4, 550-S or
550-6 probe.
. Victreen model 555 with
ImA, lOmA or 1DAS chamber

Dosemeter

Lead

300 x 300 x 2 (mm>

sh.eet

Al attenuation

200 x 200 x 19 (mm)

block
-

Al attenuat:on
Acrylic

Note:

These

phantom

are used

for

block

(Note)

thickness

(Note)

testing

200 x 200 x 38 (mm>

of Automatic

- 103 -

Exposure

5Omm

Control.

to

TOSHIBA

TOSHiBA

CORPORATlON

1385,SHIMOISHIGAMI,OTAWARA-SHI,TOCHiGl-KEN