78
METHODS
A total of forty-six consecutive critically ill patients were enrolled
from one medical and two surgical intensive care units over a period
of 6 mo. The protocol was approved by the local Ethics Committees
and written informed consent was obtained from a patients relative in
each case. Patients were eligible if already mechanically ventilated, sedated, and monitored with a pulmonary artery catheter. Those with
suspected or documented esophageal lesions (history of dysphagia or
previous esophageal surgery) or aortic dissection were excluded. Measurements did not interfere with medical or nursing procedures, and
the pattern of mechanical ventilation was not modified for the study.
V Ao ( t )dt
(1)
VOL 158
1998
(3)
Statistics
Correlation between techniques was determined using simple linear
regression analysis with Bland and Altman representation (20) to explore the possibility of systematic errors. Significance of biases was
tested using paired t test. In order to correct for the distribution of differences, data were logarithmically transformed to obtain new limits
of agreement, as recommended by Bland and Altman (20). In all patients who underwent more than one measurement of cardiac output
during the study, the variations between two time points in COTH and
COTED were compared using two-way analysis of variance (ANOVA)
for repeated measures. Correlation and agreement studies were also
performed on variations. The reproducibility of each technique was
assessed in six patients judged to be in stable hemodynamic condition.
Intra-observer variability was calculated as the standard deviation of
the differences between five measurements obtained over a short period of time divided by the mean cardiac output.
RESULTS
One hundred thirty-six paired cardiac output measurements
were made, using transesophageal Doppler and the thermodilution technique, in 46 mechanically ventilated, critically ill
patients recruited from three different ICUs. Table 1 shows
selected clinical characteristics, including information regarding inotropic support and positive end-expiratory pressure
(PEEP). Esophageal intubation with the Doppler probe was
easily accomplished in all patients, except three who required
a direct laryngoscopy. In all cases an adequate signal was obtained within a few minutes, regardless of the presence of a
gastric suction tube. The probe was left in place for up to 48 h
(median: 36 h, quartile range: 2742 h) without adverse effects.
TABLE 1
SELECTED CLINICAL CHARACTERISTICS OF PATIENTS
Patients, n
Age, yr
Male/female ratio
Main diagnosis, n
Focal pneumonia
Adult respiratory distress syndrome
Septic shock
Cardiac failure
Acute pancreatitis
Polytrauma
PEEP level . 10 cm of water, n
Catecholamine infusion, n
Dobutamine
Norepinephrine
Dopamine
Epinephrine
46
50 6 20
38:8
5
7
16
8
4
6
10
15
2
10
4
Cardiac output ranged from 1.7 to 14.8 L/min when measured using transesophageal Doppler, and from 1.4 to 15.3 L/
min using the thermodilution technique. Figure 1 demonstrates (1) the close correlation between COTED and COTH
(correlation coefficients 5 0.95; p , 0.0001), and (2) the agreement between the two techniques. The mean difference between
the paired values (COTH 2 COTED), representing the bias of
transesophageal Doppler with respect to thermodilution, was
0.24 L/min (p , 0.003), while the limits of agreement (bias 6
2 SD) were 2.04 and 21.56 L/min. The scatter of differences
tended to increase with the increase in mean cardiac output
(Figure 1B). We therefore performed a logarithmic transformation of these data and obtained a new bias of 0.03, and new
agreement limits of 0.29 and 20.23. The antilogs of these new
limits were 1.33 and 0.79, respectively, indicating that COTED
may differ from COTH within a range of 133% to 221%.
Twenty-three patients underwent more than one measurement of their cardiac output during the study, allowing us to
compare the variations between two consecutive cardiac output values as measured with the two methods in 88 instances.
The median delay between the two cardiac output determinations was 1.5 h (quartile range: 0.56.4 h). The two-way ANOVA
for repeated measures showed no statistical difference between
the variation measured by thermodilution and their paired
counterparts measured using transesophageal Doppler. The
79
Figure 2. (A) Scatterplot of paired variations in cardiac output obtained using thermodilution (DCOTH) and transesophageal Doppler (DCOTED), with the solid line representing linear regression.
Dotted lines at zero intercept delineate four quadrants: upper left
and lower right quadrants contain the points corresponding to opposite variations in COTH and COTED. (B) Bland and Altman (20) representation of the agreement between variations measured using
the two techniques. The solid line represents the mean difference
between DCOTH and DCOTED (systematic bias), and the dotted lines
define the limits of agreement (95% confidence interval).
80
VOL 158
1998
obtained using transesophageal Doppler (Figure 4A), although the agreement between the two techniques was not as
close as with the two previous methods (Figure 4B).
The results of linear regression and agreement analyses between the four techniques are presented in Table 2.
Variability in cardiac output measurement was 8% for
COTED, 12% for COTH, and 9% for COSSD. No reproducibility
study could be performed using indirect calorimetry due to
the prolonged equilibration time required for each cardiac
output determination.
Doppler. We conclude that transesophageal Doppler can provide a clinically useful estimate of cardiac output and reliably
detect its variations in mechanically ventilated, critically ill patients.
DISCUSSION
Transesophageal Doppler is a new minimally invasive method
for the continuous monitoring of cardiac output in critically ill
patients. In this multicenter trial, the cardiac output values obtained using this technique correlated well with those obtained using thermodilution. The systematic underestimation
of COTED with respect to COTH was small (, 0.25 L/min), and
the limits of agreement were 0.25 6 1.8 L/min. More importantly, variations in cardiac output with time were comparable
using both techniques. Other measures of cardiac output using
suprasternal Doppler or the Fick principle also demonstrated
excellent correlation and good agreement with transesophageal
Methodologic Considerations
The estimation of cardiac output using transesophageal Doppler is based on four assumptions: (1) an accurate measurement of the velocity of descending aortic blood flow; (2) a
flat velocity profile throughout the aorta; (3) an estimated
aortic cross-sectional area close to the mean value during systole; and (4) a constant division of blood flow between the descending aorta (70%) and the brachiocephalic and coronary
arteries (30%). The accuracy of velocity measurement requires a good alignment between the Doppler beam and
blood flow and knowledge of the angle at which the blood
flow is insonated. Alignment is best assessed subjectively by
optimizing the quality of the obtained signal with the aid of
the visual display of instantaneous velocity waveform and the
Doppler sound. The angle between the Doppler beam and
blood flow is roughly the same as that between the transducer
and the probe (458), since the esophagus and aorta are parallel
in the thorax. However, even a 10% change in the alignment
of the Doppler beam with respect to blood flow would result
81
No. of
Patients
46
17
13
17
13
13
No. of
Measurements r Value
136
53
46
53
46
46
0.95
0.94
0.89
0.98
0.91
0.91
Regression Line
Equation
p Value
(Correlation)
Bias
(L/min)
p Value
(Bias)
y 5 0.53 1 0.903
y 5 20.11 1 0.993
y 5 0 1 1.043
y 5 0.35 1 0.953
y 5 1.09 1 0.783
y 5 0.9 1 0.803
, 1024
, 1024
, 1024
, 1024
, 1024
, 1024
0.24
0.16
20.29
0
0.51
0.56
, 0.003
NS
NS
NS
, 0.003
, 1023
Limits of Agreement
(L/min)
2.04, 21.56
1.90, 21.58
2.16, 22.74
0.88, 20.88
62.70, 21.68
62.67, 21.55
Definition of abbreviations: TH 5 Thermodilution; TED 5 Transesophageal Doppler; SSD 5 Suprasternal Doppler; IC 5 indirect calorimetry (Fick principle).
82
Transesophageal Doppler is a simple technique, and most users acknowledge that it is fairly easy to achieve adequate probe
positioning and obtain reproducible results (9, 29). Freund (8)
noted a dramatic improvement in the skills of untrained operators after performing only 10 measurements. However, in the
present study only trained operators performed each type of
measurement. Interobserver variability has previously been
shown to be less than 10% using transesophageal Doppler (6).
We found intra-observer variability to be slightly less with transesophageal Doppler (8%) than with thermodilution (12%),
confirming previous observations (7, 9).
Potential side effects of transesophageal cardiac output
monitoring include esophageal damage. Some authors have
left the probe in place for over 2 wk without adverse effects
(29), but there has been no systematic fibroscopic evaluation
of esophageal mucosal injury at the time of probe removal.
A major advantage of transesophageal Doppler over other
currently available techniques is its ability to provide continuous, real-time monitoring. In contrast to continuous cardiac
output monitoring using a pulmonary artery catheter, which
only provides repeated measurements of mean cardiac output,
transesophageal Doppler displays the instantaneous aortic velocity spectrum. Probe displacement can occur during prolonged monitoring as a result of various causes (e.g., nursing
procedures, deglutition, gravity) and results in a poorly defined velocity envelope. It is mandatory to ensure an adequate
signal prior to interpreting Doppler-derived data. Failure to
reposition the probe prior to each measurement may lead to
grossly erroneous cardiac output values and poor agreement
with other techniques (23). The monitor can be set to calculate
cardiac output either on a beat-to-beat basis or after averaging
stroke volume over a number of consecutive beats (maximum
20), according to the operators choice. In addition, the device
derives parameters from the shape of the velocity envelope,
such as peak velocity, mean acceleration of blood flow, and
the rate-corrected flow time during systole. Some insight into
volemic status, peripheral resistance, and left ventricular function can be obtained from these parameters in the ascending
aorta (30, 31). Although additional studies are needed to establish the validity of these indices when obtained in the descending thoracic aorta using transesophageal Doppler, several reports support their usefulness for optimizing therapeutic
strategies in critically ill patients (3234). In a recent randomized trial, Sinclair and colleagues (35) showed that patients
with femoral neck fracture undergoing intravascular colloid
resuscitation guided by transesophageal Doppler during surgical repair had a significantly reduced hospital stay (39%) compared with control subjects (35).
In conclusion, although transesophageal Doppler cannot
replace the pulmonary artery catheter, it offers an instantaneous, reasonably accurate, and noninvasive estimate of cardiac output. In addition, the variations in cardiac output occurring spontaneously or in response to treatment are tracked
similarly by the two techniques. Transesophageal Doppler
VOL 158
1998
83
31.
32.
33.
34.
35.